Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

Ask 800

Download as pdf or txt
Download as pdf or txt
You are on page 1of 10

Contrans P ASK 800

HART Transmitter with Remote Seal for Level and


Differential Pressure
Span: 10 mbar to 40 bar
(1 kPa to 4,000 kPa)

10/15-6.35 EN

(20857.bmp)

n Isolation between Process and Transmitter with Measuring The transmitter ASK 800 combines proven, mature technology wi-
Media having: • High Temperatures and Viscosities •Corro- th a trend-setting microprocessor controlled electronic.
sive Ingredients, a Tendency to Polymerization It is based on a proven, highly-stable measuring mechanism, on
n Prevention of Deposits in the Process Connection Flanges by which a remote seal with flange is fitted directly onto on side. The
internal sensing diaphragm is slightly deflected corresponding to
Extended Remote Seals
n Mounting to various DIN or ANSI Flanges
the differential pressure present between the remote seal and the

n
equalizing side and converted into an electrical signal by the elec-
Microprocessor technology tronics. Apart from a high accuracy with an almost complete com-
n High turn down ratio 30 : 1 pensation of ambient influences, both measuring mechanism and
n Digital communication using the "HART"-Protocol without
electronic provide a substantial increase in functionality.The digital
indicator can be so configured that the measured value can be il-
disturbing the output signal 4...20 mA
n Lower range value, span and damping are externally adjust-
lustrated as a physical unit, percentage value or current. In additi-
on the measuring mechanism temperature can be indicated. The
able process wetted parts of the remote seal can be selected from dif-
n Transfer response configurable: ferent materials such as stainless steel, Hastelloy C, Monel or tan-
• linear • freely programmable • PID-controller talum and others, depending upon the required resistance to
n "Intrinsic Safety" or "Flameproof Enclosure" explosion protec-
corrosion.
Various filling liquids, for example for the food and beverage indu-
tion or mounting on "Zone 0"
n
stry, complete the spectrum of applications. The Transmitter ASK
E.M.C. according to IEC 801 800 is thus an efficient, intelligent element, a must for every type of
automation.
Data Sheet
Transmitter ASK 800 with Remote Seal
for Level and Differential Pressure 15-6.35 EN

Technical Data

Measuring Mechanism Max. permissible voltage ripple of the power supply during the
communication:
Measuring limits
• 7 Vpp at 50 Hz ≤ f ≤ 100 Hz
-100% and +100% of the max. span
• 1 Vpp at 100 Hz < f ≤ 200 Hz
Lower range value
• 0.2 Vpp at 200 Hz < f ≤ 300 Hz
Continuously adjustable between the measuring limits
The following can, for example, be configured:
Span
Digital indication of process pressure, temperature, per-
Arbitrary - continuously adjustable within the range limits.
centage value, current, arbitrary process variables
The adjusted span must not to be lower than the minimum
Damping
span. Smallest span: 10 mbar.
Transfer function
2...60 mbar (0.2...6 kPa)
Alarm signalling Imin/Imax
13.34...400 mbar (1.334...40 kPa)
PID Controller; adjustment with five parameters;
0.0834...2.5 bar (8.34...250 kPa)
- Setpoint w
0.534...16 bar (53.4...1,600 kPa)
- Proportional gain Kp
1.334...40 bar (133.4...4,000 kPa)
- Reset time Tn
Characteristic
- Derivative gain KD
Linear, rising or falling Freely programmable
- Derivative action time TD
Over-ranging limit
Write-over of parameters can be prevented by the interlock
Each side may be over-loaded up to 100% of the nominal-
switch. All stored data is secured even in the event of power
pressure
supply failure.
Measuring chamber volume (equalizing side)
A detailed description of all functions can be found in the Instruc-
approx. 14 cm3
tions 42/15-910 EN.
Isolating diaphragm (equalizing side)
1.4404 (≡316 L stainless steel) General and Safety Data
Blanking plug Ambient Conditions
1.4571 (≡316 Ti stainless steel)
Ambient temperature range
Power Supply -40°C...+80°C
Transmitter operating voltage Storage and transport temperature range
UB,max= DC 45 V UB,min= DC 11.5 V -50°C...+80°C
Humidity
Output signal ≤95%, annual mean, condensation permissible
4...20 mA
Output signal limits Amplifier Enclosure, Mounting, Weight
Imin 3.5 mA, Imax 22.5 mA Material
Standard adjustment: Imin 3.8 mA and Imax 20.5 mA Die cast copperfree aluminium (GD-AlSi)/polycarbonate
Load R Protective varnish
U 11.5V US Supply voltage Epoxy resin, gravel-grey, RAL 7032
R ≤ S kΩ Degree of Protection
I Imax 20...22.5mA (adjustable)
max
IP 65 (jet waterproof) according to EN 60 529
Digital Communication (≡ NEMA standard type 4)
Standardized communication protocol "HART" 5.1 Type of Mounting / Sealing
Minimum load for communication: 250 Ω Flanged to side of container
Recommended lead-length for communication: Seal type: non metallic
• screened multi-wire: max. 1.5 km Weight
• screened two-wire: max. 3 km Approx. 12 kg
Functional Diagram ASK 800 with Remote Seal
Measuring capsule Overload diaphragm
Remote seal ( HP) Sensing diaphragm
MICROPROCESSOR-
based electronics

4...20mA
DC 11.5...45V

TEST
INSTRUMENT

Equalizing side ( LP) Sensor body


M01060E1.EPS

Page 2 of 10 08.2000
Data Sheet
Transmitter ASK 800 with Remote Seal
for Level and Differential Pressure 15-6.35 EN

Technical Data

Application limits: permissible temperature/minimum pressure


The pressure has to be linearly interpolated between the stated temperatures
Filling-Liquid Silicone Oil Carbon Fluoride Distilled Water Vegetable Oil Vac.-proof Design
Identification IB L D PF IC
Density at 20 °C (kg/m3) 924 1880 998 920 1055
Operating Temperature (°C) -30...+250 -30...+150 +5...+70 -10...+250 -30...+200
Static Pressure (mbar abs.)
at 20 °C > 500 > 1000 > 1000 > 500 > 5
100 °C > 500 > 1000 --- > 1000 > 25
150 °C > 500 > 1000 --- > 1000 > 38
200 °C > 750 --- --- > 1000 > 50
250 °C > 1000 --- --- > 1000 ---

Transient Response at reference conditions Explosion protection, Certification


(see Instructions for further data)(1.4404 isolating diaphragm material, silicone for Zone 0
oil filling liquid).All values are limit values and refers to the output span. The
effects identified with a * are with reference to the measuring range and are to Definition of "Zone 0":
be multiplied by the turn-down factor. The ratio range/calibrated span should "Zone 0" includes all areas in which, permanently or for long periods of
be as small as possible. time, a dangerously explosive atmosphere exists.
Conformity1) including hysteresis and dead band Type Approval
terminal-based 0.1 % according to VbF "Zone 0". Gas group IIC and for flammable liquids of the
Hysteresis 0.05 % hazard classes AI, AII or B.
Power supply Instrument Mounting
voltage effect per volt 0.005 % The measuring mechanism can be directly connected onto "Zone 0"
Type Approval Certificate: PTB No. III B/S-1949
The following specifications are valid for a linear transfer function. The effect
PTB No. Ex-93.C.4004 (EEx ia IIC T6)
appearing at the output with non linear transfer functions is dependent on the
Type Approval: 08/PTB No. III B/S-1949
function and is to be calculated accordingly
for Zone 1
Static pressure effect per 100 bar When the transmitter is installed in "Zone 1", the type of protection "Intrin-
Measuring ranges ≤ 2.5 bar sic Safety EEx ia" is sufficient.
on zero 0.2 % Explosion protection Intrinsic Safety "i"
on span 0.1 % Identification Code (DIN EN 50 014): EEx ia IIC T6
Measuring ranges ≥ 16 bar Type Approval Certificate: PTB No. Ex-93.C.4004
on zero 0.5 % Connected to an intrinsically safe current circuit with the highest values:
on span 0.75 %
Ambient temperature effect acc. to DIN 16 086 Tempera- Umax Imax Pmax Max. ambient
Thermal change (-20°C...+60°C) ture class temperature
2)
on zero 0.1 % +0.6 mbar/10K
on span 0.1 % +0.1%/10K 2) T6 45 V 130 mA 0.5 W 40 °C
Temperature coefficient (-40°C...+80°C) 3) T5 45 V 130 mA 0.8 W 40 °C
on zero 0.05 %/10K +0.6 mbar/10K2)
on span 0.15 %/10K 2) T4 45 V 130 mA 1.0 W 80 °C
Long-term drift within 6 months 0.05 % T4 45 V 130 mA 1.2 W 60 °C
Rise time acc. to DIN 16 086
internal capacitance Ceq ≤ 0.010 µF, internal inductance is negligably small
dependant on range and turn-down factor 0.15...1.6s
Type of protection: Flameproof Enclosure "d"
additional adjustable time constant 0...60s
1)
Identification Code (DIN EN 50 014):EEx d IIC T6
Additionally with turn-down factor >1:10
Measuring range Type Approval Certificate: PTB No. Ex-91.C.1077
± (0.005 × 0.05) %
adjusted span Operating conditions: max. ambient temperature 75°C
2)
Measuring mechanism and remote seal with same temperature BASEEFA Type N Approval
3)
With carbon fluoride filling liquid -20°C...+80°C Ex N IIC T5 (Tamb: - 40 °C to + 80 °C)
Electrical Connections Functional Diagram-Communication
Cable gland or One Han 8U plug (option)
female thread (standard)
+ +
p
UB US
+# – –
+ IA + FSK
AS 800 Power supply
TEST

Contrans P
~181

FSK-Modem
~76

Screw terminals for


0.5...2.5 mm 2- wires Plug connection diagram
34
IBIS
Socket-insert RS 232C
1 2 (view on
sockets) M01058E2.EPS
M01063E1.EPS +

08.2000 Page 3 of 10
Data Sheet
Transmitter ASK 800 with Remote Seal
for Level and Differential Pressure 15-6.35 EN

Ordering Data

To order, the order number and the data of the adjustable values: lower and upper range values, will suffice.

Transmitter ASK 800 Order No. 1 5 9 . . - 2 -

Nominal pressure

PN 100 (10 MPa) 316 Ti St.St. (1.4571) . . . . . . . . . . . . . . . 43


Measuring range 1 5 9 9 0 -
Order No. 2 -
0 . . . 60 mbar (6 kPa)  ........................... ... . . . . . . . . . 10
0 . . . 400 mbar (40 kPa)  adjusted from ...to...mbar/kPa1) . . . . ... . . . . . . . . . 14
0 . . . 2.5 bar (250 kPa)  ........................... ... . . . . . . . . . 18
Measuring range Order No. 1 5 9 6 3 - 2 -
0 . . . 16 bar (1,600 kPa) adjusted from ...to...bar/kPa1) . . . . . . . . . . . . . . . . . 36
0 . . . 40 bar (4,000 kPa) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37

Isolating diaphragm material


316 St.St. (1.4404) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 050

Flange-O-rings
Buna . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 060
Viton (restricted operating temperature -18°C...+80°C) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 160
PTFE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 161

Process connection (equalizing side)


Flange connection and 1/4-18 NPT top and bottom . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 064
Process connection acc. to DIN 19213 and 1/4-18 NPT top and bottom (for PN100,316Ti St.St.2))191

Filling liquid
Silicone oil . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 070
Carbon fluoride (restricted operating temperature -20°C...+80°C) . . . . . . . . . . . . . . . . . . . . . . 137

Output signal
4 . . . 20 mA, linear . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 023
4 . . . 20 mA, Characteristic acc. to table of values, max. 22 pairs: Input(%);Output(%) . . . . 221

Electrical connection
One Pg 13.5 cable gland . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 044
Two Pg 13,5 cable glands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 268
Two 1/2-14 NPT female threads . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 270
One plug connector (Han 8U) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 272
Order No. 1 5 9 . . - 2 - / / / ...

If necessary, Code Nos. are to be suffixed to the Order No.

1)
Possible units: mbar, bar, Pa, kPa, mmH2O, mmHg, psi, g/cm2, kg/cm2, inches H2O, inches Hg, feet H2O.
2)
Other variations upon request.

Page 4 of 10 08.2000
Data Sheet
Transmitter ASK 800 with Remote Seal
for Level and Differential Pressure 15-6.35 EN

Additional Ordering Data for Instrument Options (Code Nos.)

Code Nos. can be continuously suffixed to the Order No.

Instrument Options Code No.

Ex-design: EEx ia IIC T6 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 557


EEx d IIC T6 (only when combined with Order No. 270) . . . . . . . . . . . . . . . . . . . . . 558
for "Zone 0" (only combined with Code No. 557) . . . . . . . . . . . . . . . . . . . . . . . . . . 264
BASEEFA Type N Approval, Ex N IIC T5 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 274

Analogue indicating instrument (Scale markings: 0...100 % linear) Normal or Ex "i" design . . . . . . . . . 206
Digital indicating instrument: Normal or Ex "i" design:
(indicates %, however with Code No. 415 process pressure) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 244
Digital indicating instrument: Ex "d" design:
(indicates %, however with Code No. 415 process pressure) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 245
Scale indication... (please define value: e.g. 0...15 m1); 4...20 mA proportional; only combinable
with Code No. 244/245, not with Code No. 415) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 416

PID Controller,Parameter: nominal value [%]; Kp (0...40); Tn [s] (0.1...3600); KD (0...40); TD [s] (1...3600)2) 415
Alarm signalling with 3.6 mA (Standard: 21 mA) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 417
Max. output current... (please state value 20...22.5 mA; Standard: 20.5 mA) . . . . . . . . . . . . . . . . . . . 418
Min. output current... (please state value 3.5...4 mA; Standard: 3.8 mA) . . . . . . . . . . . . . . . . . . . . . . . 420
Electrical damping TG=... (please state value to be adjusted: 0...60 s) . . . . . . . . . . . . . . . . . . . . . . . . 259

Tagging (on Type plate, max. 32 characters) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 205


Tagging on Tie-on plate (stainless steel, max. 32 characters) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 202

Flange screws - stainless steel (not with carbon steel flanges) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 193

Documentation
Instructions (state total number of copies)3)
German (42/15-992, no details required for one copy) . . . . . . . . . . . . . . . . . . . . . . . . . (...copies) Z2D
English (42/15-992 EN, Order-No. always required) . . . . . . . . . . . . . . . . . . . . . . . . . . . (...copies) Z2E
French (42/15-992 FR, Order-No. always required) . . . . . . . . . . . . . . . . . . . . . . . . . . . (...copies) Z2F

Digital Communication: User Interface "IBIS" (Data Sheet 15-6.96 EN), FSK-Modem (Data Sheet 15-6.97 EN)

Mounting Accessories see Data Sheet 15-8.98 EN (e.g. fixing screws, O ring seals)
Certificates see Data Sheet 15-6.99 EN

1)
Possible units: pressure units, as listed under "adjusted from... to..."; m, cm, mm, inches, feet, m3, l, barrels, gallons, yard3,
inches3, feet3, g, kg, t, pounds, short tons, long tons (further units see Technical Information: B 92 M 508 V)
2)
To optimise the process it is necessary to have the additional facilities available with IBIS and a modem. When no values
stipulated-standard values will be programmed i.e. setpoint = 50% and Kp = 1. Tn, KD and TD are switched off (0).
3)
One copy without charge

Scope of Supply
1 Instructions Supplied against special order:
1 Blanking plug Connecting pipe unions, flanges, valves, manifolds, remote seals,
1 Instrument socket with plug connector Power supply e.g. TZN 128 (Data sheet 18-8.39 EN),
Spare Parts Transmitter ASK.

08.2000 Page 5 of 10
Data Sheet
Flush Diaphragm Remote Seals DN 80 / ANSI 3"
15-6.35 EN

Ordering Data

Relevant Code Nos. can be continuously suffixed to the transmitter Order No.

Flush diaphragm remote seals DN 80 / ANSI 3" Code No.

Mounting
Direct mounting (on ASK 800 without capillary tube) . . . . . . . . . . . . . . . . . . 729

Material: Diaphragm and Body


sealing surface
1.4404(316 L st.st.) 1.4404 . . . . . . . . . . . . . . . . . . . . . . . . . . 605
Hastelloy C 1.4006 (410 st.st.) . . . . . . . . . . . . . . . . . . . . . . 606
Monel 1.4404 . . . . . . . . . . . . . . . . . . . . . . . . . . 607
Tantalum 1.4006 . . . . . . . . . . . . . . . . . . . . . . . . . . 608

Sealing surface1): Form E (DIN 2526) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 062


Form V 13 (DIN 2513) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 643
Form N (DIN 2512)(only with Code No. 605) . . . . . . . . . . . . . . . . . . . . . . . . 644

Form RF (ANSI B 16.5)


150 psi . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 604
300 psi up to 1500 psi . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 724

Form RJF (ANSI B 16.5)(only with Code No. 605)


300 up to 600 psi . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 646
1500 psi . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 645

with blank flange (DIN 2527) stainless steel 1)


Pressure rating PN 16 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 786
PN 40 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 787
PN 64 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 788
PN 100 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 789

with blank flange (ANSI B 16.5 DN 3"-RF) stainless steel 1)


Pressure rating 150 psi . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 790
300 psi . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 791
600 psi . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 792
1500 psi . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 793

Filling liquid
Silicone Oil . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 074
Carbon Fluoride . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 687
Distilled Water . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 688
Vegetable Oil . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 673
Vacuum-Proof Design . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 739

Additional Mandatory Code Nos.

with FEP coating (medium temperature ≤ 150 °C, only with Code Nos. 605/606) . . . . . . . . . . . . . 662
Ex Design for "Zone 0" . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 689

(Add the remote seal Code Nos. ... / / / ...


Order No. to the transmitter Order No.) 7 2 9

1)
The ordering data for DIN und ANSI cannot be combined with each other.

Page 6 of 10 08.2000
Data Sheet
Extended Diaphragm Remote Seals DN 80 / ANSI 3"
15-6.35 EN

Ordering Data

Relevant Code Nos. can be continuously suffixed to the transmitter Order No.

Extended diaphragm remote seals DN 80 / ANSI 3" Code No.

Mounting
Direct mounting (on ASK 800 without capillary tube) . . . . . . . . . . . . . . . . . . 729

Material: 1.4404≡316 L st.st.; 1.4571≡316 Ti st.st.


Extension and sealing surface
Diaphragm Form E, V13 Form N Form RJF
and RF Ext./seal.surf.
1.4404 1.4571/1.4404 1.4571/1.4404 1.4571/1.4404 . . . . . . . . . . . 640
Hastelloy C Hastelloy C - - . . . . . . . . . . . . . . . . 641
Hastelloy C 1.4571 - - . . . . . . . . . . . . . . . . 642

Extension length: 50 mm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 647


100 mm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 648
150 mm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 649

Sealing surface1): Form E (DIN 2526) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 062


Form V 13 (DIN 2513) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 643
Form N (DIN 2512)(only with Code No. 640) . . . . . . . . . . . . . . . . . . . . . . . . 644

Form RF (ANSI B 16.5)


150 psi . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 604
300 psi up to 1500 psi . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 724

Form RJF (ANSI B 16.5)(only with Code No. 640)


300 up to 600 psi . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 646
1500 psi . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 645

with blank flange (DIN 2527) stainless steel 1)


Pressure rating PN 16 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 786
PN 40 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 787
PN 64 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 788
PN 100 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 789

with blank flange (ANSI B 16.5 DN 3"-RF) stainless steel 1)


Pressure rating 150 psi . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 790
300 psi . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 791
600 psi . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 792
1500 psi . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 793

Filling liquid
Silicone Oil . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 074
Carbon Fluoride . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 687
Distilled Water . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 688
Vegetable Oil . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 673
Vacuum-Proof Design . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 739

Additional Mandatory Code Nos.

Ex Design for "Zone 0" . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 689

(Add the remote seal Code Nos. ... / / / ...


Order No. to the transmitter Order No.) 7 2 9

1)
The ordering data for DIN und ANSI cannot be combined with each other.

08.2000 Page 7 of 10
Data Sheet
Transmitter ASK 800 with Remote Seal
for Level and Differential Pressure 15-6.35 EN

Dimensional Diagrams

Errors and omissions excepted. All dimensions in millimeters (mm). Seals and fixing materials not supplied.
a b ~294

80
10
79 2)

DIN

ød3
ød2

ød4

105
øK
øD
Contrans
ANSI

~28
108
~161 1)

34
4 6 12 13
Design with
extended diaphragm Form V13 Sealing Surface
Ext. length
50/100/150
mm c b
a
90

ø 138

ø 120
7
1
ød3
ød1

ød4
øK
øD
41.3

3
~200 1)

M 0 10 6 4 E 1.E PS
5

9 12 13

ø 82 Remote Seal Dimensions


11 8 Sealing surface a c ∅d1
3)
∅ d2 ∅ d3
Form E (DIN 2526)
19 17 76.1 78 138
Form RF (ANSI B 16.5)
Form V 13 (DIN 2513)
DN 80 Form N (DIN 2512) 19 17 76.1 78 138
or
ANSI 3" Ring Joint Form RJF 22 22 76.1 78 146
1 1/4-18 NPT female thread. (300/600psi)(ANSI B 16.5)
2 Thread for fixing screws: 7/16 UNF, 16 mm deep. Ring Joint Form RJF 22 22 76.1 78 168
Minimum screw-in length is 12mm. (1500psi)(ANSI B 16.5)
Whith flange acc. to DIN 19 213: M10, Minimum screw-in
Flange Connection Dimensions
length acc. to DIN 19 213.
3 60 x 32mm oval flange. Nominal Screws *)
4 Electrical connections: Pg 13.5 cable gland bottom or ∅K ∅ D ∅ d4
pressure Number Thread b
two glands top and bottom or 1/2-14 NPT femal threads on
top and bottom or plug Han 8U bottom. PN 16/40 160 200 138 8 M 16 20/24
5 Type plate. DN 80 PN 64 170 215 138 8 M 20 28
6 Locking excenter (only with EEx "d"-version).
7 Measuring mechanism plate. PN 100 180 230 138 8 M 24 32
8 Enclosure cover (electr. connection/analogue instrument). 300 psi 168.3 209.5 127 8 M 20 28.6
9 Enclosure cover (connection for digital instrument). ANSI 3" 600 psi 168.3 209.5 127 8 M 20 31.8
10 Tie-on plate e.g. for tag number (optional).
11 Sealing plug. 1500 psi 203 266.7 127 8 M 24 47.6
12 Mating flange acc. to DIN/ANSI (not supplied). *) not supplied
13 Flat seal (not supplied). 1) with dig. indic. instrument +29 mm 2) with analogue indic. instr. +24 mm 3) +/-1.5%

Page 8 of 10 08.2000
08.2000 Page 9 of 10
ABB Automation Products GmbH Subject to technical changes.
Schillerstraße 72 Printed in the Fed. Rep. of Germany
10/15-6.35 EN 08.2000
D-32425 Minden
Tel. +49(0)571 830-0
Fax +49(0)571 830-1846
http://www.abb.de/automation

You might also like