Model CFA3047 Product Data Sheet

PDS 71-Phosphate3047/rev.C January 2014

On-Line Phosphate Analyzer

 One Button Operation. One button starts, cali-


brates, and puts the analyzer on-line.
 Multiple Sample Option. Measure as many as six
sample streams sequentially.
 Routine Maintenance takes five minutes every
three months.
 Self-Diagnostics. The analyzer monitors critical
functions and alerts the user when a problem
exists.
 Pre-Packages Reagents and Standards. No messy
chemical solutions to prepare and standardize.
Reagents and standards come from the factory
Analyzer shown with
ready-to-use. front covers removed

Applications
The Model CFA3047 Phosphate analyzer is intended for the Using an optional built-in stream selector, the Model CFA3047
determination of orthophosphate in boiler water. can measure as many as six sample streams sequentially. A sin-
gle analog output is standard. A stream identification marker
Features and Benefits indicates the start of each new sample. A multiple output option
in which a separate analog output is assigned to each sample
Simplicity, ease-of-use, and reliable operation characterize the
Model CFA3047 Valve-Pump phosphate analyzer. Place color- stream is also available. With the multiple output option each
coded reagent straws in color-coded reagent bottles, place the output holds the last value until updated.
blank and standard straws in the appropriate bottles, and press The Model CFA3047 Phosphate Analyzer has a pumped reagent
a button. The analyzer automatically calibrates and puts itself in delivery system. Instrument air, needed for pressurized systems,
service. Once in service, the analyzer runs a blank and full scale is unnecessary. The Model 3047 uses a stable, pre-mixed reduc-
standard twice a week. ing agent. There is no need to mix powders and liquids before
Routine maintenance is simple. Replace reagents and standards putting the reagent in service.
every three months. Replace the valve pump, the key to the reliable The Model CFA3047 uses a batch measurement technique. The
performance and low maintenance of the Model CFA3047, every update rate is approximately every 15 minutes.
two years. Total maintenance time is at most one hour per year.
Model CFA3047 January 2014

Principles of Operation
The heart of the CFA3047 Phosphate analyzer is the valve pump.
The valve pump has separate pumping elements for sample,
reagents, and standards. Each pump element consists of a chem-
ically inert flexible membrane in a cavity. Cavity volumes are 5 mL
for sample and standards and 0.2 mL for reagents. A vacuum/
pressure pump pushes and pulls on the membrane. Vacuum
causes the chamber to fill. Pressure pushes the liquid out of the
cavity into the reaction chamber. Check valves prevent backflow.
A stream of air pumped into the reaction chamber mixes the
sample and reagents. After the reaction time has passed, the
treated sample is pumped to the measurement cell. Light at 660 nm
(470 nm for 0-50 ppm) from an LED passes through the sample
and a photodiode measures the light intensity. The analyzer
automatically calculates the absorbance and translates the
results into a phosphate concentration reading. Once the meas-
urement is complete, repeated flushes with fresh sample remove
the treated sample from the measurement and reaction cells.

Single output (standard): Output corresponds to sample


Specifications currently being measured. A stream identification marker
Ranges: 0-3 ppm, 0-5 ppm, 0-10 ppm, 0-20 ppm, 0-25 ppm, on a chart recorder indicates the start of each new sample.
and 0-50 ppm, and 0-100 ppm (as PO4). Multiple output (optional): Each sample has its own out-
Method: Phosphate is determined colorimetrically as phospho- put. The output holds last value when the analyzer changes
molybdous acid or as phosphovanadomolybdic acid to the next sample stream.
(0-50 ppm); cell path length depends on measurement Outputs:
range. Analyzer measures orthophosphate only. 4-20 mA or 0-5 Vdc (non-isolated). Isolated output is optional.
Sampling Method: Batch; update rate is approximately every Load requirements for 4-20 mA output: 450 W (minimum),
15 minutes. 1200 W (maximum).
Accuracy: Load requirement for 0-5 Vdc output: 10kW (minimum).
0 — 3 ppm through 0 — 25 ppm ranges: ±5% of reading or Outputs are not scalable. The range of the outputs is the
±0.1 ppm, whichever is greater same as the measurement range.
0 — 50 ppm and 0 — 100 ppm ranges: ±5% of reading or Parallel output for printer. Standard on multi-stream
±1 ppm, whichever is greater instrument. Optional on single stream instrument.
RS232C output is standard on multi-stream instrument,
Calibration standards: zero and full-scale standard for each optional on single stream instrument.
measurement range. Alarms:
Automatic calibration: General: process alarms for streams 1 through 6, leak
Blank and full scale: every 48 hours, requires about 40 min. detection, LED failure, and calibration in progress.
Reagent consumption: 5 L of each reagent every 3 months Stream 1: separate high and low alarm relays; contacts are
Standard consumption: 5 L of standard and 5 L of baseline solu- rated 2 A at 250 Vac or 3 A at 30 Vdc; delay on/off time is 1 min.
tion every three months Streams 2 through 6 (multiple output option): one alarm
Sample requirements: relay for each stream; alarm can be configured as high or
low; contacts have dc ratings only: 100 Vdc (maximum),
Flow: 10-15 mL/min (minimum)
10 VA (maximum), 1.0 amp (maximum - carry only), 0.5
Pressure: <10 psig (<170 kPa) (overflow cup must drain to amp (maximum - switched); delay on/off time is 1 min.
atmosphere)
Display: Single line LCD; 4 1/2 digits; readout in concentration
Temperature: 0-50°C (32-122°F) units. For multi-stream option, display remains at last
Filter: 8 micron filter in sample inlet line is recommended. value until new stream reading is stable.
Multiple sample option: Status LEDs: for leak, source LED failure, hours remaining before
Up to six sample streams, measured sequentially.

2 wwww.rosemountanalytical.com
January 2014 Model CFA3047

maintenance, and calibration in progress; for multi-stream swing out access to rear panel. Allow 12 in. (305 mm) on
analyzers, status LEDs show the sample stream currently the other side for access to input/output panel and over-
valved in and the sample stream for which data is being dis- flow sampler.
played. Panel mounting: Allow 12 in. (305 mm) on both sides for
Environmental: Install in a clean, indoor environment. Ambient access to input/output panel and overflow sampler.
temperature: 10-45°C (50-113°F). Avoid temperature swings Weight: 117 lb (53 kg). Add up to 60 lb (27 kg) for full weight of
greater than ±5°C (10°F) over a twelve-hour period. reagents and standards.
Humidity: 5-95% (non-condensing). Not for outdoor use. Shipping Weight of Analyzer: 147 lb (67 kg)
Dimensions: Shipping Weight of Reagents: 64 lb (29 kg) for three
Height of analyzer cabinet: 25.75 in. (654 mm) months supply
Height of reagent cabinet: 18.25 in. (464 mm) Power:
Width: 18 in. (457 mm) includes hinges Standard: 115 Vac (±10%), 60 Hz, 3 Amp
Depth: 16 in. (406 mm) includes rear mounting bracket Optional: 230 Vac (±10%), 50 Hz, 1.5 Amp
Agency Approvals:
Space Requirements: IEC 61326
Wall mounting: Allow 18 in. (457 mm) on one side for

Ordering Information
Model CFA3047 is a batch analyzer suitable for the determination of orthophosphate in the range 0-3 ppm, 0-5 ppm, 0-10 ppm,
0-20 ppm, 0-25 ppm, and 0-50 ppm (as PO4). Range must be specified at time of ordering. A single, non-isolated output is standard.
Isolated outputs and multiple outputs (one output for each sample stream) are available as options and must be specified at time of
ordering. 115 Vac, 60 Hz is standard; 230 Vac, 50 Hz must be specified at time of ordering. The analyzer is shipped with a three-
month supply of reagents and standards.

Model 185 On-Line Analyzer


CODE Measurement ( required selection)
185-A002 On-line analyzer for phosphate

CODE Number of Sample Streams (required selection)


01C One stream
02C Two streams
03C Three streams
04C Four streams
05C Five streams
06C Six streams
CODE Output isolation
-- No isolation
N1 Isolated outputs
CODE Outputs
-- Single output
N2 Multiple outputs equal to number of sample streams
CODE RS232C interface
-- No RS232C interface
N3 RS232C interface for remote communications
CODE Power
-- 115 VAC 60 Hz
N4 230 VAC 50 Hz
185-A002 -01C -N1 -N4 Example

www.rosemountanalytical.com Rosemount Analytical 3


3
Model CFA3047 Product Data Sheet
PDS 71-Phosphate3047/rev.C January 2014

Parts and Accessories


Part Number Description
185-3047-63 CFA-3047 phosphate reagents and standards, 3 months, 0-3 ppm
185-3047-45 CFA-3047 phosphate reagents and standards, 3 months, 0-5 ppm
185-3047-40 CFA-3047 phosphate reagents and standards, 3 months, 0-10 ppm
185-3047-42 CFA-3047 phosphate reagents and standards, 3 months, 0-20 ppm
185-3047-30 CFA-3047 phosphate reagents and standards, 3 months, 0-25 ppm
185-3047-43 CFA-3047 phosphate reagents and standards, 3 months, 0-50 ppm
185-1341-08 Disposable cartridge filter (for overflow assembly)

NOTE: A biannual reagent and valve pump service contract is available and is recommended. To order the reagent and pump service
contract, provide a blanket purchase order for eight reagent and standard kits (two-year supply). The kits will be sent in regular
intervals over the 24-month period. Once during the 24-month period you will be entitled to a valve pump refurbishment.
Contact Rosemount Analytical for details concerning the valve pump refurbishment.

http://www.twitter.com/raihome http://www.facebook.com/ Scan with your smart


EmersonProcessManagement phone for more
http://www.analyticexpert.com information about
http://www.linkedin.com/company/2753
Rosemount Analytical
Model CFA3047.
Emerson Process Management
Rosemount Analytical
2400 Barranca Parkway
Irvine, CA 92606 USA 8

T 949 757 8500


F 949 474 7250
http://www.rosemountanalytical.com

© 2014 Emerson Process Management. All rights reserved.


The Emerson logo is a trademark and service mark of Emerson Electric Co. Rosemount Analytical
isa mark of one of the Emerson Process Management family of companies. All other marks are
the property of their respective owners.
The contents of this publication are presented for information purposes only, and while effort
has been made to ensure their accuracy, they are not to be construed as warranties or guaran-
tees, express or implied, regarding the products or services described herein or their use or
applicability. All sales are governed by our terms and conditions, which are available on request.
We reserve the right to modify or improve the designs or specifications of our products at any
time without notice.