LIQ PDS 3047 Phosphate
LIQ PDS 3047 Phosphate
Applications
The Model CFA3047 Phosphate analyzer is intended for the Using an optional built-in stream selector, the Model CFA3047
determination of orthophosphate in boiler water. can measure as many as six sample streams sequentially. A sin-
gle analog output is standard. A stream identification marker
Features and Benefits indicates the start of each new sample. A multiple output option
in which a separate analog output is assigned to each sample
Simplicity, ease-of-use, and reliable operation characterize the
Model CFA3047 Valve-Pump phosphate analyzer. Place color- stream is also available. With the multiple output option each
coded reagent straws in color-coded reagent bottles, place the output holds the last value until updated.
blank and standard straws in the appropriate bottles, and press The Model CFA3047 Phosphate Analyzer has a pumped reagent
a button. The analyzer automatically calibrates and puts itself in delivery system. Instrument air, needed for pressurized systems,
service. Once in service, the analyzer runs a blank and full scale is unnecessary. The Model 3047 uses a stable, pre-mixed reduc-
standard twice a week. ing agent. There is no need to mix powders and liquids before
Routine maintenance is simple. Replace reagents and standards putting the reagent in service.
every three months. Replace the valve pump, the key to the reliable The Model CFA3047 uses a batch measurement technique. The
performance and low maintenance of the Model CFA3047, every update rate is approximately every 15 minutes.
two years. Total maintenance time is at most one hour per year.
Model CFA3047 January 2014
Principles of Operation
The heart of the CFA3047 Phosphate analyzer is the valve pump.
The valve pump has separate pumping elements for sample,
reagents, and standards. Each pump element consists of a chem-
ically inert flexible membrane in a cavity. Cavity volumes are 5 mL
for sample and standards and 0.2 mL for reagents. A vacuum/
pressure pump pushes and pulls on the membrane. Vacuum
causes the chamber to fill. Pressure pushes the liquid out of the
cavity into the reaction chamber. Check valves prevent backflow.
A stream of air pumped into the reaction chamber mixes the
sample and reagents. After the reaction time has passed, the
treated sample is pumped to the measurement cell. Light at 660 nm
(470 nm for 0-50 ppm) from an LED passes through the sample
and a photodiode measures the light intensity. The analyzer
automatically calculates the absorbance and translates the
results into a phosphate concentration reading. Once the meas-
urement is complete, repeated flushes with fresh sample remove
the treated sample from the measurement and reaction cells.
2 wwww.rosemountanalytical.com
January 2014 Model CFA3047
maintenance, and calibration in progress; for multi-stream swing out access to rear panel. Allow 12 in. (305 mm) on
analyzers, status LEDs show the sample stream currently the other side for access to input/output panel and over-
valved in and the sample stream for which data is being dis- flow sampler.
played. Panel mounting: Allow 12 in. (305 mm) on both sides for
Environmental: Install in a clean, indoor environment. Ambient access to input/output panel and overflow sampler.
temperature: 10-45°C (50-113°F). Avoid temperature swings Weight: 117 lb (53 kg). Add up to 60 lb (27 kg) for full weight of
greater than ±5°C (10°F) over a twelve-hour period. reagents and standards.
Humidity: 5-95% (non-condensing). Not for outdoor use. Shipping Weight of Analyzer: 147 lb (67 kg)
Dimensions: Shipping Weight of Reagents: 64 lb (29 kg) for three
Height of analyzer cabinet: 25.75 in. (654 mm) months supply
Height of reagent cabinet: 18.25 in. (464 mm) Power:
Width: 18 in. (457 mm) includes hinges Standard: 115 Vac (±10%), 60 Hz, 3 Amp
Depth: 16 in. (406 mm) includes rear mounting bracket Optional: 230 Vac (±10%), 50 Hz, 1.5 Amp
Agency Approvals:
Space Requirements: IEC 61326
Wall mounting: Allow 18 in. (457 mm) on one side for
Ordering Information
Model CFA3047 is a batch analyzer suitable for the determination of orthophosphate in the range 0-3 ppm, 0-5 ppm, 0-10 ppm,
0-20 ppm, 0-25 ppm, and 0-50 ppm (as PO4). Range must be specified at time of ordering. A single, non-isolated output is standard.
Isolated outputs and multiple outputs (one output for each sample stream) are available as options and must be specified at time of
ordering. 115 Vac, 60 Hz is standard; 230 Vac, 50 Hz must be specified at time of ordering. The analyzer is shipped with a three-
month supply of reagents and standards.
NOTE: A biannual reagent and valve pump service contract is available and is recommended. To order the reagent and pump service
contract, provide a blanket purchase order for eight reagent and standard kits (two-year supply). The kits will be sent in regular
intervals over the 24-month period. Once during the 24-month period you will be entitled to a valve pump refurbishment.
Contact Rosemount Analytical for details concerning the valve pump refurbishment.