ASTM D516-2011_7500
ASTM D516-2011_7500
ASTM D516-2011_7500
1
D516 – 11
5000 mg/L will cause a negative interference. Aluminum, mL 95 % ethanol or isopropanol and 75 g sodium chloride
polymers, and large quantities of organic material present in (NaCl) in a container. Add 50 mL glycerol and mix.
the test sample may cause the barium sulfate to precipitate 8.5 Sulfate Solution, Standard (1 mL = 0.100 mg SO4−−)—
nonuniformly. In the presence of organic matter certain bacte- Dissolve 0.1479 g of anhydrous sodium sulfate (Na2SO4) in
ria may reduce sulfate to sulfide. To minimize the action of water, and dilute with water to 1 L in a volumetric flask. A
sulfate reducing bacteria, samples should be refrigerated at 4°C purchased stock solution of adequate purity is also acceptable.
when the presence of such bacteria is suspected.
9. Sampling
6.3 Although other ions normally found in water do not
appear to interfere, the formation of the barium sulfate suspen- 9.1 Collect the sample in accordance with Practice D1066,
sion is very critical. Determinations that are in doubt may be and Practices D3370, as applicable.
checked by a gravimetric method in some cases, or by the 10. Calibration
procedure suggested in Note 2.
10.1 Follow the procedure given in Section 11, using
7. Apparatus appropriate amounts of the standard sulfate solution prepared
in accordance with 8.5 and prepare a calibration curve showing
7.1 Photometer—One of the following which are given in sulfate ion content in milligrams per litre plotted against the
order of preference. corresponding photometer readings (Note 1). Prepare standards
7.1.1 Nephelometer or turbidimeter; by diluting with water 0.0, 5.0, 10.0, 15.0, 20.0, 30.0, and 40.0
7.1.2 Spectrophotometer for use at 420 nm with light path of mL of standard sulfate solution to 100-mL volumes in volu-
4 to 5 cm; metric flasks. These solutions will have sulfate ion concentra-
7.1.3 Filter photometer with a violet filter having a maxi- tions of 0.0, 5.0, 10.0, 15.0, 20.0, 30.0, and 40.0 mg/L (ppm),
mum near 420 nm and a light path of 4 to 5 cm. respectively.
7.2 Stopwatch, if the magnetic stirrer is not equipped with
an accurate timer. NOTE 1—A separate calibration curve must be prepared for each
photometer and a new curve must be prepared if it is necessary to change
7.3 Measuring Spoon, capacity 0.2 to 0.3 mL.
the cell, lamp, or filter, or if any other alterations of instrument or reagents
7.4 Filter photometers and photometric practices prescribed are made. Check the curve with each series of tests by running two or
in this test method shall conform to Practice E60; spectropho- more solutions of known sulfate concentrations.
tometer practices shall conform to Practice E275.
11. Procedure
8. Reagents 11.1 Filter the sample if it is turbid through a 0.45-µm
8.1 Purity of Reagents—Reagent grade chemicals shall be membrane and adjust the temperature to between 15 and 30°C.
used in all tests. Unless otherwise indicated, it is intended that 11.2 Pipette into a 250-mL beaker 100 mL or less of the
all reagents shall conform to the specifications of the Commit- clear sample containing between 0.5 and 4 mg of sulfate ion
tee on Analytical Reagents of the American Chemical Society.3 (Note 2). Dilute to 100 mL with water if required, and add 5.0
Other grades may be used, provided it is first ascertained that mL of conditioning reagent (Note 1).
the reagent is of sufficiently high purity to permit its use NOTE 2—The solubility of BaSO4 is such that difficulty may be
without lessening the accuracy of the determination. experienced in the determination of sulfate concentrations below about 5
8.2 Purity of Water—Unless otherwise indicated, reference mg/L (ppm). This can be overcome by concentrating the sample or by
to water shall be understood to mean reagent water conforming adding 5 mL of standard sulfate solution (1 mL = 0.100 mg SO4−−) to the
to Specification D1193, Type I. Other reagent water types may sample before diluting to 100 mL. This will add 0.5 mg SO4 to the sample,
which must be subtracted from the final result.
be used provided it is first ascertained that the water is of
sufficiently high purity to permit its use without adversely 11.3 Mix in the stirring apparatus.
affecting the precision and bias of the test method. Type II 11.4 While the solution is being stirred, add a measured
water was specified at the time of round robin testing of this spoonful of BaCl2 crystals (0.3 g) and begin timing immedi-
test method. ately.
8.3 Barium Chloride—Crystals of barium chloride 11.5 Stir exactly 1.0 min at constant speed.
(BaCl2·2H2O) screened to 20 to 30 mesh. To prepare in the NOTE 3—The stirring should be at a constant rate in all determinations.
laboratory, spread crystals over a large watch glass, desiccate The use of a magnetic stirrer has been found satisfactory for this purpose.
for 24 h, screen to remove any crystals that are not 20 to 30 11.6 Immediately after the stirring period has ended, pour
mesh, and store in a clean, dry jar. solution into the cell and measure the turbidity at 30-s intervals
8.4 Conditioning Reagent—Place 30 mL of concentrated for 4 min. Record the maximum reading obtained in the 4-min
hydrochloric acid (HCl, sp gr 1.19), 300 mL reagent water, 100 period.
11.7 If the sample contains color or turbidity, run a sample
blank using the procedure 11.2 through 11.6 without the
3
Reagent Chemicals, American Chemical Society Specifications, American addition of the barium chloride.
Chemical Society, Washington, DC. For Suggestions on the testing of reagents not 11.8 If interferences are suspected, dilute the sample with an
listed by the American Chemical Society, see Annual Standards for Laboratory
Chemicals, BDH Ltd., Poole, Dorset, U.K., and the United States Pharmacopeia
equal volume of water, and determine the sulfate concentration
and National Formulary, U.S. Pharmacopeial Convention, Inc. (USPC), Rockville, again. If the value so determined is one half that in the
MD. undiluted sample, interferences may be assumed to be absent.