PMMC & Moving Iron
PMMC & Moving Iron
PMMC & Moving Iron
Absolute instrument:
Secondary instrument:
This type of instrument gives the result directly by the deflection of the
instrument and dial reading only if the latter has been calibrated by
comparison with either an absolute instrument or one which has already been
calibrated . The deflection obtained is meaningless until such a calibration
has been made.
All secondary instrument can be classified according to the manner , the result or
observation is presented or displayed .The types are
i)
Indicating instrument
Recording instrument
This instrument gives continuous record of the quantity being measured over
a specific period . The variation of the quantity being measured are recorded
by a pen(attached to the moving system of the instrument) giving permanent
mark over the moving sheet of paper or chart at constant speed. Examples
are X-Y plotter , strip chart recorder etc.
iii)
Integrating instrument
This instrument totalize events over a specific period of time . The summation
, which they give is the product of time and instantaneous quantity being
measured. Examples are Ampere-hour meter , Watt-hour or energy meter .
Magnetic
Thermal
Chemical
Electrostatic
Electromagnetic Induction
Magnetic effect:
When two plates are charged with opposite or same polarity , there is a force of
attraction or repulsion between the plates . This force is utilized to move one of
the plates and the movement is coupled to the driving mechanism of the pointer
giving measurement of the quantity under measurement
Induction effect:
All secondary indicating type instrument are constructed with the subsystems as
i) moving system ii) supporting mechanism ii) Control system to obtain steady
deflection . The constructional details are described
Operating forces of Electromechanical Indicating instrument:
Three types of forces are needed for the satisfactory operation . These are :i)
Deflecting force ii) Controlling force iii) Damping force
Deflecting force:
The deflecting or operating force is required for moving the pointer from its zero
position . the system producing the deflecting force is called Moving system .
The deflecting force can be produced by utilizing any of the aforesaid effects
.Thus the moving system converts the current into mechanical force in
electromechanical type instrument.
Controlling force:
This force is required for indicating instrument in order that the current produces
a steady deflection which is proportional to the magnitude of the current flowing
the moving system. The function of the controlling force are:
a) to produce a force equal and opposite to the deflecting force at the final
steady position of the pointer in order to make the deflection definite for a
particular magnitude of current . In absence of a controlling system , the
pointer will shoot beyond the final steady position for any magnitude of
current and thus the deflection will be indefinite.
b) To bring the moving system back to zero position when the force causing
the instrument moving system to deflect is removed . In the absence of
controlling system the pointer will not come back to zero position when
current through the moving system ceases to flow .
Damping force:
When a deflecting force is applied to the moving system , it deflects and it should
come to rest at a position where the deflecting force is balanced by the
controlling force . The deflecting and controlling forces are produced by the
system which have the inertia and therefore , the moving system can not
3
immediately settle at its final position but overshoots or swings around the final
position . This is illustrated in the following fig. Suppose O is the equilibrium or
final steady position . Because of the inertia of the moving system moves to
position a . Now for any position a beyond the equilibrium position the
controlling force is more than the deflecting force and hence the moving system
swings back . Due to inertia it can not settle at O but swings to a position say b
behind the equilibrium position . At b , the deflecting force is more than the
controlling force and hence the moving system again swings . The pointer thus
oscillates about its final steady position with decreasing amplitude till its kinetic
energy is dissipated in friction and therefore , it will settle down at its final steady
position . If extra force are not provided to damp these oscillation , the moving
system will take a considerable time to settle to the final position and hence time
consumed in taking readings will be very large. Therefore damping forces are
necessary so that the moving system comes to its equilibrium position rapidly
and smoothly without any oscillation.
Moving system:
The moving parts should be light and the frictional forces should be minimum
.These requirement should be fulfilled in order that power required by the
instrument for its operation is small. The power expenditure is proportional to the
weight of the moving parts and the frictional forces opposing the movement . The
moving part can be made light by using aluminum as far as possible . The
frictional forces are reduced by using a spindle mounted between jewel bearings
and by carefully balancing the system.
Supporting system :
The force or torque developed by the moving element in electromechanical
indicating instrument is small in order that the power consumption be kept low so
that the introduction of the instrument into a circuit may cause minimum change
in the existing circuit conditions . Because of low power level , the supporting the
moving system is of great importance . With operating forces being small , the
frictional forces must be kept to a minimum in order that the instrument reads
correctly and is not erratic in action and is reliable. Several types of supports are
used , depending upon the sensitivity required and the operating conditions to be
met . Supports may be of the following types :
i) Suspension:
It consists of fine ,ribbon shaped metal filament for the upper suspension and coil
of fine wire for the lower part . The ribbon is made of a spring material like
beryllium copper or phosphor bronze . This coiling of lower part of suspension in
done is order to give negligible restraint on the moving system . The type of
suspension requires careful leveling of the instrument , so that the moving
system hangs in correct vertical position . This construction is , therefore , not
suited to field use and is employed only in those laboratory applications in which
very great sensitivity is required . In order to prevent shocks to the suspension
during transit etc. a clamping arrangement is employed for supporting the moving
system .
ii) Taut suspension:
Suspension type instruments can only be used in vertical position . The taut
suspension has a flat ribbon suspension both above and below the moving
element , with suspension kept under tension by a spring arrangement . The
advantage of this suspension is that exact leveling is not required if the moving
element is properly balanced . Ribbon suspensions , in addition to supporting the
moving element , exert a controlling torque when twisted . Taut suspensions are
customarily used in instruments which require a low friction and high sensitivity .
Thus the use of suspension results in elimination of pivots , jewels and control
springs and therefore pivot less instruments are free from many defects.
The moving system is mounted on a spindle made of hardened steel . The two
ends of the spindle are made conical and then polished to form pivots . These
ends fit conical holes in jewels located in the fixed parts of the instruments .
These jewels , which are preferable made of natural or synthetic sapphire form
the bearings . The contact surface between the spindle and bearings gives
minimum friction . It has been found that frictional torque , for jewel bearings , is
proportional to area of contact between the pivot and jewel . the contact area
5
between pivot and jewel should be small . The pivot is ground to a cone and its
tip is rounded to a hemispherical surface of small area. The jewel is ground to a
cone of somewhat larger angle . The pivot may have a radius at tips from 0.0125
mm to as high as 0.075 mm depending upon the weight of the mechanism and
the force it will encounter . The radius of the pit of jewel is somewhat longer so
that the contact is in the form of a circle. The contact area should not be too
small otherwise the stress may exceed the crushing strength of the material of
pivot . The pivots of many modern instruments are loaded nearly to yield point of
steel . The rough handing of the instrument may deform the pivot and crack the
jewel . jewels are sometimes spring mounted to decrease the like hood of
damage because of shocks , and its bottom is also rounded.
The jewel bearing should be dry . any lubricant , such as oil would cause trouble
because of gumming and collected dust . In order that the wear and tear on the
bearings becomes uniform and symmetrical the centre of gravity of weight of the
moving parts should coincide with axis of the spindle for all positions of the
pointer and there are no out of balance forces . This also ensures that the
deflection is unaffected by the position of the pointer and there are no reading
errors . The balance can be achieved by using balance weights carried on arms
attached to the moving system . The balance weight whose position is adjustable
compensates for the weight of pointer and any other part attached to the moving
system . A small weight at a larger distance increase the moment of inertia and
necessitates a larger damping torque . A larger weight at short distance , on the
other hand , increases the weight of the moving system . A compromise is usually
made.
Torque/Weight Ratio:
The frictional torque in indicating instrument depends upon the weight of the
moving parts . If the weight of the moving parts is large , the frictional torque will
be large . The frictional torque exerts a considerable influence on the
performance of an indicating instrument . If the frictional torque is large and is
comparable to a considerable fraction of the deflecting torque the deflection of
the moving system will depend upon the frictional torque to an appreciable extent
. Also the deflection will depend on the direction from which the equilibrium
position is approached and will be uncertain . On the other hand , if the fractional
torque is very small compared with the deflecting torque , its effect on deflection
is negligible . Thus the ratio of deflecting torque to fractional torque is a measure
Of reliability of the instrument indication and of the inherent quality of the design .
hence torque-weight ratio of the instrument is an index of its performance , the
higher the ratio , the better will be its performance . If the deflecting torque is
expressed in terms of a force which acting at a radius of 1 cm , produces full
scale deflection , the ratio of this torque to weight of the moving parts should not
be less than 0.1 as far as possible.
Control system:
The moving system is mounted on a pivoted spindle . The quantity being
measured produces deflecting torque proportional to its magnitude . There
should a restraining or controlling torque acting in opposite to the deflecting
torque , which will bring the moving system to steady deflected position ,
otherwise the moving system will go on moving for indefinite time. Two types of
control system are used in indicating instruments as i) Gravity control ii) Sprin g
control .
Gravity control:
Spring control:
A flat spiral spring is attached to the moving system , produced controlling torque.
One end of the spring is attached to the spindle and the outer end carries a
spigot which engages in a circular disc surrounding the jewel screw . This disc
carries an arm which is slotted and splayed out at the end. The purpose of
slotted extension arm is to allow the spring to be coiled or uncoiled slightly , so
that the pointer may be set at zero . the slotted arm is actuated by a set screw
mounted at the front of instrument and , therefore , zero setting of the instrument
can be done without removing the cover.
The spring material should be i) non-magnetic ii) proof from mechanical fatigue .
Where springs are used to lead the current into the moving system , they should
have a small resistance , their cross sectional area must be sufficient to carry
the current without temperature rise effecting their constant . they should also
have a low resistance temperature coefficient. A no. of non magnetic materials
like silicon bronze , hard rolled silver or copper , platinum silver , platinum
iridium and German silver have been used but have not been found satisfactory
owing to some reason or the other . For most applications phosphor bronze has
been found to be most suitable material except in instruments of low resistance (
) . In this case some special bronze alloys having low resistance may be used
with some sacrifice in mechanical quality .
By making the large no. of turns , the deformation per unit length is kept small on
full scale deflection . The controlling torque is thus made proportional to the angle
of deflection of the moving system . For a flat spiral spring , the controlling torque
developed by deflection of the moving system is given by :
Tc
Ebt 3
. K c N-m.
12l
Ebt 3
12l
( N-m/rad.)
The spring should be stressed well below their elastic limit at maximum deflection
of the instrument in order that there is no permanent set or that no change in
deflection (or zero shift ) will occur from inelastic yield.
6T
Maximum fibre stress f max 2c (N/m2)
bt
Sometimes a parameter length / thickness ratio is used for design purpose .
l
E
( )
Thus t f
2
max
Hence in order that the material is not over stressed the ration of length to
thickness of the spring should not be much less that
Fatigue in spring may be avoided to a great extent by proper annealing and
ageing during manufacture . In order to eliminate the effect of temperature
variations upon the length of the spring, two springs coiled in opposite directions
are used . When the moving system deflects , one spring is extended while the
other is compressed .
Comparison of spring and gravity control instrument:
i)
10
The damping forces thus created always increases with the increase in velocity
of vanes . There is no damping force when the vanes are stationary.
In the second method , a disc attached to the spindle of the moving system is
immersed in damping oil . the frictional drag developed during the motion of the
disc always opposes the motion of the moving system and becomes zero when
the disc is stationary . The suspension system of the disc should be cylindrical
and of small diameter so that the surface tension effect is negligible .
i)
11
Kl
The eddy current , e R
K (A)
wt
B 2 wlrt
Damping force , F BI e l
(N)
K
B 2 wlrt
r (N-m)
Damping torque, Td Fd r
K
B 2 wlr 2 t
B 2 r 2 At
Or Td
K
K
wl
[
, the area of the pole face]
The actual resistance of the total eddy current path depends upon the radial
position of pole and let it be K times as high as that of section under the poles
, K is always greater than unity .
The magnitude of damping torque may be varied by adjusting the radial
position of the poles. Damping torque decreases with the distance of the
magnet towards the edge of the disc and becomes zero when the centre of
the poles are at the edge of the disc .
ii)
12
velocity w radians per second. Let the thickness and width of the section of
metallic former be t and w metre respectively .
The e.m.f induced in each side of the former = Blv
d
d
Bld (V) [ v
Total induced e.m.f. in the former , e 2 Blv 2 Bl
]
2
2(l d )
Resistance of the path for eddy current , R
(ohm)
t
Eddy current
R 2(l d ) 2(l d ) (A)
wt
B 2 l 2 dwt
(N)
2 (l d )
B 2 l 2 d 2 wt
(N-m)
2 (l d )
Td
B 2 l 2 d 2 wt
(N-m/rad)
2 (l d )
Hence the damping can be varied by varying the thickness of the metal
former .
Damping constant K d
Electromagnetic damping:
The movement of a coil in a magnetic field produces a current in the coil due to
induced e.m.f. ,which interacts with the magnetic field to produce torque . This
torque opposes the movement of the coil . The magnitude of the current and
hence the damping torque is dependant on the resistance of the coil circuit .
13
Magnet system:
Special alloys steel are used in instrument magnet construction . Tungsten steel
has been popular since the early days. Cobalt chrome steel has also been used
because of their higher coercivity and where economy of weight and space is
important . But recent years , Alnico magnets has been used to an increasing
extent .Ideally an instrument magnet should have in its air gap high magnetic flux
density which may not change with time or temperature . The design of
instrument magnets involves consideration of weight and economy of space ,
expense of materials and manufacturing processes , and performance of
magnetization both as well as the field strength desired in the air gap . In most of
the applications, the field strength may be expected to be between 0.05 and 0.25
T in air gap of 1.5-2.5 mm length , depending on the size and type of instrument.
Permanet magnets are made of hard materials , i.e. materials which have a
broad hysteresis loop (large co-ercive force) so that they are not subject to self
demagnetization . In order that the volume of the permanent magnet is small ,
the (BH) , of the material used should be large .The following table gives data for
magnetic materials used for permanent magnets.
Material
10%
Carbon
Remenance
(Wb/m2)
0.9
Co-ercive
force
(A/m)
4,000
14
Value of B
for (BH)max
(T)
0.6
Value of H
for (BH)max
(BH) max
2,600
1,560
steel
6% Tungsten
steel
35% Cobalt
steel
Alnico
Alini
Alcomax-III
Alcomax-IV
1.05
5,200
0.7
3,750
2625
0.9
20,000
0.6
13,000
7800
0.8
0.55
1.25
1.1
40,000
46,000
54,000
60,000
0.56
0.36
0.95
0.8
24,000
28,000
40,000
43,000
13,500
10,000
38,000
34,000
Galvanometer:
A galvanometer is an electrical indicating type instrument used for detecting
presence of small currents or voltages in a circuit and measuring the polarity ,
15
Displacement constant:
It is the amount of deflecting torque produced due to flow unit current .e.
T
G d (N-m/A)
i
ii)
Inertia constant:
16
d 2
, where J s called inertia torque constant and is
dt 2
equal to the moment of inertia of the moving system. Thus the retarding torque
produced per unit angular acceleration is called inertia constant (N-m/Rad-sec)
Inertia torque Td J
iii)
Damping constant
Damping is provided by the friction due to motion of the moving system in the
surrounding air and also by induced electrical effects on the moving coil carrying
current . Damping torque is proportional to the angular velocity of the moving
d
D , where
system . Thus damping torque is given by the equation : TD D
dt
Control constant:
A controlling torque is produced due to elasticity of the moving system which tries
to restore the system back to its equilibrium position . Controlling torque Tc S
,where S =control constant (N-m/Rad)
Dynamic equation of motion of the galvanometer:
There are four torques acting on the moving system of the galvanometer at ant
time , since the energization of the coil as
deflecting torque ( Td ) tends to accelerate the coil
Inertia torque ( Ti ) tends to retard the moving system proportional to angular
acceleration
damping torque ( TD ) tends to retard the moving system in proportion to the
angular velocity
control torque ( Tc ) tends to retard the moving system in proportion to angular
displacement
At any time , t within the transition period , the deflecting torque must be equal
the all retarding torque , Thus Td Ti TD Tc
d 2
d
Or J 2 D
S Gi
dt
dt
Where , J = Inertia constant
D =damping constant
S =spring constant
17
G =galvanometer constant
=deflection at time t
i = current at time t
The above equation is a linear 2 nd degree differential equation , whose solution s
the sum of C.F. (complementary function representing the transient condition )
and PI( particular integral representing the steady state condition)
Solution of C.F.:
d
m , the auxiliary equation s : Jm 2 Dm S 0
dt
D D 2 4 JS
D D 2 4 JS
The roots are given by: m1
and m2
2J
2J
Assuming
The solution : Ae m1t Be m2t , where A ,B are constants , whose values are to
be determined from the initial conditions.
Solution of P.I.:
Case a)
When D 2 4 JS , The roots ( m1 , m2 ) are imaginary . Under these condition , the
motion is oscillatory
D
4 JS D 2
j
k1 jk 2
2J
2J
And m2 k1 jk 2
Now m1
Ae ( k1 jk2 )t Be ( k1 jk2 )t F
or e k1t [ Ae jk2t Be jk2t ] F
or e k1t [ A(Cosk 2t jSink 2t ) B (Cosk 2t jSink 2t )]
( A B) 2 ( A B) 2
1
and tan [
( A B)
]
j ( A B)
18
The above equation signifies that , when a steady current passes through the
coil , the angular deflection starts and will oscillate sinusoidally with exponentially
decreasing amplitude before reaching to a final steady state value as f .The
frequency of oscillation is given by : f d
d
1
4 JS D 2
2J
4J
Case b)
When D 2 4 JS , the roots of auxiliary equation are equal i.e. m1 m2 m
The solution in this case is given by : f e
d
D
[ e
dt
2J
D
t
2J
{ A Bt} Be
D
t
J
] [e
D
t
2J
d
0 0 f A
dt
D
. f
and B
2J
The solution is : f [1 e
Now D and
2J
D
t
2J
(1
D
t
J
[ A Bt ]
D
. A B Bt}]
2J
At t 0 , 0 and
A f
D
2J
D
.A B
2J
D
t )]
2J
S
c
J
t
Hence the solution is : f [1 e c (1 c t )]
The above equation signifies that the moving system quickly moves to its final
steady position without any oscillations. The motion is called critically damped
motion and the galvanometer is called critically damped . The plot of vs . t
under critically damped condition s shown below:
19
Case c)
When D 0 , the roots of the equation is m1
2 j JS
S
j
J
J
S
. Thus roots are imaginary with opposite sign
J
And m2 j
S
J
The motion is called undamped under which the moving system of the
galvanometer oscillates around the steady value F sinusoidally with frequency
f n . the plot of vs. t is shown below:
Case d)
When D 2 4 JS , then the roots are real .
20
m2 F
and
m2 m1
m1 F
m2 m1
F [1
m2
m1
e m1t
e m2t ]
m2 m1
m2 m1
Now m1
D
D 2 4 Js
D
D2 S
0 ( 2 1)
2
2J
2J
2J
4J
J
D
D 2 4 JS
0 ( 2 1)
2J
2J
D
Where D , is called relative damping factor
c
And m2
And 0
S
2
2 S
4 J 2 02
; D0 4 JS 4 J
J
J
2
2
Now D 4 J (
2
or 4 J
D02
02
S2
d2 ) 4 J 2 ( 02 d2 )
J
d
d2
d2
D2
2
1 2 d 0 1 2
or ,
,
2
2
2
D0
0
0
0
m 2 m1 0 ( 2 1) 0 ( 2 1) 2 0 2 1
Hence
f [1
f [1
the
0 ( 2 1)
2 0 2 1
e 0
complete
solution
2
0 ( 1) (
( 1 ) t
2
2 0 2 1
is
2 1 ) t
21
Damping of galvanometer:
a) Mechanical damping:
In this type of damping , air friction is produced owing to the motion of moving
coil in the air surrounding it . So damping torque due to air friction is given by :
d
Mechanical damping: Tm Dm
dt
22
i)
If the coil is mounted on the metallic former , then eddy current flows through the
former due to induced eddy e.m.f.. With the interaction of the magnetic field and
eddy current , a damping torque s produced . The voltage induced n the metallic
former s given by : e = no. of conductors X e.m.f. generated n each conductor =
d
d
2 Blv 2 Bl Bld ( )
2
dt
( Bld ) 2 d
(
)
Rf
dt
G 2 d
d
( ) D former ( )
2
N R f dt
dt
G2
N 2Rf
area
bt
bt
Ee
Bld
Blbtd
Eddy current : I e
Rf
2 (l d ) / bt 2 (l d )
Damping force :
FD BI e l B
Damping torque : TD FD d
Damping constant K D
Blbtd
B 2 l 2 btd
l
2 (l d )
2 (l d )
B 2 l 2 d 2 bt
B 2 l 2 d 2 bt
d
2 (l d )
2 (l d )
TD
B 2 l 2 d 2 bt B 2 l 2 d 2 bt
(Nm/rad-sec)
2 (l d ) 2 (l d )
23
d
bt
Actual path of eddy current is not limited to the portion of the disc under the
magnet but is greater than this . Therefore to take this factor into account the
d
actual resistance is taken as K times
bt
Re
Kd
bt
Ee
BdR
BRb
Re K d / bt
K
Damping force FD B
BRbt
B 2 Rdbt
d
K
K
B 2 R 2 dbt
Damping torque TD
K
Damping constant K D
TD
B 2 R 2 dbt
( Nm/rad-sec.)
24
It is clear that the damping torque can be changed by changing the distance of
the poles from the centre of the disc. (i.e. the radial distance of poles w.r.t. the
centre of the disc)
e 2 NBlv 2 NBl
( ) NBld ( )
2 dt
dt
The current flowing in the coil due induced e.m.f. is given by:
e NBld d
i
dt
= NBlxd NBld
NBld d
( NBld ) 2 d
G 2 d
( )
( )
( )
R
dt
R
dt
R dt
G2
R
Te Tcoil T f (
N 2Rf
dt
De
dt
25
G2
G2
2
R
N Rf
TD Tm Te Dm
dt
De
dt
( Dm De )
dt
dt
G2
Thus De R 2
N Rf
But damping due to metal former is usually small and can be neglected .So
De
G2
R
G2
R
Thus for critical damping , the total resistance of the galvanometer circuit should
be : R
G2
2
KJ .
G2
2
KJ R g
Current sensitivity:
It is defined as the full scale deflection in mm. Produced by unit current and is
given by:
S i F / i , but F G / K , S i Gi / K Rad/A
ii)
Voltage sensitivity:
The voltage sensitivity is defined as the deflection in scale divisions per unit
voltage
26
Meg-ohm sensitivity:
The meg-ohm sensitivity is the resistance of the coil circuit so that the
deflection will be 1 scale division with 1 volt impressed to the moving coil .
Thus S o resistance in meg ohm to give a deflection of one scale division
d
with 1 volt =
M-ohm / scale division
i 10 6
a) Moving coil
The moving coil is wound with many turns of enameled or silk covered
copper wire . The coil is mounted on a rectangular aluminum former which is
pivoted on jeweled bearings The coil moves freely in the field of a permanent
magnet .
27
b) Magnet system
The permanent magnet consisting of long U shape having soft iron pole
pieces is used in PMMC type meter. The material used for magnet is
Alcomax , Alnico , which have a high coercive force .The flux density varies
from 0.1 Wb/m2 to 1 Wb/ m2 . Since coercively is high , it is possible to use a
small coil having small no. of turns and hence a reduction in volume is
achieved . Alternatively in instruments having a large scale length it is
possible to increase the air gap length to accommodate large no. of turns.
c) Control mechanism
When the coil is supported between two jewel bearings , the control torque is
provided by two phosphor bronze hair springs . These springs also serves as
leads to the moving coil .
d) Damping mechanism
28
The pointer is carried by the spindle and moves over a graduated scale . The
pointer is of light-weight construction and , apart from those used in some
inexpensive instruments has the section over the scale twisted to form a
fine blade . This helps to reduce parallax errors in the reading of the scale .
The pointer blade usually moves over mirror adjacent to the scale so that
parallax can be avoided by its reflection while noting the reading from the
scale . The weight of the instrument is normally counter balanced by weight
rigidly attached at the end the ,diametrically opposite
Torque equation:
Let N = No. of turns of the moving coil
B =flux density of D.C. magnetic field produced by permanent magnet
l ==length of the coil
d =width of coil
I =the current flowing through moving coil
then the torque equation is given by: Td NBIdl GI
where G NBld = a constant
For spring control PMMC meter , steady deflection is produced when
deflecting torque becomes equal to control torque
29
Thus , Td Tc
constant
or GI K
, where
= steady deflection
and
= spring
(G / K ) I or I ( K / G )
30
I = current to be measured
Since the shunt resistance is in parallel with the moving coil , the voltage drops
across shunt and coil must be same .
The ratio of total current to the coil current is called the multiplying factor of shunt
R
I
m
1 m
Im
Rsh
resistance of shunt = Rsh Rm /(m 1)
Effect of temperature changes in Ammeter:
31
Multi-range Ammeters:
The current range of D.C. ammeter may be further extended by adding multiple
shunts , selected by a range switch . If , if there are n No. of multiplying factors as
, there should be n No. of shunt resistances as
. At a time only one shunt
resistance is to be selected by a selector switch as shown in the fig. In an
alternative arrangement , the Universal shunt (or Ayrton shunt) may be used for
multi-range ammeter. It uses a single shunt resistor with multiple tapping point so
that different values for shunt resistor can be selected (shown in fig.) by selector
switch. By setting the selector position to 1st,2nd , nth , different multiplying
factors (as ) can be set. This is illustrated in the following way :
Considering that the meter ranges have to be extended to I 1 , I 2 , I 3 ...I n . , the
range selector switch position is to be set at 1,2 , nth respectively. For switch
position 1,
I m Rm ( I 1 I m ) R1
m1
R
I1
1 m
Im
R1
or ,
32
Voltmeter multiplier:
A PMMC meter can be converted into D.C. voltmeter by connecting a series
resistance with it . This series resistance is called as multiplier . The multiplier
limits the current through the meter so that it does not exceed the value for full
scale deflection and thus prevents the moving coil being damaged .
The resistance of the multiplier can be calculated as follows:
Let Rm = internal resistance of the moving coil
Rs =multiplier resistance
v =voltage across the moving coil for current
V =Full scale voltage of the PMMC type voltmeter
Since the multiplier resistance is connected in series , then v I m Rm
And V I m ( Rm Rs )
V I m Rm
V
Rs
Rm
Im
Im
R
V I m ( Rm R s )
1 s
So the voltage multiplying factor is m
v
I m Rm
Rm
resistance of multiplier Rs (m 1) Rm
Hence for the measurement of voltage m times the voltage range of the
instrument , the series resistance should be (m-1) times the meter resistance .
The essential requirement of the multiplier are:
33
i)
ii)
iii)
iv)
Rsn ( mn 1) Rm
where Rm =meter resistance
34
b) By Potential divider :
Multiple resistances are connected in series with the PMMC instrument such
that connection to unknown voltage source is made to any one junctions of
these series resistances R1 , R2 , R3 ....Rn for selection of proper voltage range
as . The series resistances for the voltage ranges V1 , V2 ,....Vn
can be
computed as follows:
R1
V1
V1
Rm
Rm m1 Rm Rm ( m1 1) Rm
Im
v / Rm
V2
V2
Rm R1
Rm ( m1 1) Rm m2 Rm Rm ( m1 1) Rm ( m2 m1 ) Rm
Im
v / Rm
Similarly
R2
V3
Rm R1 R2 (m3 m2 ) Rm
Im
This method has the advantage that all multipliers except the first have
standard resistance values and can be obtained commercially in precision
tolerances . The range multiplier , R1 , is the only special resistor which must
be manufactured to meet specific circuit requirement .
R3
35
iii)
36
vi)
37
38
current flowing through the coil and hence the pointer deflects to only one
direction . The controlling torque is provided by springs but gravity control can be
used for panel type instrument which are vertically mounted. Damping is
provided by air friction with the help of a light aluminum piston , which moves in a
fixed chamber closed at one end or with the help of a vane attached to the
moving system.
In repulsion type construction, there are two vanes inside the coil , one is fixed
and the other is movable. Two different types of design are followed as radial
vane ii) coaxial type vane . In radial vane type , two rectangular soft iron strips
are used , one is fixed at inside surface of the coil and the other is attached to the
spindle radially .When the coil is energized by the current , the vanes are
similarly magnetized and a force of repulsion exists between them . This force
moves the movable vane away from the fixed vane. In coaxial type design ,the
fixed iron may consists of tongue shaped piece of sheet iron bent into a
cylindrical form , the moving iron being another piece of sheet iron bent and
mounted so as to move parallel to the fixed iron and towards its narrower end.
The controlling torque is provided by springs The damping torque is provided by
air friction piston moving in enclosed chamber. The operating magnetic field in
moving iron instrument is very weak due air core electromagnet and therefore
eddy current damping is never used . It is clear that whatever may be the
direction of current in the coil , the iron vanes are so magnetized that there is
always a force of attraction or repulsion . The moving iron is unpolarised ,i.e. they
are independent of the direction in which the current passes . therefore , these
instrument can be used with both a.c. and d.c.
39
Torque equation :
An expression of torque of a moving iron instrument may be derived by
considering the energy relations when there is a small increment in current
supplied to the instrument . when this happens there will be small deflection and
some mechanical work will be done .
Let Td be deflecting torque
Mechanical work done = Td d
Suppose the initial current is I , and the inductance of the coil is L
If the current increases by dI then the deflection changes by d and the
Inductances changes by dL
d
dL
dI
( LI ) I
L
dt
dt
dt
The electrical energy supplied is eIdt I 2 dL ILdI
1 2
1
I L to
( I dI ) 2 ( L dL)
The stored energy changes from
2
2
1 2
1 2
2
The change in stored energy is ( I 2 IdI dI )( L dL) I L
2
2
If
1 2
I dL
2
I 2 dL ILdI ILdI
1 2
I dL Td
2
1 2
I dL
2
or deflecting torque Td
1 2 dL
I
2 d
The moving system is provided with spring control and it comes to steady
position when the deflecting torque is balanced by the controlling torque.
Thus Td Tc or K
1 2 dL
1 I 2 dL
I
or ,
2 d
2 K d
Hence the deflection is proportional to square of the rms value of the operating
current . The deflecting torque is , therefore , unidirectional whatever may be the
polarity of the current. As the deflection is proportional to square of current , it is
evident that the scale of such an instrument is non uniform .If there is no
saturation , the change of inductance with angle of deflection is uniform (i.e.
dL
d
40
In actual instruments
dL
d
position of the moving iron and thus the scale is distorted from the square law in
a manner dependant upon the way in which inductance varies with angle of
deflection . This variation can be controlled by suitable design i.e. by choosing
proper dimensions , shape and position of iron vanes . Thus it is possible to
design and construct an instrument with a scale which is very nearly uniform over
a considerable part of its length . The necessary condition relating to
dL
d
Also
or
=a constant . Thus for a linear scale , the product (
d C
d C 2
dL
) should be a constant . This is not possible as it requires
d
to be infinite at
. In practice the scale is made linear from the maximum deflection down to about
1
dL
th of the maximum deflection . The plot of
10
d
41
or
I sh
Im
R (Lm )
R (Lsh ) 2
2
m
2
sh
Rsh 1 (
Lm 2
)
Rm
Lsh 2
)
Rsh
In order that the division of currents through coil and shunt shall remain same for
all frequencies , the ratio of time constants of two branches must be same or in
Lsh Lm
other words ,
.
Rsh Rm
Rm 1 (
=angular frequency
V
( R R s ) 2 2 L2
( R Rs ) 2 2 L2
R 2 2 L2
between
Attraction
and
42
Repulsion
type
moving
iron
Temperature error :
The effect of temperature changes on moving iron on moving iron
instruments arises chiefly from the temperature coefficient of spring .
The error may be 0.02 % per C change in temperature . In voltmeter ,
errors are caused due to self heating of coil and series resistance . The
temperature of the coil may increase by 10 to 20 C for a power
consumption of 1W . Therefore , the resistance increased by about 4 to
8 % , causing a decrease in current for a given voltage . This produces
a decreased deflection . Therefore, the series resistance should be
made of material like Maganin which has a small temperature coefficient . the value of series resistance should be very large as
compared with the coil resistance in order to minimize errors due to
self heating .
iii)
43
V
( R Rs ) 2 2 L2
The deflection of the moving iron voltmeter depends upon the current
through the coil . Therefore , the deflection for a given voltage will be less at
high frequencies than at low frequencies . To some extent , compensation to
this type of error is possible by connecting a capacitor C across the series
resistance R s
This shunting capacitor will make the circuit behave like a pure resistance so
that the frequency changes have no effect on the readings of the instrument .
Thus the compensated instrument will have a power factor nearly equal to
unity .
Rs
Rs jCRs2
jL
Now the impedance of the circuit Z jL
1 jCRs
1 2 C 2 Rs2
Since CRs 1 ,
2
2
2 2
then Z jL ( Rs jCRs )(1 C Rs )
2
2
2 3
3 4
2
2 3
2
2 2 2
= jL Rs jCRs C Rs j Rs Rs C Rs j[L CRs (1 C Rs )]
Z 2 Rs2 (1 2 C 2 Rs2 ) 2 2 ( L CRs2 ) 2
2
This must equal Rs in order that the a.c. calibration at all frequencies and
d.c. calibration is the same .
2
2
2
2 2 2
2
2 2
Thus Rs Rs (1 C Rs ) ( L CR s )
2
2
2 2
4
4 4
2
2 2
= Rs (1 2 C Rs C Rs ) ( L CRs )
2
2
2
2
2
2 2
4
4
4
= Rs (1 2 C Rs ) ( L CRs )
[ C Rs 1 ]
2
2
2 4
2 2
2
2 4
2
2
= Rs 2 C Rs L C Rs 2 LCRs
L2 2 LCR s2 C 2 Rs4 0
2
or, L 2.41CRs
1
L
L
C
( 2 ) 0.41( 2 )
2.41 Rs
Rs
44
Since the deflection of the instrument depends upon the current through the
coil , the deflection for a given voltage will be less at high frequency than at
low frequency . The error due to this problem can be eliminated somehow by
connecting a capacitor in parallel to series resistance r .The idea of shunting
the series resistor is to make the circuit behave like a pure resistor so that the
frequency changes have no effect on the readings of the instrument . Thus
the compensated instrument will have a power factor nearly equal to unity .
The changes in frequency will cause the change in eddy current induced in
metal parts of the instrument. Let the mutual inductance between the coil and
the iron parts be M . The induced voltage due to current I in the instrument
coil lags the current I by 90 . As a result of this induced voltage an eddy
MI
current flows and its magnitude is given by: I e
, where
2
Re 2 L2e
A component of this eddy current ,which will oppose the instrument current
is given by:
I e' I e Cos (90 e ) I e Sin( e )
This current will crease opposing field thus reducing the torque on the
moving system . From the phasor diagram , it is seen that
2 MLe
MI
I e' I e Sin( e )
is small
MI
'
and I e L , when Le Re , i.e. when is large.
e
Thus at low frequency , the eddy current effect increases with square of the
frequency while at high frequency , the effect is practically constant . For
these reason moving iron instruments are unsuitable for frequencies above
125 Hz.
45
Vibration galvanometer:
The instrument is used generally as a.c. bridge null detector . n construction ,
the galvanometer is similar to DArsonval type , having a moving coil
suspended between the poles of strong parmanent magnet . But since the
instrument s used with a.c. circuit , the moving coil is energsed by a.c. . When
an alternating current is passed through the coil , an torque is produced ,
which makes the coil vibrate with a frequency equal to the frequency of the
a.c. . On account of inertia of the moving parts , the amplitude of vibration
becomes small . However when the natural frequency of oscillation of the
moving part is made to equal to the frequency of a.c. , the mechanical
resonance is obtained and the moving system vibrate about its axis of rotation
with large amplitude . The damping is very small in these galvanometer n
order to obtain sharp resonance characteristic.
46
Construction:
The moving coil consists of a few turns of wire made of fine bronze or
platinum silver . This wires passes over a small pulley at the top and s pulled
tight by a spring attached to the pulley . The tension of the spring can be
adjusted by tuning a head attached to the spring . the loop of wire s stretched
over two ivory bridge pieces , the distance between the pieces s adjustable .
The moving coil carries a small mirror ,upon which a focused light beam is
incident and reflected . when the moving coil vibrate due to passage of
current , the reflected beam from the mirror throws a band of light on the
graduated scale . the effective resistance of the wire loop is about 250 ohm .
The current sensitivity s 50 mm /uA
Performance analysis :
Let the alternating current
flows through the moving coil , producing
instantaneous deflection from the equilibrium position .
The equation of motion is given by :
T TD Tc Td
where
Td = deflection torque = GI m Sin (t )
d 2
Ti = Inertia torque= J
dt 2
d
TD =Damping torque= D
dt
Tc =Controlling torque= S
G =Displacement constant
d
d
J 2 D
S GI m Sin (t )
dt
dt
47
m1
D D2 4 JS
2J
and m2
D D2 4 JS
2J
m2t
The defection under steady state condition is found by solving the particular
d 2
d
integral : J 2 D
S GI m Sin(t )
dt
dt
The P.. is of the form : f max Sin(t )
d f
d 2 f
maxCos (t ) and
max 2 Sin(t )
2
dt
dt
Substituting these values in equation of motion , we get ,
max J 2 Sin(t ) max DCos (t ) GI m Sin(t )
2
or , Sin(t ).[ max S max J ] max DCos (t ) GI m Sin(t )
Now when , at t , then max D GI m Sin( )
2
When (t )
, then max J max S GI m Cos ( )
2
Also from
GI m
( D ) ( S J 2 ) 2
2
, tan
GI m
D
D
1
)
or , tan (
2
S J
S J 2
( D ) 2 ( S J 2 ) 2
Sin(t )
D
t
2J
.F .Sin(
4 JS D 2
t )
2J
GI m
( D ) 2 ( S J 2 ) 2
Sin(t )
Since the 1st term of the above equation represent transient behaviour ,which
usually affects only the 1st few vibrations after switching on , and diminishes to
zero , the steady state deflection id given by:
(t )
GI m
( D ) 2 ( S J 2 ) 2
Sin(t )
48
GI m
( D ) 2 ( S J 2 ) 2
1
2
S
J
= undamped natural
S
0
J
So when the frequency of the a.c. signal , applied s equal to the natural
frequency of the moving system , the amplitude of vibration reaches the
maximum value and is equal to
GI m
J
.
D
S
mechanical resonance .
The deflection of the moving system at various frequencies of a.c. signal and
at constant amplitude can be calculated from which a plot of f vs. can
be generated , which is a bell shaped curve . It is found that for a
galvanometer having a sharply peaked resonance curve a slightly deviation
49
GI
f
m
[2D.2.( S J 2 )(2J )]
Thus , d 2
2
2 2
( D ) ( S J )
1 GI m .2.[D J ( S J 2 )]
3
=2
2
2 2 2
[( D ) ( S J ) ]
d f
d
S
J
At
S
J
d f
Or , d
S
J
,
GI m
S
(D 2 . )
J
3
2
the
sensitivity
GI m .D.
D 3 .(
S
J
S 3
)
J
by
GI m .J
GI m .J
GI m 1
(
)
D 2 .S
D 2 .S
D2 2 .
D J ( S J 2 ) 0
max
given
GI m .J
D 2 .S
is
or ,
, when
D
( S J 2 )
J
S D
J J2
50
or,
d f
d
S D
J J2
or
51