Unit 1
Unit 1
Unit 1
MEASURING INSTRUMENTS
1. Absolute Instruments:
These instruments give the value of the electrical quantity in terms of absolute quantities
(or Some constants) of the instruments and their deflections. In this type of instruments no
calibration or comparison with other instruments is necessary. They are generally not used in
laboratories and are seldom used in practice by electricians.
Some of the examples of absolute instruments are:
EX: Tangent galvanometer.
2. Secondary Instruments:
They are direct reading instruments. The quantity to be measured by these instruments can be
determined from the deflection of the instruments. They are often calibrated by comparing
them with either some absolute instruments or with those which have already been calibrated.
Some of the very widely used secondary instruments are: ammeters, voltmeter, wattmeter,
energy meter , ampere-hour meters etc.
secondary instruments are again classified into three types, these are.
1. Indicating instruments
2. Recording instruments
3. Integrating instruments.
1. Indicating instruments:
Indicating instruments indicate the quantity to be measured at the time of
measurement by means of a pointer which moves on a scale.
Examples are ammeter, voltmeter, wattmeter etc.
2. Recording instruments:
These instruments record continuously the variation of any electrical quantity with respect
to specified time. In principle, these are indicating instruments but so arranged that a
permanent continuous record of the indication is made on a chart or dial. The recording is
generally made by a pen on a graph paper which is rotated on a dice or drum at a uniform
speed. The amount of the quantity at any time (instant) may be read from the traced chart.
Any variation in the quantity with time is recorded by these instruments.
Examples are ECG and X-Y recorder etc.
2. Controlling torque/force:
This torque must act in the opposite direction to the deflecting torque, and the
movement will take up an equilibrium or definite position when the deflecting and
controlling torque are equal in magnitude. The controlling torque is dependent on
the magnitude of deflection produced. The moving system is deflected from zero to
such a position that the controlling torque at that deflected position is equal to the
deflecting torque. The controlling torque increases in magnitude with the deflection
till it balances the deflecting torque. That is, for a steady deflection,
Controlling torque = Deflection or operating torque, Then we will get the steady
deflection.
The controlling torque developed in an instrument has two functions:
(a) It limits the movement of the moving system and ensures that the magnitude of
the deflections always remains the same for a given value of the quantity to be measured.
(b) It brings back the moving system to its zero position when deflection force is
zero.There are two methods to provide controlling torques.
1. Spring controlling torque.
2. Gravity controlling torque.
The moving iron type instruments are one of the types of measuring instruments used for
measuring voltage or current. These instruments use a movable piece of iron placed in the magnetic
field that deflects the pointer over the scale and hence named moving iron instrument.
There are two types of moving iron (MI) instruments. They are,
1) Attraction type moving iron instruments
2) Repulsion type moving iron instruments.
Since the operating magnetic field produced by the coil winding is not much strong, the eddy current
damping which uses permanent magnets can distort the main field. Thus eddy current damping cannot
be used and fluid friction damping is not much preferred. The moving iron is made of sheet metal for
obtaining a uniform scale.
For a small increment in current (dI), the deflection is increased by dθ. We know that the energy stored
by the inductance of the meter = 1/2 LI2. Therefore, new energy storage due to increment in current,
= 1/2 (L + dL)(I + dI)2
The change in stored energy is,
Neglecting the second and third order differential terms, we get change in stored energy as,
Similar to attraction force there will be repulsion force when same magnetic poles are placed near
each other. Based upon this repulsion force between two like poles the repulsion type moving iron
instruments were developed.
It consists of two iron strips (vanes) are placed radially, in which one is fixed and the other is movable.
In this type of instrument, the deflection torque is directly proportional to the actual current in the coil,
thus making the scale uniform and readings can be obtained directly. These are the most sensitive type
of instruments.
Hysteresis error
Temperature error
Errors due to stray magnetic fields
Errors due to change in frequency
Hysteresis Error :
The main source of this error is the tendency of the iron parts in the instrument to store magnetism for
short periods. Due to this error, for decaying values of currents, the flux appears to be more than its
corresponding value. As the deflecting torque is directly proportional to the flux, the meter reads high.
Control System :
The function of the control system is to produce the necessary and sufficient controlling torque. The control
torque in PMMC instrument is provided by two phosphor bronze hairsprings. These springs also allow the
current to flow in and out of the coil. The control torque produced is equal in magnitude but opposite to
that of deflection torque, for the pointer to reach the steady-state position. Gravity control and ribbon
suspension are also used for producing controlling torque.
Damping System :
The function of the damping system is to provide the damping torque for allowing the pointer to reach the
rest position within a short duration of time. The damping produced in the PMMC instrument is
electromagnetic damping which is obtained by the movement of the aluminum former in the magnetic field
on which the core is mounted. Eddy current damping system and bearing friction also provide the damping
torque.
Whenever the coil moves in the magnetic field, the pointer shows deflection and the deflection of the
pointer is directly proportional to the coil current. The pointer is a thin device that is lightweight in
construction. It is made up of aluminum and is flattened at one end in order to make it as an edge of a knife.
The pointer is mounted on a spindle that moves a graduated scale and it is balanced by the balancing
weight which is connected to it. A mirror is placed below the scale in order to avoid parallax error.
When the current to be measured is applied to the moving coil terminals placed in the magnetic field. The
coil experience a torque and starts moving. This movement of the coil also moves the pointer attached to it
i.e., causing deflection of pointer over the scale.
F = NBIl sinα
F = NBIl sin90°
F = NBIl ...(1)
Td = F × Distance = F × d ...(2)
Substituting equation 1 in 2, we get,
Td = NBIl × d
But the area of the coil is given by,
a=l×d
Td = NBIa
∴ Td = GI (G = galvanometer constant = NBa)
The controlling torque is obtained by the spring control action.
∴ Tc = Kθ
Assuming θ as deflection and K as spring constant. At steady-state deflection, we know that Tc = Td i.e.,
Kθ = GI.
∴ Deflection, θ = GI/K
It is basically an air-cored coil with a large area of cross-section and fewer turns. The fixed coils are
divided into two sections and a moving coil is placed between them. These two coils are connected in
series with line and hence carry the full line current. The main function of the fixed coil is to produce the
required flux. The field coils are laminated and varnished as they carry heavy current and they are also
known as current coils.
Moving Coil :
The moving coil is placed between the two fixed coils and made to rotate with the help of a spindle. A
pointer is attached to the spindle that moves with the moving coil over the scale. The coil is wounded on a
metallic former and connected across the line. It is also known as the voltage coil or pressure coil. The
moving coil is made with a small cross-sectional area and more turns so as to limit the current through it.
The resistance of the voltage coil can be increased by connecting a resistance in series with it.
Springs :
Since the moving coil is not a stationary element, it cannot be connected directly to the supply. The
supply to the moving coil is given through springs. These springs also provide the necessary controlling
torque to control the deflection of the pointer.
Dampers :
This type of instrument is provided with air friction damping to damp out oscillations by providing
necessary damping torque. The moving coils carry a pair of aluminum vanes, which move in sector
shaped chamber at the bottom of the instrument.
Shielding :
When the current passes through these coils, the fixed coil, as well as the moving coil, creates a magnetic
field due to which, the deflecting torque is developed, which is proportional to power. Thus, the power
can be measured by using the electrodynamic type wattmeter.
Assume, the two currents i1 and i2 to be sinusoidal with a phase difference of Φ we get,
i1 = Im1 sinωt
i2 = Im2 sin(ωt - Φ)
Substituting, i1 and i2 values in equation 3, we get,
In dc operation of a dynamometer type wattmeter, the deflection produced in the torque is given by,
Where,
I1, I2 = RMS values of currents in fixed and moving coils respectively
cos Φ = Power factor
Hence, the deflection is proportional to the product of RMS values of the currents in both coils, power
factor, and rate of change of mutual inductance.
Therefore, for a dynamometer-type wattmeter, the scale is calibrated with the square root of the current
that is squared, which is nothing but the RMS value of the ac quantity. Hence, in this way, both ac and dc
can be measured using dynamometer type wattmeter and indicates power whether it is connected in ac or
dc circuit.
And m is called multiplying power of shunt and is defined as the ratio of totalcurrent to
the current through the coil
Multirange ammeters
The range of basic dc ammeter can be extended by using no. of shunts and a selector switch,
such a meter is called multirange ammeter and is shown in the fig
A multirange ammeter can be constructed simple by employing several values of shunt resistances,
with a rotary switch to select the desired range. Fig. shows the circuit arrangement.
Range extension of voltmeter by using Multiplier:
The resistance is required to be connected in series with basic meter to use it as a voltmeter. This
series resistance is called a multiplier. The main function of the multiplier is to limit the current
through the basic meter, so that meter current does not exceed full scale deflection value.
The multiplier resistance can be calculated as Let Rm is
the internal resistance of the coil. Rs = series multiplier
resistance
Im = full scale deflection current
V = full range voltage to be measured V = ImRm + ImRs
ImRs = V –ImRm / ImRs = V/Im - Rm
The multiplying factor for multiplier is the ratio of full range voltage to be measured
and the drop across the basic meter.
.
The DVM is a versatile and accurate instrument that can be used in many laboratory
measurement applications. Since the development and perfection DVM have
been drastically reduced so that DVMs can actively compete with conventional analog
instruments, both in portability and price. The DVM’s outstanding qualities
can best be illustrated by quoting some typical operating and performance characteristics.
Digital voltmeters can be classified according to the following broad categories:
(a) Ramp-type DVM
(b) Integrating DVM
(c) Continuous-balance DVM
(d) Successive-approximation DVM
RAMP-TYPE DVM:
The operating principle of the ramp-type DVM is based on the measurement of the time
it takes for a linear ramp voltage to rise from 0 V to the level of the input voltage, or to decrease
from the level of the input voltage to zero. This time interval is measured with an electronic time-
interval counter, and the count is displayed as a number of digits on electronic indicating tubes.
the comparator output is positive and the control circuits keep D 7 set. Similarly the rest of
the bits beginning from D7 to D0 are set and tested. Therefore, the measurement is
completed in 8 clock pulses.
At the beginning of the measurement cycle, a start pulse is applied to the start-stop
multivibrator. This sets a 1 in the MSB of the control register and a 0 in all bits (assuming
an 8-bit control) its reading would be 10000000. This initial setting of the register causes
the output of the D/A converter to be half the reference voltage, i.e. 1/2 V. This converter
output is compared to the unknown input by the comparator. If the input voltage is greater
than the converter reference voltage, the comparator output produces an output that causes
the control register to retain the 1 setting in its MSB and the converter continues to supply
its reference output voltage of 1/2 Vref..
The ring counter then advances one count, shifting a 1 in the second MSB of the control
register and its reading becomes 11000000. This causes the D/A converter to increase its
reference output by 1 increment to 1/4 V, i.e. 1/2 V + 1/4 V, and again it is compared with
the unknown input. If in this case the total reference voltage exceeds the unknown voltage,
the comparator produces an output that causes the control register to reset its second MSB
to 0. The converter output then returns to its previous value of 1/2 V and awaits another
input from the SAR. When the ring counter advances by 1, the third MSB is set to 1 and
the converter output rises by the next increment of 1/2 V + 1/8 V. The measurement cycle
thus proceeds through a series of Successive Approximation Type DVM. Finally, when the
ring counter reaches its final count, the measurement cycle stops and the digital output of
the control register represents the final approximation of the unknown input voltage.
The signal may be amplified before being applied to the Schmitt trigger. The Schmitt trigger
converts the input signal into a square wave with fast rise and fall times, which is then
differentiated and clipped. As a result, the output from the Schmitt trigger is a train of pulses,
one pulse for each cycle of the signal.
The output pulses from the Schmitt trigger are fed to a START/STOP gate. When this
gate is enabled, the input pulses pass through this gate and are fed directly to the electronic
counter, which counts the number of pulses.
When this gate is disabled, the counter stops counting the incoming pulses. The
counter displays the number of pulses that have passed through it in the time interval between
start and stop. If this interval is known, the unknown frequency can be measured.
The basic circuit for frequency measurement is as shown in Figure. The output of the
unknown frequency is applied to a Schmitt trigger, producing positive pulses at the output.
These pulses are called the counter signals and are present at point A of the main gate.
Positive pulses from the time base selector are present at point B of the START gate and at
point B of the STOP gate.
As the STOP gate is enabled, the positive pulses from the time base pass through the STOP
gate to the Set (S) input of the F/F-2 thereby setting F/F-2 to the 1 state and keeping it there.
The resulting 0 output level from Y̅ of F/F-2 is applied to terminal B of the main gate. Hence
no pulses from the unknown frequency source can pass through the main gate.
In order to start the operation, a positive pulse is applied to (read input) reset input of
F/F-1, thereby causing its state to change. Hence Y̅ = 1, Y = 0, and as a result the STOP gate
is disabled and the START gate enabled. This same read pulse is simultaneously applied to
the reset input of all decade counters, so that they are reset to 0 and the counting can start.
When the next pulse from the time base arrives, it is able to pass through the START gate to
reset F/F-2, therefore, the F/F-2 output changes state from 0 to 1, hence Y̅ changes from 0 to
1. This resulting positive voltage from Y̅ called the gating signal, is applied to input B of the
main gate thereby enabling the gate.
Now the pulses from the unknown frequency source pass through the main gate to the
counter and the counter starts counting. This same pulse from the START gate is applied to
the set input of F/F-1, changing its state from 0 to 1. This disables the START gate and
Digital Multimeter :
There are two main types of multimeters. One of the first and oldest multimeters is
the analog meter and the other, now more widely used meter is the digital multimeter .
A digital multimeter (DMM) is a multifunctional meter that displays its electrical
quantitative values on an LCD screen. A digital multimeter much like an analog meter, it is
able to read voltage, current, and resistance. What makes a digital multimeter differ from the
analog meter is its ability to display measured electrical values quickly without any
computations. Because of its design, a processor can be built into the meter which allows the
user to take measurements of frequency, the inductance of a coil, capacitance of a capacitor,
and a host of other high functional electrical measurements. There two types of digital
multimeters (DMM): scalable digital multimeter and auto-ranging digital multimeter .
When working with the scalable digital multimeter you need to have an idea of the value of
voltage, current, or resistance that you are attempting to measure. Failure to observe these
values will result in inaccurate readings and possible damage to the meter. The auto-ranging
digital multimeter is more widely used due to its ease, high functionality, and quick display
readings achieved without the user completing the calculations.
The word tachometer is derived from two Greek words: tachos mean “speed” and metron
means “to measure”. A digital tachometer is a digital device that measures and indicates the
speed of a rotating object. A rotating object may be a bike tire, a car tire or a ceiling fan,
or any other motor, and so on. A digital tachometer circuit comprises LCD or LED readout
and a memory for storage. Digital tachometers are more common these days and they provide
numerical readings instead of dials and needles.
A digital tachometer is an optical encoder that determines the angular velocity of a rotating
shaft or motor. Digital tachometers are used in different applications such as automobiles,
airplanes, and medical and instrumentation applications.
Digital tachometers are classified into four types based on data acquisition and measurement
techniques.
Based on the data acquisition technique, the tachometers are of the following types:
1. Contact type
2. Non Contact type.