4.5 Optek Brochure
4.5 Optek Brochure
4.5 Optek Brochure
Control 4000
Photometric Analyzers
inline control
Control 8000
Universal Analyzers
english
deutsch
espaol
portuguese
franais
italiano
www.optek.com
Content
C4000 / C8000 converters
03
04
06
08
09
10
11
12
14
16
18
20
22
23
24
26
System calibration
27
28
www.optek.com
C4000/C8000 - Converters I 03
Sensors
Optical sensors (optek)
C4000 C8000
1-4
1-2
pH-Probes
Communications
C4000 C8000
2/4
mA-inputs (4 - 20 mA)
0/2
Relay-outputs
Failsafe-relay (active)
Remote-IN: Zero
Remote-IN: Range
Remote-IN: Hold
Profibus PA
FOUNDATION FieldbusTM
Explosion-Proof
C4000 C8000
Ex-proof ATEX
Ex-proof FM
www.optek.com
Sensor
1
4101
4121
4151
4161
4201
4221
4251
4261
4202
4222
4252
4262
4322
4352
4362
AF16 (AS16)
AF16 (AS16)
AF16 (AS16)
AF16 (AS16)
AF26
AF26
AF45
AF45
AF45 or TF16
AF46
TF16
TF16
ASD12, 19 or 25
TF16
ASD12, 19 or 25
ASD12, 19 or 25
ASD12, 19 or 25
ASD12, 19 or 25
ASD12, 19 or 25
ASD12, 19 or 25
ASD12, 19 or 25
ASD12, 19 or 25
ASD12, 19 or 25
4422
4452
4462
Display Modes
Remote Control
1 - 4 simultaneously displayed
values (congurable)
Numeric with bar graph and alarm
setting
Trendline
Software Tools
8 parameter sets (incl. range, alarm, display, etc.)
16 linearization tables (max. 11 points)
8 offset and slope sets
Auto zero (local or remotely activated)
Factory zero (scattered light sensors only)
Password protection (3 levels and none)
Memory (non-volatile) retains all congurations and logged data
C4000 Conguration
Profibus PA
FOUNDATION FieldbusTM
www.optek.com
4221
Relay-outputs
4452
4462
4352
4362
4252
4262
4422
4251
4261
4322
Failsafe-relay (active)
4222
pH-probe
Ex-proof (optional)
PROFIBUS PA
FOUNDATION FieldbusTM
www.optek.com
VIS-NIR Absorption
AF26
AF45
UV-Absorption
AF46
Dual Channel UV
TF16
11 Scattered Light
4 Electrochemical Sensors
pH-probe
Software Tools
Remote Control
8 parameter sets
(incl. range, display etc.)
16 linearization tables
(max. 11 points)
8 offset and slope sets
Auto zero
(local or remotely activated)
Factory zero
(scattered light sensors only)
Password protection
(3 levels and none)
Memory (non-volatile) retains all
congurations and logged data
Parameter set
Zero
Display Modes
2 - 8 simultaneously displayed
values (congurable)
Numeric with bar graph
www.optek.com
1
2
8 Measurements
5 Sensors
3 Armatures
1 Converter
C8000 - Configuration
Relay-outputs
Failsafe-relay (active)
1 Converter
pH-probe
Volume: < 22 ml
Height: 96 mm (3.78 in.)
UV
Cond.
pH
C8486
5 Sensors
AF46
Dual Channel
UV Absorption
8
3
8 Measurements
UV Absorption
at 280 nm
UV Absorption
at 300 nm
Conductivity
ACF60
0 - 10 S/cm
(patented 6 electrode to 0 - 850 mS/cm
probe based on
Temperature
4 pole technique)
-10 C - 135 C (14 - 275 F)
Volume: < 44 ml
PF12
pH
Height: 96 mm (3.78 in.) (various pH
0 - 14 pH
electrodes)
Conductivity
ACF60
0 - 10 S/cm
Line size: 0.50 in.
(patented 6 electrode to 0 - 850 mS/cm
Clamp TC L14 AM7
probe based on
Temperature
PN: 0120-3508-33
4 pole technique)
-10 C - 135 C (14 - 275 F)
Volume: < 44 ml
PF12
pH
Height: 96 mm (3.78 in.) (various pH
0 - 14 pH
electrodes)
Line size: 0.50 in.
Clamp TC L14 AM7
PN: 0120-3508-33
C8480
pH
3 Armatures
Line size: 0.50 in.
Clamp TC L14 AM7
PN: 0120-3507-33
OPL: 5 mm
C8480
FOUNDATION FieldbusTM
Ex-proof (optional)
Application (example):
Profibus PA
Cond.
www.optek.com
08 I C4000/C8000 - Accessories
The PC-Transfer software allows communication between converter and PC via a
RS-232 port. Documentation and set-up
including identical set-up of multiple converters are made simple.
Converter to PC:
Parameter set
Trend data online
Data logger
PC to converter:
Parameter set
Software update
Mathematics module
(C4000 only)
C
B
A
A
S19-42
B19-42
B19-4
Wall m
mount housing (IP66)
Materia
Material: plastic (ABS)
A: 287 mm (11.3 in.)
B: 353 mm (13.9 in.)
C: 147 mm ( 5.8 in.)
D: 237 mm ( 9.4 in.)
B
C
T19-42
Front-Kit
ATEX E
EX d
Housing
Display
LCD graphic display black on white (240 x 128 pixel), LED background illuminated
Operation
18-button keyboard
System clock
LED
Data logger
Sensor-inputs
n/a
mA-inputs
optional:
2 x 4 - 20 mA (functionally galvanically isolated)
- accuracy: < 0.5 %
- resolution: < 0.05 %
- load:
< 200 Ohm
n/a
Remote-inputs
optional:
7 x 24 V (19 ... 29 V DC), typically 6.0 mA
for remote range setting, remote zero, remote hold
standard:
4 x 24 V (19 - 29 V DC), typically 6.0 mA
for remote range setting, remote zero
n/a
Profibus PA interface
C8000
Sensor lamp-outputs
mA-outputs
2 or 4 x 0/4 - 20 mA (NAMUR)
(functionally galvanically isolated)
- accuracy: < 0.5 %
- resolution: < 0.05 %
- load:
< 600 Ohm
8 x 0/4 - 20 mA (NAMUR)
(functionally galvanically isolated)
- accuracy: < 0.5 %
- resolution: < 0.05 %
- load:
< 600 Ohm
Relay-outputs
n/a
Failsafe-output
Serial communication
RS232 bi-directional interface on front panel (with software package optek PC-transfer)
- upload / download of configuration, download of data logger content
Power supply
115 / 230 V AC, selectable (93.5 - 132 / 187 - 264 V AC, 47 - 64 Hz) or
24 V AC / DC (AC: 20.4 - 26.4 V AC, 47 - 64 Hz; DC: 20.4 - 28.8 V DC)
- power consumption: < 50 VA
Ambient conditions
Software languages
www.optek.com
Technical Data
www.optek.com
Sensor Specifications
Basic Measuring Principle:
ASD
AS16
AS56
AF16
AF26
AF45
AF46
TF16
Scattering of light - 11
AS16-N
AS56-N
AF16-N
AF16-F
AS16-F
AS56-F
AF16-F
(*)
1 - 40
5 / 10
n/a
n/a
n/a
Wavelengths Used:
NIR (840 - 910 nm)
1 - 1000
1 - 160
1 - 160
40
Process Adaption:
Armature in pipe
Insertion through port
Process Ratings:
Max. pressure up to bar (psi)
(*)
20 (290)
10 (145)
(*)
100 (212)
90 (194)
120 (248)
120 (248)
70 (158)
70 (158)
120 (248)
240 (464)
240 (464)
120 (248)
120 (248)
240 (464)
VB (calibration adapter)
EX-proof ATEX
EX-proof FM
Options:
AS (e.g. AS16)
Lamp
Detector
Detector
Lamp
www.optek.com
NIR-Absorption,
single channel concentration measurement
Sensor AF16
Abs.
Abs.
Abs.
Sensor AF46
UV-Absorption,
dual channel concentration measurement
with compensation of lamp intensity
Abs.
Sensor AF45
UV-Absorption,
single channel concentration measurement
with compensation of lamp intensity
Detector(s)
Sensor AF26
VIS-Absorption,
dual channel color measurement
with turbidity compensation
Lamp
Abs.
Abs.
Sensor TF16
11
Abs.
11
2 1 2
5
4
5
7
2 1 2
2 Windows
4 Measurement detector
6 Lamp module
AF16-N
(NIR-Absorption / Turbidity)
A special tungsten lamp produces
a constant light beam that passes
through the process medium. The attenuation of the light intensity, caused
by absorption and/or scattering by
dissolved and undissolved substances,
is detected by a sealed silicon photodiode.
The AF16-N uses light from 730 - 970 nm
(NIR) to measure solids concentration
independent from color or color changes. Depending on the optical path
length, measuring ranges from high
percentage (OPL = 1 mm) to 0 - 100 ppm
(OPL = 160 mm) are possible.
7
1
3
5
7
Detectors
www.optek.com
1 Sensor body
3 Filter
5 Optics module
2 Windows
4 Detector 0 (Abs.)
6 Optics module
TF16-N
(Scattered Light / Turbidity)
Light scattered from particles (trace
suspended solids, undissolved liquids
or gas bubbles) in the medium is
detected by eight hermetically sealed
silicon photodiodes at an angle of 11.
At the same time, the unscattered light
is detected by a reference photodiode
(comparable to an AF16-N). The sensor
can be calibrated in ppm (DE), EBC or
FTU and measures extremely low particle sizes and concentrations. Additionally, high particle concentrations can
be monitored independent of color at
the direct light detector.
OPL
Special optical windows are made from
a single crystal sapphire, providing
superior resistance to all abrasive and
corrosive media. With the appropriate
choice of sensor bodies and windows
in various lengths, the optimal OPL
NIST-traceable
NIST-traceable calibration accessories
(AF16-N only) provide absolute measurement condence (for details refer to
page 27).
Typical Applications:
Separator control, pulp
concentration (AF16-N)
Filter control, oil in water (TF16-N)
optek TF16-EX-HT-N
Scattered Light Dual Channel Turbidity
Turbidi Sensor
AF16-N (Turbidity)
TF16-N (Turbidity)
Measurement principle
Measurement wavelength
Detector(s)
730 nm - 970 nm
730 nm - 970 nm
1 - 1000 mm
40 mm standard
(10 - 60 mm with reduced accuracy)
Light source
Resolution
Repeatability
Linearity
Protection
Material
Stainless steel 1.4435 (SS 316L), 1.4539, 1.4571 (SS 316Ti), 1.4462,
Titanium 3.7035 (Grade 2), Hastelloy 2.4602 (C22), Plastic TFM4215, PVC, ... others on request
Line size
Sensor Body
Process connection
Process pressure
Flanges (ASME, DIN, JIS), Clamps (TC, ISO, DIN), Female Threads (NPT, DIN),
Sanitary Threads (DIN 11851), Tube Ends (DIN, ISO, OD), Varivent, ... others on request
10 mbar to 100 bar (0.15 psi to 1450 psi) - higher on request
depending on process connection, materials and design
Windows
Window gaskets
Silicone (FDA), Viton (FDA), EPDM (FDA / USP Class VI), Kalrez 4079, ... others on request
Process temperature
permanent: -30 - 240 C (-22 - 464 F) / peak 15 min/day: -30 - 260 C (-22 - 500 F)
permanent: -30 - 120 C (-22 - 248 F) / peak 15 min/day: -30 - 150 C (-22 - 302 F)
permanent: -30 - 240 C (-22 - 464 F) / peak 15 min/day: -30 - 260 C (-22 - 500 F)
Ambient temperature
operation: 0 - 40 C ( 32 - 104 F)
operation: -30 - 40 C (-22 - 104 F) with options HT / EX / EX-HT
transport: -20 - 70 C ( -4 - 158 F)
Ex-proof
none
Sensor assembly in ex-proof version acc. ATEX (EN-D) - Approval: DMT ATEX E 176
Calibration adapter
none
n/a
n/a
Temperature Ratings
Explosion Proof
Calibration
Pressure and temperature ratings specied herein may be subject to limitations - see instruction manual.
The appropriate choice of material for all wetted parts is the sole responsibility of the user.
Data given are subject to changes without prior notice.
www.optek.com
Measurement
1- Channel absorption of light
www.optek.com
1 Sensor body
3 Filter
5 Optics module
6
2 1 2
3
9
4
1
3
5
7
9
2 Windows
4 Measurement detector
6 Lamp module
2
4
6
8
Windows
Filter A
Filter B
Optics module
2 1 2
VIS-Absorption (Color)
A special tungsten lamp produces
a constant light beam that passes
through the process medium. The attenuation of the light intensity, caused
by absorption and/or scattering by dissolved and undissolved substances, is
detected by sealed silicon photodiodes.
OPL
Special optical windows are made from
a single crystal sapphire, providing
superior resistance to all abrasive and
corrosive media. With the appropriate
choice of sensor bodies and windows
in various lengths, the optimal OPL (optical path length = distance between
the windows) can be achieved to meet
the measurement requirements, e. g.
low/high measuring ranges at highest
resolution.
Dual Wavelengths
Selected combinations of optical lters
make it possible to focus on specic
wavelengths ensuring suitable adaption to the application. While the AF16-F
uses one wavelength, an AF26 is
equipped with an internal beam splitter
making it possible to measure two
wavelengths simultaneously.
Connected to optek Control 4000 or
Control 8000 converters, the second
NIST-traceable
NIST-traceable calibration accessories
provide absolute measurement condence (for details refer to page 27).
Typical Applications:
Monitoring of various color scales
0-10 to 0-500 APHA Hazen,
30 to -16 Saybolt,
0-1 to 0-8 ASTM etc.
Measuring various concentrations
0-100mg/l chlorine,
0-5mg/l iron in hydrochloric acid,
0-100% chlorine gas,
0-10 ppm to 0-15g/l chlorine dioxide
AF16-F (Color)
AF26 (Color)
Measurement principle
385, 400, 430, 470, 525, 620, 670, 750, 1000 nm,
others on request
Measurement wavelength(s)
Detector(s)
Measuring range
Optical path length
Calibration
Light source
Resolution
Repeatability
Linearity
Protection
CU (concentration units)
application specic calibration
special incandescent tungsten lamp 5.0 V DC, 970 mA
typical life span: 3 to 5 years (25,000 to 40,000 hours)
< 0.05 % of respective measuring range
< 0.5 % of respective measuring range
< 1 % of respective measuring range (specic to application)
all optical parts have an IP rating of IP65 or higher
Sensor Body
Material
Stainless steel 1.4435 (SS 316L), 1.4539, 1.4571 (SS 316Ti), 1.4462,
Titanium 3.7035 (Grade 2), Hastelloy 2.4602 (C22), Plastic TFM4215, PVC, ... others on request
Line size
Process connection
Process pressure
Flanges (ASME, DIN, JIS), Clamps (TC, ISO, DIN), Female Threads (NPT, DIN),
Sanitary Threads (DIN 11851), Tube Ends (DIN, ISO, OD), Varivent, ... others on request
10 mbar to 100 bar (0.15 psi to 1450 psi) - higher on request
depending on process connection, materials and design
Windows
Window gaskets
Silicone (FDA), Viton (FDA), EPDM (FDA / USP Class VI), Kalrez 4079, ... others on request
Process temperature
permanent:
permanent: -30 - 240 C (-22 - 464 F) / peak 15 min/day: -30 - 260 C (-22 - 500 F)
permanent: -30 - 120 C (-22 - 248 F) / peak 15 min/day: -30 - 150 C (-22 - 302 F)
permanent: -30 - 240 C (-22 - 464 F) / peak 15 min/day: -30 - 260 C (-22 - 500 F)
Ambient temperature
operation:
0 - 40 C ( 32 - 104 F)
operation: -30 - 40 C (-22 - 104 F) with options HT / EX / EX-HT
transport: -20 - 70 C ( -4 - 158 F)
Ex-proof
none
Sensor assembly in ex-proof version acc. ATEX (EN-D) - Approval: DMT ATEX E 176
Calibration adapter
none
Temperature Ratings
0 - 120 C ( 32 - 248 F) / peak 15 min/day: 0 - 150 C ( 32 - 302 F)
Explosion Proof
Calibration
Pressure and temperature ratings specied herein may be subject to limitations - see instruction manual.
The appropriate choice of material for all wetted parts is the sole responsibility of the user.
Data given are subject to changes without prior notice.
www.optek.com
Measurement
16 I UV Sensors AF45/AF46
5
www.optek.com
1 Sensor body
3 Filter
5 Reference detector
2 1 2
9
6
4
10
1
3
5
7
9
2 Windows
4 Measurement detector
6 Lamp module (mercury)
2
4
6
8
10
Windows
Filter A
Filter B
Reference detector A
Lamp module (mercury)
2 1 2
Models AF45 and AF46 are high precision UV absorption sensors for use
in the biotech and chemical industries. The sensors are designed for
inline operation and provide accurate
concentration measurements with
remarkable repeatability, linearity and
resolution.
Modular construction of the sensors
offers maximum exibility in adapting
to various process needs. Options
include electro-polished sensor bodies,
hazardous location (explosion-proof)
capability, chemical resistant materials
(sapphire windows, titanium, Hastelloy,
etc.) and high temperature or high
pressure versions.
UV-Absorption
A special mercury lamp produces
a constant light beam that passes
through the process medium. The attenuation of the light intensity, caused
by absorption and/or scattering by dissolved and undissolved substances, is
detected by sealed silicon photodiodes.
The light intensity of the lamp itself
is measured by sealed silicon photodiodes using the same lter as the
measurement wavelength. The sealed
silicon photodiodes also compensate
for any variances in lamp intensity,
assuring the highest level of precision
and long-term performance. Specic
lamp design and capability of optek
OPL
Special optical windows are made from
a single crystal sapphire, providing
superior resistance to all abrasive and
corrosive media. With the appropriate
choice of sensor bodies and windows
in various lengths, the optimal OPL (optical path length = distance between
the windows) can be achieved to meet
the measurement requirements, e. g.
low/high measuring ranges at highest
resolution.
Dual Wavelengths
Selected combinations of optical lters
make it possible to focus on specic
wavelengths. Different peak wavelengths are available with a variety of
bandwidth options ensuring suitable
adaption to the application.
optek AF46-VB
Dual Channel UV Absorption Sensor
with Calibration Option
NIST-traceable
NIST-traceable calibration accessories
provide absolute measurement condence (for details refer to page 27).
Typical Applications:
Monitoring of chromatographic
columns (i.e. protein concentration)
Measuring concentrations of
aromatics
UV Sensors AF45/AF46 I 17
Technical Data
AF45 (UV)
AF46 (UV)
Measurement principle
254-13 / 280-13,
280-09 / 300-05,
280-09 / 313-13,
280-13 / 313-13,
others on request
Measuring detector(s)
Reference detector(s)
Measuring range
1 - 160 mm
Calibration
CU (concentration units)
application specic calibration
Light source
Resolution
Repeatability
Linearity
Protection
Material
Stainless steel 1.4435 (SS 316L), 1.4539, 1.4571 (SS 316Ti), 1.4462,
Titanium 3.7035 (Grade 2), Hastelloy 2.4602 (C22), Plastic TFM4215, PVC, ... others on request
Line size
Measurement wavelength(s)
254-13 / 313-13,
280-09 / 300-13,
280-13 / 300-13,
290-13 / 313-13 nm, ...
Sensor Body
Process connection
Process pressure
Flanges (ASME, DIN, JIS), Clamps (TC, ISO, DIN), Female Threads (NPT, DIN),
Sanitary Threads (DIN 11851), Tube Ends (DIN, ISO, OD), Varivent, ... others on request
10 mbar to 100 bar (0.15 psi to 1450 psi) - higher on request
depending on process connection, materials and design
Windows
Window gaskets
Viton (FDA), EPDM (FDA / USP Class VI), Kalrez 4079, ... others on request (do not use Silicone)
Process temperature
permanent:
permanent: -30 - 120 C (-22 - 248 F) / peak 15 min/day: -30 - 150 C (-22 - 302 F)
permanent: -30 - 70 C (-22 - 158 F) / peak 15 min/day: -30 - 135 C (-22 - 275 F)
permanent: -30 - 120 C (-22 - 248 F) / peak 15 min/day: -30 - 150 C (-22 - 302 F)
Ambient temperature
operation:
0 - 40 C ( 32 - 104 F)
operation: -30 - 40 C (-22 - 104 F) with options HT / EX / EX-HT
transport: -20 - 70 C ( -4 - 158 F)
Temperature Ratings
0 - 70 C ( 32 - 158 F) / peak 15 min/day:
0 - 135 C ( 32 - 275 F)
Explosion Proof
Ex-proof
none
Sensor assembly in ex-proof version acc. ATEX (EN-D) - Approval: DMT ATEX E 176
Calibration adapter VB
Filter adapter FH03 (detector side) for calibration lter used for sensor verication
Calibration
Pressure and temperature ratings specied herein may be subject to limitations - see instruction manual.
The appropriate choice of material for all wetted parts is the sole responsibility of the user.
Data given are subject to changes without prior notice.
www.optek.com
Measurement
www.optek.com
1 OPL
3 Lamp module
5 Detector module
5
AS16
The AS16 series offers the high-end
range of optek probe sensors. A wide
selection of different optical path
lengths and insertion depths combined
with optional calibration lters and
electro-polished stainless steel meet
all requirements of the biotechnology
industry.
AS56
The AS56, based on the same design
as the AS16 with seal-less window construction, is typically used in food and
beverage applications. Limited variations allow cost effective measurement
(i.e. phase separation).
NIR-Absorption (Turbidity)
VIS-Absorption (Color)
A special tungsten lamp produces
a constant light beam that passes
through the process medium. The attenuation of the light intensity, caused
by absorption and/or scattering by
dissolved and undissolved substances,
is detected by a sealed silicon photodi-
2 Windows
4 Optics modules (incl. lter)
OPL
Special optical windows are made from
a single crystal sapphire, providing
superior resistance to all abrasive
and corrosive media. opteks superior
manufacturing techniques allow mounting the windows without gaskets or
glue for lifetime without maintenance.
The appropriate choice of the optimal
OPL (optical path length = distance
between the windows) supports all
measurement requirements,
e. g. low/high measuring ranges at
highest resolution.
NIST-traceable
NIST-traceable calibration accessories
(AS16 only) provide absolute measurement condence (for details refer to
page 27).
Typical Applications:
Cell density in fermentation (AS16-N)
Milk/water phase separation (AS56-N)
Beverage blending (AS16-F)
Beer/water phase separation (AS56-F)
optek AS16-VB-N
Single Channel Absorption Probe with Calibration Option
AS16
AS56
Measurement principle
Detector
Measurement wavelength
AS16-N:
any measuring range between 0 - 0.05 to 6 CU
AS16-F:
any measuring range between 0 - 0.05 to 2 CU
(depending on wavelength)
1, 5, 10, 20 or 40 mm
AS56-N:
any measuring range between 0 - 0.05 to 4 CU
AS56-F:
any measuring range between 0 - 0.05 to 1.5 CU
Measuring range
5 or 10 mm
Calibration
CU (concentration units)
application specic calibration
Light source
Resolution
Repeatability
Linearity
Protection
Process Adaption
Material
Port connection
Port gasket
Insertion depth
wetted parts:
stainless steel 1.4435 (SS 316 L) dF < 1 %, BN2
surface: electro-polished Ra < 0.4 m
housing: stainless steel 1.4571 (SS 316 Ti)
thread G1-1/4 in., ISO 228/1
for port AS25 (similar Ingold-port)
diameter: 25 mm ( D= 25 H7)
O-ring groove for 30 mm and for 60 mm port length
wetted parts:
stainless steel 1.4435 (SS 316 L)
surface: electro-polished Ra < 0.8 m
housing: stainless steel 1.4571 (SS 316 Ti)
Process pressure
Windows
sapphire (seal-less)
Window gaskets
n/a
Installation accessories
weld-in ports, Varivent adapter (50.00), clamp adapter (1.5 and 2.0 in.)
optek T-pieces DIN 11850 (DN50 - DN100), optek T-pieces OD (BS4821-1) (2.0 - 4.0 in.)
Temperature Ratings
Process temperature
permanent:
peak 60 min/day:
Ambient temperature
operation:
transport:
permanent:
0 - 90 C (32 - 194 F)
peak 60 min/day: 0 - 100 C (32 - 212 F)
0 - 40 C (32 - 104 F)
-20 - 70 C ( -4 - 158 F)
Calibration
Calibration adapter
none
n/a
n/a
Pressure and temperature ratings specied herein may be subject to limitations - see instruction manual.
The appropriate choice of material for all wetted parts is the sole responsibility of the user.
Data given are subject to changes without prior notice.
Installation accessories:
Tri-Clamp
Varivent
Weld-in port 15
Weld-in port 0
T-piece
www.optek.com
Measurement
www.optek.com
Model ASD
1 OPL
3 Detector
5 LED light source
5
NIR-Absorption
A precisely dened and constant LED
light beam penetrates the process
medium. ASD12-N, ASD19-N and
ASD25-N use light from 840 nm to
910 nm (NIR). The attenuation of the
light intensity, caused by absorption by
dissolved substances and undissolved
material is detected by a hermetically
sealed photodiode.
ASD12-N
The ASD12-N probe is specically engineered for use in laboratory fermenters. The seal-less sapphire window
design eliminates crevices and gaps to
assure the highest level of sterility.
All wetted parts are electro-polished
steel. The ASD12-N is autoclavable and
mounts conveniently through any headplate with a PG13.5 threaded port.
The ASD12-N is available in three different insertion depths with either
5 or 10 mm OPLs (optical path length =
distance between the windows).
Shorter optical path lengths are typically used for dense cell cultures such
as bacterial and yeast cultures. Longer
OPLs are used for less dense cell
2 Sapphire windows
4 Daylight lter
ASD19-N
The ASD19-N, based on the same design as the ASD12-N with seal-less
window construction, is typically used
in larger fermenters or bioreactors.
Connection is through a standard
19 mm port (M26 x 1) in the fermenter
or bioreactor headplate. The ASD19-N
is available in three different insertion
depths and with OPLs of 1 mm, 5 mm,
10 mm or 20 mm.
Typical Applications:
Cell growth in mammalian cell cultures and bacterial fermentation
Algae concentration monitoring
Biomass concentrations
Crystallization process monitoring
ASD25-N / ASD25-BT-N
Models ASD25-N and ASD25-BT-N
are designed for use in pilot or production scale fermenters or bioreactors.
These extremely durable probes are
designed for sanitary bioprocess environments and can be used in CIP/ SIP
processes.
The ASD25-N is conveniently mounted
through a standard 25 mm Ingold
type port and the ASD25-BT-N through
a standard 25 mm safety port.
optek ASD12-N
Single Channel Absorption Probes
ASD12-N
ASD19-N
ASD25-N
ASD25-BT-N
Detector
1, 5, 10 or 20 mm
request
Calibration
Light source
Hybrid LED (hermetically sealed), 5.4 V DC, 100 mA, typical life span: approx 10 years
IP68
Protection
IP65
Process Adaption
Material
Port connection
Port gasket
Insertion depth
Process pressure
wetted parts:
stainless steel 1.4435 (SS 316L) dF < 1%, BN2
surface: electro-polished Ra < 0.4 m
housing: stainless steel 1.4571 (SS 316 Ti)
Fermenter head plates
diameter: 12 mm
thread: PG 13.5
110 mm + OPL
215 mm + OPL
315 mm + OPL
others on request
pressure-free
(+/- 0.5 bar) (+/- 7.25 psi)
35 mm + OPL
140 mm + OPL
with port length 60 mm
210 mm + OPL
310 mm + OPL
others on request
10 mbar to 10 bar (0.15 psi to 145 psi)
35 mm + OPL
with port length 52 mm
weld-in ports
Windows
Sapphire (seal-less)
special adapters on request
Installation accessories
permanent:
5 - 50 C (41 - 122 F)
permanent:
5 - 65 C (41 - 149 F)
peak (60 min./day): 5 - 135 C (41 - 275 F)
thermal shutdown at approximately 75 C (167 F)
weld-in ports,
Varivent adapter (50.00),
clamp adapter,
optek T-pieces DIN 11850, optek
T-pieces OD (BS4821-1)
Temperature Ratings
Process temperature
Ambient temperature
Sterilization conditions
Pressure and temperature ratings specied herein may be subject to limitations - see instruction manual.
The appropriate choice of material for all wetted parts is the sole responsibility of the user.
Data given are subject to changes without prior notice.
optek ASD19-N
Single Channel Absorption Probe
optek ASD25-N
Single Channel Absorption Probe
www.optek.com
Measurement
Measurement principle
www.optek.com
Six-Electrodes
The ACF60 conductivity sensor features
a superior six-electrode, four-pole design. The patented arrangement of the
four current electrodes around the two
potential electrodes results in a reliable
and precise measurement. This unique
design also provides greatly reduced sensitivity to sensor fouling and
polarization. The combination of optek
C8000 universal converter and ACF60
conductivity sensors allow a wide dynamic range from 0 10 S/cm up to
0 850 mS/cm with the same sensor.
Temperature Measurement
The integrated Pt1000 platinum RTD
in the tip of the ACF60 sensor provides
fast-response temperature measurement for compensation, which can be
displayed and transmitted from the
C8000 converter.
Sanitary Design
Designed for ultra-sterility, the six electrodes are sealed in the FDA (USP class
VI) compliant PEEK sensor tip without
the use of O-rings or epoxy. Mounted in
an optek inline sensor body, the design
ensures smooth and unrestricted ow
of all process uids with minimized
hold-up and hydrostatic shear. The
ACF60 sensor is suitable for CIP/SIP
applications.
Typical Applications:
Chromatography process monitoring
Monitoring of CIP processes
Measuring ultrapure water
ACF60
Conductivity Sensor
Electrodes
Port-gaskets
Line size
Process pressure
Protection
Measuring range
Accuracy
Repeatability
0 - 10 S/cm
calibrated:
1 % of measuring value 0.2 S/cm
uncalibrated: 3 % of measuring value 0.2 S/cm
0.5 %
0 - 250 mS/cm
calibrated:
1 % of measuring value 0.2 S/cm
uncalibrated: 3 % of measuring value 0.2 S/cm
0.5 %
1%
3%
Pressure and temperature ratings specied herein may be subject to limitations - see instruction manual.
The appropriate choice of material for all wetted parts is the sole responsibility of the user.
Data given are subject to changes without prior notice.
www.optek.com
improved performance
12 Degree
Sanitary Design
PF12
pH Electrode Holder
Solution Ground
The PF12 is equipped with a solution
ground (earth) connection mounted to
the sensor body. This allows the use of
pH electrodes that incorporate differential pH input techniques. Additionally,
the solution earth connection provides
a highly stable measurement while enabling sensor diagnostics including low
glass impedance, out-of sample, and
broken electrode/cable warnings.
PF12
ACF60
Volume
Volume
cond. only
pH + cond.
Material
Surface
0.25 in.
< 22 ml
< 41 ml
Solution ground
0.50 in.
< 26 ml
< 44 ml
Line size
0.75 in.
< 34 ml
< 52 ml
Electrode types
1.00 in.
< 48 ml
< 65 ml
Process temperature
Process pressure
Pressure and temperature ratings specied herein may be subject to limitations see instruction manual. The appropriate choice of material for all wetted parts is the sole
responsibility of the user. Data given are subject to changes without prior notice.
www.optek.com
S.U.C. Design
The S.U.C. is available in four different
designs. Two versions are primarily for
pre-column chromotography measurements (SUC 01/03) and two versions
for post-column measurements (SUC
05/07) with optical windows for
optek UV sensors.
Each S.U.C. is clearly labeled with data
specic to that particular cell. This data
includes the sensorr constant of the
conductivity sensor, as well as the UV
OPL (optical path length)
ngth) adaptation
data (for SUC 05/07).
7).
S.U.C. Holder
S.U.C. pH Adapter
SUC 03 and SUC 07 are equipped with
the pH adapter, which is compatible
with a broad variety of standard pH
electrodes (12 mm x 120 mm).
When no pH adapter is required (for
SUC 01 and SUC 05) a plug is used
to cover the pH input. This pH plug
has the same dimensions as the pH
electrode, keeping the hold-up
volume minimal.
8 Measurements
5 Sensors
2 Single Use Cells
1 Converter
Technical Data
SS 316 L
Measuring range
Temperature compensation
of conductivity sensor
Protection
Cleanability
Technical Data
Material (wetted)
The plastic and elastomeric wetted parts of the sensor have passed the bio-reactivity tests according to USP <87> and <88> class
VI and comply with FDA Regulations 21 CFR 177.2600. All wetted parts are of non-animal origin and any materials of animal origin
or containing animal substances are not used during manufacture. All wetted parts are of bovine-free origin and any materials of
bovine origin or containing TSE have not been used during manufacture.
Material (non wetted)
Ra < 0.8 m
Process connection
Line Size
OPL (SUC 05, SUC 07)
Hose barb
1/4 in., 3/8 in., 1/2 in., 5/8 in., ... others on request
1 mm, 2.5 mm, 10 mm, others on request
Shelf life
Process pressure
0 - 6 bar (0 - 87 psi) May be reduced in combination with pH probe. Refer to corresponding instruction manual for specications of
pH probe.
Process temperature
2 - 50 C (35.6 - 122 F) May be reduced in combination with pH probe. Refer to corresponding instruction manual for specications of pH probe.
Ambient conditions
Available Models
S.U.C. Types Conductivity pH UV
SUC 01
SUC 03
SUC 05
SUC 07
Hold-up Volume
20 ml
20 ml
Example
(OPL 1 mm): 22 ml
(OPL 2.5 mm): 23 ml
(OPL 10 mm): 25 ml
(OPL 1 mm): 22 ml
(OPL 2.5 mm): 23 ml
(OPL 10 mm): 25 ml
www.optek.com
www.optek.com
15
14
13
12
11
Sensor Body
For Variations and Details see Separate Sensor Body Datasheet
Material
Stainless steel 1.4435 (SS 316L), 1.4539, 1.4571 (SS 316Ti), 1.4462,
Titanium 3.7035 (Grade 2), Hastelloy 2.4602 (C22), Plastic TFM4215,
PVC, ... others on request
Line size
Process connection
Flanges (ASME, DIN, JIS), Clamps (TC, ISO, DIN), Female Threads (NPT,
DIN), Sanitary Threads (DIN 11851), Tube Ends (DIN, ISO, OD), Varivent,
... others on request
Process pressure
Windows
Window gaskets
10
www.optek.com
System Calibration I 27
UV Sensors
Three series of solid lters are available to ensure condence in measurements. The UV-L lter series is used
to calibrate photometric accuracy and
linearity. The UV-B lter series veries
integral blocking and the UV-S lter
series tests for long-term stability of the
sensor.
VIS/NIR Sensors
A special series of solid lters is
available for each wavelength (range)
to ensure best measurement performance. The calibration lters are used
to calibrate photometric accuracy and
linearity.
NIST-Traceable
All optek lters ship with NIST-traceable
(National Institute of Standards and
Technology) certication. The optek
laboratory is equipped with a high quality, NIST-traceable spectrophotometer
to assure quality and quick turnaround
time for recertication of lters.
Concept
Advantages of optek calibration concept include:
- Only 1 lter (set) for multiple sensors
ensures identical calibration
- Only the lter needs to be sent back
for recertication, while the sensor
remains operating
Calibration Filters UV-L
Nominal absorption:
0.45, 0.9, 1.8 and 2.4 CU
Calibration Filter UV-B
Nominal absorption:
> 3 CU
Calibration Filters UV-S
Nominal absorption:
Application specic
Calibration Filters VIS-L
Nominal absorption:
0.45, 0.9 and 1.8 CU
Calibration Filters NIR-L
Nominal absorption:
0.45, 0.9 and 1.8 CU
Calibration Case
Holds up to 7 calibration lters
Calibration Cuvette
The unique calibration cuvette,
FH03, enables product calibration
without need to interfere with the
process line.
The cuvette allows users to create a
correlation of absorption signals to
the concentration of product or an
equivalent substance, creating an
easy link from lab to process.
inline control
Germany
optek-Danulat GmbH
Emscherbruchallee 2
45356 Essen / Germany
Phone: +49 201 63409 0
Fax: +49 201 63409 999
E-Mail: info@optek.de
USA
optek-Danulat Inc.
N118 W18748 Bunsen Drive
Germantown WI 53022 / USA
Phone: +1 262 437 3600
Toll free call: +1 800 371 4288
Fax: +1 262 437 3699
E-Mail: info@optek.com
Singapore
China
optek D
optek-Danulat
Shanghai Co., Ltd.
Room 718 Building 1
No.88 Keyuan Road
Pudong Zhangjiang
Shanghai, China 201203
Phone: +86 21 2898 6326
Fax: +86 21 2898 6325
E-Mail: info@optek-danulat.cn
E-Mail: info@optek-danulat.cn
Please visit our website for contact details of our local distributors in other countries.
www.optek.com
1005-1001-02