WelchAllyn Vital Signs - User Manual
WelchAllyn Vital Signs - User Manual
WelchAllyn Vital Signs - User Manual
SYSTO
LIC (m
m
Hg)
SpO2
%
Pulse
DIAS TEMP Tone
TOLIC ERAT
(m
URE Volu
me
mHg)
PULS
E (BPM
)
Sele
Chargi ct
ng Alarm
Low B
attery
CYC
LE
Set
Pow
er
Sta
rt
Can
cel
Rev
iew
Au
to
Sile
nce
Operator’s Manual
CAUTION
4. Calibration......................................................................................................................67-68
a. Blood Pressure Calibration Check..................................................................67-68
b. Temperature Calibration Check ..........................................................................68
c. SpO2 Calibration Check ........................................................................................68
SECTION I
INTRODUCTION
1. About the Operator Manual
2. Product Overview
3. Indications/Contraindications For Use
4. Special Features
5. Supplies and Accessories
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
The information included in this manual is inclusive of all options available with the Welch
Allyn Vital Signs Monitor (ie. SpO2, Temperature and Printer). The applicability of some
sections of this operator manual will depend on the configuration of your particular unit.
The first two sections, INTRODUCTION and FUNCTIONS & SPECIFICATIONS introduce
you to the product, its applications and its capabilities. The next two sections, SYSTEM
INSTALLATION and OPERATING PROCEDURES takes you step by step through the
installation and functional operation of the monitor in a logical sequence. The final two
sections, TROUBLESHOOTING/MAINTENANCE/CALIBRATION and WARRANTY
AND SERVICE INFORMATION are resources to offer troubleshooting or other special
help as needed.
I. 2 PRODUCT OVERVIEW
The Welch Allyn Vital Signs Monitor is designed to non-invasively and automatically
measure systolic and diastolic blood pressure, pulse rate, temperature and oxygen satura-
tion (SpO2) for adult and pediatric patients over the age of 3 years. All blood pressure,
pulse, temperature and SpO2 values are displayed on large, easy-to-read displays, and
may be printed via an integrated thermal printer, as desired.
The rechargeable battery and wide variety of mounting accessories make the Welch Allyn
Vital Signs Monitor convenient for many locations. The operator may choose any combi-
nation of simultaneous measurement modalities. This flexibility, combined with features
such as programmable alarms and automatic blood pressure cycles, makes the Welch
Allyn Vital Signs Monitor ideal for a wide variety of patient monitoring needs.
The Welch Allyn Vital Signs Monitor is intended for use in a wide variety of health care
settings. This includes hospital departments, as well as patient transport within the hospi-
tal environment. The Welch Allyn Vital Signs Monitor is also intended for use in alternate
care settings, such as physician offices, freestanding ambulatory care and surgery centers,
health clinics and nursing homes. The Welch Allyn Vital Signs Monitor may also be used
during patient transport within any of these alternate care environments.
The Welch Allyn Vital Signs Monitor is not intended for the monitoring of patients during
transport external to the health care environment (eg. ambulance, helicopter transport).
The Welch Allyn Vital Signs Monitor is not intended for use in environments which are
not supervised by a health care practitioner.
1
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
The monitor is not designed for use with neonates, infants or children under the age of 3
years.
The monitor should not be used on patients who are linked to heart/lung machines.
I. 4 SPECIAL FEATURES
The following special features enhance the use of the Welch Allyn Vital Signs Monitor:
Programmable Alarms
Both visual and audible alarms indicate readings outside of operator programmable
high/low limits, and also indicate system hardware/software problems.
Non-Invasive, Oscillometric BP
Eliminates the risks associated with invasive monitoring, with no need for microphones or
external transducers.
Operator-Friendly Results
Large, easy-to-read LED displays are complemented by an integrated thermal printer.
2
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
OXICLIQ® P Pediatric Oxygen Transducer, use with OC-3 cable (case of 24)
OC-3 OXICLIQ® Sensor Cable
OXI-A/N OXIBAND® Adult/Neonatal Oxygen Transducer (1 sensor, 50 wraps)
OXI-P/I OXIBAND® Pediatric/Infant Oxygen Transducer (1 sensor, 50 wraps)
RS-10 Reflectance Oxygen Transducer (6 sensors, 6 headbands)
SRC-2 Portable Oximetry Tester
4
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
MISCELLANEOUS SUPPLIES
7052-25 Printer Paper (6 rolls/box)
5200-84 Lead-acid Battery
5200-85E Operator Manual
5200-86E Service Manual
5200-88 Plastic Monitor Covers (pack of 5)
5200-100 Welch Allyn Vital Signs Monitor Carrying Case
5200-101 AC Power Transformer (desktop transformer, line cord not included)
- North AmericanVersion
5200-102 AC Power Transformer (desktop transformer, line cord not included)
- European Version
5200-103 AC Power Transformer (desktop transformer, line cord not included)
- United Kingdom Version
5200-106 Transformer Mounting Kit (for use with the desktop transformer)
5200-110 Line Cord (United States/Canada/Japan version)
5200-111 Line Cord (European version)
5200-112 Line Cord (United Kingdom version)
5200-113 Line Cord (Australian version)
5200-114 Line Cord (Swiss version)
5200-115 Line Cord (South American version)
5
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
SECTION II
FUNCTIONS AND SPECIFICATIONS
1. Blood Pressure Operating Modes
a. Max/Min Blood Pressure and Pulse Ranges
b. Blood Pressure Manual Mode
c. Blood Pressure Auto Mode
5. Performance Specifications
6. Technical Specifications
a. Mechanical Specifications
b. Electrical Specifications
c. Environmental Specifications
d. Transportation/Storage Conditions
6
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
The cuff will immediately begin to deflate in a stepped fashion and will determine systolic
pressure and diastolic pressure from the pulses sensed by the cuff at various pressure lev-
els. This is the oscillometric method of non-invasive blood pressure monitoring.
At the completion of a measurement cycle the systolic and diastolic pressures are dis-
played. If the Monitor is in Automatic Mode, the measured values are kept on display
until the next BP measurement is initiated. When not in Automatic Mode, the measured
values are displayed for two minutes, after which time the display screen is blanked. The
most recent BP measurement may be recalled by pressing the REVIEW button.
When in Automatic Mode, if the unit is unable to determine the systolic or diastolic
value, the measurement will be automatically repeated once.
7
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
A measurement cycle may be cancelled at any time by pressing the CANCEL button. This
action immediately initiates a rapid cuff deflation.
The blood pressure measurement data will appear on the display immediately following
the measurement. The display will blank after two minutes. If the display is blanked,
pressing the REVIEW button on the front panel will recall the measurement. Up to 99
prior measurements are available for review or printing.
In Blood Pressure Manual Mode, the following alarm limits may be activated; SYSTOLIC
HIGH LIMIT, SYSTOLIC LOW LIMIT, DIASTOLIC HIGH LIMIT, DIASTOLIC LOW
LIMIT, PULSE RATE HIGH LIMIT and PULSE RATE LOW LIMIT. Blood pressure deter-
minations which activate alarms are indicated by flashing displays and a repetitive audi-
ble tone. If an alarm limit is violated, subsequent blood presssure determinations may be
made only after the alarm condition is reset by pressing any button on the Monitor's dis-
play.
Note: In the 1-minute automatic cycle mode, the Welch Allyn Vital Signs Monitor will
automatically take blood pressure measurements in 1-minute intervals for up to 15 minutes.
A measurement cycle may be cancelled at any time by pressing the CANCEL button. This
action immediately initiates a rapid cuff deflation.
The blood pressure measurement data will appear on the display immediately following
the measurement and will remain displayed until the next measurement cycle is initiated.
In Blood Pressure Auto Mode, the following alarm limits may be activated; SYSTOLIC
HIGH LIMIT, SYSTOLIC LOW LIMIT, DIASTOLIC HIGH LIMIT, DIASTOLIC LOW
LIMIT, PULSE RATE HIGH LIMIT and PULSE RATE LOW LIMIT. Blood pressure deter-
minations which activate alarms are indicated by flashing displays and a repetitive audi-
ble tone. Any alarm limit violation must be reset to continue automatically timed blood
pressure determinations. The alarm may be reset by pressing any button on the Monitor's
display.
8
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
Note: Normal mode axillary temperatures (10 seconds) are accurate only for children
under the age of four.
Note: The Welch Allyn Vital Signs Monitor is intended for use with adult and pediatric
patients, age 3 and above.
Operator selectable patient alarm limits are not available in Temperature Normal Mode.
However, temperatures which are outside of the operating range of the device will be
noted on the temperature display.
9
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
Operator selectable patient alarm limits are not available in temperature monitor mode.
However, temperatures which are outside of the normal operating range of the device will
be noted on the temperature display.
Oxygen saturation percent is calculated with each pulse detected, and thus the monitor
display is continually updated. The pulse signal bar graph is an indicator of the strength
and quality of the detected pulses.
When SpO2 is measured, the patient's pulse rate is also measured and displayed. A pulse
rate measurement from the SpO2 determination overrides a pulse rate measurement
derived from a blood pressure measurement.
When measuring SpO2 the user may enable the SpO2 Pulse Tone feature. A short audible
tone is emitted with every patient heartbeat. The frequency of the tone is based on the
patient’s SpO2 level. The lower the frequency of the tone, the lower the patients SpO2 level.
In SpO2 monitoring mode, operator selectable alarm limits for low SpO2 % may be acti-
vated. A condition which violates the SpO2 low limit alarm is indicated by a flashing dis-
play and repetitive audible tone. Should a patient alarm condition for SpO2 or pulse rate
occur, the Monitor will indicate an alarm condition (flashing & beeping) while continuing
to monitor and display the patient's current SpO2%. The alarm will automatically reset
when the patient's condition returns to within the preset alarm parameters.
Should a patient alarm condition for SpO2 or pulse rate occur, the operator may invoke
"Silence Mode" by pressing the SILENCE button. This will silence the audible tone (dis-
play will continue to flash), while the practitioner attends to the patient and the monitor.
Silence mode resets automatically after 30 seconds, or when the patient's condition returns
to within the preset alarm parameters.
10
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
Removal of the SpO2 sensor from the patient will initiate an alarm, unless the SpO2 and
pulse rate alarms are turned off. To reset the sensor alarm, press any button on the
Monitor's display.
SpO2 is generally measured via pulses detected using a finger sensor. However, for cer-
tain situations SpO2 may be measured at alternate sites including the earlobe, forehead
and toes. Special sensors must be employed in these situations.
Maximum Minimum
All Welch Allyn Vital Signs Monitors with pulse oximetry capability are equipped with a
pulse tone. This audible tone beeps in synchronization with each beat of the patient’s
heart. In addition, the frequency of the tone will vary based on the patient’s oxygen satu-
ration value. The higher the tone’s frequency, the higher the patient’s oxygen saturation
value. The user may change the pulse tone volume, or turn this feature off as necessary.
11
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
The pulse rate, in beats per minute, will be determined primarily from the SpO2 measure-
ment methodology. In the case where SpO2 is not available, or is disabled, the pulse rate
display will be driven by data from the blood pressure measurement method.
There are two operator selectable alarm limits for the pulse rate. They are PULSE RATE
HIGH LIMIT and PULSE RATE LOW LIMIT. Pulse rates which activate alarm limits are
indicated by a flashing display and a repetitive audible tone.
Should a pulse rate alarm occur when the pulse rate measurement is derived from the
blood pressure measurement, no subsequent blood pressure or pulse rate measurements
may be made until the alarm is reset. The alarm may be reset by pressing any button on
the Monitor's display.
Should a pulse rate alarm occur when the pulse rate measurement is derived from the
SpO2 measurement, the monitor will indicate an alarm condition (flashing & beeping)
while continuing to monitor and display the patient's current SpO2% and pulse rate. The
alarm will automatically reset when the patient's condition returns to within the preset
alarm parameters.
Should a patient alarm condition for pulse rate occur during SpO2 monitoring, the oper-
ator may invoke "Silence Mode" by pressing the SILENCE button. This will silence the
audible tone (display will continue to flash), while the practitioner attends to the patient
and Monitor. Silence mode resets automatically after 30 seconds, or when the patient's
condition returns to within the preset alarm parameters. If an alarm limit is violated, sub-
sequent determinations of any type may only be made after the alarm condition is reset.
Maximum Minimum
Pulse Rate 200 bpm 40 bpm
12
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
SYSTOLIC DETERMINATION
Maximum: 250 mmHg
Minimum: 60 mmHg
DIASTOLIC DETERMINATION
Maximum: 160 mmHg
Minimum: 30 mmHg
OVERPRESSURE CUTOFF
295 mmHg to 330 mmHg
TEMPERATURE DETERMINATION
Normal and Monitor Modes: Maximum 108.0°F (42.2°C)
Minimum 84.0°F (28.9°C)
TEMPERATURE ACCURACY
Temperature accuracy meets ASTM E112-86: "Standard Specification for Electronic
Thermometer for Intermittent Determination of Patient Temperature."
13
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
SpO2 ACCURACY
Nonin® Pulse Oximeter Module
Saturation (%SpO2 ± 1 Standard Deviation*)
70-99% ± 2 digits
< 70% unspecified
* Standard Deviation is a statistical measure: up to 32% of the readings may fall outside
these limits.
BATTERY CHARGING
To at least 90% capacity in 12 hours. Unit will operate and charge battery simultaneously
when connected to power source.
A fully charged battery will support 200 "typical" blood pressure determinations taken at 3
minute intervals. Battery is 90-100% charged after 12 hours of charging. The battery auto-
matically charges when the monitor is powered through the AC power transformer. The
battery will charge faster when the instrument is not in operation.
WEIGHT
Approximately 6 pounds (2.8Kg)
COLOR
Oral/Axillary Temperature Probe - Blue
Rectal Temperature Probe - Red
14
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
MOUNTING
Self-Supporting on rubber feet
IV Pole Mountable
Custom Mobile Stand
Wall Mountable
Attaches to Bed Rail
PORTABILITY
- May be hand carried when held by the recessed handle.
- When attached to an IV pole, or mounted on its custom mobile stand, the
monitor and accessories can be wheeled from patient to patient.
- When attached to the bed rail may be transported with the patient within the
hospital environment.
BATTERY
Lead acid, with external recharge capability.
OPERATING ALTITUDE
-170m to +4877m
-557ft. to +16,000ft.
RELATIVE HUMIDITY
15 to 90% (non-condensing)
15
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
SECTION III
SYSTEM INSTALLATION
1. Unpacking Checklist
2. Controls, Indicators and Connections
3. Set Up Procedure
a. Blood Pressure Hose & Cuff Connections
b. Temperature Probe Connection
c. SpO2 Sensor Connection
d. AC Power Connection
e. Charging the Battery
f. Setting Up the Printer
g. Changing the Time and Date Setting
4. Safety Warnings and Cautions
All Welch Allyn Vital Signs Monitors include the following components (A-G):
16
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
Note: A full range of cuff sizes are available (as accessory items) however, the
Large Adult cuff will fit the majority of adults and give the most accurate blood
pressure measurement.
G. Cuff Clip
Attaches to the rear of the monitor to hold the blood pressure cuff. Not used if
mobile stand, pole mount or wall mount accessories are purchased.
Certain Welch Allyn Vital Signs Monitors will include the following items, based on the
options purchased:
PRINTER OPTION
H. Box of Printer Paper
Six (6) rolls of thermal printer paper. One additional roll will be pre-loaded into the
printer.
TEMPERATURE OPTION
J. Oral Temperature Probe
The blue oral probe is used for both oral and axillary temperature determinations.
A red rectal probe is available separately.
Note: Report any signs of shipping damage to the carrier. If an item is missing or dam-
aged, contact the Welch Allyn Service Center near you. See section VI of this manual for
more information
17
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
M N P R S I
Pulse Volume
J
K Tone
Charging
L DIASTOLIC (mmHg) PULSE (BPM) Low Battery
CYCLE
Set H
Q
T
O Power Start Cancel Review Auto Silence
A B C D E F
FRONT PANEL
B. START Button
Pressing this button initiates an on-demand blood pressure cycle.
C. CANCEL Button
If a blood pressure cycle is in progress, pressing this button aborts the cycle and
immediately releases the cuff pressure. In "STAT" automatic blood pressure mode,
pressing the cancel button aborts the cycle in process, and cancels the "STAT" mode
as well.
When pressed for 3 seconds, the cancel button erases all stored data in the Welch
Allyn Vital Signs Monitor.
18
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
D. REVIEW Button
Pressing this button will display the most current set of stored data including BP &
pulse rate and/or SpO2 and/or temperature. Subsequent presses will display the
next most recent data sets.
K. SYSTOLIC Display
This red LED display shows the systolic blood pressure.
Additionally, this display can show systolic alarm limits (see G, and H above).
L. DIASTOLIC Display
This red LED display shows the diastolic blood pressure
Additionally, this display can show diastolic alarm limits (see G, and H above).
M. SpO2 Display
This red LED display shows the percent saturation of arterial hemoglobin (SpO2%).
Additionally, this display can show SpO2 low saturation alarm limits
(see G, and H above).
19
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
O. PULSE Display
This yellow LED display shows the pulse rate.
Additionally, this display can show pulse rate alarm limits (see G and H above).
Q. CYCLE Display
This red LED display shows the time (in minutes) between automatically initiated
blood pressure measurements, or "St" for STAT mode (continuous blood pressure
measurements for up to 15 minutes).
Additionally, this display identifies the high and low alarm limits for the various
parameters, during alarm select mode, and the cycle number during review mode.
R. CHARGING Indicator
This flashing yellow indicator will signify that the internal battery is being charged
when the Monitor is powered through the AC power transformer. When the Monitor
is fully charged, the flashing will be replaced with a solid yellow indicator light.
T. MEMORY Indicator
This green indicator will flash when the unit has reached its maximum capacity of
99 data sets in storage. Also, it will remain on continuously when the operator is
reviewing stored data sets.
20
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
4
7
1
5
2
11
6
10
3
7. Printer
Compartment and paper feed area for integrated thermal printer.
8. Print Button
Pressing this button initiates a print operation.
21
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
Using the illustrations below for reference, perform the following set-up procedures:
1. Inspect the black coiled pressure hose and note that one end has a connector fitting
and the other end does not. Attach the end without the connector to the pressure
hose connector on the monitor by fitting the pressure hose on to the connector as
shown. Be sure that the pressure hose is completely inserted over the connector and
into the recess, and that the fit is snug.
22
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
2. Join the other end of the black coiled pressure hose to the pneumatic tubing which is
attached to the cuff. Twist the black plastic connectors together until finger-tight. DO
NOT OVER TIGHTEN.
23
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
To install the temperature probe, press down on the tab on top of the connector and insert
the connector into the temperature probe connector port on the side of the Monitor until
the connector clicks into place. The temperature connector port on the Monitor is clearly
labeled "TEMP." The probe connector can only be inserted one way, with the tab on top.
The temperature probe should be inserted into the probe holder on the side of the Monitor.
Sele
Chargin ct
g Alarm
Low Ba
ttery
CYC
LE
Set
Pow
er
Sta
rt
Can
cel
Rev
iew
Au
to
Sile
nce
24
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
For Welch Allyn Vital Signs Monitor with the Nonin® Pulse Oximetry Option:
To install the Nonin SpO2 sensor, insert the connector end of the sensor into the SpO2 con-
nector port on the side of the Monitor as shown. The SpO2 connector port on the Monitor
is clearly labeled "SpO2." The sensor connector can only be inserted one way, matching
the shape and pin configuration of the connector to the port. Push the connector in until it
is fully seated on the port. Note: only Nonin® SpO2 sensors and accessories may be used
with this configuration of the Welch Allyn Vital Signs Monitor.
For Welch Allyn Vital Signs Monitor with the Nellcor Puritan BennettTM Pulse
Oximetry Option:
Attach the Nellcor Puritan Bennett SpO2 sensor to the pulse oximetry extension cable.
Then insert the connector end of the extension cable into the SpO2 connector port on the
side of the Monitor as shown. The SpO2 connector port on the Monitor is clearly labeled
“SpO2”. The extension cable can only be inserted one way; matching the shape and pin
configuration of the connector to the port. Push the connector in until it is fully seated on
the port. Note: Only Nellcor Puritan BennettTM SpO2 sensors and accessories may be used
with this configuration of the Welch Allyn Vital Signs Monitor.
25
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
To install the AC power transformer, insert the round transformer connector into the
power port on the side of the Monitor as shown. Insert the connector into the port until it
is fully seated. Insert the line cord into the line connector on the transformer.
To power the Welch Allyn Vital Signs Monitor, plug the line cord into the AC mains
power source.
UPON INITIALLY RECEIVING THE MONITOR, CHARGE THE BATTERY FOR SIX-
TEEN (16) HOURS OR UNTIL THE CHARGING INDICATOR LIGHT NO LONGER
FLASHES. The battery is charged by attaching the AC power transformer to the Monitor,
and plugging the AC power transformer into the AC mains power source.
While charging, the charging indicator will flash a yellow light. When the Monitor is fully
charged the flashing light will be replaced with a solid yellow indicator light.
26
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
1. Grasp the printer cover as shown, and slide the cover off of the Monitor by sliding
the cover towards you.
2. Holding a new roll of paper as shown, feed approximately 1 inch of paper into the
slot at the bottom of the paper holder. Release the paper roll so that it sits in the bot-
tom of the paper holder.
27
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
3. Press the paper "FEED" button to advance the paper 3 to 4 inches above the top
paper slot.
4. Thread the paper through the paper slot on the printer cover. Slide the cover on to
the Monitor and press down until it clicks into place as shown. The excess paper
may be torn off if desired.
Note: The operator will notice a red stripe along the edge of the printer paper when the
paper roll is almost at the end. Replace the printer paper with a new roll when the red
stripe is observed.
28
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
2 Press the review button four (4) times to advance to the Date Set Screen. The year,
month and day will appear in the systolic, diastolic and heart rate displays respec-
tively.
3 Use the SELECT ALARM button to select the date item to be changed. When select-
ed, the date item will flash.
4 Use the SET buttons (arrow up or arrow down) to change the selected date item.
5 After making all the desired date changes, press the REVIEW button to save the
changes and advance to the Time Set Screen.
6 When in the Time Set Screen the hour (in 24 hour format) and minutes will appear in
the systolic and diastolic displays respectively. Use the SELECT ALARM and SET
buttons to set the time (in the same manner as described above).
7 When the time is set as desired, press the REVIEW button to save the time and
advance to the next screen.
29
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
Conforms to
IEC 60601-1
Attention, Consult
UL 2601-1 Accompanying Document
CAN/CSA C22.2 No 601-1
Approved to Australia AS 3200.1,
Appendix Z
GENERAL WARNINGS
• The Welch Allyn Vital Signs Monitor is designed for use by medical clinicians.
Although this manual may illustrate medical monitoring techniques, this system
should only be used by a trained clinician who knows how to take and interpret a
patient's vital signs.
• The information in this operator's manual is a comprehensive guide to the operation
of the Welch Allyn Vital Signs Monitor. To achieve satisfactory results the operator
should read the manual thoroughly before attempting to use the Monitor.
• To insure patient safety, use only accessories and supplies (i.e. cuffs, hoses, tempera-
ture probes, SpO2 sensors etc.) recommended or supplied by Welch Allyn for the
Welch Allyn Vital Signs Monitor.
• Do not operate the Welch Allyn Vital Signs Monitor in the presence of flammable
anesthetics or other explosive atmosphere. An explosion may result.
• It is the operator’s responsibility to set alarm limits as appropriate for each individual
patient.
• Avoid compression of the pneumatic tubing of the Welch Allyn Vital Signs Monitor.
Compression of the tubing may cause system errors to occur in the Monitor.
• The Welch Allyn Vital Signs Monitor is safe to use in the presence of high frequency
surgical equipment
• Care should be taken to prevent water or other fluid from entering any connectors on
the Monitor. Should this occur, the connectors should be dried with warm air. All
monitoring functions should then be checked for proper operation.
30
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
• Any Monitor which has been dropped or damaged should be checked by qualified
service personnel to insure proper operation prior to use.
• Cords should periodically be checked for fraying or other damage, and replaced as
necessary.
• There are no user serviceable parts inside the Monitor other than paper replacement
and battery replacement.
• This Monitor is not designed for use with neonates, infants or children under the age
of 3 years.
• This Monitor should not be used on patients who are linked to heart/lung machines.
• This Monitor will not operate effectively on patients who are experiencing convul-
sions or tremors.
• This device complies with current required standards for electromagnetic interference
and should not present problems to other equipment or be affected by other devices.
As a precaution, avoid using this device in close proximity to other equipment.
• Blood pressure measurements may not be accurate for patients experiencing moder-
ate to severe arrhythmias.
SpO2 WARNINGS
• The operation of the SpO2 sensor in MRI environments is specifically not recom-
mended.
• Use only SpO2 sensors and accessories which are compatible with the SpO2 configu-
ration purchased. Welch Allyn Vital Signs Monitor with the Nonin® pulse oximetry
option may only be used with Nonin® brand SpO2 sensors and accessories. Welch
Allyn Vital Signs Monitor with Nellcor Puritan BennettTM pulse oximetry option may
only be used with Nellcor Puritan BennettTM brand sensors and accessories.
TEMPERATURE WARNINGS
• Single-use, disposable probe covers, available from Welch Allyn will limit patient
cross-contamination. The use of any other probe cover or the failure to use a probe
cover may produce temperature errors and is specifically not indicated.
• Oral probes (blue) are to be used for taking oral and axillary temperatures only.
Rectal probes (red) are to be used for taking rectal temperatures only. The use of the
wrong probe may produce temperature errors.
31
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
GENERAL CAUTIONS
• If the accuracy of any measurement is in question, check the patient's vital sign(s) by
an alternate method and then check the Monitor for proper functioning.
• Insure the Monitor is placed on a secure surface or use one of the optional mounting
accessories.
• If the blood pressure cuff is not at heart level, the difference in reading due to the
hydrostatic effect should be noted. The value of 1.80 mmHg must be added to the
displayed reading for every inch (2.5cm) above heart level. The value of 1.80 mmHg
must be subtracted from the displayed reading for every inch (2.5cm) below heart
level.
• Proper blood pressure cuff size and placement is essential to the accuracy of the BP
determination. See Section IV.7 for cuff sizing information.
SpO2 CAUTIONS
• The pulse oximeter is calibrated to determine the percentage of arterial oxygen satu-
ration of functional hemoglobin. Significant levels of dysfunctional hemoglobins
such as carboxyhemoglobin or methemoglobin may affect the accuracy of the mea-
surement.
• Some intravascular dyes, depending on the concentration, may affect the accuracy of
the SpO2 measurement.
• Some sensors may not be appropriate for a particular patient. If at least 10 seconds
of perfusion pulses cannot be observed for a given sensor, change sensor location or
sensor type until this condition is achieved.
TEMPERATURE CAUTIONS
• Do not allow the tip of the temperature probe to come into contact with a heat source
(i.e. hands or fingers) prior to taking a temperature measurement. If this occurs dis-
card the probe cover and start the temperature determination process again.
• Normal Mode (10 second) axillary temperatures are accurate only for children under
the age of four. Normal Mode axillary temperatures will not be accurate on older
children or adults. The Welch Allyn Vital Signs Monitor is intended for use with
adult and pediatric patients, age 3 and above.
32
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
SECTION IV
OPERATING PROCEDURES
1. Power On/System Check Procedure
2. Choosing Operating Modes
3. Setting the Programmable Alarms
a. Blood Pressure - High Systolic Limit
b. Blood Pressure - Low Systolic Limit
c. Blood Pressure - High Diastolic Limit
d. Blood Pressure - Low Diastolic Limit
e. Pulse Rate - High Limit
f. Pulse Rate - Low Limit
g. SpO2 - Low Limit
33
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
34
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
Warning: It is the operator’s responsibility to set alarm limits as appropriate for each
individual patient.
Note: Temperature default alarms are 108°F, 42.2°C (high) and 84.0°F, 28.9°C (low). The
temperature alarms are not programmable, they indicate that the patient's temperature
has exceeded the functional range of the thermometer.
*The unit will default to these factory settings only until the operator selects a different
alarm limit. When a new alarm limit is selected, the Monitor will retain this new limit set-
ting in memory, even if the Monitor is turned off.
Alarm limits may be reviewed or changed by using the SELECT ALARM and SET (arrow
up and arrow down) buttons, and viewing the CYCLE display and the display of the
parameters being changed.
The Systolic display will show the current alarm limit for low systolic blood pressure in
mmHg, or "---" if the alarm is turned off. Pressing the SET (arrow up and arrow down)
buttons will cycle through the options in 5 mmHg intervals. The operator may choose the
low systolic blood pressure limit within the range of the current high systolic alarm limit
through 65mmHg. The alarm may be turned off by choosing "---."
When the desired alarm limit is displayed, pressing SELECT ALARM again saves the dis-
played alarm limit and changes the alarm parameter. Alternatively, pressing no buttons
for approximately 10 seconds will cause the displayed value to be set as the alarm limit,
and the Monitor will automatically revert to the current operating mode.
The diastolic display will show the current alarm limit for high diastolic blood pressure in
mmHg, or "---" if the alarm is turned off. Pressing the SET (arrow up and arrow down)
buttons will cycle through the options in 5 mmHg intervals. The operator may choose the
high diastolic blood pressure limit within the range of the current low diastolic limit
through 155mmHg. The alarm may be turned off my choosing "---."
When the desired alarm limit is displayed, pressing SELECT ALARM again saves the dis-
played alarm limit and changes the alarm parameter. Alternatively, pressing no buttons
for approximately 10 seconds will cause the displayed value to be set as the alarm limit,
and the Monitor will automatically revert to the current operating mode.
36
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
The diastolic display will show the current alarm limit for low diastolic blood pressure in
mmHg, or "---" if the alarm is turned off. Pressing the SET (arrow up and arrow down)
buttons will cycle through the options in 5 mmHg intervals. The operator may choose the
low diastolic blood pressure limit within the range of 35mmHg through the current high
diastolic limit. The alarm may be turned off by choosing "---."
When the desired alarm limit is displayed, pressing SELECT ALARM gain saves the dis-
played alarm limit and changes the alarm parameter. Alternatively, pressing no buttons
for approximately 10 seconds will cause the displayed value to be set as the alarm limit,
and the Monitor will automatically revert to the current operating mode.
The pulse display will show the current alarm limit for high pulse rate in bpm, or "---" if
the alarm is turned off. Pressing the SET (arrow up and arrow down) buttons will cycle
through the options in 5 bpm intervals. The operator may choose the high pulse rate limit
within the range of the current low pulse rate limit through 195bpm. The alarm may be
turned off by choosing "---."
When the desired alarm limit is displayed, pressing SELECT ALARM again saves the dis-
played alarm limit and changes the alarm parameter. Alternatively, pressing no buttons
for approximately 10 seconds will cause the displayed value to be set as the alarm limit,
and the Monitor will automatically revert to the current operating mode.
37
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
The pulse display will show the current alarm limit for low pulse rate in bpm, or "---" if
the alarm is turned off. Pressing the SET (arrow up and arrow down) buttons will cycle
through the options in 5 bpm intervals. The operator may choose the low pulse rate limit
within the range of 45bpm through the current high pulse rate alarm limit. The alarm
may be turned off by choosing "---."
When the desired alarm limit is displayed, pressing SELECT ALARM again saves the dis-
played alarm limit and changes the alarm parameter. Alternatively, pressing no buttons
for approximately 10 seconds will cause the displayed value to be set as the alarm limit,
and the Monitor will automatically revert to the current operating mode.
The SpO2 display will show the current alarm limit for the SpO2% or "--" if the alarm is
turned off. Pressing the SET (arrow up and arrow down) buttons will cycle through the
options in 1% intervals. The operator may choose the low SpO2% limit within the range
of 98-70%, or the alarm may be turned off by choosing "--."
When the desired alarm limit is displayed, pressing SELECT ALARM again saves the dis-
played alarm limit and changes the alarm parameter. Alternatively, pressing no buttons
for approximately 10 seconds will cause the displayed value to be set as the alarm limit,
and the Monitor will automatically revert to the current operating mode.
38
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
IMPORTANT: No audible tone will indicate that the temperature is outside of the
measurement range of the Monitor. This is a visual indicator only.
Temperatures which are outside of the measurement range of the Monitor will cause the
following to be displayed. Note the symbol shown in the tenths position of the display.
39
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
The Welch Allyn Vital Signs Monitor handles patient alarms in the following manner:
NOTE: Pulse rate measurements from the SpO2 determination override pulse rate mea-
surements from the blood pressure determination.
Note: Certain alarms alert the operator to abnormal system conditions or internal sys-
tem failures. Alarm conditions, error codes and suggested operator action are reviewed
in the Troubleshooting section (Section V.1.) of this manual.
40
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
Note: In Manual Blood Pressure Operating Mode, the Monitor will inflate to the default
pressure preset level. However, if a BP measurement is initiated within one minute of a
prior BP measurement, the Monitor will inflate to a level which is 40 mmHg above the
previous systolic reading.
In Automatic Blood Pressure Operating Mode, the Monitor will inflate to default pressure
preset level for the initial BP determination only. After the initial measurement, the
Monitor will inflate to a level which is 40 mmHg above the previous systolic reading.
Research has shown that an undersized cuff will overestimate the true blood pressure by
as much as 10 to 30 mmHg. Please refer to the reference markings located inside the cuff
for correct cuff sizing. When there is an area of overlap whereby you could use a smaller
or larger cuff, it is strongly recommended that you use the larger size cuff.
You may find that the bottom of the cuff extends to the antecubital fossa (bend in the
elbow) on many people, but because the Monitor uses oscillometric technology, not aus-
cultation, this will NOT result in an inaccurate blood pressure.
Careful sizing of the cuff is important to the accuracy of blood pressure readings. If the
cuff is too small, the readings could be falsely high.
41
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
There are two simple methods for determining the correct cuff size for the patient. Use
either method (A) or (B) below.
➡ RANGE
➡ ADULT
IND
EX ➡
AR
TE
RY
31.5
cm
MA
XS
IZE
42
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
Warning: Do not place the cuff on any extremity being used for intravenous infusions, or
any area where circulation is or may be compromised.
Warning: Do not place the cuff on any extremity being used for SpO2 monitoring. Cuff
inflation during an SpO2 measurement will cause inaccurate SpO2 results.
Wrap the cuff snugly with room between the cuff and the arm for two fingers. Excessive
tightness may cause venous congestion and discoloration of the limb. If the cuff is wrapped
too loosely, it cannot be inflated properly and the measured values may be in error.
It is best to place the cuff on a bare arm. Clothing may interfere with measurement accuracy.
When wrapping the cuff, observe the mark on the cuff which must be placed over the
artery. Insure that the hose is not twisted, kinked or compressed, as this may cause mea-
surement errors.
To operate the blood pressure monitor in manual mode, follow this procedure:
1. Insure that the blood pressure cuff is properly sized and wrapped around the
patient's upper arm (or alternate site, as necessary).
2. With the Monitor powered up, press the START button. The Monitor will inflate the
cuff to the default pressure preset level.
3. The systolic display will show the pressure in the cuff as the blood pressure determi-
nation is being made.
4. When the measurement cycle is complete an audible tone will sound and the systolic,
diastolic and pulse rate* display will show the values of the reading.
5. The reading will be displayed for two minutes, and then this display will be blanked.
The reading may be recalled at any time prior to the next BP determination by press-
ing the REVIEW button.
6. Pressing CANCEL at any time during a blood pressure determination will cause the
determination to be aborted, and the cuff to rapidly deflate.
43
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
*Pulse rate, as determined from the blood pressure measurement method, will be displayed
with the BP reading only if SpO2 option is absent or disabled. If the SpO2 function is opera-
tional, all pulse rate determinations will come as a result of the SpO2 measurement method.
To operate the blood pressure monitor in automatic ("Auto") mode, follow this procedure:
1. Insure that the blood pressure cuff is properly sized and wrapped around the
patient's upper arm (or alternate site, as necessary).
2. Press the AUTO button and note two dashes "--" in the CYCLE display. This is the
default setting for AUTO mode and indicates that no time interval cycles have been
selected.
3. Select one of the ten pre-programmed cycle interval times by pressing the AUTO but-
ton until the desired interval time (in minutes) is displayed. The operator may
choose from the following time interval selections: 1, 3, 4, 5, 10, 15, 30, 45, 60 or 90
minutes. Note: These intervals represent the time from the beginning of one cycle to
the beginning of the next cycle. Note: In 1-minute automatic cycle mode, the Monitor
will automatically take BP measurements in 1 minute intervals for up to 15 minutes.
The operator may also choose "St", STAT mode. In STAT mode the Monitor takes
blood pressure readings repeatedly for up to 15 minutes. A new measurement cycle
will be initiated when the cuff pressure from the previous cycle drops below
10mmHg.
4. When the desired interval is displayed, the operator may select the interval by press-
ing any other button or by refraining from pushing the AUTO button for 10 seconds.
When the interval is stored, the AUTO indicator light will glow green.
5. The unit will take its first automatic blood pressure determination 10 seconds after
the time interval is selected. Subsequent readings will be taken according to the time
interval selected.
6. For the first automatic blood pressure determination, the Monitor will inflate the cuff
to the default pressure preset level. For all subsequent blood pressure determina-
tions, the Monitor will inflate the cuff to a pressure 40 mmHg above the prior systolic
reading.
7. The systolic display will show the pressure in the cuff as the blood pressure determi-
nation is being made.
Exception: In STAT mode, the systolic and diastolic displays will show the most current
completed BP measurement.
44
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
8. When the measurement cycle is complete an audible tone will sound and the systolic,
diastolic and pulse rate* displays will show the values of the reading.
9. The measurement will be displayed until the next measurement cycle is initiated.
10. Pressing CANCEL at any time during a blood pressure determination will cause the
determination to be aborted, and the cuff to deflate. In STAT mode, pressing CAN-
CEL will also terminate the STAT mode of operation.
11. To end an AUTO blood pressure session, press the AUTO button until two dashes "--"
appear in the CYCLE display. Auto mode will then be terminated. STAT mode may
be terminated in this manner, or by pressing the CANCEL button.
*Pulse rate, as determined from the blood pressure measurement method, will be dis-
played with the BP reading only if the SpO2 option is absent or disabled. If the SpO2
function is operational all pulse determinations will come as a result of the SpO2 measure-
ment method.
45
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
Blood Pressure Data: will be saved for both MANUAL and AUTOMATIC cycles.
Monitor Mode Temperature Data: will be saved only if concurrent with a blood pressure
cycle.
SpO2 Data: will be saved if concurrent with a blood pressure cycle, or a predictive tem-
perature measurement. Will be saved if the START button is pressed while the BP mode is
inactive. Will also be saved (one data set per minute) if a patient alarm condition occurs.
Pulse Rate Data: will be saved with every blood pressure and SpO2 data set.
To review data from prior vital signs measurements, press the REVIEW button. The most
recently obtained data will appear in the appropriate displays, and the CYCLE display
will show the total number of data sets currently in memory. Also, the MEMORY indica-
tor will be illuminated, indicating that stored data is being displayed.
Each subsequent press of the REVIEW button will display the next most recent data set.
After the earliest data set has been displayed (CYCLE display = 1) the next press of the
REVIEW button will return the display to the most recent set of measurements. At any
time during REVIEW the operator may return to normal operating mode by pressing any
other button or by refraining from pressing the REVIEW button for 10 seconds.
To erase all stored data in the Welch Allyn Vital Signs Monitor, press and hold the CAN-
CEL button for three (3) seconds until a confirmatory beep is heard. This beep indicates
that all data stored in the Monitor has been erased.
Data is also erased each time the Welch Allyn Vital Signs Monitor is turned off.
46
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
To determine the current temperature scale, remove the temperature probe from its holder
and view the TEMPERATURE display. A small green display will illuminate showing
either "°F" or "°C."
To change the temperature scale, the user must enter the Monitor's internal configuration
mode:
1. While the Monitor is turned off, press both the POWER and START buttons simulta-
neously. Hold the START button down until all of the LED display segments are illu-
minated.
2. The operator can now cycle through the Monitor's internal configuration menu by
pressing the REVIEW button. Pressing the REVIEW button three times will cycle to
the temperature options and three dashes (---) will appear in the TEMPERATURE
display.
3. The first option illuminated on the TEMPERATURE display is "°F." Pressing the
ADJUST button twice will illuminate "°C." When the desired temperature scale is
selected, press the REVIEW button again and then turn the Monitor off. By follow-
ing this procedure, the selected temperature scale will be saved into memory.
47
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
Monitor mode is normally used for longer term monitoring and when difficult situations
prevent accurate temperature from being taken in the Normal mode. In Monitor mode,
the probe must be in contact with tissue for at least 3 minutes for accurate temperature
measurement.
Note: Normal mode axillary temperatures are only accurate on children 4 years or
younger.
Note: The Welch Allyn Vital Signs Monitor is intended for use with adult and pediatric
patients, age 3 and above.
To change the operating mode for temperature determinations, the user must enter the
Monitor's internal configuration mode:
1. While the Monitor is turned off, press both the POWER and START buttons simulta-
neously. Hold the START button down until all of the LED display segments are
illuminated.
2. The operator can now cycle through the Monitor's internal configuration menu by
pressing the REVIEW button. Pressing the REVIEW button three times will cycle to
the temperature options, and three dashes (---) will appear in the TEMPERATURE
display.
3. Press the ADJUST button once to choose Monitor mode (°F), or three times to choose
Monitor mode (°C). When the desired selections are made, press the REVIEW but-
ton again and then turn the Monitor off. These choices will be saved into memory.
Note: If the temperature is in Normal mode, the user may easily switch to Monitor mode
without entering the Monitor's internal configuration mode. To do this, remove the probe
from the probe holder, attach a new probe cover, and wait one minute (do not place probe
in patients mouth, underarm or rectum at this time). After one minute the Monitor will
automatically switch to temperature Monitor mode, and the green "Monitor Mode" dis-
play will be illuminated on the temperature display. The operator may now proceed to
take the patient's temperature. After the probe is replaced in the holder, the Monitor will
revert back to Normal temperature mode.
48
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
1. Insure that the oral probe is connected to the unit. The oral probe has a BLUE tip.
Accurate oral temperatures can only be obtained by using the blue temperature
probe.
2. Remove the probe from the probe holder. A short self-test mode will be initiated
where every LED segment on the TEMPERATURE display is illuminated briefly.
Following this self-test the display will show "OrL". "OrL" = Oral Probe indicating
that the oral probe is in use. The Monitor’s display must show “OrL” prior to the
initiation of a temperature measurement.
3. Load a probe cover onto the probe by holding the probe collar with the thumb and
forefinger, being careful not to hold or press the ejection button. See illustration
below.
49
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
4. Insert the probe tip gently into the patient's slightly opened mouth. Carefully slide
the probe under the tongue on either side of the mouth to reach the sublingual pock-
et (see illustration below). Accurate temperatures can only be obtained in this loca-
tion. Temperatures in other mouth locations can vary by as much as 2°F or 1°C.
5. The probe should be held by the clinician during the entire temperature measure-
ment process to insure the probe tip maintains tissue contact.
6. During the temperature measurement cycle the TEMPERATURE display will show a
series of LED segments in a box - shaped formation. This indicates that the tempera-
ture measurement is in process.
7. When the final temperature has been reached a tone will sound and the temperature
will be displayed on the Monitor.
8. After the temperature measurement is complete, remove the probe from the patient's
mouth and eject the probe cover by firmly pressing the ejection button on the probe.
Properly dispose of the used probe cover.
9. Insert the probe into the probe holder before attempting to take another temperature
measurement.
10. The current temperature is displayed for one minute after the probe is replaced in the
holder, after which time the display will go blank. The most recent temperature can
be recalled by pressing the REVIEW button once.
Note: If a probe position error occurs during the temperature determination, the tempera-
ture display will alternate between the final predicted temperature and the letter "P" in the
display.
50
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
Note: The Welch Allyn Vital Signs Monitor is intended for use with adult and pediatric
patients, age 3 and above.
To take a axillary temperature (in either Normal or Monitor mode) follow this procedure:
1. Insure that the oral probe is connected to the Monitor. The oral probe has a BLUE tip.
Accurate axillary temperatures can only be obtained by using the blue-tipped tem-
perature probe.
2. Remove the probe from the probe holder. A short self-test mode will be initiated
where every LED segment on the TEMPERATURE display is illuminated briefly.
Following this self-test, the display will show “OrL”. OrL=Oral Probe indicating that
the oral probe is in use.
3. Press the SET-Arrow Up button once. The display will now show “ALY” indicating
that the Monitor is ready to take an axillary temperature reading. Note that subse-
quent presses of the SET-Arrow Up button allows the Monitor to toggle between the
oral and axillary modes of operation. The Monitor’s display must show “ALY” prior
to the initiation of a temperature measurement.
4. With the TEMPERATURE display showing “ALY”, load a probe cover onto the probe
by holding the probe collar with the thumb and forefinger, being careful not to hold
or press the ejection button.
5. Lift the patient’s arm so that the entire axilla is easily visualized. Place the probe as
high as possible in the axilla. Do not allow the probe tip to come into contact with
the patient until it is deliberately placed in the measurement site. Any tissue contact
before this time will activate the probe position error message and may cause inaccu-
rate tempurature readings.
6. Be sure that the probe tip will be completely surrounded by axillary tissue. Clothing
or other material touching the probe tip may cause inaccurate readings.
7. Place the arm snugly at the patient’s side. Hold the arm in this position without
movement of the arm or probe during the measurement cycle. Movement of the arm
may cause inaccurate readings.
8. In Normal Mode the Monitor will produce an audible tone and display the tempera-
ture reading when a final temperature has been reached. This generally takes about
10 seconds. In Monitor Mode the operator must allow the temperature readout to
stabilize for 5 minutes to accurately display a final temperature reading.
51
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
9. After the temperature measurement is complete, remove the probe from the patient’s
axilla and eject the probe cover by firmly pressing the ejection button on the probe.
Properly dispose of the used probe cover.
10. Insert the probe into the probe holder before attempting to take another temperature
measurement.
11. When in Normal mode, the current temperature is displayed for one minute after the
probe is replaced in the holder, after which time the display will go blank. The most
recent temperature can be recalled by pressing the REVIEW button once.
Note: If a probe position error occurs during the temperature determination, the tempera-
ture display will alternate between the final predicted temperature and the letter “P” in
the display.
1. Insure that the rectal probe is connected to the unit. The rectal probe has a RED tip.
Accurate rectal temperatures can only be obtained by using the red temperature
probe.
2. Remove the probe from the probe holder. A short self-test mode will be initiated
where every LED segment on the TEMPERATURE display is illuminated briefly.
Following this self-test the display will show "rEC". "rEC" = Rectal Probe indicating
that the rectal probe is in use. The Monitor’s display must show “rEC” prior to the
initiation of a temperature measurement.
3. Load a probe cover onto the probe by holding the probe collar with the thumb and
forefinger, being careful not to press the ejection button. See illustration below.
52
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
4. Separate the buttocks with one hand. Apply a thin coat of water-based lubricant
when necessary. Using the other hand, gently insert the probe ONLY 1cm (3/8 inch
ONLY) inside the rectal sphincter. Extreme caution should be used to avoid risk of
bowel perforation in children.
5. Tilt the probe to insure good tissue contact and continue to keep the buttocks sepa-
rated while the measurement is in process.
6. During the temperature measurement cycle, the TEMPERATURE display will show a
series of LED segments in a box-shaped formation. This indicates that the tempera-
ture measurement is in process.
7. When the final temperature has been reached, a tone will sound and the temperature
will be displayed on the Monitor.
8. After the temperature measurement is complete, remove the probe from the patient's
rectum and eject the probe cover by firmly pressing the ejection button on the probe.
Properly dispose of the used probe cover.
9. Insert the probe into the probe holder before attempting to take another temperature
measurement.
10. The current temperature is displayed for one minute after the probe is replaced in
the holder, after which time the display will go blank. The most recent temperature
can be recalled by pressing the REVIEW button once.
Note: If a probe position error occurs during the temperature determination, the tempera-
ture display will alternate between the final predicted temperature and the letter "P" in the
display.
Oxygen saturation percent is calculated with each pulse detected, and thus the Monitor
display is continually updated. The pulse signal bar graph is an indicator of the strength
and quality of the detected pulses.
SpO2 is generally measured via pulses detected using a finger clip sensor. However, for
certain situations SpO2 may be measured at alternate sites including the earlobe, forehead
and toes. Special sensors must be employed in these situations (See Appendix A for
Nonin® sensors, Appendix B for Nellcor Puritan BennettTM sensors).
53
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
The reusable finger clip pulse oximeter sensor is designed for spot-check monitoring of
arterial oxyhemoglobin saturation (SpO2%).
The finger clip sensor may also be used for continuous monitoring where patient move-
ment is not expected and the patient's finger is large enough for the sensor to fit securely.
The Nonin finger clip sensor (5200-40, 5200-54) may be used for continuous monitoring
less than 30 minutes. The Nellcor Puritan Bennett finger clip sensor (DS-lOOA) may be
used for continuous monitoring less than 4 hours.
To use the finger clip sensor, insert the patient's finger (preferably the left or right index
finger) completely into the sensor as shown. The thumb is specifically not recommended
for use with the finger clip sensor.
COR
NELL
The Nonin® Finger Clip Sensor The Nellcor Puritan BennettTM Finger Clip Sensor
(5200-40, 5200-54) (DS-100A)
Note: Sensor sites must be checked periodically to determine circulation, sensor position-
ing and skin sensitivity.
54
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
For a detailed description of the use and application of these sensors, see Appendix A of
this manual for Nonin® sensors, Appendix B for Nellcor Puritan BennettTM sensors.
Warning: Use only Nonin® brand SpO2 sensors and accessories with Welch Allyn
Vital Signs Monitors configured with Nonin® pulse oximetry module. Use only
Nellcor Puritan BennettTM brand SpO2 sensors and accessories with Welch Allyn
Vital Signs Monitors configured with the Nellcor Puritan BennettTM pulse oximetry
module.
The operator should follow this procedure for taking an SpO2 measurement:
2. The pulse signal bar graph will illuminate, indicating the relative strength and quali-
ty of the patient's pulses at the sensor site. The sensor takes approximately 10 sec-
onds to determine the initial SpO2% value and pulse rate. When the initial values
are determined they will be shown in the SpO2 display and the Pulse Rate display
respectively.
3. The SpO2% and pulse rate are updated approximately every second.
4. Removing the SpO2 sensor from the patient ends the monitoring period and immedi-
ately blanks the pulse signal bar graph. The SpO2 display will flash 2 dashes "--" and
a repetitive audible alarm will sound. To reset the sensor removal alarm, press the
SILENCE button on the Monitor's display.
Note: If the programmable alarms for low SpO2%, high pulse rate and low pulse rate are
turned off, no sensor removal alarm will occur when the SpO2 sensor is removed from the
patient.
55
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
The volume of the pulse tone may be adjusted to any of 5 volume settings, as well as a
“volume off” setting. To change the volume of the SpO2 pulse tone, follow these proce-
dures:
When the SpO2 sensor is attached to a patient, press the VOLUME button. With each sub-
sequent press of the button the volume of the pulse tone will increase, until the “Off” set-
ting is reached. Subsequent presses of the VOLUME button will again cycle through the
five volume settings. When the desired volume setting is reached, refrain from pressing
the button for three seconds and the setting will be saved. In all settings except “volume
off” the pulse tone indicator light will be illuminated. There is no visual indicator of the
volume level.
When the SpO2 sensor is NOT attached to a patient, press the VOLUME button. With each
press of the button a sample pulse tone will sound, and the SpO2 Pulse Signal Bar Graph
will provide a visual indicator of the volume setting. Subsequent presses of the VOLUME
button will cycle the unit through each of the five volume settings until the “Off” setting is
reached. Alternatively, the user can press the SET (arrow up and arrow down) buttons to
cycle through the volume choices. When the desired volume setting is reached, refrain
from pressing the button for three seconds and the setting will be saved. In all settings
except “volume off” the pulse tone indicator light will be illuminated.
56
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
The operator may control the printer using the two buttons (FEED, PRINT) located on the
top rear face of the Monitor.
The PRINT button will generate the printout of all the stored blood pressure, temperature,
pulse rate and SpO2 data in the Monitor. The unit will store and print a maximum of 99
data sets.
The following criteria are applied for saving and printing measurement data:
Blood Pressure Data: will be saved for both MANUAL and AUTOMATIC cycles.
Monitor Mode Temperature Data: will be saved only if concurrent with a blood pressure
cycle.
SpO2 Data: will be saved only if concurrent with a blood pressure cycle, or a predictive
temperature measurement. Will be saved if the START button is pressed and the BP mode
has been disabled. Will also be saved (one data set per minute) if a patient alarm condi-
tion occurs.
Pulse Rate Data: will be saved with every blood pressure and SpO2 data set.
Note: If no data is stored in the Monitor, pressing the PRINT button will cause only the
printout header to be printed.
Note: The printer uses thermal paper, which has a tendancy to fade over time, especially
when exposed to heat and light. For more permanent records, it is recommended that the
printout be photocopied.
Feed
57
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
Warning: It is the operator’s responsibility to set the Monitor’s programmable alarm lim-
its as appropriate for each individual patient. See section IV.3 for complete information.
The interface between the Monitor and the Nurse Call system will always be active whenev-
er an alarm condition occurs. Canceling or resetting an alarm on the Monitor itself will also
cancel the alarm signal to the Nurse Call System. Note that while the alarm sound will
“flash” on and off during an alarm condition, the Nurse Call output will be continuously
ON until the alarm condition is removed. Reference Appendix D of the Operator’s Manual
for the wiring diagram for the Nurse Call interface.
SECTION V
TROUBLESHOOTING/
MAINTENANCE/CALIBRATION
1. Troubleshooting: Error Indications and Interpretation
2. Troubleshooting: General Guide to Problems and Corrective Actions
3. Maintenance
a. Cleaning
b. Storage
c. Battery Removal and Replacement
4. Calibration
a. Blood Pressure Calibration Check
b. Temperature Calibration Check
c. SpO2 Calibration Check
60
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
V. 1. TROUBLESHOOTING:
ERROR INDICATIONS AND INTERPRETATION
The following table of alarm conditions and error codes is intended to provide the opera-
tor with a quick reference to the descriptions and probable causes of the error codes. For
service level troubleshooting, refer to the service manual.
When responding to a Monitor alarm, always CHECK THE PATIENT FIRST and then
proceed to check the Monitor.
C03 Blood Pressure Inflation too rapid Check for kinked hose or air
obstruction
C04 Blood Pressure Excessive inflation time Check for air leaks
61
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
Please note: Patient’s arm position Ensure patient’s arm is at heart level.
Differences of up to 10mmHg should be
considered normal and will occur for a Arm movement during blood pressure Keep arm still during blood pressure cycle:
number of reasons including intrapatient cycle • Movement may cause inaccuracies from artifact.
BP variability, observer hearing differences,
and auscultatory deflation rate. Blood pressure taken over clothing Blood pressure should be taken on a bare arm.
Incorrect reference Use the correct Korotkoff sound to determine diastolic blood
pressure.
• Many listeners incorrectly equate diastolic blood pressure with the
disappearance of sound only (phase 5). The Welch Allyn Vital Signs
Monitor was developed using the American Heart Association recom-
mendations, which state that phase 5 be used unless sound continues
to 0 mmHg, in which case the change in the quality of sound (phase
4) is to be used.
Change in blood pressure from ausculta- Check blood pressure immediately prior to Welch Allyn Vital Signs
tory reading to Welch Allyn Vital Signs Monitor Reading.
Monitor reading
Poor auscultatory sound recognition Use higher quality stethoscope. Have a different observer check
by observer patient’s blood pressure.
62
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
Connections from monitor to cuff Check all connections. (Do not over tighten).
3 NO CUFF loose
INFLATION
Probe not replaced Replace probe in holder prior to taking another temperature.
Improperly attached sensor Insert the patient’s finger completely into sensor
5 SpO2
Cable incorrectly plugged into Ensure sensor cable is correctly plugged into monitor
MALFUNCTION monitor
• Sensor in Place but No Sp02 on Sp02 disabled Ensure Sp02 is enabled (check Configuration Mode)
Display
Low battery Check connections between monitor and transformer, and trans-
7 MONITOR WILL former and wall receptacle.
NOT TURN ON Monitor not powering up Unplug unit from wall receptacle and check for breaks in cord. If
connections secure, check electrical outlet
• Charging light will be on if connections are OK and the monitor is
plugged into a working outlet
Pressure Preset too high Check default Pressure Preset setting: (in Configuration Mode)
8 CUFF TOO TIGHT • Unless patient has underlying systolic hypertension, set pressure
(Over inflation) preset at 160 mmHg. (If systolic blood pressure greater than
pressure preset, monitor will automatically increase an additional
40 mmHg)
Inappropriate size cuff Determine cuff size with the cuff markings or refer to chart in
9 CUFF POPPING Operator’s Manual for determining cuff size.
OFF • If cuff continues to pop off, notify Biomedical Department or Welch
Allyn Technical Support.
63
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
Do not have arm tight against chest wall as respirations may affect
speed and accuracy of blood pressure measurement.
Small leak in pneumatic system Check cuff and tubing for leaks.
Action EXPLANATION
1 Collect appropriate equipment. Many sphygmomanometers are inaccurate. Low quality stetho-
Use a certified accurate sphygmomanometer and quality scopes do not transmit sound well enough for blood pressure
stethoscope. sounds to be heard accurately. A cuff that is either too large or too
small will produce an inaccurate reading.
Select a blood pressure cuff of a suitable size. Use a blood
pressure cuff of the largest appropriate size for patient (see
markings on inside of cuff).
2. Have the patient assume a comfortable position with the If the arm is not at the proper level, inaccurate readings will result.
upper arm relaxed at heart level and the lower arm passively
supported.
3. Expose the area of the brachial artery by removing clothing, Clothing over the artery hinders the ability to hear and may cause
or move a sleeve, if not too tight, above the area where the inaccurate readings. Tight clothing may cause vessel congestion and
cuff will be placed. inaccurate readings.
4. Center the cuff bladder so that the lower edge is at least 1 This places the cuff in the best position for occluding the blood flow
inch (2.5cm) above the bend of inner arm of the elbow. through the brachial artery.
5. Palpate the brachial or radial pulse. Determines the most accurate location for assessment and approxi-
mation of systolic pressure.
6. Inflate the cuff until the pulsation disappears. Then continue Facilitates identification of Phase One of Korotkoff sounds.
to inflate until the pressure reads 30 mmHg above the point
where the pulse disappeared.
7. Listen carefully with stethoscope over brachial artery while One of the major sources of error in auscultatory blood pressure
controlling the release of air at a rate of 3 mmHg per second. measurement is deflating the cuff too quickly. It is a normal opera-
tion of the Welch Allyn Vital Signs Monitor to deflate at the
American Heart Association recommended 3 mmHg per second.
8. Systolic is determined by reading the manometer gauge Follows AHA recommended standards.
when the first faint but clear tapping sound is heard with
the stethoscope.
9. Diastole, in adults, is determined by reading the manometer Follows AHA recommended standards.
gauge to the closest even number when the last sound is Diastolic blood pressure in children is the point at which the sound
heard. becomes muffled.
64
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
Occasionally, as necessary, the unit may be cleaned with appropriately diluted non-stain-
ing disinfectant solution.
Note: Care should be taken to prevent water or other fluids from entering any connectors.
Should this occur, the connectors should be dried with warm air. All monitoring functions
should then be checked for proper operation.
The reusable blood pressure cuff should be cleaned by sponging with a damp cloth. If
washing is necessary, the bladder should be removed and the cuff washed with soap or
detergent/disinfectant. After washing, the cuff should be air dried.
Cabling and the pressure hose can be wiped with a damp cloth moistened in a mild deter-
gent solution. Do not immerse hoses.
The Nonin® reusable finger clip SpO2 sensor may be cleaned with isopropyl alcohol solution,
and may be sterilized using ethylene oxide (EtO), cold cycle. Do not immerse the sensor.
The Nellcor Puritan BennettTM reusable finger clip SpO2 sensor may be cleaned with 70%
alcohol or 1:10 bleach.
V. 3. B. MAINTENANCE: STORAGE
Whenever possible, the Welch Allyn Vital Signs Monitor should be stored at room temper-
ature in a dry environment.
If it is necessary to store the monitor for an extended period of time, the unit should be
packed in its original shipping carton.
Note: Insure that the printer is loaded with paper prior to storage.
65
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
V. 3. C. MAINTENANCE:
BATTERY REMOVAL & REPLACEMENT
Occasionally it may be necessary for the internal battery to be replaced. When the battery
will no longer take a charge, remove it and replace it as outlined below, with a battery of
the same part number.
1. Insure that the AC power transformer cord is disconnected from the Monitor, and
that the Monitor is turned off.
2. Use a Phillips-head screwdriver to remove the four (4) screws holding in the battery
door. Remove the battery door exposing the battery.
3. Tip the Monitor and slide the battery out. Disconnect each of the 2 connectors and
discard the old battery per local regulations.
P6
5. Slide the new battery into the battery compartment as far as it will go.
7. Connect the AC power transformer to the monitor and allow the new battery to
charge for approximately 16 hours. The Monitor may be used during this charging
period. While charging, the charging indicator will flash a yellow light. When the
Monitor is fully charged, the flashing light will be replaced with a solid yellow light.
8. The battery is a non-spillable lead-acid battery. In the USA, call 1-800-SAV-LEAD for
instructions on how to recycle. For International users, contact your local authorities
on recycling.
66
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
V. 4. A. CALIBRATION:
BLOOD PRESSURE CALIBRATION CHECK
The accuracy of the Welch Allyn Vital Signs Monitor’s pressure measurement can be veri-
fied with an accurate office mercurial or aneroid sphygmomanometer. The calibration
check is a simple yet valuable test to determine that the unit is sensing pressure accurately.
Welch Allyn recommends that blood pressure calibration be verified on an annual basis.
Equipment needed:
1. Accurate* office sphygmomanometer with inflation system
2. Calibration T-connector
3. Any rigid cylinder that approximates arm size (such as a one-pound coffee can)
Disconnect the coiled tubing from the cuff. Attach the calibration T-connector to the
coiled tubing end. Connect the male taper connector extending from the tubing of the
office sphygmomanometer into the large, open end of the T-connector. Then connect the
female taper connector extending from the office sphygmomanometer cuff to the other
side of the T-connector. Wrap the arm cuff around the rigid cylinder. Check that all con-
nections are tight.
Initiate the Monitor's internal configuration settings menu by powering up the unit while
the START button is depressed. Hold the START button down until all the LED display
segments go off. The first message displayed is the revision level of the internal software.
This will be displayed in the systolic and diastolic displays. Press the REVIEW button to
advance to the calibration check menu. Once in the calibration check menu, the display
will read "CAL" in the systolic display and the pressure will be shown in the diastolic and
pulse rate displays.
Press the START button to close the Monitor's internal valve. Inflate the cuff manually to
a pressure of about 250 mmHg by squeezing the manometer bulb. (If the cuff won't
inflate, insure that the valve on the bulb is fully closed and that you have pressed the
START button to close the Monitor's internal valve). Now, slowly deflate the cuff and stop
at 30mmHg increments to compare the pressure reading on the manometer with the pres-
sure reading displayed on the Monitor display.
The Monitor displays the pressure in mmHg units, down to the .01 mmHg resolution.
This should compare closely with the manometer reading, within 3 mmHg. (Inherent inac-
curacies of ±3mmHg in office sphygmomanometers could result in an observed and
acceptable Monitor error of ±6mmHg).
Compare several readings taken at pressures ranging from 250 to 50 mmHg. This checks
the Monitor within the range of typical blood pressure readings. If the Monitor does not
agree closely with the manometer, that is within ±3 mmHg, call the Welch Allyn technical
service center.
67
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
WARNING: Internal safety features of the Monitor will automatically open the
valve and sound an alarm if the pressure is detected to be above 15 mmHg for
longer than 165 seconds. In the event that this happens pressing either the
SILENCE or CANCEL button will reset the alarm condition. Pressing the START
button will then close the valve after which the check can be continued.
Upon completion of the calibration check, the Monitor can be powered down and pow-
ered back up to bring the device to its normal state of operation.
V. 4. B. CALIBRATION:
TEMPERATURE CALIBRATION CHECK
The accuracy of the Welch Allyn Vital Signs Monitor temperature determinations can be
verified using the Temperature Calibration Key. The calibration check is a simple yet
valuable test to determine that the unit is measuring temperature accurately.
2. Remove the temperature probe completely from the holder and detach the connector,
then insert the temperature calibration key.
3. Insert, then remove the probe from the probe holder to reset the thermometer.
4. Wait for the temperture display test to be complete. Observe and record the temper-
ture noted on the temperature display.
5. The recorded temperature should be 97.3°F ± 0.2°F (36.3°C +/- 0.1° C) if the Monitor
is properly calibrated. If the temperature displayed is out of this range, call your
Welch Allyn technical service center for assistance.
NONIN Model 8000S- for Monitors incorporating the Nonin brand pulse oximetry
module.
NELLCOR PURITAN BENNETT Model SRC2- for Monitors incorporating the Nellcor
Puritan Bennett brand pulse oximetry module.
Follow the manufacturer’s instructions for performing a pulse oximetry calibration check.
68
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
SECTION VI
WARRANTY AND
SERVICE INFORMATION
1. Warranty Information
2. Service Information
a. Service Policy
b. Technical Assistance/Service Centers
c. Service Manual/Spare Parts
d. Service Loaners
69
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
70
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
If you are advised to return a product to Welch Allyn for service or repair, schedule the
repair with the service center nearest you.
Before returning a product for repair you must obtain authorization from Welch Allyn.
An RGA (Return Goods Authorization) number will be given to you by our service per-
sonnel. Be sure to note this number on the outside of your shipping box. Returns
without an RGA number will not be accepted for delivery.
The CE Mark on this product indicates it has been tested to and conforms
with the provisions noted with the 93/42/EEC Medical Device Directive.
Also included with the service manual is a complete spare parts price list. Spare parts
may be ordered from your local Welch Allyn Service Center.
For service repairs outside of warranty or contract, loaners will be available for a nominal
daily charge and will be shipped subject to availability. Loaners will be shipped pre-paid,
however this charge will be added to the service charges.
72
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
The Welch Allyn Vital Signs Monitor features a sensor which is ideal for nearly every
application. These sensors include the Flex Sensor, the Fingerclip Sensor, the Ear Clip, the
Reflectance Sensor and two sizes of Flexi-Form single patient use sensors. Each sensor is
designed for a specific site application and specific patient size and weight range. When
selecting sensors for a particular application consider the following:
1. The best performing sensor for most patients is an appropriately sized adhesive style
sensor, either reusable or single patient use, located on the finger or toe.
2. The fingerclip sensor is recommended for spot checks or short term continuous moni-
toring. The fingerclip sensor performs best for most patients when used on fingers
other than the thumb. The finger clip sensor is not recommended where motion is
expected or for relatively long term monitoring; e.g. greater than 30 minutes.
3. The reflectance and ear clip sensors generally do not perform as well as sensors locat-
ed on the finger or toe. They are not recommended for applications where the best
possible SpO2 accuracy is important. Use the reflectance and ear clip sensors when
fingers and toes are not suitable, as with peripheral shut down, or when monitoring
central body perfusion for timing response reasons. They may also be useful in high
motion environments such as stress testing.
Clear Tape Strips or Hydrogel Tape Strips are recommended to provide additional secur-
ing for the reusable flex sensors.
Clean reusable sensors with an isopropyl alcohol wipe. Allow enough time for the sensor
to dry thoroughly before reusing. The reusable sensors may also be sterilized using ethyl-
ene oxide (EtO) (cold cycle).
73
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
COMPATIBILITY
Caution: Use only the sensors provided by Welch Allyn. These sensors are manufactured
to meet the calibration requirements for the Welch Allyn Vital Signs Monitor.
Caution: Each sensor is designed for a specific clinical application. Optimal performance
can only be attained by using each sensor appropriately.
Caution: Use of double backed adhesive strips or the Hydrogel tape strips should be dis-
continued if the patient exhibits allergic reactions to the adhesive material.
Caution: Sensor sites must be checked periodically to determine circulation, sensor posi-
tioning and skin sensitivity.
Warning: Only Nonin® brand SpO2 sensors and accessories should be used with
Welch Allyn Vital Signs Monitors configured with the Nonin® pulse oximetry option.
74
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
NOTE: If patient movement is occurring or the finger size is inappropriate, select a differ-
ent sensor that is appropriate for the patient and the monitoring environment.
Insert finger (preferably left or right index finger) completely into the sensor. See illustra-
tion below. The thumb is specifically not recommended for use with the fingerclip sensor.
NOTE: For the best results, secure the sensor cable independently from the sensor, prefer-
ably around the base of the finger. Make sure that the tape securing the cable does not
restrict the blood flow.
Position the Sensor on the top and bottom of the end of the finger or toe. Place the light
emitter portion on the finger/toe nail side and the detector on the side opposite of the
nail. In all sensor placement applications, align the windows (detector and emitter por-
tions of the sensor) over the tissue. Attach the sensor using 3M Micropore* tape, or equiv-
alent by wrapping the tape or wrap over the sensor assembly. Wrap the sensor snug but
not so tight as to restrict the blood flow. See illustration below.
75
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
NOTE: For optimum light transmission, attach the sensor on the finger or toe. For best
results, secure the cable independently from the sensor. Make sure that the tape securing
the cable does not restrict the blood flow.
Remove the backing from one side of the double back tape and apply to the flange of the
holder. Then remove the back from the other side of the tape. Press the sensor into the
foam with the windows out, and apply to the patient. Use additional tape to secure the
lead wire to the patient to avoid pulling or tipping the sensor.
Caution: The reflectance sensor is not recommended for pediatric patients because the
accuracy has not been established for pediatric use.
The preferred application site is the index finger. However, other fingers or toes may be
used where the tissue thickness is between 5 and 21 millimeters. Other sites may not give
acceptable results because of inadequate perfusion or inadequate light transmission. The
application of these sensors is the same for either the adult or the pediatric patient. The
difference is in the size of the sensor. For best results, secure the cable independently from
the sensor. Make sure that the tape securing the cable does not restrict the blood flow.
Caution: Do not stretch the tape while applying the sensor. This may cause inaccurate
readings or skin blisters.
NOTE: These sensors may be sterilized using ethylene oxide (EtO) cold cycle after
removal from the plastic shipping bag.
76
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
a. Grip the tab on the sensor's bottom adhesive cover and peel the adhesive cover off.
b. Place the patient's finger or toe into the sensor nail side up with the tip of the finger
or toe centered against the center line mark in the curved area as illustrated.
Reference the line that indicates the center of the curved area on the tape. This will
assure vertical alignment between the emitter and detector. Wrap the tape around
the finger. The fingernail should not be covered with tape during this step.
T
O
IC
P
IATR
PED
d. Fold the sensor's top over the top of the finger. Ensure that the detector and emitter
are vertically aligned as illustrated by the dotted axis line.
TO
P
RIC
PEDIAT
77
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
Nellcor Puritan Bennett, the leading manufacturer of pulse oximeters, offers a broad line
of adhesive and reusable oximetry sensors for all monitoring requirements. This appendix
contains information about selecting and applying adhesive and reusable sensors and pro-
vides tips for the optimal, cost-effective use of sensors.
Please consult the directions for use packaged with Nellcor Puritan BennettTM sensors for
additional information.
Adhesive Sensors
Nellcor Puritan Bennett's adhesive sensors are designed to enable the sensor's light source
and photodetector to be securely and properly positioned on the patient. The adhesive
stabilizes these important optical components and provides a comfortable, “second-skin”
fit.
Adhesive sensors are patient-dedicated and can travel with your patients. Single-patient-
use sensors do not present the risk of cross-contamination caused by products that are
reused from patient to patient.
Nellcor Puritan Bennett's sterile, adhesive Oxisensor® II sensors are the ideal choice when
environmental electronic noise levels are high and the patient's pulse is weak, because
special shielding in both the bandage and the cable helps protect the pulse oximetry sig-
nal. Nellcor Puritan Bennett offers Oxisensor II models to fit different patient sizes.
Nellcor Puritan Bennett offers another cost-effective option in its adhesive sensor line. The
OxiCliq® system combines a reusable cable with a patient-dedicated adhesive sensor. The
sensor clicks into place in the cable connector and is easy to detach when monitoring is
interrupted. This two-part system is less expensive than other adhesive sensors. Nellcor
Puritan Bennett offers OxiCliq models to fit different patient sizes.
78
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
Reusable Sensors
Nellcor Puritan Bennett also offers a broad line of reusable sensors designed to monitor
various sizes of relatively immobile patients, particularly when cross-contamination is less
of an issue. When short-term or intermittent monitoring is necessary, these reusable, non-
sterile sensors are an effective monitoring alternative. Nellcor Puritan Bennett offers the
following models of reusable sensors and accessories:
• Dura-Y® multisite sensor, which can be applied with disposable wraps, an ear clip for
adults, or a pediatric spot-check clip, depending on patient size.
Weight
All sensors contain a light source and photodetector, which are the essential optical com-
ponents necessary to determine arterial oxygen saturation by pulse oximetry (SpO2). All
Nellcor Puritan Bennett sensors are designed so that the light source and photodetector
are positioned a certain distance from one another to provide for proper fit over various
sizes of tissue.
Sensors should be chosen according to the patient's body weight to ensure the optical
components are properly aligned when applied to the recommended area. With the excep-
tion of the reflectance (RS-10) and nasal (R-15) sensors, all sensors must be positioned so
that the light source and photodetector are directly opposed to one another across an arte-
riolar bed (Figure 1).
Figure 1
Duration of Use
While adhesive sensors can be used for short- or long-term monitoring, reusable sensors
are generally indicated for spot-check measurements or for short-term monitoring of less
than four hours. Adhesive sensor sites should be checked for skin integrity and distal cir-
culation at least once every eight hours and changed as appropriate. Reusable sensor sites
must be checked and changed at least every four hours or as specified in the directions for
use.
79
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
Patient Activity
Adhesive sensors provide a stable, "second-skin" fit that maintains secure positioning of
the sensor's optical components and allows the sensor to move with the patient. With
active patients, these sensors provide greater monitoring reliability. Reusable sensors gen-
erally are less secure on active patients. Since the Dura-Y and the Oxiband sensors are held
in place with an adhesive wrap, they provide a more secure fit on active patients than the
Durasensor finger clip sensor.
Sterility
Adhesive Oxisensor II and OxiCliq sensors are sterile in their unopened packages. They
offer an infection control advantage for patients with suspected or confirmed infections,
and for those at greater risk for infection such as neonates or immunosuppressed patients.
Reusable sensors are nonsterile and require cleaning between patients with 70% alcohol or
1:10 bleach.
Monitoring Compatibility
Only Nellcor Puritan Bennett brand SpO2 sensors and accessories should be used with
Welch Allyn Vital Signs Monitors configured with the Nellcor Puritan BennettTM pulse
oximetry option.
This sensor is for use only with Nellcor Puritan Bennett instruments and with instruments
that contain Nellcor Puritan Bennett oximetry or are licensed to use Nellcor Puritan
Bennett sensors. Consult individual manufacturers for compatibility of particular instru-
ments and sensor models. Each instrument manufacturer is responsible for determining
whether its instruments are compatible for safe and effective use with each Nellcor
Puritan Bennett sensor model. Because most monitoring systems manufacturers offer
Nellcor Puritan Bennett sensor compatibility, many hospitals now apply an Oxisensor II
sensor to each monitored patient. The sensor can travel with the patient from department
to department throughout the hospital stay.
Latex Content
The adhesive coating of the bandage on the Oxisensor II sensor models I-20, I-20R, N-25
and N-25R, OxiCliq N, OxiCliq I and the ADH-A/N disposable wraps, may contain
a minute amount of natural latex. Please note that this material is the same type used in
commercial adhesive bandages.
With respect to the RS-10 reflectance oxygen transducer, the patient contact surface and
the double-sided adhesive tabs contain no natural latex. However, natural latex is a com-
ponent in the adhesive used to bond the layers within the sensor. These layers are above
the patient contact surface.
The adhesive materials used in these sensors have been tested using the Latex ELISA
Antigenic Protein (LEAP) Assay of Biomedical Extract testing at a nationally recognized
independent testing laboratory. No latex antigen was detected in this test. The ELISA test
has a lower detection threshold of 0.03 micrograms/milliliter.
80
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
All Sensors
• Choose your sensor according to patient size, duration of use, level of activity, and
infection control concerns.
• Many sensors have both preferred and alternative application sites, offering a broader
range of placement options.
• Observe alignment marks to ensure proper position of the light source and photodetec-
tor for each sensor.
• Sensors placed on the extremities should be positioned at heart level. When selecting a
sensor site, priority should be given to an extremity free from an arterial catheter, blood
pressure cuff, or intravenous infusion line. This will reduce challenges related to poor
pulse signals at the sensor site.
• It is possible that light from the light source may scatter throughout edematous tissue
before reaching the photodetector, which can result in inaccurate readings. Care should
be taken to position the sensor on nonedematous sites.
If extensive peripheral edema is present, the R-15 nasal sensor, RS-10 reflectance sensor,
or Dura-Y sensor with ear clip may be an acceptable alternative.
• To reduce the potential for motion artifact on an active patient, apply the sensor or an
appropriate alternative sensor to a less active, recommended site (e.g., the great toe or
forehead of an adult).
• Cover the sensor site with an opaque material in the presence of bright light sources
such as direct sunlight, surgical lamps, infrared warming lamps, and phototherapy
lights. This will minimize the potential for ambient light interference, which can create
unreliable readings.
81
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
• Check the sensor site and circulation distal to the sensor site as specified in the direc-
tions for use.
• In cases of poor perfusion, local rewarming of sensor sites (as permitted by hospital
policy) may restore adequate signal quality. Covering the sensor site with a noncon-
stricting bootie or glove, placing the site under a warming device, and/or correcting
the cause of poor perfusion can improve blood flow to the area.
• Repositioning the patient to correct compromised blood flow may restore signal quality
when readings cannot be obtained.
• Nellcor Puritan Bennett oximetry sensors should not be used during MRI scanning
because the sensors could affect the MRI image, or the MRI could affect the accuracy of
the oximetry measurements.
Adhesive Sensors
• Apply to clean, dry sites.
• Avoid applying additional tape over the sensor bandage. This will reduce the risk of
venous pulsation and inaccurate saturation measurements as well as the potential for
pressure damage at the site. However, applying tape over the cable may help prevent
the sensor from becoming dislodged (Figure 2).
Figure 2
• Check sensor site and circulation distal to the sensor at least every eight hours, and
change the sensor site as appropriate.
• Carefully remove the sensor from the patient for reapplication to another site.
• Oxisensor II and OxiCliq sensors may be reused on the same patient if the adhesive tape
attaches without slipping. Replace the sensor when its adhesive quality is depleted.
• With Oxisensor II sensors, if intermittent readings are required, leave the sensor on the
patient, secure the cable, and protect the site as much as possible when not monitoring.
82
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
• With OxiCliq sensors, if intermittent readings are required, leave the sensor on the
patient and either disconnect the OxiCliq sensor cable (OC-3) from the sensor (Figure 3)
or secure the cable with tape. Protect the site as much as possible when not monitoring.
Figure 3
• If the sensor is removed between readings, store the sensor on a clean, dry surface,
such as the original sensor liner, and label with the patient's name (Figure 4).
Figure 4
• Keep sensors with the patient's belongings until discharge, and notify others that the
patient has a "dedicated" sensor.
• The Oxisensor II I-20 and N-25 and OxiCliq N and I sensors have six clear, double-sided
adhesive "dots" that can be applied over the light source and photo detector to restore
adhesive quality and extend the life of the sensor.
• In cases of poor perfusion, adhesive sensors may obtain more reliable readings than
reusable sensors. The R-15 nasal sensor or RS-10 reflectance sensor may provide read-
ings when peripheral pulse signals are poor.
• The Oxisensor II D-25 and D-20 sensors have an 18-inch cable, while the I-20, N-25, and
D-25L come with a 36-inch cable to accommodate application in a variety of situations.
The OxiCliq sensor cable (OC-3) is a 36-inch cable linking the OxiCliq sensor to the
pulse oximeter.
83
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
Reusable Sensors
• Check sensor site and circulation distal to the sensor, and move the sensor to another
site at least every four hours or as specified in the directions for use.
• Avoid applying additional tape to minimize the risk of impaired perfusion and tissue
injury. However, applying tape over the cable may help prevent the sensor from
becoming dislodged.
• Carefully remove the adhesive wrap from the D-YS, OXI-P/I, OXI-A/N, and RS-10 to
avoid damage to the patient's skin and the sensor.
• Secure the RS-10 reflectance sensor with the headband provided. Avoid use on patients
in Trendelenburg, supine, or lateral positions, and with patients on mechanical ventila-
tors.
• Apply "Do not throw away" labels to cables of reusable sensors to prevent inadvertent
disposal.
• "If the cable is gray, don't throw it away": Nellcor Puritan Bennett manufactures
Durasensor, Oxiband, and Dura-Y sensors and the EC-4, EC-8, and OC-3 cables with gray
components to identify them as reusable rather than disposable items.
Contact your Nellcor Puritan Bennett Regional Oximetry Specialist, Account Manager, or
Clinical Consultant at 1-800-NELLCOR for information and assistance with all Nellcor
Puritan Bennett sensor products. They will gladly assist you, your department, and your
hospital with educational presentations and any questions regarding sensors.
84
WELCH ALLYN VITAL SIGNS MONITOR
QUICK GUIDE FOR NELLCOR PURITAN BENNETTTM SENSORS
Oxisensor II adhesive sensors – sterile, single-patient use
MODEL QUANTITY APPLICATION CHARACTERISTICS
D-25/ Case of 24 Adult • For patients who weigh more than 30 kg:
D-25L - Preferred application site is index finger.
- Alternative sites are thumb, smaller finger, great or
second toe.
• D-25 has an 18-inch cable.
• D-25L has a 36-inch cable.
• Check at least every 8 hours.
D-20 Case of 24 Pediatric • For patients who weigh between 10 and 50 kg:
- Preferred application site is index finger.
- Alternative sites are thumb, smaller finger, great or
second toe.
• Check at least every 8 hours.
I-20 Case of 24 Infant • For infants who weigh between 3 and 20 kg:
85
R-15 Case of 24 Adult nasal • For patients who weigh more than 50 kg:
- Only application site is bridge of nose.
• For use in no-motion environments.
• Check at least every 8 hours.
OPERATOR MANUAL
OxiCliq adhesive sensors – sterile, single patient use
A Case of 24 Adult • For patients who weigh more than 30 kg:
(OC-3 cable sold - Preferred application site is index finger.
separately) - Alternative sites are thumb, smaller finger, great toe.
• Must be used with OxiCliq OC-3 sensor cable.
• Check site at least every 8 hours.
OPERATOR MANUAL
86
• Pediatrics (15-40 kg)
- Preferred application site is index finger, with cable
running along top of hand.
WELCH ALLYN VITAL SIGNS MONITOR
OPERATOR MANUAL
- Alternative sites are thumb or other small finger.
• Infants (3-15 kg)
- Preferred application site is great toe, with cable
running along sole of foot.
• Adhesive wraps are disposable.
• Recommend changing site at least every 4 hours.
• For short-term use only. If long-term monitoring is
required, consider using an Oxisensor II D-20 or
OxiCliq P sensor for pediatrics and an Oxisensor II-20
or OxiCliq I sensor for infants.
.
OPERATOR MANUAL
DS-100 1 per package Adult • For patients who weigh more than 40 kg:
- Preferred application site is index finger.
- Alternative site is smaller finger.
• Do not use on thumb or toe.
• Recommend changing site at least every 4 hours.
• For short-term use only. If long-term monitoring is
required, consider using an Oxisensor II D-25 or an
OxiCliq A sensor.
.
RS-10 reflectance sensor
RS-10 Case of 6 sensors Adult/reflectance • For patients who weigh more than 40 kg:
with 6 headbands - Preferred application site is forehead above the brow
88
or below the hairline.
- Alternative site is the temple.
• Recommend changing site at least every 4 hours.
WELCH ALLYN VITAL SIGNS MONITOR
INDICATIONS/CONTRAINDICATIONS
The Nellcor Puritan Bennett® Durasensor® adult oxygen transducer, model DS-1OOA, is indicated
for use when continuous noninvasive arterial oxygen saturation and pulse rate monitoring are
required for patients weighing greater than 40 kg.
This sensor is for use only with Nellcor Puritan Bennett instruments and with instruments that con-
tain Nellcor Puritan Bennett oximetry or are licensed to use Nellcor Puritan Bennett sensors.
Consult individual manufacturers for compatibility of particular instruments and sensor models.
Each instrument manufacturer is responsible for determining whether its instruments are compati-
ble for safe and effective use with each Nellcor Puritan Bennett sensor model.
The DS-1OOA is contraindicated for use on active patients or for prolonged use. It is not designed
for long-term monitoring. It must be moved every 4 hours (or more often, if indicated by circulato-
ry status and/or skin integrity) and reapplied to a different site. If long-term monitoring is required,
use an Oxisensor® II, oxygen transducer, model D-25, or D-25L.
89
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
Note: If the sensor does not track the pulse reliably, it may be incorrectly positioned—or the sensor
site may be too thick, thin, or deeply pigmented to permit appropriate light transmission. If any of
these situations occurs, reposition the sensor or choose an alternate Nellcor Puritan Bennett sensor.
CLEANING
The DS-1OOA may be surface-cleaned by wiping it with a solution such as 70% isopropyl alcohol.
If low-level disinfection is required, use a 1:10 bleach solution. Do not use undiluted bleach (5%-
5.25% sodium hypochlorite) or any cleaning solution other than those recommended here because
permanent damage to the sensor could occur.
1) Saturate a clean, dry gauze pad with the cleaning solution. Wipe all surfaces of the sensor and
cable with this gauze pad.
2) Saturate another clean, dry gauze pad with sterile or distilled water. Wipe all surfaces of the sen-
sor and cable with this gauze pad.
3) Dry the sensor and cable by wiping all surfaces with a clean, dry gauze pad.
WARNINGS
1) Failure to apply the DS-1OOA properly may cause incorrect measurements.
2) Using the DS-1OOA in the presence of bright lights may result in inaccurate measurements. In
such cases, cover the sensor site with an opaque material.
3) Reusable sensors must be moved to a new site at least every 4 hours. Because individual skin
condition affects the ability of the skin to tolerate sensor placement, it may be necessary to
change the sensor site more frequently with some patients. If skin integrity changes, move the
sensor to another site.
4) Intravascular dyes may lead to inaccurate measurements.
5) The performance of the DS-1OOA is compromised by motion, use of this sensor is contraindi-
cated for active patients.
6) Do not apply tape to secure the sensor in place or to tape it shut; venous pulsations may lead to
inaccurate saturation measurements.
7) As with all medical equipment, carefully route cables to reduce the possibility of patient entan-
glement or strangulation.
8) Do not use the DS-lOOA or other oximetry sensors during MRl scanning. Conducted current
may cause burns. Also, the DS-lOOA may affect the MRI image, and the MRI unit may affect
the accuracy of oximetry measurements.
9) Do not alter or modify the DS-lOOA. Alterations or modifications may affect performance or
accuracy.
90
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
If you have questions regarding any of this information, contact the Technical Services Department,
or your local Nellcor representative.
WARRANTY
To obtain information about a warranty, if any, for this product, contact Nellcor Puritan Bennett
Technical Services or your local Nellcor Puritan Bennett representative.
ACCURACY SPECIFICATIONS
For the accuracy specification range when used with Nellcor Puritan Bennett monitors, refer to
information provided with the monitor, or (in the U.S.), contact Nellcor Puritan Bennett’s Technical
Services Department. Outside the U.S., contact your local Nellcor Puritan Bennett representative.
For the accuracy specification range when used with a monitor other than those manufactured by
Nellcor Puritan Bennett, consult the information provided by the manufacturer of that instrument.
Note: Refer to the instrument operator’s manual for complete instructions for use of the sensor with
that monitor.
CAUTION: Federal law (U.S.) restricts this device to sale by or on the order of a physician.
This sensor is sold under the following U.S. Patents and foreign equivalents: 4,621,643; 4,685,464;
and 4,700,708.
Oxisensor 11, and Durasensor are trademarks of Nellcor Puritan Bennett Inc.
91
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
92
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
93
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
94
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
95
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
96
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
97
WELCH ALLYN VITAL SIGNS MONITOR OPERATOR MANUAL
Note: See Section IV. 19. A of the Operator’s Manual for complete information on the Nurse Call
interface.
The Welch Allyn Vital Signs Monitor provides a switch closure output between two pins of the
serial connector (pins 7 and 8). The output is “universal” in that it is compatible with a wide vari-
ety of different systems with no polarity dependence. The Nurse Call output is ohmically isolated
from all circuitry. The output is rated for 1 Amp at 240 VAC or 30VDC. During an alarm condi-
tion the output is closed, otherwise the output is open. During a power off or power failure condi-
tion the output is open.
98
Welch Allyn Vital Signs Monitor ______________________________________
Serial Number:
Date of Purchase:____________________________________________________
Telephone:__________________________________________________________
Telephone:__________________________________________________________
Welch Allyn
Sales Representative: ________________________________________________
Tel.: 315-685-4100 or
1-800-535-6663
Fax.: 315-685-3361