User's Guide
User's Guide
User's Guide
Model 38389
Safety
International Safety Symbols
This symbol, adjacent to another symbol or terminal,
indicates the user must refer to the manual for further
information.
Double insulation
SAFETY NOTES
• Do not exceed the maximum allowable input range of any function
• Do not apply voltage to meter when resistance function is selected.
• Set the function switch OFF when the meter is not in use.
• Remove the battery if meter is to be stored for longer than 60 days.
WARNINGS
• Set function switch to the appropriate position before measuring.
• When measuring volts do not switch to current/resistance modes.
• Do not measure current on a circuit whose voltage exceeds 600V.
• When changing ranges using the selector switch always disconnect
the test leads from the circuit under test.
• Do not exceed the maximum rated input limits.
UL
A UL mark does not indicate that this product has been evaluated for
the accuracy of its readings.
Input Limits
Function Maximum Input
A AC, A DC 600A
V DC, V AC 600V DC/AC
Resistance, Capacitance, Diode Test 250V DC/AC
Temperature 60V DC, 24V AC
1. Current clamp
2. Data Hold button 1
3. Peak Hold button
4. Mode select button
2
5. ZERO button
3 8
6. LCD display
4 9
7. COM input jack 5
8. Clamp trigger
6
9. Rotary Function switch
10. V/Ω/CAP/TEMP jack 7 10
Capacitance Measurements
WARNING: To avoid electric shock, discharge the capacitor under test
before measuring.
1. Set the function switch to the •)))
Ω CAP Position.
2. Press the multifunction button to select the
capacitance function.
3. Insert the black test lead banana plug into
the negative (COM) jack and the red test
lead banana plug into the positive jack.
4. Touch the test probe tips across the part
under test.
5. Read the capacitance in the display.
6. The display will indicate the proper CAP
decimal point and value.