ELECTRICAL SAFETY TESTER ELECTRICAL SAFETY TESTER

USER MANUAL USER MANUAL

EC Declaration of Conformity
We
GOOD WILL INSTRUMENT CO., LTD.
WARNING No.7-1, Jhongsing Road, Tucheng City, Taipei County 236, Taiwan
GOOD WILL INSTRUMENT (SUZHOU) CO., LTD.
MANUFACTURER ASSUMES NO LIABILITY No. 69 Lushan Road, Suzhou New District Jiangsu, China.
IF UNIT OPERATED IN AN UNSAFE declares that the below mentioned product
GPT-805/815, GPI-825/826
MANNER. are herewith confirmed to comply with the requirements set out in the Council
Directive on the Approximation of the Law of Member States relating to
Electromagnetic Compatibility (89/336/EEC,92/31/EEC) and Low Voltage
Equipment Directive (73/23/EEC, 93/68/EEC).
For the evaluation regarding the Electromagnetic Compatibility and Low Voltage
Equipment Directive, the following standards were applied:

EN 61326-1: Electrical equipment for measurement, control and laboratory use ––


EMC requirements (1997+A1: 1998+A2:2001)
Conducted and Radiated Emissions Electrostatic Discharge
WARNING EN 55011 Group I class A: 1998 EN 61000-4-2: 1995+A1 :1998
Current Harmonic Radiated Immunity
THIS INSTRUMENT GENERATES AND EN 61000-3-2: 2000 EN 61000-4-3: 1996+A1 :1998
DELIVERS A HAZARDOUSLY HIGH Voltage Fluctuation Electrical Fast Transients
EN 61000-3-3: 1995 EN 61000-4-4: 1995
VOLTAGE (5kV). BE EXTREMELY Surge Immunity
----------------------------------------------
CAREFUL WHEN USING THIS EN 61000-4-5: 1995
Conducted Susceptibility
INSTRUMENT. BE SURE TO READ ----------------------------------------------
EN 61000-4-6: 1996
MANUAL SECTION 3 PRECAUTIONS ---------------------------------------------- Power Frequency Magnetic Field
IEC 61000-4-8: 1993
Voltage Dips/ Interrupts
----------------------------------------------
EN 61000-4-11: 1994

Low Voltage Equipment Directive 73/23/EEC & amended by 93/68/EEC


Safety Requirements IEC/EN 61010-1: 2001
Remark: Also complied with Continuity of Protective Bonding Tester, Insulation
Resistance Test, Voltage Test, and Residual Voltage Test in accordance with
the Sub-Clauses 19.2, 19.3, 19.4 and 19.5 of EN 60204-1: 1997

i ii
ELECTRICAL SAFETY TESTER ELECTRICAL SAFETY TESTER
USER MANUAL USER MANUAL

CONTENTS PAGE
1. PRODUCT INTRODUCTION............................................ 1 SAFETY TERMS AND SYMBOLS
1-1.Description…………………………..…………………. 1 These terms may appear in this manual or on the product:
1-2.Feature……………………………..…………………… 2

2. SPECIFICATION………………………………………….. 3 WARNING. Warning statements identify condition or


practices that could result in injury or loss of life.
3. PRECAUTIONS BEFORE OPERATION……...………... 6
3-1.Unpacking the instrument……..…………….………... 6
3-2.Checking the Line Voltage..…………………..……….. 6
CAUTION. Caution statements identify conditions
3-3.Environment……………………………………..…….. 8
or practices that could result in damage to this
4. PANEL INTRODUCTION……………………..…………. 9 product or other property.
4-1.Front Panel……………………………………………... 11
4-2.Rear Panel……………………………………….……... 12

5. OPERATION METHOD……..…………………………… 13 The equipment shall not be used for measurements


5-1.AC Withstanding Test…………………………………. 13 within category II, III and IV.
5-2.DC Withstanding Test…….…………………………… 14
5-3.Insulation Test…………...…..…………………………. 16
The following symbols may appear in this manual or on
6. MAINTENANCE………..…………………………………. 17
6-1.Fuse rating and type.…………………………………... 17 the product:
6-2.Cleaning………………..……………………………….. 17

DANGER ATTENTION Protective Earth (ground) Frame or


High Voltage refer to Manual Conductor Terminal Chassis
Terminal Terminal

iii iv
ELECTRICAL SAFETY TESTER ELECTRICAL SAFETY TESTER
USER MANUAL USER MANUAL

FOR UNITED KINGDOM ONLY The wire which is coloured Brown must be connected to the
terminal marked with the letter L or P or coloured Brown or
Red.
NOTE: This lead/appliance must only be wired by competent If in doubt, consult the instructions provided with the equipment
or contact the supplier.
persons
This cable/appliance should be protected by a suitably rated and
WARNING: THIS APPLIANCE MUST BE EARTHED approved HBC mains fuse: refer to the rating information on
the equipment and/or user instructions for details. As a guide,
IMPORTANT: The wires in this lead are coloured in accordance
cable of 0.75mm2 should be protected by a 3A or 5A fuse. Larger
with the following code:
conductors would normally require 13A types, depending on the
connection method used.
Green/ Yellow: Earth
Any moulded mains connector that requires removal
Blue: Neutral
/replacement must be destroyed by removal of any fuse & fuse
Brown: Live (Phase)
carrier and disposed of immediately, as a plug with bared wires
is hazardous if a engaged in live socket. Any re-wiring must be
As the colours of the wires in main leads may not correspond
carried out in accordance with the information detailed on this
with the colours marking identified in your plug/appliance,
label.
proceed as follows:
The wire which is coloured Green & Yellow must be connected
to the Earth terminal marked with the letter E or by the earth

symbol or coloured Green or Green & Yellow.

The wire which is coloured Blue must be connected to the


terminal which is marked with the letter N or coloured Blue or
Black.

v vi
ELECTRICAL SAFETY TESTER ELECTRICAL SAFETY TESTER
USER MANUAL USER MANUAL

1. PRODUCT INTRODUCTION 1-2. Features


1-1. Description The GPT/GPI-800 series offer several other features:
1) Current and output voltage setting
The GPT/GPI-800 Electrical Safety Testers (EST) are designed for
A safe way to set the current and output voltage without high
AD/DC Withstanding Voltage test and Insulation Resistance test (IR) in
voltage activated.
order to provide a safe and accurate test environment for the operator.
2) Easily and quickly setting with front panel control
With thoughtful design described in 1-2. Feature insures a safe operation
A user-friendly interface provides user an easy and quick way to
environment of high voltage test to protect user from hazardous impact.
set all parameters.
3) Testing time
The Electrical Safety Testers comply with the requirement of the
Use micro processor to control testing time.
electrical equipment & appliance control ordinances and JIS, CSA, UL,
4) Adjustable ARC detect level
BS and other overseas standards as well. The testers can be used for
The ARC detect level can be adjusted with front panel control.
withstanding voltage test of the various types of electrical and equipment
5) Adjustable output voltage during test
and components.
The output voltage can be adjusted during testing that add an
The GPT/GPI-800 series are based on the family of GW
operation flexibility.
withstanding voltage tester including AC Withstanding Voltage test, DC
6) An alert indicator for high voltage
Withstanding Voltage test and Insulation Resistance test. Please refer to
A flashing red LED indicates dangerous situation during high
the table as follows:
voltage output is activated.
7) Remote I/O controller
Function
AC DC IR In addition to the 9 pin remote I/O controller to control
Model
GPT-805 (500VA) ˇ START/RESET, it also provides PASS/FAIL/TEST signal for
GPT-815 (500VA) ˇ ˇ professional use.
GPI-825 (500VA) ˇ ˇ
GPI-826 (100VA) ˇ ˇ

1 2
ELECTRICAL SAFETY TESTER ELECTRICAL SAFETY TESTER
USER MANUAL USER MANUAL

2. SPECIFICATION (15℃~35℃ RH≦75%)


500VA
1) AC Hi-Pot Specifications (only for GPT-805/815, GPI-825) 100VA
Voltage Regulation 15% 3) AC Hi-Pot Specifications (only for GPI-826)
Voltage Range 0.200~5.00kV Voltage Regulation 15%
Voltage Accuracy ±3% of reading ±3 counts Voltage Range 0.200~5.00kV
Current cut-off range 0.3mA~100mA Voltage Accuracy ±3% of reading ±3 counts
Current Accuracy ≤1mA ±(5% of reading +40µA) Current cut-off range 0.3mA~20mA
>1 mA ±(5% of reading +20µA) Current Accuracy ≤1mA ±(5% of reading +40µA)
>1 mA ±(5% of reading +20µA)
Warning: The test time by using 5000V 100mA is within
180s, it needs 15 minutes interval after 180s of continuous Warning: The test time by using 5000V 20mA is within
testing. If proceed continuity test, the test voltage and 180s and it needs 15 minutes interval after 180s of
current must be at 5000V 50mA. continuous testing. If proceed continuity test, the test
voltage and current must be at 5000V 10mA.
2) DC Hi-Pot Specifications (only for GPT-815)
Voltage Regulation 20% 4) Insulation Resistance Specifications (only for GPI-825/826)
Voltage Range 0.200~5.00kV DC Voltage 500V/1000V
Voltage Accuracy ±3% of reading ±3 counts Resistance Range 1~2000MΩ
Current cut-off range 0.3mA~10mA 1~500MΩ: ±5% of reading ± 2counts
Resistance Accuracy
Current Accuracy ≤1mA ±(5% of reading +40µA) 501~2000MΩ: ±10% of reading
>1 mA ±(5% of reading +20µA)
Warning: The main purpose provided by the series of the
Warning: The test time by using 5000V 10mA is within 180s, instruments is for Puncture Testing.
it needs 15 minutes interval after 180s of continuous testing.
If proceed continuity test, the test voltage and current must
be at 5000V 5mA.

3 4
ELECTRICAL SAFETY TESTER ELECTRICAL SAFETY TESTER
USER MANUAL USER MANUAL

Warning: Unlike power supply which can be used on the 3. PRECAUTIONS BEFORE OPERATION
extreme low load range, when the series are used on the AC 3-1. Unpacking the instrument
voltage, the suggestive load should be more than 500VA at The product has been fully inspected and tested before shipping
50kΩ and 100VA at 250kΩ, while used on the DC voltage,
from the factory. Upon receiving the instrument, please unpack and
more than 500kΩ load is suggested.
inspect it to check if there is any damage caused during
transportation. If any sign of damage is found, notify the bearer
5) ARC Detect
and/or the dealer immediately.
500VA : 0.3~100mA
Detect Current
100VA : 0.3~20mA
3-2. Checking the Line Voltage
6) Interface
The instruments can be applied with any kind of line voltage shown
Terminal type Terminal seat
in the table below. Before connecting the power plug to an AC line
7) General Specification
outlet, make sure the voltage selector of the rear panel is set to the
Power Source AC115V/ 230V±10%, 50/60Hz
correct position corresponding to the line voltage. It might be
Indoor use, altitude up to 2000m.
Operation Relative Humidity 80% (Maximum). damaged the instrument if connected to the wrong AC line voltage.
Environment Installation category II
Pollution Degree 2 WARNING: To avoid electrical shock the power
Operation Temperature &
0℃ ~ 40℃, <70% cord protective grounding conductor must be
Humidity
Storage temperature & connected to ground.
-10℃ ~ 70℃, <70%
Humidity
Power cord × 1,
Accessories Test lead (GHT-105A) × 1, The equipment shall not be used for
Instruction manual ×1 measurements within category II, III and IV.
446(L) × 330(W) × 149(H) (m/m)
Dimension
GPT/GPI-8X5: 20kgs, GPI-826: 13kgs approx.
When line voltages are changed, replace the required fuses shown as
below:
* All specifications are guaranteed under the distortion of AC power
source less than 3%. Line voltage Range Fuse Line voltage Range Fuse
T 6.3A T 3.15A
115V 105-125V 230V 207-250V
250V 250V

5 6
ELECTRICAL SAFETY TESTER ELECTRICAL SAFETY TESTER
USER MANUAL USER MANUAL

Secure Testing
WARNING: To avoid personal injury, disconnect the Never operate the tester in the place with electric circuit device
power cord before removing the fuse holder. around.
The earth lead should be well connected in accordance with
Operator’s Precaution instruction. The Return Lead has to be connected to the tested object
(1) With immense high output voltage and current of the puncture first before linking up test probe. Do not plug the high voltage test
tester, only qualified person can operate the tester in order to probe to the high voltage output terminal before doing the testing.
avoid fatal electric shock. Also, do not touch the electric conductor of test probe and the
(2) On-job training is required for operator to better use the tester operator has to fully control the power on/off by using switch or
smoothly and safely. remote control, which should not be lay aside carelessly.
(3) The operator is prohibited to dress with metal ornaments or wear
metal decoration in order to avoid electric shock. WARNING: During the testing, do not touch the
(4) The person with cardiac or wear a pacemaker must not to operate tested object or any other connected objects.
the tester.
(5) Be sure to wear electric protective gloves whenever operating
this instrument, in order to guard against electric shock hazards.
3-3. Environment
(6) Before turning on the power switch, make sure that the TEST
The normal ambient temperature range of this instrument is from 0°
VOLTAGE dial is in the counterclockwise extreme position(Min.
to 40°C (32° to 104°F). To operate the instrument over this specific
voltage).
temperature range may cause damage to the circuits.
(7) Be sure to turn off the power switch each time the instrument is not
Do not use the instrument in a place where strong magnetic or
used even for a short period of time or when the operator leaves
electric field exists as it may disturb the measurement.
the instrument.
(8) Before touching the leads or output terminals, be sure to turn off
WARNING: This is a Class A product. In a domestic
the instrument
environment this product may cause radio
interference in which case the user may be required
to take adequate measures.

7 8
ELECTRICAL SAFETY TESTER ELECTRICAL SAFETY TESTER
USER MANUAL USER MANUAL

4. PANEL INTRODUCTION

Front Panel Rear Panel

9 10
ELECTRICAL SAFETY TESTER ELECTRICAL SAFETY TESTER
USER MANUAL USER MANUAL

4-1. Front Panel 4-2. Rear Panel


20. Fuse Holder with To change AC source voltage, pull the fuse holder and
1 Model Number Model number and description
Voltage Selector replace with an adequate fuse to connect with AC
2. Display Indicate all messages about test procedure.
power cord.
3. CAUTION During test the red LED will flash to indicate 21. Ground Terminal Connect Ground terminal to the earth ground.
Indicator LED dangerous.
22. Remote I/O Control Remote I/O output joint point:
4. START Button Press the green button to start a test procedure. Test signal: Two test holes has to be shorted when
5. RESET Button Press the red button to reset/stop a test procedure. getting into test mode by pressing Start
6. ARC VR Control ARC range. key, and it won’t return to open status
7. Current VR Control the range of Cut-off current. until the test result of PASS or FAIL
8. I Set/ARC Set : I Set, : ARC Set comes available, or the RESET key is
9. ARC Con./Stop : A signal for stop test. when the ARC is in use, pressed.
the test will be stopped. PASS signal: At the time when the test result comes
: A signal for continuous test. out to be PASS, two PASS holes has to
10. Range Switch over the range. be shorted, and return to open until the
11. Time VR Control test time. RESET key is pressed.
12. TIMER ON/OFF : Power on and the test time can be adjusted. FAIL signal: At the time when the test result comes
: Power off and the test time can not be adjusted. out to be FAIL, two FAIL holes has to
13. Test Voltage Test voltage adjustment. be shorted, and return to open until the
14. Setting (Current, : Can adjust Current/ARC VR to set current and RESET key is pressed.
ARC, & Test) ARC. RESET: The function is same as the RESET key on
: Stand by status for testing. the front panel.
15. AC/DC or AC/IR : AC withstand voltage test. START: The function is same as the START key on
: DC withstand voltage test (only for GPT-815) the front panel.
or Insulation Test (only for GPI-825/826). 23. Current Monitor The cutoff current can be directly monitored by
16. 500V/1000V : 500V, : 1000V(for GPI-825/826 only) disconnecting the shorting bar from these terminals
17. High Voltage Output High voltage output terminal and connecting a milliammeter (DC/AC) between
Seat them. The milliammeter should be capable of
18. RETURN terminal The test return terminal. measuring the cutoff current. Be sure to connect the
19. Power switch Power on to start test. shorting bar when no milliammeter is connected
between these terminals.

11 12
ELECTRICAL SAFETY TESTER ELECTRICAL SAFETY TESTER
USER MANUAL USER MANUAL

5. OPERATION adding test, adjust “Voltage knob” to maintain the value at 5.00.
5-1. AC Withstanding Test 15) If the test on the EUT is approval, the PASS indicator on the
Take GPT-805 as example, if want to output 5000VAC, set the cut- panel will be lighted up.
off current to 100mA and ARC to 50mA. During the 60s test time, 16) If the test on the EUT is disapproval, the FAIL indicator on the
the instrument will keep on testing even when the ARC appears. panel will be lighted up and the buzzer will blare out a warning.
1) Connect the Withstanding/Insulation Tester to the ground. Now remove the high pot probe from the EUT, and press
2) Adjust the “Voltage Knob” counter-clockwise to the lowest RESET to turn off the buzzer and FAIL light, then the
voltage position. instrument will be back to the initial status.
3) Set power switch to ON position.
4) Set CURRENT/ARC to position. 5-2. DC Withstanding Test
5) Set Range to 100mA range. Take GPT-815 as example, if want to output 5000VDC, set the cut-
6) Set to I Set position by pressing (8) key, then adjust the off current to 10mA and ARC to 10mA. During the 180s test time,
Current VR until the value of 100.0 is displayed on the panel. the instrument will stop testing when the ARC appears.
7) Set to ARC Set by pressing (8) key, then adjust ARC VR 1) Connect the Withstanding/Insulation Tester to the ground.
until the value of 50.0 is displayed on the panel. 2) Adjust the “Voltage Knob” counter-clockwise to the lowest
8) Set to Test by pressing (14) key. voltage position.
9) Set Timer to ON position, adjust the knob until the timer 3) Set power switch to ON position.
value is at 60s. 4) Set the key (15) to DC position.
10) Connect a test lead from the Return terminal (18) to the EUT 5) Set CURRENT/ARC of SETTING to position.
(equipment under test). 6) Set Range to 10mA range.
11) Connect a red high pot probe from “High voltage output 7) Set to I Set position by pressing (8) key, then adjust the
terminal” (17) to the EUT (equipment under test). Current VR until the value of 10.0 is displayed on the panel.
12) Press START key once to appear STBY(Stand by) on the 8) Set to ARC Set by pressing (8) key, then adjust ARC VR
display, then adjust (13) the voltage knob until the value of 5.00 until the value of 10.0 is displayed on the panel.
is display on the panel. 9) Set to Test by pressing (14) key.
13) Then, press START key to proceed testing, during testing, the 10) Set Timer to ON position, adjust the knob until the timer
warning indicator(3) will be flashing continuously. value is at 180s.
14) If the value displayed on the panel drops because of the load 11) Connect a test lead from the Return terminal (18) to the EUT

13 14
ELECTRICAL SAFETY TESTER ELECTRICAL SAFETY TESTER
USER MANUAL USER MANUAL

(equipment under test). 5-3. Insulation Test


12) Connect a red high pot probe from “High voltage output Take GPT-825 as example, set output to 1000V, but without
terminal (17)” to the EUT (equipment under test). specifying the test time.
13) Press START key once to appear STBY (Stand by) on the 1) Connect the Withstanding/Insulation Tester to the ground.
display, then adjust (13) the voltage knob until the value of 5.00 2) Adjust the “Voltage Knob” counter-clockwise to the lowest
is display on the panel. voltage position.
14) Then, press START key to proceed testing, during testing, the 3) Set power switch to ON position.
warning indicator (3) will be flashing continuously. 4) Set the key (15) to IR position.
15) If the value displayed on the panel drops because of the load 5) Set the Timer to OFF position.
adding test, adjust “Voltage knob (13)” to maintain the value at 6) Connect a test lead from the Return terminal (18) to the EUT
5.00. (equipment under test).
16) If the test on the EUT is approval, the PASS indicator on the 7) Connect a red high pot probe from “High voltage output
panel will be lighted up. terminal (17)” to the EUT (equipment under test).
17) During testing, if the ARC is occurred, both the lights of ARC 8) Press START key once to appear STBY (Stand by) on the
and FAIL will be on, and the buzzer will blare out a warning, display, then, press START key to proceed testing. During
now remove the high pot probe from the EUT, and press RESET testing, the warning indicator (3) will be flashing continuously.
to turn off the buzzer, ARC and FAIL lights, then the instrument 9) If the insulation resistance of the EUT is within 2000MΩ, thus
will be back to the initial status. the panel will display the insulation resistance value of the EUT,
18) If the test is over the jump-out current, the FAIL indicator on the and the testing is kept going until the RESET key is pressed.
panel will be lighted up and the buzzer will blare out a warning. 10) If the test on the EUT is disapproval, the FAIL indicator on the
Now remove the high pot probe from the EUT, and press panel will be lighted up and the buzzer will blare out a warning.
RESET to turn off the buzzer and FAIL light, then the Now remove the high pot probe from the EUT, and press
instrument will be back to the initial status. RESET to turn off the buzzer and FAIL light, then the
instrument will be back to the initial status.

15 16
ELECTRICAL SAFETY TESTER
USER MANUAL

6. MAINTENANCE

The following instructions are used by qualified person only to avoid


electrical shock, do not perform any service other than contained in the
operation instructions unless you are qualified to do so.

6-1. Fuse Rating and type


If the fuse blows, the product will not operate. Try to determine and
correct the cause of the blown fuse, then replace the fuse with
correct rating and type shown as below:

WARNING: For continued fire protection, replace only


with 250V fuse of the specified type and rating, and
disconnect the power cord before proceeding fuse
replacement.

6-2. Cleaning
To keep the instrument clean, wipe the case with a damp cloth and
detergent. Do not use abrasives or solvents.