General Specifications: Panel Mounted Converter For PH and Redox (ORP) Model PH150
General Specifications: Panel Mounted Converter For PH and Redox (ORP) Model PH150
FEATURES
● Compact panel mount design
● Interactive touch screen interface
● Trend displays pH, ORP & temperature
● Process temperature compensation
● pH and ORP simultaneous measurement
● Sensor diagnostic system
● HART communications (DD available)
● Event logbooks
● Programmable security codes
● Adjustable output damping
● IP65 (front panel)
● English language interface
● French, German, Spanish or Japanese as second language.
SYSTEM CONFIGURATION
GS 12B6B4-E-H
2nd Edition
2
General specifications of EXAxt PH150 K) Shipping details
A) Inputs specifications : One high impedance input Package size w x h x d : 180 x 161 x 243 mm
(≥ 1013Ω). One low impedance (7.1 x 6.3 x 9.6 inch)
input (≥ 108Ω). One temperature Package weight : app. 1.1 kg (2.4lbs)
input. One liquid earth input. L) Housing
B) Input ranges Enclosure : High quality chemical resistant
pH : -2 to 16 pH plastic front 96x96 mm.
ORP : -1500 to 1500 mV SS housing behind the panel
rH : 0 to 55 rH depth 98 mm behind the panel
Temperature : Pt1000, -30 to 140ºC (121 mm including cover)
: Pt100, -30 to 140ºC Mounting : Panel-mounted design in a standard
: 350Ω (DKK), -30 to 140ºC DIN-size 92x92 mm cutout.
: 5k1, -30 to 140ºC M) Power supply : 85-265 VAC (47-63 Hz) 10VA max
: 6k8, -30 to 140ºC 9.6-30 VDC 10W max
: NTC10k, -20 to 140ºC N) Regulatory compliance
: NTC 8k55, -10 to 120ºC EMC : Meets directive 89/336/EEC
: 3kBalco, -30 to 140ºC Emission conform EN 55022 class A
C) Accuracy Immunity conform IEC 61000-6-2
pH input : ≤ 0.01 pH Low Voltage : Meets directive 73/23/EEC
ORP input : ≤ 1 mV Conform IEC 61010-1, UL/cUL
Temperature : ≤ 0.3 ºC (≤ 0.4 ºC for Pt100) 3101-1 and CSA 22.2 No. 1010,
mA output circuits : ≤ 0.02 mA Installation category II, Pollution
Ambient temperature degree 2
influence : ± 0.05% /ºC Certification pending for cCSAus,
D) Transmission Signals Kema Keur and Geprüfte Sicherheit
General : Two isolated outputs of 4-20 mA. O) Environment and operational conditions
DC with common negative. Ambient temperature : -20 to +55 ºC
Maximum load 600Ω. Storage temperature : -30 to +70 ºC
Control function : Linear or 21-step table for pH, Humidity : Up to 90% RH at 40 ºC
temperature, ORP or rH. (non-condensing)
PID control. Environmental protection : IP65 (NEMA 4X) front panel, IP20
Burn up (21.0mA) or burn down behind the panel
(3.6mA) to signal failure. Data protection : EEPROM for configuration data
E) Contact outputs and logbook. Lithium cell for clock.
General : Two SPDT relay contacts with Watchdog timer : Checks microprocessor.
display indicators. Power down : Reset to measurement.
Switch capacity : Maximum values 100 VA, Automatic safeguard : Auto return to measuring mode
250 VAC, 5 Amps. when touchscreen is untouched
Maximum values 50 Watts, for 10 min.
250 VDC, 5 Amps.
Status : High/Low process alarms, Model Suffix Option Description
Code Code Code
selected from pH, ORP, rH and
PH150 Panel mount pH converter
temperature. Configurable delay -A 85 - 265 VAC power supply
time and hysteresis. -D 9.6 - 30 VDC power supply
Control function : On / Off -D Second language - German
PID duty cycle or pulsed frequency -F Second language - French
control. -J Second language - Japanese
FAIL alarm -S Second language - Spanish
F) Temperature compensation -AA Always AA
Function : Automatic or manual. /TAG Tagnumber
Compensation to Nernst
equation. Connection Diagram
Process compensation by
configurable temperature coefficient, High Voltage Signal
POWER SUPPLY OUTPUTS
NEN6411 for strong acids/bases or 61
programmable matrix.
Power 100-240VAC/10VA
mA1
+ mA Outputs
1 - 62
G) Calibration : Semi-automatic 1/2/3 point Supply L
2 65
calibration using pre-configured N +
mA2 66
NIST, US, DIN buffer tables 4, 7 & 3 -
63
9, or with user defined buffer tables, Shield
with automatic stability check.
Manual adjustment to grab sample. HIGH VOLTAGE SEE MANUAL
H) Serial Communication : Bi-directional HART digital Sensor Inputs
CONTACTS INPUTS
communication, superimposed on 11
mA1 (4-20) signal. 31 +
Relay C Temp
- 12
Device description available. Contacts 32
13
NC S1 Ref
I) Logbook : Software record of important 33
17
events and diagnostic data readily NO
Shield
41
available in the display. C 14
LE
42
J) Display : Graphical Quarter VGA (320 x 240 NC
S2
pixels) LCD with LED backlight 43
NO 15
and touchscreen. Plain (English) pH
240V/5A AC/DC 16
language messages, with choice 100VA/50W Shield
of alternative languages.
GS 12B6B4-E-H
3
13
40ºC
80ºC
OUTPUT AND ALARM FUNCTIONS
Two isolated mA outputs are provided, and can be set for linear
pH 11.00 at 25ºC or scaled output signals. Alternatively PID analogue control is
11
available on either or both mA outputs. The transmitted or control
parameter may be pH, ORP, rH or temperature. Control settings
9 becomes 9.80 pH at 80ºC are fully configurable.
0
pH Duty cycle
DISPLAYS AND OPERATING INTERFACE Proportional
The display is a large clear graphical LCD with LED backlight and Setpoint range % of output range Time
QVGA resolution. Operation is by touch screen. Graphical keys
on the right and other areas of the screen respond to contact as
2) PID pulse frequency control
virtual push buttons.
The pulsing frequency is regulated to control electrical valve
opening or pump stroke.
Main Screen In each case the setpoint, PB, I and D terms are all easily
adjustable in the PH150.
-Go
-Gototo
graph screen
trend screen
% controller output
-Go
-Gototodetail screen
detail screen
100
-Go
-Gototoinfoinfo
screen
screen Maximum pulse frequency
-Go
-Gototosetup screen
setup screen 0.3 s
Trend Screen
50
-Scale
-Scale 50 % pulse frequency
-Trend
-TrendLine
line
-Live
-LiveReading
reading
0
No pulses
-Time
-TimeScale
scale Proportional pH
Setpoint range % of output range Time
S S
GS 12B6B4-E-H
MAINTENANCE & CALIBRATION A) N.I.S.T. International Standard
It is important that a pH system should be well maintained. °C 5 10 15 20 25 30
The electrodes should be kept clean, and calibration needs to be 1.68pH NIST 1.668 1.670 1.672 1.675 1.679 1.683
checked at regular intervals. To ensure a high accuracy 4.01pH NIST 3.999 3.998 3.999 4.002 4.008 4.015
6.87pH NIST 6.951 6.923 6.900 6.881 6.865 6.853
measurement, careful calibration must be performed using
9.18pH NIST 9.395 9.332 9.276 9.225 9.180 9.139
accurate and stable buffers, taking into account the temperature
dependence of the solutions. B) DIN German Industry Standard
The EXAxt PH150 has 3 sorts of primary buffer solutions pre- °C 5 10 15 20 25 30
configured in memory. The user selects from these standards the 4.65 pH DIN 4.665 4.660 4.655 4.650 4.650 4.650
solutions to be used in the plant. The tables below are extracted 6.79 pH DIN 6.865 6.840 6.820 6.800 6.790 6.780
from the data programmed in the PH150. They show how much 9.23 pH DIN 9.425 9.370 9.320 9.270 9.230 9.180
variation exists, and why it is compensated. EXAxt can also have
C) Technical Standards (common in US)
three user defined buffer tables.
°C 5 10 15 20 25 30
4.0 pH US 3.998 3.997 3.998 4.001 4.005 4.001
7.0 pH US 7.090 7.060 7.040 7.020 7.000 6.990
10.0 pH US 10.245 10.179 10.118 10.06 10.01 9.97
Safety
25 (1")
Cover
Cut-Out
92x 92
Tol. +0,5/-0
(3.63" x 3.63"
Tol. +0.02/-0)
C/L
118 (4.6")
96 (3.8")
96 (3.8")
YOKOGAWA HEADQUARTERS YOKOGAWA CORPORATION OF AMERICA Yokogawa has an extensive sales and
9-32, Nakacho 2-chome, 2 Dart Road
Musashinoshi Newnan GA 30265 distribution network.
Tokyo 180 United States
Japan Tel. (1)-770-253-7000 Please refer to the European web-site
Tel. (81)-422-52-5535 Fax (1)-770-251-2088
Fax (81)-422-55-1202 E-mail: info@yca.com (www.yokogawa-europe.com) to contact
E-mail: webinfo@mls.yokogawa.co.jp www.yca.com
www.yokogawa.com your nearest representative.
GS 12B6B4-E-H
Subject to change without notice Printed in The Netherlands, 02-301 (A) Q