Qatar Petroleum Ngl-1 Operating Manual
Qatar Petroleum Ngl-1 Operating Manual
Qatar Petroleum Ngl-1 Operating Manual
SEC. : 9
NGL-1 OPERATING MANUAL PAGE : 1
ANALYTICAL CONTROL
1. GENERAL
2. ANALYTICAL INSTRUMENTS AND LOCATIONS
3. SAMPLE SCHEDULES
4. SAMPLING TECHNIQUES, SAFETY PRECAUTION AND PERPARATION
a) General
b) Adip Unit
c) Merox Unit
5. SAMPLE POINTS
6. AUTOMATIC QMI INSTRUMENTS
7. PORTABLE INSTRUMENTS
a) Operating Principles of Gas Detectors
b) MSA Portable Combustible Gas Detectors
1. General description
2. Principles of operation
3. MSA gas detector models
4. MSA model 2E OR 2 A portable gas detector
5. MSA model 40 portable gas detector
6. MSA tankscope portable gas indicator
7. Applications
8. Gas free
9. Operating instructions for Model 2E and 2 A
10. Operating instructions for Model 40
11. Interpretation of readings
12. Procedure for gas testing
13. Gas testing - points to look out for
1. Monocolor MC 1001
2. Monocolor MC 1002
QATAR PETROLEUM VOL : 1
SEC. : 9
NGL-1 OPERATING MANUAL PAGE : 2
Analytical Control
1. GENERAL
Within the NGL complex there are numerous requirements for determining the quality of
products and streams of process components.
The quality of the NGL feed, intermediate streams and final products is ascertained by
laboratory and on line analysis.
This is required in order to meet the product specifications and to monitor plant
performance.
Chemical compositions and concentrations must also be determined from the various
systems or substances.
To assess these qualities and quantities the substances have to be sampled, analysed
and reported.
The systems and methods, apparatus and collection procedures are varied, but should
follow the accepted standards of test methods.
The overall aspect and monitoring of the systems is controlled by the Senior Process
Technologist.
The Instrument Department are responsible for the servicing and maintenance of field
analysers and detection equipment, directed by the Instrument Engineer through the
quality measuring section of the Instrument Department [QMI].
The process and operating department, directed by the Senior Process Supervisor, will
ensure that the samples are collected safely, that equipment and systems are in
operational condition and that all safety procedural requirements are complied with.
Since the unit does not have the tankage provision for 'off- grade' products, it is of the
utmost importance that the quality of products is carefully monitored.
QATAR PETROLEUM VOL : 1
SEC. : 9
NGL-1 OPERATING MANUAL PAGE : 3
Analytical instruments used for quality control should be checked twice weekly against
standard samples.
A record of any deviation to the standard samples should be kept in the control room
and applied as required.
13-QR-4- A gas chromatograph, which is provided in the gas line from the
de-ethaniser reflux accumulator V- 1301 in order to analyse the C3 content
of the stream.
13-QR-5- An oxygen analyser, which is provided in the flue gas from heater F-1305
in order to analyse the gas for O2 content.
13-QR-11- A gas chromatograph, which is provided in the overhead gas line from the
depropaniser C-1302 in order to analyse the C2 and C4 content of the
propane stream.
13-QR-16- A gas chromatograph, which is provided in the overhead gas line from the
debutaniser C-1303 in order to analyse the C3 and C5 content of the
butane stream.
3. SAMPLE SCHEDULES
2. LPG to Extractor
H2S* Lead acetate paper 1/shift
H2S, RSH, COS UOP 212 Daily
3. LPG to Storage
RSH, COS UOP 212 Daily
Total sulphur ASTM D-2784 Daily
4. Prewash-Caustic
Gravity Hydrometer Daily
Total Alkalinity UOP 210 Daily
%Spent UOP 210 Daily
-----------------------------------------------------------------------------------------------------------------------------------
Stream/Component Method Frequency
------------------------------------------------------------------------------------------------------------------------------------
2. Gasoline to Extractor
H2S* Lead acetate paper 1/shift
H2S, RSH UOP-163 Daily
4. Prewash Caustic
Gravity Hydrometer Daily
Total Alkalinity UOP-210 Daily
% Spent UOP-210 Daily
1. Feed to Reactor
Distillation ASTM D-86 Daily
H2S and RSH UOP-163 Daily
specific gravity ASTM D-1298 Daily
Particulates ASTM-2276 Weekly
2. Treated Product
Doctor Test* UOP-41 3/daily
RSH UOP-163 Daily
***Peroxide number UOP-33 Daily
***Existent gum** ASTM D-381 Daily
3. Injection Caustic
Gravity Hydrometer 3/weekly
Total Alkalinity UOP-210 3/weekly
4. Aqueous Disposal
pH pH meter Daily
Total Alkalinity UOP-210 Daily
Acid oils misc. As required
Gravity Hydrometer Daily
a) General
What appears to be a relatively simple task can very quickly become extremely
hazardous to personnel and equipment.
All liquid sample points should be close to the ground to avoid spillage and to
assist in easy sampling. They should also be close to a drainage system.
Sampling should always be carried out with the utmost care and attention.
The discharge of LPG from a sampling bomb can cause static electricity. Sample
bombs must therefore be earthed effectively, whilst sampling or emptying.
Bonded copper wire is used for this purpose.
It should also be ascertained that drainage facilities are routed to a safe location.
Similarly gaseous sample systems should have excess vents routed to a safe
location and flare. Venting vapours to atmosphere should be eliminated. Such
sampling should be done slowly and methodically. The absence of sources of
ignition must always be checked before and during sampling.
The procedures laid down for sampling and storage of gases and low boiling
temperature liquids should be strictly adhered to. Regard should be taken of the
following recommendations.
1) Stainless steel high pressure 'bombs' with a capacity of 500 ml and 250 ml
should be used.
Samples containing sulphur compounds should not come into contact with
rubber, copper, iron or mercury.
Samples should not be taken from points other than designated sample
points : should it be absolutely necessary to take a sample, other than from
the normal point, precautions should be taken and authority given from the
Supervisor in-charge, with a designated operator standing by.
b) Adip Unit
For the sampling of lean adip solvent, it is advisable to use dark coloured glass
bottles or heat resistant pyrex bottles.
The bottle should be filled almost completely through an inlet tube reaching to the
bottom of the bottle.
The bottle should be well stoppered and kept away from direct sunlight.
Fat adip solvent samples should be taken under full plant pressure and avoid
losses of H2S and CO2.
The method used is to connect a stainless steel sample bomb, provided with two
valves, to the sample point and hold in a vertical position, inlet valve down.
Open the sample line valve and the inlet valve completely.
Carefully open the outlet valve to control the flow of solvent up through the bomb.
Flush the bomb thoroughly.
After filling the bomb, close the valves in reverse order i.e. first the outlet valve,
then the inlet valve and finally the sample line valve.
QATAR PETROLEUM VOL : 1
SEC. : 9
NGL-1 OPERATING MANUAL PAGE : 10
Depressurise the trapped liquid between sample line valve and the inlet valve of
the sample bomb, before disconnecting the sample bomb.
When sampling C5+ the bottle cap should be kept loose, unless it is being kept in
an ice bath. e.g. when sampling C5 only for colour.
c) Merox Unit
The propane, after removal of H2S and COS in the adip unit will carry
approximately 1000 ppmw of Mercaptan.
The following lists show the exact figures and the product specifications :-
Capacity
* Feed rates are based on a wellstream rate of 22.7 mmm3(st) / day (30mmm3
(st)/day)
[Merox unit]
-------------------------------------------------------------------------------------------------------------------------------------
PRODUCT CHARACTERISTICS PROPANE BUTANE TEST METHODS
-------------------------------------------------------------------------------------------------------------------------------------
Composition Liquid Vol.% ASTM D-2163
Ethane 2.5 max. Trace
Propane 95.0 min. 2.0 max.
iso - Butane )
n - Butane ) 2.5 max. 96.0 min.
Pentane & heavier 2.0 max.
Volatile Sulphur 30 ppmwt max. 30 ppmwt max. ASTM D-2784/SMS
Hydrogen Sulphide Negative Negative ASTM D-2420
(mod)
Copper corrosion No. 1 max. No. 1 max. ASTM D-1838
Residual matter 0.05 ml max. - ASTM D-2158
To be reported To be reported ASTM D-165
Vapour pressure
[@ 100oF, PSIG] 208 max. 70 max. ASTM D-1267
Moisture content Pass - ) GPA-propane dryness
test
) (Cobalt Bromide) or
ASTM )
) D-2713
Free Water Content - None )
Specifications for products when feed is from alternate source (Onshore or NGL-2) can be
obtained from laboratory.
QATAR PETROLEUM VOL : 1
SEC. : 9
NGL-1 OPERATING MANUAL PAGE : 13
[*] The Licensor has stated that the unit will reduce the mercaptan sulphur to
about 10 ppmw but this performance is not guaranteed.
QATAR PETROLEUM VOL : 1
SEC. : 9
NGL-1 OPERATING MANUAL PAGE : 14
There are various corrosive and potentially dangerous chemicals within the merox
system as follows :-
Sulphuric Acid
Caustic Soda
Acetic Acid
Ammonia
Reference should be made to the section dealing with the toxicological effects of H 2S in
Volume 4 of the manual.
The safety precautions for all of the corrosive chemicals must be strictly adhered to.
Eye protection, gloves and all other safety clothing must be used.
When dealing with any corrosive chemical, the operator should wear acid resistant suits.
Sampling connections are located throughout the system to allow the taking of gaseous
samples in bombs or liquid samples through a cooling coil.
QATAR PETROLEUM VOL : 1
SEC. : 9
NGL-1 OPERATING MANUAL PAGE : 15
QATAR PETROLEUM VOL : 1
SEC. : 9
NGL-1 OPERATING MANUAL PAGE : 16
5. SAMPLE POINTS
Various connections and designated sample points are located throughout the unit.
By referring to the P & ID's, information can be derived of official sampling locations and
types.
e.g. P & ID 6304-00-014 describes a typical sample point for taking gasoline samples
together with the location and cooling water supply.
P & ID 6304-00-015 gives a small sketch of a basic sample connection with a ball valve
and gate valve situation.
Other P & ID's give the above information and also the sample point location together
with where the sample vapours and excess liquid should go to.
Sample points should have valve facility for depressurising trapped liquids.
Analysers have been described in this section and a list of locations given. The
information received in the control room will determine the action to be taken.
17-AT-002 samples % O2 in the vent gas from V-1703 disulphide separator and gives an
audible alarm in control room on high (H) % O2 at 14%.
17-AT-005 analyses % O2 in the vent gas from V-1703 disulphide separator and gives
an audible alarm in the control room and also a trip signal to the plant air control
17-FIC-005 (17-XV-001) on a high high HH % O2 at 15.0%.
17-FIC-005 (17-XV-001) would also be tripped on a high high temperature on the inlet to
V-1703.
Rich caustic passes through a hot water heater E-1301 and is heated to 52C as
indicated on 17-TI-015.
Plant air through 17-FIC-005 joins the caustic flow and mixes on route to the oxidiser
V-1702.
The mixture of caustic, excess air and disulphide passes to the disulphide separator
V-1703.
The disulphides, regenerated caustic and excess air are separated in V-1703.
The mixture from V-1702 to V-1703 passes through 17-TA-007 [HH] which will give an
audible alarm in the control room on high high temperature and trip the plant air
controller 17-FIC-005.
The reason for this is that the higher temperature would reduce separation and leave
disulphide remaining in the regenerated caustic.
The excess air from V-1703 with nitrogen from the air used, flows via 17-PCV-007,
mixed with sweet fuel gas via 17-FIC- 006, flows to F-1305 heater fire box via solenoid
valve 17-XV-004.
When F-1305 is stopped the flow is diverted to V-1704, vent vessel via solenoid valve
17-XV-002, which opens. 17-XV-004 will close and also solenoid valve 17-XV-003 on
sweet fuel gas will close.
QATAR PETROLEUM VOL : 1
SEC. : 9
NGL-1 OPERATING MANUAL PAGE : 18
7. PORTABLE INSTRUMENTS
In various situations, operations personnel and laboratory personnel will be called upon
to use various kinds of portable instruments.
The instruments can be used to detect the presence of explosive gases, toxic vapour,
corrosive vapour etc. Other instruments are used to ascertain the presence of H 2S and
mercaptans and others to determine the amount of O2 in the system.
When drying out systems, especially those where the temperature may be low with
possibility of freezing, moisture analysers may be used.
Every day sampling like pH [acidity/alkalinity] and weathering [boil test] for propane or
butane can be carried out by operations personnel on a 24 hour basis.
As previously stated there are various reasons and requirements for determining the
quantity, quality and concentration of products or substances within the plant systems
and areas.
The correct use and application of the equipment used for these determinations is vitally
important for many reasons but predominately for SAFETY.
The user should be well versed in the proper methods of using the various models.
Another vitally important factor is the care and maintenance of the apparatus. The
following points should be emphasised :-
Don't abuse
Don't leave exposed to atmospheric conditions
Don't use for wrong application
Switch off to conserve batteries
Keep in a safe location
Keep a register of service, maintenance etc.
Deliver to laboratory on a regular basis for recharging and changing of batteries.
The following pages give principles and descriptions of some types of portable analysers
that are used in the plant, also their mode of operation.
Testing for flammable, explosive or hazardous gas/air mixtures with the aim to
prevent sires, explosions and injuries to personnel by suffocation and intoxication
is a daily practice in all Petroleum production/Refining operations.
It is therefore essential that gas-testers are fully acquainted with the working and
limitations of gas testing instruments.
Most hydrocarbon vapours and gases when mixed with air form a flammable or
explosive mixture.
These ratios are known for most vapours/gases and are expressed as
The L.E.L. is the minimum concentration of the vapour/gas in air required to form
an explosive mixture whilst the U.E.L. is the maximum concentration of
vapour/gas in air to form an explosive mixture.
This range between the L.E.L. and the U.E.L. is called the EXPLOSIVE range.
The mixture in the mid-range of the two limits will give the most violent explosion.
Unfortunately no single Gas detection instrument can be used to cover the whole
range of hydrocarbon mixtures. In Crude oil refineries for instance special
instruments are required where LEADED spirits [petrol] are concerned. Also
special precautions have to be taken when testing for hydrocarbons in INERT
atmospheres.
The following notes are on MSA gas detectors which are in use almost
universally within the Petroleum industry.
QATAR PETROLEUM VOL : 1
SEC. : 9
NGL-1 OPERATING MANUAL PAGE : 20
1) General Description
The MSA range of portable Combustible Gas Indicators are intended for
use in measuring the concentration of combustible gases or vapours in air.
2) Principles of Operation
Readings in this area should be treated with extreme care as the addition of
air will make the atmosphere violently explosive.
The Model 40 uses the same principal as the Model 2E, indeed it is exactly
the same except for one difference, it has two indicating scales :
When using this instrument ALWAYS use the 0 to 100% scale in the first
instance, if the reading obtained is above 10% L.E.L. DO NOT USE the 0 to
10% scale until the readings obtained fall below the 10% mark.
The 0 to 10% scale can be used to monitor atmospheres below 10% L.E.L.
This is not to say the Model 40 is more accurate than the Model 2E, it just
has a finer scale and is therefore more sensitive below 10% L.E.L. having a
full scale deflection from 0 to 10%.
The Tankscope gas indicator has been designed specifically to detect and
measure the concentration of hydrocarbon vapours in INERT atmospheres,
that is to say atmospheres that have been purged from a total hydrocarbon
condition with N2.
The Tankscope instrument has two filaments each one being part of an
electrical circuit. The instrument is balanced, i.e. "zeroed" in fresh air, which
is free of hydrocarbon gases. When the gas sample is drawn into the
instrument only one filament is exposed to it. If hydrocarbon gas is present
it will unbalance the circuit and so indicate volume percentage hydrocarbon
gas present.
There are two points about this instrument which must be understood by
any gas tester.
7) Applications
There are three primary reasons for wanting to know gas concentrations
If work had to be done within a vessel the reading obtained with the Model
2E would still be safe for the Hot work, but now a different or additional
standard must be met. To do the work a man or men must enter the vessel
and more rigid standards must be applied. An atmosphere in which men
have to work has to meet a standard called the "Threshold Limit Value" or
T.L.V.
Both above applications are used to determine the percentage L.E.L. of the
atmosphere being tested therefore air [normal oxygen concentrations] is
present.
8) Gas Free
b) Valve isolate then fill to a slightly elevated pressure with N2 and again
release to flare. If this is done two or three times it dilutes the
hydrocarbon atmosphere with N2 making it safer to SPADE isolate.
c) Spade isolate the equipment. This ensures that no means remains for
anything to enter the equipment except for the N2 inlet and the flare
outlet. Once spade isolation is done pressure and depressure with N 2
three or four times, then when equipment is at atmospheric pressure,
sample for combustible gas concentration. As the atmosphere in the
equipment is N2 plus hydrocarbon gases neither the MSA Model 2E or
40 is applicable. The instrument applicable here is the Tankscope.
The MSA Model 2E gas detector is set in its proper operating conditions by
the adjustment of a single control. This control is a rheostat regulating the
current to the meter measuring circuit. The rheostat knob is held in the off
position by a locking bar. This bar must be lifted before the knob can be
turned from the off position.
a) Lift the end of the rheostat knob on-off bar and turn the rheostat knob
one quarter turn clockwise.
The circuit of the detector must be balanced with fresh air free of
combustible gases or vapours surrounding the detector filament. Five
squeezes of the aspirator bulb is sufficient to flush the combustion
chamber. If a sampling line is attached an additional two squeezes will
be required for each 10 feet [2.5 metres] of line.
c) Adjust the rheostat knob until the meter pointer rests at zero, holding
the meter against wind direction.
d) Place end of sampling line at, or transport the meter to, the point where
the sample is to be drawn from.
f) Connect the sampling line and aspirate sample through meter until
highest reading is obtained.
g) To turn the instrument off : flush with fresh air so that combustion
chamber is clear of combustible gases or vapours, and indicator returns
to ' 0' reading. Rotate rheostat knob counterclockwise until the locking
bar drops into position in its slot.
This places the meter across the active filament where it will
indicate the voltage across the filament.
This operation closes the battery circuit, and the meter pointer will
be seen to rise on the scale towards the "check" point.
To accomplish this turn the "volt adj” in the direction which the
meter pointer must move to come to "check". This establishes the
proper operating voltage to the bridge.
Turn the "zero adj" knob so that the meter pointer slightly
overshoots zero then turn the knob slowly in the opposite direction
until the meter pointer is on zero.
10. Place the end of the sampling line in the area to be tested.
11. Aspirate sample through the indicator. The aspirator bulb should
be operated until the meter pointer comes to rest on the scale.
12. If after 10 squeezes the reading obtained is below 10%, turn the
selector switch to the "10% L.E.L." scale. Continue aspirating and
take readings from top scale, ["10% L.E.L."].
Caution: Always take the first sample with the selector switch in
the 100% L.E.L. position.
13. If the meter pointer does not rise on the scale within 15 squeezes
of the aspirator bulb the point of sample may be considered "gas
free".
QATAR PETROLEUM VOL : 1
SEC. : 9
NGL-1 OPERATING MANUAL PAGE : 27
No details available
When interpreting meter readings, the indicator is set for its maximum
range when the selector switch is in the 100% L.E.L. position. With the
selector switch in this position,
The indicator is set for its maximum sensitivity when the selector switch is
in the "10% L.E.L." position. Meter readings on this scale indicate that only
traces of "combustibles" are present, usually not above 10% of the lower
explosive limit. With the switch in this position the concentration of many
combustible gases and vapours may be measured to give an indication of
the toxicity of the atmosphere as related to their Threshold Limit Value
(T.L.V.).
Test Site
a) Check wind direction to give indication where vapours are likely to drift
over the work area.
b) Inspect all sewers, beehives and drains. Ensure that they are properly
sealed, and test over them.
c) Test vents, open lines, flanges, drain cocks and valve glands or any
other points where there is the possibility of gas leakage [vapour or
liquid].
d) In the case of proposed burning through lines, gas tests must be taken
at frequent intervals in the line and there should always be a nil reading
indicated by the explosimeter. It may be necessary to drill test holes in
the line prior to the hot work.
QATAR PETROLEUM VOL : 1
SEC. : 9
NGL-1 OPERATING MANUAL PAGE : 28
e) When testing : vessels, drums or tanks take tests from as many points
as possible.
1. 6" from the tank bottom [take care the sampling tube is not in liquid].
FLOATING ROOF TANKS, Test through the bottom manways, and test
all compartments in the pontoon. When tank is gas free check that no
vapours are drifting into the area from other sources, e.g. tanks, sewers,
vents, etc. Take tests on top of adjacent tank firewalls.
In the case of hot work on the floor plates of tanks, holes should be
drilled in the floor to ensure that the space between the ground and the
tank bottom is gas free.
f) Do not take tests in lines, vessels, tanks etc., while steam is present as
it is impossible to obtain an accurate reading and the liquid trap in the
explosimeter will become filled with condensed vapour.
General Points
Check the fittings. Do not use "home made" extensions. The sample lines
supplied are either flexible non-absorbent type or stainless steel.
Non-standard connections can allow air to enter and dilute the sample. A
tightness check can be made by squeezing the bulb and closing the inlet
with a finger, the bulb should not reinflate.
These notes are intended as guides when gas testing. They cannot hope to
cover all eventualities. Should you be in doubt or feel unsure about results
obtained ..."ask for help". Do not allow work to start until everything has
been checked out satisfactorily.
3. First, indication "OXY" + 20.8% vol. appears [for oxygen tests only].
6. Next two pressings indicates, TWA & Hr. [ignore these, because these are
time base analysis, required during vessel entry].
7. Your final pressing of SELECT switch gives you "EXP" 0% i.e. explosivity %
LEL.
9. Keep the exotox in its case and allow the gas to flash in the case from the
loose end of the tube connected to the high pressure line.
Allow gas entry at very low pressure till a stable reading is obtained.
10. Before turning the instrument off, flush it with fresh air (Instrument Air).
Note
1) General
More than 50 years ago, this desire led to the development in the USA of a
simple analysis unit with which it was attempted to develop a laboratory
process to such an extent that it could be used as a complete analysis
when required.
The stimulus for this came from the Mining industry. An instrument was
required which would give an early warning of the presence of poisonous
carbon monoxide. Laboratory analysis took too long and it was no longer
sufficient to use small animals - white mice and canaries - as a means of
warning.
This was a great step forward but it was not until the 1930's that a paper
was published for the detection of H2S [by a Mr. Littlefield of the USA].
However the detector tube was still not in general use at that time because
of problems associated with the developing of suitable reagents which
would exist within glass tubes for some time without deteriorating.
QATAR PETROLEUM VOL : 1
SEC. : 9
NGL-1 OPERATING MANUAL PAGE : 32
Without getting too technical, the problem is based on the minute amount of
reactant chemical which is used within the tube. This small amount of pure
chemical is "carried" on another substance which is neutral in the reaction.
Not only has the "carrier" to remain neutral during the sampling reaction but
it has to remain neutral to the reactant chemical itself during the length of
time a tube is kept in storage awaiting use. If these two conditions are not
met then the tube would be useless. That is why it is extremely important,
even with modern-day tubes, to ensure they are not "out of date" before
using them.
Most Drager tubes have a storage life of two years but this may vary as
there are now some 100 different Drager detection tubes. The expiry date is
clearly shown on each box.
1. Hydrogen sulphide
2. Oxygen
3. SO2
4. CO
5. CO2
6. Ammonia
7. Hydrocarbons
For normal operations work where vessel entry or detection of toxic gases
is concerned Nos. 1, 2, and 7 above are those mainly used.
As stated above the Drager gas detection tube uses a reagent chemical
within a glass tube to detect the presence of another chemical substance.
Whilst this would say the substance is there if a complete colour change
took place it would by no means tell us how much of that substance was
present. A Drager detection tube does this using two principles
By having the reactant chemical mixed uniformly with its neutral carrier and
this total amount of material distributed along a tube of a known volume it
only then remains for the pump to intake a set volume of the air being
sampled for the reactant to be discoloured progressively from the sample
inlet point. The tube can then be graduated so that as the discolouration
progresses along the tube for a certain volume of sample intake. The
contaminant quantity can be directly determined from the tube
discolouration length [see attached diagram of "before" and "after" tube].
The gas detector bellows pump [see diagrams of pump and pump with
sample tube attached] is made from neoprene and opens automatically
after compressing and releasing the bellows. This opening process is
effected by two steel springs built into the pump. The end of the suction
process is reached when the limit chain is taut.
The gas detector pump has only one valve, which is closed when the gas
sample is sucked in and opens again on squeezing the bellows. The pump
head has an aperture into which the Drager tube to be used is inserted.
The time from releasing the bellows [after squeezing] until the limit chain is
taut is termed the opening time of the gas detector pump: hence the
opening time is the duration of one pump stroke.
The opening time depends on the flow resistance of the Drager tube
inserted, which is a function of the filling preparation used. The reaction
time for various reagents and gases differ. Therefore the opening time of
the gas detector pump will vary. With some tubes, H 2S for instance, it will
open within about 20 seconds, with others it could be up to 40 seconds.
Generally speaking the tubes used within the petroleum industry open in
about the same times. The opening times are indicated on the instruction
leaflet in the box for each tube type.
In most cases tubes should not be opened until at the sampling site and the
test started within ten to fifteen seconds of opening the tube. Read the
instructions given in each box of tubes carefully before carrying out tests,
and be especially careful when evaluating test results. These instructions
are a general guide to the use of the Drager sampling instrument but each
type of detector tube has a very definite individual measurement
characteristic determined by the number of strokes and the opening time for
each stroke. Full instructions are given on the sheet within each box.
QATAR PETROLEUM VOL : 1
SEC. : 9
NGL-1 OPERATING MANUAL PAGE : 34
Each test made by a gas tester is used to determine whether or not work
can take place or a man can enter a space, the result of the test is recorded
on the Work Permit. In the case of the MSA gas detector no permanent
record of the test result can be made, those responsible for issuing a Work
Permit should always be sure the Performing Authority witnesses the actual
test and signifies this on the permit. In the case of the Drager test a record
can be kept of the test. When the test has been completed the ends of the
test tube should be sealed with the rubber caps provided and the tube
labelled clearly as to place, time, and tester together with performing
authority's witness to authenticity.
Before using the Drager gas detection unit perform a leak test. To do this
seal the pump with an unopened DRAGER tube and completely compress
the bellows. The pump is sufficiently air tight if the bellows has not
expanded again completely after ten minutes i.e. the limit chain is not taut,
it will then deliver the volume required for completing a test.
Break off both tips of the Drager tube using the break-off eyelet [refer to
attached diagrams].
Insert the opened Drager tube in the pump head SO THAT THE ARROW
POINTS TOWARDS THE PUMP. The tube must fit firmly and tightly in the
pump head stopper so that no bypass air can be sucked in.
Hold the pump in the hand with the pump between the thumb and the base
of the index finger and the fingers resting on the front plate [Fig 6].
Compress the bellows completely [as in Fig 7, not as Fig 8], then release it
[as in Fig 9].
When the bellows are compressed, the air in it escapes through the outlet
valve and not through the tube, since the tube has a higher resistance than
the outlet valve. The pump suction operation commences when the fingers
are relaxed. The compression springs inside the bellows, which are placed
under stress when the bellows are compressed, extend and the outlet valve
is closed through the vacuum created in the bellows.
The air flows through the Drager tube into the bellows while the latter
returns to its original volume [Fig 9]. The volume of air sucked in by the
Drager tube is defined by the dimensions and stroke of the bellows. It
amounts to 100 cm3 [one hundred cubic centimeters] per stroke. The end
of the suction movement is reached when the limit chain is completely taut.
Since the suction of the pump is caused only by the relaxation of the
springs, any subjective influence is excluded.
Consequently, the rate of flow of the air in the Drager tube, which is
decisive for the accuracy of measurement, is determined exclusively by the
spring force of the pump and the resistance of the tube filling, which is fixed
QATAR PETROLEUM VOL : 1
SEC. : 9
NGL-1 OPERATING MANUAL PAGE : 35
during manufacture. To check that testing takes place correctly, check the
"Opening Time" given on the Instruction sheet within each packet of tubes.
When testing, make the number of pump strokes specified in the operating
Instruction for the tube concerned.
b) Break off both tips of the drager tube at test site with the break off eyelet
[Fig 3].
c) Tightly insert the drager tube in the pump head with the arrow pointing
towards the pump [Fig 5].
Test for the highly toxic gases firstly e.g. [ 1 ] H2S, (2 ] CO, [ 3 ] CO2 .
When testing unknown gas concentrations Breathing Apparatus [B.A.]
should be used-and a lifeline worn with a watch being kept by standby
personnel in a safe area.
Remember that when testing atmospheres for man entry, after the
absence of TOXIC AND HYDROCARBON GASES are proven, oxygen
capable of supporting life MUST be established and proven before entry
is permissable without B.A.
6) Limitations
ii) Allow the gas to run into an open receptacle (hand glove will do)
and take the reading by inserting the Drager tube inside the
receptable which is at atmospheric pressure.
QATAR PETROLEUM VOL : 1
SEC. : 9
NGL-1 OPERATING MANUAL PAGE : 37
QATAR PETROLEUM VOL : 1
SEC. : 9
NGL-1 OPERATING MANUAL PAGE : 38
QATAR PETROLEUM VOL : 1
SEC. : 9
NGL-1 OPERATING MANUAL PAGE : 39
QATAR PETROLEUM VOL : 1
SEC. : 9
NGL-1 OPERATING MANUAL PAGE : 40
This is an extremely sensitive but nevertheless robust gas analyser for the
determination of traces and also higher concentrations of H 2S in gases or
air. The exceeding of the maximum permissible concentration of H 2S can
be signalled by visible or audible alarms.
a) Measuring Principle
The photoelectric cell [L1] receives the light reflected by the unstained
paper in the reference chamber, whereas light reflected by the more or
less stained paper in the measuring chamber [M] is received by the
photoelectric cell [L2]. The resulting differential current is proportional to
the H2S concentration in the test gas and measured by the instrument
[E].
QATAR PETROLEUM VOL : 1
SEC. : 9
NGL-1 OPERATING MANUAL PAGE : 41
b) Mode of operation
A second similar chamber [V] communicates with the ambient air and
serves as a comparison chamber. The principle of measurement is
described earlier. One roll of sensitized paper is sufficient for 15 days.
Shortly before the paper is moved forward a contact is closed which
electromagnetically actuates the chopper bar of the recorder for
recording the measured value.
The recorder is designed for flush panel mounting and uses a moving
coil system with a special chopper bar actuator which is controlled by
the transmitter. The chart paper is 140/120 mm wide and has a feed
rate of 20 mm/hour. It is graduated from 0 to 25. The actual reading is
determined by the scale of the recorder, for example 0...25 mg/m3 [0-16
ppm] or 0…25 x 0.1 g for 0…25 g/m3 [0-1600 ppm] or 0…25 x 0.2 g/m3
for 0…5 g/m3 [0-0.32%], etc.
QATAR PETROLEUM VOL : 1
SEC. : 9
NGL-1 OPERATING MANUAL PAGE : 42
One chart roll is sufficient for 1 1/2 months. The chart is wound up
automatically.
[Ex] type : The MONOCOLOR can be supplied as ex- type [fig. 6] for
use in hazardous areas. This type has been approved in accordance
with PTB certificate No. 17358 of October, 67 for the protection class
[Ex] d 2 G 5.
c) Installation
The connecting line between the gas sampling point and the analyser
should be Monoflex tubing, 3.5 x 9 mm [a flexible plastic] run in a conduit.
d) Technical Data
Test paper : impregnated with lead acetate. One test paper roll will
last about 15 days
Weight : 27.5 kg
[Ex] type : 38 kg
2) Monocolor MC 1002
a) Mode of Operation
For measuring H2S in gases containing SO2 for example in flue gas, the
MONOCOLOR MC 1002 is used. This instrument corresponds
essentially to the preceding description and is different from the MC
1001 only in the following respect.
1. The test paper is impregnated with silver nitrate rather than with
lead acetate for eliminating the interfering influence of SO 2 on the
measurement. The shelf life of this test paper is however limited to
about 3 months in view of its light sensitivity.
2. The test paper feed is not fixed at 1 mm per minute as the test
paper is moved forward in steps of a full gas exposure field [about
16 mm]. By varying the exposure time [from 1 min. to 16 min.] by
means of a time switch, different measuring ranges can be realised
[from 6.25 to 100 ppm].
b) Installation
The sample gas is drawn from the stack through a gas sampling probe
with heated filter Ga 54. The gas sampling tube made of stainless steel
[V4A] should be installed in downward slope [about 15o] so that any
deposits of ashes will drop out of the tube. The ceramic filter attached to
the end of the sampling tube is electrically heated to about 170C in
order to avoid clogging of the filter by condensing acid.
The sample gas line between the heated filter and the sample gas
cooler must also be heated to about 60 oC in order to avoid condensate
in this section of the line. For this purpose, the sample gas line should
be run in a conduit together with heating strap of about 100 watt.
The corrosion resistant electrical gas cooler Fi 57 cools the gas to about
+ 7C so that most of the water vapour is condensed. The condensate
accumulates in the water separator Fi 19 and is automatically drained.
The relative error as a result of the dissolving of H 2S in the condensate,
is only 0.17% for sample gas water saturated at 40C or 0.49% for
sample gas water saturated at 60C; hence negligibly small.
The gas pump Gf 21 delivers the sample gas from the gas cooler
through a fine dust filter Fi 46, which is essential for protecting the
analyser, into the MONOCOLOR MC 1002.
Monoflex tubing Le 29 or Teflon tube should be used for the sample line
and installed in a conduit as mechanical protection.
QATAR PETROLEUM VOL : 1
SEC. : 9
NGL-1 OPERATING MANUAL PAGE : 49
QATAR PETROLEUM VOL : 1
SEC. : 9
NGL-1 OPERATING MANUAL PAGE : 50
c) Technical Data
Power : 220V + 150; 50Hz
Upon enquiry for other connection
Test paper : Impregnated with silver nitrate, Ve 55a, shelf life about 3
months. Depending on the measuring range, a test paper
roll will last for one to 15 days.
Weight : 27.5 kg