PT Compact PT Compact Plus: Description Features
PT Compact PT Compact Plus: Description Features
PT Compact PT Compact Plus: Description Features
PT Compact
PT Compact plus
Screw - in resistance thermometer
Description Features
The resistance thermometer „PT Compact“ is designed for O compact dimensions
temperature measurement in processes with medium and O easy handling
low pressure. Measurement ranges between -200°C and
O cost-efficient
+600°C cover the majority of your temperature measuring
tasks. O short delivery times
O optional with transmitter 4-20mA or 0-10V
O service friendly
The thermometers are available with platinum sensors O customized solutions
PT100 or PT1000 in 2 -, 3 -, 4 or 2x2-wire connections. If
the measuring signal has to be transmitted over longer
distances, the PT Compact plus is available with an
integrated transmitter. This transmitter converts the
Models
resistance of the platinum sensor into a linear temperature
O -50°C up to +200°C (-60...400°F)
signal of 4-20 mA or 0-10 V.
O -50°C up to +400°C (-60...750°F)
O -50°C up to +600°C (-60...1100°F)
Different process connections, as well as adjustable O -200°C up to +600°C (-300...1100°F)
compression fittings underline the variability of this
measuring instrument. To achieve fast reaction times a
version with a tapered stem is also available. Measuring ranges (with transmitter)
O 0...50°C; 0...100°C; 0...120°C
The electrical connection is realized with an L-plug O 0...200°F; 0...500°F; -50...300°F
according to DIN EN 175301-803A. Optionally an M12x1
O according to customer requirements
connection is available.
Models: TEP11
TES11, TES12, TES13, TES14
tecsis GmbH DE 1127 l
Carl-Legien Str. 40-44 01/2014
D-63073 Offenbach / Main Sales national Sales international e-Mail: info@tecsis.de
Tel.: +49(0) 69 / 5806-0 Fax: +49(0) 69 / 5806-8002394 Fax: +49(0) 69 / 5806-8002394 Internet: www.tecsis.de p. 1 / 4
Technical data
0-10 V, 3-wire
supply voltage: 12 – 30 V DC,
ripple < 10%
Error signal sensor burnout: 23mA
sensor short circuit: 3.3 mA
Temperature standard: -50°C up to +200°C (-60...400°F)
range -50°C up to +400°C (-60...750°F)
-50°C up to +600°C (-60...1100°F)
-200°C up to +600°C (-300...1100°F)
Measuring range see temperature range see temperature range selectable measuring range:
0°C up to +50°C
0°C up to +100°C
0°C up to +120°C
standard version:
minimum range 50K
maximum range 250K
•
1)
special solid machined parts for pressure up to 600 bar
Accuracy transmitter: <0.5% of measuring range
Ambient at connector max. 125°C
temperature with transmitter max. 85°C
Storage -40°C up to +85°C
temperature
Electrical L-plug acc. to DIN EN 175301-803 form A
connection optional: round connector, 4-pin, M12x1
EMC-resistance emitted interference acc. DIN EN 61326
immunity acc. to DIN EN 61326
Protection class IP65 acc. to DIN EN 60529 / IEC 529
1)
Pressure ranges refer to static pressure; Rating depends on:
- process medium
- process pressure and temperature
- flow rate
- Stem design (length, diameter, wall thickness)
DE 1127 l
p. 2 / 4
Dimensions
L-plug acc. to
DIN EN 175301-803A
~28
silicone seal
~80
~52
adjustable
compression
fixed thread acc. fitting
to DIN 3852
stem length EL
tapered stem ∅ 3
for fast response-time
Wiring Diagrams
4 1
4..20mA: 1=Ub+/S+ 4 3 2
3=0V/S-
0..10V: 1=Ub+
3=0V/Ub- 1 1 3
4=S+
4
DE 1127 l
p. 3 / 4
Configuration PT-Compact