Manual Dm5e
Manual Dm5e
Manual Dm5e
DM5E
Operating Manual
DM5E
Operating Manual
GESensingInspection.com
2009 General Electric Company. All rights reserved. Technical content subject to change without notice.
Contents
iii
Contents
iv
Contents
Chapter 7. Specifications
Contents
Chapter 8. Maintenance
8.3 Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
vi
Preface
Important Notice
The following information must be read and understood by any user of a GE Sensing & Inspection Technologies ultrasonic thickness gauge. Failure to follow these instructions can lead to errors in thickness measurements or other test results. Decisions based on erroneous results can, in turn, lead to property damage, personal injury or death.
General Warnings
This operating manual provides instruction in the basic setup and operation of the thickness gauge. There are, however, additional factors which affect the use of ultrasonic test equipment. Specific information regarding these additional factors is beyond the scope of this manual. The operator should refer to textbooks on the subject of ultrasonic testing for more detailed information.
vii
Preface
Operator Training
Operators must receive adequate training before using ultrasonic test equipment. Operators must be trained in general ultrasonic testing procedures and in the setup and performance required by a particular test. Operators must understand:
Behavior of the sound wave where two different materials are in contact
More specific information about operator training, qualification, certification, and test specifications is available from various technical societies, industry groups, and government agencies.
viii
Preface
Testing Limitations
In ultrasonic testing, information is obtained only from within the limits of the sound beam. Operators must exercise great caution in making inferences about the test material outside the limits of the sound beam. For example, when testing large materials it may be impossible or impractical to inspect the entire test piece.
When a less-than-complete inspection is to be performed, the operator must be shown the specific areas to inspect. Inferences about the condition of areas not inspected, based on data from the evaluated areas, should only be attempted by personnel fully trained in applicable statistical and probability techniques. In particular, materials subject to erosion or corrosion, in which conditions can vary significantly in any given area, should only be evaluated by fully trained and experienced operators.
Sound beams reflect from the first interior surface encountered. Because of part geometry and overlapped flaws or overlapped surfaces, thickness gauges may measure the distance to an internal flaw rather than to the back wall of the material. Operators must take steps to ensure that the entire thickness of the test material is being examined.
ix
Preface
The following operating procedures must be observed by all users of ultrasonic thickness gauges in order to minimize errors in test results.
The principle of operation of an ultrasonic thickness gauge is that the instrument measures the time of flight of an ultrasonic pulse through the test piece and multiplies this time by the velocity of sound in the material. Thickness measuring error is minimized by ensuring that the sound velocity to which the instrument is calibrated is the sound velocity of the material being tested. Actual sound velocities in materials often vary significantly from the values found in published tables. In all cases, best results are obtained if the instrument is calibrated on a velocity reference block made from the same material as the test piece; this block should be flat and smooth and as thick as the maximum thickness of the test piece.
Operators should also be aware that the sound velocity may not be constant in the material being tested; heat treating, for example, can cause significant changes in sound velocity. This must be considered when evaluating the accuracy of the thickness provided by this instrument. Instruments should always be calibrated before testing, and the calibration should be checked after testing, to minimize testing errors.
Before performing calibration with a contact probe, the probe zero procedure must be performed first. If the zero procedure is not performed or if it is performed improperly, inaccurate thickness readings will result.
Preface
Temperature variations change the sound velocity of materials and transducer delay lines and, therefore, calibrations. All calibrations should be performed on-site and with test blocks at the same temperature as the test piece, to minimize errors due to temperature variations.
4. Transducer Selection
The transducer used in testing must be in good condition, without noticeable wear of the front surface. Badly worn transducers will have a reduced effective measuring range. The specified range of the transducer must include the complete range of thicknesses to be tested. The temperature of the material to be tested must be within the transducer's temperature range.
5. Use of Couplants
Operators must be familiar with the use of ultrasonic couplants. Testing skills must be developed so that couplant is used and applied in a consistent manner to minimize variations in couplant layer thickness and errors in test results. Calibration and actual testing should be performed under similar coupling conditions, using a minimum amount of couplant and applying consistent pressure on the transducer.
xi
Preface
6. Doubling
Ultrasonic thickness gauges will, under certain conditions, display readings which are twice (or in some cases, three times) the actual material thickness being measured. This effect, commonly known as doubling, can occur below the minimum specified range of the transducer. If the transducer being used is worn, doubling is possible at a thickness greater than the minimum of the specified range.
When using a new transducer, any reading which is less than twice the minimum specified range of the transducer may be a doubled reading, and the thickness of the material being tested should be verified by the use of other methods. If the transducer shows any sign of wear, doubling may occur at a second echo, or other echo signal combinations may produce a readable signal. The instrument reading and apparent thickness are up to about twice the actual value, resulting in a thickness greater than twice the minimum of the specified range. This thickness should be determined by calibrating the instrument/transducer combination on reference blocks that represent the complete range of possible thicknesses that may be encountered in testing. This is particularly important when a test piece is being ultrasonically measured for the first time or in any case where the history of thickness of the test specimen is unknown.
xii
Preface
Safety Information
IMPORTANT: The DM5E is an instrument for materials testing. Any use for medical applications or other purposes is not allowed. The DM5E may only be used in industrial environments.
IMPORTANT: For battery operation of the DM5E, two AA size batteries are needed. You should only use the products recommended by us for battery operation. Use of well-known battery brands is highly recommended.
IMPORTANT: Disposal of all components of the DM5E and its batteries must be done in compliance with all applicable regulations. See Appendix B, Regulatory Compliance, for detailed information.
Software
According to the current state of the art, software is never completely free from errors. Before using any software-controlled test equipment, please make sure that the required functions operate perfectly in the intended combination.
xiii
Preface
If you have reason to believe that safe operation of your DM5E is no longer possible, you must disconnect the instrument and secure it against unintentional reconnection. Remove the batteries if necessary.
The instrument has been subjected to prolonged storage under adverse conditions such as exceptional temperatures and/or especially
Service
Every effort has been made to provide you with a reliable product. However, should service become necessary, GE Sensing & Inspection Technologies, has established a number of Factory Trained Service Centers. For the location of the nearest facility, see the back cover of this manual.
xiv
The DM5E is an ultrasonic corrosion thickness measurement device. The instrument is available in three versions:
DM5E Basic
This chapter introduces the key features of the DM5E, and the next chapter explains the keypad functions and display-menu contents. Closely reviewing the material in these two chapters will help you make better use of the more detailed information found in subsequent chapters of this manual.
The instrument is powered by two AA size batteries, providing up to approximately 100 hours of service life. To install the batteries, remove the battery-compartment cover as shown in Figure 1 on page 2.
IMPORTANT: You cannot charge or power the DM5E through its USB port.
Rear of Instrument
Figure 1: DM5E Connections and Batteries (shown with battery cover removed)
DM5E Operating Manual
When the battery indicator is in the last quarter, as indicated by the display-screen symbol , replace the batteries as soon as possible. The DM5E automatically shuts OFF when the batteries are too weak for reliable operation. However, your settings are saved and restored when the instrument is turned ON again. When testing in remote locations, always carry spare batteries.
Importing of data recorder files directly into Microsoft Excel (DM5E DL)
Normal (thickness), MinScan, MaxScan, and Differential/Rate of Reduction Measurement modes are standard
Timed B-Scan
Optional data recorder (DM5E DL) stores up to 50,000 thickness measurements in data recorder file
2 AA batteries
Couplant sample
Certificate of conformity
Note: A mini-USB to PC cable is only included with the DM5EDL version. This cable can be purchased separately for use with the DM5E Basic or DM5E version.
This DM5E Operating Manual is divided into eight chapters and one appendix. All chapters apply to all instrument versions, except for two functions:
Data Recorder, which is described in Chapter 5, applies only to the DM5E DL version
Dual Multi mode applies only to the DM5E with Dual Multi functionality and to the DM5E DL
The DM5E's display, keypad, and functional commands are easy to interpret and use. In this chapter, you'll find a brief explanation of all display and keypad features. You'll also find references to manual sections where more detailed information is available.
Note: The DM5E's display screen contents varies depending on the optional features installed and the operational settings selected. The display screens shown in this chapter are typical of those found in all the instrument configurations.
The instrument's keypad includes a combination of dedicated keys and virtual function keys. The key functions, as illustrated in Figure 2 on page 8, are as follows:
Arrow Keys - Use to adjust the value of the selected parameter and for on-screen navigation.
Function Keys - Use to activate any functional control shown in the display directly above the key.
Cal/On Key - Press once to launch the calibration process or to make on screen selections. Press and hold to power the instrument
ON and OFF.
Mode Key - Press to activate the standard display modes and any optional display modes that are installed.
2.1
Select a probe setup Press to activate the functional control (when present) above the key.
Press to launch the calibration process or make on-screen selections. Press and hold to power the instrument ON or OFF. Press to adjust the value of the selected parameter and for on-screen navigation.
This section describes the layout of the DM5E's primary display modes including:
Measurement Display Mode - Indicates measured thickness, contains status icons, displays data recorder location and file name (when
this option is installed and activated). Figure 3 on page 10 shows four representative display screens. The display's appearance varies based on the installed instrument options as well as instrument display settings. Additional measurement view modes are available when the Dual Multi or Data Recorder options are installed.
Probe Setup Display Mode - Allows the user to select a standard preloaded instrument setup, each of which is matched to a specific
File Display Mode - Allows the user to create and store thickness readings in data recorder files. This feature is only available when the
data recorder option is installed and activated (see Figure 5 on page 12). When installed and activated, the data recorder option allows users to create data recorder files, store thickness readings, and navigate through file contents.
Configuration Display Mode - Instrument controls are accessed through this screen. The controls listed on the configuration display
screen vary depending on the installed options (see Figure 6 on page 13). The contents of the Configuration display depends on the instrument's installed-option configuration.
2.2
10
2.2
11
2.2
12
2.2
13
By repeatedly pressing
, you can access the Configuration display and you can adjust the following DM5E settings:
VIEW MODE - In base-model units this parameter offers five settings. See Section 4.1 on page 36 to select the view to be displayed.
DATA RECORDER - When the optional data recorder is installed, setting this parameter to ON activates the file recording capability and
MINIMUM ALARM - Enables and sets the minimum alarm value (see Section 3.4 on page 30).
MAXIMUM ALARM - Enables and sets the maximum alarm value (see Section 3.4 on page 30).
UNITS - Sets the units of measurement to inches or millimeters (see Section 3.2 on page 21).
NOMINAL THICKNESS - Sets the nominal thickness value that is then used to calculate and the differential and rate of reduction
measurement values when selected by the VIEW MODE parameter (see Section 3.2.3 on page 25).
RESOLUTION - Controls the format (number of decimal places) of the thickness measurement that is displayed and stored in a Data
Recorder file.
UPDATE RATE - Sets the rate at which the measurement reading is updated (see Section 3.2.2 on page 24).
14
2.3
BACKLIGHT - Turns the display backlight feature to ON, OFF, or AUTO (AUTO switches backlighting ON whenever a key is pressed or
LANGUAGE - Sets the instrument display language (see Section 3.2 on page 21).
BSCAN MINIMUM - Sets the minimum value that can be drawn when BSCAN is processing (see Section 3.2.1 on page 23).
BSCAN MAXIMUM - Sets the maximum value that can be drawn when BSCAN is processing (see Section 3.2.1 on page 23).
DUAL MULTI - Used to switch between Dual Multi and IP test modes, when Dual Multi option is allowed (see Section 3.2 on page 21).
READING HOLD - Can be set to hold the last measurement value when the probe is uncoupled in measurement mode (see Section 3.2 on
page 21).
15
2.3
TIME - Used to set the current time (see Section 3.2 on page 21).
DATE - Used to set the current date (see Section 3.2 on page 21).
GAIN - Used to select a different level of gain (LOW/HIGH/AUTO/DEFAULT) (see Section 3.2.1 on page 23).
RADIX - Selects a period (.) or comma (,) to be used as a decimal point (see Section 3.2 on page 21).
AUTO POWER DOWN - When set to ON, this control conserves battery life by powering the instrument OFF if no key presses or
is pressed and
measurements occur for the set number of minutes. When set to OFF, the instrument is only powered OFF when held.
16
Prior to measuring thickness, the instrument must be properly setup. This chapter explains the steps that must be taken prior to measurement, to ensure that the display appearance, ultrasonic settings, and measurement mode match the desired configuration. The steps described in this chapter apply to all instruments, regardless of whether or not they incorporate the Dual Multi or Data Recorder options.
Most fundamental instrument parameters are automatically set to a default value based on the Setup file selected by the user. Each instrument is supplied with standard Setup files, and naming these to match the probe is the first step in setting up the instrument.
Install a probe and configure the instrument by loading a Setup file (see Section 3.1 on page 18).
Set up the instrument display appearance (see Section 3.2 on page 21).
Set the instrument's Gain and Update Rate (see Section 3.2 on page 21).
Set maximum and minimum alarm values (see Section 3.4 on page 30).
Lock and unlock the instrument controls (see Section 3.6 on page 32).
17
Prior to measuring thickness, you must connect a probe to the instrument (see Figure 7 below) and select a Setup file that is compatible with the probe.
18
3.1
After a probe is connected, press to activate the Probe Setup display. The Probe Setup display, which is shown in Figure 8 below, allows the user to select one of the standard Setup files.
3.1
The following instrument settings are automatically adjusted when the Setup file is activated:
View Mode
Gain
Measurement Mode
20
Before using the DM5E, the parameters listed on the next page must be programmed at the Configuration display (see Figure 9) as follows:
1.
2. to
Press
or
3.
4.
Press
or
5. to
Repeat steps 2-4 until all remaining parameters have been set.
After all of the settings have been configured, press return to the measurement display.
21
3.2
Configuring the Instrument (cont.) BSCAN MINIMUM - BSCAN display of minimum value BSCAN MAXIMUM - BSCAN display of maximum value Auto Power Down - Select ON and then choose the desired time
Using the procedure described on the previous page, configure the following parameters:
4 or 8 Hz
to automatically turn the instrument OFF 5/10/15/30 minutes after the last key press (no data will be lost) or select OFF to allow the instrument to remain ON until manually powered OFF.
point
Contrast - Adjusts display contrast Backlight - Adjusts the display backlighting. A setting of ON
increases battery usage, a setting of AUTO causes the backlighting to illuminate and remain on for five seconds after each key press.
22
The instrument gain level may set to AUTO, DEFAULT, HIGH, or to a noise-reducing LOW setting. To select the instrument gain setting:
1.
Press
3.
Press
4.
Press
23
Measurements are updated at a rate of 4 or 8 Hz, as selected by the user. To select the measurement update rate:
1.
Press
3.
Press
4.
Press
24
When the VIEW parameter is set to DIFF/RR%, the displayed thickness includes two differential values (in the instruments units and as a percentage) that represent the variation from user specified nominal thickness (see Section 4.4). To manually input or modify the nominal thickness:
1.
Press
3.
Press
4.
Press
25
Before using the DM5E, the instrument and the connected probe must be calibrated using one or more standards of known thickness. It is important that the Setup file corresponding to the probe be selected prior to launching the calibration process. Then, the user can select either 1pt, 2pt, or AUTO calibration mode in the LOCKOUT menu. To complete the calibration, follow the instructions on your instrument's display screen and in Figure 10 on page 27 (1pt mode), Figure 11 on page 28 (2pt mode) or Figure 12 on page 29 (Auto mode).
Note: Proper calibration is essential for accurate operation of the instrument. The instrument should be calibrated if there is a change in the type or temperature of the material being tested, if the attached probe is changed, after performing any parameter adjustment, or at intervals specified in your test procedures.
26
3.3
3.3
28
3.3
When the thickness measurement value exceeds either the minimum or maximum alarm setting, the measurement value alternates between solid and hollow font type displays, except in B-SCAN mode. When an alarm appears in B-SCAN mode, the measurement value flashes ON and OFF in 1/2 second intervals.
1.
Press
2. to activate it. or
Press
or
3.
4.
Press
or
position. To switch the alarm setting to OFF from any other value, simultaneously press
5.
Press
after the alarm thickness has been set to the desired value.
30
The default BSCAN display range is 0-20 inch. We highly recommend adjusting the range according to the target, because setting an appropriate range will display a clearer scan graph.To adjust the range, complete the steps below.
Note: If the BSCAN value exceeds the preset range, that point will be graphed as equal to the range limit.
1.
Press
2. to activate it. or
Press
or
3.
4.
Press
or
position. To switch the alarm setting to OFF from any other value, simultaneously press
5.
Press
after the alarm thickness has been set to the desired value.
31
The DM5E allows operators to lockout specific functions by using the lockout display, which is only accessible from the measurement screen. Follow the instructions in Figure 13 on page 33 to access this display and lockout (or enable) some or all of the function keys.
Note: The DATA RECORDER and DUAL MULTI functions are optional and are only available with advanced instrument versions. However, an upgrade of the basic version is possible by purchasing a license key.
In the Lockout menu, the operator can set the probe zeroing and calibration mode (CAL/ZERO MODE) by selecting 1pt, 2pt or AUTO. One of the following calibration reminder styles can also be selected:
REMIND TIME - The calibration reminder is displayed after a preset elapsed time from the last thickness calibration. The setting
CAL READINGS - The calibration reminder is displayed after the user presses the SEND function the number of times indicated by the
setting value.
CAL POWER ON - The calibration reminder is displayed each time that the instrument is powered ON when the value is set to ON.
32
3.6
34
After setting up the instrument as described in Chapter 3, additional adjustments can be made by selecting the type of measurement made and the way in which it is displayed. You will also find information related to configuring and working with the BSCAN mode. The specific topics include how to:
Select the measured value to view (see Section 4.1 on page 36)
Work with Normal measurement mode (see Section 4.2 on page 39)
Use the Min Scan and Max Scan measurement modes (see Section 4.3 on page 40)
Interpret the Differential/Rate-of-Reduction measurement mode (see Section 4.4 on page 41)
Work with the B-Scan measurement mode (see Section 4.5 on page 42)
Use the Dual Multi measurement mode (see Section 4.6 on page 43)
Note: Load a Setup file corresponding to the connected probe prior to making the adjustments described in this chapter. Loading a Setup file after these adjustments are made will force certain settings back to the default value stored in the file.
35
Base model DM5E instruments offer four display modes. In addition, two more display modes are available depending on the installed options. These display modes include:
NORMAL (Thickness Only) - The thickness value appears as large digits in the center of the display.
MIN SCAN (Minimum Thickness Scan) - Allows the user to continuously evaluate material thickness, provided the probe remains
coupled or is uncoupled only briefly. After the evaluation period is complete, the minimum material thickness is displayed. During the evaluation period, the thickness is displayed along with the minimum and maximum observed thickness values and the calculated difference between the two. The display also includes a time-out bar, which begins to fill as soon as the probe is uncoupled. Recoupling the probe before the time-out period expires allows the user to continue with the same evaluation period. After the evaluation period is complete, as indicated by the filling of the time-out bar, the minimum thickness observed during the evaluation period is displayed.
MAX SCAN (Maximum Thickness Scan) - Allows the user to continuously evaluate material thickness, provided the probe remains
coupled or is uncoupled only briefly. After the evaluation period is complete, the maximum material thickness is displayed. During the evaluation period, thickness is displayed along with the minimum and maximum observed thickness values and the calculated difference between the two. The display also includes a time-out bar, which begins to fill as soon as the probe is uncoupled. Recoupling the probe before the time-out period expires allows the user to continue with the same evaluation period. After the evaluation period is complete, as indicated by the filling of the time-out bar, the maximum thickness observed during the evaluation period is displayed.
36
4.1
Note: During the placement of a very sensitive probe on the part surface, excess couplant may cause erroneous results. To eliminate the under HOME. This restarts the scanning session
effect of excessive couplant, couple the probe firmly to the part, then press while the probe is properly in contact with the part.
DIFF/RR% (Differential/Rate-of-Reduction) - The currently measured and user-specified nominal thicknesses appear along the top of
the display, while the difference between these two values (measured minus nominal values, expressed in percentage and dimensional terms) appears in the middle of the display.
B-SCAN (B Scan) - Allows the user to continuously evaluate material thickness by graphing 1 point per second. During the evaluation
period, the measured thickness value is displayed along with the minimum and maximum observed thickness values. The display has a hidden 5-second time-out bar, which starts as soon as the probe is uncoupled. Recoupling the probe before the time-out period expires allows the user to continue with the same evaluation period. Pressing the SEND key stores the minimum thickness value after the evaluation period is complete, as indicated by the filling of the time-out bar.
Note: The displayed differences can have either positive or negative values (see Section 4.4 on page 41).
37
4.1
1.
Press
2.
Press
or
3.
Press
4.
Press
or
38
When NORMAL view mode is active, the display shows only a thickness reading, as shown in Figure 14 below.
Note: If the Data Recorder file is active, the data location and stored value will be displayed at the top left corner of the display, instead of the material velocity value.
The MIN SCAN and MAX SCAN modes allow the user to continuously evaluate material thickness then, after the evaluation period is complete, display the extreme (minimum or maximum) material thickness observed. During the evaluation period, thickness is displayed along with the minimum and maximum observed thickness values (see Figure 15 below). A time-out feature allows a momentarily uncoupling of the probe during the evaluation period. In these modes, the instrument will continue to collect and compare readings for as long as the time-out period is not exceeded.
40
This view mode (see Figure 16 below) displays both the currently measured thickness and a nominal thickness input by the user, along with the difference between these two values (measured minus nominal) expressed in percentage and dimensional terms.
Note: The displayed differences can have either positive or negative values.
The B-SCAN mode (see Figure 17 below) allows the user to continuously display the material thickness in a graphical mode. During the evaluation period, the thickness measurement is displayed along with the minimum and maximum thickness values. A time-out feature allows a momentarily uncoupling of the probe during the evaluation period. In this mode, the instrument will continue to collect and compare readings for as long as the time-out period is not exceeded. Pressing the SEND key will store the minimum value in the active file.
42
Protective coatings, including paint, contribute significant error to the thickness measurement of the underlying metal when using conventional measurement techniques. The Dual-Multi mode eliminates coating layers from the measurement by measuring between successive back wall reflections of the metal. The measurement range for each probe will vary depending upon the type and thickness of the coating, the thickness of the metal, and the quality of the bond between the coating and the metal.
To activate or deactivate the Dual-Multi mode, you can either use the Configuration menu or press and hold
measurement mode. A single-peak icon ( )above the battery icon indicates IP mode, while a dual-peak icon ( icon indicates Dual Multi mode (see Figure 18 on page 44).
Note: The Basic version of the DM5E does not have the Dual Multi function. Only specially-equipped DM5E and DM5E DL versions offer the Dual Multi function.
43
4.6
44
Note: The instructions in this chapter apply only to instruments equipped with a Data Recorder.
When installed and activated, the Data Recorder option allows users to create Data Recorder files, store thickness readings (and the accompanying B-SCAN, when desired), and navigate through file contents. In this chapter you will learn how to:
Recall and erase Data Recorder files (see Section 5.2 on page 49)
Store thickness readings in Data Recorder files (see Section 5.3.1 on page 51)
Navigate Data Recorder file locations to store or replace readings (see Section 5.3.2 on page 51)
45
To create a new Data Recorder file, you need only specify a file name. In most cases, however, you will wish to enter additional information. Parameters that can be adjusted during the file-naming process include:
The numbers of ROWS and COLUMNS in the Data Recorder file (see Figure 20 on page 48).
Note: Once created, the file name, size and ADVANCE DIRECTION parameters cannot be edited. Only the NOTES contents can be modified. To accomplish this, first select the file then activate and modify the parameter.
Follow Steps 1 through 5 in Figure 19 on page 47 to create a new Data Recorder file.
46
5.1
5.1
Figure 20: Defining the Size and Advance Direction of a Data Recorder File
DM5E Operating Manual
48
Stored Data Recorder files can be recalled or erased at any time. Once recalled, thickness measurements can be stored in available file locations, but the advance direction cannot be changed.
1.
Press
as many times as necessary to access the Configuration display and then set DATA RECORDER to ON.
2.
Press
again to launch the File Display mode. below FILES. A file list will appear on the display. to confirm the selection.
3.
4.
Press
or
49
1.
Press
2. below ERASE.
If necessary, press
3.
Press
or
to select the stored file you wish to erase, then press and hold
IMPORTANT: Once erased, the Data Recorder file and any stored thickness measurements cannot be retrieved.
The Data Recorder must be enabled (turned ON) via the CONFIGURATION display. Once the Data Recorder is turned on, the contents of the active Data Recorder file are displayed in the File Navigation Grid shown in Figure 20 on page 48.
To store a thickness reading in the selected file position, press below SEND. Remember that the units of the thickness value and the number of decimal places to which the thickness is recorded are determined by the UNITS and RESOLUTION settings as described in Section 3.2 on page 21.
50
When in B-Scan mode, the minimum value is stored in the active position of the Data Recorder by pressing
When the Data Recorder is turned ON, the contents of the active Data Recorder file are displayed in the File Navigation Grid shown in
Figure 20 on page 48. After the navigation grid is activated, , , and can be used to select any file position. A heavier grid box appears around the selected file position. Once a file position has been selected:
51
52
Note: The instructions in this chapter apply to all versions of the DM5E instrument.
The DM5E can communicate with external devices through its built-in I/O port. A mini-USB to PC cable is available for connection to this virtual COM port (see Figure 21 below for the location). With the UltraMATE, UltraMATE Lite, or even a simple Excel Macro, a PC can import data recorder files from the instrument.
The instruments I/O port can be connected to a PC using a mini-USB cable. Be sure to install the virtual COM port driver on the PC before making the connection. Then, follow the instructions on your PC screen to install the new hardware.
54
Chapter 7. Specifications
Chapter 7. Specifications
This chapter lists the features and specifications of your DM5E including:
Features of the optional Data Recorder (see Section 7.2 on page 60)
55
Chapter 7. Specifications
Operating Principle:
Measuring Range:
Depends on probe and material, 0.60mm to 508mm (0.025in to 20.00in) in IP to 1st BW Measurement Mode, 2.00mm to 127.0mm (0.079in to 5.00in) in Dual-Multi Measurement Mode, the coating thickness range shall be 0.3mm to 2.50mm (0in to 0.098in).
Measuring Resolution:
0.001 inch default (selectable 0.001, 0.01 inch) 0.01 mm default (selectable 0.01, 0.1 mm)
Units:
inch or millimeter
Calibration:
56
Chapter 7. Specifications
7.1
Pulser:
Excitation Pulse: Spike Pulser Voltage: 120 V into 50 ohm load, using 20 MHz oscilloscope
Receiver:
Display Type:
High-resolution graphical LCD, 64 x 128 Pixels, 53.0mm x 27.0mm with backlight and adjustable contrast
Update Rate:
NORMAL view mode: 5 digit, 0.4" (10.6 mm) high B-SCAN view mode: 5 digit, 0.1" (2.55 mm) high
Setups:
57
Chapter 7. Specifications
7.1
Alarm Settings:
Minimum and Maximum alarms, Range of 0.010 in to 20.00 in, 0. (0.25 mm to 508 mm) Reading alternates between solid and hollow when alarms are enabled and violated
Power Requirements:
2 AA size batteries
Instrument Shut-0ff:
Language:
Selectable English, German, French, Spanish, Italian, Russian, Japanese and Chinese
I/O Connectors:
Temperature:
Operating: +10F to +120F (-10C to +50C) Storage: -10F to +140F (-20C to +60C)
Weight:
58
Chapter 7. Specifications
7.1
Size:
Shock:
IEC 68-2-27 Ea, as per Mil Std 810C Method 516.2 Procedure I with a 15g 11ms impulse half sinusoidal wave applied 6 times per axis
Sealing:
IEC529 / IP54, Dust proof/dripping water proof as per IEC 529 specifications for IP54 classification
59
Chapter 7. Specifications
Capacity:
50,000 readings
File Structures:
Grid file
Number of Rows:
1 to 50,000
Number of Columns:
1 to 223
File Naming:
Optional Software:
60
Chapter 7. Specifications
Model DA501
Frequency:
5 MHz
Probe Style:
standard
-10 to 70C
Contact Diameter:
15 mm (.590)
3.0 to 25 mm
N/A
61
Chapter 7. Specifications
62
Chapter 8. Maintenance
Chapter 8. Maintenance
Clean the instrument housing and display using a soft cloth lightly dampened with water or a mild window-cleaning product. Never use harsh solvents as they could cause the plastic to become brittle or damaged.
Grasp the cable only by the connectors when connecting or disconnecting it from the instrument.
8.3 Batteries
Periodically inspect the batteries for signs of leakage or corrosion. If either occurs, remove and replace ALL the batteries. Take care to properly dispose of the defective batteries (see Battery Disposal on page 75).
63
Chapter 8. Maintenance
64
Resetting the instruments operating software (see Section A.1 on page 66)
Upgrading the instrument's operating software (see Section A.2 on page 67)
Manufacturing and Service Center locations (see Section A.5 on page 71)
65
IMPORTANT: The following procedure will set all parameters to their factory default values. Any files you wish to retain should be downloaded using the UltraMATE PC software before resetting the operating software. However, any Data Recorder files stored in the instrument will not be deleted during this procedure.
1.
Turn the instrument OFF. for approximate 3 seconds, until the display indicates that the reset is complete.
2.
3.
A successful reset is acknowledged when the RESET COMPLETE message briefly appears at the bottom center of the display screen.
66
The DM5E operating software can be upgraded using an internet connection and an upgrade utility program. Each instrument is shipped with a CD containing the DM5E Upgrade Utility Program. To perform the update, proceed as follows:
1.
Install the Upgrade Utility Program on a PC that has an active internet connection.
2.
Run the Upgrade Utility Program from the Program Menu on your PC. For a typical MS Windows operating system, this is accomplished by clicking the START button, clicking the RUN button, selecting PROGRAMS and then selecting the DM5E Upgrade Utility (see Figure 22 on page 68).
3.
Select the Browse local disk button if the new firmware is stored on a local PC drive, or select the Download new file button to connect to the GE Sensing & Inspection Technologies FTP site (you must be connected to the internet prior to this operation). The program checks the firmware version you have on your machine against the version that is on the FTP site. If the firmware version on the FTP site is newer, the utility will download the updated DM5E operation code.
Note: If the firmware versions are the same, the utility will not download the code and will exit the download routine.
4.
5.
67
68
6.
When prompted, press and hold the and keys for 3 seconds. The DM5E will enter re-programming mode. When the update has been completed, the DM5E will automatically shut OFF.
IMPORTANT: All stored custom settings will be lost during the next step. However, stored Data Recorder files will not be affected.
7.
The following information is available on the GE Sensing & Inspection Technologies web site at www.geinspectiontechnologies.com:
Specifications
New features
Literature
69
Instrument Description: The DM5E is an ultrasonic thickness gauge capable of measuring materials using ultrasonic transducers operating at frequencies from 5 to 15 MHz. This instrument displays a thickness calculated by the measured time of flight in the material at the velocity programmed into the instrument. The instrument is intended for use as a portable thickness gauge in an industrial environment. The instrument is not intended to be part of a closed control loop in an automated process control system.
Compliance Statement:
EN55011: 2007
EN61000-6-2: 2005
70
The DM5E is manufactured according to state-of-the-art methods using high-quality components. In-process testing and a quality management system certified to ISO 9001 ensure optimum quality of instrument conformance. The DM5E is manufactured by:
GE Sensing & Inspection Technologies China Q Section, 4th Floor, #185, Tai-Gu Road, Waigaoqiao Free Trade Zone, Shanghai, China
Should you experience a problem or require technical assistance, visit the product service page accessible at www.geinspectiontechnologies.com to locate your local GE Inspection Technologies representative, or contact one of the Customer Support Centers listed on the back cover of this manual.
71
72
73
GE Sensing & Inspection Technologies is an active participant in Europes Waste Electrical and Electronic Equipment (WEEE) take-back initiative, directive 2002/96/EC.
The equipment that you bought has required the extraction and use of natural resources for its production. It may contain hazardous substances that could impact health and the environment.
In order to avoid the dissemination of those substances in our environment and to diminish the pressure on the natural resources, we encourage you to use the appropriate take-back systems. Those systems will reuse or recycle most of the materials of your end life equipment in a sound way.
The crossed-out wheeled bin symbol invites you to use those systems.
If you need more information on the collection, reuse and recycling systems, please contact your local or regional waste administration.
Visit www.ge.com/inspectiontechnologies for take-back instructions and more information about this initiative.
74
This product contains a battery that cannot be disposed of as unsorted municipal waste in the European Union. See the product documentation for specific battery information. The battery is marked with this symbol, which may include lettering to indicate cadmium (Cd), lead (Pb), or mercury (Hg). For proper recycling return the battery to your supplier or to a designated collection point.
Batteries and accumulators must be marked (either on the battery or accumulator or on its packaging, depending on size) with the separate collection symbol. In addition, the marking must include the chemical symbols of specific levels of toxic metals as follows:
75
Your participation is an important part of the effort to minimize the impact of batteries and accumulators on the environment and on human health. For proper recycling you can return this product or the batteries or accumulators it contains to your supplier or to a designated collection point.
Some batteries or accumulators contain toxic metals that pose serious risks to human health and to the environment. When required, the product marking includes chemical symbols that indicate the presence toxic metals: Pb for lead, Hg for mercury, and Cd for cadmium.
Cadmium poisoning can result in cancer of the lungs and prostate gland. Chronic effects include kidney damage, pulmonary
emphysema, and bone diseases such as osteomalcia and osteoporosis. Cadmium may also cause anemia, discoloration of the teeth, and loss of smell (anosmia).
Lead is poisonous in all forms. It accumulates in the body, so each exposure is significant. Ingestion and inhalation of lead can cause
severe damage to human health. Risks include brain damage, convulsions, malnutrition, and sterility.
Mercury creates hazardous vapors at room temperature. Exposure to high concentrations of mercury vapor can cause a variety of severe
symptoms. Risks include chronic inflammation of mouth and gums, personality change, nervousness, fever, and rashes.
76
The technical specifications for the DM5E according to EN15317 are listed in Table 1 below.
Description General Characteristics 138mm x 32mm x 75mm 223 Grams with batteries
Size
mm
Display Type
Pixel
Display Size
mm
Hz
Power Supply
2 "AA" Alkaline Batteries 2 "00" LEMO Dual Multi Method 2 45 2.15 50 2.1 2.2 60 3 Battery level shown in a 9 step icon
Probe Connection
Volts
Hours
Volts
77
Operational Current Range -10 -10 DataLogger 50,000 readings DataLogger connected via USB Transmitter (Spike) 3.2 6.4 19.2 30 400 -105.6 Receiver Automatic 1 12 -96 -86.4 500 600 24 28.8 8 9.6 High Min Scan and B Scan Modes 4 4.8 Low 50 50
Table 1: EN15317 Technical Specifications (cont.) milliAmps 95 145 Update rate set at 8hz, With USB cable connected to PC
oC
oC
Calibration Methods
Data Storage
Data Output
PRF
Hz
PRF
Hz
PRF
Hz
nSec
nSec
Volts
Gain Control
Frequency Range
MHz
78
Table 1: EN15317 Technical Specifications (cont.) Performance 1 5 0.6 1 125 with DA590 60 with DA512 300 with DA503 200 with DA501
mm
mm
mm
mm
79
80
Index
Backlight, Adjusting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 Base Instrument, Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Batteries Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75 Installing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Level Indicator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63 Operating Life. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58 Size . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58 B-Scan Display Range, Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31 Measurement Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42 Storing Minimum in Data Recorder File . . . . . . . . . . . . . . . 51
Cables Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 I/O Port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53 Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63 Calibration Procedure, 1pt . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27 Procedure, 2 pt . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28 Procedure, Auto . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 When to Perform . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 COM Port see I/O Port Configuration Adjusting Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 Display Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Display Mode Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
81
Index
Connections I/O Device. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18 Probe Cable. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Contrast, Adjusting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 Controls, Locking & Unlocking. . . . . . . . . . . . . . . . . . . . . . . . . . 32 Customer Support Centers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Display Mode Configuration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 File . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Options. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 Disposal Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75 Electronic Waste . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74 DM5E Key Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Keypad. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 Power Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Powering ON/OFF. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Setting Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 Versions Available . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1, 6 Dual Multi Measurement Mode . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Data Recorder Creating a New File . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46 Navigating Through File. . . . . . . . . . . . . . . . . . . . . . . . . . . . 51 Recalling & Erasing Files . . . . . . . . . . . . . . . . . . . . . . . . . . . 49 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60 Storing B-Scan Minimum . . . . . . . . . . . . . . . . . . . . . . . . . . . 51 Storing Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50 Date Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . i Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 Differential/Rate-of-Reduction Measurement Mode . . . . . . . . . . 41
E
EMC Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70 EN15317 Technical Specifications. . . . . . . . . . . . . . . . . . . . . . . . 77
82
Index
L
Language Options. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58 Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 Locking & Unlocking the Controls . . . . . . . . . . . . . . . . . . . . . . . 32
Features, Key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 File Data Recorder, Creating . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46 Data Recorder, Navigating . . . . . . . . . . . . . . . . . . . . . . . . . . 51 Data Recorder, Recalling & Erasing. . . . . . . . . . . . . . . . . . . 49 Display Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 Setup, Loading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18 Firmware, Updating . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67 Function Keys, Locking & Unlocking . . . . . . . . . . . . . . . . . . . . . 33
Gain, Setting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Maintenance Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63 Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63 Instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63 Manufacturing Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71 Measurement Mode B-Scan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42 Differential/Rate-of-Reduction . . . . . . . . . . . . . . . . . . . . . . . 41 Dual Multi . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43 Min Scan & Max Scan. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40 Normal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39 Options. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36 Selecting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
83
Index
R
Resetting the Operating Software . . . . . . . . . . . . . . . . . . . . . . . . . 66 10 50 35 40
Measurements Display Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Storing in a Data Recorder File . . . . . . . . . . . . . . . . . . . . . . Thickness . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Min Scan & Max Scan Measurement Modes . . . . . . . . . . . . . . . Mode see Measurement Mode or Display Mode
Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xiii Setup Display, Probe. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19 Setup File Loading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18 Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 Size, Instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59 Software, Resetting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66 Specifications Data Recorder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60 EN15317 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77 Instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56 Probe/Transducer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
T
Temperature Operating . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58 Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Part Number, Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . i PC, Connecting to . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54 Power for DM5E . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Powering ON/OFF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Probe Connecting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2, 18 Display Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 Setup Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
84
Index
Thickness Measuring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35 Nominal, Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 Time, Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 Training, Operator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . viii Transducer see Probe 22 24 67 67
Waste Disposal Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75 Electronic Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74 Web Site Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69 WEEE Directive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74 Weight, Instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Units, Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Update Rate, Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Updating the Firmware . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Upgrade Utility Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
85
Index
86
North/South America 50 Industrial Park Road Lewistown, PA 17044 U.S.A. Tel: 866 243 2638 (toll-free) 717 242 0327
France 68, Chemin des Ormeaux Limonest 69760 France Tel: +33 47 217 9216
United Kingdom/Ireland 892 Charter Avenue Canley Coventry CV4 8AF England Tel: +44 845 130 3925
Germany Robert Bosch Str. 50354 Hrth Germany Tel: +49 2233 601 0
Spain San Maximo,31, Planta 4A, Nave 6 Madrid 28041 Spain Tel: +34 195 005 990
2009 General Electric Company. All rights reserved. Technical content subject to change without notice.