DX IIFU
DX IIFU
DX IIFU
EC REP
Beckman Coulter Ireland, Inc.
Lismeehan, O’Callaghan’s Mills,
Co. Clare, Ireland
Tel: +(353) (0) 65 683 1100
C96782-AA
December 2022
Printed in U.S.A.
© 2022 Beckman Coulter, Inc. All rights reserved.
Beckman Coulter, Inc. • 250 S. Kraemer Blvd. • Brea, CA 92821 • U.S.A.
https://www.beckmancoulter.com
Publication Notes
This manual, C96782-AA, release date 2022-12, supports:
• UniCel DxI published system software version 5, including full versions 5.1, 5.2, 5.3.0, 5.3.1, 5.5.0,
5.6.0, and 5.7.0.
• UniCel DxI published system software version 7, including full version 7.0.0.
Revision History:
C94099-AA. October 2022 UniCel DxI software versions 5.1, 5.2, 5.3.0, 5.3.1, 5.5.0, 5.6.0, and 5.7.0
C85243-AA, April 2022 UniCel DxI software versions 5.1, 5.2, 5.3.0, 5.3.1, 5.5.0, 5.6.0, and 5.7.0
C73201-AA, March 2021 UniCel DxI software versions 5.1, 5.2, 5.3.0, 5.3.1, 5.5.0, 5.6.0, and 5.7.0
C48059-AA, August 2020 UniCel DxI software versions 5.1, 5.2, 5.3.0, 5.3.1, 5.5.0, and 5.6.0
C38266-AB, March 2019 UniCel DxI software versions 5.1, 5.2, 5.3.0, 5.3.1, and 5.5.0
C38266, January 2019 UniCel DxI software versions 5.1, 5.2, 5.3.0, 5.3.1, and 5.5.0
Beckman Coulter, the stylized logo, and the Beckman Coulter product and service marks mentioned herein are trademarks or registered trademarks
of Beckman Coulter, Inc. in the United States and other countries.
ProClin™ is a trademark of The Dow Chemical Company ("Dow") or an affiliated company of Dow.
Microsoft, Windows, and the Windows logo are either registered trademarks or trademarks of Microsoft Corporation in the United States and/or
other countries.
Monovette is a trademark of Sarstedt A.G. & Co.
Contrad is a trademark of Decon Laboratories, Inc.
Citranox is a trademark of Alconox, Inc.
Quidel is a trademark of Quidel Corporation.
Triage is a trademark of Quidel Corporation.
Bio-Rad is a registered trademark of Bio-Rad Laboratories, Inc.
All other trademarks are the property of their respective owners.
UniCel DxI Instructions for Use Table of Contents
Table of Contents
1 System Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
• Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
• Scope of Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
• Reference Materials. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
• Technical Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
• System Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
• Status Indicator Lights. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
• System Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
• System Status Buttons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
• Help Button . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
• System Command Buttons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
• Main Menu Workflow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
• Precautions and Hazards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-10
• Regulatory Symbols and Statements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-12
• Other Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-13
• System Specifications and Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-13
• Electrical Requirements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-14
• Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-17
• Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-17
3 Supplies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
• System Status Panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
• Changing an Empty or Expired Substrate Bottle . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
• Changing an Empty Bulk Wash Buffer Container . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
• Adding RVs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
• Changing a Full Bulk Liquid Waste Container . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
• Changing a Full Solid Waste Container . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
• Loading a Reagent Pack . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
• Unloading a Reagent Pack. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
6 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
• Maintenance Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
• Daily Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
• Special Weekly Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6
• Test Interval Maintenance: 5,000 Tests. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-7
• Test Interval Maintenance: 10,000 Tests. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-10
• Maintenance Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-15
7 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
• Event Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
• Troubleshooting Events. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
• QC Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
• Test Result Flags and Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-4
1 System Overview
Intended Use
The UniCel DxI Access Immunoassay System is an in vitro diagnostic device used for the quantitative,
semi-quantitative, or qualitative determination of various analyte concentrations found in human body fluids.
Scope of Manual
The UniCel DxI Instructions for Use manual is designed for use after you have become familiar with the
UniCel DxI system. This manual contains short instructions for everyday use and routine maintenance. It also
contains general information about the UniCel DxI system, such as theory of operation, system specifications,
safety labeling, and troubleshooting.
Reference Materials
Additional UniCel DxI system documentation is listed in Appendix B. For more information, contact your
Beckman Coulter representative.
Technical Support
For technical assistance with the UniCel DxI Access Immunoassay System:
• In the U.S.A. or Canada, contact Beckman Coulter Technical Support by phone at 1-800-854-3633, or
online at www.beckmancoulter.com/. Before using online support the first time, you will need to
register online.
• Outside the U.S.A. and Canada, contact your technical support representative.
System Description
The UniCel DxI system consists of two major subsystems: the instrument, which performs all sample
processing functions, and the system console, which provides the human interface. There are several points of
operator interaction:
21 7
1
20 8
2
3
9 10
19
11
4
18 5
12
6
13
14
15
17 16
UniCel DxI 800 System
2055A.eps
19 7
1
18 8
2
3
9 10
4
17 5
11
6
12
13
14
16 15
UUniCel DxI 600 System
2055B.eps
2091A.bmp
1 Red: Not ready. Indicates either that the system has stopped, or that initialization is in progress.
2 Amber, steady: Supplies required. One or more system supply areas are low, or waste containers
are almost full. The system will continue to process samples and schedule new tests.
Amber, blinking: Supplies required. One or more system supplies are out, or an area requires
attention. The system will not schedule new tests, but will complete tests in progress.
3 Green: Running. System is processing tests or performing a maintenance routine.
4 Blue: Ready. No processing operations are in progress, but the system is ready to begin processing.
SPU operations such as aliquoting of samples can take place while in the Ready mode.
System Modes
The UniCel DxI system operates in one of four system modes. The current mode is displayed in the upper left
corner of each screen. When the system is in the Running mode, the estimated completion time for the
scheduled tests is displayed as a text line above the three system command buttons.
Ready The system is ready to begin processing samples. SPU operations such as aliquoting
of samples can take place while in the Ready mode.
Paused No new tests are scheduled, but currently scheduled tests continue processing.
Not Ready The system is not ready to process samples. The system requires initialization, or it is
checking the status of subsystems, initializing motors, or homing movable parts.
NOTE
The icon color reflects the
condition currently displayed on the
screen with the highest concern
level. If there is at least one
instrument condition that meets a
red icon state, that condition takes
precedence and the Supplies
Required icon displays red.
Help Button
Select the Help button to display a topic with information about the screen you are on, a View Screen link to a
picture with descriptions of the screen, and a list of related topics. From the screen topic, you can navigate to
anywhere in the Help system.
The Help button also provides links to procedures for critical instrument conditions displayed on the Supplies
Required screen.
Button Screen
Description
Display
Stop Select to stop the instrument. The system stops processing and cancels any tests in
progress. The system requires initialization before tests can be run again.
Pause Select to pause the instrument. The system stops aliquoting after it finishes the
current aliquot. No new tests are scheduled. Processing continues on samples
already in progress.
Resume Select to resume processing when the system is in the Paused mode.
The following flowchart outlines the workflow for each of the Main Menu function buttons.
F1 In Progress
F1 Views F2 Requested F1 Rack ID
F3 Daily F2 Delete Sample
F4 Exceptions F3 New Request
F4 Test Results
F5 Request QC
F2 Edit Request (Requested and Daily views) F6 Cancel Request
Clear Exceptions (Exceptions view) F7 Clear All Samples
Sample Manager F8 More Options
F1 F1 Patient/QC Requests
F3 New Request F2 Calibration
F1 Rack ID
F3 Maintenance
F3 New Request
F4 Test Results (N/A for
F4 Test Results F1 Filter maintenance requests)
F5 Work Pending F2 Sample Manager F6 Cancel Request
F6 Find F6 Delete
F8 Details F7 Print
F8 Options
F1 Filter
F2 Find
F3 Rerun Test
F4 Sample Details
F5 Send to LIS
Test Results
F6 Copy to Disk F1 Report
F2
F7 Print F2 Completed Samples
F1 Refresh Screen
F8 More Options F2 Configure Screen
F3 Delete Data
F8 Reagent Inventory
Supplies
F3
F1 Reagent Supplies
F1 Bulk Supplies
F2 Unload Reagent Pack
F2 Unload Reagent Packs
F5 Delete Pack
F3 Unload All Reagent Packs
F7 Print
F6 Details
F2 Reagent Supplies F8 Filter
F7 Print
F8 Reagent Inventory
F1 Filter F1 Filter
F2 Review Chart F5 Delete Point
and Data F6 Copy to Disk
F7 Print
Quality Control
F4 F3 Compare Charts F1 Save Comparison F1 Chart 1
F4 Delete Comparison F2 Review Chart and Data F2 Chart 2
F4 Edit Date Range F3 Chart 3
F1 Filter F1 Active
F2 Review Data F1 Select Curve/Limits F2 Previous Active
F2 Switch Active Curve F3 Last Run
F3 Comment
Calibration F7 Print
F5
3003C.wmf
F1 RV Feeder
F1 Initialize System F2 RV Mixer
F2 Prime Fluidics F3 Pick and Places
F3 Subsystem F4 Reagent Storage
Diagnostics F5 Volume Checks
F1 Home All
F1 Mechanics F2 Alignments
F1 Pressure
F2 Wash Buffer
F2 Analog Devices F1 Analog Data F3 Liquid Waste
F2 Sensor Calibration F4 Wash Buffer Flow
F4 Device Diagnostics
F3 Digital Devices F1 Panel Lights
F2 Peristaltic and
F1 Read
Diagnostics
Vacuum Pumps
F2 Load/Read Vessel
F7 F3 Digital Inputs
F3 Save LED Reading
F4 Digital Outputs
F4 Luminometer F4 Save High Voltage
Control
F5 Dark Count Check
F8 Cancel
F1 Ultrasonic Settings
F5 Ultrasonics F2 Ultrasonics Adjustments F1 SPU
F6 Exerciser
F2 LAS
F7 RF Level Sense
3004J.wmf
Interlock Switches
The UniCel DxI system is equipped with interlock switches to protect you from injury. If you open the covers
of the instrument, the interlock switches stop all moving parts. If you open the covers while the system is
processing samples, the system cancels all tests in progress.
Safety Symbols
Certain areas of the UniCel DxI instrument present a risk of personal injury or damage to the instrument if
proper safety procedures are not followed. These areas are marked with one or more safety symbols to identify
the hazard. These symbols are defined in the Access Immunoassay Systems Symbol Glossary, available at
techdocs.BeckmanCoulter.com.
Laser Symbol
The laser symbol indicates areas of the instrument where laser light is used. Do not
stare into the laser beam.
CAUTION
Do not remove the laser warning labels or the protective housings that shield the lasers. The lasers
are accessible if the protective housings are removed. Only a trained Beckman Coulter technical
support representative should service the lasers. Do not stare directly into a laser beam if the
instrument cover is opened or removed.
CAUTION
VORSICHT
11-110008-01
LASER RADIATION
LASERSTRAHLUNG, NICHT IN DEN
DO NOT STARE INTO BEAM LICHTSTRAHL BLICKEN. SICHTBARE
CLASS 2 LASER PRODUCT LASERSTRAHLUNG.
670nm 1.0mW 101~186µs
LASERKLASSE 2
IEC 60825-1 Ed. 2 (2007)
CAUTION - CLASS 2 LASER RADIATION WHEN OPEN-DO NOT STARE INTO THE BEAM.
PRODUCT COMPLIES WITH 21 CFR 1040.10 AND 1040.11 EXCEPT FOR DEVIATIONS
PURSUANT TO LASER NOTICE NO. 50, DATED JUNE 24, 2007
These labels are attached to the housing that shields the reagent
storage bar code reader.
&
These labels are attached to the back of the UniCel DxI instrument.
Safety Statements
The following statements describe general safety concerns and provide information about attention symbols
with no accompanying text.
WARNINGS
• The UniCel DxI instrument has moving parts and uses high voltage in the ultrasonic transducers.
Both present an injury hazard. Do not operate the UniCel DxI instrument with the covers or
doors open.
• Reagents, calibrators and controls used with the system may contain small quantities of sodium
azide preservative. Sodium azide preservative may form explosive compounds in metal drain
lines. Refer to National Institute for Occupational Safety and Health Bulletin: Explosive Azide
Hazards (8/18/76).
• Always plug the UniCel DxI system into a grounded three-conductor outlet. DO NOT bypass the
grounding prong on the plug.
• Do not defeat the safety interlock switches on the covers.
CAUTION
Replace substrate bottles only with the main upper covers closed to avoid spillage into the
instrument.
Notice to User
For a patient/user/third party in the European Union and in countries with identical regulatory regime
(Regulation 2017/746/EU or In vitro Diagnostic Medical Devices): if, during the use of this device or as a result
of its use, a serious incident has occurred, please report it to the manufacturer and/or its authorized
representative and to your national authority.
CAUTIONS
• This equipment has been designed and tested to CISPR 11 Class A. In a domestic environment it
could cause radio interference, in which case you may need to take measures to mitigate the
interference.
• Prior to operation of this device, the electromagnetic environment should be evaluated. Do not
use this device in close proximity to sources of strong electromagnetic radiation (for example,
unshielded intentional RF sources) as these could interfere with proper operation.
• If you suspect interference between the UniCel DxI system and other equipment, you must take
whatever action is required to correct the interference. Beckman Coulter suggests the following
actions:
- Move the equipment so there is a greater distance between the equipment and the
UniCel DxI system.
- Re-orient the equipment with respect to the UniCel DxIsystem.
- Be sure that the equipment is operating from a different power service connector than that of
the UniCel DxI system.
Other Symbols
In addition to safety and regulatory symbols, other symbols are placed on the instrument to identify functional
elements such as network jacks and the air filter. These symbols are defined in the Access Immunoassay
Systems Symbol Glossary, available at techdocs.BeckmanCoulter.com.
Instrument: Covers and Width = 67.5 inches (171 cm) Width = 61.5 inches (156 cm)
Drawers Closed Height = 67 inches (170 cm) Height = 67 inches (170 cm)
Depth = 37.5 inches (97 cm) Depth = 37.5 inches (97 cm)
Instrument: Covers and Width = 67.5 inches (171 cm) Width = 61.5 inches (156 cm)
Drawers Opened Height = 75.2 inches (191 cm) Height = 75.2 inches (191 cm)
Depth = 67.7 inches (172 cm) Depth = 67.7 inches (172 cm)
Instrument Clearance Rear = 11.8 inches (30 cm) Rear = 11.8 inches (30 cm)
Required for Ventilation Top = 8.5 inches (22 cm) Top = 8.5 inches (22 cm)
Instrument Clearance Front = 15.5 inches (39 cm) Front = 15.5 inches (39 cm)
Required for Opening Rear = 14.5 inches (37 cm) Rear = 14.5 inches (37 cm)
Covers and Drawers
Top = 8.5 inches (22 cm) Top = 8.5 inches (22 cm)
Service Clearance Required Front = 38.5 inches (98 cm) Front = 36.5 inches (93 cm)
for Opening SPU
System Console (Cart with Width = 29.8 inches (76 cm) Width = 29.8 inches (76 cm)
External Computer and Height = 67.5 inches (171 cm) Height = 67.5 inches (171 cm)
Peripherals)
Depth = 37.8 inches (96 cm) Depth = 37.8 inches (96 cm)
UniCel DxI 800 Instrument (before supplies 1390 pounds (630 kg)
and samples are added)
UniCel DxI 600 Instrument (before supplies 1065 pounds (483.1 kg)
and samples are added)
External Computer See documentation provided by the manufacturer
Monitor See documentation provided by the manufacturer
Printer See documentation provided by the manufacturer
Electrical Requirements
The UniCel DxI system uses two separate power supplies: one for the instrument, and one for the external
computer. Each power supply must meet specific requirements.
Line Power Supply 200-240 VAC at 1,100 VA, at 50/60 hertz (Hz),
single phase power
Line Dedication Dedicated (UniCel DxI instrument is the only
equipment connected to the electrical line)
Line Power Supply 110-120 VAC at 6 amps, 50/60 Hz, or 220-240 VAC at 3 amps, 50/60 Hz
Surge Suppressors
Beckman Coulter recommends that you do not use a surge suppressor with the UniCel DxIinstrument. The
instrument has built-in protection similar to that provided by a surge suppressor.
NOTE
Consult the documentation provided by the UPS unit manufacturer for proper and safe operation, including
ventilation clearance requirements, of UPS units.
NOTES
• Contact Beckman Coulter Technical Support for a recommended backup power supply.
• Consult the documentation provided by the UPS unit manufacturer for proper and safe operation,
including ventilation clearance requirements, of UPS units.
LIS Interface
The UniCel DxI system can be directed by a laboratory information system (LIS) through the LIS interface.
When connected to an LIS, the UniCel DxI instrument receives test requests from, and sends test results to, the
LIS. The LIS interface consists of two major components:
• The physical, or hardware, interface, which is a port located on the external computer
• The logical, or software, interface, which includes the frame-layer protocols and message formats for
sending and receiving messages
For information about setting up the LIS interface on the UniCel DxI system, see Help topic LIS Setup or the
LIS Vendor Information document.
Integrated Workstations
The UniCel DxI system can be integrated with the UniCel DxC (chemistry) system and the UniCel CTA
(closed tube aliquoter) to form the UniCel DxC Integrated Workstation. There are four versions of the
integrated workstation, depending on which versions of the DxI and DxC systems are used.
For information about integrated workstations, refer to the UniCel DxC Synchron Access Integrated
Workstation Instructions for Use manual.
LAS Interface
The UniCel DxI system can be connected to a laboratory automation system (LAS), When connected to an
LAS, the UniCel DxI instrument receives samples from a track in addition to samples that are loaded on the
sample presentation unit. The track can route samples to multiple instruments in the laboratory.
The LAS interface on the UniCel DxI system consists of a sample pipettor mechanism that extends horizontally
beyond the back of the instrument to aspirate the required volume of sample directly from the sample tube on
the connector unit.
For information about connecting the UniCel DxI system to an LAS, contact your Beckman Coulter
representative.
Installation
The UniCel DxI Access Immunoassay System must be installed by a qualified Beckman Coulter technical
support representative. Do not remove the instrument from the shipping crate until a technical support
representative is present.
Warranty
The UniCel DxI Access Immunoassay System is covered by and subject to the provisions of the warranty
included in your contractual agreement for the system or its reagents.
The customer is responsible for routine preventive maintenance procedures. Repairs arising from the failure to
perform these maintenance procedures at the indicated time intervals will be made at the discretion of
Beckman Coulter, and at the customer’s expense.
CAUTION
If you do not follow these procedures, you may damage the instrument or corrupt the system
database.
Shut down and restart the PC if you are directed to do so by a technical support representative or by the system
documentation.
NOTE
If you shut down and restart the PC, the instrument continues processing samples. The test data is sent
automatically when the PC re-establishes communication with the instrument.
Restarting an instrument after shut down resets the software, and returns all instrument devices to their home
positions.
Use the UI software for standard shut downs. Use the keyboard only when the UI is not available. If you use the
computer keyboard to shut down the PC, you may require additional assistance from Technical Support to clean
up your database.
NOTE
If you are shutting down the PC and the instrument at the same time, shut down the instrument first.
NOTES
• Depending on the version of your PC, you may need the system password to restart the PC. If you do
not know the password, contact the lab supervisor.
• In emergencies you can shut down the PC when the instrument is in the Running or Paused mode.
1. From the Configure menu, select PC Admin F7 to display the PC Admin screen.
2. From the PC Admin screen, select System Admin F8 and then select Shut Down PC F1 to display the
UniCel DxI Shut Down window.
WARNING
Do not select the Shut down the instrument software box.
(Optional) To shut off the power to the PC, press and hold the power switch for at least 10 seconds. Wait at
least 20 seconds before restarting the PC.
NOTES
• Depending on the version of your PC, you may need the system password to restart the PC. If you do
not know the password, contact the lab supervisor.
• Only use the computer keyboard to shut down the UI if you have no other alternative.
• If the keyboard does not respond during the performance of this procedure, shut down the PC by
turning off the power. Press and hold the power switch for at least 10 seconds. Then wait at least 20
seconds before restarting the PC.
2379A.bmp
1. Simultaneously press the [Ctrl], [Alt], and 1. If necessary, press the Windows® key [®] on
[Delete] keys on the computer keyboard. the keyboard to display the Start screen.
2. Select Shut Down. 2. Move the mouse cursor to the lower-right corner
3. Select one of the shut down options, depending of the Start screen.
upon whether you want to immediately restart A narrow vertical panel of icons is displayed
the PC software. along the right side of the screen.
4. Select OK. 3. Select the Settings icon.
2380A.bmp
The Settings panel is displayed.
Depending on the version of your PC, you may need the system password to restart the PC. If you do not
know the password, contact the lab supervisor.
2. Wait until the UniCel DxI Main Menu is displayed. If this procedure fails to restart the PC or the UI
software, contact Technical Support.
3. If the PC was shut down for more than 30 minutes, and the instrument was processing tests, it may take a
few minutes for the instrument to send test results to the PC. Do not use the system until the PC receives all
of the test results.
NOTE
To be sure that all test results are sent, display the Test Results screen and filter the results by completion
time. Watch the Result and Comp. Time columns for the system to stop sending results. If you have any
questions, contact Technical Support.
NOTES
• You need the system password to use this feature. If you do not know the password, contact your lab
supervisor.
• If you are shutting down the PC and the instrument at the same time, shut down the instrument first.
• Shutting down the instrument turns off the refrigeration in the reagent storage area.
2. From the PC Admin screen, select System Admin F8 and then select Shut Down Instrument F2 to display
the Shut Down Instrument Software window.
NOTE
If there is no connection between the UI software and your UniCel DxI instrument, a message is displayed
informing you that the connection was lost. The system cancels the shut down operation. If this occurs,
contact Technical Support before attempting to shut down the instrument with the power switch.
4. When the software has shut down, a message is displayed instructing you to turn off the instrument power
switch. Select OK F1 to exit the message window.
5. The instrument power switch is behind the lower right door as you face the instrument. Open the door and
locate the power switch.
6. Press the lower part of the switch to turn the power off (O position).
The system mode area of the system software screen turns red, but no system mode is displayed.
WARNINGS
• You will come in contact with potentially infectious materials during these procedures. Handle
and dispose of biohazard materials according to proper laboratory procedures. Proper hand,
eye, and facial protection is required.
• If you plan to have the instrument moved, make sure that the new location is properly plumbed.
Reagents, calibrators, and controls used with the system may contain small quantities of sodium
azide preservative. Sodium azide preservative may form explosive compounds in metal drain
lines. Refer to National Institute for Occupational Safety and Health Bulletin: Explosive Azide
Hazards (8/18/76).
NOTE
Shutting down the instrument turns off the refrigeration in the reagent storage chamber and other areas.
2. Remove all racks and sample containers from the onboard and offboard areas of the sample presentation
unit (SPU).
4. Change the bulk liquid waste containers if your system uses them.
6. Remove the on-board substrate bottles and replace with empty substrate bottles. Discard the in-use
substrate bottle in the biohazard waste. Refrigerate the second bottle if it has not been used.
7. Remove and refrigerate the bottles equilibrating in the substrate equilibration area.
8. Shut down the PC and the instrument. This includes shutting off the power switch for the instrument and
the PC.
CAUTION
Do not select any buttons on the PC touchscreen or press any keys while the instrument restarts and
initializes. After you start this procedure, the system pauses for approximately two minutes as it
resets the software. Then the system enters the Not Ready mode and begins the system initialization
process.
NOTE
If the PC is on, the system mode area of the system software screen is red, but no system mode is displayed.
1. Verify that the console PC is shut down. For more information, see the Shutting Down the PC procedure.
3. Locate the instrument power switch behind the lower right door as you face the instrument.
4. Press the top part of the switch to turn the power on ( | position) and then wait 90 seconds.
5. Press and hold the console PC power switch for two seconds to turn power on and start the UI software.
7. Verify that the system is in the Ready mode and no message is displayed in the system mode area. If the
instrument does not initialize successfully, contact Technical Support.
8. Wait for the system to restore the internal temperatures. If the instrument was shut down for a short period
of time, it will take 15-20 minutes for the system to restore the internal temperatures. Do not load samples
on the instrument until all temperature zones are in range.
9. Continue normal operation.
WARNINGS
• Wash buffer contains a preservative which may cause sensitization by skin contact. After contact
with skin, wash immediately with soap and water. Wear suitable gloves.
• You will come in contact with potentially infectious materials during this procedure. Handle and
dispose of biohazard materials according to proper laboratory procedures. Proper hand, eye,
and facial protection is required.
NOTE
Wait for the system to restore the internal temperatures. Because the system was shut down for an extended
period of time, it will take up to one hour for the system to restore the internal temperatures. Do not load
samples on the instrument until all temperature zones are in range.
3. Verify that the wash buffer supply is adequate. If necessary, change the bulk wash buffer containers.
4. Empty the solid waste container if it is full. If your system uses bulk liquid waste containers, empty them if
they are full. Check the RV supply. If necessary, add more RVs.
NOTE
Before you load substrate onto the instrument, the bottles must equilibrate at room temperature for the time
specified in the reagent instructions for use.
6. From the Diagnostics menu, select Prime Fluidics F2 to display the Prime Fluidics window.
7. Select All F3. All of the components are selected for priming.
8. Enter the following number of priming cycles in the Cycles Requested field for each component:
• Substrate In Use - Enter 6 cycles.
9. Select Start Priming F2. When the priming is complete, a message is displayed.
12. Depending upon how long the system has been shut down, perform the following procedures.
10,000 Test
X
Interval Maintenance
Special Clean Routine X X X
Run QC X X X
3 Supplies
System Status Panel
(UniCel DxI 800 System Only; not all systems are equipped)
Some UniCel DxI 800 instruments have a system status panel located directly below the status indicator lights.
The panel has six icons that correspond to specific supply areas. A panel icon is lit if a supply is low,
approaches its expiration date, or a waste container is almost full. The icon blinks if a supply runs out, expires,
or if a waste container is full.
The Console panel icon is lit if one or more of the following system status buttons turns yellow or red:
• Rack Exceptions
• Work Pending
• Supplies Required
• Quality Control
• Event Log
The Console panel icon blinks if the Event Log button turns red.
2033A.bmp
1 RVs icon
2 Substrate icons
3 Console icon
4 Wash Buffer icons
5 Solid Waste icon
6 Liquid Waste icons
Place an unequilibrated
bottle in the substrate
equilibration area
2017D.svg
CAUTION WARNING
Pull out the wash buffer
To avoid contaminating supply drawer Wash buffer contains a
the wash buffer supply, preservative which may
do not touch any tubes cause sensitization by skin
Remove the perforated
that enter the container. panels from a new container contact. After contact with
and gently mix the container
Handle the assembly by skin, wash immediately
the cap only. with soap and water. Wear
Extend the neck of the new suitable gloves.
container and remove the
cap and inner seal
Is there a
UniCel DxI system with disconnect on the No UniCel DxI 600 or 800
cap and tube System
System Status Panel assembly?
Consult the Bulk Supplies
If equipped, check the Wash screen.
Yes Remove the shim from the
Buffer panel icons. empty container and slide it If the Wash Buffer
Blinking: Change all low wash into place on the new
Estimated % Full field is
container
buffer containers. If there is only red, change all low wash
one on-board container, load buffer containers. If there is
two new containers. Press the disconnect button Remove the empty container only one on-board container,
Solid: Change the wash buffer on the empty container to from the drawer and set it load two new containers.
release the tubing beside the new container
container that is not in use. If the Wash Buffer
If you cannot determine the Estimated % Full field is
in-use container, consult the Remove the empty container
Remove the cap and tube yellow, change the wash
assembly from the empty
Bulk Supplies screen. from the drawer and replace
container and insert it into
buffer container that is not in
it with the new container
the new container use.
2018E.svg
Adding RVs
Check the RV status to determine how many bags of RVs to add to the vessel hopper. When the vessel hopper
is empty, the system completes any in progress tests, but does not schedule any new tests until you add more
RVs.
WARNING
To avoid contaminating the RV supply, only add full, previously unopened bags to the hopper
whenever it is possible. If you add a previously opened bag of RVs to the hopper, be sure the bag was
closed to protect the RVs from dust and other contaminants.
CAUTION
Only add UniCel DxI system RVs to the hopper. If you add other RVs or containers, the instrument
will malfunction.
2019D.svg
WARNINGS
• You will come in contact with potentially infectious materials during this procedure. Handle and
dispose of biohazard materials according to proper laboratory procedures. Proper hand, eye,
and facial protection is required.
• Thoroughly flush any decontaminant from the bulk liquid waste container before you store the
container. If you reuse a container without flushing it, any remaining decontaminant may react
with chemicals dispensed into the waste container. The resulting chemical reactions may create
gases harmful to you or the instrument.
• Reagents, calibrators, and controls used with the system may contain small quantities of sodium
azide preservative. Sodium azide preservative may form explosive compounds in metal drain
lines. Refer to National Institute for Occupational Safety and Health Bulletin: Explosive Azide
Hazards (8/18/76).
System Mode: Any Mode
2020D.svg
WARNING
You will come in contact with potentially infectious materials during this procedure. Handle and
dispose of biohazard materials according to proper laboratory procedures. Proper hand, eye, and
facial protection is required.
CAUTIONS
• If you change a full solid waste container while the system is in the Running or Paused mode, be
sure to complete the procedure within five minutes. Solid waste may accumulate and obstruct the
waste chute if it takes longer than five minutes to change the container.
• Do not overfill the solid waste container. Excess solid waste may obstruct the waste chute.
• Do not open the solid waste door while the system is initializing, running a Daily Clean System
routine, or running a Special Clean routine. Opening the solid waste door will cancel the routine
and the system will enter the Not Ready mode.
System Mode: Any Mode
2021D.svg
WARNINGS
• You will come in contact with potentially infectious materials during this procedure. Handle and
dispose of biohazard materials according to proper laboratory procedures. Proper hand, eye,
and facial protection is required.
• Do not load partial packs from another UniCel DxI or Access system onto your instrument.
• If you load a used pack, make sure the number of tests displayed for the pack is less than 50.
Erroneous but believable results may be obtained if a used pack is incorrectly identified as a new
pack containing 50 tests. If a used pack is identified as a new pack, unload it and use this
procedure to load a fresh pack.
CAUTION
To prevent damaging the reagent packs, be sure they are properly seated in the tray positions.
Yes
2022D.wmf
WARNING
You will come in contact with potentially infectious materials during this procedure. Handle and
dispose of biohazard materials according to proper laboratory procedures. Proper hand, eye, and
facial protection is required.
NOTES
• You can unload up to four reagent packs at a time on the Reagent Supplies screen for the
UniCel DxI 800 instrument (up to two packs on the UniCel DxI 600 instrument). You can only unload
packs one at a time on the Reagent Inventory screen.
• If a reagent pack is in use, you cannot unload it from the system (identified by the in-use [padlock] icon
on the Supplies screen).
(Any Screen)
Bulk Supplies
2023C.svg
Racks
There are four sample rack sizes: 13 x 75 mm, 13 x 100 mm, 16 x 75 mm, and 16 x 100 mm.
Sample racks are identified by four different labels: Rack bar code label, Container type label, Rack ID label,
and the Reserve Volume label.
1 4
3
2
3
13
x 1
51
0
4
7
R
6
13x
75
5
2149B.eps
7 Rack ID label
The rack bar code label includes the bar code and rack ID number. The container type label includes an
illustration of the only sample container type you can use with the rack. The Rack ID label distinguishes one
sample rack from another, and also identifies the accepted sample container type for the rack. The Reserve
Volume label indicates what type of reserve volume is designated for the rack.
NOTE
You can find the rack ID ranges set up for each type of sample container from the System Setup screen.
The instrument recognizes the type of sample containers held in the sample rack when it scans the rack bar code
label.
Use this equation whenever you need to calculate the minimum sample volume for a container in a rack:
A B C D E
Minimum sample
Sample assay + System dead + Reserve volume + Sample pipettor + Sample container = volume required
volume volume (if set up) overdraw dead volume
2223B.wmf
A The sample assay volume is the sum of the sample volumes for each requested test. To find
the sample volume for each test, see the corresponding reagent instructions for use.
B The system dead volume is the amount of sample that cannot be drawn from an RV inside
of the instrument. The system dead volume is 60 µL for each RV.
C (If set up) The reserve volume is the amount of sample the system stores for additional
testing.
D The sample pipettor overdraw is 20 µL or 5% of the volume of sample in the RV (A + B +
C), whichever is greater.
E The sample container dead volume is the amount of sample that cannot be aliquoted from
the sample container.
NOTE
If the sum of the sample and reserve volumes (A + C) is greater than 500 µL, the system aliquots the
reserve volume into 1-2 additional reaction vessels (RVs). For each additional RV, include the system dead
volume and sample pipettor overdraw calculation.
Example
Sample volume aliquot in one RV - The PSA assay has a sample volume of 25 µL, and the fPSA assay
has a sample volume of 25 µL. The system is set up to draw 200 µL of reserve volume. Here is how to
calculate the minimum sample volume required for these two assays if you are using a 2 mL cup.
Example
Sample volume aliquot in two RVs - The PSA assay has a sample volume of 25 µL, and the fPSA assay
has a sample volume of 25 µL. The system is set up to remove 500 µL of reserve volume. Here is how to
calculate the minimum sample volume required for these two assays if you are using a 2 mL cup.
.
Sample Containers
The following table lists the sample containers that are accepted on the UniCel DxI instrument, along with their
dead volume requirements.
WARNING
Racks are configured to accept only one type of sample container. The sample containers used must
match the ID configured for the rack. Placing an incorrect container in a rack may damage the
system and compromise the integrity of your test results. You can find the rack ID ranges set up for
each type of container from the System Setup screen.
CAUTION
Remove caps from all sample containers before loading on the UniCel DxI system.
Glass
• Dead volume: 500 µL
• Dead volume: 200 µL • Sample rack: 13x75 mm
• Sample rack: 13x75 mm
12/13
x75
12/13
x75 13x100
16x75
Insert
Insert
a 13x75 tube a 13x100 mm tube
• Dead volume: 350 µL • Dead volume: 400 µL
• Sample rack: 13x75 mm • Sample rack: 13x100 mm
1 mL in 1 mL in
13x75 13x100
Beckman Coulter 0.5 mL cup Beckman Coulter 2 mL cup
• Dead volume: 80 µL • Dead volume: 150 µL
• Sample rack: 13x75 mm or • Sample rack: 13x75 mm or
13x100 mm 13x100 mm
2 mL
5 Sample Manager
Sample processing begins when the UniCel DxI system queries the laboratory information system (LIS) for
sample test requests, downloads test requests from the LIS, or when a test request is entered on the Test
Requests screen.
WARNING
You will come in contact with potentially infectious materials during this procedure. Handle and
dispose of biohazard materials according to proper laboratory procedures. Proper hand, eye, and
facial protection is required.
NOTES
• When downloading test requests, be sure that LIS host query is not enabled.
• When setting up new tests to run on the instrument, enable the tests before downloading test requests
from the LIS. The system rejects an LIS test request if the associated test is not enabled.
(Any Screen)
Work Pending
2095C.wmf
NOTES
• Be sure that LIS host query is enabled.
• When setting up new tests to run on the instrument, enable the tests before querying the LIS for test
requests. The system rejects an LIS test request if the associated test is not enabled.
NOTE
If the reserve volume feature is enabled, be sure to place containers with sufficient volume in racks that are
set up to draw reserve volume.
2. Load the racks. Load any racks that contain STAT samples first.
3. The system scans the bar code labels for each sample container in the rack, queries the LIS for the
associated test requests, and then waits for the LIS to send the requests.
NOTES
• You set the length of time the system waits for a request in the LIS Setup window.
• If the system does not receive a test request within the specified time period, the system does not
remove an aliquot for processing and moves to the next container in the rack. The container with no
assigned test requests is not placed in the work pending list. The container must be reloaded after the
system aliquots the remaining samples in the rack.
4. When the system receives the test requests from the LIS, it updates the Sample Manager screen and scans
the bar code label on the next sample container in the rack.
WARNING
You will come in contact with potentially infectious materials during this procedure. Handle and
dispose of biohazard materials according to proper laboratory procedures. Proper hand, eye, and
facial protection is required.
Sample Manager
NOTE
Manually requested patient test
results are evaluated against
default ranges, even if your
New Request F3
laboratory has established
reference, critical, or LIS ranges.
Patient/QC Requests F1 For more information, see Ranges
Setup in the Help system.
Test Requests
Yes Yes
2096C.svg
WARNING
You will come in contact with potentially infectious materials during this procedure. Handle and
dispose of biohazard materials according to proper laboratory procedures. Proper hand, eye, and
facial protection is required.
NOTE
Before you can enter a calibration request, the calibrator must be set up, the test must be enabled, and
sufficient reagent must be in inventory.
Sample Manager
New Request F3
Calibration F2
Request Calibration
OK F1
Test Requests
2101B.wmf
WARNING
You will come in contact with potentially infectious materials during this procedure. Handle and
dispose of biohazard materials according to proper laboratory procedures. Proper hand, eye, and
facial protection is required.
NOTE
You can run tests on patient samples and QC samples in the same rack, even if different tests are requested
for each sample.
Sample Manager
New Request F3
Patient/QC Requests F1
Test Requests
Request QC F5
Request QC
OK F1
Test Requests
2057B.wmf
6 Maintenance
Maintenance Overview
Routine maintenance for the UniCel DxI system includes the following maintenance procedures:
• Daily maintenance
• Special weekly maintenance
• Maintenance after a 5,000 test interval
• Maintenance after a 10,000 test interval
• The Utility routine
Daily maintenance consists of procedures for preparing the system to process samples each day. When the
10,000 and 5,000 test interval procedures are scheduled on the same day, always end by performing daily
maintenance.
Special weekly maintenance is only for laboratories that run tests for HIV antibody levels. Special weekly
maintenance consists of a procedure for installing clean aspirate probes.
Maintenance after a 5,000 test interval consists of a procedure for replacing the duck bill valve.
Maintenance after a 10,000 test interval consists of cleaning and closely inspecting internal components and
tubing connections. The 10,000 test interval procedures include a step to perform 5,000 test interval
maintenance before the 10,000 interval procedures are complete.
The Utility routine primes the reagent pipettors, aspirate probes, and dispense probes. The system automatically
runs the routine every four hours if the system is not processing samples. You can also run the Utility routine
manually. The Utility routine should always be enabled. If the Utility routine is disabled, you should enable the
routine and prime fluidics before running a clean routine as part of your daily maintenance procedures.
A technical support representative will schedule periodic preventive maintenance procedures on your
UniCel DxI instrument in accordance with the terms of your service agreement, if applicable.
Begin 10,000 test maintenance Begin the remaining 10,000 1 - Remove and inspect the dispense
at the Diagnostics menu test interval maintenance probes
procedures ** 2 - Inspect pump connections at the
wash buffer reservoir
Device Diagnostics F4 3 - Inspect the upper aspirate peristaltic
pump connections
Close the right 4 - Inspect the reagent pipettors and
Device Diagnostics main upper cover clean the wash towers
5 - Inspect for fallen RVs
Install clean*
aspirate probes **
1 - Back up the system
(weekly if testing for HIV
levels or at 10,000 test Perform daily maintenance ** 2 - Record the test count
interval maintenance) 3 - Shake down the solid waste
container
4 - Run a clean routine
Replace the duck bill valve **
Run QC and evaluate any
(5,000 test interval
out-of-range results to verify
maintenance)
system performance **
* Clean the used aspirate probes after ** Record that the procedure is
you complete all other maintenance complete on the Maintenance Log
procedures
2270G.wmf
Daily Maintenance
Beckman Coulter recommends that you perform daily maintenance once every 24 hours. Perform daily
maintenance prior to beginning your daily workload. If your laboratory runs the instrument continuously,
perform maintenance at the end of the longest period of inactivity. Even if the system is not used routinely,
perform daily maintenance each day.
Perform daily maintenance to prepare the instrument to process samples for the day.
• Back up the system
• Record the test count
• Shake down the solid waste container
• Run a clean routine
Perform one of two clean routines each day to clean the sample, reagent, and aspirate probes. Run the Special
Clean routine daily if your laboratory has run the Vitamin B12 assay during the last 24-hour time period.
Otherwise, run the Daily Clean System routine.
NOTE
You can load patient sample racks while a clean routine is in process. Samples are aspirated after the clean
routine rack is aspirated, and are held in the sample wheel until the clean routine is complete.
If the special weekly maintenance procedure is scheduled, perform the daily maintenance procedures after you
complete the weekly procedure. If the 10,000 or 5,000 test interval maintenance procedures are scheduled,
perform the daily maintenance procedures after you complete the test interval procedures.
Required Materials
• A backup tape (if your DxI system uses a tape drive for backup)
• Sample containers that hold at least 4 mL of solution
• Sample rack with the appropriate rack ID for the container used
• UniCel DxI wash buffer
• Contrad 70 cleaning solution
• 1:5 dilution of Citranox cleaning solution - mix 1 part Citranox and 4 parts deionized water
• 70% ethanol solution—mix 7 parts of 95% ethanol and 3 parts wash buffer (alternatively, you can
substitute methanol for the 95% ethanol) (Special Clean routine only)
• Maintenance Log
WARNINGS
• You will come in contact with potentially infectious materials during these procedures. Handle
and dispose of biohazard materials according to proper laboratory procedures. Proper hand,
eye, and facial protection is required.
• Citranox cleaning solution is acidic and may cause eye or skin irritation. Handle according to
proper laboratory procedures. See the manufacturer’s label for details.
• Contrad 70 cleaning solution is alkaline and may cause severe eye irritation or mild skin
irritation. Handle according to proper laboratory procedures. See the manufacturer’s label for
details.
• Ethanol and methanol are extremely flammable. Do not use near heat or flame. Do not ingest.
Avoid contact with eyes, skin, and clothing. Use with adequate ventilation.
CAUTION
Racks are configured to accept only one type of sample container. The sample containers used must
match the ID configured for the rack. Placing an incorrect sample container in a rack may damage
the system. You can find the rack ID ranges set up for each type of sample container from the System
Setup screen.
WARNINGS
• You will come in contact with potentially infectious materials during these procedures. Handle
and dispose of biohazard materials according to proper laboratory procedures. Proper hand,
eye, and facial protection is required.
• Ethanol and methanol are extremely flammable. Do not fill sample tubes on the instrument or in
a rack positioned in the sample presentation unit. Remove sample tubes from the instrument
immediately after completing the Special Clean routine.
System mode: Ready
Maintenance Requests
Record the instrument test
count on the Maintenance
Log
Enter the rack ID and press
[Enter]
Time to Completion
The system begins the clean
Daily Clean System: Approximately 10 minutes
routine *
Special Clean: Approximately 13 minutes
New Request F3
2105F.svg
Laboratories that install clean aspirate probes during special weekly maintenance are not required to install
clean probes during 10,000 test interval maintenance.
After you complete the special weekly maintenance procedure, perform daily maintenance and verify system
performance.
Device Diagnostics
WARNINGS
• You will come in
Exit the Device Diagnostics
contact with potentially screen and initialize
infectious materials the system
2301D.wmf
If the 10,000 and 5,000 test interval maintenance procedures are scheduled for the same day, begin with the
10,000 test interval maintenance procedures. The 10,000 test interval procedures include a step to perform
5,000 test interval maintenance. Perform the daily maintenance procedures after you complete the 10,000 and
5,000 test interval procedures.
Device Diagnostics
WARNINGS
• You will come in
Exit the Device Diagnostics
contact with screen and initialize
potentially the system
infectious materials
during these
procedures. Handle
and dispose of
Open the right main upper Perform daily maintenance *
biohazard materials cover
according to proper
laboratory
procedures. Proper
hand, eye, and facial
protection is Verify system performance *
Replace the duck bill valve *
required.
• Never wear exposed
metal that can come
into contact with
Close the right main * Record that the procedure
electrical
upper cover is complete on the
components of the Maintenance Log
instrument.
2300D.wmf
WARNINGS
• Never wear exposed metal that can come into contact with electrical components of the
instrument.
• You will come in contact with potentially infectious materials during this procedure. Handle and
dispose of biohazard materials according to proper laboratory procedures. Proper hand, eye,
and facial protection is required.
CAUTIONS
• Opening the covers abruptly shuts off power to the upper cabinet, and may cause damage if the
system is not in the proper mode and the user interface is not displaying the correct screen.
Follow this procedure carefully.
• Before you begin this procedure, verify that you have a replacement valve in your CARE kit. If
there is no replacement valve, do not perform this procedure. Order a new set of valves and
replace the duck bill valve at your earliest opportunity.
Required Materials
• An empty sample rack
• Several clean lint-free tissues
• New duck bill valve
WARNINGS
• You will come in contact with potentially infectious materials during this procedure. Handle and
dispose of biohazard materials according to proper laboratory procedures. Proper hand, eye,
and facial protection is required.
• Never wear exposed metal that can come into contact with electrical components of the
instrument.
For 5,000 test maintenance, begin this If continuing from 10,000 test
procedure at the Diagnostics menu maintenance, begin the procedure here
Gently lift the dispense probe Record that the procedure is Perform daily maintenance
Insert a new valve into
plate and slide a sample rack complete on the and verify system
the retaining nut
beneath the plate Maintenance Log performance
Yes
Inspect the duck bill valve
2240K.wmf
NOTE
If your laboratory runs tests for HIV antibody levels, you will install clean aspirate probes each week as
part of special weekly maintenance instead of installing clean probes during 10,000 test interval
maintenance.
The 10,000 test interval maintenance procedures include a step to perform 5,000 test interval maintenance
before the 10,000 interval procedures are complete. Perform the daily maintenance procedures after you
complete the 10,000 and 5,000 test interval procedures.
Required Materials
• Several clean, lint-free tissues
• Three clean aspirate probes
• Proper hand, eye, and facial protection
• Small beakers (2)
• Contrad 70 cleaning solution
• Disposable aspirate probe brush
• Deionized water
• Syringe and syringe fitting assembly
• Alcohol wipe
• Hand-held vacuum device
WARNINGS
• You will come in contact with potentially infectious materials during this procedure. Handle and
dispose of biohazard materials according to proper laboratory procedures. Proper hand, eye,
and facial protection is required.
• Never wear exposed metal that can come into contact with electrical components of the
instrument.
Install clean Open the left Close the left and right
aspirate probes * Perform daily maintenance *
main upper cover main upper covers
(unless your lab installs
clean probes weekly
for HIV testing)
2260G.svg
WARNINGS
• Never wear exposed metal that can come into contact with electrical components of the
instrument.
• You will come in contact with potentially infectious materials during this procedure. Handle and
dispose of biohazard materials according to proper laboratory procedures. Proper hand, eye,
and facial protection is required.
Yes
Are you
Are you performing Is 5,000 test Continue with
only installing No special weekly No No Yes Remove and inspect the
maintenance 10,000 test
clean aspirate maintenance? dispense probes
scheduled? maintenance?
probes?
Yes Yes No
Exit the Device Diagnostics
screen and initialize
the system
For more Confirm that the dispense Remove the sample rack
Close the right
information, see probe tubing is routed from beneath the dispense
main upper cover
page 6-14 correctly probe plate
2266J.wmf
2267C.wmf
WARNINGS
• Never wear exposed metal that can come into contact with electrical components of the
instrument.
• You will come in contact with potentially infectious materials during this procedure. Handle and
dispose of biohazard materials according to proper laboratory procedures. Proper hand, eye,
and facial protection is required.
Device Diagnostics
Is the
Moisten a lint-free tissue with
probe free of No
deionized water and wipe
crystalline
down the probe exterior
deposits?
Yes
2268J.svg
DAILY MAINTENANCE 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31
Back Up the System
Record the Test Count
(Turn form to record
number)
C96782-AA
Shake the Solid Waste
Container
Run the Daily Clean
Routine (D) D D D D D D D D D D D D D D D D D D D D D D D D D D D D D D D
Maintenance Log
S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S
Routine (S)
Tech Initials
5,000 TEST INTERVAL MAINTENANCE On Date On Date On Date On Date On Date On Date
Replace the Duck Bill Valve
Verify System Performance (Run Clean Routine and QC)
Tech Initials
Complete the maintenance procedures in the order listed in
10,000 TEST INTERVAL MAINTENANCE On Date On Date On Date each table. When a maintenance procedure is complete, draw
a check () or circle the code in the corresponding box.
Install Clean Aspirate Probes
When all procedures are complete for a scheduled
Remove and Inspect Dispense Probes maintenance, add your initials. Add the date and test count as
Inspect the Pump Connections at the Wash Buffer Reservoir appropriate.
Inspect the Upper Aspirate Peristaltic Pump Connections If you schedule the 10,000 and 5,000 test interval
Inspect the Reagent Pipettors and Clean the Wash Towers maintenance procedures consecutively, perform the 10,000
test interval procedures first. Perform the daily maintenance
Inspect for Fallen RVs
procedures after you complete the 10,000 and 5,000 test
If your system uses bulk liquid waste containers, inspect the interval procedures.
Liquid Waste Drawer
When you combine daily maintenance with special weekly
Inspect the Wash Buffer Supply Drawer or test interval maintenance, you only need to run a clean
Clean the Air Filter routine once in your final maintenance step.
6: Maintenance
6-15
Verify System Performance (Run Clean Routine and QC)
Tech Initials
This page may be reproduced for laboratory use.
UniCel DxI Instructions for Use 7: Troubleshooting
7 Troubleshooting
Event Log
The Event Log is a list of events the UniCel DxI system generates as it monitors the status of various system
parameters. You can use these events to keep informed of system operations and to assist with troubleshooting.
Troubleshooting Events
You can view technical information about an event in the Details window. The event details can be useful for
troubleshooting. Caution and Warning events may include suggestions for fixing the problem.
Before you contact Technical Support for assistance, either print the event details or write down all of the
information displayed in the Details window.
QC Troubleshooting
QC results can fail for a variety of reasons including past-due maintenance, cold substrate, QC material
stability, reagent stability, calibrator stability, and instrument hardware or software problems.
NOTE
For assays that have more than one reagent pipettor assigned, any troubleshooting step that produces
acceptable QC results should be repeated on all pipettors that previously produced failed QC results.
1. Identify event log errors. Contact Technical Support if you need help troubleshooting event log errors.
3. Verify that the sample container has sufficient volume, and has been loaded in the correct rack and the
correct sample container position. If not, correct the problem and repeat the test.
4. Verify that routine maintenance has been performed. If it has not, perform the required maintenance tasks
and then repeat the QC test.
5. Verify that the substrate was equilibrated to room temperature before being loaded. If it was not
equilibrated, repeat the QC test after the substrate has equilibrated. For the recommended temperature and
equilibration time, see the substrate reagent instructions for use.
6. To eliminate random errors as a cause of the QC failure, pipette a fresh sample from the vial in use and
rerun the test. Use the Rerunning a QC Test procedure to ensure that the test is repeated with the same
reagent pipettor and reagent pack.
NOTE
Statistically, even with an appropriate mean and 2SD range, 1 of 20 QC results will be out of range, and
1 of 333 QC results will be out of the 3SD range.
Quality control material stability 1. Prepare new quality control according to the procedure
provided by the manufacturer.
2. Load the freshly prepared quality control and repeat the
test.
Wrong quality control loaded onto the sample rack 1. Load the correct quality control.
2. Repeat the test.
Wrong quality control lot number selected for the test Repeat the test using the correct lot number.
Wrong mean and/or standard deviation information 1. Review the QC information using Edit Control F2 from
entered when setting up the quality controls the QC Setup screen.
2. If necessary, edit the information.
3. Repeat the test.
WARNING
System Check Solution contains ProClin™ 300 preservative, which may cause sensitization by skin
contact. After contact with skin, wash immediately with soap and water. Wear suitable gloves.
8. Run the System Check routines. If results are out of range, contact Technical Support.
9. Identify any reagent pipettor that is producing failed QC results for all assays that run on that pipettor.
a. If QC failures are coming from one pipettor, clear the assignment of that pipettor in the test setup for
each failing assay.
b. If the pipettor that became unassigned in step a was the only pipettor assigned to a test, assign a
different pipettor to that test.
c. Repeat the QC test by entering a new QC test request.
NOTE
Because the reagent pipettor that was used for the failed QC test is no longer assigned to that test, you
cannot rerun the test by using the Rerunning a QC Test procedure.
d. If QC results from other pipettors are satisfactory, disable the failing pipettor on the Supplies Setup
screen until repairs can be made.
10. Recalibrate the assay to compensate for subtle changes in the instrument or reagent.
11. If none of the preceding steps resolve the problem, contact Technical Support.
Rerunning a QC Test
If a quality control test produces questionable results, use this procedure to rerun the test. The system uses the
same reagent pipettor and reagent pack to eliminate random errors as a cause of the questionable result.
NOTE
Only select a single result. If you select a group of results and then select Rerun Test F3, only the result
indicated with the Arrow button is rerun.
2. Select Rerun Test F3. A message informs you that the test has been reordered, and the test is added to the
work pending list.
NOTE
After the test has been added to the work pending list, the test will be applied automatically to the next
request for that QC.
3. Go to the Sample Manager screen and select New Request F3, then select Patient/QC Requests F1 to
display the Test Requests screen.
6. Select the quality control to rerun. The test is automatically removed from the work pending list, and is
added to the test list for the quality control.
8. Place the QC sample in the rack entered in step 4, and load the rack on the instrument.
NOTE
If the reagent pipettor or the reagent pack are unavailable, the sample status will be changed to Supply Wait
and the test will not rerun until both the reagent pipettor and the reagent pack are available. When the
supply condition has been cleared, the test will rerun automatically.
WARNING
You will come in contact with potentially infectious materials during this procedure. Handle and
dispose of biohazard materials according to proper laboratory procedures. Proper hand, eye, and
facial protection is required.
NOTE
Before performing corrective actions, make sure that you are familiar with the troubleshooting process.
Fatal Flags
Non-Fatal Flags
8 Theory of Operation
Sample Processing
The UniCel DxI Access Immunoassay System is an automated analyzer that performs a wide variety of
immunoassays on body fluid samples. The system starts sample processing when it receives a test request either
from the laboratory information system (LIS), or from the system console.
Once a test request is entered, you place the sample in a rack and load it in the sample presentation unit (SPU)
onload area. The instrument automatically moves the rack to the aliquot station, scans the bar codes on the rack
and sample tube, aliquots enough sample to complete all requested tests and replicates into one or more reaction
vessels (RVs), and stores the RVs in the refrigerated sample storage area. The rack is then moved to the SPU
offload area, where you can unload and reclaim any unused quantity of sample for further testing.
Reserve Volume
The system can be set up to draw an additional volume of sample that can be used for system generated reflex
tests, LIS downloaded reflex tests, reruns, or late LIS requests. The additional volume is called a reserve
volume. When the system is set up to draw reserve volumes, a reserve volume may be drawn from patient and
QC samples in racks that are designated for reserve volume.
Individual sample racks can be designated by rack ID for no reserve volume, or for one of two types of reserve
volume: flexible or standard.
• For samples in flexible reserve volume racks, the system calculates the reserve volume for each sample
based on the tests requested for that sample and the reflex conditions set up for the requested tests. A
flexible reserve volume is drawn only when at least one reflex condition is defined and enabled for at
least one of the tests requested for the sample.
• For samples in standard reserve volume racks, the system always draws a fixed volume, regardless of
other considerations.
NOTE
If the instrument is connected to a laboratory automation system (LAS), you can also set up the system to
draw flexible or standard reserve volumes from all sample containers presented to the instrument by the
LAS. This function is independent of rack designations, and has no effect on the way samples in racks are
handled by the SPU.
You enable the reserve volume feature and set up the standard quantity of sample to draw in the Reserve
Volume Setup window.
The system calculates the dead volume and overdraw and draws enough sample to make the entire reserve
volume available for tests. If the volume required for requested tests plus the reserve volume will not fit in one
RV, all of the reserve volume is dispensed into one or more additional RVs.
The system draws reserve volumes only from sample containers in racks that are designated for reserve
volumes. You designate racks for reserve volumes and select the type of reserve volume to be used for each
rack in the Add/Edit Racks window. Racks should be labeled according to their designation, with either a
Flexible Reserve or a Standard Reserve label, or, if no reserve volume is to be drawn, with a No Reserve label.
Unused reserve volume is discarded when the sample expires, or when sample storage capacity is reached and
more room is needed to process samples, whichever comes first.
Onboard Dilutions*
Assays with the onboard dilution feature provide two tests in the assay protocol file (APF): the parent test,
which analyzes a neat (undiluted) sample, and the onboard dilution test, which uses the same protocols as the
parent test, but performs a dilution step before processing. The dilution factor used for onboard dilution tests is
fixed in the APF.
NOTE
The parent test can be diluted manually, in which case the dilution factor for the sample must be entered in
the Test Requests screen.
See the reagent instructions for use to determine if onboard dilutions are supported for a particular assay.
You can use the reflex test feature, with reserve volume, to automatically request onboard dilution tests of
samples that have analyte concentrations that exceed the calibration limits of the assay.
If you expect the analyte concentration in a sample to be high, you can request onboard dilution tests directly
from the console or with a LIS.
Onboard dilution tests use the same processes and calibration curves as the corresponding parent tests. Onboard
dilution tests have test names beginning with d and they require separate setup, including ranges, units of
measure, default sample type, and decimal places.
NOTE
Another type of onboard dilution is the special dilution. Special dilution tests work similarly to onboard
dilution tests, but only a small number of assays support this feature. Special dilution tests have names
beginning with Dil-, and must be set up on the system by a technical support representative.
The system also verifies that the substrate, reagent pack, and assay calibration for the requested test are not
expired. If a calibration or a time-sensitive supply is expired, the system prompts you to replace the supply or
recalibrate the assay.
* The availability of the UniCel DxI onboard dilution feature in your country depends on the status of submissions to local regulatory
agencies. Please contact you Beckman Coulter representative if you have questions about the availability in your geography.
NOTES
• You can run tests with expired supplies or calibrations by overriding the conditions, but the system
applies appropriate flags to the test results.
• Quality control tests run with expired supplies or calibrations will produce results, but the results will
not be added to the QC database.
The system notifies you when it cannot run a test because of certain critical instrument conditions. The Supplies
Required screen identifies the affected instrument subsystem and the procedure that is required to correct the
condition. Select the Help button in the notification to view related information and instructions.
1 STAT samples
3 QC samples
4 Patient samples
5 Maintenance sets
The priority is assigned when the sample is aliquoted. Sets of calibrator and maintenance samples are prioritized
after all of the samples in the set are aliquoted.
If two or more samples have the same priority, the system gives priority to the first sample aliquoted. If a
higher-priority test cannot start because a system resource (such as a reagent pack) is not available, a
lower-priority test may be processed first.
Reagent Addition
When the system is ready to process a sample, the RV is moved by a robotic module to one of the reagent
carriages, where a dedicated reagent pipettor transfers the required volume of sample into another RV. If
required by the assay protocol file (APF), a dilution is made at this time.
The reagent pipettor transfers specified quantities of reagents into the RV. The robotic module moves the RV to
the incubator, where it remains for a period of time specified in the APF. During the incubation period,
paramagnetic particles from the reagents bind with analyte from the sample.
In two-step assays, the RV returns to one of the reagent carriages for additional reagents, followed by another
incubation period.
Device Calibration
Luminometer Calibration
The luminometer is calibrated in the factory against a luminous standard. Thereafter, the UniCel DxI system
adjusts the luminometer automatically at system-defined intervals. The system uses an on-board reference
standard to maintain luminometer reading consistency. The system calculates a drift correction factor based on
the reference standard reading and applies this factor to RLU output.
Ultrasonics Calibration
The ultrasonic transducers on the reagent pipettors are calibrated in the factory.
Because of potential variations in the length and diameter of substrate tubing, the volume of the tubing must be
measured by the system whenever maintenance is performed on the substrate probe or tubing. The measurement
is performed in conjunction with the procedure for priming the substrate line.
Assay Characteristics
See the reagent instructions for use for each assay for information about assay characteristics including:
• Reagent composition, storage, and handling precautions
• Analytical performance
• Sample handling
• Calibrator characteristics and calibration interval
• Calibrator and control traceability
Quantitative In general, calibrator test results provide a multi-point calibration curve. The
system uses the calibration curve to convert a measured response in RLUs to an
analyte concentration and then expresses the result in numerical units.
Semi-Quantitative Calibrator test results provide a multi-point calibration curve. The system uses the
calibration curve to convert a measured response in RLUs to an analyte
concentration and then expresses the result in numerical units.
These assays may report their quantitative result as a qualitative interpretation,
such as positive, negative, or equivocal.
Qualitative Calibrator test results provide a cutoff value based on a formula defined in the
APF. The system compares a test RLU value to the cutoff value and then classifies
the result as reactive or non-reactive for the analyte.
In the case of a reactive test result, a confirmatory test may be indicated, and
automatically requested as a reflex test. The confirmatory assay produces a result
of confirmed or not confirmed.
Acceptance Criteria
For quantitative and semi-quantitative assays, the system uses the precision profile method to determine if the
calibration meets acceptance criteria. The precision profile method consists of three steps:
1. Fitting calibration data using the math model defined in the APF for that assay.
3. Comparing predicted precision and the limits defined in the APF for that assay.
NOTE
Some assays have further acceptance criteria, as defined in the APF.
2 4
1
2239A.eps
A Temperature-Sensitive Assays
The Access assays listed in the table below are affected by changes in the room temperature of the UniCel DxI
system. The result for a given sample test may shift if the room temperature changes significantly from the
temperature at which the assay was calibrated. Depending on the assay, an increase or decrease in room
temperature causes test results to increase or decrease. The magnitude and direction of the result shift for each
assay is shown in the table.
If the change in room temperature exceeds the restricted calibration temperature range, the assay must be
recalibrated at the new temperature before samples are tested. Recalibrating these assays at the new room
temperature limits the magnitude of the shift on sample results.
For each of the affected assays, Beckman Coulter has established a restricted calibration temperature range in
which the assay should be calibrated and run. Within this range, the change in assay results due to temperature
is expected to be within the allowable performance characteristics of the assay.
Restricted
Allowable% If the temperature
calibration
Part #* Product* Change in increases, the
temperature
result assay result…
range
Restricted
Allowable% If the temperature
* * calibration
Part # Product Change in increases, the
temperature
result assay result…
range
The UniCel DxI system does not monitor the room temperature or alert the operator if room temperature
changes from the original calibration temperature for assays identified in the table. You should ensure that your
laboratory has established a procedure to monitor and review temperature during system operation. Quality
control may not detect temperature related change in assay results and cannot be used as a substitute for
temperature monitoring.
If your laboratory is unable to maintain the required temperature ranges, do not report results out of the
laboratory. Please contact Beckman Coulter for additional suggestions regarding temperature monitoring and
control.
B Ordering Information
Assay-Specific Reagents
The availability of these assays in your country depends on the status of submissions to local regulatory
agencies. In addition, an assay may not be available for use on all systems. Contact your Beckman Coulter
representative if you have questions about the availability of particular assays.
# Tests or
Access Assay # Calibrations
Analyte Part # Description Volume*
Name (@ 500 µL/
Calibration)
Adrenal/ Pituitary
Cortisol Cortisol 33600 Reagent 2 x 50 tests 100
33605 Calibrator 6 x 4.0 mL 8
33606 Diluent (S0) 1 x 4.0 mL
Anemia
Active-B12† / Holotranscobalamin C01772 Reagent 2 x 50 tests 100
Active-B12- (holoTC); Active B12 C01773 Calibrator 6 x 2.0 mL 4
Diluted
C01774 QC 2 levels, 3.0 mL each
EPO/ Erythropoietin A16364 Reagent 2 x 50 tests 100
EPO-Diluted A16365 Calibrator 6 total; S0 @ 10 mL 5
S1–S5 @ 2.5 mL
A79783 Diluent Pack 2 x 90 tests 180
Ferritin / Ferritin 33020 Reagent 2 x 50 tests 100
Diluted Ferritin / 33025 Calibrator 6 x 4.0 mL 8
Ferritin-Diluted
A79784 Diluent Pack 2 x 102 tests 204
Folate / Folic acid; Folate A98032 Reagent 2 x 50 tests 100
Red Blood Cell A98033 Calibrator 6 x 4.0 mL 8
(RBC) Folate /
A99250 Diluent (S0) 1 x 4.0 mL
Folate-Diluted
A14206 Lysing Agent 2 x 100 mL
A79784 Diluent Pack 2 x 174 tests 348
Intrinsic Factor Intrinsic factor 387992 Reagent 2 x 50 tests 100
Ab blocking antibody 387993 Calibrator 1 level; 2 x 4.0 mL 16
387999 QC 2 levels; 3 x 4.0 mL each
sTfR / Soluble transferrin A32493 Reagent 2 x 50 tests 100
sTfR-Diluted receptor A32494 Calibrator 6 total; S0 @ 4 mL 5
S1–S5 @ 2.5 mL
B11056 QC1 1 level; 2 x 2.5 mL each
B11057 QC2 & QC3 2 levels; 2 x 2.5 mL each
A79784 Diluent Pack 2 x 70 tests 140
# Tests or
Access Assay # Calibrations
Analyte Part # Description Volume*
Name (@ 500 µL/
Calibration)
Vitamin B12 / Vitamin B12; 33000 Reagent 2 x 50 tests 100
Vitamin B12- Cobalamin 33005 Calibrator 6 x 4.0 mL 8
Diluted
33006 Diluent (S0) 1 x 4.0 mL
A79783 Diluent Pack 2 x 102 tests 204
Blood Virus
HAV Ab† Hepatitis A virus 34200 Reagent 2 x 50 tests 100
antibody 34205 Calibrator 5 x 2.0 mL 4
34209 QC 2 levels; 3 x 3.5 mL each
HAV IgM† Hepatitis A virus IgM 34210 Reagent 2 x 50 tests 100
antibody 34215 Calibrator 2 x 1.0 mL 2
34219 QC 2 levels; 3 x 2.5 mL each
HBc Ab† Hepatitis B virus core 34240 Reagent 2 x 50 tests 100
antibody 34245 Calibrator 2 x 1.0 mL 2
34249 QC 2 levels; 3 x 2.0 mL each
HBc IgM† Hepatitis B virus core 34250 Reagent 2 x 50 tests 100
IgM antibody 34255 Calibrator 2 x 1.0 mL 2
34259 QC 2 levels; 3 x 2.5 mL each
HBs Ab† Hepatitis B virus A24296 Reagent 2 x 50 tests 100
surface antibody A24297 Calibrator 6 x 2.5 mL 5
A24298 QC 2 levels; 3 x 3.5 mL each
HBs Ag† Hepatitis B virus A24291 Reagent 2 x 50 tests 100
surface antigen A24292 Calibrator 2 x 2.7 mL 5
A24294 QC 2 levels; 3 x 4.0 mL each
HBs Ag Hepatitis B virus A24295 Confirmatory 2 x 50 tests 100 (50 patient
Confirmatory† surface antigen (2 x 25 patient samples) samples)
confirmatory
HCV Ab Anti-hepatitis C virus 34330 Reagent 2 x 50 tests 100
PLUS†,** antibodies 34335 Calibrator 2 x 1.0 mL 2
34339 QC (Pos/Neg) 2 levels; 3 x 2.5 mL each
HCV Ab V3†, ** Anti-hepatitis C virus B33458 Reagent 2 x 50 tests 100
antibodies B33459 Calibrator 2 levels; 1 x 1.0 mL each 4
B33460 QC 2 levels; 2 x 3.5 mL each
HIV Combo†,** HIV p24 antigen and A59428 Reagent 2 x 50 tests 100
antibodies to A59429 Calibrator 2 x 1.7 mL 3
HIV-1/O/2 (Pos/Neg)
A59430 QC1, QC2, and 3 levels; 2 x 4.4 mL each
QC3
B22822 QC4 and QC5 2 levels; 2 x 4.4 mL each
Bone Metabolism
Intact PTH / Parathyroid A16972 Reagent 2 x 50 tests 100
Intact PTH- hormone, intact A16953 Calibrator 6 x 1.0 mL 2
Diluted (routine and A79783 Diluent Pack 2 x 60 tests 120
intraoperative)
# Tests or
Access Assay # Calibrations
Analyte Part # Description Volume*
Name (@ 500 µL/
Calibration)
Ostase / Bone specific alkaline 37300 Reagent 2 x 50 tests 100
Ostase-Diluted phosphatase B83876 Calibrator 6 x 2.5 mL 5
B83877 QC 2 levels; 4.0 mL each
A79783 Diluent Pack 2 x 123 tests 246
Ultrasensitive Growth hormone 33580 Reagent 2 x 50 tests 100
hGH / 33585 Calibrator 6 x 2.0 mL 4
Ultrasensitive
A79783 Diluent Pack 2 x 174 tests 348
hGH-Diluted
25(OH) 25(OH) vitamin D A98856 Reagent 2 x 50 tests 100
Vitamin D Total† A98857 Calibrator 6 x 1.4 mL 2
Cardiovascular
BNP†† / B-type natriuretic 98200 Reagent 2 x 50 tests 100
BNP-Diluted peptide (BNP) 98202 Calibrator 6 x 1.5 mL 3
98201 QC 3 levels; 2 x 2.5 mL each
CK-MB / Creatine kinase, 386371 Reagent 2 x 50 tests 100
CK-MB-Diluted isoenzyme MB 386372 Calibrator 6 x 2.0 mL 4
A79783 Diluent Pack 2 x 174 tests 348
Digoxin / Digoxin 33710 Reagent 2 x 50 tests 100
Digoxin-Diluted 33715 Calibrator 6 x 4.0 mL 8
33716 Diluent (S0) 1 x 4.0 mL
A79783 Diluent Pack 2 x 174 tests 348
hsTnI / High sensitivity B52699 Reagent 2 x 50 tests 100
hsTnI-Diluted troponin I B52700 Calibrator 7 total; S0-S2 @ 1.5 mL 2
(Outside US only) S3-S6 @ 1.0 mL
C26909 Calibrator 7 total; S0-S2 @ 1.5 mL 2
(US only) S3-S6 @ 1.0 mL
A79783 Diluent Pack 2 x 60 tests 120
Myoglobin / Myoglobin 973243 Reagent 2 x 50 tests 100
Myoglobin- 973244 Calibrator 6 x 1.0 mL 2
Diluted
A79783 Diluent pack 2 x 60 tests 120
Diabetes
C-Peptide / C-peptide; C33451 Reagent 2 x 50 tests 100
C-Peptide Connecting peptide C31860 Calibrator 6 x 2.0 mL 4
Onboard
A79783 Diluent pack 2 x 60 tests 120
Pre-Dilution
Ultrasensitive Insulin 33410 Reagent 2 x 50 tests 100
Insulin / 33415 Calibrator 6 x 2.0 mL 4
Ultrasensitive
A79783 Diluent Pack 2 x 60 tests 120
Insulin-Diluted
Infectious Disease
CMV IgG† Cytomegalovirus A40702 Reagent 2 x 50 tests 100
antibody, IgG A40703 Calibrator 6 x 1.0 mL 2
A40704 QC 2 levels; 3 x 2.5 mL each
CMV IgM† Cytomegalovirus A40705 Reagent 2 x 50 tests 100
antibody, IgM A40706 Calibrator 2 x 1.0 mL 2
A40707 QC 2 levels; 3 x 2.5 mL each
# Tests or
Access Assay # Calibrations
Analyte Part # Description Volume*
Name (@ 500 µL/
Calibration)
Rubella IgG Rubella antibody, IgG 34430 Reagent 2 x 50 tests 100
34435 Calibrator 6 x 1.0 mL 2
34439 QC (Pos/Neg) 2 levels; 3 x 2.5 mL each
Rubella IgM† Rubella antibody, A32937 Reagent 2 x 50 tests 100
IgM 34445 Calibrator 4 x 1.0 mL 2
34449 QC (Pos/Neg) 2 levels; 3 x 2.5 mL each
SARS-CoV-2 SARS-CoV-2 IgG C58961 Reagent 2 x 100 tests 200
IgG*** C58963 Calibrator 2 x 2.0 mL 4
C58964 QC (Pos/Neg) 2 levels; 3 x 4.0 mL each
SARS-CoV-2 SARS-CoV-2 IgG C69057 Reagent 2 x 100 tests 200
IgG II*** C69058 Calibrator 6 x 2.0 mL 4
C69059 QC 2 levels; 3 x 4.0 mL each
SARS-CoV-2 SARS-CoV-2 IgM C58957 Reagent 2 x 100 tests 200
IgM*** C58958 Calibrator 2 x 2.0 mL 4
C58959 QC (Pos/Neg) 2 levels; 3 x 4.0 mL each
Toxo IgG Toxoplasma gondii A31588 Reagent 2 x 50 tests 100
antibody, IgG A31589 Calibrator 6 x 1.0 mL 2
A31590 QC (Pos/Neg) 2 levels; 3 x 2.5 mL each
Toxo IgM II Toxoplasma gondii 34470 Reagent 2 x 50 tests 100
antibody, IgM 34475 Calibrator 2 x 1.5 mL 3
34479 QC (Pos/Neg) 2 levels; 3 x 3.5 mL each
Inflammation
IL-6*** / Interleukin-6 A16369 Reagent 2 x 50 tests 100
IL-6-Diluted A16370 Calibrator 6 total; S0 @ 4.0 mL 5
S1–S5 @ 2.5 mL
A16371 QC 3 levels; 2 x 2.5 mL each
A79783 Diluent Pack 2 x 123 tests 246
Reproductive
AFP / Alpha-fetoprotein 33211 300 test kit 6 x 50 tests 300
Diluted AFP / (Cals included) Cals: 7 x 2.5 mL 5
AFP-Diluted
33210 100 test kit 2 x 50 tests 100
(outside US only)
33215 Calibrators 7 x 2.5 mL 5
(for 100 test kit)
33216 Diluent 1 x 14.0 mL
A79784 Diluent Pack 2 x 132 tests 264
AMH Anti-Müllerian B13127 Reagent 2 x 50 tests 100
hormone (AMH) B13128 Calibrator 6 x 2.0 mL 4
B13129 QC 3 levels; 2 x 2 mL each
AMH Advanced† Anti-Müllerian C62997 Reagent 2 x 50 tests 100
hormone (AMH) B13128 Calibrator 6 x 2.0 mL 4
B13129 QC 3 levels; 2 x 2 mL each
# Tests or
Access Assay # Calibrations
Analyte Part # Description Volume*
Name (@ 500 µL/
Calibration)
DHEA-S / Dehydroepiandro- A10826 Reagent 2 x 50 tests 100
DHEA-S-Diluted sterone sulfate A10827 Calibrator 6 x 2.0 mL 4
A79784 Diluent Pack 2 x 60 tests 120
hFSH / Follicle stimulating 33520 Reagent 2 x 50 tests 100
hFSH-Diluted hormone 33525 Calibrator 6 x 4.0 mL 8
A79783 Diluent Pack 2 x 174 tests 348
hLH / Luteinizing hormone 33510 Reagent 2 x 50 tests 100
hLH-Diluted 33515 Calibrator 6 x 4.0 mL 8
A79783 Diluent Pack 2 x 174 tests 348
Inhibin A / Inhibin A A36097 Reagent 2 x 50 tests 100
Inhibin A- A36098 Calibrator 7 x 2.5 mL 5
Diluted
A36100 QC 3 levels; 2 x 2.5 mL each
A79783 Diluent Pack 2 x 123 tests 246
PAPP-A† / Pregnancy- A48571 Reagent 2 x 50 tests 100
PAPP-A-Diluted associated A48572 Calibrator 6 x 1.0 mL 2
plasma protein A
A48573 QC 3 levels; 2 x 2.5 mL each
A79784 Diluent Pack 2 x 174 tests 348
Progesterone Progesterone 33550 Reagent 2 x 50 tests 100
33555 Calibrator 6 total; S0 @ 4.0 mL 5
S1–S5 @ 2.5 mL
33556 Diluent (S0) 1 x 4.0 mL
Prolactin / Prolactin 33530 Reagent 2 x 50 tests 100
Prolactin- 33535 Calibrator 6 total; S0 @ 4.0 mL 5
Diluted S1–S5 @ 2.5 mL
A79783 Diluent Pack 2 x 60 tests 120
Sensitive Estradiol B84493 Reagent 2 x 50 tests 100
Estradiol / B84494 Calibrator 6 total; S0 @ 4.0 mL 4
Sensitive S1-S5 @ 2.0 mL
Estradiol-
A79783 Diluent Pack 2 x 174 tests 348
Diluted
SHBG/ Sex hormone binding A48617 Reagent 2 x 50 tests 100
SHBG-Diluted globulin A48618 Calibrator 6 x 1.0 mL 2
A48619 QC 2 levels; 3 x 2.0 mL each
A79784 Diluent Pack 2 x 60 tests 120
Testosterone Testosterone, total 33560 Reagent 2 x 50 tests 100
33565 Calibrator 6 x 2.5 mL 5
Total hCG (5th hCG; Beta-hCG A85264 Reagent 2 x 50 tests 100
IS) / Diluted B11754 Calibrator 6 x 4.0 mL 8
Total hCG (5th
A79784 Diluent Pack 2 x 135 tests 270
IS) / Total hCG
(5th IS)-Diluted
Unconjugated Unconjugated estriol 33570 Reagent 2 x 50 tests 100
Estriol 33575 Calibrator 7 total; S0 @ 4.0 mL 5
S1-S6 @ 2.5 mL
# Tests or
Access Assay # Calibrations
Analyte Part # Description Volume*
Name (@ 500 µL/
Calibration)
Sepsis
PCT / Procalcitonin C22593† Reagent 2 x 50 tests 100
PCT-Diluted (Outside US only) C22594 †
Calibrator 7 x 2.0 mL 4
Procalcitonin C53987 Reagent 2 x 50 tests 100
(US only) C53986 Calibrator 7 x 2.0 mL 4
Thyroid
Free T3 Triiodothyronine, free A13422 Reagent 2 x 50 tests 100
A13430 Calibrator 6 x 2.5 mL 5
Free T4 Thyroxine, free 33880 Reagent 2 x 50 tests 100
33885 Calibrator 6 x 2.5 mL 5
Thyroglobulin Thyroglobulin 33860 Reagent 2 x 50 tests 100
33865 Calibrator 6 x 2.0 mL 4
33866 Diluent 1 x 14.0 mL
Thyroglobulin Thyroglobulin A32898 Reagent 2 x 50 tests 100
Antibody II antibody A36920 Calibrator 6 total; S0 @ 4.0 mL 5
S1–S5 @ 2.5 mL
Total T3 Triiodothyronine 33830 Reagent 2 x 50 tests 100
33835 Calibrator 6 x 4.0 mL 8
Total T4 Thyroxine 33800 Reagent 2 x 50 tests 100
33805 Calibrator 6 x 4.0 mL 8
Thyroid Uptake Thyroxine-binding 33810 Reagent 2 x 50 tests 100
capacity 33815 Calibrator 1 level; 6 x 1.0 mL 12
TPO Antibody/ Thyroperoxidase A12985 Reagent 2 x 50 tests 100
TPO Antibody- Antibody A18227 Calibrator 6 x 2.0 mL 4
Diluted
A79783 Diluent Pack 2 x 60 tests 120
TSH (3rd IS) / Thyroid-stimulating B63284 Reagent 2 x 100 tests 200
Diluted TSH hormone; Thyrotropin B63285 Calibrator 6 x 2.5 mL 5
(3rd IS)
Tumor Markers
AFP / Alpha-fetoprotein 33211 300 test kit 6 x 50 tests 300
Diluted AFP/ (Cals included) Cals: 7 x 2.5 mL 5
AFP-Diluted 33210 100 test kit 2 x 50 tests 100
(outside US only)
33215 Calibrators 7 x 2.5 mL 5
(for 100 test kit)
33216 Diluent 1 x 14.0 mL
A79784 Diluent Pack 2 x 132 tests 264
BR Monitor / Cancer antigen 15-3 387620 Reagent 2 x 50 tests 100
BR Monitor- 387647 Calibrator 6 x 1.5 mL 3
Diluted
A79783 Diluent Pack 2 x 60 tests 120
CEA Carcinoembryonic 33200 Reagent 2 x 50 tests 100
antigen 33205 Calibrator 6 x 2.5 mL 5
33206 Diluent 1 x 4.0 mL
33209 QC 2 levels; 3 x 2.5 mL each
# Tests or
Access Assay # Calibrations
Analyte Part # Description Volume*
Name (@ 500 µL/
Calibration)
GI Monitor / Cancer antigen 19-9 387687 Reagent 2 x 50 tests 100
GI Monitor- 387688 Calibrator 6 x 2.5 mL 5
Diluted
A79783 Diluent Pack 2 x 60 tests 120
Hybritech PSA / Prostate-specific 37200 Reagent 2 x 50 tests 100
Hybritech PSA- antigen 37205 Calibrator 6 x 2.5 mL 5
Diluted
37206 Diluent 1 x 14.0 mL
37209 QC 3 levels; 1 x 5.0 mL each
A79784 Diluent Pack 2 x 60 tests 120
Hybritech free Free prostate- 37210 Reagent 2 x 50 tests 100
PSA specific antigen 37215 Calibrator 6 total; S0 @ 5.0 mL 5
S1 - S5 @ 2.5 mL
37219 QC 2 levels; 1 x 5.0 mL each
Hybritech [-2]proPSA A49752† Reagent 2 x 50 tests 100
p2PSA A49753† Calibrator 7 x 2.1 mL 4
B03704 Reagent 2 x 50 tests 100
B03705 Calibrator 7 x 2.1 mL 4
A56934 QC 3 levels; 1 x 5.0 mL each
OV Monitor / Cancer antigen 125 386357 Reagent 2 x 50 tests 100
OV Monitor- 386358 Calibrator 6 x 2.5 mL 5
Diluted
A79783 Diluent Pack 2 x 63 tests 126
Research Use Only
IL-6 (RUO) / Interleukin-6 A30945 Reagent 2 x 50 tests 100
IL-6 (RUO)- A30944 Calibrator 6 total; S0 @ 4.0 mL 5
Diluted S1–S5 @ 2.5 mL
A30946 QC 3 levels; 2 x 2.5 mL each
A79783 Diluent Pack 2 x 123 tests 246
PAPP-A (RUO) Pregnancy- A49209 Reagent 2 x 50 tests 100
/PAPP-A (RUO)- associated A49210 Calibrator 6 x 1.0 mL 2
Diluted plasma protein A
A49211 QC 3 levels; 2 x 2.5 mL each
A79784 Diluent Pack 2 x 174 tests 348
* All onboard dilution assays use diluent pack A79783 or A79784. The volume of diluent used for each test is assay
dependent. Consequently, the number of tests available per diluent pack differs by assay.
† Not available in the US.
** Distributed by Beckman Coulter for Bio-Rad for use on Beckman Coulter immunoassay systems.
†† Sale and distribution of part number 98200 have moved from Quidel Corporation to Beckman Coulter. Customers
in some countries may continue to receive product labeled as Quidel Triage BNP during a transition period. The
formulation of both products is the same.
*** In the U.S.A., this assay is for use under the Emergency Use Authorization (EUA) only.
CONDITIONS OF AUTHORIZATION FOR THE LABORATORY
The Access IL-6, Access SARS-CoV-2 IgG, Access SARS-CoV-2 IgG II, and Access SARS-COV-2 IgM assays,
along with the authorized Fact Sheet for Healthcare Providers, the authorized Fact Sheet for Patients and authorized
labeling, are available on the FDA website: https://www.fda.gov/medical-devices/emergency-situations-medical-
devices/emergency-use- authorizations#covid19ivd or at https://www.beckmancoulter.com. Authorized laboratories
using the Access IL-6, Access SARS-CoV-2 IgG, Access SARS-CoV-2 IgG II, or Access SARS-COV-2 IgM assays
must adhere to the Conditions of Authorization indicated in the Letter of Authorization.
System Supplies
System supplies are organized alphabetically in two tables. The first table contains supplies ordered through
Customer Service. The second table contains supplies ordered through Customer Technical Support.