NDT-SA-SATORP-UTT-60-RV00 Date-15 Aug-2023
NDT-SA-SATORP-UTT-60-RV00 Date-15 Aug-2023
NDT-SA-SATORP-UTT-60-RV00 Date-15 Aug-2023
INDEX
SL# CONTENT PAGE NO
1 Scope 2
2 Purpose 2
3 Reference Standard 2
4 Personnel Qualification 2
5 Equipment 2
5.1 Ultrasonic Flaw Detector: A-Scan and D-meter with A-Scan Display 2
5.2 Transducer Sclection 2
5.3 Couplant 2
5.4 Calibration Block 3
6 Surface preparation 3
7 Equipment Calibration 3
7.1 General 3
7.2 Calibration 3
7.3 Straight beam calibration for A-Scan Flaw detector 4
7.4 Straigh beam calibration for Lamination check 4
7.5 Calibration for D-meter WITH a-Scan display 4
7.6 Stright Beam for Lamination checks 5
7.7 Calibration for D-Meters with A-Scan display 6
8 Measurement 6
9 Post Cleaning 7
10 Reporting 7
Attachment-I: Velocity Chart 8
Attachment-2: UTT Report Format 9
Page 1 of 9
NDT Corrosion Control Services Co.
Dammam – Saudi Arabia Date: 15th AUG 2023
1 SCOPE
This procedure shall be applicable to all structural materials pipeline materials, piping
spools and joints, tanks, vessels, and plates made of carbon steel and alloyed steel.
2 PURPOSE
This procedure describes the techniques adopted for measurement of material
thickness and laminations checking by Ultrasonic Method using flaw detectors
equipped with A-Scan display.
3 REFERENCE STANDARD
5 EQUIPMENT
5.1 ULTRASONIC FLAW DETECTOR: (A-SCAN), ULTRASONIC D-METER WITH A-SCAN
The examination shall be conducted with a pulse echo ultrasonic instrument OLYMPUS
EPOCH XT, GE USM 36 or equivalent capable of generating frequencies over the
ranges of at least 1 MHz to 5 MHz or equivalent and the instrument shall be equipped
with a stepped gain control calibrated in units of 2 dB or less. For thru coat thickness
measurements instruments such as Olympus Model 38DL+, MG2XT (having software for
thru coat) with compatible transducers such as D 790 can be used. For more details
user’s manual need to be referred.
Other probes are MSCB4 Dual crystal and D1 401 Dual crystal Transduces are selected
for normal temperatures.
5.3 COUPLANTS:
Motor Oil, grease, cellulose gel shall be used as a couplant as indicated in the
operating manuals of UT Machines. couplants shall be selected with due consideration
towards the temperature of the examination surface. High Temperature couplant shall
Echo 8 HT & Sono 900 and other equivalent.
Page 2 of 9
NDT Corrosion Control Services Co.
Dammam – Saudi Arabia Date: 15th AUG 2023
10 mm
8 mm
6 mm
4 mm
2 mm
The temperature of the calibration block and the surface to be examined shall be
within 14 degrees Celsius of each other.
Step wedges shall have a thickness step near the minimum range to be examined, a
step near the maximum range to be examined and a step in between the minimum
and maximum.
6 SURFACE PREPARARION
6.1 The base metal on each side of the weld shall be free of weld spatter, surface irregularities, of
foreign matter that night interfere with the examination.
6.2 Ensure that the surfaces to be examined are clean and free of any materials which will interfere
with the transmission of the sound into the material under examination.
6.3 The condition of the examination surface shall be determined by the examiner to be acceptable
for UT examination.
6.4 The temperature of the examination surface shall be determined to ensure that the correct
transducer and couplant has been selected.
7 EQIPMENT CALIBRATION
7.1 GENERTAL
7.1.1 The requirements of screen height linearity and amplitude control linearity shall be met with
intervals not to Exceed Three months for analog type instruments and one year for digital type
instruments, or prior to first use.
7.1.2 The screen height linearity evaluation and amplitude control linearity evaluation shall be
performed in accordance with procedure WI 111-1
(Work instruction-111-1)
7.2 GAIN CONTROL LNEARITY
The ultrasonic instrument shall provide linear vertical presentation within +/-5% of the
full screen height for 20% to 80% of the calibrated screen height. The screen height
linearity evaluation shall be met with intervals not to exceed three months for analog
type instruments and one year for digital type instruments, prior to first use. During
Inspection Technicians shall be varify gain control linearity every 4 hours.
Table: 6.2
Indication limit of % full screen Height
Gain [dB] Expected screen height [%]
(%)
+2 101 Not Less than 95
0 80 Reference Line
- 6 dB 40 37% to 43
Page 3 of 9
NDT Corrosion Control Services Co.
Dammam – Saudi Arabia Date: 15th AUG 2023
- 12 dB 20 17 to 23
-18 dB 10 8 to 12
Page 4 of 9
NDT Corrosion Control Services Co.
Dammam – Saudi Arabia Date: 15th AUG 2023
7.5.4 Take any other lower step of known thickness to on the block and couple the probe on it. With
range and delay control knobs set the left edge of the echo at … appropriate on horizontal
position trace on CRT.
7.5.5 Now horizontal range of the machine has been set to the desired range. Adjust the gate such
that the left edge of the 1st switch wall echo from the material whose thickness is to be measured
is just outside the right edge of the gate.
7.5.6 During scanning sound alarm is switched on such that comes the gate alarm will go on. That will
indicate in that particular area thickness is less than the nominal wall thickness. Extent of
reduction on thickness can be calculated from the machine.
7.5.7 A-Scan instrument with digital display
7.5.8 Determine the anticipated thickness to be examined and select a calibration block which
includes that thickness range.
7.5.9 Select proper unit of length on the machine setup (either in inch or mm) depending on
calibration block as well as the unit in which final measurement is to be done.
7.5.10 Set the proper sound velocity in the instrument i.e., 5920 meter/sec. for plain carbon steel.
7.5.11 Place the transducer on the thinner section of the calibration block and adjust the delay until
the left edge of the indication aligns with the correct CRT increment.
7.5.12 Place the transducer on the thicker section of the calibration block and adjust the range control
until the left edge of the indication aligns with the correct CRT.
7.5.13 Replace the above steps until both indications appear on their respective increments.
7.5.14 Adjust the gate to display the thickness of the selected step. If the thickness displayed is not
correct repeat the above steps.
7.6 STRAIGHT BEAM FOR LAMINATION CHECK
7.6.1 The scanning of adjacent base metal shall be performed to detect reflector that might affect
interpretation of angle beam result. Location and areas of such reflectors shall be recorded.
7.6.2 The base metal shall be scanned, where required by the referencing code section, to the extent
possible with the scanning zone until angle beam search unit. The scanning shall be performed
at a gain setting of at least 2 times the primary reference level. Evaluation shall be performed
with respect to the primary reference level.
7.6.3 As far as practicable, subject the entire volume of the plate lamination to ultrasonic examination.
Because of radii at change of sections and other local configurations, it may be impossible to
examine some sections of a plate lamination.
7.6.4 Perform the ultrasonic examination after heat treatment for mechanical properties (exclusive of
stress-relief treatments) but prior to drilling holes, cutting keyways, tapers, grooves, or machining
sections to contour. If the configuration of the Plate required for the treatment for mechanical
properties prohibits a subsequent complete examination of the plate lamination, it shall be
permissible to examine prior to treatment for mechanical properties. In such cases, reexamine
the Plate ultrasonically as completely as possible after heat treatment.
7.6.5 To ensure complete coverage of the plate volume, index the search unit with at least 15 %
overlap with each pass. For manual scanning, do not exceed a scanning rate of 6 in./s [150
mm/s]. if found any laminations shall be evaluated by 6dB/20db method as per the client
requirement, If not technicians shall be prefer 20dB method for sizing the defects.
Page 5 of 9
NDT Corrosion Control Services Co.
Dammam – Saudi Arabia Date: 15th AUG 2023
8.1 For proper sound coupling the transducer must be pressed firmly against the test surface in the
same way it was followed during calibration. on small diameter cylindrical surfaces such as pipes,
hold the transducer so that the sound barrier material visible on the probe face is aligned
perpendicular to the central axis of the pipe.
8.2 While firm hand pressure on the transducer is necessary for good readings, the probe should
never be scraped along or twisted against a rough metal surface. this will scratch the face of
the transducer and eventually degrade performance. the safest technique for moving a
transducer along a rough surface is to pick it up and reposition it for each measurement, not to
slide it along.
8.3 Examinations shall be conducted from the od surface or id surface, as applicable.
8.4 If the surface temperature is below 32°f (0°c) or above 215°f (101°c), the surface temperature
shall be measured and recorded.
8.5 Apply couplant to the probe or the specimen surface. for high temperature measurements,
couplant always applied on probe space.
8.6 The transducer acoustic barrier shall be placed perpendicular to the centre axis when taking
measurements on pipe or round objects up to 24 inches in diameter. the acoustic barrier
orientation is not important on pipe and round objects greater than 24 inches in diameter.
8.7 The transducer shall be held flat on the surface then slowly rocked tangentially after contact to
find the lowest reading. all digital readings shall be verified by viewing the a-scan presentation
on the instrument as doubling effects can give false readings on digital outputs. the lowest
reading shall be recorded. for difficult readings, additional gain may be required to avoid
grinding the transducer into the surface.
8.8 Valid reading is one that can be maintained on the gauge display for several seconds and the
value is steady within ± 0.05 mm.
8.9 Thickness readings shall be measured and recorded to three decimal places.
8.10 High temperature materials up to about 540°c can be measured with specially designed
instruments with high temperature compensation, search unit assemblies and couplants.
normalization for apparent thickness readings for elevated temperature is required. a rule of
thumb often used is as follows. the apparent thickness reading obtained from steel walls having
elevated temperatures is high (too thick) by a factor about 1% per 55°c. thus if the instrument
Page 6 of 9
NDT Corrosion Control Services Co.
Dammam – Saudi Arabia Date: 15th AUG 2023
was standardized on a piece of similar material at 20°c, and if the reading was obtained with a
surface temperature of 460°, the apparent reading should be reduced by 8 %. this correction is
an average one for many types of steel. in short, the difference of environmental temperature &
the job surface temperature needs to be considered. the below formula may be used for
temperature correction.
TA = TM X [ 1.0038-(0.00018 X Temp) ]
where TA = Actual thickness of the part in mm.
TM = Measured thickness of the part in mm.
Temp = Surface temperature in °C.
8.11 The transducer face shall be cleaned before each measurement at temperatures greater than
400°f or 204°celsius. as evaporated couplant can leave deposits on the transducer face.
8.12 Use the "touch test" to monitor the transducer temperature. if the transducer face is too hot to
be held against your skin, then it shall be cooled in air or water prior to taking readings.
8.13 For thru coat thickness measurements instruments such as Olympus model 37 DL plus, 45MG 2XT,
MG 2X (having software for thru coat) with compatible transducers can be used. for more details
user’s manual need to be referred
9 POST CLEANING
All couplant applied on the surface shall be completely removed from the specimen by wiping
with a dry absorbent cloth or paper towel, fresh water wash etc.
10 REPORTING
All measurements are recorded in Ultrasonic Thickness Measurement report, a format of which is
shown in attachement-1.
After filling up the complete report it should be countersigned by client’s inspector and filed for
future reference.
On the basis of customer requirement, customer report format can be used for spot thickness
measurement reporting
Page 7 of 9
NDT Corrosion Control Services Co.
Dammam – Saudi Arabia Date: 15th AUG 2023
Page 8 of 9
NDT Corrosion Control Services Co.
Dammam – Saudi Arabia Date: 15th AUG 2023
UT M/C Model Sr. No.: Calibration due Probe Description: Frequency: Size:
on:
ORIGINAL
TML NO. THICKNESS AS OBSERVED THICKNESS READING (mm) REMARKS
(mm)
Other
0 90 180 270 E W S N
Location
Note: 0°, 90°, 180° & 270° are in clockwise direction with respect to the flow
Page 9 of 9