ABB ME ASURE ME NT & AN AL YT ICS | D AT A SHE ET

VA Master FAM540
Metal Cone Variable Area Flowmeter
2 VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E


Measurement made easy


For measuring liquids, gases and steam


Graphical 2-line LCD display


Magnet stick operation
• Configuration also possible with closed housing


Easily adjustable for new operating conditions


Approvals for explosion protection
• ATEX
• IECEx
• FM, CSA
VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E 3

Change from one to two columns


Change from two to one column


Overview

Model FAM541 FAM544


Standard design Hygienic design

Measured error in accordance with VDE / VDI 3513 1.6 % qg = 50 % 1.6 % qg = 50 %


Reproducibility 0.25 % of measured value 0.25 % of measured value
Process connection Flange in accordance with DIN, ASME, JIS, female Thread DIN 11851, SMS 1145
thread
Nominal connection diameters DN 15 (½ in) to DN 80 (3 in) DN 25 (1 in) to DN 80 (3 in)
Maximum measuring medium temperature 400 °C (752 °F) 140 °C (284 °F)
Maximum pressure rating PN 400 / class 2500 PN 40
Wetted material Stainless steel 1.4404 (316L), 1.4571 (316Ti) Stainless steel 1.4404 (316L), 1.4571 (316Ti)
Housing material Stainless steel 1.4404 (316L) Stainless steel 1.4404 (316L)
Gasket material Viton A (DN 15 only) Viton A (DN 25 only)

Indicator / transmitter
IP degree of protection in accordance with IP 66, IP 67, NEMA 4X
EN 60529
Mechanical indicator Analog indicator with or without alarm signaling unit
Electronic indicator Analog indicator with transmitter 4 to 20 mA, with or without LCD indicator
Communication HART® protocol (only for analog indicators with transmitters)
Power supply Without alarm signalling unit: no power supply
Analog indicator with alarm signaling unit: 8 V DC via switch amplifier
Analog indicator with transmitter: 10 to 46 V DC (Ex: 10 to 30 V DC)
Indicator housing material Al Si 12 material number 3.2582 (copper content 0.1 %), stainless steel 1.4408
Paint Epoxy paint 80 to 100 μm; bottom color: RAL 7012, cover color: RAL 9002
(No housing paint for stainless steel indicator housing)

Approvals
Ex approvals ATEX / IECEx: Zone 0 / 1 / 2 / 21
FM / cCSAus: XP, IS, DIP, NI, FM Zone 1 + 2
Sealing concept Dual sealing in accordance with ANSI / ISA-12.27.01
SIL approval Analog indicator with alarm signaling unit: SIL 2
Analog indicator with transmitter: FMEDA evaluation

Change from one to two columns


4 VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E


Introduction and basics
Installation conditions
• The installation recommendations of VDI / VDE Directive
3513 must be observed.
• The flowmeter is installed vertically in the piping. The
measuring media must flow from bottom to top.
• Keep the device as far away as possible from pipe
vibrations and powerful magnetic fields.
• The nominal diameter of the piping should be the same as
the connection nominal diameter.
• Inlet and outlet sections are generally not required. Care
should be taken to avoid flow turbulence, pulsations,
pressure shocks and other flow instabilities in order to
prevent measuring inaccuracies, increased wear or
damage.
• When selecting devices, pay close attention to the
chemical resistance of the wetted parts of the device and
the process connection gaskets in relation to the
measuring medium.
• Avoid pulsating flow of the measuring medium. Use the Figure 1: Installation of the flowmeter (example)

optional float damping if necessary.


• For gaseous measuring media we recommend an Refer to VDI/VDE Directive 3513 sheet 3, Selection and
undisturbed inlet length of five times the inside diameter Installation Recommendations for Variable Area Flowmeters.
of the piping in accordance with VDI / VDE Directive 3513
sheet 3. Additional measures such as flow straighteners Sensor insulation
or perforated plates may be necessary for highly
unbalanced flow profiles.
• Avoid contamination of gaseous measuring media (refer
to BGR 132-7.3.2.2.2).
• For liquid measuring media, the nominal diameter of the
piping should be dimensioned as large as possible (if
economically viable).
• Avoid gas inclusions in liquid measuring media.
• Use valves which open slowly.
• If the flowmeter is installed in a pipeline where
decommissioning is impossible or inexpedient, a bypass
line should be provided.
• Stop and throttle valves should preferably be attached to
the outlet of the flowmeter.
Figure 2: Insulation of the flowmeter

As shown in Figure 2, the flowmeter may only be insulated up


to the flange diameter.
VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E 5

Operating conditions
A variable area flowmeter is specified for a defined set of Pressure loss
operating conditions of the measuring medium. For liquids The available operating pressure at the measuring point must
and gases, these are pressure and temperature-related be higher than the pressure loss listed for the flowmeter in
properties (density and viscosity) under operating the specifications.
conditions. It is important to also consider the pressure loss downstream
For gases, in particular, this means operating at a specific from the flowmeter due to losses in the piping and other
operating pressure and operating temperature. The specified fittings.
accuracy of the device always refers to the operating
conditions underlying the specification. Prevention of compression oscillations when measuring
gases
Temperature Data During low flow amounts and low operating pressure, so-
The following diagram shows the maximum permissible called compression oscillations of the float can occur.
measuring medium temperature depending on the ambient If the maximum upstream pressure listed in the
temperature. specifications is not reached, the flowmeter can optionally be
equipped with a gas damper.

To prevent self-generated compression oscillations, note the


following information from VDI / VDE 3513 Sheet 3:
• Select a flowmeter with the lowest possible pressure
loss.
• Minimize the piping length between the flowmeter
and the nearest upstream or downstream throttling
location.
• Set the limit of the regular measuring range from the
usual 10 to 100 % to 25 to 100 %.
• When setting the flow rate value, always start by
assuming larger values.
• Increase the operating pressure and consider its
effect on the flow rate changes due to the change in
gas density in the operating conditions.
• Minimize non-throttled, free volumes upstream and
downstream of the device.
1 Devices with alarm output −20 to 70 °C (−4 to 158 °F)
Pressure shocks
2 Devices with current output −40 to 70 °C (−40 to 158 °F)
Especially when measuring gases, it is possible that pressure
3 With insulation
or shock waves can occur when fast opening solenoid valves
Figure 3: Medium temperature (Tmedium), ambient temperature (Tamb) are employed and the piping cross-sections are not throttled,
or if there are gas bubbles in liquids.
Note As a result of the sudden expansion of the gas in the piping,
When using in potentially explosive atmospheres, observe the the float is forcibly driven against the upper floatstop.
temperature information in the ATEX / IECEx limit values Under certain conditions, this can lead to destruction of the
starting from Safety specifications ATEX / IECEx on page 25 device.
and the FM / cCSAus limit values tables starting from Safety Gas damping is not suited to compensating for pressure
specifications FM, cCSAus on page 34! shocks!
6 VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E


… Introduction and basics
Solids content in the measuring medium Measuring range limits
Variable area flowmeters have only limited suitability for
measuring media containing solids.
Depending on the concentration, particle size and type of
solid, increased mechanical abrasion may occur, especially at
the critical measuring edge of the float.
In addition, solidified deposits on the float can change its
weight and shape.
These effects can lead to erroneous measurement results,
depending on the float type.
In general, the use of appropriate filters is recommended in
such applications.
For the flow measurement of measuring media containing
magnetic particles, we recommend the installation of a
magnetic separator upstream of the variable area flowmeter.
1 Pressure loss (dP in mbar)
2 Diameter of the float weight (mm)
3 x 1000 l/h water
Float designs
Figure 5: Flow rate depending on float shape and weight (example)

For measuring range limits depending on nominal size and


float type, refer to the measuring range tables.

Calculating the viscosity influence number (VIN)


The floats are designed geometrically so as to be
independent of viscosity within broad limits. This means that
the measuring medium viscosity can vary within these limits
without resulting in a change of scale.
Figure 4: Float designs The viscosity influence number (VIN) is given in the device
measuring range tables.
Float ‘S’: • If the calculated VIN is greater than the listed VIN,
Basic shape of float. viscosity has no influence on the measured values.
Low flow rates, minimal pressure losses, essentially • If the calculated VIN is greater than the listed viscosity,
independent of viscosity; lower upstream pressure required the flowmeters are equipped from the factory with a scale
for gas measurement. adjusted to the viscosity of the measuring medium.

Float ‘N’: VUZ = η ⋅


(ρ s −1 ) ⋅ 1
Higher flow ranges, average pressure losses, well suited to
liquids with minimum viscosity; higher minimum upstream
(ρ s1 − ρ1 ) ⋅ ρ1
pressure requirements for gas measurements.
Legend
Float ‘X’: η Dynamic viscosity of the measuring medium [mPa s]
Highest flow rates, maximum pressure losses, well suited to ρs Density of the float in accordance with the selection in the
liquids with minimum viscosity; higher minimum upstream measuring range table (r = 8.02 g/cm3)
pressure requirements for gas measurements. ρs1 Density of the float being used
ρ1 Density of the measuring medium
VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E 7


Indicator and transmitter
A magnet in the float translates the height of the float as a Product highlights
measurement for the flow to the decouple-proof magnet
• Decouple-proof and hysteresis-free magnet follower
follower system of the flowmeter, which is directly connected
system.
to the indicator pin.
• The device conforms to NAMUR Recommendations NE43,
For analog indicators, the flow value is shown by the indicator
NE53, NE107.
on the scale. For indicators with a two-wire transmitter, the
• The installation and disassembly of the analog indicator
indicator position is additionally measured directly at the
on the meter tube is possible without opening the
axis.
indicator housing.
A standard eddy-current brake is used to dampen indicator
• Reading repeatability: ±0.25 % of scale end value.
vibrations and ensure outstanding readability.
• The alarm signaling unit can be retrofitted as an alarm
The indicator/transmitter unit is attached to the flowmeter
module.
with 2 screws in a reproducible manner. To facilitate
• The switching points of the alarm signaling units can be
installation, the indicator can be removed. A name plate on
adjusted without removing the scale.
the flowmeter bracket allows you to assign a unique identifier
• The set switching points of the alarm signaling units are
at a later date.
visible from the exterior.

Alarm signaling unit


Analog indicator with or without alarm
A switching amplifier is needed to operate the alarm
signaling unit signaling units.
The analog indicators are available with or without an alarm See Main ordering information FAM541 on page 41 for
signaling unit. The alarm signaling units are housed on an further information.
alarm module that can also be retrofitted. It is available as a
single (min. or max. alarm) or dual (min. or max. alarm) alarm
module. Alarm signaling unit specifications
Operating mode bistable
Reproducibility ±0.5% of scale end value
Nominal voltage 8 V DC (Ri approx. 1 kΩ)
Operating voltage 5 to 25 V DC
Switching frequency, max. 3 kHz

Switching amplifier
For analog indicators with alarm signaling units
(model FAM54xB/C/Dx)

Specifications
Power supply 230 V AC, +10 % / −15 %, 45 to 60 Hz
115 V AC, +10 % / −15 %, 45 to 60 Hz

1 Analog indicator 2 Alarm module (option) 24 V DC, +10 % / −15 %


Output One or two switching relays with potential-free
Figure 6: Analog indicator
changeover contacts
Switching capacity Maximum 250 V, maximum 4 A, maximum 500 VA
Maximum permissible Between the switching amplifier and alarm
cable length signalling unit:
300 m (984 ft)
Permissible ambient −20 to 60 °C (−4 to 140 °F)
temperature range
Electrical connection Screw terminals, maximum 2.5 mm2 (14 AWG)
Type of assembly 35 mm top-hat rail in accordance with
EN 60715:2001
IP rating IP 20 in accordance with EN 60529
Weight approx. 150 g (0.3 lb)
8 VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E


… Indicator and transmitter
Analog indicator with transmitter
The electronic transmitter uses two-wire technology. An Product highlights
optional LCD indicator offers the possibility of adjusting on- • 4 to 20 mA current output.
site to changed measuring medium parameters. • In the event of an alarm, current output can be adjusted
For models with an LCD indicator, ABB recommends that you to 21 to 23 mA (NAMUR NE43).
use a bar graph scale to avoid differences between a product • Programmable binary output (min. / max. device alarm,
scale and the flow rate shown on the LCD indicator. general alarm or pulse output).
• Parameterization using HART communication via
handheld terminal or DSV401 (SMART VISION).
• Damping: can be adjusted 1 to 100 s (5 τ).
• Low flow cut-off: 0 to 5 % for current and pulse output.
• Measuring medium parameters can be changed at any
time (pressure and temperature influence, density, units,
etc.).
• Simulation of current and binary output (manual process
execution).
• EEPROM storage of counter readings and parameters
specific to the measuring point (over 10 years).
• LCD indicator can be retrofitted.

LCD Indicator
1 Button ◄ 4 Button ▼ • High-contrast LCD indicator.
2 Button ▲ 5 Button ►
• Flow rate and flow totalizer value display.
• Plain text fault diagnostics
3 LCD indicator 6 Points for magnet stick
operation • Menu-guided parameterization with four buttons.
• Parameterization of the device using a magnet stick
Figure 7: LCD indicator with operating buttons and magnet stick markings
without opening the housing.
VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E 9


Electrical connections
Note Analog indicator with transmitter
When using in potentially explosive atmospheres, observe the Model FAM54xE/Fx
electrical data in the ATEX / IECEx limit value tables starting
from Safety specifications ATEX / IECEx on page 25 and the
FM / cCSAus limit value tables starting from Safety
specifications FM, cCSAus on page 34!

Analog indicator with alarm signaling unit


Model FAM54xB/C/Dx

F Flowmeter 2 Minimum alarm signalling unit


1 Maximum alarm signalling unit 3 Switch amplifier

Figure 8: Alarm signalling unit

Alarm signaling unit specifications


Operating mode bistable
Reproducibility ±0.5% of scale end value A Central power supply 2 Power supply

Nominal voltage 8 V DC (Ri approx. 1 kΩ) B Power supply via power supply 3 Power supply unit
unit
Operating voltage 5 to 25 V DC 4 HART® modem
Switching frequency, max. 3 kHz C HART® output
1 Flowmeter

Figure 9: Analog indicator with transmitter

Legend
UB Operating voltage
US Input terminal voltage
RB Maximum permissible load for power supply unit (e.g. indicator)
R Maximum permissible load for output circuit; is determined by
power supply unit
Functional earth
10 VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E


… Electrical connections
Cable HART output

Maximum cable length 1500 m, AWG 24 twisted and shielded. Model FAM54xE/Fx

Power supply / current output


Specifications
Terminals 31 / 32 serve both as a connection for the power
Terminals 31 / 32
supply and as a 4 to 20 mA current output for the transmitter.
The current output is also used for HART® communication. Configuration • Directly on the device
• Using DAT200 Asset Vision Basic software
and HART-DTM
Power supply
Transmission FSK modulation on current output 4 to 20 mA in
Terminals 31 / 32
accordance with the Bell 202 standard
Voltage Standard: 10 to 46 V DC
Baud rate 1200 baud
Explosion-proof design: 10 to 30 V DC.
Display Logic 1: 1200 Hz
Residual ripple maximum 5 % or. ± 1.5 Vss
Logic 0: 2200 Hz
Power consumption <1W
Maximum signal 1.2 mAss
amplitude
Current output Load (RB) at current 250 to 1500 Ω
Terminals 31 / 32 output
Output 4 to 20 mA, can be configured to 21 to 23 mA for an
alarm (in accordance with NAMUR NE43) See the separate interface description for detailed
Load Minimum > 250 Ω, maximum 1500 Ω information.
(for I at alarm = 23.0 mA)
At zero-point: ≤ 8 µ A/K System integration
Power consumption <1W In conjunction with the DTM (Device Type Manager) available
for the device, the corresponding framework applications in
accordance with FDT 0.98 or 1.2 (DAT200 Asset Vision Basic)
can be used for communication (configuration,
parameterization).
Other tool or system integrations (e.g. Emerson AMS /
Siemens PCS7) upon request.
The necessary DTMs and other files can be downloaded from
www.abb.com/flow.

Programmable binary output


Terminals 41 / 42 are used as a primary programmable binary
output. The pulse output, general alarm, min./max. alarm and
general alarm functions, as well as "no function" can be
configured using the software.
Figure 10: Current output load diagram
Binary output
Terminals 41 / 42
Output • NAMUR contact (DIN 19234) or
• Standard optoelectronic coupler (UH =
16 to 30 V DC)
Switching behavior As normally closed or normally open contacts
Internal resistance With contact open > 10 kΩ
Switching current Maximum 15 mA
Output Voltage Minimum Us 2 V DC

Pulse output
Terminals 41 / 42
Pulse width 5 to 256 ms, maximum 50 % of the period
Frequency fmax Maximum 50 Hz
VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E 11


Model FAM541 - Standard design
Temperature limits °C (°F)
Medium temperature range
−55 to 400 °C (−67 to 752 °F)

Ambient temperature range


−40 to 85 °C (−40 to 185 °F)

Figure 11: Model FAM541 Note


• For use in potentially explosive atmospheres, observe the
Scale temperature information in the ATEX / IECEx limit values
Design optionally with percentage scale or product scale. starting from Safety specifications ATEX / IECEx on
page 25 and the FM / cCSAus limit values tables starting
Gas damping from Safety specifications FM, cCSAus on page 34!
Option, for preventing compression oscillations in case of • Additionally for specifications refer to Temperature Data
gas measurements with low operating pressure. on page 5!

SIL approval
• SIL2 declaration of conformity for devices with alarm Process connections
function. • Flange in accordance with EN 1092-1 (PN 16, PN 40)
• Manufacturer's declaration (SIL1) in accordance with IEC • Flange in accordance with DIN 2501 (PN 63, PN 100)
61508 / IEC61511 for devices with a current output of • Flange with groove in accordance with DIN 2501
4 to 20 mA. • Flange in accordance with ASME B16.5
• Female thread G 1 in (½ in meter tube)
• Female thread G 1 ½ in (1 in meter tube)
Measuring accuracy
1.6 % qg = 50 % in accordance with VDE / VDI 3513, sheet 2,
optional 1 % of upper range value. Pressure ratings
• Standard pressure rating PN 40
• Flange in accordance with DIN / EN: PN 16, PN 40, PN 63,
Measuring range limits and span limits PN 100
Measuring span: 10:1 • Flange in accordance with ASME: CL 150, CL 300, CL 600
For detailed information about measuring ranges, see
Measuring range table on page 13. Other designs and pressure ratings are available upon
request.
See also Material load on page 12.

Maximum permissible operating pressure


64 bar, 100 bar, 160 bar, 250 bar
(CL 600 /CL 900 /CL 1500 / CL 2500)
12 VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E


… Model FAM541 - Standard design
Materials Material load
Materials for the indicator housing

Component Material
Indicator housing Standard: aluminum, powder-coated
Option: stainless steel 1.4408
Powder coating Epoxy paint 80 to 100 μm
Bottom color: RAL 7012
Cover color: RAL 9002
Housing gasket (O-ring) Buna N®
Sight glass Shatterproof glass

Materials for wetted parts


Figure 12: DIN flange made from stainless steel

Component Material
Meter tube Stainless steel 1.4404 (316 L)
Measuring element Stainless steel 1.4571 (316 Ti), only DN 15 (½)
Flange Stainless steel 1.4404 (316 L)
Variable area Standard: stainless steel 1.4571 (316 Ti)
Gas damping (option) Stainless steel 1.4571 (316 Ti)

Figure 13: ASME flange made from stainless steel


VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E 13

Change from two to one column

Measuring range table


Note
The ‘FlowCalc’ program, which is available as a free download at www.abb.com/flow can be used to calculate flow rates,
pressure losses, and upstream pressure requirements.
The data listed applies for water at 20 °C (68 °F), 1 kg/dm3 (62.43 lb/ft3), 1 mPas (1 cP) and for air at 0 °C (70 °F), 1013 mbar
(14.7 psia).

DN1 Upper range value2 for Meter tube / VIN4 Pressure loss5 Minimum upstream pressure for gas metering6
Water Air nominal flow rate [mbar (psi)] Without gas damping With gas damping7
[l/h (USgal/h)] [m3/h (scfh)] H2O [l/h] / Order [bar abs (psia)] [bar abs (psia)]
code3
DN 15 (½ in) 28 to 32 0.83 to 0.95 ½ in / 30 / A78) 6 80 (1.16) 4.0 (58.02) 1.0 (14.5)
(7.4 to 8.5) (37 to 45)
DN 25 (1 in) 37 to 43 1.10 to 1.28 ½ in / 40 / B78) 6 80 (1.16) 4.0 (58.02) 1.0 (14.5)
G1 in (10 to 11.4) (52 to 59)
44 to 55 1.30 to 1.63 ½ in / 50 / C78) 6 80 (1.16) 4.0 (58.02) 1.0 (14.5)
(11.6 to 14.5) (54 to 66)
56 to 64 1.66 to 1.90 ½ in / 60 / D78) 6 80 (1.16) 4.0 (58.02) 1.0 (14.5)
(14.5 to 17) (68 to 80)
77 to 83 2.29 to 2.47 ½ in / 80 / E7 16 40 (0.58) 3.0 (43.51) 1.0 (14.5)
(20.5 to 21.5) (84 to 90)
96 to 104 (25.5 to 27) 2.85 to 3.09 ½ in / 100/ F7 16 45 (0.65) 3.2 (46.41) 1.0 (14.5)
(104 to 112)
115 to 125 3.42 to 3.72 ½ in / 120 / G7 16 50 (0.73) 3.5 (50.76) 1.0 (14.5)
(30 to 33) (125 to 135)
144 to 156 (38 to 41) 4.28 to 4.64 ½ in / 150 / H7 16 60 (0.87) 3.8 (55.11) 1.0 (14.5)
(155 to 165)
188 to 212 (50 to 56) 5.59 to 6.30 ½ in / 200 / J7 16 60 (0.87) 4.0 (58.02) 1.0 (14.5)
(205 to 230)
235 to 265 (62 to 70) 6.98 to 7.88 ½ in / 250 / K7 16 65 (0.94) 4.2 (60.92) 1.0 (14.5)
(255 to 285)
282 to 318 (74 to 84) 8.38 to 9.45 ½ in / 300 / L7 16 70 (1.02) 4.4 (63.82) 1.0 (14.5)
(310 to 340)
376 to 424 11.17 to 12.60 ½ in / 400 / M7 16 75 (1.09) 4.6 (66.72) 1.0 (14.5)
(100 to 112) (410 to 460)
470 to 530 13.97 to 15.75 ½ in / 500 / N7 16 75 (1.09) 4.8 (69.62) 1.0 (14.5)
(125 to 140) (510 to 570)
565 to 635 16.79 to 18.87 ½ in / 600 / P7 16 80 (1.16) 5.0 (72.52) 1.0 (14.5)
(150 to 165) (620 to 680)
750 to 850 22.29 to 25.26 ½ in / 800 / R7 16 85 (1.23) 5.4 (78.32) 1.0 (14.5)
(200 to 220) (820 to 920)
14 VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E


… Model FAM541 - Standard design

DN1 Upper range value2 for Meter tube / VIN4 Pressure loss5 Minimum upstream pressure for gas metering6
Water Air nominal flow rate [mbar (psi)] Without gas damping With gas damping7
[l/h (USgal/h)] [m3/h (scfh)] H2O [l/h] / Order [bar abs (psia)] [bar abs (psia)]
code3
DN 25 (1 in) 280 to 656 (74 to 170) 8.32 to 19.50 1 in / 400 / A1 13 to 21 20 to 76 2.9 to 3.1 1.1 to 1.4
G1 ½ in (310 to 700) (0.3 to 1.1) (42.1 to 45.0) (16.0 to 20.3)
393 to 870 11.70 to 25.85 1 in / 600 / B1 7 to 10 27 to 76 3.0 to 3.4 1.2 to 1.4
(104 to 220) (430 to 940) (0.4 to 1.1) (43.5 to 49.3) (17.4 to 20.3)
660 to 1600 19.38 to 50.80 1 in / 1000 / C1 16 to 22 20 to 76 3.3 to 4.3 1.1 to 1.4
(170 to 450) (720 to 1850) (0.3 to 1.1) (48.0 to 62.4) (16.0 to 20.3)
975 to 2370 28.98 to 70.44 1 in / 1600 / D1 8 to 10 27 to 82 3.3 to 5.3 1.2 to 1.5
(260 to 620) (1060 to 2550) (0.4 to 1.2) (48.0 to 77.0) (17.4 to 21.8)
1650 to 4020 49.04 to 119.50 1 in / 2500 / E1 17 to 6 20 to 76 4.2 to 6.4 1.1 to 1.4
(440 to 1060) (1800 to 4300) (0.3 to 1.1) (61.0 to 92.8) (16.0 to 20.3)
2585 to 6170 76.83 to 183.50 1 in / 4000 / F1 8 to 10 27 to 82 5.2 to 8.0 1.2 to 1.5
(680 to 1600) (2800 to 6600) (0.4 to 1.2) (75.4 to 116) (17.4 to 21.8)
DN 40 (1 ½ in) 4220 to 12130 125.40 to 360.50 2 in / 8000 / A2 21 to 38 11 to 62 3.1 to 4.5 1.1 to 1.4
(1120 to 3200) (4600 to 13000) (0.2 to 0.9) (45.0 to 65.3) (16.0 to 20.3)
DN 50 (2 in) 7940 to 18460 236.0 to 584.6 2 in / 12000 / B2 13 to 17 24 to 74 3.8 to 6.2 1.1 to 1.4
(2100 to 4800) (8600 to 20000) (0.3 to 1.1) (55.1 to 90.0) (16.0 to 20.3)
11760 to 24200 349.50 to 720.00 2 in / 18000 / C2 3 to 4 28 to 72 4.4 to 7.5 1.1 to 1.4
(3100 to 6400) (13000 to 27000) (0.4 to 1.0) (63.8 to 109) (16.0 to 20.3)
DN 80 (3 in) 7000 to 21010 208.00 to 624.40 3 in / 12000 / A3 22 to 54 6 to 48 3.4 to 5.4 1.1 to 1.3
(1850 to 5500) (7600 to 22000) (0.1 to 0.7) (49.3 to 78.3) (16.0 to 18.9)
18090 to 35010 537.70 to 1040 3 in / 25000 / B3 18 to 25 24 to 65 4.8 to 7.4 1.1 to 1.4
(4800 to 9200) (19500 to 38000) (0.3 to 1.0) (69.6 to 107) (16.0 to 20.3)
29750 to 53810 795 to 1600 3 in / 40000 / C3 4 to 5 26 to 68 6.0 to 9.2 1.1 to 1.4
(7000 to 14000) (29000 to 58000) (0.4 to 1.0) (87.0 to 133) (16.0 to 20.3)

1 Connection nominal diameter


2 The upper range value can be selected anywhere within these limits. The measuring span is 10:1.
Example: upper range value 12 m3/h (USgal/h) water, measuring span of the device 1.2 to 12 m3/h (USgal/h) water.
3 Order code for meter tube / float construction (versions 18, 19), see Main ordering information FAM541 on page 41.
4 Viscosity influence number (VIN), see Calculating the viscosity influence number (VIN) on page 6.
5 The listed pressure loss values are based on the respective upper range value.
6 The minimum required static pressure (abs) to prevent compression oscillations (float bounce) in the meter tube. The specifications for the minimum
required pressures with and without damping are based on average installation conditions. Some installations may have higher values (high free
volume upstream / downstream of the device).
7 These meter tube / float combinations always require gas piston damping for gas measurements.
VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E 15

Dimensions

1 Threaded bushing ½ in NPT 4 N number of holes


2 Cable entry M20 × 1.5 5 Ground terminal
3 Plug M25 × 1.5 (model FAM541-A only) 6 LCD display only for FAM541-F

Figure 14: Dimensions for flange devices in mm (in), tolerances ±2 mm

Meter size Pressure DN ØD Øk ØL N A C G


rating [mm (in)] [mm (in)] [mm (in)] [mm (in)] [mm (in)] [mm (in)]
½ in PN 40 DN 15 95.0 (3.74) 65.0 (2.56) 14.0 (0.55) 4 250.0 (9.84) 87.0 (3.43) 118.0 (4.65)
PN 63 / 100 105.0 (4.13) 75.0 (2.95) 14.0 (0.55) 4 258.0 (10.16) 87.0 (3.43) 118.0 (4.65)
CL 150 ½ in 89.0 (3.50) 60.3 (2.37) 15.9 (0.63) 4 250.0 (9.84) 87.0 (3.43) 118.0 (4.65)
CL 300 95.2 (3.75) 66.7 (2.63) 15.9 (0.63) 4 250.0 (9.84) 87.0 (3.43) 118.0 (4.65)
CL 600 95.2 (3.75) 66.5 (2.62) 15.7 (0.62) 4 260.0 (10.24) 87.0 (3.43) 118.0 (4.65)
1 in PN 40 DN 25 115.0 (4.53) 85.0 (3.35) 14.0 (0.55) 4 250.0 (9.84) 87.0 (3.43) 118.0 (4.65)
PN 63 / 100 140.0 (5.51) 100.0 (3.94) 18.0 (0.71) 4 262.0 (10.31) 87.0 (3.43) 118.0 (4.65)
CL 150 1 in 107.9 (4.25) 79.4 (3.13) 15.9 (0.63) 4 250.0 (9.84) 87.0 (3.43) 118.0 (4.65)
CL 300 123.8 (4.87) 88.9 (3.50) 19.0 (0.75) 4 250.0 (9.84) 87.0 (3.43) 118.0 (4.65)
CL 600 124.0 (4.88) 88.9 (3.50) 19.0 (0.75) 4 262.0 (10.31) 87.0 (3.43) 118.0 (4.65)
2 in PN 40 DN 40 150 (5.91) 110 (4.33) 18.0 (0.71) 4 250.0 (9.84) 102.0 (4.02) 130.0 (5.12)
CL 150 1 ½ in 127 (5.0) 98.4 (3.87) 15.7 (0.62) 4 250.0 (9.84) 102.0 (4.02) 130.0 (5.12)
CL 300 1 ½ in 155.3 (6.11) 114.3 (4.5) 22.2 (0.87) 4 250.0 (9.84) 102.0 (4.02) 130.0 (5.12)
PN 40 DN 50 165.0 (6.50) 125.0 (4.92) 18.0 (0.71) 4 250.0 (9.84) 102.0 (4.02) 130.0 (5.12)
PN 63 180.0 (7.09) 135.0 (5.31) 22.0 (0.87) 4 262.0 (10.31) 102.0 (4.02) 130.0 (5.12)
PN 100 195.0 (7.68) 145.0 (5.71) 26.0 (1.02) 4 266.0 (10.47) 102.0 (4.02) 130.0 (5.12)
CL 150 2 in 152.4 (6.00) 120.6 (4.75) 19.0 (0.75) 4 250.0 (9.84) 102.0 (4.02) 130.0 (5.12)
CL 300 165.1 (6.50) 127.0 (5.00) 19.0 (0.75) 8 250.0 (9.84) 102.0 (4.02) 130.0 (5.12)
CL 600 165.1 (6.50) 127.0 (5.00) 19.0 (0.75) 8 274.0 (10.79) 102.0 (4.02) 130.0 (5.12)
16 VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E


… Model FAM541 - Standard design

Meter size Pressure DN ØD Øk ØL N A C G


rating [mm (in)] [mm (in)] [mm (in)] [mm (in)] [mm (in)] [mm (in)]
3 in PN 40 DN 80 200.0 (7.87) 160.0 (6.30) 18.0 (0.71) 8 250.0 (9.84) 132.0 (5.20) 144.0 (5.67)
PN 63 215.0 (8.46) 170.0 (6.69) 22.0 (0.87) 8 258.0 (10.16) 132.0 (5.20) 144.0 (5.67)
PN 100 230.0 (9.06) 180.0 (7.09) 26.0 (1.02) 8 272.0 (10.71) 132.0 (5.20) 144.0 (5.67)
CL 150 3 in 190.5 (7.50) 152.4 (6.00) 19.0 (0.75) 4 250.0 (9.84) 132.0 (5.20) 144.0 (5.67)
CL 300 209.5 (8.25) 168.3 (6.63) 22.2 (0.87) 8 250.0 (9.84) 132.0 (5.20) 144.0 (5.67)
CL 600 209.5 (8.25) 168.1 (6.62) 22.2 (0.87) 8 278.0 (10.94) 132.0 (5.20) 144.0 (5.67)

Overall length for devices with female thread connectors

Process connection Installation length [mm (in)]


Female thread 1 in 296 (11.65)
Female thread 1 ½ in 304 (11.97)

Weight in kg (lb)

Device size (meter tube)


Indicator housing material DN 15 (½ in) DN 25 (1 in) DN 50 (2 in) DN 80 (3 in)
Al Si 12 4.5 / (9.9) 5.8 / (12.8) 9.5 / (20.9) 15.7 / (34.6)
Stainless steel 1.4408 7.0 / (15.4) 8.3 / (18.3) 12.0 / (26.4) 18.2 / (40.1)

Change from one to two columns


VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E 17


Model FAM544 - hygienic design
Temperature limits °C (°F)
Medium temperature range
−55 to 400 °C (−67 to 752 °F)

Ambient temperature range


−40 to 85 °C (−40 to 185 °F)

Figure 15: Model FAM544 in hygienic design


Note
• For use in potentially explosive atmospheres, observe the
This model has been specially constructed with threaded temperature information in the ATEX / IECEx limit values
connections in accordance with DIN 11851 to meet the more starting from Safety specifications ATEX / IECEx on
stringent hygienic requirements of the food and beverage page 25 and the FM / cCSAus limit values tables starting
industry in respect of biological cleaning. from Safety specifications FM, cCSAus on page 34!
All parts of the flowmeter that come into contact with the • Additionally for specifications refer to Temperature Data
measuring medium are welded and polished. There are no on page 5!
gaps or other empty areas. The device is suitable for cleaning
or sterilization with steam, acids and alkali. The device is also
CIP-cleanable. Process connections
• Threaded spud in accordance with DIN 11851 (SC 25 to SC
Scale 80)
Design optionally with percentage scale or product scale. • Threaded spud in accordance with SMS 1145
(DN 38 to DN 76)
Gas damping
Not available. Connection dimensions

SIL approval
Nominal Measuring tube SMS 1145 DIN 11851
• SIL2 declaration of conformity for devices with alarm
diameter sizes
function.
DN 25 ½ in SC 25 —
• Manufacturer's declaration (SIL1) in accordance with
DN 40 1 in SC 40 DN 38
IEC 61508 / IEC 61511 for devices with a current output of
DN 50 2 in SC 50 DN 51
4 to 20 mA.
DN 80 3 in SC 80 DN 76

Measuring accuracy
Pressure ratings
1.6 % qg = 50 % in accordance with VDE / VDI 3513, sheet 2,
optional 1 % of upper range value. Threaded spuds in accordance with DIN 11851:
• PN 25 for DN 50 to DN 80 (2 to 3 in)
• PN 40 for DN 25 to DN 40 (1 to 1 ½ in)
Measuring range limits and span limits
Threaded spud in accordance with DIN 1145:
Measuring span: 10:1 • PN 6 for (DN 38 to DN 76)
For detailed information about measuring ranges, see
Measuring range table on page 19. See also Material load on page 18.

Maximum permissible operating pressure


Refer to Material load on page 18.
18 VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E


… Model FAM544 - hygienic design
Materials Material load
Materials for the indicator housing
Process Nominal diameter PSmax TSmin TSmax
connection
Component Material DIN 11851 DN 15 to DN 40 40 bar −40 °C 140 °C
Indicator housing Standard: aluminum, powder-coated (½ to 1 ½ in) (580 psi) (-40 °F) (284 °F)
Option: stainless steel 1.4408 DN 50 to 80 25 bar −40 °C 140 °C
Powder coating Epoxy paint 80 to 100 μm (2 to 3 in) (362 psi) (-40 °F) (284 °F)
Bottom color: RAL 7012 SMS 1145 DN 38 to 76 6 bar −40 °C 140 °C
Cover color: RAL 9002 (1 ½ to 3 in) (87 psi) (-40 °F) (284 °F)
Housing gasket (O-ring) Buna N®
Sight glass Shatterproof glass

Materials for wetted parts

Component Material
Meter tube Stainless steel 1.4404 (316 L)
Measuring element Stainless steel 1.4571 (316 Ti), DN 15 (½ in) only
Variable area Stainless steel 1.4571 (316 Ti)
VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E 19

Change from two to one column

Measuring range table


Note
The ‘FlowCalc’ program, which is available as a free download at www.abb.com/flow can be used to calculate flow rates,
pressure losses, and upstream pressure requirements.
The data listed applies for water at 20 °C (68 °F), 1 kg/dm3 (62.43 lb/ft3), 1 mPas (1 cP) and for air at 0 °C (70 °F), 1013 mbar
(14.7 psia).

DN1 Upper range value2 for Meter tube / nominal flow rate VIN4 Pressure loss5 Minimum upstream
Water Air H2O [l/h] / Order code3 [mbar (psi)] pressure for gas
[l/h (USgal/h)] [m3/h (scfh)] metering6
[bar abs (psia)]
DN 25 28 to 32 — ½ in / 30 / A7 6 80 (1.16) —
(1 in7) (7.4 to 8.5)
37 to 43 — ½ in / 40 / B7 6 80 (1.16) —
(10 to 11.4)
44 to.55 (11.6 to 14.5) — ½ in / 50 / C7 6 80 (1.16) —
56 to 64 — ½ in / 60 / D7 6 80 (1.16) —
(14.5 to 17)
77 to 83 (20.5 to 21.5) 2.29 to 2.47 (84 to 90) ½ in / 80 / E7 16 40 (0.58) 3.0 (43.51)
96 to 104 (25.5 to 27) 2.85 to 3.09 ½ in / 100/ F7 16 45 (0.65) 3.2 (46.41)
(104 to 112)
115 to 125 3.42 to 3.72 ½ in / 120 / G7 16 50 (0.73) 3.5 (50.76)
(30 to 33) (125 to 135)
144 to 156 4.28 to 4.64 ½ in / 150 / H7 16 60 (0.87) 3.8 (55.11)
(38 to 41) (155 to 165)
188 to 212 5.59 to 6.30 ½ in / 200 / J7 16 60 (0.87) 4.0 (58.02)
(50 to 56) (205 to 230)
235 to 265 6.98 to 7.88 ½ in / 250 / K7 16 65 (0.94) 4.2 (60.92)
(62 to 70) (255 to 285)
282 to 318 8.38 to 9.45 ½ in / 300 / L7 16 70 (1.02) 4.4 (63.82)
(74 to 84) (310 to 340)
376 to 424 11.17 to 12.60 ½ in / 400 / M7 16 75 (1.09) 4.6 (66.72)
(100 to 112) (410 to 460)
470 to 530 13.97 to 15.75 ½ in / 500 / N7 16 75 (1.09) 4.8 (69.62)
(125 to 140) (510 to 570)
565 to 635 16.79 to 18.87 ½ in / 600 / P7 16 80 (1.16) 5.0 (72.52)
(150 to 165) (620 to 680)
750 to 850 22.29 to 25.26 ½ in / 800 / R7 16 85 (1.23) 5.4 (78.32)
(200 to 220) (820 to 920)
20 VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E


… Model FAM544 - hygienic design

DN1 Upper range value2 for Meter tube / nominal flow rate VIN4 Pressure loss5 Minimum upstream
Water Air H2O [l/h] / Order code3 [mbar (psi)] pressure for gas
[l/h (USgal/h)] [m3/h (scfh)] metering6
[bar abs (psia)]
DN 40 280 to 656 (74 to 170) 8.32 to 19.50 1 in / 400 / A1 13 to 21 20 to 76 2.9 to 3.1
(1 ½ in) (310 to 700) (0.3 to 1.1) (42.1 to 45.0)
393 to 870 11.70 to 25.85 1 in / 600 / B1 7 to 10 27 to 76 3.0 to 3.4
(104 to 220) (430 to 940) (0.4 to 1.1) (43.5 to 49.3)
660 to 1600 19.38 to 50.80 1 in / 1000 / C1 16 to 22 20 to 76 3.3 to 4.3
(170 to 450) (720 to 1850) (0.3 to 1.1) (48.0 to 62.4)
975 to 2370 28.98 to 70.44 1 in / 1600 / D1 8 to 10 27 to 82 3.3 to 5.3
(260 to 620) (1060 to 2550) (0.4 to 1.2) (48.0 to 77.0)
1650 to 4020 49.04 to 119.50 1 in / 2500 / E1 17 to 6 20 to 76 4.2 to 6.4
(440 to 1060) (1800 to 4300) (0.3 to 1.1) (61.0 to 92.8)
2585 to 6170 76.83 to 183.50 1 in / 4000 / F1 8 to 10 27 to 82 5.2 to 8.0
(680 to 1600) (2800 to 6600) (0.4 to 1.2) (75.4 to 116)
DN 50 4220 to 12130 125.40 to 360.50 2 in / 8000 / A2 21 to 38 11 to 62 3.1 to 4.5
(2 in) (1120 to 3200) (4600 to 13000) (0.2 to 0.9) (45.0 to 65.3)
7940 to 18460 236.0 to 584.6 2 in / 12000 / B2 13 to 17 24 to 74 3.8 to 6.2
(2100 to 4800) (8600 to 20000) (0.3 to 1.1) (55.1 to 90.0)
11760 to 24200 349.50 to 720.00 2 in / 18000 / C2 3 to 4 28 to 72 4.4 to 7.5
(3100 to 6400) (13000 to 27000) (0.4 to 1.0) (63.8 to 109)
DN 80 7000 to 21010 208.00 to 620 3 in / 12000 / A3 22 to 54 6 to 48 3.4 to 5.4
(3 in) (1850 to 5500) (7600 to 22000) (0.1 to 0.7) (49.3 to 78.3)
18090 to 35010 537.70 to 1040 3 in / 25000 / B3 18 to 25 24 to 65 4.8 to 7.4
(4800 to 9200) (19500 to 38000) (0.3 to 1.0) (69.6 to 107)
29750 to 53810 795 to 1550 3 in / 40000 / C3 4 to 5 26 to 68 6.0 to 9.2
(7000 to 14000) (29000 to 58000) (0.4 to 1.0) (87.0 to 133)

1 Connection nominal diameter


2 The upper range value can be selected anywhere within these limits. The measuring span is 10:1.
Example: upper range value 12 m3/h (USgal/h) water, measuring span of the device 1.2 to 12 m3/h (USgal/h) water.
3 Order code for meter tube / float construction (versions 18, 19), see Main ordering information FAM544 on page Fehler! Textmarke nicht definiert..
4 Viscosity influence number (VIN), see Calculating the viscosity influence number (VIN) on page 6.
5 The listed pressure loss values are based on the respective flow rate end value.
6 The minimum required static pressure (abs) to prevent compression oscillations (float bounce) in the meter tube. The specifications for the minimum
required pressures are based on average installation conditions. Some installations may have higher values (high free volume upstream / downstream
of the device).
7 Only for threaded spuds in accordance with DIN 11851
VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E 21

Dimensions

1 Threaded bushing ½ in NPT 4 Ground terminal


2 Cable entry M20 × 1.5 5 LCD display only for FAM541-F
3 Plug M25 × 1.5 (model FAM541-A only)

Figure 16: Dimensions in mm (in), tolerances ±2 mm

Dimensions for devices with threaded spuds in accordance with DIN 11851

Meter size DN Pressure rating ØD A* B C G


[mm (in)] [mm (in)] [mm (in)] [mm (in)] [mm (in)]
½ in 25 (1 in) PN 40 Rd 52 × ¹⁄₆ in 270.0 (10.63) 51.5 (2.03) 87.0 (3.43) 118.0 (4.65)
1 in 40 (1 ½ in) PN 40 Rd 65 × ¹⁄₆ in 270.0 (10.63) 51.5 (2.03) 87.0 (3.43) 118.0 (4.65)
2 in 50 (2 in) PN 25 Rd 78 × ¹⁄₆ in 270.0 (10.63) 51.5 (2.03) 102.0 (4.02) 130.0 (5.12)
3 in 80 (3 in) PN 25 Rd 110 × ¹⁄₆ in 272.0 (10.71) 52.5 (2.07) 132.0 (5.20) 144.0 (5.67)
* Dimension A: design with threaded spud DIN 11851
Dimension M: hygienic design with threaded spud in accordance with DIN 11851 (available upon request)
22 VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E


… Model FAM544 - hygienic design
Dimensions for devices with threaded spuds in accordance with SMS 1145

Meter size DN [mm] Pressure rating ØD A [mm (in)] B [mm (in)] C [mm (in)] G [mm (in)]
1 in 38 PN 6 Rd 60 × ¹⁄₆ in 300.0 (11.81) 66.5 (2.62) 87.0 (3.43) 118.0 (4.65)
2 in 51 Rd 70 × ¹⁄₆ in 102.0 (4.02) 130.0 (5.12)
3 in 76 Rd 98 × ¹⁄₆ in 132.0 (5.20) 144.0 (5.67)

Weight in kg (lb)

Device size (meter tube)


Indicator housing material DN 15 (½ in) DN 25 (1 in) DN 50 (2 in) DN 80 (3 in)
Al Si 12 4.5 (9.9) 5.8 (12.8) 9.0 (19.8) 15.7 (34.6)
Stainless steel 1.4408 7.0 (15.4) 8.3 (18.3) 11.5 (25.3) 18.2 (40.1)
VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E 23


Use in potentially explosive atmospheres in accordance with ATEX and IECEx
Note

Further information on the approval of devices for use in potentially explosive atmospheres can be found in the type
examination certificates or the relevant certificates at www.abb.com/flow.

Device overview
The devices are designed for maximum versatility. This is achieved through a combination of several types of protection within
each device. All devices are suitable for use in potentially explosive atmospheres with combustible dust. For detailed
installation instructions and terminal assignments, refer to Electrical connections on page 9.

Zone 1 Zone 2 Standard / No explosion protection Order code*

Potential equalization

* Order code ‘Explosion protection and approvals’ (versions 9, 10); refer to Ordering Information on page 41.
24 VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E


… Use in potentially explosive atmospheres in accordance with ATEX and IECEx
Ex marking

Note
• Depending on the design, a specific marking in accordance with ATEX or IECEx applies.
• ABB reserves the right to modify the Ex-marking. Refer to the name plate for the exact marking.

Model FAM54xAx (analog indicator without alarm signaling unit)


Marking Type of protection Order code* Limit value table
ATEX II 1/2 G Ex h T6 ... T1 Gb Constructional safety A4, A9, B1 Table 5 on page 30
II 2 D Ex h T85°C ... Tmedium Db Constructional safety
II 2 D Ex h tb IIIC T85°C ... Tmedium Db Dust-explosion protection
* Order code ‘Explosion protection and approvals’ (versions 9, 10); refer to Ordering Information on page 41.

Model FAM54xB/C/Dx (analog indicator with alarm signaling unit)


Marking Type of protection Order code* Limit value table
ATEX II 1/2 G Ex h ia IIC T6 ... T1 Ga/Gb Intrinsically Safe A4 Table 2 on page 27
II 1/3 G Ex ec h IIC T6 ... T1 Ga/Gc Non-sparking materials Table 4 on page 29
II 2D Ex h tb IIIC T85°C ... Tmedium Db Dust-explosion protection Table 2 on page 27,
Table 4 on page 29

IECEx Ex h ia IIC T6 ... T1 Ga/Gb Intrinsically Safe Table 2 on page 27


Ex ec h IIC T6 ... T1 Ga/Gc Non-sparking materials Table 4 on page 29
Ex h tb IIIC T85°C ... Tmedium Db Dust-explosion protection Table 2 on page 27,
Table 4 on page 29
ATEX II 1/2 G Ex db h IIC T6 ... T1 Ga/Gb Flameproof / Explosion-Proof A9 Table 3 on page 28
II 1/2 G Ex h ia IIC T6 ... T1 Ga/Gb Intrinsically Safe Table 2 on page 27
II 1/3 G Ex ec h IIC T6 ... T1 Ga/Gc Non-sparking materials Table 4 on page 29
II 2 D Ex h tb IIIC T85°C ... Tmedium Db Dust-explosion protection Table 2 on page 27,
Table 3 on page 28,
Table 4 on page 29
IECEx Ex db h IIC T6 ... T1 Ga/Gb Flameproof / Explosion-Proof Table 3 on page 28
Ex h ia IIC T6 ... T1 Ga/Gb Intrinsically Safe Table 2 on page 27
Ex ec h IIC T6 ... T1 Ga/Gc Non-sparking materials Table 4 on page 29
Ex h tb IIIC T85°C ... Tmedium Db Dust-explosion protection Table 2 on page 27,
Table 3 on page 28,
Table 4 on page 29

ATEX II 1/3 G Ex ec h IIC T6 ... T1 Ga/Gc Non-sparking materials B1 Table 4 on page 29


II 2 D Ex h tb IIIC T85°C ... Tmedium Db Dust-explosion protection Table 4 on page 29
IECEx Ex ec h IIC T6 ... T1 Ga/Gc Non-sparking materials Table 4 on page 29
Ex h tb IIIC T85°C ... Tmedium Db Dust-explosion protection Table 4 on page 29
* Order code ‘Explosion protection and approvals’ (versions 9, 10); refer to Ordering Information on page 41.
VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E 25

Model FAM54xE/Fx (analog indicator with transmitter and with or without LCD indicator)
Marking Type of protection Order code* Limit value table
ATEX II 1/2 G Ex h ia IIC T4 ... T1 Ga/Gb Intrinsically Safe A4 Table 1 on page 26
II 1/3 G Ex ec h ic T6 ... T1Ga/Gc Non-sparking materials
II 2 D Ex h tb IIIC T85°C ... Tmedium Db Dust explosion protection
IECEx Ex h ia IIC T4 ... T1 Ga/Gb Intrinsically Safe
Ex ec h ic IIC T6 ... T1Ga/Gc Non-sparking materials
Ex h tb IIIC T85°C ... Tmedium Db Dust explosion protection
ATEX II 1/2 G Ex db h IIC T6 ... T1 Ga/Gb Flameproof / Explosion-Proof A9 Table 1 on page 26
II 1/2 G Ex h ia IIC T4 ... T1 Ga/Gb Intrinsically Safe
II 1/3 G Ex ec h ic IIC T6 ... T1 Ga/Gc Non-sparking materials
II 2 D Ex h tb IIIC T85°C ... Tmedium Db Dust explosion protection
IECEx Ex db h IIC T6 ... T1 Ga/Gb Flameproof / Explosion-Proof
Ex h ia IIC T4 ... T1 Ga/Gb Intrinsically Safe
Ex ec h ic IIC T6 ... T1 Ga/Gc Non-sparking materials
Ex h tb IIIC T85°C ... Tmedium Db Dust explosion protection
ATEX II 1/3 G Ex ec h ic IIC T6 ... T1 Ga/Gc Non-sparking materials B1 Table 4 on page 29
II 2 D Ex h tb IIIC T85°C ... Tmedium Db Dust explosion protection
IECEx Ex ec h ic IIC T6 ... T1 Ga/Gc Non-sparking materials
Ex h tb IIIC T85°C ... Tmedium Db Dust explosion protection
* Order code ‘Explosion protection and approvals’ (versions 9, 10); refer to Ordering Information on page 41.

Safety specifications ATEX / IECEx


Special conditions for type of protection ‘Ex td’ (dust explosion protection)

Model FAM54xB/C/D/E/Fx
The following differing temperature data applies to usage in areas with combustible dust.

Order code Ambient temperature Tamb Maximum permissible measuring medium temperature Tmedium
A4, A9, B1 −50 °C to +60 °C 250 °C
−50 °C to +40 °C 340 °C
−50 °C to +20 °C 430 °C
26 VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E


… Use in potentially explosive atmospheres in accordance with ATEX and IECEx
Table 1: Analog indicator with transmitter, with / without LCD indicator

Model FAM54xE/Fx
Type of protection: flameproof enclosure, intrinsic safety, non-sparking equipment, dust explosion protection.

Order Marking Terminals Electrical values Tamb Temp. class Tmedium Insulation
code* −20 °C to (−50 °C to ) Maximum
A4, A9 ATEX: 31 / 32** Ui = 30 V 40 °C T1 440 °C No
II 1/2 G Ex h ia IIC T4 ... T1 Ga/Gb Ii = 110 mA 40 °C T1 375 °C yes
II 1/3 G Ex ec h ic T6 ... T1Ga/Gc Pi = 770 mW 40 °C T1 260 °C yes
II 2 D Ex h tb IIIC T85°C ... Tmedium Db Ci = 5.3 nF 50 °C T1 300 °C yes
Li = 266 µH 50 °C T2 290 °C yes
IECEx: 50 °C T2 220°C yes
Ex h ia IIC T4 ... T1 Ga/Gb 41 / 42** Ui = 30 V 60 °C T2 320 °C No
Ex ec h ic IIC T6 ... T1Ga/Gc Ii = 30 mA 60 °C T2 230 °C yes
Ex h tb IIIC T85°C ... Tmedium Db Pi = 115 mW 60 °C T3 170 °C yes
Ci = 4.8 nF 70 °C T3 195 °C No
Li = 133 µH 70 °C T3 150 °C yes
70 °C T4 125 °C yes
A9 ATEX: 31 / 32*** Umax = 46 V 40 °C T1 440 °C No
II 1/2 G Ex db h IIC T6 ... T1 Ga/Gb 40 °C T1 375 °C yes
II 1/2 G Ex h ia IIC T4 ... T1 Ga/Gb 40 °C T1 260 °C yes
II 1/3 G Ex ec h ic IIC T6 ... T1 Ga/Gc 50 °C T1 300 °C yes
II 2 D Ex h tb IIIC T85°C ... Tmedium Db 50 °C T2 290 °C yes
50 °C T2 220°C yes
IECEx: 41 / 42*** Umax = 30 V 60 °C T2 320 °C No
Ex db h IIC T6 ... T1 Ga/Gb Imax = 30 mA 60 °C T2 230 °C yes
Ex h ia IIC T4 ... T1 Ga/Gb Pmax = 115 mW 60 °C T3 170 °C yes
Ex ec h ic IIC T6 ... T1 Ga/Gc
60 °C T4 130 °C yes
Ex h tb IIIC T85°C ... Tmedium Db
60 °C T5 95 °C yes
60 °C T6 80 °C yes
A4, A9, B1 ATEX: 31 / 32*** Umax = 46 V 40 °C T1 440 °C No
II 1/3 G Ex ec h ic IIC T6 ... T1 Ga/Gc 40 °C T1 375 °C yes
II 2 D Ex h tb IIIC T85°C ... Tmedium Db 40 °C T1 260 °C yes
50 °C T1 300 °C yes
IECEx: 50 °C T2 290 °C yes
Ex ec h ic IIC T6 ... T1 Ga/Gc 50 °C T2 220°C yes
Ex h tb IIIC T85°C ... Tmedium Db 41 / 42*** Umax = 30 V 60 °C T2 320 °C No
Imax = 30 mA 60 °C T2 230 °C yes
Pmax = 115 mW 60 °C T3 170 °C yes
70 °C T3 195 °C No
70 °C T3 150 °C yes
70 °C T4 130 °C yes
* Order code ‘Explosion protection and approvals’ (versions 9, 10); refer to Ordering Information on page 41.
** for connection to an intrinsically safe circuit
*** for connection to a non-intrinsically safe circuit If the device is to be operated subsequently with type of protection "intrinsically safe", then Umax = 60
V may not be exceeded.
VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E 27

Table 2: Analog indicator with alarm signaling unit

Model FAM54xB/C/Dx
Type of protection: intrinsic safety, dust explosion protection.

Order Marking Terminals Electrical values Tamb Temp. class Tmedium Insulation
code* −20 °C to (−50 °C to ) Maximum
A4, A9 ATEX: 41 / 42** Ui = 16 V 40 °C T1 440 °C No
II 1/2 G Ex h ia IIC T6 ... T1 Ga/Gb 51 / 52** Ii = 25 mA 40 °C T1 375 °C yes
II 2D Ex h tb IIIC T85°C ... Tmedium Db Pi = 64 mW 40 °C T1 260 °C yes
Ci = 50 nF 50 °C T1 300 °C yes
IECEx: Li = 250 µH 50 °C T2 290 °C yes
Ex h ia IIC T6 ... T1 Ga/Gb 50 °C T2 220°C yes
Ex h tb IIIC T85°C ... Tmedium Db 60 °C T2 320 °C No
60 °C T2 230 °C yes
60 °C T3 170 °C yes
70 °C T3 195 °C No
70 °C T3 150 °C yes
70 °C T4 130 °C yes
70 °C T5 95 °C yes
60 °C T6 80 °C yes
41 / 42** Ui = 16 V 40 °C T1 440 °C No
51 / 52** Ii = 52 mA 40 °C T1 375 °C yes
Pi = 169 mW 40 °C T1 260 °C yes
Ci = 50 nF 50 °C T1 300 °C yes
Li = 250 µH 50 °C T2 290 °C yes
50 °C T2 220°C yes
60 °C T2 320 °C No
60 °C T2 230 °C yes
60 °C T3 170 °C yes
70 °C T3 195 °C No
70 °C T3 150 °C yes
70 °C T4 130 °C yes
60 °C T5 60 °C yes
50 °C T5 90 °C No
40 °C T6 60 °C yes
41 / 42** Ui = 16 V 40 °C T1 440 °C No
51 / 52** Ii = 76 mA 40 °C T1 310 °C yes
Pi = 242 mW 40 °C T2 190 °C yes
Ci = 50 nF 50 °C T2 340 °C No
Li = 250 µH 50 °C T2 230 °C yes
60 °C T2 230°C No
60 °C T3 160 °C yes
70 °C T4 120 °C No
70 °C T4 100 °C yes
40 °C T5 60 °C yes
30 °C T6 30 °C yes
* Order code ‘Explosion protection and approvals’ (versions 9, 10); refer to Ordering Information on page 41.
** for connection to an intrinsically safe circuit
28 VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E


… Use in potentially explosive atmospheres in accordance with ATEX and IECEx
Table 3: Analoganzeiger mit Grenzwertgeber

Model FAM54xB/C/Dx
Type of protection: flameproof enclosure, dust explosion protection.

Order Marking Terminals Electrical values Tamb Temp. class Tmedium Insulation
code* −20 °C to (−50 °C to ) Maximum
A9 ATEX: 41 / 42** Umax = 16 V 40 °C T1 440 °C No
II 1/2 G Ex db h IIC T6 ... T1 Ga/Gb 51 / 52** Imax = 25 mA 40 °C T1 375 °C yes
II 2 D Ex h tb IIIC T85°C ... Tmedium Db Pmax = 64 mW 40 °C T1 260 °C yes
50 °C T1 300 °C yes
IECEx: 50 °C T2 290 °C yes
Ex db h IIC T6 ... T1 Ga/Gb 50 °C T2 220°C yes
Ex h tb IIIC T85°C ... Tmedium Db 60 °C T2 320 °C No
60 °C T2 230 °C yes
60 °C T3 170 °C yes
70 °C T3 195 °C No
70 °C T3 150 °C yes
70 °C T4 130 °C yes
70 °C T5 95 °C yes
60 °C T6 80 °C yes
41 / 42** Umax = 16 V 40 °C T1 440 °C No
51 / 52** Imax = 52 mA 40 °C T1 375 °C yes
Pmax = 169 mW 40 °C T1 260 °C yes
50 °C T1 300 °C yes
50 °C T2 290 °C yes
50 °C T2 220°C yes
60 °C T2 320 °C No
60 °C T2 230 °C yes
60 °C T3 170 °C yes
70 °C T3 195 °C No
70 °C T3 150 °C yes
70 °C T4 130 °C yes
60 °C T5 60 °C yes
50 °C T5 90 °C No
40 °C T6 60 °C yes
41 / 42** Umax = 16 V 40 °C T1 440 °C No
51 / 52** Imax = 76 mA 40 °C T1 310 °C yes
Pmax = 242 mW 40 °C T2 190 °C yes
50 °C T2 340 °C No
50 °C T2 230 °C yes
60 °C T2 230°C No
60 °C T3 160 °C yes
70 °C T4 120 °C No
70 °C T4 100 °C yes
40 °C T5 60 °C yes
30 °C T6 30 °C yes
* Order code ‘Explosion protection and approvals’ (versions 9, 10); refer to Ordering Information on page 41.
** for connection to a non-intrinsically safe circuit If the device is to be operated subsequently with type of protection "intrinsically safe", then Umax = 60
V may not be exceeded.
VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E 29

Table 4: Analoganzeiger mit Grenzsignalgeber

Model FAM54xB/C/Dx
Type of protection: non-sparking materials, dust explosion protection.

Order Marking Terminals Electrical values Tamb Temp. class Tmedium Insulation
code* −20 °C to (−50 °C to ) Maximum
A4, A9, B1 ATEX: 41 / 42** Umax = 16 V 40 °C T1 440 °C No
II 1/3 G Ex ec h IIC T6 ... T1 Ga/Gc 51 / 52** Imax = 25 mA 40 °C T1 375 °C yes
II 2D Ex h tb IIIC T85°C ... Tmedium Db Pmax = 64 mW 40 °C T1 260 °C yes
50 °C T1 300 °C yes
IECEx: 50 °C T2 290 °C yes
Ex ec h IIC T6 ... T1 Ga/Gc 50 °C T2 220°C yes
Ex h tb IIIC T85°C ... Tmedium Db 60 °C T2 320 °C No
60 °C T2 230 °C yes
60 °C T3 170 °C yes
70 °C T3 195 °C No
70 °C T3 150 °C yes
70 °C T4 130 °C yes
70 °C T5 95 °C yes
60 °C T6 80 °C yes
41 / 42** Umax = 16 V 40 °C T1 440 °C No
51 / 52** Imax = 52 mA 40 °C T1 375 °C yes
Pmax = 169 mW 40 °C T1 260 °C yes
50 °C T1 300 °C yes
50 °C T2 290 °C yes
50 °C T2 220°C yes
60 °C T2 320 °C No
60 °C T2 230 °C yes
60 °C T3 170 °C yes
70 °C T3 195 °C No
70 °C T3 150 °C yes
70 °C T4 130 °C yes
60 °C T5 60 °C yes
50 °C T5 90 °C No
40 °C T6 60 °C yes
A4, A9, B1 ATEX: 41 / 42** Umax= 16 V 40 °C T1 440 °C No
II 1/3 G Ex ec h IIC T6 ... T1 Ga/Gc 51 / 52** Imax = 76 mA 40 °C T1 310 °C yes
II 2 D Ex h tb IIIC T85°C ... Tmedium Db Pmax = 242 mW 40 °C T2 190 °C yes
50 °C T2 340 °C No
IECEx: 50 °C T2 230 °C yes
Ex ec h IIC T6 ... T1 Ga/Gc 60 °C T2 230°C No
Ex h tb IIIC T85°C ... Tmedium Db 60 °C T3 160 °C yes
70 °C T4 120 °C No
70 °C T4 100 °C yes
40 °C T5 60 °C yes
30 °C T6 30 °C yes
* Order code ‘Explosion protection and approvals’ (versions 9, 10); refer to Ordering Information on page 41.
** for connection to a non-intrinsically safe circuit If the device is to be operated subsequently with type of protection "intrinsically safe", then Umax = 60
V may not be exceeded.
30 VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E


… Use in potentially explosive atmospheres in accordance with ATEX and IECEx
Table 5: Analoganzeiger ohne Grenzsignalgeber

Model FAM54xAx
Type of protection: constructional safety, dust explosion protection

Order Marking Terminals Electrical values Tamb Temp. class Tmedium Insulation
code* −20 °C to (−50 °C to ) Maximum
A4, A9, B1 ATEX: n.a. n. a 70 °C T1 440 °C yes
II 1/2 G Ex h T6 ... T1 Gb 70 °C T2 290 °C yes
II 2 D Ex h T85°C ... Tmedium Db 70 °C T3 190 °C yes
II 2 D Ex h tb IIIC T85°C ... Tmedium Db 70 °C T4 130 °C yes
IECEx: 70 °C T5 95 °C yes
Ex h tb IIIC T85°C ... Tmedium Db 70 °C T6 80°C yes
* Order code ‘Explosion protection and approvals’ (versions 9, 10); refer to Ordering Information on page 41.
Change from one to two columns

Electrical connections
Potentially explosive atmosphere Non-hazardous area

Figure 18: Terminals 31 / 32, power supply / load

UV Power supply

RB Maximum permissible load in the power supply (z. B. indicator)

The minimum voltage UV = 0 V is based on a load of 0 Ω.

1 Namur switching amplifier 3 Potential equalization


2 Safety barrier

Figure 17: ATEX / IECEx electrical connection

Terminal Function
31 / 32 Power supply / current output / HART output
41 / 42 Programmable binary output
Alarm signaling unit (min.)
51 / 52 Alarm signaling unit (max.)
VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E 31

Signal cable Earthing


The Ex calculations are based on temperatures of 80 °C
(176 °F) at the cable input. For this reason, cables with a
specification of 80 °C (176 °F) must be used.
For cables limited to 70 °C (158 °F), a maximum ambient
temperature of Tamb = 60 °C (140 °F) applies.
The resulting new maximum measuring medium
temperatures are determined as follows:
• Calculate a new ambient temperature:
Tamb new = Tamb + 10 °C (18 °F).
• Use the calculated ambient temperature Tamb new to
determine the new associated permissible measuring
medium temperatures in the tables.
• Please use the limit value tables with the original 1 Ground terminal
ambient temperature Tamb to determine the relevant
2 Potential equalization in accordance with EN 60079-0
temperature class. See Safety specifications
ATEX / IECEx on page 25, Tables 1 to 5. Figure 19: Grounding

Example: The FAM540 housing must be correctly earthed in order to


• Tamb = 50 °C (122 °F) becomes Tamb new = 60 °C (140 ensure proper function and safe operation.
°F). Copper wires with a minimum cross-section of 6 mm2
• Determine the measuring medium temperature Tamb = (AWG 10) must be used to connect to the potential
60 °C (140 °F). equalization.
• Determine the temperature class for Tamb = 50 °C (122
°F). Note
The operator must make sure that when connecting the
protective ground (PE), there are no potential differences
between protective ground (PE) and potential equalization,
even in the event of a fault.
Change from two to one column
32 VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E


Use in potentially explosive atmospheres in accordance with FM and cCSAus
Note

Further information on the approval of devices for use in potentially explosive atmospheres can be found in the type
examination certificates or the relevant certificates at www.abb.com/flow.

Device overview
The devices are designed for maximum versatility. This is achieved through a combination of several types of protection within
each device. All devices are suitable for use in potentially explosive atmospheres with combustible dust. For detailed
installation instructions and terminal assignments, refer to Electrical connections on page 9.

Class 1 Division 1 / Zone 1 Class 1 Division 2 / Zone 2 Standard / No explosion protection Order code*

Potential equalization

* Order code ‘Explosion protection and approvals’ (versions 9, 10); refer to Ordering Information on page 41.
VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E 33

Ex marking

Note
• Depending on the design, a specific marking in accordance with FM applies.
• ABB reserves the right to modify the Ex-marking. Refer to the name plate for the exact marking.

Model FAM54xAx (analog indicator without alarm signaling unit and model FAM54xB/C/Dx (analog indicator with alarm signaling unit)
Marking Type of protection Order code* Limit value table
FM XP / CL I / DIV 1 / GP ABCD / T6 … T1 Explosionproof F3 Table 2 on page 35
CL I, ZN 1 AEx d IIC T6 … T1
IS / CL I,II,III / DIV 1 / GP ABCDEFG / T6 … T1 Intrinsic Safety F3, F4 Table 1 on page 34, Table 2 on page 35
CL I, ZN 1 AEx ia IIC T6 … T1
DIP / CL II, III / DIV 1 / GP EFG / T6 … T1 Dust-Ignitionproof
NI /CL I,II / DIV 2 / GP ABCDFG / T5 … T1 Non-Incendive F3, F4 Table 1 on page 34, Table 2 on page 35,
NI / CL III T5 … T1 Table 3 on page 36
CL II, ZN 2 AEx ec II T5 … T1
cCSAus XP / CL I / DIV 1 / GP BCD / T6 … T1 Explosionproof F3 Table 2 on page 35
Ex d IIC T6 … T1
IS / CL I,II,III / DIV 1 / GP ABCDEFG / T6 … T1 Intrinsic Safety F3, F4 Table 1 on page 34,
Ex ia IIC T6 … T1
DIP / CL II, III / DIV 1 / GP EFG / T6 … T1 Dust-Ignitionproof F3, F4 Table 1 on page 34, Table 2 on page 35,
DIP A21 TA 85°C … Tmedium Table 3 on page 36
NI /CL I,II / DIV 2 / GP ABCDFG / T5 … T1 Non-Incendive F3, F4 Table 3 on page 36
NI / CL III T5 … T1
Ex ec II T5 to T1
* Order code ‘Explosion protection and approvals’ (version digit no. 9, 10); see Ordering Information on page 41.

Model FAM54xE/Fx (analog indicator with transmitter / with or without LCD display)
Marking Type of protection Order code* Limit value table
FM XP / CL I / DIV 1 / GP ABCD / T6 … T1 Explosionproof F3 Table 4 on page 37
CL I, ZN 1 AEx d IIC T6 … T1
IS / CL I,II,III / DIV 1 / GP ABCDEFG / T4 … T1 Intrinsic Safety F3, F4
CL I, ZN 1 AEx ia IIC T4 … T1
DIP / CL II, III / DIV 1 / GP EFG / T6 … T1 Dust-Ignitionproof
NI /CL I,II / DIV 2 / GP ABCDFG / T4 … T1 Non-Incendive
NI / CL III T4 … T1
CL II, ZN 2 AEx ec IIC T6 … T1
cCSAus XP / CL I / DIV 1 / GP BCD / T6 … T1 Explosionproof F3
Ex d IIC T6 … T1
IS / CL I,II,III / DIV 1 / GP ABCDEFG / T4 … T1 Intrinsic Safety F3, F4
Ex ia IIC T4 … T1
DIP / CL II, III / DIV 1 / GP EFG / T6 … T1 Dust-Ignitionproof
DIP A21 TA 85°C to Tmedium
NI /CL I,II / DIV 2 / GP ABCDFG / T4 … T1 Non-Incendive
NI / CL III T4 … T1
Ex ec IIC T6 … T1
* Order code ‘Explosion protection and approvals’ (version digit no. 9, 10), see Ordering Information on page 41.
34 VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E


… Use in potentially explosive atmospheres in accordance with FM and cCSAus
Safety specifications FM, cCSAus

Table 1: Analog indicator with alarm signaling unit (temperature data for FM in °F, for cCSAus in °C)
Order Marking Terminals Electrical Tamb Tamb Temp. class Tmedium Insulation
code* values −58 °F to −50 °C to Maximum
F3, F4 FM 41 / 42*** Ui = 16 V 104 °F 40 °C T1 824 °F 440 °C No
IS / CL I,II,III / DIV 1 / GP ABCDEFG / 51 / 52*** Ii = 25 mA 104 °F 40 °C T1 707 °F 375 °C yes
T6 ... T1** Pi = 64 mW 104 °F 40 °C T1 500 °F 260 °C yes
Ci = 50 nF 122 °F 50 °C T1 572 °F 300 °C yes
DIP / CL II, III / DIV 1 / GP EFG / T6 ... T1 Li = 250 µH 122 °F 50 °C T2 554 °F 290 °C yes
122 °F 50 °C T2 428 °F 220°C yes
CL I, ZN 1 AEx ia IIC T6 ... T1 140 °F 60 °C T2 608 °F 320 °C No
140 °F 60 °C T2 446 °F 230 °C yes
cCSAus
140 °F 60 °C T3 338 °F 170 °C yes
IS / CL I,II,III / DIV 1 / GP ABCDEFG /
158 °F 70 °C T3 383 °F 195 °C No
T6 ... T1**
158 °F 70 °C T3 302 °F 150 °C yes
158 °F 70 °C T4 266 °F 130 °C yes
DIP / CL II, III / DIV 1 / GP EFG / T6 ... T1
158 °F 70 °C T5 203 °F 95 °C yes
140 °F 60 °C T6 176 °F 80 °C yes
Ex ia IIC T6 ... T1
41 / 42*** Ui = 16 V 104 °F 40 °C T1 824 °F 440 °C No
51 / 52*** Ii = 52 mA 104 °F 40 °C T1 707 °F 375 °C yes
DIP A21 TA 85°C to Tmedium
Pi = 169 mW 104 °F 40 °C T1 500 °F 260 °C yes
Ci = 50 nF 122 °F 50 °C T1 572 °F 300 °C yes
Li = 250 µH 122 °F 50 °C T2 554 °F 290 °C yes
122 °F 50 °C T2 428°F 220°C yes
140 °F 60 °C T2 608 °F 320 °C No
140 °F 60 °C T2 446 °F 230 °C yes
140 °F 60 °C T3 338 °F 170 °C yes
158 °F 70 °C T3 383 °F 195 °C No
158 °F 70 °C T3 302 °F 150 °C yes
158 °F 70 °C T4 266 °F 130 °C yes
140 °F 60 °C T5 140 °F 60 °C yes
122 °F 50 °C T5 194 °F 90 °C No
104 °F 40 °C T6 140 °F 60 °C yes
41 / 42*** Ui = 16 V 104 °F 40 °C T1 824 °F 440°C No
51 / 52*** Ii = 76 mA 104 °F 40 °C T1 590 °F 310 °C yes
Pi = 242 mW 104 °F 40 °C T2 374 °F 190 °C yes
Ci = 50 nF 122 °F 50 °C T2 644 °F 340 °C No
Li = 250 µH 122 °F 50 °C T2 446 °F 230 °C yes
140 °F 60 °C T2 446 °F 230 °C No
140 °F 60 °C T3 320 °F 160 °C yes
158 °F 70 °C T4 248 °F 120 °C No
158 °F 70 °C T4 212 °F 100 °C yes
104 °F 40 °C T5 140 °F 60 °C yes
86 °F 30 °C T6 86 °F 30 °C yes
* Order code ‘Explosion protection and approvals’ (versions 9, 10); refer to Ordering Information on page 41.
** ‘IS’ installation in accordance with Installation Drawing SDM-10-A0253.
*** for connection to an intrinsically safe circuit
VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E 35

Table 2: Analog indicator with alarm signaling unit (temperature data for FM in °F, for cCSAus in °C)
Order Marking Terminals Electrical values Tamb Tamb Temp. class Tmedium Insulation
code* −58 °F to −50 °C to Maximum
F3 FM 41 / 42** Umax = 16 V 104 °F 40 °C T1 824 °F 440 °C No
XP / CL I / DIV 1 / GP ABCD / T6 ... T1 51 / 52** Imax = 25 mA 104 °F 40 °C T1 707 °F 375 °C yes
Pmax = 64 mW 104 °F 40 °C T1 500 °F 260 °C yes
DIP / CL II, III / DIV 1 / GP EFG / 122 °F 50 °C T1 572 °F 300 °C yes
T6 ... T1 122 °F 50 °C T2 554 °F 290 °C yes
122 °F 50 °C T2 428 °F 220°C yes
CL I, ZN 1 AEx d IIC T6 ... T1 140 °F 60 °C T2 608 °F 320 °C No
140 °F 60 °C T2 446 °F 230 °C yes
cCSAus
140 °F 60 °C T3 338 °F 170 °C yes
XP / CL I / DIV 1 / GP BCD / T6 ... T1
158 °F 70 °C T3 383 °F 195 °C No
158 °F 70 °C T3 302 °F 150 °C yes
DIP / CL II, III / DIV 1 / GP EFG /
158 °F 70 °C T4 266 °F 130 °C yes
T6 ... T1
158 °F 70 °C T5 203 °F 95 °C yes
140 °F 60 °C T6 176 °F 80 °C yes
Ex d IIC T6 ... T1
41 / 42** Umax = 16 V 104 °F 40 °C T1 824 °F 440 °C No
51 / 52** Imax = 52 mA 104 °F 40 °C T1 707 °F 375 °C yes
DIP A21 TA 85°C ... Tmedium
Pmax = 169 mW 104 °F 40 °C T1 500 °F 260 °C yes
122 °F 50 °C T1 572 °F 300 °C yes
122 °F 50 °C T2 554 °F 290 °C yes

122 °F 50 °C T2 428°F 220°C yes

140 °F 60 °C T2 608 °F 320 °C No

140 °F 60 °C T2 446 °F 230 °C yes


140 °F 60 °C T3 338 °F 170 °C yes
158 °F 70 °C T3 383 °F 195 °C No
158 °F 70 °C T3 302 °F 150 °C yes
158 °F 70 °C T4 266 °F 130 °C yes
140 °F 60 °C T5 140 °F 60 °C yes
122 °F 50 °C T5 194 °F 90 °C No
104 °F 40 °C T6 140 °F 60 °C yes
41 / 42** Umax = 16 V 104 °F 40 °C T1 824 °F 440°C No
51 / 52** Imax = 76 mA 104 °F 40 °C T1 590 °F 310 °C yes
Pmax = 242 mW 104 °F 40 °C T2 374 °F 190 °C yes
122 °F 50 °C T2 644 °F 340 °C No
122 °F 50 °C T2 446 °F 230 °C yes
140 °F 60 °C T2 446°F 230 °C No
140 °F 60 °C T3 320 °F 160 °C yes
158 °F 70 °C T4 248 °F 120 °C No
158 °F 70 °C T4 212 °F 100 °C yes
104 °F 40 °C T5 140 °F 60 °C yes
86 °F 30 °C T6 86 °F 30 °C yes
* Order code ‘Explosion protection and approvals’ (versions 9, 10); refer to Ordering Information on page 41.
** for connection to a non-intrinsically safe circuit
36 VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E


… Use in potentially explosive atmospheres in accordance with FM and cCSAus

Table 3: Analog indicator with / without alarm signaling unit (temperature data for FM in °F, for cCSAus in °C)
Order Marking Terminals Electrical values Tamb Tamb Temp. class Tmedium Insulation
code* −58 °F to −50 °C to Maximum
F3**, F4** FM 41 / 42*** Umax = 16 V 104 °F 40 °C T1 824 °F 440 °C No
NI /CL I,II / DIV 2 / GP ABCDFG / 51 / 52*** Imax = 25 mA 104 °F 40 °C T1 707 °F 375 °C yes
T5 ... T1 Pmax = 64 mW 104 °F 40 °C T1 500 °F 260 °C yes
122 °F 50 °C T1 572 °F 300 °C yes
NI / CL III / T5 ... T1 122 °F 50 °C T2 554 °F 290 °C yes
122 °F 50 °C T2 428°F 220°C yes
CL II, ZN 2 AEx ec II T5 ... T1 140 °F 60 °C T2 608 °F 320 °C No
140 °F 60 °C T2 446 °F 230 °C yes
cCSAus
140 °F 60 °C T3 338 °F 170 °C yes
NI /CL I,II / DIV 2 / GP ABCDFG /
158 °F 70 °C T3 383 °F 195 °C No
T5 ... T1
158 °F 70 °C T3 302 °F 150 °C yes
158 °F 70 °C T4 266 °F 130 °C yes
NI / CL III / T5 ... T1
158 °F 70 °C T5 203 °F 95 °C yes
41 / 42*** Umax = 16 V 104 °F 40 °C T1 824 °F 440 °C No
Ex ec II T6 ... T1
51 / 52*** Imax = 52 mA 104 °F 40 °C T1 707 °F 375 °C yes
Pmax = 169 mW 104 °F 40 °C T1 500 °F 260 °C yes
DIP A21 TA 85°C to Tmedium
122 °F 50 °C T1 572 °F 300 °C yes
122 °F 50 °C T2 554 °F 290 °C yes
122 °F 50 °C T2 428°F 220°C yes
140 °F 60 °C T2 608 °F 320 °C No
140 °F 60 °C T2 446 °F 230 °C yes
140 °F 60 °C T3 338 °F 170 °C yes
158 °F 70 °C T3 383 °F 195 °C No
158 °F 70 °C T3 302 °F 150 °C yes
158 °F 70 °C T4 266 °F 130 °C yes
140 °F 60 °C T5 140 °F 60 °C yes
122 °F 50 °C T5 194 °F 90 °C No
41 / 42*** Umax= 16 V 104 °F 40 °C T1 824 °F 440°C No
51 / 52*** Imax = 76 mA 104 °F 40 °C T1 590 °F 310 °C yes
Pmax = 242 mW 104 °F 40 °C T2 374 °F 190 °C yes
122 °F 50 °C T2 644 °F 340 °C No
122 °F 50 °C T2 446 °F 230 °C yes
140 °F 60 °C T2 446°F 230 °C No
140 °F 60 °C T3 320 °F 160 °C yes
158 °F 70 °C T4 248 °F 120 °C No
158 °F 70 °C T4 212 °F 100 °C yes
104 °F 40 °C T5 140 °F 60 °C yes
* Order code ‘Explosion protection and approvals’ (versions 9, 10); refer to Ordering Information on page 41.
** For connection in Division 2 or Zone 2.
*** for connection to a non-intrinsically safe circuit
VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E 37

Table 4: Analog indicator with transmitter, with or without LCD indicator (temperature data for FM in °F, for cCSAus in °C)
Order Marking Terminals Electrical values Tamb Tamb Temp. Tmedium Insulation
code1 −58 °F to −50 °C to class Maximum
F3, F4 FM 31 / 322 3 Ui = 30 V 104 °F 40 °C T1 824 °F 440 °C No
IS / CL I,II,III / DIV 1 / GP ABCDEFG / Ii = 110 mA 104 °F 40 °C T1 707 °F 375 °C yes
T4 ... T1 Pi = 770 mW 104 °F 40 °C T1 500 °F 260 °C yes
DIP / CL II, III / DIV 1 / GP EFG / Ci = 5.3 nF 122 °F 50 °C T1 572 °F 300 °C yes
T6 ... T1 Li = 266 µH 122 °F 50 °C T2 554 °F 290 °C yes
CL I, ZN 1 AEx ia IIC T4 ... T1 122 °F 50 °C T2 428°F 220°C yes
cCSAus 41 / 422 Ui = 30 V 140 °F 60 °C T2 608 °F 320 °C No
IS / CL I,II,III / DIV 1 / GP ABCDEFG / Ii = 30 mA 140 °F 60 °C T2 446 °F 230 °C yes
T4 ... T1 2 Pi = 115 mW 140 °F 60 °C T3 338 °F 170 °C yes
DIP / CL II, III / DIV 1 / GP EFG / Ci = 4.8 nF 158 °F 70 °C T3 383 °F 195 °C No
T6 ... T1 Li = 133 µH 158 °F 70 °C T3 302 °F 150 °C yes
Ex ia IIC T6 ... T1
158 °F 70 °C T4 257 °F 125 °C yes
DIP A21 TA 85°C to Tmedium
F3 FM 31 / 324 Umax = 46 V 104 °F 40 °C T1 824 °F 440 °C No
XP / CL I / DIV 1 / GP ABCD / T6 ... T1 104 °F 40 °C T1 707 °F 375 °C yes
DIP / CL II, III / DIV 1 / GP EFG / 104 °F 40 °C T1 500 °F 260 °C yes
T6 ... T1 122 °F 50 °C T1 572 °F 300 °C yes
CL I, ZN 1 AEx d IIC T6 ... T1 122 °F 50 °C T2 554 °F 290 °C yes
cCSAus 122 °F 50 °C T2 428°F 220 °C yes
XP / CL I / DIV 1 / GP BCD / T6 ... T1 41 / 424 Umax = 30 V 140 °F 60 °C T2 608 °F 320 °C No
DIP / CL II, III / DIV 1 / GP EFG / Imax = 30 mA 140 °F 60 °C T2 446 °F 230°C yes
T6 ... T1 Pmax = 115 mW 140 °F 60 °C T3 338 °F 170 °C yes
Ex d IIC T6 ... T1
140 °F 60 °C T4 266 °F 130 °C yes
DIP A21 TA 85°C to Tmedium
140 °F 60 °C T5 203 °F 95 °C yes
140 °F 60 °C T6 176 °F 80 °C yes
1) Order code ‘Explosion protection and approvals’ (versions 9, 10); refer to Ordering Information on page 41.
2) For connection to an intrinsically safe circuit.
3) Installation in accordance with Installation Drawing SDM-10-A0253.
4) For connection to a non-intrinsically safe circuit.
38 VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E


… Use in potentially explosive atmospheres in accordance with FM and cCSAus

Continuation of Table 4: Analog indicator with transmitter, with or without LCD indicator (temperature data for FM in °F, for cCSAus in °C)
Order Marking Terminals Electrical values Tamb Tamb Temp. Tmedium Insulation
code1 −58 °F to −50 °C to class Maximum
F35, F45 FM 31 / 324 Umax = 46 V 104 °F 104 °F T1 824 °F 440 °C No
NI /CL I,II / DIV 2 / GP ABCDFG / 104 °F 104 °F T1 707 °F 375 °C yes
T4...T1 104 °F 104 °F T1 500 °F 260 °C yes
NI / CL III / T4...T1 122 °F 122 °F T1 572 °F 300 °C yes
CL II, ZN 2 AEx ec IIC T4...T1 122 °F 122 °F T2 554 °F 290 °C yes
122 °F 122 °F T2 428°F 220 °C yes
cCSAus 140 °F 140 °F T2 608 °F 320 °C No
NI /CL I,II / DIV 2 / GP ABCDFG /
41 / 424 Umax = 30 V 140 °F 60 °C T2 446 °F 230 °C yes
T4…T1
Imax = 30 mA 140 °F 60 °C T3 338 °F 170 °C yes
NI / CL III / T4…T1
Pmax = 115 mW 158 °F 70 °C T3 383 °F 195 °C No
Ex ec IIC T4…T1
158 °F 70 °C T3 302 °F 150 °C yes
DIP A21 TA 85°C to Tmedium
158 °F 70 °C T4 266 °F 130 °C yes
158 °F 70 °C T5 203 °F 95 °C yes
86 °F 30 °C T6 77 °F 25 °C yes
1) Order code ‘Explosion protection and approvals’ (versions 9, 10), see Ordering Information on page 41.
2) For connection to an intrinsically safe circuit.
3) Installation in accordance with Installation Drawing SDM-10-A0253.
4) For connection to a non-intrinsically safe circuit.
5) For connection in Division 2 or Zone 2.

Change from one to two columns


VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E 39

Electrical connections
Installation instructions
The concept of intrinsic safety enables multiple intrinsic
safety devices with FM or CSA approval to be interconnected,
without entity parameters being examined specifically,
subject to observation of the following conditions:
• Uo or Voc or Vt ≤ 0 V max, Io or Isc or It ≤ I max, Ca or Co ≥ Ci
+ Ccable, La or Lo ≥ Li + Lcable, Po ≤ Pi.
• For installation in Class II and III environments, dust-
proof ignition blocks must be used.
• Devices connected to such equipment must not use or
generate rms or DC voltage in excess of 250 V.
• The installation must meet the requirements in
accordance with ANSI / ISA RP 12.6 ‘Installation of
Intrinsically Safe Systems for Hazardous Locations’
1 FM Approved IS Barrier 2 Potential equalization
and the National Electrical Code (ANSI / NFPA 70)
Figure 20: FM / cCSAus electrical connection sections 504, 505 and CEC.
• The configuration of the associated apparatus must
Terminal Function have Factory Mutual Research and CSA approval in
accordance with the entity concept.
31 / 32 Power supply / current output / HART output
• Devices must be installed in accordance with the
41 / 42 Binary output
manufacturer-supplied installation drawing of the
Alarm signaling unit (min.)
associated equipment.
51 / 52 Alarm signaling unit (max.)
• Changes to drawings are only permitted subject to
prior approval from Factory Mutual Research and CSA.
• Only shielded twisted pair cables may be used (see
above).

Note
For intrinsically safe installations, the FAM540 must be
installed as illustrated in the "Intrinsic Safety Control Drawing
SDM-10-A0253".

Figure 21: Terminals 31 / 32, power supply / load

UV Power supply

RB Maximum permissible load in the power supply (z. B. indicator)

The minimum voltage UV = 0 V is based on a load of 0 Ω.


40 VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E


… Use in potentially explosive atmospheres in accordance with FM and cCSAus
Earthing Power supply

Installation must comply with the requirements of the


National Electric Code® (ANSI / NFPA70).
Unless specified otherwise in regional or national standards,
power supply lines must be dimensioned to AWG 20.
Installation must be carried out as outlined in the latest
edition of the manufacturer's instruction manual.

A power supply with the following requirements must be


used to provide power:
• SELV (safety extra-low voltage) with LPS (current-
limited source) and double or reinforced insulation.
• Maximum output current of 8 A (current-limited
output).
1 Ground terminal • In accordance with National Electric Code®
2 Potential equalization in accordance with EN 60079-0 (ANSI / NFPA70) connected to NEC class 2.
Figure 22: Grounding

The FAM540 housing must be correctly earthed in order to


ensure proper function and safe operation.
Copper wires with a minimum cross-section of 6 mm2
(AWG 10) must be used to connect to the potential
equalization.

Note
The operator must make sure that when connecting the
protective ground (PE), there are no potential differences
between protective ground (PE) and potential equalization,
even in the event of a fault.

Signal cable
For ambient temperatures below 5 °C (41 °F) or above
40 °C (104 °F), signal cables that are suited for the
minimum/maximum ambient temperatures in question must
be used.
Only use signal cables made from copper, copper-coated
aluminum, or aluminum.
The recommended tightening torque for the terminals is
0.8 Nm (7 in lb) or higher, in accordance with the
specification.
Change from two to one column
VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E 41


Change from two to one column

Ordering Information
Main ordering information FAM541

VA Master FAM541 Metal Cone Variable Area Flowmeter, standard, for FAM541 X X XX XX X X XX X XX
general purposes
Display Type / Output Signal
Analog indicator / Without A
Analog indicator / Min-Alarm B
(Amplifier D163A011U01 bis U03 as a separate item)
Analog indicator / Max-Alarm C
(Amplifier D163A011U01 bis U03 as a separate item)
Analog indicator / Min- and Max-Alarm D
(Amplifier D163A011U04 bis U06 as a separate item)
Analog indicator / 4 bis 20 mA with HART protocol E
Analog indicator with LCD display / 4 bis 20 mA with HART protocol F
Housing Material / Cable Gland
Aluminium / Cable gland M20 × 1.5 (Indicator type A with blind plug M25 × 1.5) 1
Aluminium / Thread ½ in NPT (Indicator type A with blind plug M25 × 1.5) 2
Stainless steel / Cable gland M20 × 1.5 (Indicator type A with blind plug M25 × 1.5) 31
Stainless steel / Thread ½ in NPT (Indicator type A with blind plug M25 × 1.5) 41
Explosion Protection Certification and Approvals
Without Y0
ATEX / IECEx, (Zone 2 / 21), Ex ec B12
ATEX / IECEx, (Zone 1 / 21), Ex i, Ex ec A42
ATEX / IECEx, (Zone 1 / 21), Ex d, Ex i A93
FM / CSA, Class I, Div 1, 2 (Zone 1, 2), XP + IS, NI F34
FM / CSA, Class I, Div 1, 2 (Zone 1, 2), IS, NI F44
Process Connection
Flange F1
Flange with groove (DIN 2512) F2
Female thread metric (DN 25 [1 in] = PN 100, DN 40 [1 ½ in] = PN 40) T15
Connection Size
DN 15 (½ in) A
DN 25 (1 in) B
DN 40 (1-½ in) N
DN 50 (2 in) C
DN 80 (3 in) D
Float Design
Standard 1
With gas damping 3
Continued see next page
1 Not available with Ex d or XP applications
2 Version with analog indicator without output signal: Type of protection "c" only available with ATEX approval
3 With Ex d approved cable gland. Not available for analog indicator without output signal
4 Only with cable entry ½ in NPT
5 Not available with FM / CSA approval, only available in 1 in and 1 ½ in size
42 VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E


… Ordering Information

VA Master FAM541 Metal Cone Variable Area Flowmeter, standard, for XX X XX


general purposes
Pressure Rating
PN 16 D2
PN 40 D4
PN 63 D5*
PN 100 D6*
ASME CL 150 A1
ASME CL 300 A3
ASME CL 600 A6*
JIS 10K J1*
Design Level
(Specified by ABB) X
Tube / Float Combination
(Specified by ABB) See flow range tables. XX
* Not available with DN 40 (1 ½ in)

Additional ordering information

VA Master FAM541 Metal Cone Variable Area Flowmeter, standard, for general purposes XX XX XX XX XX XX
Primary
Without, secondary only Y0*
Standard Y1**
Measuring Deviation
Class 1.6 acc. VDE / VDI 3513 A1**
1 % of full scale AA
4 % of full scale (high viscosity, without calculation) AK
4 % of full scale (high viscosity, with calculation) AL
Material: 3.1, 3.2 Inspection / NACE
Material monitoring with inspection certificate 3.1 acc. EN 10204 C2
Material monitoring with inspection certificate 3.2 acc. EN 10204 C3
Material monitoring NACE MR 01-75 with inspection certificate 3.1 acc. EN 10204 CN
Material: 2.1 Compliance
Declaration of compliance with the order 2.1 acc. EN 10204 C4
Certificates: 3.1 Visual, Dimensional and Functional Test
Inspection certificate 3.1 acc. EN 10204 of visual, dimensional and functional test C6
Certificates: 3.1 PMI Test
Inspection certificate 3.1 acc. EN 10204 of positive material identification PMI CA
* For Ex devices upon request
** Standard, automatically specified by ABB if not specified by customer
VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E 43

Additional ordering information XX XX XXX XX XX


Certificates: 3.1 Pressure Test
Pressure test acc. AD2000 CB
Test package (pressure test, non-destructive test, welder and welding procedure CP
certificate) (Non-destructive test = X-ray for welding neck flanges, dye penetration
test for slip-on flanges)
Certificates: Calibration, Test Report
Certificate of compliance for calibration 2.1 acc. EN 10204 CM
Inspection certificate 3.1 acc. EN 10204 of calibration, with test report CE
Other Explosion Protection Certifications and Approvals
INMETRO (Brazil) EB1
Documentation Language
German M1
English M5*
Language package Western Europe / Scandinavia (Languages: DE, EN, DA, ES, FR, IT, NL, PT, FI, SV) MW
Language package Eastern Europe (Languages: DE, EL, CS, ET, LV, LT, HU, PL, SK, SL, RO, BG) ME
Applications
Degreased for oxygen applications P1
* Standard, automatically specified by ABB if not specified by customer

Additional ordering information XX XX XX XX


Device Identification Plate
Stainless steel plate with TAG no. T0
Adhesive label with TAG no. TC*
Float Material
AISI 316Ti SST (1.4571) F1*
Scale Design
Direct reading scale SD*
Percentage scale SP
Bargraph SB**
Ambient Temperature Range
Extended -40 bis 70 °C (-40 bis 158 °F) / -40 bis 60 °C (-40 bis 140 °F) with Ex protection Exd + Ex tD R5***
Standard -20 bis 70 °C (-4 bis 158 °F) / -20 bis 60 °C (-4 bis 140 °F) with Ex protection Exd + Ex tD R6*
* Standard, automatically specified by ABB if not specified by customer
** Recommended for indicator with transmitter 4 bis 20 mA with LCD display
*** Not available with analog indicator with alarm
44 VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E


… Ordering Information
Main ordering information FAM544

VA Master FAM544 Metal Cone Variable Area Flowmeter, hygienic design, for specific food and FAM544 X X XX XX X X XX X XX
pharmacy applications
Display Type / Output Signal
Analog indicator / Without A
Analog indicator / Min-Alarm (Amplifier D163A011U01 to U03 as a separate item) B
Analog indicator / Max-Alarm (Amplifier D163A011U01 to U03 as a separate item) C
Analog indicator / Min- and Max-Alarm (Amplifier D163A011U04 to U06 as a separate item) D
Analog indicator / 4 to 20 mA with HART protocol E
Analog indicator with LCD display / 4 to 20 mA with HART protocol F
Housing Material / Cable Gland
Aluminium / Cable gland M20 × 1.5 (Indicator type A with blind plug M25 × 1.5) 1
Aluminium / Thread ½ in NPT (Indicator type A with blind plug M25 × 1.5) 2
Stainless steel / Cable gland M20 × 1.5 (Indicator type A with blind plug M25 × 1.5) 31
Stainless steel / Thread ½ in NPT (Indicator type A with blind plug M25 × 1.5) 41
Explosion Protection Certification and Approvals
Without Y0
ATEX / IECEx, (Zone 2 / 21), Ex nA, c, Ex tD B12
ATEX / IECEx, (Zone 1 / 21), Ex ia, Ex nA, c, Ex tD A42
ATEX / IECEx, (Zone 1 / 21), Ex d, Ex ia, Ex nA, c, Ex tD A93
FM / CSA, Class I, Div 1, 2 (Zone 1, 2), XP + IS, NI F34
FM / CSA, Class I, Div 1, 2 (Zone 1, 2), IS, NI F44
Process Connection
Thread DIN 11851 S1
SMS 1145 S2
Connection Size
DN 25 (1 in) B
DN 40 (1 ½ in) N
DN 50 (2 in) C
DN 80 (3 in) D
DN 100 (4 in) E
Float Design
Standard 1
Pressure Rating
PN 6 D0
PN 25 D35
PN 40 D46
Design Level
(Specified by ABB) X
Tube / Float Combination
(Specified by ABB.) See flow range tables. XX
1) Not available with Ex d or XP applications
2) Version with analog indicator without output signal: Type of protection "c" only available with ATEX approval
3) With Ex d approved cable gland. Not available for analog indicator without output signal
4) Only with cable entry ½ in NPT
5) For DN 50 and DN 80 (2 in and 3 in)
6) For DN 25 and DN 40 (1 in and 1-½ in)
VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E 45

Additional ordering information FAM544

VA Master FAM544 Metal Cone Variable Area Flowmeter, hygienic design, for specific XX XX XX XX XX XX XX XX
food and pharmacy applications
Primary
Without, secondary only Y0*
Standard Y1**
Measuring Deviation
Class 1.6 acc. VDE / VDI 3513 A1**
1 % of full scale AA
4 % of full scale (high viscosity, without calculation) AK
4 % of full scale (high viscosity, with calculation) AL
Material: 3.1, 3.2 Inspection / NACE
Material monitoring with inspection certificate 3.1 acc. EN 10204 C2
Material monitoring with inspection certificate 3.2 acc. EN 10204 C3
Material monitoring NACE MR 01-75 with inspection certificate 3.1 acc. EN 10204 CN
Material: 2.1 Compliance
Declaration of compliance with the order 2.1 acc. EN 10204 C4
Certificates: 3.1 Visual, Dimensional and Functional Test
Inspection certificate 3.1 acc. EN 10204 of visual, dimensional and functional test C6
Certificates: 3.1 PMI Test
Inspection certificate 3.1 acc. EN 10204 of positive material identification PMI CA
Certificates: 3.1 Pressure Test
Pressure test acc. AD2000 CB
Test package (pressure test, non-destructive test, welder and welding procedure certificate) CP
(Non-destructive test = X-ray for welding neck flanges, dye penetration test for slip-on flanges)
Certificates: Calibration, Test Report
Certificate of compliance for calibration 2.1 acc. EN 10204 CM
Inspection certificate 3.1 acc. EN 10204 of calibration, with test report CE
* For Ex devices upon request
** Standard, automatically specified by ABB if not specified by customer
46 VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E


… Ordering Information

Additional ordering information XXX XX XX XX XX XX XX


Other Explosion Protection Certifications and Approvals
INMETRO (Brazil) EB1
Documentation Language
German M1
English M5*
Language package Western Europe / Scandinavia (Languages: DE, EN, DA, ES, FR, IT, NL, PT, FI, SV) MW
Language package Eastern Europe (Languages: DE, EL, CS, ET, LV, LT, HU, HR, PL, SK, SL, RO, BG) ME
Applications
Degreased for oxygen applications P1
Device Identification Plate
Stainless steel plate with TAG no. T0
Adhesive label with TAG no. TC*
Float Material
AISI 316Ti SST (1.4571) F1*
Scale Design
Direct reading scale SD*
Percentage scale SP
Bargraph SB**
Ambient Temperature Range
Extended -40 to 70 °C (−40 to 158 °F) / -40 to 60 °C (-40 to 140 °F) with Ex protection Exd + Ex tD R5***
Standard -20 to 70 °C (-4 to 158 °F) / -20 to 60 °C (-4 to 140 °F) with Ex protection Exd + Ex tD R6*
* Standard, automatically specified by ABB if not specified by customer
** Recommended for indicator with transmitter 4 to 20 mA with LCD display
*** Not available with analog indicator with alarm

Trademarks
HART is a registered trademark of FieldComm Group, Austin, Texas,
USA
Buna-N is a registered trademark of DuPont Dow Elastomers.
™ Hastelloy C-2 is a Haynes International trademark
VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E 47

Change from one to two columns


Questionnaire

Customer: Date:

Ms./Mr.: Department:

Telephone: Fax:

Measuring medium _______________________________


liquid contaminated Yes
gaseous No
transparent Solids content Yes
non-transparent No
translucent Size n
Flow rate: min.______________________________ norm. __________ max. ______________n
l/min l/h m3/min m3/min
cm3/min kg/min g/min kg/h
other _________________________
If a vol. gas measurement refers to standard state, (Qv)s should be added to the flowrate unit.
Operating temperature: standard __________ max. ___________n
Operating pressure: Upstream pressure (P1) Downstream pressure (P1) max. _______________
______________________ _____________________

Gases: Standard density (ρn) ______________ Viscosity _________ mPa s


kg/m3
Fluid: Concentration ________________ Vol. % _________ Weight %
Density at operating temperature _________ kg/dm3
Viscosity at operating temperature _________ mPa s
Materials: Meter tube ______________________________________________________________________________________
Float ________________________________________________________________________________
Gaskets ____________________________________________________________________________________
Other wetted parts ___________________________________________________________________
Connections: Thread Flange other ____________n
Threaded spuds in accordance with DIN
11851
Installation type: Cable assembly Wall mounting

Display on device: % scale Directly readable product scale


Transmitter, electrical: 4 ... 20 mA Ex Non-Ex
Limit switch: Min. contact Max. contact Min. and max. contact
Notes:

ABB Limited ABB Inc.
Measurement & Analytics Measurement & Analytics
Howard Road, St. Neots 125 E. County Line Road
Cambridgeshire, PE19 8EU Warminster, PA 18974
UK USA
Tel: +44 (0)870 600 6122 Tel: +1 215 674 6000
Fax: +44 (0)1480 213 339 Fax: +1 215 674 7183
Email: enquiries.mp.uk@gb.abb.com

ABB Automation Products GmbH


Measurement & Analytics
Schillerstr. 72
32425 Minden
Germany
Tel: +49 571 830-0
Fax: +49 571 830-1806

abb.com/flow
DS/FAM540-EN Rev. E 07.2019


We reserve the right to make technical changes or modify the contents of this document without
prior notice. With regard to purchase orders, the agreed particulars shall prevail.
ABB does not accept any responsibility whatsoever for potential errors or possible lack of
information in this document.

We reserve all rights in this document and in the subject matter and illustrations contained therein.
Any reproduction, disclosure to third parties or utilization of its contents – in whole or in parts – is
forbidden without prior written consent of ABB.

© ABB 2019 3KXF154001R1001