Metal Cone Variable Area Flowmeter
2 VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E
—
Measurement made easy
—
For measuring liquids, gases and steam
—
Graphical 2-line LCD display
—
Magnet stick operation
• Configuration also possible with closed housing
—
Easily adjustable for new operating conditions
—
Approvals for explosion protection
• ATEX
• IECEx
• FM, CSA
VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E 3
—
Overview
Indicator / transmitter
IP degree of protection in accordance with IP 66, IP 67, NEMA 4X
EN 60529
Mechanical indicator Analog indicator with or without alarm signaling unit
Electronic indicator Analog indicator with transmitter 4 to 20 mA, with or without LCD indicator
Communication HART® protocol (only for analog indicators with transmitters)
Power supply Without alarm signalling unit: no power supply
Analog indicator with alarm signaling unit: 8 V DC via switch amplifier
Analog indicator with transmitter: 10 to 46 V DC (Ex: 10 to 30 V DC)
Indicator housing material Al Si 12 material number 3.2582 (copper content 0.1 %), stainless steel 1.4408
Paint Epoxy paint 80 to 100 μm; bottom color: RAL 7012, cover color: RAL 9002
(No housing paint for stainless steel indicator housing)
Approvals
Ex approvals ATEX / IECEx: Zone 0 / 1 / 2 / 21
FM / cCSAus: XP, IS, DIP, NI, FM Zone 1 + 2
Sealing concept Dual sealing in accordance with ANSI / ISA-12.27.01
SIL approval Analog indicator with alarm signaling unit: SIL 2
Analog indicator with transmitter: FMEDA evaluation
—
Introduction and basics
Installation conditions
• The installation recommendations of VDI / VDE Directive
3513 must be observed.
• The flowmeter is installed vertically in the piping. The
measuring media must flow from bottom to top.
• Keep the device as far away as possible from pipe
vibrations and powerful magnetic fields.
• The nominal diameter of the piping should be the same as
the connection nominal diameter.
• Inlet and outlet sections are generally not required. Care
should be taken to avoid flow turbulence, pulsations,
pressure shocks and other flow instabilities in order to
prevent measuring inaccuracies, increased wear or
damage.
• When selecting devices, pay close attention to the
chemical resistance of the wetted parts of the device and
the process connection gaskets in relation to the
measuring medium.
• Avoid pulsating flow of the measuring medium. Use the Figure 1: Installation of the flowmeter (example)
Operating conditions
A variable area flowmeter is specified for a defined set of Pressure loss
operating conditions of the measuring medium. For liquids The available operating pressure at the measuring point must
and gases, these are pressure and temperature-related be higher than the pressure loss listed for the flowmeter in
properties (density and viscosity) under operating the specifications.
conditions. It is important to also consider the pressure loss downstream
For gases, in particular, this means operating at a specific from the flowmeter due to losses in the piping and other
operating pressure and operating temperature. The specified fittings.
accuracy of the device always refers to the operating
conditions underlying the specification. Prevention of compression oscillations when measuring
gases
Temperature Data During low flow amounts and low operating pressure, so-
The following diagram shows the maximum permissible called compression oscillations of the float can occur.
measuring medium temperature depending on the ambient If the maximum upstream pressure listed in the
temperature. specifications is not reached, the flowmeter can optionally be
equipped with a gas damper.
—
… Introduction and basics
Solids content in the measuring medium Measuring range limits
Variable area flowmeters have only limited suitability for
measuring media containing solids.
Depending on the concentration, particle size and type of
solid, increased mechanical abrasion may occur, especially at
the critical measuring edge of the float.
In addition, solidified deposits on the float can change its
weight and shape.
These effects can lead to erroneous measurement results,
depending on the float type.
In general, the use of appropriate filters is recommended in
such applications.
For the flow measurement of measuring media containing
magnetic particles, we recommend the installation of a
magnetic separator upstream of the variable area flowmeter.
1 Pressure loss (dP in mbar)
2 Diameter of the float weight (mm)
3 x 1000 l/h water
Float designs
Figure 5: Flow rate depending on float shape and weight (example)
—
Indicator and transmitter
A magnet in the float translates the height of the float as a Product highlights
measurement for the flow to the decouple-proof magnet
• Decouple-proof and hysteresis-free magnet follower
follower system of the flowmeter, which is directly connected
system.
to the indicator pin.
• The device conforms to NAMUR Recommendations NE43,
For analog indicators, the flow value is shown by the indicator
NE53, NE107.
on the scale. For indicators with a two-wire transmitter, the
• The installation and disassembly of the analog indicator
indicator position is additionally measured directly at the
on the meter tube is possible without opening the
axis.
indicator housing.
A standard eddy-current brake is used to dampen indicator
• Reading repeatability: ±0.25 % of scale end value.
vibrations and ensure outstanding readability.
• The alarm signaling unit can be retrofitted as an alarm
The indicator/transmitter unit is attached to the flowmeter
module.
with 2 screws in a reproducible manner. To facilitate
• The switching points of the alarm signaling units can be
installation, the indicator can be removed. A name plate on
adjusted without removing the scale.
the flowmeter bracket allows you to assign a unique identifier
• The set switching points of the alarm signaling units are
at a later date.
visible from the exterior.
Switching amplifier
For analog indicators with alarm signaling units
(model FAM54xB/C/Dx)
Specifications
Power supply 230 V AC, +10 % / −15 %, 45 to 60 Hz
115 V AC, +10 % / −15 %, 45 to 60 Hz
—
… Indicator and transmitter
Analog indicator with transmitter
The electronic transmitter uses two-wire technology. An Product highlights
optional LCD indicator offers the possibility of adjusting on- • 4 to 20 mA current output.
site to changed measuring medium parameters. • In the event of an alarm, current output can be adjusted
For models with an LCD indicator, ABB recommends that you to 21 to 23 mA (NAMUR NE43).
use a bar graph scale to avoid differences between a product • Programmable binary output (min. / max. device alarm,
scale and the flow rate shown on the LCD indicator. general alarm or pulse output).
• Parameterization using HART communication via
handheld terminal or DSV401 (SMART VISION).
• Damping: can be adjusted 1 to 100 s (5 τ).
• Low flow cut-off: 0 to 5 % for current and pulse output.
• Measuring medium parameters can be changed at any
time (pressure and temperature influence, density, units,
etc.).
• Simulation of current and binary output (manual process
execution).
• EEPROM storage of counter readings and parameters
specific to the measuring point (over 10 years).
• LCD indicator can be retrofitted.
LCD Indicator
1 Button ◄ 4 Button ▼ • High-contrast LCD indicator.
2 Button ▲ 5 Button ►
• Flow rate and flow totalizer value display.
• Plain text fault diagnostics
3 LCD indicator 6 Points for magnet stick
operation • Menu-guided parameterization with four buttons.
• Parameterization of the device using a magnet stick
Figure 7: LCD indicator with operating buttons and magnet stick markings
without opening the housing.
VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E 9
—
Electrical connections
Note Analog indicator with transmitter
When using in potentially explosive atmospheres, observe the Model FAM54xE/Fx
electrical data in the ATEX / IECEx limit value tables starting
from Safety specifications ATEX / IECEx on page 25 and the
FM / cCSAus limit value tables starting from Safety
specifications FM, cCSAus on page 34!
Nominal voltage 8 V DC (Ri approx. 1 kΩ) B Power supply via power supply 3 Power supply unit
unit
Operating voltage 5 to 25 V DC 4 HART® modem
Switching frequency, max. 3 kHz C HART® output
1 Flowmeter
Legend
UB Operating voltage
US Input terminal voltage
RB Maximum permissible load for power supply unit (e.g. indicator)
R Maximum permissible load for output circuit; is determined by
power supply unit
Functional earth
10 VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E
—
… Electrical connections
Cable HART output
Maximum cable length 1500 m, AWG 24 twisted and shielded. Model FAM54xE/Fx
Pulse output
Terminals 41 / 42
Pulse width 5 to 256 ms, maximum 50 % of the period
Frequency fmax Maximum 50 Hz
VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E 11
—
Model FAM541 - Standard design
Temperature limits °C (°F)
Medium temperature range
−55 to 400 °C (−67 to 752 °F)
SIL approval
• SIL2 declaration of conformity for devices with alarm Process connections
function. • Flange in accordance with EN 1092-1 (PN 16, PN 40)
• Manufacturer's declaration (SIL1) in accordance with IEC • Flange in accordance with DIN 2501 (PN 63, PN 100)
61508 / IEC61511 for devices with a current output of • Flange with groove in accordance with DIN 2501
4 to 20 mA. • Flange in accordance with ASME B16.5
• Female thread G 1 in (½ in meter tube)
• Female thread G 1 ½ in (1 in meter tube)
Measuring accuracy
1.6 % qg = 50 % in accordance with VDE / VDI 3513, sheet 2,
optional 1 % of upper range value. Pressure ratings
• Standard pressure rating PN 40
• Flange in accordance with DIN / EN: PN 16, PN 40, PN 63,
Measuring range limits and span limits PN 100
Measuring span: 10:1 • Flange in accordance with ASME: CL 150, CL 300, CL 600
For detailed information about measuring ranges, see
Measuring range table on page 13. Other designs and pressure ratings are available upon
request.
See also Material load on page 12.
—
… Model FAM541 - Standard design
Materials Material load
Materials for the indicator housing
Component Material
Indicator housing Standard: aluminum, powder-coated
Option: stainless steel 1.4408
Powder coating Epoxy paint 80 to 100 μm
Bottom color: RAL 7012
Cover color: RAL 9002
Housing gasket (O-ring) Buna N®
Sight glass Shatterproof glass
Component Material
Meter tube Stainless steel 1.4404 (316 L)
Measuring element Stainless steel 1.4571 (316 Ti), only DN 15 (½)
Flange Stainless steel 1.4404 (316 L)
Variable area Standard: stainless steel 1.4571 (316 Ti)
Gas damping (option) Stainless steel 1.4571 (316 Ti)
DN1 Upper range value2 for Meter tube / VIN4 Pressure loss5 Minimum upstream pressure for gas metering6
Water Air nominal flow rate [mbar (psi)] Without gas damping With gas damping7
[l/h (USgal/h)] [m3/h (scfh)] H2O [l/h] / Order [bar abs (psia)] [bar abs (psia)]
code3
DN 15 (½ in) 28 to 32 0.83 to 0.95 ½ in / 30 / A78) 6 80 (1.16) 4.0 (58.02) 1.0 (14.5)
(7.4 to 8.5) (37 to 45)
DN 25 (1 in) 37 to 43 1.10 to 1.28 ½ in / 40 / B78) 6 80 (1.16) 4.0 (58.02) 1.0 (14.5)
G1 in (10 to 11.4) (52 to 59)
44 to 55 1.30 to 1.63 ½ in / 50 / C78) 6 80 (1.16) 4.0 (58.02) 1.0 (14.5)
(11.6 to 14.5) (54 to 66)
56 to 64 1.66 to 1.90 ½ in / 60 / D78) 6 80 (1.16) 4.0 (58.02) 1.0 (14.5)
(14.5 to 17) (68 to 80)
77 to 83 2.29 to 2.47 ½ in / 80 / E7 16 40 (0.58) 3.0 (43.51) 1.0 (14.5)
(20.5 to 21.5) (84 to 90)
96 to 104 (25.5 to 27) 2.85 to 3.09 ½ in / 100/ F7 16 45 (0.65) 3.2 (46.41) 1.0 (14.5)
(104 to 112)
115 to 125 3.42 to 3.72 ½ in / 120 / G7 16 50 (0.73) 3.5 (50.76) 1.0 (14.5)
(30 to 33) (125 to 135)
144 to 156 (38 to 41) 4.28 to 4.64 ½ in / 150 / H7 16 60 (0.87) 3.8 (55.11) 1.0 (14.5)
(155 to 165)
188 to 212 (50 to 56) 5.59 to 6.30 ½ in / 200 / J7 16 60 (0.87) 4.0 (58.02) 1.0 (14.5)
(205 to 230)
235 to 265 (62 to 70) 6.98 to 7.88 ½ in / 250 / K7 16 65 (0.94) 4.2 (60.92) 1.0 (14.5)
(255 to 285)
282 to 318 (74 to 84) 8.38 to 9.45 ½ in / 300 / L7 16 70 (1.02) 4.4 (63.82) 1.0 (14.5)
(310 to 340)
376 to 424 11.17 to 12.60 ½ in / 400 / M7 16 75 (1.09) 4.6 (66.72) 1.0 (14.5)
(100 to 112) (410 to 460)
470 to 530 13.97 to 15.75 ½ in / 500 / N7 16 75 (1.09) 4.8 (69.62) 1.0 (14.5)
(125 to 140) (510 to 570)
565 to 635 16.79 to 18.87 ½ in / 600 / P7 16 80 (1.16) 5.0 (72.52) 1.0 (14.5)
(150 to 165) (620 to 680)
750 to 850 22.29 to 25.26 ½ in / 800 / R7 16 85 (1.23) 5.4 (78.32) 1.0 (14.5)
(200 to 220) (820 to 920)
14 VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E
—
… Model FAM541 - Standard design
DN1 Upper range value2 for Meter tube / VIN4 Pressure loss5 Minimum upstream pressure for gas metering6
Water Air nominal flow rate [mbar (psi)] Without gas damping With gas damping7
[l/h (USgal/h)] [m3/h (scfh)] H2O [l/h] / Order [bar abs (psia)] [bar abs (psia)]
code3
DN 25 (1 in) 280 to 656 (74 to 170) 8.32 to 19.50 1 in / 400 / A1 13 to 21 20 to 76 2.9 to 3.1 1.1 to 1.4
G1 ½ in (310 to 700) (0.3 to 1.1) (42.1 to 45.0) (16.0 to 20.3)
393 to 870 11.70 to 25.85 1 in / 600 / B1 7 to 10 27 to 76 3.0 to 3.4 1.2 to 1.4
(104 to 220) (430 to 940) (0.4 to 1.1) (43.5 to 49.3) (17.4 to 20.3)
660 to 1600 19.38 to 50.80 1 in / 1000 / C1 16 to 22 20 to 76 3.3 to 4.3 1.1 to 1.4
(170 to 450) (720 to 1850) (0.3 to 1.1) (48.0 to 62.4) (16.0 to 20.3)
975 to 2370 28.98 to 70.44 1 in / 1600 / D1 8 to 10 27 to 82 3.3 to 5.3 1.2 to 1.5
(260 to 620) (1060 to 2550) (0.4 to 1.2) (48.0 to 77.0) (17.4 to 21.8)
1650 to 4020 49.04 to 119.50 1 in / 2500 / E1 17 to 6 20 to 76 4.2 to 6.4 1.1 to 1.4
(440 to 1060) (1800 to 4300) (0.3 to 1.1) (61.0 to 92.8) (16.0 to 20.3)
2585 to 6170 76.83 to 183.50 1 in / 4000 / F1 8 to 10 27 to 82 5.2 to 8.0 1.2 to 1.5
(680 to 1600) (2800 to 6600) (0.4 to 1.2) (75.4 to 116) (17.4 to 21.8)
DN 40 (1 ½ in) 4220 to 12130 125.40 to 360.50 2 in / 8000 / A2 21 to 38 11 to 62 3.1 to 4.5 1.1 to 1.4
(1120 to 3200) (4600 to 13000) (0.2 to 0.9) (45.0 to 65.3) (16.0 to 20.3)
DN 50 (2 in) 7940 to 18460 236.0 to 584.6 2 in / 12000 / B2 13 to 17 24 to 74 3.8 to 6.2 1.1 to 1.4
(2100 to 4800) (8600 to 20000) (0.3 to 1.1) (55.1 to 90.0) (16.0 to 20.3)
11760 to 24200 349.50 to 720.00 2 in / 18000 / C2 3 to 4 28 to 72 4.4 to 7.5 1.1 to 1.4
(3100 to 6400) (13000 to 27000) (0.4 to 1.0) (63.8 to 109) (16.0 to 20.3)
DN 80 (3 in) 7000 to 21010 208.00 to 624.40 3 in / 12000 / A3 22 to 54 6 to 48 3.4 to 5.4 1.1 to 1.3
(1850 to 5500) (7600 to 22000) (0.1 to 0.7) (49.3 to 78.3) (16.0 to 18.9)
18090 to 35010 537.70 to 1040 3 in / 25000 / B3 18 to 25 24 to 65 4.8 to 7.4 1.1 to 1.4
(4800 to 9200) (19500 to 38000) (0.3 to 1.0) (69.6 to 107) (16.0 to 20.3)
29750 to 53810 795 to 1600 3 in / 40000 / C3 4 to 5 26 to 68 6.0 to 9.2 1.1 to 1.4
(7000 to 14000) (29000 to 58000) (0.4 to 1.0) (87.0 to 133) (16.0 to 20.3)
Dimensions
—
… Model FAM541 - Standard design
Weight in kg (lb)
—
Model FAM544 - hygienic design
Temperature limits °C (°F)
Medium temperature range
−55 to 400 °C (−67 to 752 °F)
SIL approval
Nominal Measuring tube SMS 1145 DIN 11851
• SIL2 declaration of conformity for devices with alarm
diameter sizes
function.
DN 25 ½ in SC 25 —
• Manufacturer's declaration (SIL1) in accordance with
DN 40 1 in SC 40 DN 38
IEC 61508 / IEC 61511 for devices with a current output of
DN 50 2 in SC 50 DN 51
4 to 20 mA.
DN 80 3 in SC 80 DN 76
Measuring accuracy
Pressure ratings
1.6 % qg = 50 % in accordance with VDE / VDI 3513, sheet 2,
optional 1 % of upper range value. Threaded spuds in accordance with DIN 11851:
• PN 25 for DN 50 to DN 80 (2 to 3 in)
• PN 40 for DN 25 to DN 40 (1 to 1 ½ in)
Measuring range limits and span limits
Threaded spud in accordance with DIN 1145:
Measuring span: 10:1 • PN 6 for (DN 38 to DN 76)
For detailed information about measuring ranges, see
Measuring range table on page 19. See also Material load on page 18.
—
… Model FAM544 - hygienic design
Materials Material load
Materials for the indicator housing
Process Nominal diameter PSmax TSmin TSmax
connection
Component Material DIN 11851 DN 15 to DN 40 40 bar −40 °C 140 °C
Indicator housing Standard: aluminum, powder-coated (½ to 1 ½ in) (580 psi) (-40 °F) (284 °F)
Option: stainless steel 1.4408 DN 50 to 80 25 bar −40 °C 140 °C
Powder coating Epoxy paint 80 to 100 μm (2 to 3 in) (362 psi) (-40 °F) (284 °F)
Bottom color: RAL 7012 SMS 1145 DN 38 to 76 6 bar −40 °C 140 °C
Cover color: RAL 9002 (1 ½ to 3 in) (87 psi) (-40 °F) (284 °F)
Housing gasket (O-ring) Buna N®
Sight glass Shatterproof glass
Component Material
Meter tube Stainless steel 1.4404 (316 L)
Measuring element Stainless steel 1.4571 (316 Ti), DN 15 (½ in) only
Variable area Stainless steel 1.4571 (316 Ti)
VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E 19
DN1 Upper range value2 for Meter tube / nominal flow rate VIN4 Pressure loss5 Minimum upstream
Water Air H2O [l/h] / Order code3 [mbar (psi)] pressure for gas
[l/h (USgal/h)] [m3/h (scfh)] metering6
[bar abs (psia)]
DN 25 28 to 32 — ½ in / 30 / A7 6 80 (1.16) —
(1 in7) (7.4 to 8.5)
37 to 43 — ½ in / 40 / B7 6 80 (1.16) —
(10 to 11.4)
44 to.55 (11.6 to 14.5) — ½ in / 50 / C7 6 80 (1.16) —
56 to 64 — ½ in / 60 / D7 6 80 (1.16) —
(14.5 to 17)
77 to 83 (20.5 to 21.5) 2.29 to 2.47 (84 to 90) ½ in / 80 / E7 16 40 (0.58) 3.0 (43.51)
96 to 104 (25.5 to 27) 2.85 to 3.09 ½ in / 100/ F7 16 45 (0.65) 3.2 (46.41)
(104 to 112)
115 to 125 3.42 to 3.72 ½ in / 120 / G7 16 50 (0.73) 3.5 (50.76)
(30 to 33) (125 to 135)
144 to 156 4.28 to 4.64 ½ in / 150 / H7 16 60 (0.87) 3.8 (55.11)
(38 to 41) (155 to 165)
188 to 212 5.59 to 6.30 ½ in / 200 / J7 16 60 (0.87) 4.0 (58.02)
(50 to 56) (205 to 230)
235 to 265 6.98 to 7.88 ½ in / 250 / K7 16 65 (0.94) 4.2 (60.92)
(62 to 70) (255 to 285)
282 to 318 8.38 to 9.45 ½ in / 300 / L7 16 70 (1.02) 4.4 (63.82)
(74 to 84) (310 to 340)
376 to 424 11.17 to 12.60 ½ in / 400 / M7 16 75 (1.09) 4.6 (66.72)
(100 to 112) (410 to 460)
470 to 530 13.97 to 15.75 ½ in / 500 / N7 16 75 (1.09) 4.8 (69.62)
(125 to 140) (510 to 570)
565 to 635 16.79 to 18.87 ½ in / 600 / P7 16 80 (1.16) 5.0 (72.52)
(150 to 165) (620 to 680)
750 to 850 22.29 to 25.26 ½ in / 800 / R7 16 85 (1.23) 5.4 (78.32)
(200 to 220) (820 to 920)
20 VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E
—
… Model FAM544 - hygienic design
DN1 Upper range value2 for Meter tube / nominal flow rate VIN4 Pressure loss5 Minimum upstream
Water Air H2O [l/h] / Order code3 [mbar (psi)] pressure for gas
[l/h (USgal/h)] [m3/h (scfh)] metering6
[bar abs (psia)]
DN 40 280 to 656 (74 to 170) 8.32 to 19.50 1 in / 400 / A1 13 to 21 20 to 76 2.9 to 3.1
(1 ½ in) (310 to 700) (0.3 to 1.1) (42.1 to 45.0)
393 to 870 11.70 to 25.85 1 in / 600 / B1 7 to 10 27 to 76 3.0 to 3.4
(104 to 220) (430 to 940) (0.4 to 1.1) (43.5 to 49.3)
660 to 1600 19.38 to 50.80 1 in / 1000 / C1 16 to 22 20 to 76 3.3 to 4.3
(170 to 450) (720 to 1850) (0.3 to 1.1) (48.0 to 62.4)
975 to 2370 28.98 to 70.44 1 in / 1600 / D1 8 to 10 27 to 82 3.3 to 5.3
(260 to 620) (1060 to 2550) (0.4 to 1.2) (48.0 to 77.0)
1650 to 4020 49.04 to 119.50 1 in / 2500 / E1 17 to 6 20 to 76 4.2 to 6.4
(440 to 1060) (1800 to 4300) (0.3 to 1.1) (61.0 to 92.8)
2585 to 6170 76.83 to 183.50 1 in / 4000 / F1 8 to 10 27 to 82 5.2 to 8.0
(680 to 1600) (2800 to 6600) (0.4 to 1.2) (75.4 to 116)
DN 50 4220 to 12130 125.40 to 360.50 2 in / 8000 / A2 21 to 38 11 to 62 3.1 to 4.5
(2 in) (1120 to 3200) (4600 to 13000) (0.2 to 0.9) (45.0 to 65.3)
7940 to 18460 236.0 to 584.6 2 in / 12000 / B2 13 to 17 24 to 74 3.8 to 6.2
(2100 to 4800) (8600 to 20000) (0.3 to 1.1) (55.1 to 90.0)
11760 to 24200 349.50 to 720.00 2 in / 18000 / C2 3 to 4 28 to 72 4.4 to 7.5
(3100 to 6400) (13000 to 27000) (0.4 to 1.0) (63.8 to 109)
DN 80 7000 to 21010 208.00 to 620 3 in / 12000 / A3 22 to 54 6 to 48 3.4 to 5.4
(3 in) (1850 to 5500) (7600 to 22000) (0.1 to 0.7) (49.3 to 78.3)
18090 to 35010 537.70 to 1040 3 in / 25000 / B3 18 to 25 24 to 65 4.8 to 7.4
(4800 to 9200) (19500 to 38000) (0.3 to 1.0) (69.6 to 107)
29750 to 53810 795 to 1550 3 in / 40000 / C3 4 to 5 26 to 68 6.0 to 9.2
(7000 to 14000) (29000 to 58000) (0.4 to 1.0) (87.0 to 133)
Dimensions
Dimensions for devices with threaded spuds in accordance with DIN 11851
—
… Model FAM544 - hygienic design
Dimensions for devices with threaded spuds in accordance with SMS 1145
Meter size DN [mm] Pressure rating ØD A [mm (in)] B [mm (in)] C [mm (in)] G [mm (in)]
1 in 38 PN 6 Rd 60 × ¹⁄₆ in 300.0 (11.81) 66.5 (2.62) 87.0 (3.43) 118.0 (4.65)
2 in 51 Rd 70 × ¹⁄₆ in 102.0 (4.02) 130.0 (5.12)
3 in 76 Rd 98 × ¹⁄₆ in 132.0 (5.20) 144.0 (5.67)
Weight in kg (lb)
—
Use in potentially explosive atmospheres in accordance with ATEX and IECEx
Note
Further information on the approval of devices for use in potentially explosive atmospheres can be found in the type
examination certificates or the relevant certificates at www.abb.com/flow.
Device overview
The devices are designed for maximum versatility. This is achieved through a combination of several types of protection within
each device. All devices are suitable for use in potentially explosive atmospheres with combustible dust. For detailed
installation instructions and terminal assignments, refer to Electrical connections on page 9.
Potential equalization
* Order code ‘Explosion protection and approvals’ (versions 9, 10); refer to Ordering Information on page 41.
24 VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E
—
… Use in potentially explosive atmospheres in accordance with ATEX and IECEx
Ex marking
Note
• Depending on the design, a specific marking in accordance with ATEX or IECEx applies.
• ABB reserves the right to modify the Ex-marking. Refer to the name plate for the exact marking.
Model FAM54xE/Fx (analog indicator with transmitter and with or without LCD indicator)
Marking Type of protection Order code* Limit value table
ATEX II 1/2 G Ex h ia IIC T4 ... T1 Ga/Gb Intrinsically Safe A4 Table 1 on page 26
II 1/3 G Ex ec h ic T6 ... T1Ga/Gc Non-sparking materials
II 2 D Ex h tb IIIC T85°C ... Tmedium Db Dust explosion protection
IECEx Ex h ia IIC T4 ... T1 Ga/Gb Intrinsically Safe
Ex ec h ic IIC T6 ... T1Ga/Gc Non-sparking materials
Ex h tb IIIC T85°C ... Tmedium Db Dust explosion protection
ATEX II 1/2 G Ex db h IIC T6 ... T1 Ga/Gb Flameproof / Explosion-Proof A9 Table 1 on page 26
II 1/2 G Ex h ia IIC T4 ... T1 Ga/Gb Intrinsically Safe
II 1/3 G Ex ec h ic IIC T6 ... T1 Ga/Gc Non-sparking materials
II 2 D Ex h tb IIIC T85°C ... Tmedium Db Dust explosion protection
IECEx Ex db h IIC T6 ... T1 Ga/Gb Flameproof / Explosion-Proof
Ex h ia IIC T4 ... T1 Ga/Gb Intrinsically Safe
Ex ec h ic IIC T6 ... T1 Ga/Gc Non-sparking materials
Ex h tb IIIC T85°C ... Tmedium Db Dust explosion protection
ATEX II 1/3 G Ex ec h ic IIC T6 ... T1 Ga/Gc Non-sparking materials B1 Table 4 on page 29
II 2 D Ex h tb IIIC T85°C ... Tmedium Db Dust explosion protection
IECEx Ex ec h ic IIC T6 ... T1 Ga/Gc Non-sparking materials
Ex h tb IIIC T85°C ... Tmedium Db Dust explosion protection
* Order code ‘Explosion protection and approvals’ (versions 9, 10); refer to Ordering Information on page 41.
Model FAM54xB/C/D/E/Fx
The following differing temperature data applies to usage in areas with combustible dust.
Order code Ambient temperature Tamb Maximum permissible measuring medium temperature Tmedium
A4, A9, B1 −50 °C to +60 °C 250 °C
−50 °C to +40 °C 340 °C
−50 °C to +20 °C 430 °C
26 VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E
—
… Use in potentially explosive atmospheres in accordance with ATEX and IECEx
Table 1: Analog indicator with transmitter, with / without LCD indicator
Model FAM54xE/Fx
Type of protection: flameproof enclosure, intrinsic safety, non-sparking equipment, dust explosion protection.
Order Marking Terminals Electrical values Tamb Temp. class Tmedium Insulation
code* −20 °C to (−50 °C to ) Maximum
A4, A9 ATEX: 31 / 32** Ui = 30 V 40 °C T1 440 °C No
II 1/2 G Ex h ia IIC T4 ... T1 Ga/Gb Ii = 110 mA 40 °C T1 375 °C yes
II 1/3 G Ex ec h ic T6 ... T1Ga/Gc Pi = 770 mW 40 °C T1 260 °C yes
II 2 D Ex h tb IIIC T85°C ... Tmedium Db Ci = 5.3 nF 50 °C T1 300 °C yes
Li = 266 µH 50 °C T2 290 °C yes
IECEx: 50 °C T2 220°C yes
Ex h ia IIC T4 ... T1 Ga/Gb 41 / 42** Ui = 30 V 60 °C T2 320 °C No
Ex ec h ic IIC T6 ... T1Ga/Gc Ii = 30 mA 60 °C T2 230 °C yes
Ex h tb IIIC T85°C ... Tmedium Db Pi = 115 mW 60 °C T3 170 °C yes
Ci = 4.8 nF 70 °C T3 195 °C No
Li = 133 µH 70 °C T3 150 °C yes
70 °C T4 125 °C yes
A9 ATEX: 31 / 32*** Umax = 46 V 40 °C T1 440 °C No
II 1/2 G Ex db h IIC T6 ... T1 Ga/Gb 40 °C T1 375 °C yes
II 1/2 G Ex h ia IIC T4 ... T1 Ga/Gb 40 °C T1 260 °C yes
II 1/3 G Ex ec h ic IIC T6 ... T1 Ga/Gc 50 °C T1 300 °C yes
II 2 D Ex h tb IIIC T85°C ... Tmedium Db 50 °C T2 290 °C yes
50 °C T2 220°C yes
IECEx: 41 / 42*** Umax = 30 V 60 °C T2 320 °C No
Ex db h IIC T6 ... T1 Ga/Gb Imax = 30 mA 60 °C T2 230 °C yes
Ex h ia IIC T4 ... T1 Ga/Gb Pmax = 115 mW 60 °C T3 170 °C yes
Ex ec h ic IIC T6 ... T1 Ga/Gc
60 °C T4 130 °C yes
Ex h tb IIIC T85°C ... Tmedium Db
60 °C T5 95 °C yes
60 °C T6 80 °C yes
A4, A9, B1 ATEX: 31 / 32*** Umax = 46 V 40 °C T1 440 °C No
II 1/3 G Ex ec h ic IIC T6 ... T1 Ga/Gc 40 °C T1 375 °C yes
II 2 D Ex h tb IIIC T85°C ... Tmedium Db 40 °C T1 260 °C yes
50 °C T1 300 °C yes
IECEx: 50 °C T2 290 °C yes
Ex ec h ic IIC T6 ... T1 Ga/Gc 50 °C T2 220°C yes
Ex h tb IIIC T85°C ... Tmedium Db 41 / 42*** Umax = 30 V 60 °C T2 320 °C No
Imax = 30 mA 60 °C T2 230 °C yes
Pmax = 115 mW 60 °C T3 170 °C yes
70 °C T3 195 °C No
70 °C T3 150 °C yes
70 °C T4 130 °C yes
* Order code ‘Explosion protection and approvals’ (versions 9, 10); refer to Ordering Information on page 41.
** for connection to an intrinsically safe circuit
*** for connection to a non-intrinsically safe circuit If the device is to be operated subsequently with type of protection "intrinsically safe", then Umax = 60
V may not be exceeded.
VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E 27
Model FAM54xB/C/Dx
Type of protection: intrinsic safety, dust explosion protection.
Order Marking Terminals Electrical values Tamb Temp. class Tmedium Insulation
code* −20 °C to (−50 °C to ) Maximum
A4, A9 ATEX: 41 / 42** Ui = 16 V 40 °C T1 440 °C No
II 1/2 G Ex h ia IIC T6 ... T1 Ga/Gb 51 / 52** Ii = 25 mA 40 °C T1 375 °C yes
II 2D Ex h tb IIIC T85°C ... Tmedium Db Pi = 64 mW 40 °C T1 260 °C yes
Ci = 50 nF 50 °C T1 300 °C yes
IECEx: Li = 250 µH 50 °C T2 290 °C yes
Ex h ia IIC T6 ... T1 Ga/Gb 50 °C T2 220°C yes
Ex h tb IIIC T85°C ... Tmedium Db 60 °C T2 320 °C No
60 °C T2 230 °C yes
60 °C T3 170 °C yes
70 °C T3 195 °C No
70 °C T3 150 °C yes
70 °C T4 130 °C yes
70 °C T5 95 °C yes
60 °C T6 80 °C yes
41 / 42** Ui = 16 V 40 °C T1 440 °C No
51 / 52** Ii = 52 mA 40 °C T1 375 °C yes
Pi = 169 mW 40 °C T1 260 °C yes
Ci = 50 nF 50 °C T1 300 °C yes
Li = 250 µH 50 °C T2 290 °C yes
50 °C T2 220°C yes
60 °C T2 320 °C No
60 °C T2 230 °C yes
60 °C T3 170 °C yes
70 °C T3 195 °C No
70 °C T3 150 °C yes
70 °C T4 130 °C yes
60 °C T5 60 °C yes
50 °C T5 90 °C No
40 °C T6 60 °C yes
41 / 42** Ui = 16 V 40 °C T1 440 °C No
51 / 52** Ii = 76 mA 40 °C T1 310 °C yes
Pi = 242 mW 40 °C T2 190 °C yes
Ci = 50 nF 50 °C T2 340 °C No
Li = 250 µH 50 °C T2 230 °C yes
60 °C T2 230°C No
60 °C T3 160 °C yes
70 °C T4 120 °C No
70 °C T4 100 °C yes
40 °C T5 60 °C yes
30 °C T6 30 °C yes
* Order code ‘Explosion protection and approvals’ (versions 9, 10); refer to Ordering Information on page 41.
** for connection to an intrinsically safe circuit
28 VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E
—
… Use in potentially explosive atmospheres in accordance with ATEX and IECEx
Table 3: Analoganzeiger mit Grenzwertgeber
Model FAM54xB/C/Dx
Type of protection: flameproof enclosure, dust explosion protection.
Order Marking Terminals Electrical values Tamb Temp. class Tmedium Insulation
code* −20 °C to (−50 °C to ) Maximum
A9 ATEX: 41 / 42** Umax = 16 V 40 °C T1 440 °C No
II 1/2 G Ex db h IIC T6 ... T1 Ga/Gb 51 / 52** Imax = 25 mA 40 °C T1 375 °C yes
II 2 D Ex h tb IIIC T85°C ... Tmedium Db Pmax = 64 mW 40 °C T1 260 °C yes
50 °C T1 300 °C yes
IECEx: 50 °C T2 290 °C yes
Ex db h IIC T6 ... T1 Ga/Gb 50 °C T2 220°C yes
Ex h tb IIIC T85°C ... Tmedium Db 60 °C T2 320 °C No
60 °C T2 230 °C yes
60 °C T3 170 °C yes
70 °C T3 195 °C No
70 °C T3 150 °C yes
70 °C T4 130 °C yes
70 °C T5 95 °C yes
60 °C T6 80 °C yes
41 / 42** Umax = 16 V 40 °C T1 440 °C No
51 / 52** Imax = 52 mA 40 °C T1 375 °C yes
Pmax = 169 mW 40 °C T1 260 °C yes
50 °C T1 300 °C yes
50 °C T2 290 °C yes
50 °C T2 220°C yes
60 °C T2 320 °C No
60 °C T2 230 °C yes
60 °C T3 170 °C yes
70 °C T3 195 °C No
70 °C T3 150 °C yes
70 °C T4 130 °C yes
60 °C T5 60 °C yes
50 °C T5 90 °C No
40 °C T6 60 °C yes
41 / 42** Umax = 16 V 40 °C T1 440 °C No
51 / 52** Imax = 76 mA 40 °C T1 310 °C yes
Pmax = 242 mW 40 °C T2 190 °C yes
50 °C T2 340 °C No
50 °C T2 230 °C yes
60 °C T2 230°C No
60 °C T3 160 °C yes
70 °C T4 120 °C No
70 °C T4 100 °C yes
40 °C T5 60 °C yes
30 °C T6 30 °C yes
* Order code ‘Explosion protection and approvals’ (versions 9, 10); refer to Ordering Information on page 41.
** for connection to a non-intrinsically safe circuit If the device is to be operated subsequently with type of protection "intrinsically safe", then Umax = 60
V may not be exceeded.
VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E 29
Model FAM54xB/C/Dx
Type of protection: non-sparking materials, dust explosion protection.
Order Marking Terminals Electrical values Tamb Temp. class Tmedium Insulation
code* −20 °C to (−50 °C to ) Maximum
A4, A9, B1 ATEX: 41 / 42** Umax = 16 V 40 °C T1 440 °C No
II 1/3 G Ex ec h IIC T6 ... T1 Ga/Gc 51 / 52** Imax = 25 mA 40 °C T1 375 °C yes
II 2D Ex h tb IIIC T85°C ... Tmedium Db Pmax = 64 mW 40 °C T1 260 °C yes
50 °C T1 300 °C yes
IECEx: 50 °C T2 290 °C yes
Ex ec h IIC T6 ... T1 Ga/Gc 50 °C T2 220°C yes
Ex h tb IIIC T85°C ... Tmedium Db 60 °C T2 320 °C No
60 °C T2 230 °C yes
60 °C T3 170 °C yes
70 °C T3 195 °C No
70 °C T3 150 °C yes
70 °C T4 130 °C yes
70 °C T5 95 °C yes
60 °C T6 80 °C yes
41 / 42** Umax = 16 V 40 °C T1 440 °C No
51 / 52** Imax = 52 mA 40 °C T1 375 °C yes
Pmax = 169 mW 40 °C T1 260 °C yes
50 °C T1 300 °C yes
50 °C T2 290 °C yes
50 °C T2 220°C yes
60 °C T2 320 °C No
60 °C T2 230 °C yes
60 °C T3 170 °C yes
70 °C T3 195 °C No
70 °C T3 150 °C yes
70 °C T4 130 °C yes
60 °C T5 60 °C yes
50 °C T5 90 °C No
40 °C T6 60 °C yes
A4, A9, B1 ATEX: 41 / 42** Umax= 16 V 40 °C T1 440 °C No
II 1/3 G Ex ec h IIC T6 ... T1 Ga/Gc 51 / 52** Imax = 76 mA 40 °C T1 310 °C yes
II 2 D Ex h tb IIIC T85°C ... Tmedium Db Pmax = 242 mW 40 °C T2 190 °C yes
50 °C T2 340 °C No
IECEx: 50 °C T2 230 °C yes
Ex ec h IIC T6 ... T1 Ga/Gc 60 °C T2 230°C No
Ex h tb IIIC T85°C ... Tmedium Db 60 °C T3 160 °C yes
70 °C T4 120 °C No
70 °C T4 100 °C yes
40 °C T5 60 °C yes
30 °C T6 30 °C yes
* Order code ‘Explosion protection and approvals’ (versions 9, 10); refer to Ordering Information on page 41.
** for connection to a non-intrinsically safe circuit If the device is to be operated subsequently with type of protection "intrinsically safe", then Umax = 60
V may not be exceeded.
30 VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E
—
… Use in potentially explosive atmospheres in accordance with ATEX and IECEx
Table 5: Analoganzeiger ohne Grenzsignalgeber
Model FAM54xAx
Type of protection: constructional safety, dust explosion protection
Order Marking Terminals Electrical values Tamb Temp. class Tmedium Insulation
code* −20 °C to (−50 °C to ) Maximum
A4, A9, B1 ATEX: n.a. n. a 70 °C T1 440 °C yes
II 1/2 G Ex h T6 ... T1 Gb 70 °C T2 290 °C yes
II 2 D Ex h T85°C ... Tmedium Db 70 °C T3 190 °C yes
II 2 D Ex h tb IIIC T85°C ... Tmedium Db 70 °C T4 130 °C yes
IECEx: 70 °C T5 95 °C yes
Ex h tb IIIC T85°C ... Tmedium Db 70 °C T6 80°C yes
* Order code ‘Explosion protection and approvals’ (versions 9, 10); refer to Ordering Information on page 41.
Change from one to two columns
Electrical connections
Potentially explosive atmosphere Non-hazardous area
UV Power supply
Terminal Function
31 / 32 Power supply / current output / HART output
41 / 42 Programmable binary output
Alarm signaling unit (min.)
51 / 52 Alarm signaling unit (max.)
VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E 31
—
Use in potentially explosive atmospheres in accordance with FM and cCSAus
Note
Further information on the approval of devices for use in potentially explosive atmospheres can be found in the type
examination certificates or the relevant certificates at www.abb.com/flow.
Device overview
The devices are designed for maximum versatility. This is achieved through a combination of several types of protection within
each device. All devices are suitable for use in potentially explosive atmospheres with combustible dust. For detailed
installation instructions and terminal assignments, refer to Electrical connections on page 9.
Class 1 Division 1 / Zone 1 Class 1 Division 2 / Zone 2 Standard / No explosion protection Order code*
Potential equalization
* Order code ‘Explosion protection and approvals’ (versions 9, 10); refer to Ordering Information on page 41.
VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E 33
Ex marking
Note
• Depending on the design, a specific marking in accordance with FM applies.
• ABB reserves the right to modify the Ex-marking. Refer to the name plate for the exact marking.
Model FAM54xAx (analog indicator without alarm signaling unit and model FAM54xB/C/Dx (analog indicator with alarm signaling unit)
Marking Type of protection Order code* Limit value table
FM XP / CL I / DIV 1 / GP ABCD / T6 … T1 Explosionproof F3 Table 2 on page 35
CL I, ZN 1 AEx d IIC T6 … T1
IS / CL I,II,III / DIV 1 / GP ABCDEFG / T6 … T1 Intrinsic Safety F3, F4 Table 1 on page 34, Table 2 on page 35
CL I, ZN 1 AEx ia IIC T6 … T1
DIP / CL II, III / DIV 1 / GP EFG / T6 … T1 Dust-Ignitionproof
NI /CL I,II / DIV 2 / GP ABCDFG / T5 … T1 Non-Incendive F3, F4 Table 1 on page 34, Table 2 on page 35,
NI / CL III T5 … T1 Table 3 on page 36
CL II, ZN 2 AEx ec II T5 … T1
cCSAus XP / CL I / DIV 1 / GP BCD / T6 … T1 Explosionproof F3 Table 2 on page 35
Ex d IIC T6 … T1
IS / CL I,II,III / DIV 1 / GP ABCDEFG / T6 … T1 Intrinsic Safety F3, F4 Table 1 on page 34,
Ex ia IIC T6 … T1
DIP / CL II, III / DIV 1 / GP EFG / T6 … T1 Dust-Ignitionproof F3, F4 Table 1 on page 34, Table 2 on page 35,
DIP A21 TA 85°C … Tmedium Table 3 on page 36
NI /CL I,II / DIV 2 / GP ABCDFG / T5 … T1 Non-Incendive F3, F4 Table 3 on page 36
NI / CL III T5 … T1
Ex ec II T5 to T1
* Order code ‘Explosion protection and approvals’ (version digit no. 9, 10); see Ordering Information on page 41.
Model FAM54xE/Fx (analog indicator with transmitter / with or without LCD display)
Marking Type of protection Order code* Limit value table
FM XP / CL I / DIV 1 / GP ABCD / T6 … T1 Explosionproof F3 Table 4 on page 37
CL I, ZN 1 AEx d IIC T6 … T1
IS / CL I,II,III / DIV 1 / GP ABCDEFG / T4 … T1 Intrinsic Safety F3, F4
CL I, ZN 1 AEx ia IIC T4 … T1
DIP / CL II, III / DIV 1 / GP EFG / T6 … T1 Dust-Ignitionproof
NI /CL I,II / DIV 2 / GP ABCDFG / T4 … T1 Non-Incendive
NI / CL III T4 … T1
CL II, ZN 2 AEx ec IIC T6 … T1
cCSAus XP / CL I / DIV 1 / GP BCD / T6 … T1 Explosionproof F3
Ex d IIC T6 … T1
IS / CL I,II,III / DIV 1 / GP ABCDEFG / T4 … T1 Intrinsic Safety F3, F4
Ex ia IIC T4 … T1
DIP / CL II, III / DIV 1 / GP EFG / T6 … T1 Dust-Ignitionproof
DIP A21 TA 85°C to Tmedium
NI /CL I,II / DIV 2 / GP ABCDFG / T4 … T1 Non-Incendive
NI / CL III T4 … T1
Ex ec IIC T6 … T1
* Order code ‘Explosion protection and approvals’ (version digit no. 9, 10), see Ordering Information on page 41.
34 VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E
—
… Use in potentially explosive atmospheres in accordance with FM and cCSAus
Safety specifications FM, cCSAus
Table 1: Analog indicator with alarm signaling unit (temperature data for FM in °F, for cCSAus in °C)
Order Marking Terminals Electrical Tamb Tamb Temp. class Tmedium Insulation
code* values −58 °F to −50 °C to Maximum
F3, F4 FM 41 / 42*** Ui = 16 V 104 °F 40 °C T1 824 °F 440 °C No
IS / CL I,II,III / DIV 1 / GP ABCDEFG / 51 / 52*** Ii = 25 mA 104 °F 40 °C T1 707 °F 375 °C yes
T6 ... T1** Pi = 64 mW 104 °F 40 °C T1 500 °F 260 °C yes
Ci = 50 nF 122 °F 50 °C T1 572 °F 300 °C yes
DIP / CL II, III / DIV 1 / GP EFG / T6 ... T1 Li = 250 µH 122 °F 50 °C T2 554 °F 290 °C yes
122 °F 50 °C T2 428 °F 220°C yes
CL I, ZN 1 AEx ia IIC T6 ... T1 140 °F 60 °C T2 608 °F 320 °C No
140 °F 60 °C T2 446 °F 230 °C yes
cCSAus
140 °F 60 °C T3 338 °F 170 °C yes
IS / CL I,II,III / DIV 1 / GP ABCDEFG /
158 °F 70 °C T3 383 °F 195 °C No
T6 ... T1**
158 °F 70 °C T3 302 °F 150 °C yes
158 °F 70 °C T4 266 °F 130 °C yes
DIP / CL II, III / DIV 1 / GP EFG / T6 ... T1
158 °F 70 °C T5 203 °F 95 °C yes
140 °F 60 °C T6 176 °F 80 °C yes
Ex ia IIC T6 ... T1
41 / 42*** Ui = 16 V 104 °F 40 °C T1 824 °F 440 °C No
51 / 52*** Ii = 52 mA 104 °F 40 °C T1 707 °F 375 °C yes
DIP A21 TA 85°C to Tmedium
Pi = 169 mW 104 °F 40 °C T1 500 °F 260 °C yes
Ci = 50 nF 122 °F 50 °C T1 572 °F 300 °C yes
Li = 250 µH 122 °F 50 °C T2 554 °F 290 °C yes
122 °F 50 °C T2 428°F 220°C yes
140 °F 60 °C T2 608 °F 320 °C No
140 °F 60 °C T2 446 °F 230 °C yes
140 °F 60 °C T3 338 °F 170 °C yes
158 °F 70 °C T3 383 °F 195 °C No
158 °F 70 °C T3 302 °F 150 °C yes
158 °F 70 °C T4 266 °F 130 °C yes
140 °F 60 °C T5 140 °F 60 °C yes
122 °F 50 °C T5 194 °F 90 °C No
104 °F 40 °C T6 140 °F 60 °C yes
41 / 42*** Ui = 16 V 104 °F 40 °C T1 824 °F 440°C No
51 / 52*** Ii = 76 mA 104 °F 40 °C T1 590 °F 310 °C yes
Pi = 242 mW 104 °F 40 °C T2 374 °F 190 °C yes
Ci = 50 nF 122 °F 50 °C T2 644 °F 340 °C No
Li = 250 µH 122 °F 50 °C T2 446 °F 230 °C yes
140 °F 60 °C T2 446 °F 230 °C No
140 °F 60 °C T3 320 °F 160 °C yes
158 °F 70 °C T4 248 °F 120 °C No
158 °F 70 °C T4 212 °F 100 °C yes
104 °F 40 °C T5 140 °F 60 °C yes
86 °F 30 °C T6 86 °F 30 °C yes
* Order code ‘Explosion protection and approvals’ (versions 9, 10); refer to Ordering Information on page 41.
** ‘IS’ installation in accordance with Installation Drawing SDM-10-A0253.
*** for connection to an intrinsically safe circuit
VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E 35
Table 2: Analog indicator with alarm signaling unit (temperature data for FM in °F, for cCSAus in °C)
Order Marking Terminals Electrical values Tamb Tamb Temp. class Tmedium Insulation
code* −58 °F to −50 °C to Maximum
F3 FM 41 / 42** Umax = 16 V 104 °F 40 °C T1 824 °F 440 °C No
XP / CL I / DIV 1 / GP ABCD / T6 ... T1 51 / 52** Imax = 25 mA 104 °F 40 °C T1 707 °F 375 °C yes
Pmax = 64 mW 104 °F 40 °C T1 500 °F 260 °C yes
DIP / CL II, III / DIV 1 / GP EFG / 122 °F 50 °C T1 572 °F 300 °C yes
T6 ... T1 122 °F 50 °C T2 554 °F 290 °C yes
122 °F 50 °C T2 428 °F 220°C yes
CL I, ZN 1 AEx d IIC T6 ... T1 140 °F 60 °C T2 608 °F 320 °C No
140 °F 60 °C T2 446 °F 230 °C yes
cCSAus
140 °F 60 °C T3 338 °F 170 °C yes
XP / CL I / DIV 1 / GP BCD / T6 ... T1
158 °F 70 °C T3 383 °F 195 °C No
158 °F 70 °C T3 302 °F 150 °C yes
DIP / CL II, III / DIV 1 / GP EFG /
158 °F 70 °C T4 266 °F 130 °C yes
T6 ... T1
158 °F 70 °C T5 203 °F 95 °C yes
140 °F 60 °C T6 176 °F 80 °C yes
Ex d IIC T6 ... T1
41 / 42** Umax = 16 V 104 °F 40 °C T1 824 °F 440 °C No
51 / 52** Imax = 52 mA 104 °F 40 °C T1 707 °F 375 °C yes
DIP A21 TA 85°C ... Tmedium
Pmax = 169 mW 104 °F 40 °C T1 500 °F 260 °C yes
122 °F 50 °C T1 572 °F 300 °C yes
122 °F 50 °C T2 554 °F 290 °C yes
—
… Use in potentially explosive atmospheres in accordance with FM and cCSAus
Table 3: Analog indicator with / without alarm signaling unit (temperature data for FM in °F, for cCSAus in °C)
Order Marking Terminals Electrical values Tamb Tamb Temp. class Tmedium Insulation
code* −58 °F to −50 °C to Maximum
F3**, F4** FM 41 / 42*** Umax = 16 V 104 °F 40 °C T1 824 °F 440 °C No
NI /CL I,II / DIV 2 / GP ABCDFG / 51 / 52*** Imax = 25 mA 104 °F 40 °C T1 707 °F 375 °C yes
T5 ... T1 Pmax = 64 mW 104 °F 40 °C T1 500 °F 260 °C yes
122 °F 50 °C T1 572 °F 300 °C yes
NI / CL III / T5 ... T1 122 °F 50 °C T2 554 °F 290 °C yes
122 °F 50 °C T2 428°F 220°C yes
CL II, ZN 2 AEx ec II T5 ... T1 140 °F 60 °C T2 608 °F 320 °C No
140 °F 60 °C T2 446 °F 230 °C yes
cCSAus
140 °F 60 °C T3 338 °F 170 °C yes
NI /CL I,II / DIV 2 / GP ABCDFG /
158 °F 70 °C T3 383 °F 195 °C No
T5 ... T1
158 °F 70 °C T3 302 °F 150 °C yes
158 °F 70 °C T4 266 °F 130 °C yes
NI / CL III / T5 ... T1
158 °F 70 °C T5 203 °F 95 °C yes
41 / 42*** Umax = 16 V 104 °F 40 °C T1 824 °F 440 °C No
Ex ec II T6 ... T1
51 / 52*** Imax = 52 mA 104 °F 40 °C T1 707 °F 375 °C yes
Pmax = 169 mW 104 °F 40 °C T1 500 °F 260 °C yes
DIP A21 TA 85°C to Tmedium
122 °F 50 °C T1 572 °F 300 °C yes
122 °F 50 °C T2 554 °F 290 °C yes
122 °F 50 °C T2 428°F 220°C yes
140 °F 60 °C T2 608 °F 320 °C No
140 °F 60 °C T2 446 °F 230 °C yes
140 °F 60 °C T3 338 °F 170 °C yes
158 °F 70 °C T3 383 °F 195 °C No
158 °F 70 °C T3 302 °F 150 °C yes
158 °F 70 °C T4 266 °F 130 °C yes
140 °F 60 °C T5 140 °F 60 °C yes
122 °F 50 °C T5 194 °F 90 °C No
41 / 42*** Umax= 16 V 104 °F 40 °C T1 824 °F 440°C No
51 / 52*** Imax = 76 mA 104 °F 40 °C T1 590 °F 310 °C yes
Pmax = 242 mW 104 °F 40 °C T2 374 °F 190 °C yes
122 °F 50 °C T2 644 °F 340 °C No
122 °F 50 °C T2 446 °F 230 °C yes
140 °F 60 °C T2 446°F 230 °C No
140 °F 60 °C T3 320 °F 160 °C yes
158 °F 70 °C T4 248 °F 120 °C No
158 °F 70 °C T4 212 °F 100 °C yes
104 °F 40 °C T5 140 °F 60 °C yes
* Order code ‘Explosion protection and approvals’ (versions 9, 10); refer to Ordering Information on page 41.
** For connection in Division 2 or Zone 2.
*** for connection to a non-intrinsically safe circuit
VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E 37
Table 4: Analog indicator with transmitter, with or without LCD indicator (temperature data for FM in °F, for cCSAus in °C)
Order Marking Terminals Electrical values Tamb Tamb Temp. Tmedium Insulation
code1 −58 °F to −50 °C to class Maximum
F3, F4 FM 31 / 322 3 Ui = 30 V 104 °F 40 °C T1 824 °F 440 °C No
IS / CL I,II,III / DIV 1 / GP ABCDEFG / Ii = 110 mA 104 °F 40 °C T1 707 °F 375 °C yes
T4 ... T1 Pi = 770 mW 104 °F 40 °C T1 500 °F 260 °C yes
DIP / CL II, III / DIV 1 / GP EFG / Ci = 5.3 nF 122 °F 50 °C T1 572 °F 300 °C yes
T6 ... T1 Li = 266 µH 122 °F 50 °C T2 554 °F 290 °C yes
CL I, ZN 1 AEx ia IIC T4 ... T1 122 °F 50 °C T2 428°F 220°C yes
cCSAus 41 / 422 Ui = 30 V 140 °F 60 °C T2 608 °F 320 °C No
IS / CL I,II,III / DIV 1 / GP ABCDEFG / Ii = 30 mA 140 °F 60 °C T2 446 °F 230 °C yes
T4 ... T1 2 Pi = 115 mW 140 °F 60 °C T3 338 °F 170 °C yes
DIP / CL II, III / DIV 1 / GP EFG / Ci = 4.8 nF 158 °F 70 °C T3 383 °F 195 °C No
T6 ... T1 Li = 133 µH 158 °F 70 °C T3 302 °F 150 °C yes
Ex ia IIC T6 ... T1
158 °F 70 °C T4 257 °F 125 °C yes
DIP A21 TA 85°C to Tmedium
F3 FM 31 / 324 Umax = 46 V 104 °F 40 °C T1 824 °F 440 °C No
XP / CL I / DIV 1 / GP ABCD / T6 ... T1 104 °F 40 °C T1 707 °F 375 °C yes
DIP / CL II, III / DIV 1 / GP EFG / 104 °F 40 °C T1 500 °F 260 °C yes
T6 ... T1 122 °F 50 °C T1 572 °F 300 °C yes
CL I, ZN 1 AEx d IIC T6 ... T1 122 °F 50 °C T2 554 °F 290 °C yes
cCSAus 122 °F 50 °C T2 428°F 220 °C yes
XP / CL I / DIV 1 / GP BCD / T6 ... T1 41 / 424 Umax = 30 V 140 °F 60 °C T2 608 °F 320 °C No
DIP / CL II, III / DIV 1 / GP EFG / Imax = 30 mA 140 °F 60 °C T2 446 °F 230°C yes
T6 ... T1 Pmax = 115 mW 140 °F 60 °C T3 338 °F 170 °C yes
Ex d IIC T6 ... T1
140 °F 60 °C T4 266 °F 130 °C yes
DIP A21 TA 85°C to Tmedium
140 °F 60 °C T5 203 °F 95 °C yes
140 °F 60 °C T6 176 °F 80 °C yes
1) Order code ‘Explosion protection and approvals’ (versions 9, 10); refer to Ordering Information on page 41.
2) For connection to an intrinsically safe circuit.
3) Installation in accordance with Installation Drawing SDM-10-A0253.
4) For connection to a non-intrinsically safe circuit.
38 VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E
—
… Use in potentially explosive atmospheres in accordance with FM and cCSAus
Continuation of Table 4: Analog indicator with transmitter, with or without LCD indicator (temperature data for FM in °F, for cCSAus in °C)
Order Marking Terminals Electrical values Tamb Tamb Temp. Tmedium Insulation
code1 −58 °F to −50 °C to class Maximum
F35, F45 FM 31 / 324 Umax = 46 V 104 °F 104 °F T1 824 °F 440 °C No
NI /CL I,II / DIV 2 / GP ABCDFG / 104 °F 104 °F T1 707 °F 375 °C yes
T4...T1 104 °F 104 °F T1 500 °F 260 °C yes
NI / CL III / T4...T1 122 °F 122 °F T1 572 °F 300 °C yes
CL II, ZN 2 AEx ec IIC T4...T1 122 °F 122 °F T2 554 °F 290 °C yes
122 °F 122 °F T2 428°F 220 °C yes
cCSAus 140 °F 140 °F T2 608 °F 320 °C No
NI /CL I,II / DIV 2 / GP ABCDFG /
41 / 424 Umax = 30 V 140 °F 60 °C T2 446 °F 230 °C yes
T4…T1
Imax = 30 mA 140 °F 60 °C T3 338 °F 170 °C yes
NI / CL III / T4…T1
Pmax = 115 mW 158 °F 70 °C T3 383 °F 195 °C No
Ex ec IIC T4…T1
158 °F 70 °C T3 302 °F 150 °C yes
DIP A21 TA 85°C to Tmedium
158 °F 70 °C T4 266 °F 130 °C yes
158 °F 70 °C T5 203 °F 95 °C yes
86 °F 30 °C T6 77 °F 25 °C yes
1) Order code ‘Explosion protection and approvals’ (versions 9, 10), see Ordering Information on page 41.
2) For connection to an intrinsically safe circuit.
3) Installation in accordance with Installation Drawing SDM-10-A0253.
4) For connection to a non-intrinsically safe circuit.
5) For connection in Division 2 or Zone 2.
Electrical connections
Installation instructions
The concept of intrinsic safety enables multiple intrinsic
safety devices with FM or CSA approval to be interconnected,
without entity parameters being examined specifically,
subject to observation of the following conditions:
• Uo or Voc or Vt ≤ 0 V max, Io or Isc or It ≤ I max, Ca or Co ≥ Ci
+ Ccable, La or Lo ≥ Li + Lcable, Po ≤ Pi.
• For installation in Class II and III environments, dust-
proof ignition blocks must be used.
• Devices connected to such equipment must not use or
generate rms or DC voltage in excess of 250 V.
• The installation must meet the requirements in
accordance with ANSI / ISA RP 12.6 ‘Installation of
Intrinsically Safe Systems for Hazardous Locations’
1 FM Approved IS Barrier 2 Potential equalization
and the National Electrical Code (ANSI / NFPA 70)
Figure 20: FM / cCSAus electrical connection sections 504, 505 and CEC.
• The configuration of the associated apparatus must
Terminal Function have Factory Mutual Research and CSA approval in
accordance with the entity concept.
31 / 32 Power supply / current output / HART output
• Devices must be installed in accordance with the
41 / 42 Binary output
manufacturer-supplied installation drawing of the
Alarm signaling unit (min.)
associated equipment.
51 / 52 Alarm signaling unit (max.)
• Changes to drawings are only permitted subject to
prior approval from Factory Mutual Research and CSA.
• Only shielded twisted pair cables may be used (see
above).
Note
For intrinsically safe installations, the FAM540 must be
installed as illustrated in the "Intrinsic Safety Control Drawing
SDM-10-A0253".
UV Power supply
—
… Use in potentially explosive atmospheres in accordance with FM and cCSAus
Earthing Power supply
Note
The operator must make sure that when connecting the
protective ground (PE), there are no potential differences
between protective ground (PE) and potential equalization,
even in the event of a fault.
Signal cable
For ambient temperatures below 5 °C (41 °F) or above
40 °C (104 °F), signal cables that are suited for the
minimum/maximum ambient temperatures in question must
be used.
Only use signal cables made from copper, copper-coated
aluminum, or aluminum.
The recommended tightening torque for the terminals is
0.8 Nm (7 in lb) or higher, in accordance with the
specification.
Change from two to one column
VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E 41
—
Change from two to one column
Ordering Information
Main ordering information FAM541
VA Master FAM541 Metal Cone Variable Area Flowmeter, standard, for FAM541 X X XX XX X X XX X XX
general purposes
Display Type / Output Signal
Analog indicator / Without A
Analog indicator / Min-Alarm B
(Amplifier D163A011U01 bis U03 as a separate item)
Analog indicator / Max-Alarm C
(Amplifier D163A011U01 bis U03 as a separate item)
Analog indicator / Min- and Max-Alarm D
(Amplifier D163A011U04 bis U06 as a separate item)
Analog indicator / 4 bis 20 mA with HART protocol E
Analog indicator with LCD display / 4 bis 20 mA with HART protocol F
Housing Material / Cable Gland
Aluminium / Cable gland M20 × 1.5 (Indicator type A with blind plug M25 × 1.5) 1
Aluminium / Thread ½ in NPT (Indicator type A with blind plug M25 × 1.5) 2
Stainless steel / Cable gland M20 × 1.5 (Indicator type A with blind plug M25 × 1.5) 31
Stainless steel / Thread ½ in NPT (Indicator type A with blind plug M25 × 1.5) 41
Explosion Protection Certification and Approvals
Without Y0
ATEX / IECEx, (Zone 2 / 21), Ex ec B12
ATEX / IECEx, (Zone 1 / 21), Ex i, Ex ec A42
ATEX / IECEx, (Zone 1 / 21), Ex d, Ex i A93
FM / CSA, Class I, Div 1, 2 (Zone 1, 2), XP + IS, NI F34
FM / CSA, Class I, Div 1, 2 (Zone 1, 2), IS, NI F44
Process Connection
Flange F1
Flange with groove (DIN 2512) F2
Female thread metric (DN 25 [1 in] = PN 100, DN 40 [1 ½ in] = PN 40) T15
Connection Size
DN 15 (½ in) A
DN 25 (1 in) B
DN 40 (1-½ in) N
DN 50 (2 in) C
DN 80 (3 in) D
Float Design
Standard 1
With gas damping 3
Continued see next page
1 Not available with Ex d or XP applications
2 Version with analog indicator without output signal: Type of protection "c" only available with ATEX approval
3 With Ex d approved cable gland. Not available for analog indicator without output signal
4 Only with cable entry ½ in NPT
5 Not available with FM / CSA approval, only available in 1 in and 1 ½ in size
42 VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E
—
… Ordering Information
VA Master FAM541 Metal Cone Variable Area Flowmeter, standard, for general purposes XX XX XX XX XX XX
Primary
Without, secondary only Y0*
Standard Y1**
Measuring Deviation
Class 1.6 acc. VDE / VDI 3513 A1**
1 % of full scale AA
4 % of full scale (high viscosity, without calculation) AK
4 % of full scale (high viscosity, with calculation) AL
Material: 3.1, 3.2 Inspection / NACE
Material monitoring with inspection certificate 3.1 acc. EN 10204 C2
Material monitoring with inspection certificate 3.2 acc. EN 10204 C3
Material monitoring NACE MR 01-75 with inspection certificate 3.1 acc. EN 10204 CN
Material: 2.1 Compliance
Declaration of compliance with the order 2.1 acc. EN 10204 C4
Certificates: 3.1 Visual, Dimensional and Functional Test
Inspection certificate 3.1 acc. EN 10204 of visual, dimensional and functional test C6
Certificates: 3.1 PMI Test
Inspection certificate 3.1 acc. EN 10204 of positive material identification PMI CA
* For Ex devices upon request
** Standard, automatically specified by ABB if not specified by customer
VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E 43
—
… Ordering Information
Main ordering information FAM544
VA Master FAM544 Metal Cone Variable Area Flowmeter, hygienic design, for specific food and FAM544 X X XX XX X X XX X XX
pharmacy applications
Display Type / Output Signal
Analog indicator / Without A
Analog indicator / Min-Alarm (Amplifier D163A011U01 to U03 as a separate item) B
Analog indicator / Max-Alarm (Amplifier D163A011U01 to U03 as a separate item) C
Analog indicator / Min- and Max-Alarm (Amplifier D163A011U04 to U06 as a separate item) D
Analog indicator / 4 to 20 mA with HART protocol E
Analog indicator with LCD display / 4 to 20 mA with HART protocol F
Housing Material / Cable Gland
Aluminium / Cable gland M20 × 1.5 (Indicator type A with blind plug M25 × 1.5) 1
Aluminium / Thread ½ in NPT (Indicator type A with blind plug M25 × 1.5) 2
Stainless steel / Cable gland M20 × 1.5 (Indicator type A with blind plug M25 × 1.5) 31
Stainless steel / Thread ½ in NPT (Indicator type A with blind plug M25 × 1.5) 41
Explosion Protection Certification and Approvals
Without Y0
ATEX / IECEx, (Zone 2 / 21), Ex nA, c, Ex tD B12
ATEX / IECEx, (Zone 1 / 21), Ex ia, Ex nA, c, Ex tD A42
ATEX / IECEx, (Zone 1 / 21), Ex d, Ex ia, Ex nA, c, Ex tD A93
FM / CSA, Class I, Div 1, 2 (Zone 1, 2), XP + IS, NI F34
FM / CSA, Class I, Div 1, 2 (Zone 1, 2), IS, NI F44
Process Connection
Thread DIN 11851 S1
SMS 1145 S2
Connection Size
DN 25 (1 in) B
DN 40 (1 ½ in) N
DN 50 (2 in) C
DN 80 (3 in) D
DN 100 (4 in) E
Float Design
Standard 1
Pressure Rating
PN 6 D0
PN 25 D35
PN 40 D46
Design Level
(Specified by ABB) X
Tube / Float Combination
(Specified by ABB.) See flow range tables. XX
1) Not available with Ex d or XP applications
2) Version with analog indicator without output signal: Type of protection "c" only available with ATEX approval
3) With Ex d approved cable gland. Not available for analog indicator without output signal
4) Only with cable entry ½ in NPT
5) For DN 50 and DN 80 (2 in and 3 in)
6) For DN 25 and DN 40 (1 in and 1-½ in)
VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E 45
VA Master FAM544 Metal Cone Variable Area Flowmeter, hygienic design, for specific XX XX XX XX XX XX XX XX
food and pharmacy applications
Primary
Without, secondary only Y0*
Standard Y1**
Measuring Deviation
Class 1.6 acc. VDE / VDI 3513 A1**
1 % of full scale AA
4 % of full scale (high viscosity, without calculation) AK
4 % of full scale (high viscosity, with calculation) AL
Material: 3.1, 3.2 Inspection / NACE
Material monitoring with inspection certificate 3.1 acc. EN 10204 C2
Material monitoring with inspection certificate 3.2 acc. EN 10204 C3
Material monitoring NACE MR 01-75 with inspection certificate 3.1 acc. EN 10204 CN
Material: 2.1 Compliance
Declaration of compliance with the order 2.1 acc. EN 10204 C4
Certificates: 3.1 Visual, Dimensional and Functional Test
Inspection certificate 3.1 acc. EN 10204 of visual, dimensional and functional test C6
Certificates: 3.1 PMI Test
Inspection certificate 3.1 acc. EN 10204 of positive material identification PMI CA
Certificates: 3.1 Pressure Test
Pressure test acc. AD2000 CB
Test package (pressure test, non-destructive test, welder and welding procedure certificate) CP
(Non-destructive test = X-ray for welding neck flanges, dye penetration test for slip-on flanges)
Certificates: Calibration, Test Report
Certificate of compliance for calibration 2.1 acc. EN 10204 CM
Inspection certificate 3.1 acc. EN 10204 of calibration, with test report CE
* For Ex devices upon request
** Standard, automatically specified by ABB if not specified by customer
46 VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E
—
… Ordering Information
Trademarks
HART is a registered trademark of FieldComm Group, Austin, Texas,
USA
Buna-N is a registered trademark of DuPont Dow Elastomers.
™ Hastelloy C-2 is a Haynes International trademark
VA MASTER FAM540 METAL CONE VARIABLE AREA FLOWMETER | DS/FAM540-EN REV. E 47
—
Questionnaire
Customer: Date:
Ms./Mr.: Department:
Telephone: Fax:
abb.com/flow
DS/FAM540-EN Rev. E 07.2019
—
We reserve the right to make technical changes or modify the contents of this document without
prior notice. With regard to purchase orders, the agreed particulars shall prevail.
ABB does not accept any responsibility whatsoever for potential errors or possible lack of
information in this document.
We reserve all rights in this document and in the subject matter and illustrations contained therein.
Any reproduction, disclosure to third parties or utilization of its contents – in whole or in parts – is
forbidden without prior written consent of ABB.