Datasheet

RADII™

KL1604E

Stationary Anode X-ray Tube

 Especially designed for security check in


railway station, airport, etc.
 This tube has focus 0.8 and is available for
maximum tube voltage 160 kV.
 Installed in the same enclosure with the
high voltage transformer
 A special designed hood anode structure
for the special beam angle.

Hangzhou Kailong Medical Instruments Co., Ltd.


Address: 6, No.6 Road, Dongzhou Industrial Zone, Fuyang, Hangzhou, China 311401
Tel: +86 571 87195007 Email: info@kailongx-ray.com
KL1604E Datasheet No. SA0055 Version Number A2 Page 1 of 7
KL1604E

General Data
Electrical:
Circuit:
High Voltage Generator………………..…Constant Potential High-Voltage Generator
Grounding………………………………………………………….……. Center-grounded
Nominal X-ray Tube Voltage (IEC60613)….………………………....…..……....…..160 kV
Nominal Focal Spot Value (IEC60336)………………………………….…………………0.8

Mechanical:
Dimensions
Overall Length………………………………………………...…See dimensional outline
Maximum Diameter ………………………………...….……… See dimensional outline
Target:
Anode Angle (IEC/TR 60788) ...................................................................10 degrees
Beam angle
Vertical………………………………………….…………..…………….…...20 degrees
Horizontal………………………………………………………….…………120 degrees
Material ….................................................................................................... Tungsten
Inherent Filtration ......................................................................0.1 mm Ti, 1.7 mm glass
Weight (Approx.)………………………………………………………………..……......1.1 kg
Cooling Method………………..……Oil immersed (60°C Max.) and convection oil cooling
Tube Holding ..................................... Holding the glass envelope of the anode end and
cathode end or the screw of the anode shank.

Hangzhou Kailong Medical Instruments Co., Ltd.


Address: 6, No.6 Road, Dongzhou Industrial Zone, Fuyang, Hangzhou, China 311401
Tel: +86 571 87195007 Email: info@kailongx-ray.com
KL1604E Datasheet No. SA0055 Version Number A2 Page 2 of 7
KL1604E

Absolute Maximum and Minimum Ratings

(At any time, these values must not be exceeded.)


Maximum X-ray Tube Voltage (IEC60613) …………..…………………….……...….…..160 kV
Maximum X-ray Tube Current(IEC60613) ...........................................................40kV@8 mA
Maximum Filament Current..............................................................................................3.7 A
Filament Voltage (At maximum filament current 3.7 A) ............................................3.2~3.8 V
Maximum Anode Input Power ……….…………………………...……...…….…..…….….640 W
Wire length…………………………………………………………………...……...……….205 mm

Environmental Limits
Operating Limits (in dielectric oil):
Oil Temperature ................................................................................................10 ~ 60 °C
Oil Pressure ................................................................................................. 70 ~ 106 kPa
Shipping and Storage Limits:
Temperature..................................................................................................... -40~ 70 °C
Humidity ........................................................................................................... 10 ~ 90 %
(No condensation)
Atmospheric Pressure ................................................................................. 50 ~ 106 kPa

Hangzhou Kailong Medical Instruments Co., Ltd.


Address: 6, No.6 Road, Dongzhou Industrial Zone, Fuyang, Hangzhou, China 311401
Tel: +86 571 87195007 Email: info@kailongx-ray.com
KL1604E Datasheet No. SA0055 Version Number A2 Page 3 of 7
KL1604E

Maintaining seasoning schedule


Prior to usage, season the tube in accordance with the seasoning schedule given below until
the required tube voltage is attained. Example given – needs to revise by the manufacturer
and specified in the data sheet of the part:

Initial incoming seasoning schedule for idle period (more than 6 months)
Circuit: DC (Center grounded)

Tube Voltage [kV] Tube Current [mA] Exp. Time [s]


60 1 60
70 1 60
80 1 60
90 1 60
100 1 60
110 1 60
120 1 60
130 1 60
140 1 60
150 1 60
160 1 60

When the tube current is unstable in seasoning, immediately switch off the tube voltage and
after an interval of 5 minutes or more, increase the tube voltage gradually from the low
voltage while making sure that the tube current is stable.

The withstand voltage performance of the tube unit will be lowered as the exposure time and
the number of operation increase. Stain-like impact traces might appear on the x-ray tube
target surface by slight discharge in the course of the seasoning. These phenomena are one
process to recover the withstanding voltage performance at that time.

Therefore, if it is in stable operation at the maximum tube voltage of seasoning subsequent


to them, the tube unit can be used without any interference to its electrical performance
which in use.

Hangzhou Kailong Medical Instruments Co., Ltd.


Address: 6, No.6 Road, Dongzhou Industrial Zone, Fuyang, Hangzhou, China 311401
Tel: +86 571 87195007 Email: info@kailongx-ray.com
KL1604E Datasheet No. SA0055 Version Number A2 Page 4 of 7
KL1604E

X-Ray Tube Dimensional Drawings– KL1604E

Hangzhou Kailong Medical Instruments Co., Ltd.


Address: 6, No.6 Road, Dongzhou Industrial Zone, Fuyang, Hangzhou, China 311401
Tel: +86 571 87195007 Email: info@kailongx-ray.com
KL1604E Datasheet No. SA0055 Version Number A2 Page 5 of 7
KL1604E

Cautions!!!
X-ray tube will emit X–ray when it is energized with high voltage, Special knowledge should be
required and cautions need to be taken when handling it.

1. Only a qualified specialist with X-Ray tube knowledge should assemble, maintain and remove
the tube. When mounting tube inserts adopt proper caution, in order to avoid glass bulb breaking and
fragments projection. Please use protective gloves and glasses.
2. Tube insert connected to H.V. supply is a radiation source: be sure to take all necessary safety
cautions.
3. Wash thoroughly with alcohol the external surface of tube insert (care of fire risk). Avoid contact
of dirty surfaces with cleaned tube insert.
4. Clamp system inside housing or self-contained units must not mechanically stress the tube.
5. After installation, check the right working of the tube (no fluctuation of tube current nor crackling).
6. Comply with insert thermal parameters, planning and programming the exposure parameters
and cooling pauses. Housing or self-contained units must be provided with an adequate thermic
protection.
7. Voltages indicated in charts are valid for transformer supplied with ground center.
8. It is extremely important to observe the connection diagram and the grid resistor value. Any
change could modify the dimensions of the focal spot, also varying diagnostic performances or
overloading anode target.
9. Tube inserts contain environment polluting materials, particularly lead liner tubes. Please apply
to qualified operator for waste disposal, according to local regulation requirements.
10. When any abnormalities are found during operation, immediately switch off the power supply
and contact the service engineer.

Hangzhou Kailong Medical Instruments Co., Ltd.


Address: 6, No.6 Road, Dongzhou Industrial Zone, Fuyang, Hangzhou, China 311401
Tel: +86 571 87195007 Email: info@kailongx-ray.com
KL1604E Datasheet No. SA0055 Version Number A2 Page 6 of 7
KL1604E

Notes
• This high vacuum product is produced according to state-of-the–art technology. To prevent

implosion please handle with care and use protective devices, e.g. glasses!

• In the interest of complying with legal requirements concerning the environmental compatibility of

our products (protection of natural resources, avoidance of waste) we endeavor to reuse

components and to return them to the production cycle. We guarantee the functioning, quality and

life of these components by taking extensive quality assurance measures, just as for factory-new

components.

• All IEC standards are the update versions.

The Hangzhou Kailong Medical instruments Co., Ltd.


is ISO 13485 certified, manufactures in accordance with
the Quality System Regulations (QSR) as defined by the
Food and Drug Administration (FDA) and endeavors to
comply with legal requirements concerning the
environmental compatibility of its products.
The reproduction, transmission or use of this
document
or its contents is not permitted without express written
consent. Offenders will be liable for damages. Kailong
reserves the right to modify the design and specifications
contained herein without prior notice. All rights reserved,
particularly in connection with patent applications or
registrations of utility model or design.
© Hangzhou Kailong Medical instruments
Co., Ltd.

For further information, please contact:


Hangzhou Kailong Medical instruments Co., Ltd.
Address:6, No. 6 Road, Dongzhou Industrial Zone,
Fuyang, Hangzhou, China 311401
Fax: +86571 87195003
http://www.kailongx-ray.com

Hangzhou Kailong Medical Instruments Co., Ltd.


Address: 6, No.6 Road, Dongzhou Industrial Zone, Fuyang, Hangzhou, China 311401
Tel: +86 571 87195007 Email: info@kailongx-ray.com
KL1604E Datasheet No. SA0055 Version Number A2 Page 7 of 7