Transmissores Microprocessados

Mod. EQ 350

PROFIBUS PA ver. 3.0

FOUNDATION

Automatic switch between protocols

Basic or LAS capability with F.F.

TM

Application

Fieldbus ver. ITK 4.51

Standard signal

Function

Level

Mounting

DIN form B sensor head

Input

Lin. R 0..100 kohm

Output

Profibus PA
FOUNDATION Fieldbus

Supply

Bus-powered

Channels

1 channel

Configuration

Profibus
FOUNDATION Fieldbus

Features

PID controller
Customer linearisation
Alarm/control

Electrical specifications:
Specifications range:
-40 C to +85C
Common specifications:
Supply voltage..................................... 9...32 VDC
Consumption ............................ < 11 mA
Isolation Voltage, test / operation......... 1.5 kVAC / 50 VAC
Signal / noise ratio................................ Min. 60 dB
Response time (programmable).... 160s
Updating time.... < 400ms
Execution time, PID controller.... < 200ms
Execution time, analogue input.. < 50ms
Signal dynamics, input ......................... 24 bit
Calibration temperature... 2028C
Accuracy, the greater of general and basic values:
General values
Absolute
accuracy

Input type

0.05% of reading

All

Temperature
coefficient
0.002% of reading / C

Basic values
Input type
Pt100 and PT1000
Ni100
Cu10
Lin. R.

Vibration ......................................................................... IEC 68-2-6 Test FC


Lloyds specification no. 1............................................... 4 g / 2...100 Hz
Humidity ......................................................................... < 95% RH (non-cond.)
Dimensions...................................................................... 44 x 20.2 mm
Tightness (enclosure / terminal) ..................................... IP68 / IP00
Weight............................................................................. 55 g
Electrical specifications, input:
RTD and linear resistance input:
RTD
type
Pt25Pt1000
Ni25NI1000
Cu10Cu1000
Lin.Resistance
Potentiometer

Min.
value
-200C
-60C
-50C
0W
0W

Max.
value
+850C
+250C
+200C
10k W
100k W

Norm
IEC60751/JIS C 1604
DIN 43760

= 0.00427
-

Cable resistance per wire.................................................50 W


Sensor current..................................................................nom. 0,2 mA
Effect of sensor cable resistance
(3- / 4-wire).......................................................................< 0.002 W / W
Sensor error detection......................................................Yes
Short circuit detection.......................................................< 15 W

TC / mV input:
Type
B
E
J
K
L
N
R
S
T
U
W3
W5
Ext. CJC
mV

Min.
Temperature
+400C
-100C
-100C
-180C
-100C
-180C
-50C
-50C
-200C
-200C
0C
0C
-40C
-800

Max.
Temperature
+1820C
+1000C
+1200C
+1372C
+900C
+1300C
+1760C
+1760C
+400C
+600C
+2300C
+2300C
+135C
+800

Norm
IEC584
IEC584
IEC584
IEC584
DIN 43710
IEC584
IEC584
IEC584
IEC584
DIN 43710
ASTM E988-90
ASTM E988-90
IEC60751
-

Basic
accuracy

Temperature
coefficient

0.1C

0.002C / C

1.3C
0.05W

0.02C / C
0.002W / C

Cold junction compensation (CJC)..............................< 0,5 C


Sensor error detection...............................yes
Sensor error current:
When detecting.............................................................Nom. 4A
Else................................0 A
Short circuit detection.............................< 3mV
Output:

FOUNDATIONTM Fieldbus connection:


TM
FOUNDATION Fieldbus version.............................. ITK 4.51
TM

0.15

0.002C / C

Volt

10V

TC type:
E, J, K, L, N, T, U
TC type:
B, R, S, W3, W5

0.2 V /C

0.5C

0.010C / C

1C

0.025C / C

EMC immunity influence .............< 0.1% of reading


Extended EMC immunity:
NAMUR NE 21, A criterion, burst..< 1% of reading

www.equipe-termopar.com.br

FOUNDATION F. capability.........................................Basic or LAS


TM
FOUNDATION F. function blocks.................................2 analogue and 1 PID
PROFIBUS PA connection:
PROFIBUS PA protocol standard..................................EN 50170 vol. 2
PROFIBUS PA function blocks......................................2 analogue
PROFIBUS PA address (at delivery)..............................126