Perkin Elmer Manual Operacion Series 200 VacuumDegasser PDF
Perkin Elmer Manual Operacion Series 200 VacuumDegasser PDF
Perkin Elmer Manual Operacion Series 200 VacuumDegasser PDF
User’s Manual
Release History
Any comments about the documentation for this product should be addressed to:
User Assistance
PerkinElmer
710 Bridgeport Avenue
Shelton, CT 06484-4794
Notices
The information contained in this document is subject to change without notice.
Except as specifically set forth in its terms and conditions of sale, PerkinElmer makes no
warranty of any kind with regard to this document, including, but not limited to, the
implied warranties of merchantability and fitness for a particular purpose.
PerkinElmer shall not be liable for errors contained herein for incidental consequential
damages in connection with furnishing, performance or use of this material.
Copyright Information
This document contains proprietary information that is protected by copyright.
All rights are reserved. No part of this publication may be reproduced in any form
whatsoever or translated into any language without the prior, written permission of
PerkinElmer, Inc.
Trademarks
Registered names, trademarks, etc. used in this document, even when not specifically
marked as such, are protected by law.
Europe
All information concerning EMC standards is in the Declaration of Conformity and these
statements will change as the European Union adds new requirements.
S-2 Safety and Regulatory Information Symbols and Conventions used in this Manual
Figure S-1. Location of Warning Labels on the Series 200 Vacuum Degasser
S-4 Safety and Regulatory Information Warnings on the Instrument
For protection against fire hazard, replace only with same type
and rating of fuse.
The following warning label covers the fuse receptacle located on the rear panel of the
instrument:
Electrical Warnings
Each instrument is certified to meet its functional and performance specification upon
release to shipment. The integrity of this quality system is routinely audited and is
certified by a registrar.
While the Universal Test Mix (UTM) Part No. 0089-0893, can be used for a system check,
we recommend that you develop a separate system suitability test and acceptance criteria
for each of your assays.
S-8 Safety and Regulatory Information Hazardous Chemicals
Hazardous Chemicals
Before using mobile phase solvents, you should be thoroughly familiar with all hazards
and safe handling practices. Observe the manufacturer’s recommendations for use, storage
and disposal. These recommendations are normally provided in the material safety data
sheets (MSDS) supplied with the solvents.
Operator. “Person operating equipment for its intended purpose.” [per IEC 1010-1,
Amendment 2].
Installation Category
Pollution Degree
This equipment will operate safely in environments that contain nonconductive foreign
matter up to Pollution Degree 2 in IEC 1010-1.
Storage Conditions
The Series 200 Vacuum Degasser may be stored under the following conditions:
Inside the unit, the solvent flows through a short length of Teflon AF® tubing which
is located in a vacuum chamber. Within this chamber a partial vacuum is maintained
by a constantly running, low RPM vacuum pump. Dissolved gasses migrate across
the tubing wall under a concentration gradient produced by the vacuum as the
solvent flows within the coil. Gasses removed are expelled, and the chamber is
maintained at a constant, preset vacuum level by varying the vacuum pump speed as
needed.
A special port in the vacuum pump continually flushes the pump head with a small
“bleed” of air to remove any solvent vapors which may enter the pump from the
vacuum chamber. This air bleed eliminates the need for any solenoid valves within
the system. This patented* design results in zero vacuum “hysteresis”. Previous
designs allowed the vacuum chamber pressure to fluctuate, with the pump cycling on
and then off in response to the vacuum level.
*Thisproduct is protected under U.S. patents 6,494,938; 6,248,157; 5,340,384 and 5,006,382. Other patents
pending.
ZHCR® is a registered trademark of Systec, Inc., New Brighton, Minnesota.
Teflon AF® is a registered trademark of E.I. du Pont de Nemours and Company.
1-2 Introduction
Immediately upon turning on the instrument, all 3 LED’s will momentarily flash on
then turn off. This confirms that the microprocessor has examined the vacuum sensor
and found that it is within the expected range of 0 to 810 mm Hg absolute pressure.
Following the startup test, the microprocessor ramps the vacuum pump to high RPM,
to quickly exhaust atmosphere from the vacuum chamber. The left-hand LED
illuminates as the pump can be heard to run at high RPM. As vacuum is established
in the chamber, the middle yellow LED illuminates, indicating the vacuum within the
chamber has reached a value of less than 80 mm Hg absolute. As the vacuum pump
continues to remove air within the vacuum chamber, the right-hand green LED will
illuminate and the middle yellow LED extinguish, indicating that the vacuum
chamber has reached a vacuum of 54 mm Hg absolute, where the performance of the
degasser meets its specified level. When the vacuum level passes through 47 mm Hg
absolute, the microprocessor ramps the pump speed to the low RPM, continuous run
mode. At this low RPM, the vacuum will continue to fall to a constant value,
wherein the fluctuation in the vacuum is less than 1mm Hg. This “zero hysteresis
constant run” (ZHCR) mode is necessary, due to the extremely low mass, high
response, Teflon AF® degassing channel. The effect of vacuum changes within the
vacuum chamber can easily be seen in the UV absorbance of methanol (for example)
at 215 nM. Only a ZHCR design ensures a baseline which is unaffected by the
degasser.
An additional benefit of the constantly running pump is that a pressure rise in the
vacuum chamber can be observed. This “smart leak detection” is a benefit of the
patent-pending design of the degasser, and is best described as follows: Due to the
fixed rate at which the pump runs, the gasses crossing the membrane pass from the
chamber at a fixed rate as well. If a leak occurs within the chamber, the
microprocessor will see this as a rise in pressure related to the vapor pressure of any
leaking solvent. If the pressure within the chamber rises above 54 mm Hg absolute,
the yellow LED will once again illuminate, indicating a possible leak condition
within the chamber. If the pressure rises above 80 mm Hg, the system will shut down
and go into a “safe” mode.
Introduction 1-3
POWER (Green)
Indicates when power is applied to the Vacuum Degasser (plugged in and Power
switch ON).
STATUS (Yellow)
• Flashing on and off in even 1-second intervals: pump was not able to reach
vacuum set point, indicating a possible leak in the system.
• Flashing on for 1 second and off for 2 seconds indicates a vacuum signal
error.
VACUUM (Green)
Exhaust Port
The gas pumped out of the vacuum chamber leaves the unit through the
exhaust port.
Power Receptacle
The power receptacle accepts the DC plug of the supplied AC Adapter. See
Specifications section for further power requirement information.
Power Switch
This On/Off rocker type switch applies line power to the Vacuum Degasser. The
rocker indicators are as follows: “O” = Off and “|” = On.
Scope of the Manual Introduction 1-5
l Fuses
l Nuts
l Ferrules
The items listed above are considered to be routinely replaced as part of the normal
wear of the instrument.
Conventions
The following conventions are used in this manual:
Solvents
• Wear appropriate eye protection at all times when handling chemicals. Use
safety glasses with side shields, goggles, or full-face shields, according to the
types of chemicals you will be handling.
• Wear suitable protective clothing, including gloves that are specifically resistant
to the chemicals being handled.
• Always use clean solvents. Solvents that have been distilled in glass (HPLC
Grade) are highly recommended.
• Filter the solvents through a 0.5-micron medium as an additional precaution.
• Carefully use and store flammable solvents, which may form hazardous by-
products when the instrument is shut down, by following the recommended
shutdown procedure.
• Check compatibility of solvent(s) with the type of column(s) being used.
• Know the relative polarity and miscibility of the solvents being used.
1-8 Introduction Warnings and Precautions
Buffers
Exercise care when using buffers in conjunction with organic solvents. NEVER
LEAVE BUFFERS IN THE SYSTEM OVERNIGHT. Buffers left in the system can
form salt crystals, which may cause premature pump seal failure, interfere with
proper check valve operation, and plug the connecting tubing or the detector
flowcell.
To remove buffers, flush the system first with water, then with methanol or
isopropanol. Change the mobile phase from methanol or isopropanol to water before
using buffers.
Corrosion
All parts that contact the mobile phase are made of PEEK, Kel-F®, Tefzel® or Teflon
AF®. PEEK is sensitive to Sulfuric acid and certain solvents. If you have questions
about your mobile phase, contact a PerkinElmer representative.
-1 . S o lv e n t s W h ic h M a y C o r r o d e t h e S e r ie s 2 0 0 V a c u u m D e g a s s e r
Air Bubbles
To prevent air from entering the system, and to ensure that pressure fluctuations do
not occur, observe the following precautions:
• Ensure that the fittings are tightened on the degasser input and output ports.
• Ensure that the pump's solvent inlet filter is below the solvent level in the solvent
reservoir.
• If the pump has not been used for an extended period of time, remove air bubbles
by connecting a priming syringe to the drain valve on the pump, opening the
drain valve, and fast-flushing the system to prime the pump. After priming the
pump, close the drain valve completely.
Pressure Buildup
Over time you may notice a gradual increase in the system operating pressure. If you
observe pressure readings greater than 3.45 MPa (500 psi) above the normal operating
pressure of your analysis, check the following items:
• If you are injecting “clean” samples and there is no other apparent cause for the
pressure buildup then remove, disassemble and clean the injector.
• If you are injecting “dirty” samples, the injector may be clogged or the column
packing material may have retained contaminated particulates. Remove and
clean the injector, and replace the column.
• The column end fitting or column frit may be plugged. Replace the end fitting or
frit to relieve excess pressure. (Refer to the instructions supplied with the
column.)
• In-line filters may be plugged. Replace the filter element.
• With the pump connected directly to the detector input, if you still experience
excessive pressure, the problem could be a plugged flowcell. You can flush, reverse-
flush, or rebuild the flowcell.
1-10 Introduction Warnings and Precautions
Specifications
T a b le 1 - 2 . S p e c if ic a t io n s o f t h e S e r ie s 2 0 0 V a c u u m D e g a s s e r
Specification Description
Dimensions: Height: 5.0 in.
Width: 2.87 in.
Depth: 9.81 in.
Weight 6 lb.
1-5 independent
Channels
Degassing Process
Gas permeation through a fluoropolymer membrane
1
Maximum Recommended Flow Rate 3.0 mL/min.
Pressure Drop2 1.37 mm Hg/mL/min.
~25% dissolved gases remaining in 60:40 MeOH/Water mixture at
Degassing Capacity 1 mL/min.
Dead Volume ~480 microliters per channel for standard channel
Materials contacting solvents PEEK, Glass-filled PTFE, Teflon AF®
Power:
Power requirement if using supplied AC 100 to 240 VAC (±10%), 1A, 50 to 60 Hz (±3 Hz)
Adapter
Power Requirement if not using supplied AC 15 to 24 VDC at 0.85 A maximum (0.5 A typical)
Adapter
Wall Sockets 4 supplied with AC Adapter, interchangeable: North
America/Japan, U.K., Continental Europe, Australia
Installation Over-Voltage Category II
Validation Output:
Signal 5 mVDC / 1 mm Hg absolute from 20 to 800 mm Hg
(0.100 VDC at 20 mm Hg; 4.000 VDC at 800 mm Hg)
Accuracy ±1.0% of reading ±0.010 VDC from 20 to 800 mm Hg
Operating Conditions:
Ambient Temperature 10 to 35 ºC
Ambient Relative Humidity (RH) 20 to 80 % RH (without condensation)
Altitude 0 to 2000 Meters
Indoor vs. Outdoor Use Indoor
Pollution Degree 2
Storage Conditions:
Ambient temperature -20 to +60 ºC
Ambient Relative Humidity 20 to 80% RH (without condensation)
Altitude 0 to 12000 M
1 Maximum recommended flow rate to prevent a 60:40 MeOH/Water mixture from outgassing. The estimate assumes low pressure
mixing and low flow restriction prior to the HPLC pump. MeOH/Water mixing represents the worst outgassing case and maximum
flow rate will likely increase with Acetonitrile/Water mixtures. High pressure mixing will also increase the maximum flow rate.
Degassing is still recommended.
2 Calculated tubing pressure per unit change in flow assuming laminar flow with a viscosity of 1.0 cP. Inlet and outlet bulkheads may
contribute to the overall pressure, but are not included in the estimate.
Installation 2
This chapter describes how to install your Series 200 Vacuum Degasser. The following
installation topics are covered in this chapter:
• Flammable • Caustic
• Explosive • Corrosive
• Toxic
Make sure that your laboratory environment consists of the following temperature and
humidity levels:
• Ambient temperature between 10 and 35 °C
• Constant relative humidity between 20 and 80%, without condensation
Use care when working with hazardous solvents, or solvents that produce hazardous by-
products.
Solvents/Mobile Phase
Use only HPLC grade solvents in all analyses. HPLC grade water and methanol (1 liter
each) are required for performance verification.
Corrosion
All parts that contact the mobile phase are made of PEEK, Glass-filled PTFE or Teflon
AF®. PEEK is sensitive to Sulfuric acid and certain solvents.
Space Requirements
The Series 200 Vacuum Degasser is designed to sit on a bench top, and is plumbed into
the LC system between the solvent supply and pump. A space 5 ¼-in high and 3-in wide
is sufficient. The case is 10-in deep (front to back), but additional space is required both
in front, to accommodate the tubing connected to the unit, and behind to accommodate
the power cord.
2-4 Installation Unpacking Your Series 200 Vacuum Degasser
A Start-Up Kit is supplied with the Series 200 Vacuum Degasser. Use Table 2-1 to
inventory the Start-Up Kit parts.
Table 2-1. Start-Up Kit for the Series 200 Vacuum Degasser
Item Quantity
AC Adapter (including cord) 1
Interchangeable Wall Plugs 4 (North America/Japan, U.K., Continental Europe,
Australia)
Operator’s Manual 1
Plug, ¼-28, Rheodyne 6118 2
Electrical Requirements
The AC Adapter supplied with the Vacuum Degasser incorporates a universal AC input,
switching regulator. This allows the instrument to operate at any AC line voltage from
100 to 240 VAC (±10%) with a line frequency range of 47 to 63 Hz. The switching
regulator senses the incoming line voltage and automatically adjusts its operation
accordingly. The switching regulator assembly incorporates its own on-board line
voltage fuse. This fuse is not user serviceable. In the event that this fuse blows, it will
be necessary to replace the AC Adapter.
Locate the AC Adapter with power cord and the set of four interchangeable wall plugs.
Plug the round connector at the other end of the AC Adapter’s cord into the Power jack
on the Vacuum Degasser rear panel. From the set of four plugs, select the one
appropriate for the local electrical socket and install it onto the AC adapter. With the
power switch off, insert plug into the AC supply. The AC adapter should be positioned
for easy disconnection.
1. Run a line of 1/8” O.D. x 1/16” I.D. Teflon chromatography tubing from the solvent
supply to the Vacuum Degasser.
2. Push the tubing through a PEEK 1/8” male ¼-28 fitting and slide a ferrule over the
tubing end (see figure below). Cut the Teflon tubing so the end is flat.
3. Screw the ¼-28 fitting into one port on the front of the Vacuum Degasser (Channel
A, for example). The direction of flow through the Vacuum Degasser is not critical.
Plastic connectors should be tightened by hand. Overtightening them will damage
the threads.
5. Once all desired solvent lines have been connected to the Vacuum Degasser, any and
all unused ports should be plugged. Use the plugs supplied. Press in by hand.
2-6 Installation Connecting the Tubing
6. Prime each degassing membrane by pulling the solvent from the reservoir through
the degassing system. This can be done by connecting a syringe to the tubing or LC
pump priming port and drawing air and/or mobile phase into the syringe until no air
remains in the tubing, approximately 5 milliliters.
The following illustration shows the tubing connections that are typically made between
the Vacuum Degasser and other instruments in an LC system. The direction of flow
through the Vacuum Degasser is not critical.
Installation Connecting the Tubing 2-7
During initial pump-down, the yellow Status LED will be lit. Once the vacuum has
reached normal operating level, the yellow LED will extinguish and the green
Vacuum LED will illuminate. If you want to confirm that the pump is running,
beyond the front panel LEDs, the slight vibration caused by the microstepping of the
motor driving the vacuum pump may be felt by placing your hand on the instrument.
2. S ta r t s o lv e n t flo w th r o u g h th e s y s te m a n d c h e c k fo r le a k s a r o u n d th e ¼ -2 8
c o n n e c to r s .
If a leak occurs at the connection, tighten the fitting an additional 1/8 turn. If the
leak persists, disconnect the leaking fitting and inspect it. If the nut and ferrule
2-8 Installation Connecting the Tubing
appear to be in good condition, reconnect the fitting. If the leak persists, replace the
nut and ferrule and repeat the procedure until you achieve leak-free operation.
3. Turn off the Vacuum Degasser when the LC to which it is connected is not in use.
The vacuum chamber(s) will slowly return to atmospheric pressure when the unit is
powered off. This is accomplished by a small, in-line vacuum bleed and reduces the
possibility of solvent vapors condensing in the vacuum tubing or pump head.
4. When flushing a line of solvent, the single lumen coil inside the chamber
contains a very small amount of solvent (approximately 480 microliters).
When changing from one solvent to another where the final solvent is
immiscible with the first, use an intermediate solvent miscible with both the
initial and final solvent. Carryover from solvent to solvent is much less than
previous PTFE designs. Once air bubbles have been cleared from the solvent
line, any further bubbles observed will be coming from the solvent reservoir or
from a leaking fitting.
Before beginning, make a final check of all tubing connections to and from the Series
200 Vacuum Degasser.
Principles of Operation
The Series 200 Vacuum Degasser consists of a vacuum chamber, degassing tube,
variable speed vacuum pump, microprocessor controller, sensor, and check valves. The
solvent (mobile phase) flows into a degassing tube, which is inside a vacuum chamber.
Decreased pressure in the chamber causes the outward movement of gas dissolved in the
mobile phase across the tube wall, in accordance to Henry’s Law, thus degassing the
mobile phase. The pressure in the vacuum chamber is established by the vacuum pump
and monitored by the microprocessor through an integrated absolute pressure sensor.
Degassed mobile phase exits the vacuum degasser and enters the pump.
3-2 Using the Series 200 Vacuum Degasser Principles of Operation
Immediately upon turning on the instrument, the microprocessor examines the vacuum
sensor signal to confirm that it is within an expected range. Following the startup test,
the microprocessor ramps the vacuum pump to high RPM, to quickly exhaust atmosphere
from the vacuum chamber. As the vacuum level approaches the preset control value, the
pump RPM will slowly ramp down to a low speed (typically 40 to 60 RPM).
Afterwards, the pump RPM will vary slightly, as needed under the changing degassing
load, to maintain a virtually constant vacuum level (50 ±0.5 mm Hg). This "zero
hysteresis, constant run" (ZHCR®) mode is necessary, due to the extremely low mass,
high response, Teflon AF® degassing tubing. Only a ZHCR® design ensures a baseline
which is unaffected by the degasser.
This relatively recent addition to the field of degassing has properties not found in other
fluoropolymers. The fully amorphous nature of this fluoropolymer and its molecular
structure creates a molecular level porosity unlike the mechanically induced porosity in
PTFE extruded tubing. In addition, unlike the process used in extruding PTFE, no
extrusion agents are needed (like kerosene, etc.) which contaminate mobile phases until
they are extracted by the mobile phase over time. Likewise, this molecular structure,
combined with the very small surface areas required to degas the mobile phase, reduces
the possibility of carryover from one solvent or mobile phase to another to virtually zero.
Teflon AF® is so non-polar that it is both solvophobic and hydrophobic. This feature of
Teflon AF® reduces the possibility of cross-channel contamination from one channel to
another, and when combined with the ultra-low internal volumes of Teflon AF® channels
needed for HPLC flow rates, all but eliminates this cross contamination concern by the
chromatographer. Teflon AF® has been used in certain optical systems associated with
HPLC for a few years without concern for normal HPLC solvents. However, Teflon
3-4 Using the Series 200 Vacuum Degasser Principles of Operation
AF® is soluble in certain solvents (see cautionary statements) and must not be used to
degas these types of solvents.
Teflon AF® is permeable to some degree to water vapor whereas PTFE is not. While the
vacuum pump in the Vacuum Degasser contains internal provisions for sweeping water
or solvent vapor from the pump continuously, it is possible that over time, high
concentration buffers may form crystals within the channel due to the loss of water
within the channel. The same precautions should be taken to prevent crystallization
within these channels as are taken for the HPLC pump. See the “Short-term Shutdown”
procedures.
Operating Summary Using the Series 200 Vacuum Degasser 3-5
Operating Summary
1. Select and fill each solvent reservoir with the mobile phase for your analysis.
2. Verify that the Series 200 Vacuum Degasser is properly installed as described in
Chapter 2, “Installation.”
3. Verify that the tubing to your injector, column, and detector is properly connected.
Also verify that plugs are installed in the unused ports.
4. Disconnect and remove the tubing that is connected to the output port, connect the
priming syringe to this port, and pull the solvent through the degasser until bubbles
no longer appear. Then reconnect the tubing to the output port.
5. Switch on the degasser and start the pump at 1.0 mL/min. Allow the system to
equilibrate for 5-10 minutes. The small volumes contained in the Series 200 Vacuum
Degasser should only be considered in chromatograph equilibration time when flow
rates less than 1 mL/min are used.
To more thoroughly degas a mobile phase, connect the outlet of the organic
channel to the inlet a second channel and the outlet of the second channel to the
pump. This places the two channels in series and doubles the degassing
capacity for the organic portion of the mobile phase.
Shutdown Using the Series 200 Vacuum Degasser 3-7
Shutdown
1. Remove harmful mobile phases from the vacuum degasser and other instruments in
the system.
2. Flush the column according to the instructions supplied with the column. Flush
buffer salts from the system with water. Evaporation leaves salt crystals that may
form harmful deposits. Remove chloroform or solvents that can decompose to form
hydrochloric acid from the system.
3. After removing harmful mobile phases, prepare the detector for most mobile phases
by flushing it with isopropanol. To avoid contaminating the system, refilter or
discard solvents (including water) that were exposed to the environment for
more than 24 hours before use.
Long-Term Shutdown
2. Remove the column and direct the pump output tubing to a beaker. Flush the
vacuum degasser first with water and then with isopropanol.
3. Turn off the vacuum degasser. Then disconnect the tubing between the vacuum
degasser and solvent reservoirs and the vacuum degasser and pump. Plug all of the
ports on the vacuum degasser.
5. Before using the vacuum degasser, completely purge it with the correct solvent for
the column before reconnecting the column and restarting the system.
Maintenance 4
You can experience long and trouble-free performance from your Series 200 Vacuum
Degasser by performing both routine and preventive maintenance procedures.
Preventive Maintenance
Perform preventive maintenance to ensure that your Series 200 Vacuum Degasser will
perform consistently at an optimal level. To maintain the Degasser in the best condition,
the following measures are recommended:
• Use only high-purity solvents (preferably HPLC Grade) for mobile phases. (Water
should be bottled HPLC grade, or filtered and deionized tap water.)
• Filter all solvents to prevent particulate contamination and tubing blockages.
• Ensure that all new tubing (stainless steel) is passivated and thoroughly flushed
before making pump connections. (The tubing available from PerkinElmer is
passivated.)
• Follow the short- and long-term shutdown procedures that are described in Chapter
3 of this manual.
4-2 Maintenance Routine Maintenance
Routine Maintenance
Routine maintenance is defined as replacing the normal wear items when you notice
degradation in performance.
If you have a problem, you can save time and money by referring to the Troubleshooting
Guide (on the next page) before calling your local PerkinElmer Service Engineer. Your
problem may be minor and you may be able to correct it yourself by using the
Troubleshooting Guide to pinpoint the cause.
Troubleshooting Guide
The following graph illustrates the performance of the Series 200 Vacuum Degasser.
Figure 4-1. Graphical representation of the performance of the Series 200 Vacuum
Degasser
To place an order for supplies and many replacement parts, request a free e-ssentials catalog or
ask for information, visit our website at www.perkinelmer.com
The most up-to-date information on part numbers, product brochures, systems modules, spare
parts, and application notes are in the PerkinElmer website.
Telephone:
• U.S. only: Call toll free 1-800-762-4000, 8:30 a.m. to 7 p.m. EST. Your order will be
shipped usually within 24 hours.