Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

06 Ion Pumps

Download as pdf or txt
Download as pdf or txt
You are on page 1of 53

AGILENT

ION Pumps

2-3 Ion Pumps Features and Benefits


4,5 Controllers Features and Benefits
6,7 SEM Pump Controller Features
and Benefits
8-11 Typical Applications
12-40 Pump Models and Controllers
41-51 Technical Notes
52-53 Service and Support Plan
Agilent Ion Pumps Features and Benefits

Ion Pump Evolution


● Since the late 1950’s, when
the ion pump was invented at
Varian, now Agilent Technologies,
many changes and technical Wide Pumping Speed Range
improvements have taken place. • Miniature/Appendage pumps from
Virtually all of the major innovations have come from us, from the first 0.2 to 10 l/s
Diode VacIon pump to the Triode, then to the StarCell series pumps, and • Small/Medium pumps from 20 to 75 l/s
the VacIon Plus. • Large size pumps from 150 to 2500 l/s
• TSP/Combi pumps
VacIon Plus • Custom solutions with special pumping
speed, size, magnetic field
● VacIon Plus is a complete family of ion pumps, controllers, options,
and accessories, designed to provide solutions to every application.
Parameters such as operating pressure, the gas mixture to be pumped,
the starting pressure, etc. can vary so dramatically that Varian, today
Agilent, decided to develop dedicated ion pump solutions (including
controllers and all other accessories) for different applications.
● The VacIon Plus family includes Diode, Noble Diode, and StarCell pump
versions that allow Agilent to provide the best technology for each field
of application. The family is complemented by the new 4UHV Ion Pump
controllers, that provide different power levels and interface capabilities. Vacuum Processing
In order to ensure cleanliness, all pumps are:
Titanium Sublimation Combination Pumps (TSP) • Factory processed at high temperature
(450°) in ultra-high vacuum for a
● The titanium sublimation creates extra high getterable gas pumping thorough outgassing of the body and all
speed while the ion pumping mechanisms handle the non-getterable internal components
gases such as argon and methane. • Shipped under vacuum, and an RGA
The combination pump includes the cylindrical cryopanel and TSP source spectrum can be provided with
mounted to the extra port. each pump
Customized pump configurations are also available. • HE-Leak check

Application
Specific
Solutions
for SEM: a complete line of Element Cells and Insulators
ion getter pumps dedicated to • Cells’ sizes and geometries are
Electron Microscopy. optimized in order to:
- maximize the discharge intensity
- maximize the pumping speed
• The special design of the ceramic
insulators allows:
- no buildup of sputtered
conductive coating
- longer pump life

2
ION Pumps

Custom Design
The pump body can be configured to meet
optional requirements including:
• Cryopanel and TSP side or bottom
mounted
• Integral heaters
• Additional roughing ports

Pumps
Ion
Feedthroughs
• Eliminate corrosion
• Implement the “High Voltage
Cable Interlock”
• Provide an easy connection
• Prevent unintentional extraction
• Minimize overall dimensions

Pumping Elements Cables Heaters


Three different types of pumping elements • Agilent cables have an “HV Safety • The pump can be supplied with heaters
are available to cover all possible gas Interlock” that prevents any chance of designed to perform the pump baking
mixtures and optimize the application electrical shock • Minimize operational costs
specific performances:
• If the cable is disconnected from the
• Diode pump, the voltage is automatically cut off
• Noble Diode • Available in different lengths
• Our unique StarCell • Robust, flexible metal-shielded cable
• 107 Gy radiation tolerance

3
The New Agilent 4UHV Ion Pump Controller Features and Benefits

More Choice and Flexibility


● The VacIon Plus pump family is complemented by the new 4UHV Ion Pump
controller, that provides different power levels and interface capabilities.
● The compact MiniVac controller and a dedicated TSP controller
are also available.
● The new series of IPCU controller units completes Agilent’s offer
(see next pages).
● The existing range of ion pump controllers offers more choice and flexibility
than ever before. With the latest in design features, they are simple and easy
to operate; just select the right controller to fit your specific application.

Versatility
The 4UHV is available in different 4UHV - For Ultra High Vacuum
configurations, in order to independently
power, control and monitor any combination The new state-of-the-art Agilent 4UHV Ion Pump Controller operates up to four
of multiple pumps of different sizes, from pumps simultaneously and independently. The 4UHV starts and controls ion
one to four pumps, from 10 to 2500 l/s. pumps of any type (Diode, Noble Diode, StarCell) and size (from 10 to 2500 l/s).
For each number of pumps to be operated A large four-line LCD display allows simultaneous reading of individual pump
several options are available, voltage, current and pressure. The variable voltage feature ensures optimum
120 W or 200 W for a single pump, pumping speed and pressure reading throughout the operating pressure range.
2 x 80 W or 2 x 200 W for two pumps,
Built-in set points, remote operation and RS232/485 computer interface are
2 x 80 + 200 W for three pumps,
standard. Ethernet and Profibus are options.
4 x 80 W for four pumps.

Optimized Pumping Speed


The 4UHV will select the right operating applied HV, reduced by the space charge
voltage to optimize the pumping speed of effect due to the electron cloud present in
your ion pumps. By applying High Voltage in the ion pump cell. Since the space charge
accordance with operating pressure, effect is pressure related, a variable HV is
pumping speed performance is improved. applied to maintain optimum bombardment
This is because the energy with which the energy, resulting in the best possible
Intelligence ion bombards the cathode is the nominal pumping performance at any pressure.
To access the unit you can use analog
or RS232/485 ports. The controller uses
the same protocol as our other intelligent Pumping Speed vs Pressure at Different Voltages
vacuum devices (Navigator turbo pump 100
Controller and Inverter scroll & rotary vane
pumps), giving you fast, convenient access
to all elements of the vacuum system.
% of Nominal Speed

Profibus and Ethernet communications


available on request, please call Agilent
50
for details.

0
10 -11 10 -10 10 -9 10 -8 10 -7 10 -6 10 -5 10 -4
Pressure (mbar)

4
ION Pumps

Low Noise
For SEM applications especially, the
remaining AC component of the HV output
was reduced to a minimum. It is much lower
than in any other existing unit, eliminating
the need for additional filters completely in
many cases.

Pumps
Ion
Typical Current vs Pressure Curve

Typical Leakage Current Area


1000 µ with Conventional Controller

100 µ
Current (A)

Safety Pressure Reading


To protect you against high voltage the The 4UHV is preprogrammed to 10 µ
cable is equipped with an interlock system automatically convert current reading 4UHV Reading
which immediately shuts down the high of any Vaclon Plus pump into pressure. 1µ
voltage when the plug is removed from the Thanks to its ability to detect ion current
pump. The protect mode limits the current as low as 10 nA, it allows pressure 0,1 µ
to protect the pump and the controller. measurement in the 10-10 mbar range. 10-10 10-9 10-8 10-7 10-6 10-5
To ensure reliable pressure reading down Pressure (mbar)
to the UHV region, the 4UHV optimizes
the applied high voltage as a function of
pressure. As a result, the leakage current
of the ion pump is eliminated, thereby
providing more accurate pressure readings.

5
Agilent SEM Ion Pumps Features and Benefits

The Agilent Advantage:


Dedicated Solutions for SEM Applications
● Agilent is the only manufacturer to offer specially designed SEM ion
pumps. These pumps are ideal for the high vacuum guns where stable
Wide, Dedicated Range
vacuum and low leakage current is required to control and preserve the
charged particle filament. • A complete range of SEM ion pumps
from 10 to 75 l/s, tailored to your
● The key to this superior performance is Agilent’s patented anode design specific vacuum needs
which uses contoured cells and simplified electrical elements. • Small footprint for easier system
This insures stable current readings and lower particle generation. integration
● When combining the SEM ion pump on the gun with a StarCell ion
pump on the lower column, Agilent ion pumps can offer a powerful
combination optimised for modern E-beam columns.

SEM Ion Pumps are available on request; please ask Agilent


for technical details.

Dedicated Heaters
• Dedicated Heater for every pump size
• The new heaters are designed to perform
a more effective pump baking
• Lower power and operational costs

Innovative SEM Anode Geometry Very Compact Design IPCU3 / IPCU 2 Power Supplies
• Better current stability Improved design for: Two versions: 3 or 2 supply channels
• Lowest leakage current in the industry • Lighter pump weight • Special low noise electronics for better
(< 10 nA) • Fast magnets replacement SEM imaging
• Double Shielded Ceramics • Easy maintenance • Battery backup (optional): up to 30 days
24/7 of battery life
• Longer pump life
• Optional display and front panel
• Longer pressure stability
• Maximum uptime New 4UHV Controller
• Special low noise electronics for better
SEM imaging

6
ION Pumps

Dedicated Magnetic Stray Shields Agilent Feedthrough and Cables Higher Pumping Speed
• External magnetic shields for stray • The HV Safety Interlock prevents any • Optimized magnetic circuit for max.
magnetic field reduction available chance of electrical shock performance in a very compact package
individually on request • The voltage is automatically cut off as • Faster pump down
soon as the cable is disconnected from
the pump
• Safer pump operation

Available in Round Shape

Pumps
Ion
Integrated ion pump/column allows:
• Optimum mass balancing
• Improved pumping conductance
• Compact and modular design
• Simmetric weight distribution for rugged
column integration

RGA Guaranteed Ultimate Vacuum Optical Baffle Battery Power Supply


• The pump is vacuum processed Buit-in Optical Baffle (optional) for: • Enables service without breaking
at 450 °C to outgas most of gases out of • Minimized particle emission vacuum
the pump body • Minimum conductance reduction • Allows for column shipping under
• The pump is shipped under vacuum • Total column protection vacuum
• An RGA (Residual Gas Analysis) • Maximum e-gun life
spectrum is available for each pump,
to guarantee its performance and the
cleanliness of the manufacturing process

7
Typical Applications for Agilent VacIon Plus Pumps

Research and Development


Particle Accelerators & Synchrotron Light Sources
In these machines, electrically charged particles
(electrons for the production of synchrotron light or ions for
particle accelerators) are forced to follow a curved trajectory
in a ring called a storage ring. Charged particles circulate for
hours in the storage ring, at constant energy, in an ultra-high
vacuum environment.
Before their injection into the storage ring, the particles first
have to be accelerated inside an injection system composed
of one or two accelerators (the Linac and the Booster).
All along their path within the machine, the particles
(electrons or ions) have to circulate inside a vacuum chamber.
Otherwise, they would collide with the air molecules and
would be absorbed very rapidly.
• Linac
The linac is a linear accelerator. The charged particles enter
into a first RF cavity which accelerates them and at the
same time groups them into bunches. They are then
accelerated by a succession of RF cavities throughout the
length of the linac. Vacuum within the linac can be created
Courtesy: LBNL Advanced Light Source. by Agilent VacIon Plus pumps from 20 l/s to 70 l/s.
• Booster
Charged particles, which have already been accelerated in
the linac, are accelerated even more strongly by the
booster. The acceleration is produced by RF cavities
through which the charged particles pass many times,
gaining in energy at each pass. Once the level of maximum
energy has been reached, the beam of particles is
transferred from the booster to the storage ring.
Vacuum in the booster is generally produced by small pumps.
Small Agilent VacIon Plus pumps fit this application perfectly.
• Storage Ring
Charged particles circulate inside the storage ring at
constant energy. All along the ring there are curved
sections as well as straight sections. The storage ring is
placed inside a tunnel with very thick concrete walls in
order to contain emitted radiation in case of beam loss.
Ultra high vacuum is an absolute necessity in this part of
the machine since the particles travel through the storage
ring for hours. The less residual gas there is, the more
focused the beam remains. Large Agilent VacIon Plus
pumps, in the 300 - 500 l/s range are used for this
demanding application.
• Front Ends
Courtesy PSI SLS. The front end is the pipe work which transports the
particles under a vacuum from the extraction zone up to the
beamline outside of the tunnel of the ring. There you can
find a beam shutter as well as devices allowing the
isolation of the vacuum of the ring from that of the beamline,
which is often of lower pressure. Agilent large pumps, as in
for the storage ring, can be used in this part of the machine.
8
ION Pumps

Courtesy of MedAustron. Courtesy Pacific Northwest National Laboratory.

Pumps
Ion
• Beam Lines • Miscellaneous Projects
The experimental hall, around the storage ring, houses the Some fundamental research projects that use very sensitive
beamlines built tangentially to the ring. The beamlines are equipment (necessitating ultra-high vacuum with no
usually specialized in a field of research (such as biology, mechanical vibration) will find the solution in Agilent
polymers, and magnetism) or an experimental method (such VacIon Plus pumps. The new gravitational waves detectors
as diffraction, EXAFS, and imaging). Some of the longest (GWD) such as VIRGO in Italy and LIGO in the USA use
beamlines are built outside the experimental hall. Agilent pumps to produce and maintain the required
Generally, large pumps are used in this part, from 300 l/s to vacuum.
500 l/s. They can be combined with TSP and cryopanel in
order to pump even the lightest molecules.

Courtesy P. Ginter - ESRF Grenoble. Courtesy P. Ginter - ESRF Grenoble.

9
Agilent Ion Pumps Typical Applications

Mass Spectrometry Nanotechnologies


• Analytical systems that use focused charged particle • Agilent`s line of high performance VacIon ion pumps are
beams (CPB) and certain types of mass spectrometers such well suited for the vacuum requirements of
as magnetic sector or Fournier Transfer often require Transmission Electron Microscopes (TEM),
ultra-high vacuum. Scanning Electron Microscopes (SEM),
• These applications have very stringent performance Focused Ion Beam (FIB) and Surface
requirements for sensitivity, resolution, sample throughput Analysis equipments.
and measurement repeatability. These requirements are • Agilent is the only manufacturer to offer
driven by the need to analyze ever-smaller samples, SEM application-specific ion pumps with unique
especially in semiconductor, manufacturing, and other anode design
high-tech applications. • The Diode SEM pump with its extremely low leakage
• In general these applications require very clean vacuum current is ideal for the gun section of the column.
pumping, and only VacIon pumps can certify the required • The StarCell pump elment with its unique design is the
level of cleanliness because Agilent is the only ion pump ideal solution for the high pressure operation of the
manufacturer that bakes each pump in a vacuum furnace, columns. StarCell is also the best pump
and supply each pump with an RGA scan. for noble gases or hydrogen.
• Agilent offers a full range of pumps, from 0.2 l/s up to • Agilent completes its offering to the microscope
2500 l/s, as well as combination and custom pumps so manifacture with a full line of controller/power supplies
analytical system designers can meet all their vacuum including the low cost power supplies and the full feature,
requirements from one supplier. multiple controllers.
• Over 50 years of ion pump experience makes Agilent • With the addition of Agilent complete line of oil-free, low
uniquely qualified to supply customized solutions for vibration turbo pumps - ideal for sample chamber vacuum
special applications. requirements – roughing pumps and vacuum gauges,
Agilent can supply all the vacuum components required for
electron microscopes.

Industrial Vacuum Processes


Different applications in industrial sectors such as
telecommunication, defense, medical and others make use of
VacIon pumps to process and maintain essential components
under vacuum. Most of the core equipment in these sectors
requires the use of different electron devices including:
• Microwave tubes and devices
• Power grid tubes
• X-ray imaging tubes and devices
• X-ray sources
In the processing cycle of these types of electron devices,
small ion pumps from 10 l/s to 50 l/s are being used.
Special tube sizes or special applications may require even
bigger ion pumps up to 300-500 l/s. VacIon pumps are often
being used in combination with Turbo Molecular Pumps,
backing pumps and other components out of the wide range
of Agilent products.
After processing, frequently these electron devices are being
equipped with so-called appendage ion pumps in the range of
0.2 l/s to 10 l/s for the purpose of maintaining the electron
tube under vacuum for its operational lifetime.
The first ion pump was invented in the late 1950’s, for the
production of high-quality vacuum tubes used in radar
technology. From this point on, VacIon pumps set the pace in

10
ION Pumps

Courtesy CPI.

• Broadcast
Radio and TV broadcasters and transmitter manufacturers
use electron devices in their equipment to enable high power
transmission or digital broadcasting. Power grid tubes or
microwave tubes are being used in AM radio, FM radio,
VHF TV, UHF TV or digital TV transmitters and amplifiers.
• Medical
Radiological equipment manufacturers use electron devices

Pumps
Ion
for critical components in the radiological chain, such as
Courtesy University of Modena.
X-ray image intensifiers and radiological imaging units for
diagnostics, as well as medical linear accelerators for
radiation therapy. Typical fields of application include:
the industrial field for a wide range of applications using - Medical Imaging (X-ray image tubes and devices)
vacuum processed electron devices used worldwide in: - Radiation Therapy (high power Klystrons, LINAC)
• Space - Magnetic Resonance Imaging.
Electron devices are essential to various space programs, • Defense
from satellite services and earth observation satellites to Microwave tubes and devices are key components in
space probes. Microwave tubes and devices are vital tools, equipment and systems used for different defense applications:
which link people and satellites in order to enable global - Radar (ground based or airborne)
communications. Spaceborne tubes are the power - Electronic Countermeasures (ECM)
generators for the transponders carried on satellites. - Smart Weapons & Electronic Warfare
They retransmit TV or telecommunication signals back to - Missile guidance & Missile seekers
the ground. The electron tubes used for this application are
Travelling Wave Tubes. In these tubes, amplification is • Industrial and Others
produced under vacuum by the interaction between a beam Several industrial processes make use of RF & Microwave
of electrons and the Radio-Frequency (RF) wave. tubes. Some examples are heat treating, Plastic welding,
Food processing, Textile manufacturing, Film curing
• Telecommunications & drying. Other applications make use of X-ray tubes for
Microwave tubes, devices like Traveling Wave Tubes non-destructive testing methods.
(TWTs) and Klystrons (powerful radio vacuum tubes) are
widely used in civil and military telecommunication
networks and equipment used for:
- Satellite and terrestrial communication
- High data-rate transmissions for High Speed Internet and
Wireless Cable
- Broadband high speed data
- Point-to-point and point-to-multipoint microwave links.

11
Agilent Ion Pump Models

Miniature Pump 2 l/s Pump 10 l/s Pump Vaclon Plus 20

Inlet flange 2 ¾” OD CFF (NW 35)

Star Noble
Element type Diode Diode Diode Diode
Cell Diode

Pumping speed (l/s)


(saturated pump at 1-6 mbar) 0.2 2 10 20 22 27
Nitrogen

Operating life (hours) (at 1-6 mbar) N/A 8,000 40,000 80,000 50,000 50,000

Maximum starting pressure (mbar) 1 x 10-4 1 x 10-4 ≤ 1 x 10-4 <10-2 <10-3 <10-3

400 (without magnet) 400 (without magnet)


Maximum baking temperature (°C) 350 350 350 350
150 (with magnet) 150 (with magnet)

Net 0.3 (0.66) Net 0.3 (0.66) Without magnet Net 7 (15)
Weight kg (Ibs)
Shipping 0.6 (1.33) Shipping 0.6 (1.33) 4 (9) Shipping 11 (24)

SEM version available on request – – + +

12
ION Pumps
Vaclon Plus 40 Vaclon Plus 55 Vaclon Plus 75 Vaclon Plus 150 Vaclon Plus 300 Vaclon Plus 500

2 ¾” OD CFF (NW 35) 4 ½” OD CFF (NW 63) 6" OD CFF (NW 100) 6" OD CFF (NW 100) 8" OD CFF (NW 150) 8" OD CFF (NW 150)

Star Noble Star Noble Star Noble Star Noble Star Noble Star Noble
Diode Diode Diode Diode Diode Diode
Cell Diode Cell Diode Cell Diode Cell Diode Cell Diode Cell Diode

34 36 40 50 53 55 65 68 75 125 135 150 240 260 300 410 440 500

80,000 50,000 50,000 80,000 50,000 50,000 80,000 50,000 50,000 80,000 50,000 50,000 80,000 50,000 50,000 80,000 50,000 50,000

<10-2 <10-3 <10-3 <10-2 <10-3 <10-3 <10-2 <10-3 <10-3 <10-2 <10-3 <10-3 <10-2 <10-3 <10-3 <10-2 <10-3 <10-3

350 350 350 350 350 350 350 350 350 350 350 350 350 350 350 350 350 350

Net 17 (37) Net 18 (39) Net 19 (42) Net 43 (94) Net 69 (149) Net 120 (264)
Shipping 21 (46) Shipping 22 (48) Shipping 23 (51) Shipping 53 (110) Shipping 84 (185) Shipping 135 (297)

+ + + – – –

Pumps
Ion

13
Agilent Ion Pump Models

► Agilent VacIon Plus 20

2 3/4" CFF (NW35)

152 (6.00) REF


24 (0.90) REF 133 (5.20) REF

224 (8.80) REF 121 (4.80) REF

Dimensions: millimeters (inches)

Technical Specifications
StarCell Noble Diode Diode
Nominal pumping speed for Nitrogen (*) (l/s) 20 22 27
Operating life at 1x10-6 mbar (hours) 80,000 50,000 50,000
Maximum starting pressure (mbar) ≤ 5x10-2 ≤ 1x10-3 ≤ 1x10-3
Ultimate pressure Below 10-11
Inlet flange 2 ¾” CFF (NW 35) AISI 304 ESR SST
Maximum baking temperature (°C) 350
Weight, kg (Ibs) (with ferrite magnet) Net 7 (15), Shipping 11 (24)
(*) Tested according to ISO/DIS 3556-1-1992

SEM version available on request

14
ION Pumps
VacIon Plus 20 - Pumping Speed vs Pressure

40

30
Pumping Speed (l/s)

20

10

0
10 -11 10 -10 10 -9 10 -8 10 -7 10 -6 10 -5 10 -4
Pressure (mbar)

Nitrogen unsaturated Diode Nitrogen saturated Diode Argon saturated StarCell

Ordering Information
Description Part Number Description Part Number

Pumps
Ion
Pumps HV Cables
Diodewith Ferrite magnets 9191115 HV Bakeable cable, radiation resistant,
Diode without magnets 9191114 4 m (13 ft.) long, with interlock 9290705
StarCell with Ferrite magnets 9191145 HV Bakeable cable, radiation resistant,
StarCell with Sm-Co magnets 9191146 7 m (23 ft.) long, with interlock 9290707
StarCell without magnets 9191144 HV Bakeable cable, radiation resistant,
10 m (33 ft.) long, with interlock 9290708
Noble Diode versions available on request HV Bakeable cable, radiation resistant,
20 m (66 ft.) long, with interlock 9290709
4UHV Controller
120 W negative 9299100 Replacement Parts
120 W positive 9299101 HV Feedthrough with interlock 9595125
200 W neg 9299010 VacIon Plus 20 Diode Ferrite magnet assembly 9191001
200 W pos 9299011 VacIon Plus 20 Noble Diode Ferrite magnet assembly 9191002
2 x 80 W neg 9299200 VacIon Plus 20 StarCell Ferrite magnet assembly 9191004
2 x 80 W pos 9299201 Heaters*  (Input Power 140 W) 120 V 9191110
2 x 200 W neg 9299020 Heaters*  (Input Power 140 W) 220 V 9191111
2 x 200 W pos 9299021 * To order heaters for replacement or upgrading of existing pumps,
1 x 200 W pos & 1 x 200 W neg 9299022 please contact your local Agilent Vacuum Technologies representative.
4 x 80 W neg 9299400
4 x 80 W pos 9299401
2 x 80 W pos & 2 x 80 W neg 9299402
2 x 80 W neg & 1 x 200 W neg 9299210
2 x 80 W pos & 1 x 200 W pos 9299211 Diode, Noble Diode positive
2 x 80 W pos & 1 x 200 W neg 9299212 StarCell, Triode negative
2 x 80 W neg & 1 x 200 W pos 9299213 (see page 42)

15
Agilent Ion Pump Models

► Agilent VacIon Plus 40 2 3/4" CFF (NW35)

217.50 (8.56) REF

57 (2.24) REF
160 (6.30) REF = = 60 (2.36) REF = =

4 HOLES M6
310 (12.20) REF

160.50 (6.32) REF

136.50 (5.37) REF


= =
Dimensions: millimeters (inches)

Technical Specifications
StarCell Noble Diode Diode
Nominal pumping speed for Nitrogen (*) (l/s) 34 36 40
Operating life at 1x10-6 mbar (hours) 80,000 50,000 50,000
Maximum starting pressure (mbar) ≤ 5x10-2 ≤ 1x10-3 ≤ 1x10-3
Ultimate pressure Below 10-11
Inlet flange 2 ¾” CFF (NW 35) AISI 304 ESR SST
Maximum baking temperature (°C) 350
Weight, kg (Ibs) 17 (37)
(*) Tested according to ISO/DIS 3556-1-1992

SEM version available on request

16
ION Pumps
VacIon Plus 40 - Pumping Speed vs Pressure

60

45
Pumping Speed (l/s)

30

15

0
10 -11 10 -10 10 -9 10 -8 10 -7 10 -6 10 -5 10 -4
Pressure (mbar)

Nitrogen unsaturated Diode Nitrogen saturated Diode Argon saturated StarCell

Ordering Information
Description Part Number Description Part Number

Pumps
Ion
Pumps HV Cables
Diode 9191210 HV Bakeable cable, radiation resistant,
Diode with additional 2 ¾” CFF port 9191213 4 m (13 ft.) long, with interlock 9290705
Diode without magnets 9191214 HV Bakeable cable, radiation resistant,
StarCell 9191240 7 m (23 ft.) long, with interlock 9290707
StarCell with additional 2 ¾” CFF port 9191243 HV Bakeable cable, radiation resistant,
10 m (33 ft.) long, with interlock 9290708
StarCell without magnets 9191244
HV Bakeable cable, radiation resistant,
Noble Diode versions available on request 20 m (66 ft.) long, with interlock 9290709
4UHV Controller Replacement Parts
120 W negative 9299100 HV Feedthrough with interlock 9595125
120 W positive 9299101 Heaters*/**  (Input Power 250 W) 120 V 9190071
200 W neg 9299010 Heaters*/**  (Input Power 250 W) 220 V 9190070
200 W pos 9299011 * To order heaters for replacement or upgrading of existing pumps,
2 x 80 W neg 9299200 please contact your local Agilent Vacuum Technologies representative.
2 x 80 W pos 9299201 ** cCSAus marked version available on request.
2 x 200 W neg 9299020
2 x 200 W pos 9299021
1 x 200 W pos & 1 x 200 W neg 9299022
Diode, Noble Diode positive
4 x 80 W neg 9299400 StarCell, Triode negative
4 x 80 W pos 9299401 (see page 42)
2 x 80 W pos & 2 x 80 W neg 9299402
2 x 80 W neg & 1 x 200 W neg 9299210
2 x 80 W pos & 1 x 200 W pos 9299211
2 x 80 W pos & 1 x 200 W neg 9299212
2 x 80 W neg & 1 x 200 W pos 9299213

17
Agilent Ion Pump Models

► Agilent VacIon Plus 55 41/2" CFF (NW63)

231.66 (9.12) REF

71.16 (2.80) REF


160 (6.30) REF 4 HOLES M6 = = 60 (2.36) REF

310 (12.20) REF

160.50 (6.32) REF

136.50 (5.37) REF


Dimensions: millimeters (inches)

Technical Specifications
StarCell Noble Diode Diode
Nominal pumping speed for Nitrogen (*) (l/s) 50 53 55
Operating life at 1x10-6 mbar (hours) 80,000 50,000 50,000
Maximum starting pressure (mbar) ≤ 5x10-2 ≤ 1x10-3 ≤ 1x10-3
Ultimate pressure Below 10-11
Inlet flange 4 ¾” CFF (NW 63) AISI 304 ESR SST
Maximum baking temperature (°C) 350
Weight, kg (Ibs) 18 (39)
(*) Tested according to ISO/DIS 3556-1-1992

SEM version available on request

18
ION Pumps
VacIon Plus 55 - Pumping Speed vs Pressure

100

75
Pumping Speed (l/s)

50

25

0
10 -11 10 -10 10 -9 10 -8 10 -7 10 -6 10 -5 10 -4
Pressure (mbar)

Nitrogen unsaturated Diode Nitrogen saturated Diode Argon saturated StarCell

Ordering Information
Description Part Number Description Part Number

Pumps
Ion
Pumps HV Cables
Diode 9191310 HV Bakeable cable, radiation resistant,
Diode with additional 2 ¾” CFF port 9191313 4 m (13 ft.) long, with interlock 9290705
Diode without magnets 9191314 HV Bakeable cable, radiation resistant,
StarCell 9191340 7 m (23 ft.) long, with interlock 9290707
StarCell with additional 2 ¾” CFF port 9191343 HV Bakeable cable, radiation resistant,
10 m (33 ft.) long, with interlock 9290708
StarCell without magnets 9191344
HV Bakeable cable, radiation resistant,
Noble Diode versions available on request 20 m (66 ft.) long, with interlock 9290709
4UHV Controller Replacement Parts
120 W negative 9299100 HV Feedthrough with interlock 9595125
120 W positive 9299101 Heaters*/**  (Input Power 250 W) 120 V 9190071
200 W neg 9299010 Heaters*/**  (Input Power 250 W) 220 V 9190070
200 W pos 9299011 * To order heaters for replacement or upgrading of existing pumps,
2 x 80 W neg 9299200 please contact your local Agilent Vacuum Technologies representative.
2 x 80 W pos 9299201 ** cCSAus marked version available on request.
2 x 200 W neg 9299020
2 x 200 W pos 9299021
1 x 200 W pos & 1 x 200 W neg 9299022
Diode, Noble Diode positive
4 x 80 W neg 9299400 StarCell, Triode negative
4 x 80 W pos 9299401 (see page 42)
2 x 80 W pos & 2 x 80 W neg 9299402
2 x 80 W neg & 1 x 200 W neg 9299210
2 x 80 W pos & 1 x 200 W pos 9299211
2 x 80 W pos & 1 x 200 W neg 9299212
2 x 80 W neg & 1 x 200 W pos 9299213

19
Agilent Ion Pump Models

► Agilent VacIon Plus 75 6" CFF (NW100)

242.10 (9.53) REF

81.60 (3.21) REF


160 (6.30) REF = = 4 HOLES M6 = = 60 (2.36) REF

310 (12.20) REF

160.50 (6.32) REF

136.50 (5.37) REF


= =
Dimensions: millimeters (inches)

Technical Specifications
StarCell Noble Diode Diode
Nominal pumping speed for Nitrogen (*) (l/s) 65 68 75
Operating life at 1x10-6 mbar (hours) 80,000 50,000 50,000
Maximum starting pressure (mbar) ≤ 5x10-2 ≤ 1x10-3 ≤ 1x10-3
Ultimate pressure Below 10-11
Inlet flange 6” CFF (NW 100) AISI 304 ESR SST
Maximum baking temperature (°C) 350
Weight, kg (Ibs) 19 (42)
(*) Tested according to ISO/DIS 3556-1-1992

SEM version available on request

20
ION Pumps
VacIon Plus 75 - Pumping Speed vs Pressure

160

120
Pumping Speed (l/s)

80

40

0
10 -11 10 -10 10 -9 10 -8 10 -7 10 -6 10 -5 10 -4
Pressure (mbar)

Nitrogen unsaturated Diode Nitrogen saturated Diode Argon saturated StarCell

Ordering Information
Description Part Number Description Part Number

Pumps
Ion
Pumps HV Cables
Diode 9191410 HV Bakeable cable, radiation resistant,
Diode with additional 2 ¾” CFF port 9191413 4 m (13 ft.) long, with interlock 9290705
Diode without magnets 9191414 HV Bakeable cable, radiation resistant,
StarCell 9191440 7 m (23 ft.) long, with interlock 9290707
StarCell with additional 2 ¾” CFF port 9191443 HV Bakeable cable, radiation resistant,
10 m (33 ft.) long, with interlock 9290708
StarCell without magnets 9191444
HV Bakeable cable, radiation resistant,
Noble Diode versions available on request 20 m (66 ft.) long, with interlock 9290709
4UHV Controller Replacement Parts
120 W negative 9299100 HV Feedthrough with interlock 9595125
120 W positive 9299101 Heaters*/**  (Input Power 250 W) 120 V 9190071
200 W neg 9299010 Heaters*/**  (Input Power 250 W) 220 V 9190070
200 W pos 9299011 * To order heaters for replacement or upgrading of existing pumps,
2 x 80 W neg 9299200 please contact your local Agilent Vacuum Technologies representative.
2 x 80 W pos 9299201 ** cCSAus marked version available on request.
2 x 200 W neg 9299020
2 x 200 W pos 9299021
1 x 200 W pos & 1 x 200 W neg 9299022
Diode, Noble Diode positive
4 x 80 W neg 9299400 StarCell, Triode negative
4 x 80 W pos 9299401 (see page 42)
2 x 80 W pos & 2 x 80 W neg 9299402
2 x 80 W neg & 1 x 200 W neg 9299210
2 x 80 W pos & 1 x 200 W pos 9299211
2 x 80 W pos & 1 x 200 W neg 9299212
2 x 80 W neg & 1 x 200 W pos 9299213

21
Agilent Ion Pump Models

► Agilent VacIon Plus 150 6" O.D. CFF (NW100)

47.80 (1.88) REF


362.80 (14.28) REF
100 (3.94) REF THREADING 1/2"-20 2B 170 (6.69) REF

272.50 (10.73) REF

205 (8.07) REF

244.50 (9.62) REF


Dimensions: millimeters (inches)

Technical Specifications
StarCell Noble Diode Diode
Nominal pumping speed for Nitrogen (*) (l/s) 125 135 150
Operating life at 1x10-6 mbar (hours) 80,000 50,000 50,000
Maximum starting pressure (mbar) ≤ 5x10-2 ≤ 1x10-3 ≤ 1x10-3
Ultimate pressure Below 10-11
Inlet flange 6” CFF (NW 100) AISI 304 ESR
Maximum baking temperature (°C) 350
Weight, kg (Ibs) 43 (94)
(*) Tested according to ISO/DIS 3556-1-1992

22
ION Pumps
VacIon Plus 150 - Pumping Speed vs Pressure

300

Pumping Speed (l/s) 250

200

150

100

50

0
10 -11 10 -10 10 -9 10 -8 10 -7 10 -6 10 -5 10 -4
Pressure (mbar)

Nitrogen unsaturated Diode Nitrogen saturated Diode Argon saturated StarCell

Ordering Information
Description Part Number Description Part Number

Pumps
Ion
Pumps 4UHV Controller (cont’d)
Diode 9191510 2 x 80 W neg & 1 x 200 W neg 9299210
Diode with heaters installed 120 V 9191511 2 x 80 W pos & 1 x 200 W pos 9299211
Diode with heaters installed 220 V 9191512 2 x 80 W pos & 1 x 200 W neg 9299212
Diode double-ended 9191550 2 x 80 W neg & 1 x 200 W pos 9299213
Diode double-ended with heaters installed 120 V 9191551
HV Cables
Diode double-ended with heaters installed 220 V 9191552
HV Bakeable cable, radiation resistant,
StarCell 9191540 4 m (13 ft.) long, with interlock 9290705
StarCell with heaters installed 120 V 9191541 HV Bakeable cable, radiation resistant,
StarCell with heaters installed 220 V 9191542 7 m (23 ft.) long, with interlock 9290707
StarCell double-ended 9191580 HV Bakeable cable, radiation resistant,
StarCell double-ended 10 m (33 ft.) long, with interlock 9290708
with heaters installed 120 V 9191581 HV Bakeable cable, radiation resistant,
StarCell double-ended 20 m (66 ft.) long, with interlock 9290709
with heaters installed 220 V 9191582
Replacement Parts
Noble Diode versions available on request HV Feedthrough with interlock 9595125
4UHV Controller Pumping element*  Diode 9199040
120 W negative 9299100 Pumping element*  Noble Diode 9199045
120 W positive 9299101 Pumping element*  StarCell 9199030
200 W neg 9299010 Heaters**  (Input Power 480 W) 120 V 9190073
200 W pos 9299011 Heaters**  (Input Power 480 W) 220 V 9190072
2 x 80 W neg 9299200 * Quantity required: 2.
2 x 80 W pos 9299201 ** To order heaters for replacement or upgrading of existing pumps,
2 x 200 W neg 9299020 please contact your local Agilent Vacuum Technologies
2 x 200 W pos 9299021 representative.
1 x 200 W pos & 1 x 200 W neg 9299022
4 x 80 W neg 9299400 Diode, Noble Diode positive
4 x 80 W pos 9299401 StarCell, Triode negative
2 x 80 W pos & 2 x 80 W neg 9299402 (see page 42)

23
Agilent Ion Pump Models

► Agilent VacIon Plus 300 8" O.D. CFF (NW150)

55.20 (2.17) REF


344.70 (13.57) REF
130 (5.11) REF 170 (6.69) REF

THREADING 1/2"-20
450 (17.71) REF

244.50 (9.62) REF


Dimensions: millimeters (inches)

Technical Specifications
StarCell Noble Diode Diode
Nominal pumping speed for Nitrogen (*) (l/s) 240 260 300
Operating life at 1x10-6 mbar (hours) 80,000 50,000 50,000
Maximum starting pressure (mbar) ≤ 5x10-2 ≤ 1x10-3 ≤ 1x10-3
Ultimate pressure Below 10-11
Inlet flange 8” CFF (NW 150) AISI 304 ESR
Maximum baking temperature (°C) 350
Weight, kg (Ibs) 69 (149)
(*) Tested according to ISO/DIS 3556-1-1992

24
ION Pumps
VacIon Plus 300 - Pumping Speed vs Pressure

600

Pumping Speed (l/s) 500

400

300

200

100

0
10 -11 10 -10 10 -9 10 -8 10 -7 10 -6 10 -5 10 -4
Pressure (mbar)

Nitrogen unsaturated Diode Nitrogen saturated Diode Argon saturated StarCell

Ordering Information
Description Part Number Description Part Number

Pumps
Ion
Pumps 4UHV Controller (cont’d)
Diode 9191610 2 x 80 W neg & 1 x 200 W neg 9299210
Diode with heaters installed 120 V 9191611 2 x 80 W pos & 1 x 200 W pos 9299211
Diode with heaters installed 220 V 9191612 2 x 80 W pos & 1 x 200 W neg 9299212
Diode double-ended 9191650 2 x 80 W neg & 1 x 200 W pos 9299213
Diode double-ended with heaters installed 120 V 9191651
HV Cables
Diode double-ended with heaters installed 220 V 9191652
HV Bakeable cable, radiation resistant,
StarCell 9191640 4 m (13 ft.) long, with interlock 9290705
StarCell with heaters installed 120 V 9191641 HV Bakeable cable, radiation resistant,
StarCell with heaters installed 220 V 9191642 7 m (23 ft.) long, with interlock 9290707
StarCell double-ended 9191680 HV Bakeable cable, radiation resistant,
StarCell double-ended 10 m (33 ft.) long, with interlock 9290708
with heaters installed 120 V 9191681 HV Bakeable cable, radiation resistant,
StarCell double-ended 20 m (66 ft.) long, with interlock 9290709
with heaters installed 220 V 9191682
Replacement Parts
Noble Diode versions available on request HV Feedthrough with interlock 9595125
4UHV Controller Pumping element*  Diode 9199040
120 W negative 9299100 Pumping element*  Noble Diode 9199045
120 W positive 9299101 Pumping element*  StarCell 9199030
200 W neg 9299010 Heaters**  (Input Power 580 W) 120 V 9190075
200 W pos 9299011 Heaters**  (Input Power 580 W) 220 V 9190074
2 x 80 W neg 9299200 * Quantity required: 4.
2 x 80 W pos 9299201 ** To order heaters for replacement or upgrading of existing pumps,
2 x 200 W neg 9299020 please contact your local Agilent Vacuum Technologies
2 x 200 W pos 9299021 representative.
1 x 200 W pos & 1 x 200 W neg 9299022
4 x 80 W neg 9299400 Diode, Noble Diode positive
4 x 80 W pos 9299401 StarCell, Triode negative
2 x 80 W pos & 2 x 80 W neg 9299402 (see page 42)

25
Agilent Ion Pump Models

► Agilent VacIon Plus 500

107.25 (4.22) REF


8" O.D. CFF (NW150)

46.25 (1.82) REF


524.70 (20.65) REF

350 (13.78) REF


130 (5.11) REF 230 (9.05) REF

450 (17.71) REF THREADING 1/2"-20


145 (5.7) REF

296.8 (11.68) REF


305 (12) REF Dimensions: millimeters (inches)

Technical Specifications
StarCell Noble Diode Diode
Nominal pumping speed for Nitrogen (*) (l/s) 410 440 500
Operating life at 1x10-6 mbar (hours) 80,000 50,000 50,000
Maximum starting pressure (mbar) ≤ 5x10-2 ≤ 1x10-3 1x10-3
Ultimate pressure Below 10-11
Inlet flange 8” CFF (NW 150) AISI 304 ESR
Maximum baking temperature (°C) 350
Weight, kg (Ibs) 120 (264)
(*) Tested according to ISO/DIS 3556-1-1992

26
ION Pumps
VacIon Plus 500 - Pumping Speed vs Pressure

800

700

600
Pumping Speed (l/s)

500

400

300

200

100

0
10 -11 10 -10 10 -9 10 -8 10 -7 10 -6 10 -5 10 -4
Pressure (mbar)

Nitrogen unsaturated Diode Nitrogen saturated Diode Argon saturated StarCell

Ordering Information
Description Part Number Description Part Number

Pumps
Ion
Pumps 4UHV Controller (cont’d)
Diode 9191710 2 x 80 W neg & 1 x 200 W neg 9299210
Diode with heaters installed 120 V 9191711 2 x 80 W pos & 1 x 200 W pos 9299211
Diode with heaters installed 220 V 9191712 2 x 80 W pos & 1 x 200 W neg 9299212
Diode double-ended 9191750 2 x 80 W neg & 1 x 200 W pos 9299213
Diode double-ended with heaters installed 120 V 9191751
HV Cables
Diode double-ended with heaters installed 220 V 9191752
HV Bakeable cable, radiation resistant,
StarCell 9191740 4 m (13 ft.) long, with interlock 9290705
StarCell with heaters installed 120 V 9191741 HV Bakeable cable, radiation resistant,
StarCell with heaters installed 220 V 9191742 7 m (23 ft.) long, with interlock 9290707
StarCell double-ended 9191780 HV Bakeable cable, radiation resistant,
StarCell double-ended 10 m (33 ft.) long, with interlock 9290708
with heaters installed 120 V 9191781 HV Bakeable cable, radiation resistant,
StarCell double-ended 20 m (66 ft.) long, with interlock 9290709
with heaters installed 220 V 9191782
Replacement Parts
Noble Diode versions available on request HV Feedthrough with interlock 9595125
4UHV Controller Pumping element*  Diode 9199040
120 W negative 9299100 Pumping element*  Noble Diode 9199045
120 W positive 9299101 Pumping element*  StarCell 9199030
200 W neg 9299010 Heaters**  (Input Power 780 W) 120 V 9190077
200 W pos 9299011 Heaters**  (Input Power 780 W) 220 V 9190076
2 x 80 W neg 9299200 * Quantity required: 8.
2 x 80 W pos 9299201 ** To order heaters for replacement or upgrading of existing pumps,
2 x 200 W neg 9299020 please contact your local Agilent Vacuum Technologies
2 x 200 W pos 9299021 representative.
1 x 200 W pos & 1 x 200 W neg 9299022
4 x 80 W neg 9299400 Diode, Noble Diode positive
4 x 80 W pos 9299401 StarCell, Triode negative
2 x 80 W pos & 2 x 80 W neg 9299402 (see page 42)

27
Agilent Ion Pump Models

► Agilent VacIon Plus Agilent VacIon Plus 2500 l/s

Large Size Pumps

91 (3.58) REF
161/2" CFF O.D. x 14" I.D.

760 (29.92) REF


30 (1.18) REF
(2.75)
5/8"-18 UNF

REF
70
8" CFF x 6" 2 3/4" CFF x 11/2"

857.5 (33.76) REF


450 (17.71) REF
Larger Ion Pumps up to 2500 l/s, in StarCell and Diode
370 (14.56) REF
versions, are available on request. Please contact Agilent for 857.5 (33.76) REF
technical specifications and ordering information.

Agilent VacIon Plus 960 l/s


12" CFF

HV FISCHER
166 (6.53) REF
686 (27.00) REF

450 (17.72) REF

48 (1.89) REF

Ø 22.0 (0.86)
==

640 (25.20) REF

672 (26.45) REF


250 (9.84) REF
42.5 (1.67) REF
69.8 (2.75) REF

320 (12.60) REF

640 (25.20) REF M20x16

672 (26.45) REF


23.8 (0.93)

250 (9.84) REF


18 (0.71)
REF

REF

= =
350 (13.78) REF
= =
2.5 (0.10) REF 635 (25.00) REF
465.9 (18.34) REF

= =

M20x16

Dimensions: millimeters (inches)

28
ION Pumps
► Agilent VacIon Plus Vaclon Plus 150 Combination Pump (side-mounted TSP)

Combination Pumps 6" O.D. CFF (NW100)

123.50 (4.86) REF

337.80 (13.30) REF


1/2"-20 2B

25.00 (.98) REF


100.00 (3.94) REF 170.00 (6.69) REF
245.50 (9.67) REF

205 (8.07) REF


124.35
(4.90) REF 200 (7.87) REF 272.50 (10.73) REF

Vaclon Plus 300 Combination Pump (side-mounted TSP)


8" O.D. CFF (NW150)

Titanium Sublimation Combination Pumps


348.65 (13.73) REF

(2.18) REF
55.45

• Ion-Sublimation combination pumps have been a popular


choice for many years for creating ultra high vacuum

Pumps
environments. The titanium sublimation creates extra high

Ion
getterable gas pumping speed while the ion pumping 130 (5.12) REF 170
(4.86) REF

N4 HOLES Ø1/2"-20 UNC (6.69) REF


123.50

mechanisms handle the non-getterable gases such as 244.45


(9.62) REF
148.50
(5.85) REF
124.35
(4.90) REF

argon and methane.


• This combination pump is a VacIon Plus 150, 300, or 500
with an extra side or bottom-mounted 8” ConFlat port.
The combination pump includes the cylindrical cryopanel
and TSP source mounted to the extra port. Getterable gases
enter the end of the cylindrical cryopanel and are pumped
by being combined with the freshly-deposited titanium 449.92 (17.71) REF

there. Liquid nitrogen cooling the cryopanel increases the


efficiency of the gettering process and adds greatly to the
water pumping speed.
• The Agilent VacIon Plus series combination pumps allow Vaclon Plus 500 Combination Pump (side-mounted TSP)
addition of a cryopanel from the bottom of the pump or
from the side. This can be a significant advantage in 8" O.D. CFF (NW150)

situations where height restrictions are present.


367.95 (14.49) REF

Customized pump configurations are also available.


525 (20.67) REF

Technical Specifications
142
(5.59) REF
Nominal net pumping speed at 20 °C (l/s) with StarCell
450 (17.72) REF 297 (11.69) REF
124.35
elements (water cooled cryopanel) (4.90) REF

VacIon Plus 150 N2 – 610


H2 – 1,380
VacIon Plus 300 N2 – 720
H2 – 1,580
VacIon Plus 500 N2 – 880
H2 – 1,930

29
Agilent Ion Pump Models

► Agilent VacIon Plus


Combination Pumps

Vaclon Plus 300 Combination Pump (bottom-mounted TSP) Vaclon Plus 500 Combination Pump (bottom-mounted TSP)
8" O.D. CFF (NW150)
8" O.D. CFF (NW150)
388.40 (15.29) REF

564.50 (22.22) REF


124.35 (4.90) REF

124.35
(4.89)
REF
450 (17.72) REF
450 (17.72) REF
244.50 (9.63) REF

296.80 (11.69) REF


305 (12.01) REF

Combination Pumps Pumping Speed

2000
Vaclon Plus 500
1750
Vaclon Plus 300
1500
Pumping Speed (l/s)

1250 Vaclon Plus 150

1000
Vaclon Plus 500
750 Vaclon Plus 300

500 Vaclon Plus 150

250

0
10 -11 10 -10 10 -9 10 -8 10 -7 10 -6 10 -5 10 -4
Pressure (mbar)

Nitrogen StarCell Hydrogen StarCell

30
ION Pumps
Ordering Information
The VacIon Plus 500, 300, and 150 pumps can be supplied with the sublimation cryopanel factory-installed and TSP cartridge
included. Cables and controllers are to be ordered separately. For basic pump part number, see pages 22 - 27.

Description Part Number


Vaclon Plus 500, 300, or 150 combination pump
VacIon Plus 150 Diode, with Side-Mounted Cryopanel, with TSP Cartridge and with Installed Heater 120V 9192510
VacIon Plus 150 Diode, with Side-Mounted Cryopanel, with TSP Cartridge and with Installed Heater 220V 9192511
VacIon Plus 150 Noble Diode, with Side-Mounted Cryopanel, with TSP Cartridge and with Installed Heater 120V 9192520
VacIon Plus 150 Noble Diode, with Side-Mounted Cryopanel, with TSP Cartridge and with Installed Heater 220V 9192521
VacIon Plus 150 Starcell, with Side-Mounted Cryopanel, with TSP Cartridge and with Installed Heater 120V 9192540
VacIon Plus 150 Starcell, with Side-Mounted Cryopanel, with TSP Cartridge and with Installed Heater 220V 9192541
VacIon Plus 300 Diode, with Side-Mounted Cryopanel, with TSP Cartridge and with Installed Heater 120V 9192610
VacIon Plus 300 Diode, with Side-Mounted Cryopanel, with TSP Cartridge and with Installed Heater 220V 9192611
VacIon Plus 300 Noble Diode, with Side-Mounted Cryopanel, with TSP Cartridge and with Installed Heater 120V 9192620
VacIon Plus 300 Noble Diode, with Side-Mounted Cryopanel, with TSP Cartridge and with Installed Heater 220V 9192621
VacIon Plus 300 Starcell, with Side-Mounted Cryopanel, with TSP Cartridge and with Installed Heater 120V 9192640
VacIon Plus 300 Starcell, with Side-Mounted Cryopanel, with TSP Cartridge and with Installed Heater 220V 9192641
VacIon Plus 300 Diode, with Bottom-Mounted Cryopanel, with TSP Cartridge and with Installed Heater 120V 9192612
VacIon Plus 300 Diode, with Bottom-Mounted Cryopanel, with TSP Cartridge and with Installed Heater 220V 9192613
VacIon Plus 300 Noble Diode, with Bottom-Mounted Cryopanel, with TSP Cartridge and with Installed Heater 120V 9192622
VacIon Plus 300 Noble Diode, with Bottom-Mounted Cryopanel, with TSP Cartridge and with Installed Heater 220V 9192623
VacIon Plus 300 Starcell, with Bottom-Mounted Cryopanel, with TSP Cartridge and with Installed Heater 120V 9192642
VacIon Plus 300 Starcell, with Bottom-Mounted Cryopanel, with TSP Cartridge and with Installed Heater 220V 9192643

Pumps
VacIon Plus 500 Diode, with Side-Mounted Cryopanel, with TSP Cartridge and with Installed Heater 120V 9192710

Ion
VacIon Plus 500 Diode, with Side-Mounted Cryopanel, with TSP Cartridge and with Installed Heater 220V 9192711
VacIon Plus 500 Noble Diode, with Side-Mounted Cryopanel, with TSP Cartridge and with Installed Heater 120V 9192720
VacIon Plus 500 Noble Diode, with Side-Mounted Cryopanel, with TSP Cartridge and with Installed Heater 220V 9192721
VacIon Plus 500 Starcell, with Side-Mounted Cryopanel, with TSP Cartridge and with Installed Heater 120V 9192740
VacIon Plus 500 Starcell, with Side-Mounted Cryopanel, with TSP Cartridge and with Installed Heater 220V 9192741
VacIon Plus 500 Diode, with Bottom-Mounted Cryopanel, with TSP Cartridge and with Installed Heater 120V 9192712
VacIon Plus 500 Diode, with Bottom-Mounted Cryopanel, with TSP Cartridge and with Installed Heater 220V 9192713
VacIon Plus 500 Noble Diode, with Bottom-Mounted Cryopanel, with TSP Cartridge and with Installed Heater 120V 9192722
VacIon Plus 500 Noble Diode, with Bottom-Mounted Cryopanel, with TSP Cartridge and with Installed Heater 220V 9192723
VacIon Plus 500 Starcell, with Bottom-Mounted Cryopanel, with TSP Cartridge and with Installed Heater 120V 9192742
VacIon Plus 500 Starcell, with Bottom-Mounted Cryopanel, with TSP Cartridge and with Installed Heater 220V 9192743

Description Weight kg (lbs) Part Number


Replacement Parts and Accessories
TSP Filament Cartridge on 2 ¾” CFF 2.7 (6.0) 9160050
Replacement filaments, package of 12 each 0.4 (2.0) 9160051
Titanium Sublimation Pump Control Unit (Order cables separately) 120V 17.7 (39.0) 9290022
Titanium Sublimation Pump Control Unit (Order cables separately) 220V 17.7 (39.0) 9290023
TSP Cartridge cable, 3.5 m (12 ft.) 9.1 (20.0) 9240730
Sublimation Cryopanel on 8” CFF 10.5 (23.0) 9190180

31
Agilent Ion Pump Controllers

► Agilent Mini Vac Controller


106.40 (4.19) REF
200 (7.87) REF
91.40 (3.60) REF

HIGH P
O
LOAD L

110.50 (4.35) REF


A

128.50 (5.06) REF

122.50 (4.82) REF


10 2 R
I
T
9 Y
O
P 8 U
U T
M 7 P
U
P 3 T
6
C 5
U V
R
R
4 O
L
E 4 T
N 3 A
T G
2 5 E

1 I
mA kV

H.V. O
ON
1V=1mA

MODEL

SERIAL
OUTPUT VOLT
5000 VDC
A Hz

made in Italy
varian Spa Lein Ï
1
6

9
5

120
J001
ACCESSORY

Dimensions: millimeters (inches)

The MiniVac Ion Pump Controller is designed to economically Large pumps (Vaclon Plus 150 to 500) can be operated at any
operate any Vaclon Plus type and size: Diode, Noble Diode, pressure below 2 x 10-6 mbar (continuous operation).
and StarCell, from Miniature to 500 l/s pumps. The MiniVac is The MiniVac is designed to withstand continuous operation at
very compact and light, can be operated in local or remote short circuit conditions, without damaging the ion pump or
mode, and is suitable for high radiation environments. itself. A 24 V battery-operable version is available for
Medium pumps (Vaclon Plus 20 to 75) can be operated at any portable applications.
pressure below 1 x 10-5 mbar (continuous operation).

Technical Specifications
Input Conformity to Norms
90 to 130 VAC or 180 to 240 VAC or 24 VDC Safety: EN61010-1
EMI/EMC: EN61000-4-2, EN61000-4-3, EN61000-4-4
Output
Voltage: ± 5000 VDC (open load) (factory pre-set is negative) Rear Panel
Current: 15 mA (short circuit) Nine pin “D” type connector with following available signals and
Maximum Power: 21 W (3 kV at 7 mA) commands Recorder outputs:
•  0 to +5 VDC, linear proportional to HV (1 V = 1 kV)
Front Panel •  0 to +10 VDC, linear proportional to current (10 V = 10 mA)
HV ON, HIGH LOAD, and POLARITY LEDs •  0 to +10 VDC, linear proportional to current (10 V= 1 mA)
LED bargraph linear scale for current and voltage indication HV ON confirm signal: Contact rating – 1 A at 250 VAC; 0.2 A at 30 VDC
Recorder Output 0 to +10 VDC linear proportional to current Remote HV ON/OFF (interlock) command
(10 V = 10 mA)
HV connector: Fischer type 105 or King type, 10 kV
Safety Marks
CE

32
ION Pumps
Ordering Information
Description Weight kg (Ibs) Part Number

MiniVac Controller
With any VacIon Plus pump MiniVac, FISCHER HV connector, US plug, 120 V preset 2.3 (5.0) 9290191
MiniVac, FISCHER HV connector, European plug, 220 V preset 2.3 (5.0) 9290290
MiniVac, FISCHER HV connector, 24 VDC 2.3 (5.0) 9290196
With small VacIon pumps (2 and 10 l/s) MiniVac, KING HV connector, US plug, 120 V preset 2.3 (5.0) 9290190
MiniVac, KING HV connector, European plug, 220 V preset 2.3 (5.0) 9290291
MiniVac, KING HV connector, 24 VDC 2.3 (5.0) 9290197

Accessories and Cables


19” Rack adapter 4.5 (10.0) 9699191
With any VacIon Plus pump HV bakeable cable, radiation resistant, 4 m (13 ft.) long, with Interlock 0.9 (2.0) 9290705
HV bakeable cable, radiation resistant, 7 m (23 ft.) long, with Interlock 1.6 (3.5) 9290707
HV bakeable cable, radiation resistant, 10 m (33 ft.) long, with Interlock 2.2 (5.0) 9290708
HV bakeable cable, radiation resistant, 20 m (66 ft.) long, with Interlock 4.4 (10.0) 9290709
With small VacIon pumps (2 and 10 l/s) HV bakeable cable, radiation resistant, 4 m (13 ft.) long, with Interlock
for 2 l/s pump 0.9 (2.0) 9290706
HV cable, 3 m (10 ft.) long, for 8 l/s VacIon pumps 0.9 (2.0) 9240741

Diode, Noble Diode positive


StarCell, Triode negative
(see page 42)

Pumps
Ion

33
Agilent Ion Pump Controllers

► Agilent 4UHV Ion Pump Controller

How much power do I need


for my ion pumps?
Power requirement depends on the pump
size and starting pressure; the larger the
pump and higher the starting pressure,
the higher the power consumption.
The largest standard Ion Pump
configuration, 500 l/s, can be easily
started with 200W up to 10-5 mbar, while
a medium size pump (75 l/s) needs less
than 80 W to be started at the same
pressure, and 80 W are sufficient
to operate a 500 l/s in the typical Ion
Pump operating range (below 2x10-6 mbar).

Why was the higher power rating


necessary in the past?
In the past ion pumps were started with
the aid of sorption pumps, able to reach
10-4 mbar only. As a consequence, much
larger and more powerful Ion pumps
controller were needed. The resulting life
of Ion Pumps started at such high
pressures was much shorter (1 minute of
operation at 10-4 mbar is equivalent to 2
months at 10-9 mbar). Today’s oil-free
turbo pumps, backed by oil-free primary
pumps, achieve lower pressures, thereby
reducing the starting pressure of the ion
pump. This reduces the maximum power
requirement of the ion pump controller
and extends the lifetime of the ion pump.

Negative or positive?
The requirement of negative or positive
potential depends on the pumping
element installed in the ion pump. Diode
style elements (Diode & Noble Diode)
need positive voltages, while Triode style
elements (old style Triode & StarCell)
need negative voltages for operation.

34
ION Pumps

211.40 (8,32) 199.60 (7,86)


141.20 (5,56)

164.30 (6,47)
177.00 (6,97)
400.50 (15,77)

Dimensions: millimeters (inches)

Technical Specifications
Input voltage 100 - 240 Vac (+/-10%) Modes of operation Local / Serial / Remote
Input frequency 50/60 Hz Front panel readings Voltage, Pressure, Current, Status

Pumps
Dimensions 400.5 x 211.4 x 177.0 mm (l x w x h) Safety marks CE, C_CSA_US

Ion
Display 4 rows with 20 characters Current measurement range 10 nA to 100 mA
Available configurations 1 x 120 W, 1 x 200 W, 2 x 80 W, Input signals On/off; Protect; Step Mode;
2 x 200 W, 4 x 80 W, 2 x 80 W +1 x 200 W Output signals Analog Out; NC Set-point; NO Set-point
Minimum configuration One HV card with 120 W, 200 W or 2x80 W HV connector Fischer Type 105
Output voltage (Open Circuit) 3.5 and 7 kV Output Power Maximum 400 W
Output current (Short Circuit) 40 mA for 80 W channel, Communications RS232/485 standard
100 mA for 200 W channel Profibus or Ethernet optional

Ordering Information
Description Part Number Description Part Number

4UHV Controller Accessories and Cables *


120 W negative 9299100 HV Bakeable cable, radiation resistant,
120 W positive 9299101 4 m (13 ft.) long, with interlock 9290705
200 W neg 9299010 HV Bakeable cable, radiation resistant,
200 W pos 9299011 7 m (23 ft.) long, with interlock 9290707
2 x 80 W neg 9299200 HV Bakeable cable, radiation resistant,
10 m (33 ft.) long, with interlock 9290708
2 x 80 W pos 9299201
HV Bakeable cable, radiation resistant,
2 x 200 W neg 9299020 20 m (66 ft.) long, with interlock 9290709
2 x 200 W pos 9299021 Rack adapter 19” 9290064
1 x 200 W pos & 1 x 200 W neg 9299022 Mains cable NEMA Plug, 3 m (10 ft.) long 9699958
4 x 80 W neg 9299400 Mains cable European Plug, 3 m (10 ft.) long 9699957
4 x 80 W pos 9299401
(*) The unit does not include the power cable, please order the cable
2 x 80 W pos & 2 x 80 W neg 9299402 separately.
2 x 80 W neg & 1 x 200 W neg 9299210
2 x 80 W pos & 1 x 200 W pos 9299211
2 x 80 W pos & 1 x 200 W neg 9299212
2 x 80 W neg & 1 x 200 W pos 9299213

35
Agilent Ion Pump Models

► Agilent Miniature and Small


VacIon Pumps and Controllers

Agilent offers a wide variety of small size ion pumps designed off under vacuum, which allows the vacuum integrity to be
especially for electron device and detector applications. The verified by the user just before use. Non-processed pumps are
Miniature Vaclon pump is a diode configuration and provides tested for no vacuum leaks and minimum leakage current.
approximately 0.4 l/s of nitrogen pumping speed. The 2 l/s Modified and Customized Versions
model is a modified diode configuration to enhance starting Modified versions of standard pumps can be provided when
at low pressure. The 10 l/s pump is a noble gas optimized different inlet tube lengths, angles, and diameters are
diode configuration with high efficiency for residual gases required. These pumps can also be customized with different
such as hydrogen. The pumping speed for noble gases is high voltage feedthroughs, body geometries, and pumping cell
about 20% of the nominal speed. arrangements. Special testing procedures can be quoted for
Pumps that are processed are baked to 400 °C and pinched customers who have specific requirements in this area.

Ordering Information
See pages 32 - 33 for controllers, cables, and accessories.

Description Weight kg (lbs) Part Number


Miniature Pump
With ⅜” OD 180° stainless steel tube 0.5 (1.0) 9130038
With ⅜” OD 90° stainless steel tube 0.5 (1.0) 9130041
With ⅜” OD 180° copper tube, vacuum processed 0.5 (1.0) 9130049
With ⅜” OD 90° copper tube, vacuum processed 0.5 (1.0) 9130050
Magnet for Miniature pump 0.5 (1.0) 9130042
HV cable, 2.4 m (8 ft.) long, for Mini VacIon pumps 0.9 (2.0) 9240122

2 l/s Pump
With ¾” OD 180° stainless steel tube 0.9 (2.0) 9190521
With ¾“ OD 180° copper tube, vacuum processed 0.9 (2.0) 9190522
With ¾” OD 180° stainless steel tube, vacuum processed 0.9 (2.0) 9190523
With ¾” OD 90° stainless steel tube, tee style 0.9 (2.0) 9190524
With 1 1/3” CFF 180° vacuum processed 0.9 (2.0) 9190520
Magnet for 2 l/s pump 0.9 (2.0) 9190038
HV bakeable cable, radiation resistant, 4 m (13 ft.) long with interlock for 2 l/s pump 0.9 (2.0) 9290706

10 l/s Pump
10 l/s Vaclon pump, processed, with 2 ¾” CFF 3.6 (8.0) 9195005
Magnet assembly for 10 l/s Vaclon pump 5.0 (11.0) 9110030
HV cable, 3 m (10 ft.) long, for 10 l/s VacIon pumps 0.9 (2.0) 9240741
Magnets must be ordered separately.
36
ION Pumps
Miniature Pump (90 deg. config.) Miniature Pump (180 deg. config.)

27.0 (1.06) REF


70.1 (2.8) REF

Ø 9.4 (0.4) REF


25 21.0 (.83) REF 20.1 (0.8) REF

EF
.0 27.2 (1.1) REF

R
(0

)
.9

.0
8)

(1
RE 25.7 (1.0) REF

.9
24
F 25.0 (.98) REF

Ø
54.0 (2.13) REF
Ø 9.5 (0.37) REF

PINCH OFF
19.1 (0.8) REF APPROX.

31.8 (1.3) REF


63.0 (2.5) REF 54.1 (2.1) REF
43.0 (1.69) REF 20.0 (.79) REF 54.0 (2.13) REF (VACUUM
42.9 (1.7) REF PROCESSED)

2 l/s Pump 10 l/s Pump


95.00 (3.70) REF
2 3/4" O.D. CFF (NW35)
55.673 (2.2)
Ø 19.500 (0.8)

153.00 (6.70) REF


60.400 (2.4)
30.200 (1.2)

Pumps
Ion
40.000 (1.6)
35.800 (1.4) 40.000 (1.6)
46.000 (1.8)
32.173 (1.3) 40.000 (1.6)

111.705 (4.4) 133.00 (5.24) REF 78.70 (3.04) REF


206.00 (8.11) REF

Dimensions: millimeters (inches)

The Miniature and 2 l/s pumps are available with copper or


stainless steel inlet tubes in 90- or 180-degree configurations,
relative to the high voltage feedthrough.

37
Agilent Titanium Sublimation Pumps and Controllers

► TSP Cartridge
130.4 (5.13) REF
3 TITANIUM
SUBLIMATION 23/4" O.D. CFF
FILAMENT (NW35)

STAINLESS STEEL
GROUND ROD

281.9 (11.10) REF

324.3 (12.77) REF

Dimensions: millimeters (inches)

Titanium sublimation pumps (TSPs) are typically used as an TSP Cartridge Filament Source
effective way to pump getterable gases such as hydrogen and The popular TSP cartridge is provided on a 2 ¾” OD ConFlat
nitrogen in UHV systems. TSPs are often combined with ion flange and contains three titanium-molybdenum filaments,
pumping, since the ion pump is effective with non-getterable each with 1.1 grams of usable titanium. The cartridge
UHV gases such as argon and methane. The TSP can be added assembly is bakeable to 400 °C. Maximum sublimation is
to the inside of the ion pump, or as a separate pumping unit. achieved at 300 watts of source power.
If the TSP is used in conjunction with a liquid nitrogen-cooled
cryopanel, extra high water pumping speed will be achieved.
Agilent offers two different types of titanium sublimation
Technical Specifications
pumps: filament and titanium ball sources. Filament-type TSP Usable titanium (per filament) 1.1 grams
sources are most popular with UHV systems since they can Total usable 3.3 grams
be turned off between sublimations and thus do not add Operating range 10-4 to 10-12 mbar
thermally-induced outgassing. The ball-type sources contain
larger amounts of titanium which means longer life when
operated under conditions that use more titanium, such as
higher operating pressures. However, the ball sources require
standby power between sublimations to prevent cracks from ► Mini Ti-Ball
forming in the titanium ball. 23/4" O.D. CFF
MINI TI-BALL MINI TI-BALL
The factors affecting titanium sublimation pumping efficiency SOURCE ASSEMBLY SOURCE HOLDER (NW35)

include sublimation rate, frequency, surface area, and


temperature. Sublimation pumping speed is generally a
constant value below 10-7 mbar.
21.3 (.84) REF

61.4 (2.42) REF

185.6 (7.31) REF 41.1 (1.62) REF

Dimensions: millimeters (inches)

Mini Ti-Ball Source


The Mini Ti-Ball source provides 15.2 grams of usable
titanium and is mounted on a 2 ¾” OD ConFlat flange.
Maximum sublimation is achieved at 300 watts of source
power, while the Standby power requirement is 100 watts.

Technical Specifications
Usable titanium 15.2 grams
Operating range 10-4 to 10-12 mbar

38
ION Pumps
► Agilent TSP Cryopanel
8" O.D. CFF (NW150)

120.0 (4.72) REF


Ø 140.0 (5.51)
276.6 (10.89) REF 124.3 (4.89) REF

Dimensions: millimeters (inches)

TSP Cryopanel Pumping Speed vs Pressure at different evaporation rates


Designed for use with the TSP cartridge source*, this
sublimation cryopanel is mounted to an 8” OD ConFlat flange. 1250
It can operate with water cooling, liquid nitrogen, or uncooled H2 0,1 gr/h
if used at UHV. This cryopanel can be mounted to double- 1000
ended or side-ported ion pumps and can also be used
Pumping Speed (l/s)

independently in any 8” (NW 150) CFF port with 11 inch 750


depth/clearance. H2O

The Cryopanel does not include the cartridge. 500

Pumps
N2

Ion
250
Technical Specifications 0,01 gr/h

N2 H2 H20 0
Pumping speed at 20 °C 10 -11 10 -10 10 -9 10 -8 10 -7 10 -6 10 -5 10 -4
water-cooled (l/s) 515 1200 575 Pressure (mbar)
Inner pumping surface (cm2) 826
Main flange 8” CF (NW150)
Reservoir volume (liters) 1.8
Cooling connection ⅜” G
Titanium source flange 2 ¾” CF

39
Agilent Titanium Sublimation Pumps and Controllers

► Agilent TSP Controller 361.70 (14.24)


211.33 (8.3) 337.30 (13.28 )
141.22 (5.6) 3.99 (.16)

177.04 (7.0)
164.3 (6.5)
19.00 (.75) 204.50 (8.05)

6.0 (.3)

200.0 (7.9)
Dimensions: millimeters (inches)

• Pump cable lengths up to 50 m • To operate Agilent TSP filament cartridge


• Compact design ½ standard rack or Agilent Mini Ti-Ball
• Selectable for automatic or manual operation • Safety Marks: CE, cCSAus
• Remote control (standard) via RS232/485

Technical Specifications
Input 100, 120, 220, 240 Vac ±10% 1-phase In compliance EN 61010-1 (2001)
Voltage • 90 to 110 Vac - 1 phase (use setting 100 Vac) with norms EN 61326-1 (2006) - Class A
(selectable • 110 to 130 Vac - 1 phase (use setting 120 Vac) (industrial application)
at the rear • 190 to 230 Vac - 1 phase (use setting 220 Vac) Protection category IP 20
of the case) • 230 to 265 Vac - 1 phase (use setting 240 Vac) Internal use only
Frequency 47 to 63 Hz Installation category II
Power 1400 VA (maximum, see note) Pollution degree 2
Fuse 2 x T15A (slow blow) 250V 5x20 fuses 3 meters long
(only for model 929-0032) Power cable • 3 wires Ø AWG14, NEMA plug
2 x T8A (slow blow) 250V 5x20 fuses (only for model 929-0032)
(only for model 929-0033) • 3 wires Ø 0.75 mm2 European plug
Temperature 0 to + 45 °C (only for model 929-0033)
Humidity 90 % maximum non condensing humidity Weight 12 kg (26.5 lbs)
Storage temperature -20 °C to +70 °C
NOTE • When the controller is powered by means of a transformer, the transformer power must be at least 3000 VA to avoid a distortion
of the power waveform.

Ordering Information
Titanium Sublimation Pumps Part Number Accessories Part Number
TSP Filament Cartridge on a 2 ¾” CFF 9160050 Cable for TSP pump (3.5 mt) 9240730
Replacement filaments, package of 12 each 9160051 Cable for TSP pump (7 mt) 9240730M002
Mini Ti-Ball Source with Holder on a 2 ¾” CFF 9160009 Cable for TSP pump (10 mt) 9240730M001
Replacement Mini Ti-Ball Source 9160008 Cable for TSP pump (30 mt) 9240730M017
Cable for TSP pump (40 mt) 9240730M015
Titanium Sublimation Cryopanel Part Number
Cable for TSP pump (50 mt) 9240730M013
Sublimation Cryopanel on an 8” CFF 9190180
Cable for Mini Ti-Ball pump 9240752
Controllers Part Number Rack adapters 9290064
Sublimation Controller set for 220 Vac 9290033
Sublimation Controller set for 110 Vac 9290032

40
Agilent Ion Pump Technical Notes ION Pumps
Historical Notes or electrode of that material, which is in turn sputtered and
eroded by gas ions formed under the influence of the high
Ion pumping is used to remove gases from systems in order voltage. These electrical potentials are usually in the range of
to create ultra-high vacuum environments. The earliest 3,000 to 7,000 Vdc.
evidence of ion pumping was reported by J. Plucker Most ionization devices operate in the same way. Gas
(1858 - Germany) who found that it took ever-increasing molecules are bombarded by high energy electrons when a
voltages to maintain a current in a gas discharge tube. collision occurs. A molecule may lose one or more of its own
This, he rightly concluded, is due to a reduction of pressure in electrons and thereby is left as a positively charged ion. Under
the tube by some mechanism involving the cathode. the influence of a strong electric field, the ion is accelerated
Later, as an offshoot of his work on electrical discharge in into the titanium cathode. The force of this collision is
gases, F. Penning (1937 - Holland) developed a cold cathode sufficient to cause atoms to be ejected from the cathode and
ionization gauge for measuring pressures in the range of 10-3 “sputtered” onto the adjacent walls of the pump. Freshly
to 10-5 Torr. Due to the sputtering effect of the high voltage, sputtered titanium is extremely reactive and will chemically
ions were both buried in and “gettered” by the cathode react with active gases. The resulting compounds accumulate
material. (Gettering is the chemical combination of active on surfaces of the pump elements and pump walls.
gases with a suitably reactive substance). Active gases are those such as oxygen, nitrogen, CO, CO2,
The result of this pumping action was a noticeable pressure and water, as opposed to the noble gases like helium, neon,
reduction. The Penning cell has been used as a commercially argon, krypton, and xenon, which are nonreactive. The latter
available vacuum gauge ever since, but it was not until the are pumped by “ion burial” (ion burial is the “plastering over”
late 1950’s that its pumping characteristics were exploited, of inert gas atoms by the sputtered getter atoms).
resulting in the invention of the ion pump. This was done in The simplest form of ion pump is the Penning cell, which was
order to improve the life and performance of microwave tubes originally conceived as a cold cathode vacuum gauge. It
by continuous pumping with “appendage” ion pumps. consists of a central anode wire which is at positive high
The invention of the sputter ion pump ushered in the era of voltage. In an ion pump the anode can either be a short
ultrahigh vacuum, just in time to make a large contribution to section of metal tubing or a square, box-like structure, open

Pumps
the space age.

Ion
at each end like a unit of an egg crate. Opposite each open
The availability of vacuum systems that could routinely end is a plate of titanium that is connected to the ground to
achieve pressures in the low 10-11 Torr range enhanced R&D form the cathode structure. An external permanent magnetic
efforts. Space hardware and space-compatible materials were circuit generates a magnetic field, usually ranging from 800 to
tested by simulating many of the conditions they would 2,000 G, parallel to the anode cell axis. A cell configured in
encounter. Today, ion pumps are used in both research and this way is said to be a diode pump (Figure 1). It is then
industrial applications wherever a pristine, oil-free, vibration- packaged in a suitable enclosure and the assembly becomes
free, cost-effective environment is required. a pump.
To make a higher speed pump, it is now simply a matter of
Operation making a package containing more cells with a larger cathode
(Figure 2).
Vacuum pumps in general operate on the basis of maintaining
a lower gas density within themselves than exists in the The function of the anode cell structure is to contain a
environment they are pumping. This results in a net gas “cloud” of high energy electrons which are constrained by the
migration into the pump due to the random motion of the magnetic field.
molecules under molecular flow conditions. Once in the This field causes the electrons to move in oscillating spiral
pumps, few escape and they are either displaced or captured, paths (Figure 3) that increase their chances of striking gas
depending on the type of pump. molecules and thereby create positive ions. These ions are
Rather than being a displacement pump that actually moves accelerated away from the positive anode voltage and collide
molecules of gas through it to the atmosphere, the ion pump into the titanium cathode plates (Figure 3). This results in the
instead captures and stores them. As a result, at some point removal of titanium atoms by “sputtering”.
in time the pump must be reconditioned or replaced. This is The sputtered titanium is deposited on the internal surfaces
generally required only after many years of use. of the pump where it reacts with absorbed active gases to
The generic name Sputter Ion Pump (or Ion Getter Pump) form stable compounds.
comes from the fact that some of the gas molecules undergo In summary, the pumping efficiency depends on the electron
ionization and cause sputtering of the gettering agent. This “cloud” density (which determines the number of ions
material chemically reacts with the active gases to form produced) and on the sputtering yield (which determines the
stable compounds that are deposited on the internal walls of quantity of active getter material produced).
the pump. The getter, usually titanium, is provided by a plate The electron cloud density mainly depends on the Penning

41
Agilent Ion Pump Technical Notes

Figure 1 Ion pump cutaway

Figure 2 Figure 4

Figure 3 Figure 5

42
ION Pumps
cell geometry and on the electric and magnetic field If not cleaned properly, the ion pump itself can be a source
strengths. By adjusting these parameters, the pump of gas at UHV.
performance can be modified according to the application. In order to ensure cleanliness, VacIon Plus Pumps are factory
In particular, using an “intelligent” high voltage power supply, processed at high temperature in ultraclean vacuum for a
the right voltage can be automatically selected as the thorough outgassing of the body and all internal components.
pressure changes. The cleanliness of the ion pump element at high temperature
The sputtering efficiency depends on the cathode geometry (450 °C) in ultraclean is even more critical, due to the
and material and on the gas species. Thus the cathode continuous cathode bombardment. Any gas trapped on the
configuration can also be optimized for the gas used in the surface or in the bulk of the cathode will eventually be released.
application. Ion Pump Outgassing System
Agilent offers three different cathode configurations that will The ion pump outgassing system is a thermal process of the
satisfy a wide variety of applications involving different gases pump, completely computer controlled and able to provide an
and operative pressures. automatic final test of the achieved pump specifications.
It should be noted that a few types of gases do not need to be The bake-out of the pump is done in a nitrogen-controlled
ionized in order to be pumped. Hydrogen, for instance, forms atmosphere to protect the external pump body from oxidation.
a solid solution directly with the titanium cathode plates as See Figure 6 for System outline.
well as the sputtered film.
It is the function of the ions to maintain a fresh supply of
“gettering” material. In this regard, the ion pump is self- Figure 6 - System outline
regulating; it only sputters as much getter material as is
needed at that particular pressure.
In this way, at low pressures cathode plates are not wasted
and electric power is conserved.
Some of the noble gas atoms are pumped as a result

Pumps
of being ionized.

Ion
In this case, they are implanted (at least temporarily) in the
cathodes by the force of the accelerating voltage. Others are
pumped by burial in the sputtered titanium film (Figure 4).
Generally, pumping noble gases does not pose a problem
because they are present in such small quantities. When it is
necessary to deal with considerable amounts of noble gases,
a pump of the triode configuration should be used (Figure 5).
In the triode pump, the cathode is at negative potential and
built with slits that permit grazing incidence sputtering.
Thus, they tend not to be implanted to any appreciable extent
and the un-sputtering of previously buried noble gases is
largely eliminated. Instead, the gases either react with or are The system is based on the principle of thermal outgassing
buried by the sputtered film on the walls of the pump body the ion pump internal surfaces through the control of the their
and on the anode. intrinsic outgassing. Therefore pressure, not time, is the
driving factor of the overall process.
Selecting VacIon Plus Pumps The bake-out time depends on the internal cleaning of the
pump components and all the pumps will have, in this way,
lon pumps are commonly used to create Ultra High Vacuum the same final outgassing rate and base pressure.
(UHV), due to their cleanliness, ability to pump different Figure 7 shows the working principle. The green curve
gases, and maintenance- and vibration-free operation. Long represents the temperature and the yellow one the pressure
operating life and ability to read pressure are other important read by the gauge located on the vacuum control system
features of ion pumps. placed on the bottom of the outgassing station (see Fig. 6).
The Vaclon Plus family has been designed to enhance all of The temperature is maintained at the set point level until the
these characteristics, and thus offers the most advanced and pressure does not change any more, which means the
valuable solution to any ion pumping requirement. outgassing of the pump for the defined bake-out temperature
Cleanliness is completed.
To reach very low pressures (i.e. 10-11 mbar) in any system, At the end of the thermal process, once room temperature
both the chamber and pump outgassing must be minimized. is achieved, an RGA is performed.
43
Agilent Ion Pump Technical Notes

Figure 7 - Thermal process working principle Long Operating Life


All VacIon Plus pumps have rated lives in excess of many
10 -3 500
thousand of hours at a pressure of 1x10-6 mbar (50,000 hours
10 -4
400
for the Diode pump, and 80,000 hours for the StarCell).
-5
With many ion pumps, maintenance may be required well

Temperature (°C)
10
before the rated life, due to metallization of insulators or
Pressure (mb)

300

10 -6 pumping element distortion.


10 -7
200
All VacIon Plus elements are designed to minimize cathode
distortion (even after repetitive bake-outs and starting at high
100
10 -8 pressure), and the insulators are protected from sputtered
10 -9 0
titanium by using a double re-entrant design and a cap shield.
0 1 2 3 4
Time (hh) Pressure Reading
Manifold Bell Manifold
The ability to read pressures using an ion pump is due to the
Vacuum Temperature Temperature direct proportionality between pump current and operating
pressure.
The gas analyser, placed on the vacuum system, provides The reliability of pressure readings at very low pressure is
the spectrum of the different gases degassed by the pump. limited by leakage current, and the leakage current from field
If H2 and the other peaks normally present in a well-baked emission is heavily dependent on the voltage applied
vacuum system exceed the acceptance levels, the pump is to the pump.
baked again. Otherwise, it is pinched off and its base pressure The Dual controller, designed for use with any VacIon Plus
monitored. The base pressure is evaluated through the ion pump, provides the unique ability to adjust the voltage in
current reading. accordance with the operating pressure. By doing this, the
The current decrease is computer monitored, and the pump is leakage current is minimized at low pressure, providing a
ready to be shipped only after the base current is reached. reliable pressure reading down to the 10-10 mbar range.
Figure 8 shows the result of the residual gas analysis Custom Design and Flexibility
performed at the end of the bake-out. All ion pumps can be mounted in any position, and do not
need any isolation valve from the system in case of venting or
Figure 8 - Residual gas analysis power failure.
10 -7
VacIon Plus pumps are the most compact ion pumps in each
speed range. The pumps can be configured with additional
flanges and can accomodate other pumping systems (like TSP
10 -8
= Titanium Sublimation Pump), allowing the best use of
available space.
Pressure (mb)

Agilent Feedthrough
10 -9 The whole Ion Pump line is provided with an
innovative feedthrough.
The improved feedthrough of the ion pump line is a major
10 -10
0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 100 breakthrough. The greatest advantages of the ion pumps are
Mass listed here below:
• Corrosion Free
Pumping of Different Gases The design of the feedthrough will reduce drastically the
In general, all ion pumps can pump all gases to some corrosion that may occur when the pump is used in
degree. To obtain the best performance and base pressure, humid environments.
different types of ion pumps have been developed with Our tests and experience have demonstrated that the
optimized performance in different pressure ranges and with corrosion starts and increases with the presence of humidity
different gases. between feedthrough and connector. The high voltage
Agilent’s VacIon Plus is a complete product family that offers during the pump operation ionizes the entrapped water
the choice among three different elements: Diode, Noble vapor; the ions react with the brazing alloy and corrode it.
Diode and StarCell. The design of the feedthrough aims to solve these issues:
Whatever the application, there is a VacIon Plus pump – On one hand the design structure of the feedthrough
designed for it. presents a negligible quantity of air.

44
ION Pumps
– On the other hand, the brazing is done on the vacuum side As with the diode configuration, the Noble Diode retains a
so that the brazing alloy surface exposed to air is minimal. consistent pumping speed for all gases at very low pressures.
• High Voltage Cable Interlock However, pumping speed for H2 and getterable gases is lower
than for the corresponding Diode. The Noble Diode Vaclon
The feedthrough has been designed for the “HV cable
Plus is typically used in UHV applications where a mixture of
safety interlock” implementation. This feature avoids any
gas is to be pumped and where the pressure is quite constant
electrical shock since the voltage is automatically cut off as
(i.e., no sudden gas burst or systematic high pressure cycling).
soon as the cable is disconnected from the pump.
Its characteristics of consistent speed for almost any gas
Our Ion Pump control unit (Dual) and the HV cable are
even at very low pressures make it ideal whenever the ion
already able to support this safety feature when connected
pump alone is used to obtain UHV pressures. This is often the
to an ion pump with the feedthrough.
situation in particle accelerators or synchrotron rings, as well
• Easy Connection as in surface analysis applications.
The connection of the cable connector to the feedthrough Other Vaclon Plus versions are suggested whenever the
requires simply inserting and pushing the connector. application requires cycling to higher pressures, pumping of
No use of retention screws is required. large amounts of H2, or when the ion pump is combined with
• Safety against Unintentional Extraction other UHV pumps such as Titanium Sublimation Pumps or
When the HV cable connector is inserted into the new Non-Evaporable Getters.
patented feedthrough, it is firmly and mechanically secured StarCell Vaclon Plus
to it. A latch on the cable prevents disconnection. The StarCell Vaclon Plus element is the latest variation of the
• Compactness Triode configuration. Its design makes this ion pump the only
The feedthrough design allows a significant gain of space one that can handle a high amount of noble gases (better
for the customer. than Noble Diode) and hydrogen (comparable to the Diode).
Furthermore, this pump provides the highest speed and
capacity for methane, argon, and helium. Its high total
The VacIon Plus Family capacity for all the different gases, together with its very good

Pumps
Ion
speed performance at relatively higher pressures, makes the
Diode Vaclon Plus StarCell Vaclon Plus ideal for applications requiring constant
The diode version of the Vaclon Plus pump has the highest operation at 10-8 mbar or above. This typically includes
pumping speed among all ion pumps for oxygen (O2), nitrogen Electron Microscopes and Mass Spectrometers.
(N2), carbon dioxide (CO2), carbon monoxide (CO), and any
other getterable gas. It provides the highest pumping speed Figure 9 - StarCell pumping element
and capacity for hydrogen (H2) as well.
Its simple mechanical structure allows a reliable current/
pressure reading down to very low pressures, as well as
absolutely vibration-free operation. Its geometric and
electrical configuration allows it to be used in the vicinity of
electron detectors or similar devices.
Diode Vaclon Plus pumps are therefore widely and
successfully used in general purpose UHV systems, for
evacuating electron devices and in the most sensitive
electron microscopes. However, diodes are not suggested for
applications where noble gases such as argon (Ar), helium
(He) and methane (CH4) are to be pumped.
Noble Diode Vaclon Plus
The Noble Diode Vaclon Plus element is a version of the Its high pumping speed for argon, helium, and methane (the
diode element, in which a tantalum cathode is substituted in highest of any ion pump at any pressure) has made StarCell
place of a titanium cathode. This substitution allows a higher the standard for any application where the ion pump is used
pumping speed and stability for pumping noble gases (mainly in combination with TSP or NEG pumps, where its pumping
argon and helium). performance is enhanced. The lowest attainable pressure
The element is otherwise equivalent to the Diode Vaclon Plus. has been obtained with combinations of StarCell Vaclon Plus
Noble Diode Vaclon Plus pumps are used in any application and TSP/NEG pumps, thanks to the optimized characteristics
where pumping of noble gases is an important characteristic. of these combinations. Most existing particle accelerators

45
Agilent Ion Pump Technical Notes

and synchrotron sources, beam lines, transfer lines, and In an ion pumped UHV system with a proper bakeout
similar devices have used and are successfully using these procedure (and consequent pump regeneration), a pressure in
combinations to obtain the maximum speed for all the gas the 10-11 mbar range is possible. At this pressure, the ion
species. pump will work at the higher (unsaturated) pumping speed
For more specific information regarding the pumping speeds values for a few years before it is saturated.
for different element configurations relevant to different When pump saturation is reached, the pumping speed is
gases, refer to the various curves published in the Vaclon constant and no longer depends on the quantity of pumped
Plus Pumping Speed section. The aim of this section is to gas. The values of the pumping speed curve after saturation
help you choose the best Vaclon Plus configuration. are the lowest obtainable at any given pressure.
Whatever the application, there is a Vaclon Plus version Generally, the ion pump works in an intermediate condition
suited for it. between the unsaturated and saturated extremes. The
Do not hesitate to contact your local Agilent representative if “nominal” pumping speed is defined as the maximum point
you need more assistance in making the right choice. on the pumping speed curve for a saturated pump; the
For VacIon Plus and NEG integrated pumps, please ask Agilent reference gas is usually nitrogen.
for a special quotation. The nominal pumping speed thus defines only a part of the
characteristics of an ion pump. A more comprehensive
VacIon Plus Pumping Speed description of an ion pump performance is given by the
complete pumping speed vs pressure curve. By using these
The most common parameter used to express the capability plots and keeping in mind the application, the best pump
of a pump to remove molecules from a given volume is can be chosen.
pumping speed. It is usually measured in liters per second
and expresses the volume of gas (at a given pressure) Figure 10 - Saturation effect
removed per time unit.
In an ion pump, the net pumping effect results from the sum 200

of different phenomena:
% of Nominal Pumping Speed

• The pumping action of the getter film produced by the


sputtering of cathode material by ion bombardment. 1d
1M 1h
• The pumping action due to the ion implantation and 100
1Y
diffusion into the cathode. 1 min
1O Y
• Gas burial on the anodes and pump walls.
• The gas re-emission from the cathode due to cathode
heating and erosion. 1
10 -11 10 -10 10 -9 10 -8 10 -7 10 -6 10 -5 10 -4
When an ion pump is new or has been regenerated, for Pressure (mbar)
example by baking, the surface layer of the cathode is clean
and the gas re-emission from it is negligible. In this condition,
the ion pump is called “unsaturated” and the pumping effect
is due both to the gettering effect as well as to ion Active Gases (N2, 02, CO, CO2,…)
implantation and diffusion. As the number of gas molecules A characteristic of these gases is their ability to easily react
implanted into the cathode increases, the re-emission of them with most metals forming stable compounds. In an ion pump,
due to the ion bombardment increases. these active gas molecules react with the fresh titanium film
As a consequence, the net pumping speed decreases until an produced by the sputtering of the cathode material. These
equilibrium condition between ion implantation and gas active gas molecules do not diffuse deeply in the cathode.
re-emission is reached. In this condition, the ion pump is The saturation effect, due to the re-emission of these
“saturated” and the net pumping speed, due only to the molecules trapped on the cathode surface, is very strong.
gettering action of the material sputtered from the cathode, is Diode and Noble Diode elements show a higher pumping
about half the pumping speed of the unsaturated pump. speed at low pressure while StarCell elements perform better
Since the saturation effect depends on the quantity of gas at higher pressure, because the Penning discharge is better
molecules implanted into the cathode, the time required to confined inside the element (Figures 11, 12).
saturate an ion pump is inversely proportional to the
pressure at which the pump is operated. Thus, the lower the
pressure, the longer the time before the pump saturation
occurs (Figure 10).

46
ION Pumps
Figure 11 - Nitrogen pumping speed before saturation Figure 13 - Hydrogen pumping speed
200
250

150 200
% of Nominal Speed

% of Nominal Speed
150
100

100
50

50

0
10 -11 10 -10 10 -9 10 -8 10 -7 10 -6 10 -5 10 -4 0
10 -11 10 -10 10 -9 10 -8 10 -7 10 -6 10 -5 10 -4
Pressure (mbar)
Pressure (mbar)

Diode Noble Diode StarCell


Diode StarCell

Figure 12 - Nitrogen pumping speed after saturation Noble Gases (He, Ne, Ar,…)
The main characteristic of noble gases is that they do not
100
react with any other element. Therefore, the film produced by
the sputtering of cathode material does not provide getter
pumping for helium and argon. Moreover, since these gases
% of Nominal Speed

do not tend to diffuse into the cathode, the pumping effect


50 due to the ion implantation is not permanent.
Nevertheless, all the ion pump elements have some capacity
to remove these gases.

Pumps
Ion
0
10 -9 10 -8 10 -7 10 -6 10 -5 10 -4
Figure 14 - Argon and Helium pumping speed
10 -11 10 -10
40
Pressure (mbar)

30
Diode Noble Diode StarCell
% of Nominal Speed

Hydrogen 20

Hydrogen is an active gas but, due to its very small mass,


the sputtering rate is very low. In spite of this fact, the 10

pumping speed for H2 is very high because it quickly diffuses


into the cathode with negligible re-emission. When pumping 0
10 -11 10 -10 10 -9 10 -8 10 -7 10 -6 10 -5 10 -4
H2, the ion pump always works in the unsaturated condition.
Pressure (mbar)
As a result, the nominal speed for H2 is about twice the
corresponding value for nitrogen.
StarCell, Unsat Noble Diode, Unsat
Furthermore, if some traces of heavier gases are present, the StarCell, Sat Noble Diode, Sat
increased sputtering rate produces an even higher hydrogen
pumping speed.
Noble gases are pumped by being buried by titanium. Noble
The Diode element shows a higher pumping speed than the
gas ions can be neutralized and scattered from the cathode
Noble Diode since the H2 solubility in the tantalum cathode is
without losing their energy. These neutral atoms maintain
lower than in a titanium cathode. StarCell elements combine
enough energy to implant or stick on the anode and on the
good performance at higher pressures with enhanced
pump walls where they will be buried by sputtered titanium
capacity for H2.
and thus permanently pumped.

47
Agilent Ion Pump Technical Notes

In the Diode configuration, the neutralization and back Methane


scattering probability is very small, thus the pumping speed Although methane is not a noble gas, it does not react with
for noble gases is only a small percentage of the N2 pumping any getter material. It is always present to some degree in
speed. Moreover, when operating at a relatively high argon UHV systems as a reaction product of hydrogen and carbon
partial pressure (i.e., higher than 10-8 mbar), sudden bursts of present in the vacuum system walls. Methane is a particular
pressure due to the re-emission of temporarily implanted problem in electron accelerators where it is the main cause of
argon in the cathode is observed. After this occurs, a Diode the beam decay.
pump is not able to pump more argon until its source is
stopped. This phenomenon is known as “argon instability”.
Figure 16 - Methane pumping speed
Figure below shows that the StarCell pump is superior in
comparison to diode pumps with differential cathodes. The test
150
was performed at 10-5 mbar. The differential cathode diode showed
instability after pumping about 3 mbar l of argon while the Starcell
remained stable after pumping about 70 mbar l of argon, at which
point the test was stopped. 100

% of Nominal Speed
Fig. 15 - Starcell and Noble Diode instability at P=1 10-5 mbar
50
0,4

0,3 0
10 -11 10 -10 10 -9 10 -8 10 -7 10 -6 10 -5 10 -4
SAr/SN2(nom)

Pressure (mbar)
0,2

StarCell, Unsat Noble Diode, Unsat


0,1 StarCell, Sat Noble Diode, Sat

0,0
0,1 1,0 10,0 100,0 1000,0 Due to the Penning discharge in ion pumps, the methane
Time (hours) molecule (as well as other hydrocarbon molecules)
is cracked and transformed into smaller getterable
Noble Diode StarCell
compounds (C, CH3, … H).
The result is that the pumping speed for methane and light
hydrocarbons is always higher than the speed for N2.
In the Noble Diode element, one titanium cathode is replaced
with one tantalum cathode. The high nuclear mass of
tantalum increases the back-scattering probability and
consequently the pumping speed for noble gases.
The best results in terms of pumping speed of noble gases
are obtained using the open cathode structure typical of
StarCell elements. In these configurations, the flat cathode
structure has been replaced with a structure that allows
glancing collisions with ions.
These are neutralized and then forward scattered toward the
pump wall or the anode with a much higher probability than
in the flat cathode case. The result is a pumping speed for
noble gases of up to 60% of N2. Furthermore, due to the
unique design that allows optimal use of all the available
titanium, the operating life of a StarCell pump is about 50%
longer than all the other pumps.

48
ION Pumps
Basic Performance Factors
• Pumping speed, as in the case of any high vacuum pump,
is one of the factors determining the ultimate (base)
pressure of a system. However, ion pumps pump various
gases at different rates – from very fast for hydrogen to
quite slow for argon. Thus, it is necessary to check
specifications to match the pump properly to the
application. The pump history in terms of gas load after a
bakeout also influences pumping speed. Agilent pumps are
rated by their equilibrium or “saturated” speed levels.
Unsaturated pumps temporarily deliver higher speeds,
especially at UHV. A typical pumping speed vs pressure
curve is shown on page 46.
• Maximum throughput of an ion pump is located in the 10-4
to 10-5 mbar pressure range. Since ion pumps are generally
operated at much lower pressures, throughput is generally
not a major consideration.
• Starting pressure is the pressure to which the ion pump
must first be rough-pumped before the glow discharge will
be confined to the anode cell structure and the pumping
effect begins. To prevent pump overheating and possible
damage when the pump is saturated at higher pressure, the
controllers automatically reduce the output voltage to
maintain power below a safety level.

Pumps
• Pump life is determined by various factors, including

Ion
cathode life; this is limited by cathode erosion, which is
proportional to the average operating pressure. At 10-6
mbar, cathode life ranges from 35,000 to 80,000 hours.
• Bakeability. In general, baking a UHV system and its ion
pump to 200 to 250 °C is adequate for outgassing, attaining
low pressures after atmospheric exposure. Some processes
require higher bakeout temperatures. In general, ion pump
bodies can be baked at 450 °C when the magnets are
removed, or 350 °C when the magnets are on the pump.
When bakeable cables are connected to the ion pumps,
temperature should not exceed 220 °C.

49
Agilent Titanium Sublimation Pump Technical Notes

Titanium Sublimation Pumps As the gas molecules react with the surface Ti atoms, the
number of active sites decreases and, as a consequence, the
Thin films of reactive materials have been used for “gas pumping speed decreases. A plot of the specific pumping
cleanup” or “gettering” for over a century. The early electron speed vs time at different pressures is reported in Figure 17.
tube makers were only able to mechanically pump to about Using this plot it is possible to estimate how frequently the Ti
1x10-4 mbar, but through the use of “getters” flashed on the film has to be renewed.
internal surfaces, pressures in the low 10-7 mbar scale were It must be noted that the actual pumping speed S of a TSP
attained. These getters were typically metals like barium, depends on the conductance C between the active surface
titanium, zirconium, or thorium. Gettering materials are still and the vacuum vessel according to the following equation:
used in tubes today even though pressures of 1 x 10-8 mbar 1/S = 1/C + 1/Si
are readily attained by the pumps in the
manufacturing process.
Gettering was not employed extensively in vacuum systems Figure 17 - Pumping speed vs time at different pressures
until the 1960’s, when it was found to be highly compatible 100
with ion pumping. Titanium was the metal commonly used 90
because of its availability and its ability to sublime readily 80

% of Maximum Pumping Speed


1E-12 mbar
over a moderate temperature range. 70
1E-11 mbar
60
Applications

for Nitrogen
1E-10 mbar
50
Due to cleanliness, bakeability, low power consumption, 40 1E-9 mbar
1E-8 mbar
vibration-free operation, long pumping life, and high pumping 30
1E-7 mbar
speed, Titanium Sublimation Pumping (TSP) is the ideal cost- 20
10
effective companion to ion pumping in ultrahigh vacuum. 0
5E-7 mbar

Applications for this pumping mode are found in many areas, 1 10 10 2 10 3 10 4 10 5 10 6

such as: Time (minutes)

• Auger electron spectrometry


• Electron spectroscopy for chemical analysis
• Electron tube manufacturing • Throughput. When the impingement rate of the gas
• Mass spectrometers molecules on the active film becomes higher than the Ti
• Materials science conductor R&D sublimation rate (excess of gas molecules respect the
• Nuclear physics available Ti atoms), the pumping speed does not depend
• Outer space simulation any more on the sticking coefficient. It is simply controlled
• Particle accelerators by the quantity of the available Ti atoms according to
• Secondary ion mass spectroscopy stoichiometric reaction.
• Solid state semiconductors If n Ti atoms need to pump a gas molecule (example:
2Ti + N2 →← 2TiN, n = 2), the gas throughput Q is given by:
Basic Performance Factors

[]
• Pumping Speed. The pumping speed of a Ti film is
proportional to the film area and to the sticking coefficient, Q [mbar l/s] = ____
0.13 __
R
gr
n h
that is the probability that an impinging gas molecule reacts
with Ti forming a stable compound. The pumping speeds Pluswhere R is the Ti sublimation rate. In this condition the
per unit area of a fresh evaporated Titanium film are pumping speed is not constant but it depends on the pressure
reported in Table 1. Using these coefficients the intrinsic and is directly proportional to the sublimation rate (Figure 18).
pumping speed (Si) of a Ti film can be evaluated using the
following equation:
Si[l/s] = Coefficient x Surface.

Table 1
Typical pumping speeds per square inch (per square centimeter)
of titanium sublimation surface for various gases (liters per second)
Gas H2 N2 O2 CO CO2 H20 CH4 Ar He
Surface Temperature 20 °C 20  (3.1) 30  (4.7) 60  (9.3) 60  (9.3) 50 (7.8) 20  (3.1) 0 0 0
–195 °C 65 (10.1) 65 (10.1) 70 (10.9) 70 (10.9) 60 (9.3) 90 (13.9) 0 0 0

50
ION Pumps
After this time the filament cartridge should be replaced.
Figure 18 - Pumping speed vs pressure
Design Features
100000
70 cm dia
gas: nitrogen • All sources are mounted on 2 ¾” Agilent ConFlat Flanges
10000
temperature: 20 ° C and fit through 1 ½” ports.
Pumping Speed (l/sec)

40 cm dia mode: continuous


sublimation • The three-filament source contains 3.3 grams of useful
1000
10 cm dia
titanium.
100
• The Agilent Mini Ti-Ball source contains 15.2 grams of
0,1 gr/h
useful titanium.
0,01 gr/h
10

1
10 -11 10 -10 10 -9 10 -8 10 -7 10 -6 10 -5 10 -4 10 -3 10 -2

Pressure (mbar)

• Other factors. The overall performance of a titanium


sublimation pump is a function of several variables,
including gas species, pressure, gas temperature, getter
film temperature, getter film area, the geometry of the area,
sublimation rate, sticking coefficient, and the conductance
from the film to the area being evacuated. For further
information, write for “Predicting and Evaluating Titanium
Sublimation Pump Performance” by D.J. Harra, 1974
(Vacuum Report VR-88).
Operation

Pumps
Ion
Titanium Sublimation Pumping is accomplished by coating the
inner surfaces of a vacuum system with sublimed titanium
films. Since it involves a chemical reaction, this kind of
pumping is useful where mainly active gases are present.
The pumping speed of a unit area varies with various reactive
gas species as shown in the following table. It can also be
seen that cooling the substrate to liquid nitrogen temperature
markedly increases the speed for hydrogen and water.
The gases thus “gettered” form stable compounds with
titanium and are stored in the system as such until they are
removed by cleaning. Since there is generally un-reacted
pyrophoric titanium buried in the deposits, caution should be
used in cleaning. If the desired gas throughput is known
(Q = pumping speed x pressure) the maximum theoretical
operating time is given by:
0.13 T [gr]
____
Operating time [h] = ___

n Q [mbar l/s]

where T is the usable Titanium


For example, using our cryopanel at 1 x 10-8 mbar with a three
filaments cartridge Ti source, the theoretical operating time is
given by:

0.13
___ 3.6 [gr]
_________________ = 46,800 hrs = about 5 years
2 500 [l/s] x 10-8 [mbar]

51
Agilent Ion Pump Service and Support Plan

Agilent Vacuum Technologies Products offers one of the most • End User Extended Warranty
comprehensive service and support plans in the industry. The End User Extended Warranty is a Service Agreement
As a worldwide, first class company we strive to offer added that provides coverage extension beyond the standard 12
value to our Customers wherever they may be located. months warranty of a product. It extends for additional 12
Excellence in Service and Support is a key factor in months the coverage from failures due to defects in
added value. material and workmanship.
Support can be accomplished with qualified Engineers and an By extending the warranty of purchased products, post-
appropriate logistic infrastructure. sales satisfaction is guaranteed. This is the result of the
Technical skills and infrastructure are two areas where we relevant cost decrease while cost predictability is
are continuously making medium and long-term investments. increased.
Our daily customer support philosophy is focused on two The End User Extended Warranty has been designed to
important factors: quick response and no hassle. cover the needs of the High Energy Physics Ultra High
Agilent Vacuum Products is committed to providing our Vacuum applications.
Customers with the most comprehensive service plans • Ordering Information for End User Extended Warranty
possible. High Energy Physics UHV Applications
The following pages detail the standard sections of our Ion
Pump Service and Support Plan. For more information, or if
you have a need for customized solutions, please contact Pumps Part Number
your Agilent representative. Extd. Warr. HEP 12M Ion Pump 20 l/s class EW919VIP20
Often, excellent products are not enough: Agilent’s constant Extd. Warr. HEP 12M Ion Pump 40 l/s class EW919VIP40
presence with Customer Service as a responsive partner is Extd. Warr. HEP 12M Ion Pump 55 l/s class EW919VIP55
what really makes a difference. Extd. Warr. HEP 12M Ion Pump 75 l/s class EW919VIP75
Extd. Warr. HEP 12M Ion Pump 150 l/s class EW919VIP150
Extd. Warr. HEP 12M Ion Pump 300 l/s class EW919VIP300
Worldwide Parts Sourcing Plan Extd. Warr. HEP 12M Ion Pump 500 l/s class EW919VIP500
Extd. Warr. HEP 12M Ion Pump (other config.) EW919VIP
• Exchange and Repair Programs
Agilent’s Ion Pumps and Controllers offer unmatched Controllers Part Number
reliability, performance and cleanliness. Extd. Warr. 12M Minivac Controller EW929MINIVAC
To maximize uptime, and for those occasions where time is Extd. Warr. HEP 12M 4UHV CTRL 1-2 Pumps EW919UHV2P
essential, Agilent offers Ion pump controller exchange units Extd. Warr. HEP 12M 4UHV CTRL 3-4 Pumps EW919UHV4P
for advance shipment. Extd. Warr. 12M TSP Controller EW929TSP
Exchange units are fully remanufactured to the same strict
standards as new products.
If requested, your Exchange unit can be shipped for
overnight delivery. This efficiency is due to our logistic
infrastructure and is the result of excellent teamwork
among our Service and Customer Service Teams!
The Repair Program is available for those situations where
the control of assets is important and when the requested
turn around time is less critical.
The Exchange and Repair Programs allow global OEMs and
End-Users consistent delivery time, pricing, part numbers,
and order processing procedures.

52
ION Pumps
Technical Assistance Americas
North America: Toll-Free +1 800-882-7426
As a Agilent Customer, you can rely on an attentive and Fax +1 781-860-5437
professional support staff, dedicated to providing you with the
Central and South America: Tel. +1 781-861-7200
easiest and most personalized service possible.
Fax +1 781-860-5437
• Customer Support Asia
Our toll-free lines, with mother tongue Technical Support Japan: Toll-Free +81 120-477-111
Engineers at worldwide locations, allow us to provide you Fax +81 120-880-598
with quick, corrective responses to your needs.
Korea: Toll-Free 080-222-2452
Whenever a new problem is identified and solved by our
Fax +82 (0) 2-3452-3947
support personnel, it is entered into our technical support
system and becomes available to all Agilent technical China: Beijing Office Toll-Free 800-820-3278
support centers. This system allows all Agilent locations to Fax +86 (0)10-6439-2765
provide excellent first and second level technical support to Shanghai Office Toll-Free 800-820-3278
customers worldwide. In addition, the technical support Fax +86 (0)21-6628-5169
centers are in daily contact with our R&D departments for Guangzhou Office Toll-Free 800-820-3278
third level support. Fax +86 (0)20-8669-5861
Call us today for more information. Shenzhen Office Toll-Free 800-820-3278
Fax +86 (0)755-8276-3181
Taiwan: Toll-Free 0800-018-768
Singapore: Toll-Free 1800-276-2622
Malaysia: Toll-Free 1800-880-805
India Toll-Free 1800-1801-517

Pumps
Ion
Europe and Israel
Austria, Belgium, Finland, France, Germany, Holland, Ireland,
Israel (*), Italy, Portugal, Spain, Switzerland, UK:
Toll-Free 00-800-234-234-00
Toll-Free-Fax 00-800-345-345-00
(*) From Israel dial 012 instead of initial 00

Other Countries
Tel. +39 011-9979-369
Fax +39 011-9979-330

• Application Support - Application Training


Agilent has a leadership position in vacuum technology
with continuous search for innovative solutions through
research and development.
To meet the most demanding needs, our Application
Engineer Team can bring our knowledge into your factory.
Application Support is a project-based activity where our
experts assist you towards the solution of your application
issues that might arise both at pre and post-sale level.
By designing solutions that meet Customers’ needs,
Agilent aims to create a positive and synergetic
relationship with customers.
Our experts can keep you well-informed and up-to-date
on industrial and scientific applications with the goal of
optimizing the use of our products on your system as well
as the development of new vacuum techniques.

53