1550c Umeng0100 Ohm Meter
1550c Umeng0100 Ohm Meter
1550c Umeng0100 Ohm Meter
Insulation Tester
Users Manual
Title Page
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
How to Contact Fluke . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
Before You Start . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
The Tester . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
Pushbuttons. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Power On/Off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Charge the Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Guard Terminal Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
Connect to the Circuit Under Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Before an Insulation Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Preset Test Voltage Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Program a Test Voltage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Select a Ramp or Steady-State Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Set a Timed Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Polarization Index (PI) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Dielectric Absorption Ratio . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Capacitance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
i
1550C/1555
Users Manual
Insulation Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Store Test Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
View Test Results Stored in Memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Download Test Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Delete Test Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Replaceable Parts and Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
General Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Environmental Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Electrical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Principles of Measurement and Resistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
ii
Introduction How to Contact Fluke
The Fluke 1550C and 1555 Insulation Testers (the Tester or Product) To contact Fluke, use one of these telephone numbers:
are high-voltage insulation testers to validate general circuits, such as
• USA: 1-800-760-4523
switchgear, motors, and cables.
• Canada: 1-800-36-FLUKE (1-800-363-5853)
The Tester features:
• Europe: +31 402-675-200
• Large liquid crystal display (LCD) • Japan: +81-3-6714-3114
• Preset test voltages: 250 V, 500 V, 1000 V, 2500 V, 5000 V, • Singapore: +65-6799-5566
(10 000 V 1555 only)
• China: +86-400-921-0835
• Programmable test voltages: 250 V to 10 000 V (50/100 V steps)
• Brazil: +55-11-3530-8901
• Resistance measurement: 200 kΩ to 2 TΩ
• Anywhere in the world: +1-425-446-5500
• Polarization Index (PI)
Or, visit Fluke's website at www.fluke.com.
• Dielectric Absorption Ratio (DAR or DAR [CN])
To register your Product, visit http://register.fluke.com.
• Ramp mode that linearly increases (100 V/s) the applied test
voltage To view, print, or download the latest manual supplement, visit
http://us.fluke.com/usen/support/manuals.
• Test timer and storage for test results with user-defined ID tag
• Breakdown voltage indication
• Rechargeable lead-acid battery
• Auto shutoff after 30 minutes of inactivity
• Infrared (IR) port for downloading test data
• PC software (supplied)
1
1550C/1555
Users Manual
2
Insulation Tester
Safety Information
• Remove all probes, test leads, and accessories that are • Do not disconnect the test leads before a test has been
not necessary for the measurement. completed and the test voltage at the terminals has
returned to zero. This ensures that any charged
• Keep fingers behind the finger guards on the probes.
capacitance is fully discharged.
• Use the correct terminals, function, and range for
measurements.
• Use the guard terminal only as specified in this manual.
Do not allow other foreign objects to contact the guard
• Place test leads in proper input terminals. terminals as safety can be compromised.
• Do not work alone. • Remove the input signals before you clean the
• Do not use in distribution systems with voltages higher Product.
than 1100 V. • Use only specified replacement parts.
• Use only recommended test leads. • Repair the Product before use if the battery leaks.
• Remove all power from the circuit under test and • Do not operate the Product with covers removed or the
discharge circuit capacitance before testing resistance case open. Hazardous voltage exposure is possible.
or capacitor with the tester.
• Have an approved technician repair the Product.
• Results of measurement can be adversely affected by
the impedances of additional operating circuits
connected in parallel or by transient currents.
• Before and after testing, confirm that the Product does
not indicate the presence of a hazardous voltage. If a
hazardous voltage is shown on the display, remove
power from the circuit under test or allow the
installation capacitance to fully discharge.
3
1550C/1555
Users Manual
WARNING. Do not apply greater than 1100 Volts. s Ramp mode indicator
Measurement Category II is applicable to test and measuring circuits connected directly to utilization points (socket outlets and
similar points) of the low-voltage MAINS installation.
Measurement Category III is applicable to test and measuring circuits connected to the distribution part of the building’s low-voltage
MAINS installation.
Measurement Category IV is applicable to test and measuring circuits connected at the source of the building’s low-voltage MAINS
installation.
This product complies with the WEEE Directive marking requirements. The affixed label indicates that you must not discard this
electrical/electronic product in domestic household waste. Product Category: With reference to the equipment types in the WEEE
~ Directive Annex I, this product is classed as category 9 "Monitoring and Control Instrumentation" product. Do not dispose of this
product as unsorted municipal waste.
4
Insulation Tester
Before You Start
5
6 2
1
7
Item Description
Quick Reference Guide and Safety Information
5
AC Power Cord 7
6
5
1550C/1555
Users Manual
6
Insulation Tester
The Tester
12
Item Description
Guard Terminal Use Figure 3 shows how to prevent surface current leakage with a lead
connected from the Guard terminal to a conductor that surrounds the
Note
inner insulation. The surface leakage current is directed to the Guard
Insulation resistance is measured between the (+) and (-)
terminal. This removes the leakage current from the measurement
output connections. The Guard terminal (G) is at the same
path between the positive and negative terminals, and improves the
potential as the negative (-) terminal but is not in the
accuracy of the test readings.
measurement path.
For most tests, only two test leads are used. Connect the positive (+)
and negative (-) test leads to the corresponding inputs on the Tester. Shield
Shield
Connect the test lead probes to the circuit under test. The Guard (G) ( )
terminal is left unconnected.
For the best accuracy when you measure very high resistances, use Optional
Optional
(G )
three-wire measurements and the Guard terminal. The Guard terminal ( )
is at the same potential as the negative (-) terminal, and can be used Leakage
Leakage
to prevent surface leakage or other unwanted leakage currents from Current
Current
degrading the accuracy of the insulation resistance measurement.
Figure 2 shows how to measure the resistance from one of the Figure 3. Guard Terminal Connection
conductors to the outer shield. In this case, there is a leakage current
along the surface of the inner insulation near the cables end. This Figure 4 shows how to make the measurement setup better. Connect
leakage adds to the current that the negative terminal senses, and the Guard terminal to the unused wire and attach it to the inner
causes the Tester to read a lower resistance than it should. insulation. This ensures that the Tester measures the leakage
between the selected conductor and the outer shield, but removes the
leakage path between conductors.
Shield
Shield
( )
Shield
Shield
( )
( )
Surface Leakage (G )
Surface Leakage
( )
Current
Current Leakage
Leakage
Current
Current
Figure 2. Surface Leakage Current
Figure 4. Improved Guard Terminal Connection
8
Insulation Tester
Measurements
Measurements
Table 6. Test Lead Connections
Common measurement procedures are discussed in this section.
9
1550C/1555
Users Manual
10
Insulation Tester
Measurements
R @ 10 min
PI =
R @ 1 min
11
1550C/1555
Users Manual
12
Insulation Tester
Measurements
13
1550C/1555
Users Manual
14
Insulation Tester
Measurements
Power On/Off
Note
Before the USB-IR cable can be used, software drivers must
be installed on your Windows PC. See the USB-IR
Installation Guide for more information.
15
1550C/1555
Users Manual
The Fluke Connect app works with Apple and Android mobile Maintenance
products. The app is available for download from the Apple App Store
and Google Play. XW Warning
To set up: To prevent possible electric shock, fire, or personal
injury:
1. Connect the ir3000 FC BLE-IR adapter to the IR port on the Tester.
• Do not repair or service your Product beyond what is
2. Turn on the adapter.
described in this manual.
On your mobile device: • Have an approved technician repair the Product.
• There are no user-replaceable parts inside the Product.
1. Go to Settings > Bluetooth. Verify that Bluetooth is turned on.
2. Go to the Fluke Connect app and in the list of connected Fluke Cleaning
tools, select 155x FC.
XW Warning
3. Follow the prompts in the app to continue.
To prevent possible electric shock, fire, or personal
Delete Test Results injury, remove excess water from the cloth before you
clean the Product to ensure that water does not enter
To delete all saved test results:
any terminal.
1. Push to call the Function menu.
Periodically wipe the case with a damp cloth and mild detergent. Do
2. Push or to select the menu item DELETE RESULT. not use abrasives or solvents to clean the Product.
3. Push to call the menu item.
4. Push . REALLY DEL? shows on the display.
5. Push to confirm the deletion or push to return to
Test Voltage.
Note
The Delete function deletes all stored test results. Individual
test locations cannot be deleted but are overwritten.
16
Insulation Tester
Maintenance
17
1550C/1555
Users Manual
General Specifications
Display....................................................................475 mm x 105 mm
Power......................................................................12 V lead-acid rechargeable battery, 2.6 Ahr
Typical Battery Charge Capacity
Number of tests ....................................................... 4100 @ 250 V
3600 @ 500 V
3200 @1 kV
2500 @ 2.5 kV
1000 @ 5 kV
500 @ 10 kV
At temperature extremes.....................................charge the battery more frequently
Charger Input (AC) ................................................85 V to 250 V ac, 50/60 Hz, 20 VA
This Class II (double insulated) instrument is supplied with a Class 1 (grounded) power cord. The protective
earth terminal (ground pin) is not connected internally. The extra pin is for added plug retention only.
Dimensions (H x W x L) ........................................170 mm x 242 mm x 330 mm (6.7 in. x 9.5 in. x 13.0 in.)
Weight ....................................................................3.6 kg (7.94 lb)
Tamper Protection.................................................Kensington lock
Environmental Specifications
Operating Temperature.........................................-20 °C to +50 °C (-4 °F to +122 °F)
Storage Temperature ............................................-20 °C to +65 °C (-4 °F to +149 °F)
Relative Humidity ..................................................80 % to 31 °C decreasing linearly to 50 % at 50 °C
Altitude ...................................................................2000 m
IP Rating.................................................................IEC 60529: IP40
18
Insulation Tester
Environmental Specifications
Safety .....................................................................IEC 61010-1: 600 V CAT IV / 1000 V CAT III Pollution Degree 2
Electromagnetic Compatibility (EMC)
International ........................................................IEC 61326-1: Portable
CISPR 11: Group 1, Class A
Group 1: Equipment has intentionally generated and/or uses conductively-coupled radio frequency
energy that is necessary for the internal function of the equipment itself.
Class A: Equipment is suitable for use in all establishments other than domestic and those directly
connected to a low-voltage power supply network that supplies buildings used for domestic purposes.
There may be potential difficulties in ensuring electromagnetic compatibility in other environments due
to conducted and radiated disturbances.
Caution: This equipment is not intended for use in residential environments and may not provide adequate
protection to radio reception in such environments.
Emissions that exceed the levels required by CISPR 11 can occur when the equipment is connected to
a test object.
Korea (KCC)........................................................Class A Equipment (Industrial Broadcasting & Communication Equipment)
Class A: Equipment meets requirements for industrial electromagnetic wave equipment and the seller
or user should take notice of it. This equipment is intended for use in business environments and not to
be used in homes.
USA (FCC) ..........................................................47 CFR 15 subpart B. This product is considered an exempt device per clause 15.103.
Wireless Radio with Adapter
Frequency Range................................................2412 MHz to 2462 MHz
Output Power ......................................................<100 mW
19
1550C/1555
Users Manual
Electrical Specifications
The Tester accuracy is specified for 1 year after calibration at operating temperatures of 0 °C to 35 °C. For operating temperatures outside the range
(-20 °C to 0 °C and 35 °C to 50 °C), add ±0.25 % per °C, except on the 20 % bands add ±1 % per °C.
Insulation
Test Voltage (DC) Insulation Resistance Range Accuracy (±reading)
<250 kΩ unspecified
250 kΩ to 5 GΩ 5%
250 V
5 GΩ to 50 GΩ 20 %
>50 GΩ unspecified
<500 kΩ unspecified
500 kΩ to 10 GΩ 5%
500 V
10 GΩ to 100 GΩ 20 %
>100 GΩ unspecified
<1 MΩ unspecified
1 MΩ to 20 GΩ 5%
1000 V
20 GΩ to 200 GΩ 20 %
>200 GΩ unspecified
<2.5 MΩ unspecified
2.5 MΩ to 50 GΩ 5%
2500 V
50 GΩ to 500 GΩ 20 %
>500 GΩ unspecified
<5 MΩ unspecified
<5 MΩ to 100 GΩ 5%
5000 V
100 GΩ to 1 TΩ 20 %
>1 TΩ unspecified
<10 MΩ unspecified
10 MΩ to 200 GΩ 5%
10 000 V
200 GΩ to 2 TΩ 20 %
>2 TΩ unspecified
Bar graph range: 0 to 2 TΩ
Insulation test voltage accuracy: -0 %, +10 % at 1 mA load current
Induced ac mains current rejection: 2 mA maximum
Charging rate for capacitive load: 5 s/μF
Discharge rate for capacitive load: 1.5 s/μF
20
Insulation Tester
Principles of Measurement and Resistance
Range Resolution
Timer Setting: 1 minute
0 to 99 minutes
Indication: 1 second
21
1550C/1555
Users Manual
22