Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                
Download as pdf or txt
Download as pdf or txt
You are on page 1of 4

ECG200+

General description

The ECG+ series electrocardiographs are part of centralised patient records to automatically retrieve
Cardioline's new range of devices intended for both their demographics data and collect ECG exams.
private physicians and clinical institutions.
The ECG+ series is designed to offer high quality ECG200+ is easy to use thanks to its user-friendly
services at a reasonable price. graphical interface, the large alphanumeric keypad,
the dedicated menu keys to activate the main
ECG200+ has a high resolution A4 printer, a large 7” functions and the high quality, high resolution 7”
colour display and a large keypad, making it the colour display to show ECG traces in real time.
perfect tool for professional use in wards requiring With the high resolution A4 printer you can print
intensive use. The long-life rechargeable battery ECGs both in manual mode, with up to 12 channels,
and the exclusive ECG acquisition technology ensure and in automatic mode, with a number of formats to
it can be used on a daily basis in wards for patients choose from. You can also retrieve exams saved in
confined to bed. the device's internal memory to open and print
The interpretation program, developed at Glasgow them again directly from the device or to download
University, can analyse electrocardiograms from them to a computer (with a USB connection).
adult and paediatric patients and diagnose ECG
anomalies, such as infarction and acute myocardial
infarction (AMI), ventricular hypertrophy, ST-T and
rhythmic anomalies, as well as conduction faults and
QT interval anomalies, which make ECG200+ the
perfect instrument for emergency scenarios.

ECG200+ can be connected to the most common


data management interfaces (DICOM, HL7, XML)
and can be fully integrated in existing workflows in
healthcare facilities, including the interface with

Main features

Latest-generation front end that ensure high Various export formats and protocols SCP
quality, steady traces, accurate measurements, (standard available), XML-PDF-GDT (included in
quick and precise ECG acquisition exceeding the standard connectivity option), DICOM (included
strictest ECG acquisition standards (AAMI, ANSI, in DICOM connectivity option) and HL7 (optional)
AHA, ACC). to communicate with all standard management
Glasgow interpretation algorithm to guarantee systems.
top performance levels when analysing the ECG Large, high quality colour display to show traces
of adults, children and babies at rest. in real time and to ensure a quick and safe
A number of connectivity features (USB, LAN) for acquisition of the graph.
interaction and bidirectional communication with High resolution A4 thermal printer with
the main management systems for clinical and numerous printing formats.
demographic data. Long-life battery for intensive use in wards.

www.cardioline.com
www.cardioline.com
Easy to use thanks to its user-friendly interface
and alphanumeric keypad.

www.cardioline.com
www.cardioline.com
Sp_ECG200+_Cardiolinespa_05_Eng1.Docx 23/09/2015
Technical specifications

ECG leads................................................
12-leads (I, II, III, aVR-L-F, V1-6)

CMRR................................
CMRR....................................................
....................................................
> 100 dB

DC input impedance .............................


No lead-off 100 MΩ

A/D converter .......................................


.......................................
24 bit, 32 kHz

Sampling frequency ..............................


500 Hz

A/D conversion .......................................


20 bit

< 1uV/LSB
Resolution ..............................................

Bandwidth ..............................................
0.05 – 150 Hz

Pacemaker detection ............................


Hardware detection coupled with convolution digital
filtering

Filters ...................................................
...................................................
Diagnostic fully digital high pass filter; adaptive digital
AC interference filter (50/60 Hz); digital low pass filter
muscolar filter 25 and 40 Hz (only for display and
printing)

Defibrillator protected ..........................


AAMI/IEC standards

Front-
Front-end performance ........................
ANSI/AAMI IEC 60601-2-25:2011

Safety ...................................................
...................................................
EN 60601-1 internal power equipment - class I on
external AC/DC power supply
ANSI/AAMI ES1
CE0476

ECG storage ............................................


Internal memory 100 ECGs, expansion for 1000
additional ECGs (optional)

Display ....................................................
7" backlit LCD colour display, displays the ECG wave in
real time

Thermal printer ......................................


8 dot/mm; 216mm; A4 Z-fold

Manual printing......................................
3, 6 or 12 channels -,5/10/25/50 mm/s

Automatic printing ................................Standard or Cabrera; 3, 3+1, 3+3, 6 or 12 channels,


Patient Demographic, Global Measurements, Optional
Interpretation (Glasgow University – Prof.
Registered office
MacFarlane)
Via Linz,19-20-21
Adult, Paediatric, STEMI
38121 Spini di Gardolo
Keypad ....................................................
Mechanical keypad with alphanumeric keys and (TN) Italy
special function keys T. +39 0463 850125
F. +39 0463 850088
Connectivity ...........................................
USB device, LAN (optional)
Sales Office:
Patient cable...........................................
cable Standard 15D, 10 electrodes
Via F.lli Bronzetti, 8
SCP (standard format), XML-PDF-GDT (included in
Data Export............................................. 20129 Milan, Italy
standard connectivity option), DICOM (included in T. +39 02 94750470
F. +39 02 94750471

www.cardioline.com
www.cardioline.com
Sp_ECG200+_Cardiolinespa_05_Eng1.Docx 23/09/2015
DICOM connectivity option), HL7 (optional)

Power supply ..........................................


Medical grade AC power-supply (100-240 VAC 50/60
Hz); internal rechargeable battery (NiMH)

Dimensions .............................................
396 x 290 x 80 mm

Weight ....................................................
2.6 kg with battery

Registered office
Via Linz,19-20-21
38121 Spini di Gardolo
(TN) Italy
T. +39 0463 850125
F. +39 0463 850088

Sales Office:
Via F.lli Bronzetti, 8
20129 Milan, Italy
T. +39 02 94750470
F. +39 02 94750471

www.cardioline.com
www.cardioline.com
Sp_ECG200+_Cardiolinespa_05_Eng1.Docx 23/09/2015