Experiment 08: Procedure
Experiment 08: Procedure
Experiment 08: Procedure
Procedure:
1. Connect power supply cord to the main power plug & to trainer ST2502 .
2. Make the connections as shown in figure 8.1.
a)Function Generator 1 KHz sine wave output to input socket of emitter Circuit via 4 mm lead.
b)Connect 0.5 m optic fiber between emitter output and detectors input.
c)Connect Detector output to amplifier input socket via 4mm lead.
3. Switch On the power supply of the trainer and oscilloscope.
4. Set the Oscilloscope channel 1 to 0.5 V/ Div and adjust 4-6 div amplitude by using X 1 probe
with the help of variable pot in function generator Block at input of Emitter.
5. Observe the output signal from detector (TP8) on CRO.
6. Adjust the amplitude of the received signal as that of transmitted one with the help of gain
adjusts potentiometer in AC amplifier block. Note this amplitude and name it V 1 .
7. Wind the fiber optic cable on the mandrel and observe the corresponding AC amplifier output
on CRO, it will be gradually reducing, showing loss due to bends
Result
The experiment is to measure the bending loss in the optical fiber.
After Set the Oscilloscope amplitude We see the amplitude of power input and output are
different, because of bending loss