Acquisition System
Resolution .............................................. 8 bits
Bandwidth (3 dB) ................................ 100 MHz max
20 MHz typical with bandwidth
limit enabled
Number of channels ............................... 2 simultaneously sampled,
single ended
Max real-time sample rate...................... 100 MS/s
Max random interleaved sampling
(RIS) sample rate ................................... 2.5 GS/s
Onboard sample memory ....................... 16 MB or 32 MB per channel,
depending on memory option
purchased
Calibrated vertical ranges....................... 25 mV to 25 V in 10% steps
With compliments
Timebase System
Reference clock ......................................10 MHz square wave
Clock accuracy (as master).....................50 ppm
Clock input tolerance (as slave)..............1% minimum
Clock input levels ...................................TTL
Sampling clock frequency ......................100 MHz fixed, data can be
decimated by n where 1<n<100e6
Reference clock sources .........................Backplane 10 MHz reference
clock (NI PXI-5112 only),
RTSI clock line, PFI lines,
and onboard 10 MHz clock.
NI PXI/PCI-5112 Specifications
ni.com
Triggering System
Modes..................................................... Edge, hysteresis, window, digital
Source..................................................... CH 0, CH 1, TRIG, PFI<1..2>,
RTSI<0..6>, PXI Star
(NI PXI-5112 only)
Slope....................................................... Rising/falling
Coupling................................................. DC or AC on CH 0, CH 1, TRIG
Pretrigger depth...................................... Up to 16 MB or 32 MB per
channel, depending on memory
option purchased
Posttrigger depth .................................... Up to 16 MB or 32 MB per
channel, depending on memory
option purchased
Holdoff time........................................... 100 s to 171.79 s
Trigger resolution................................... >1,000 steps in full-scale
voltage range
DC accuracy
CH 0, CH 1 ..................................... 2.5% of range setting
0.5% of offset setting
TRIG ............................................... 500 mV
TRIG input range ................................... 10 V
TRIG input impedance........................... 1 M || 30 pF or 50 ,
software-selectable
TRIG input protection ............................ 42 V (DC + peak AC) for 1 M,
5 Vrms for 50
Acquisition methods
random interleaved sampling (RIS) ....... 200 MS/s to 2.5 GS/s effective
sample rate for repetitive signals
only
Real-time sampling ................................ 1 S/s to 100 MS/s sample rate for
transient and repetitive signals
NI PXI/PCI-5112 Specifications
Calibration
Self-calibration .......................................Self-calibration is done by
software command. The
calibration involves gain, offset,
frequency response, triggering,
and timing adjustment for all
input ranges.
Interval.............................................24 hours, or any time temperature
changes beyond 2 C from
temperature at which last internal
calibration was performed
External calibration.................................Internal reference requires
external recalibration
Interval.............................................1 year
Warm-up time.........................................15 minutes
Power Requirements
+3.3 VDC.................................................0.5 A
+5 VDC...................................................1.5 A
+12 VDC.................................................80 mA
12 VDC.................................................120 mA
Physical Characteristics
Dimensions
PCI...................................................10 cm by 17.5 cm
(3.9 in. by 6.8 in.)
PXI...................................................10 cm by 16 cm
(3.9 in. by 6.3 in.)
I/O Connectors
Analog inputs CH 0, CH 1......................BNC female
Analog trigger TRIG ..............................BNC female
Digital trigger PFI 1................................SMB female
NI PXI/PCI-5112 Specifications
ni.com
Environmental Requirements
Operating temperature............................ 0 C to 40 C
Storage temperature ............................... 20 C to 70 C
Humidity ................................................ 10% to 90%, noncondensing
Maximum altitude .................................. 2,000 m
Pollution Degree .................................... 2
Indoor use only
Safety
This product is designed to meet the requirements of the following
standards of safety for electrical equipment for measurement, control,
and laboratory use:
UL 61010-1
Note For UL and other safety certifications, refer to the product label or visit
ni.com/certification, search by model number or product line, and click the
Electromagnetic Compatibility
Emissions ............................................... EN 55011 Class A at 10 m
FCC Part 15A above 1 GHz
Immunity................................................ EN 61326:1997 + A2:2001,
Table 1
CE, C-Tick, and FCC Part 15 (Class A) Compliant
For full EMC compliance, operate this device with shielded cabling. In addition,
all covers and filler panels must be installed.
Note
NI PXI/PCI-5112 Specifications
CE Compliance
This product meets the essential requirements of applicable European
Directives, as amended for CE marking, as follows:
Low-Voltage Directive (safety).............. 73/23/EEC
Electromagnetic Compatibility
Directive (EMC) ..................................... 89/336/EEC
Refer to the Declaration of Conformity (DoC) for this product for any additional
regulatory compliance information. To obtain the DoC for this product, visit
ni.com/certification, search by model number or product line, and click the
appropriate link the Certification column.
Note
With compliments
NI PXI/PCI-5112 Specifications
ni.com
National Instruments, NI, ni.com, and RTSI, are trademarks of National Instruments
Corporation. Product and company names mentioned herein are trademarks or trade names of their
respective companies. For patents covering National Instruments products, refer to the appropriate
location: HelpPatents in your software, the patents.txt file on your CD, or ni.com/patents.
20032004 National Instruments Corp. All rights reserved.
373495B-01
Oct04