Batch VUV4 Characterization for the SBC-LAr10 scintillating bubble chamber
Authors:
H. Hawley-Herrera,
E. Alfonso-Pita,
E. Behnke,
M. Bressler,
B. Broerman,
K. Clark,
J. Corbett,
C. E. Dahl,
K. Dering,
A. de St. Croix,
D. Durnford,
P. Giampa,
J. Hall,
O. Harris,
N. Lamb,
M. Laurin,
I. Levine,
W. H. Lippincott,
X. Liu,
N. Moss,
R. Neilson,
M. -C. Piro,
D. Pyda,
Z. Sheng,
G. Sweeney
, et al. (6 additional authors not shown)
Abstract:
The Scintillating Bubble Chamber (SBC) collaboration purchased 32 Hamamatsu VUV4 silicon photomultipliers (SiPMs) for use in SBC-LAr10, a bubble chamber containing 10~kg of liquid argon. A dark-count characterization technique, which avoids the use of a single-photon source, was used at two temperatures to measure the VUV4 SiPMs breakdown voltage ($V_{\text{BD}}$), the SiPM gain (…
▽ More
The Scintillating Bubble Chamber (SBC) collaboration purchased 32 Hamamatsu VUV4 silicon photomultipliers (SiPMs) for use in SBC-LAr10, a bubble chamber containing 10~kg of liquid argon. A dark-count characterization technique, which avoids the use of a single-photon source, was used at two temperatures to measure the VUV4 SiPMs breakdown voltage ($V_{\text{BD}}$), the SiPM gain ($g_{\text{SiPM}}$), the rate of change of $g_{\text{SiPM}}$ with respect to voltage ($m$), the dark count rate (DCR), and the probability of a correlated avalanche (P$_{\text{CA}}$) as well as the temperature coefficients of these parameters. A Peltier-based chilled vacuum chamber was developed at Queen's University to cool down the Quads to $233.15\pm0.2$~K and $255.15\pm0.2$~K with average stability of $\pm20$~mK. An analysis framework was developed to estimate $V_{\text{BD}}$ to tens of mV precision and DCR close to Poissonian error. The temperature dependence of $V_{\text{BD}}$ was found to be $56\pm2$~mV~K$^{-1}$, and $m$ on average across all Quads was found to be $(459\pm3(\rm{stat.})\pm23(\rm{sys.}))\times 10^{3}~e^-$~PE$^{-1}$~V$^{-1}$. The average DCR temperature coefficient was estimated to be $0.099\pm0.008$~K$^{-1}$ corresponding to a reduction factor of 7 for every 20~K drop in temperature. The average temperature dependence of P$_{\text{CA}}$ was estimated to be $4000\pm1000$~ppm~K$^{-1}$. P$_{\text{CA}}$ estimated from the average across all SiPMs is a better estimator than the P$_{\text{CA}}$ calculated from individual SiPMs, for all of the other parameters, the opposite is true. All the estimated parameters were measured to the precision required for SBC-LAr10, and the Quads will be used in conditions to optimize the signal-to-noise ratio.
△ Less
Submitted 22 July, 2024; v1 submitted 28 May, 2024;
originally announced May 2024.
-
CUTE: A Cryogenic Underground TEst Facility at SNOLAB
Authors:
Philippe Camus,
Jonathan Corbett,
Sean Crawford,
Koby Dering,
Eleanor Fascione,
Gilles Gerbier,
Richard Germond,
Muad Ghaith,
Jeter Hall,
Ziqing Hong,
Andrew Kubik,
Adam Mayer,
Serge Nagorny,
Payam Pakarha,
Wolfgang Rau,
Silvia Scorza,
Ryan Underwood
Abstract:
Low-temperature cryogenics open the door for a range of interesting technologies based on features like superconductivity and superfluidity, low-temperature phase transitions or the low heat capacity of non-metals in the milli-Kelvin range. Devices based on these technologies are often sensitive to small energy depositions as can be caused by environmental radiation. The Cryogenic Underground TEst…
▽ More
Low-temperature cryogenics open the door for a range of interesting technologies based on features like superconductivity and superfluidity, low-temperature phase transitions or the low heat capacity of non-metals in the milli-Kelvin range. Devices based on these technologies are often sensitive to small energy depositions as can be caused by environmental radiation. The Cryogenic Underground TEst facility (CUTE) at SNOLAB is a platform for testing and operating cryogenic devices in an environment with low levels of background. The large experimental chamber ($\mathcal{O}$(10) L) reaches a base temperature of $\sim$12 mK; it can hold a payload of up to $\sim$20 kg and provides a radiogenic background event rate as low as a few events/kg/keV/day in the energy range below about 100 keV, as well as a negligible muon rate ($\mathcal{O}$(1)/month). CUTE was designed and built in the context of the Super Cryogenic Dark Matter Search experiment (SuperCDMS) which uses cryogenic detectors to search for interactions of dark matter particles with ordinary matter. The facility has been used to test SuperCDMS detectors since its commissioning in 2019. In 2021, it was handed over to SNOLAB to become a SNOLAB user facility after the completion of the testing of detectors for SuperCDMS. The facility will be available for projects that benefit from these special conditions, based on proposals assessed for their scientific and technological merits. This article describes the main design features and operating parameters of CUTE.
△ Less
Submitted 30 January, 2024; v1 submitted 11 October, 2023;
originally announced October 2023.
-
Precision Measurement of the Specific Activity of $^{39}$Ar in Atmospheric Argon with the DEAP-3600 Detector
Authors:
P. Adhikari,
R. Ajaj,
M. Alpízar-Venegas,
P. -A. Amaudruz,
J. Anstey,
G. R. Araujo,
D. J. Auty,
M. Baldwin,
M. Batygov,
B. Beltran,
H. Benmansour,
C. E. Bina,
J. Bonatt,
W. Bonivento,
M. G. Boulay,
B. Broerman,
J. F. Bueno,
P. M. Burghardt,
A. Butcher,
M. Cadeddu,
B. Cai,
M. Cárdenas-Montes,
S. Cavuoti,
M. Chen,
Y. Chen
, et al. (125 additional authors not shown)
Abstract:
The specific activity of the beta decay of $^{39}$Ar in atmospheric argon is measured using the DEAP-3600 detector. DEAP-3600, located 2 km underground at SNOLAB, uses a total of (3269 $\pm$ 24) kg of liquid argon distilled from the atmosphere to search for dark matter. This detector with very low background uses pulseshape discrimination to differentiate between nuclear recoils and electron recoi…
▽ More
The specific activity of the beta decay of $^{39}$Ar in atmospheric argon is measured using the DEAP-3600 detector. DEAP-3600, located 2 km underground at SNOLAB, uses a total of (3269 $\pm$ 24) kg of liquid argon distilled from the atmosphere to search for dark matter. This detector with very low background uses pulseshape discrimination to differentiate between nuclear recoils and electron recoils and is well-suited to measure the decay of $^{39}$Ar. With 167 live-days of data, the measured specific activity at the time of atmospheric extraction is [0.964 $\pm$ 0.001 (stat) $\pm$ 0.024 (sys)] Bq/kg$_{\rm atmAr}$ which is consistent with results from other experiments. A cross-check analysis using different event selection criteria provides a consistent result.
△ Less
Submitted 10 October, 2023; v1 submitted 27 February, 2023;
originally announced February 2023.
-
Snowmass 2021 Scintillating Bubble Chambers: Liquid-noble Bubble Chambers for Dark Matter and CE$ν$NS Detection
Authors:
E. Alfonso-Pita,
M. Baker,
E. Behnke,
A. Brandon,
M. Bressler,
B. Broerman,
K. Clark,
R. Coppejans,
J. Corbett,
C. Cripe,
M. Crisler,
C. E. Dahl,
K. Dering,
A. de St. Croix,
D. Durnford,
K. Foy,
P. Giampa,
J. Gresl,
J. Hall,
O. Harris,
H. Hawley-Herrera,
C. M. Jackson,
M. Khatri,
Y. Ko,
N. Lamb
, et al. (20 additional authors not shown)
Abstract:
The Scintillating Bubble Chamber (SBC) Collaboration is developing liquid-noble bubble chambers for the quasi-background-free detection of low-mass (GeV-scale) dark matter and coherent scattering of low-energy (MeV-scale) neutrinos (CE$ν$NS). The first physics-scale demonstrator of this technique, a 10-kg liquid argon bubble chamber dubbed SBC-LAr10, is now being commissioned at Fermilab. This dev…
▽ More
The Scintillating Bubble Chamber (SBC) Collaboration is developing liquid-noble bubble chambers for the quasi-background-free detection of low-mass (GeV-scale) dark matter and coherent scattering of low-energy (MeV-scale) neutrinos (CE$ν$NS). The first physics-scale demonstrator of this technique, a 10-kg liquid argon bubble chamber dubbed SBC-LAr10, is now being commissioned at Fermilab. This device will calibrate the background discrimination power and sensitivity of superheated argon to nuclear recoils at energies down to 100 eV. A second functionally-identical detector with a focus on radiopure construction is being built for SBC's first dark matter search at SNOLAB. The projected spin-independent sensitivity of this search is approximately $10^{-43}$ cm$^2$ at 1 GeV$/c^2$ dark matter particle mass. The scalability and background discrimination power of the liquid-noble bubble chamber make this technique a compelling candidate for future dark matter searches to the solar neutrino fog at 1 GeV$/c^2$ particle mass (requiring a $\sim$ton-year exposure with non-neutrino backgrounds sub-dominant to the solar CE$ν$NS signal) and for high-statistics CE$ν$NS studies at nuclear reactors.
△ Less
Submitted 29 September, 2022; v1 submitted 21 July, 2022;
originally announced July 2022.
-
The NEWS-G detector at SNOLAB
Authors:
L. Balogh,
C. Beaufort,
A. Brossard,
J. F. Caron,
M. Chapellier,
J. M. Coquillat,
E. C. Corcoran,
S. Crawford,
A. Dastgheibi-Fard,
Y. Deng,
K. Dering,
D. Durnford,
C. Garrah,
G. Gerbier,
I. Giomataris,
G. Giroux,
P. Gorel,
M. Gros,
P. Gros,
O. Guillaudin,
E. W. Hoppe,
I. Katsioulas,
F. Kelly,
P. Knights,
L. Kwon
, et al. (16 additional authors not shown)
Abstract:
The New Experiments With Spheres-Gas (NEWS-G) collaboration intends to achieve $\mathrm{sub-GeV/c^{2}}$ Weakly Interacting Massive Particles (WIMPs) detection using Spherical Proportional Counters (SPCs). SPCs are gaseous detectors relying on ionization with a single ionization electron energy threshold. The latest generation of SPC for direct dark matter searches has been installed at SNOLAB in C…
▽ More
The New Experiments With Spheres-Gas (NEWS-G) collaboration intends to achieve $\mathrm{sub-GeV/c^{2}}$ Weakly Interacting Massive Particles (WIMPs) detection using Spherical Proportional Counters (SPCs). SPCs are gaseous detectors relying on ionization with a single ionization electron energy threshold. The latest generation of SPC for direct dark matter searches has been installed at SNOLAB in Canada in 2021. This article details the different processes involved in the fabrication of the NEWS-G experiment. Also outlined in this paper are the mitigation strategies, measurements of radioactivity of the different components, and estimations of induced background event rates that were used to quantify and address detector backgrounds.
△ Less
Submitted 4 January, 2023; v1 submitted 30 May, 2022;
originally announced May 2022.
-
EXCESS workshop: Descriptions of rising low-energy spectra
Authors:
P. Adari,
A. Aguilar-Arevalo,
D. Amidei,
G. Angloher,
E. Armengaud,
C. Augier,
L. Balogh,
S. Banik,
D. Baxter,
C. Beaufort,
G. Beaulieu,
V. Belov,
Y. Ben Gal,
G. Benato,
A. Benoît,
A. Bento,
L. Bergé,
A. Bertolini,
R. Bhattacharyya,
J. Billard,
I. M. Bloch,
A. Botti,
R. Breier,
G. Bres,
J-. L. Bret
, et al. (281 additional authors not shown)
Abstract:
Many low-threshold experiments observe sharply rising event rates of yet unknown origins below a few hundred eV, and larger than expected from known backgrounds. Due to the significant impact of this excess on the dark matter or neutrino sensitivity of these experiments, a collective effort has been started to share the knowledge about the individual observations. For this, the EXCESS Workshop was…
▽ More
Many low-threshold experiments observe sharply rising event rates of yet unknown origins below a few hundred eV, and larger than expected from known backgrounds. Due to the significant impact of this excess on the dark matter or neutrino sensitivity of these experiments, a collective effort has been started to share the knowledge about the individual observations. For this, the EXCESS Workshop was initiated. In its first iteration in June 2021, ten rare event search collaborations contributed to this initiative via talks and discussions. The contributing collaborations were CONNIE, CRESST, DAMIC, EDELWEISS, MINER, NEWS-G, NUCLEUS, RICOCHET, SENSEI and SuperCDMS. They presented data about their observed energy spectra and known backgrounds together with details about the respective measurements. In this paper, we summarize the presented information and give a comprehensive overview of the similarities and differences between the distinct measurements. The provided data is furthermore publicly available on the workshop's data repository together with a plotting tool for visualization.
△ Less
Submitted 4 March, 2022; v1 submitted 10 February, 2022;
originally announced February 2022.
-
Measurements of the ionization efficiency of protons in methane
Authors:
NEWS-G Collaboration,
:,
L. Balogh,
C. Beaufort,
A. Brossard,
J. -F. Caron,
M. Chapellier,
J. -M. Coquillat,
E. C. Corcoran,
S. Crawford,
A. Dastgheibi-Fard,
Y. Deng,
K. Dering,
D. Durnford,
C. Garrah,
G. Gerbier,
I. Giomataris,
G. Giroux,
P. Gorel,
M. Gros,
P. Gros,
O. Guillaudin,
E. W. Hoppe,
I. Katsioulas,
F. Kelly
, et al. (19 additional authors not shown)
Abstract:
The amount of energy released by a nuclear recoil ionizing the atoms of the active volume of detection appears "quenched" compared to an electron of the same kinetic energy. This different behavior in ionization between electrons and nuclei is described by the Ionization Quenching Factor (IQF) and it plays a crucial role in direct dark matter searches. For low kinetic energies (below…
▽ More
The amount of energy released by a nuclear recoil ionizing the atoms of the active volume of detection appears "quenched" compared to an electron of the same kinetic energy. This different behavior in ionization between electrons and nuclei is described by the Ionization Quenching Factor (IQF) and it plays a crucial role in direct dark matter searches. For low kinetic energies (below $50~\mathrm{keV}$), IQF measurements deviate significantly from common models used for theoretical predictions and simulations. We report measurements of the IQF for proton, an appropriate target for searches of Dark Matter candidates with a mass of approximately 1 GeV, with kinetic energies in between $2~\mathrm{keV}$ and $13~\mathrm{keV}$ in $100~\mathrm{mbar}$ of methane. We used the Comimac facility in order to produce the motion of nuclei and electrons of controlled kinetic energy in the active volume, and a NEWS-G SPC to measure the deposited energy. The Comimac electrons are used as reference to calibrate the detector with 7 energy points. A detailed study of systematic effects led to the final results well fitted by $\mathrm{IQF}~(E_K)= E_K^α~/~(β+ E_K^α)$ with $α=0.70\pm0.08$ and $β= 1.32\pm0.17$. In agreement with some previous works in other gas mixtures, we measured less ionization energy than predicted from SRIM simulations, the difference reaching $33\%$ at $2~\mathrm{keV}$
△ Less
Submitted 25 June, 2022; v1 submitted 24 January, 2022;
originally announced January 2022.
-
Quenching factor measurements of neon nuclei in neon gas
Authors:
L. Balogh,
C. Beaufort,
A. Brossard,
J. -F. Caron,
M. Chapellier,
J. -M. Coquillat,
E. C. Corcoran,
S. Crawford,
A. Dastgheibi Fard,
Y. Deng,
K. Dering,
D. Durnford,
C. Garrah,
G. Gerbier,
I. Giomataris,
G. Giroux,
P. Gorel,
M. Gros,
P. Gros,
O. Guillaudin,
E. W. Hoppe,
I. Katsioulas,
F. Kelly,
P. Knights,
L. Kwon
, et al. (25 additional authors not shown)
Abstract:
The NEWS-G collaboration uses Spherical Proportional Counters (SPCs) to search for weakly interacting massive particles (WIMPs). In this paper, we report the first measurements of the nuclear quenching factor in neon gas at \SI{2}{bar} using an SPC deployed in a neutron beam at the TUNL facility. The energy-dependence of the nuclear quenching factor is modelled using a simple power law: $α$E…
▽ More
The NEWS-G collaboration uses Spherical Proportional Counters (SPCs) to search for weakly interacting massive particles (WIMPs). In this paper, we report the first measurements of the nuclear quenching factor in neon gas at \SI{2}{bar} using an SPC deployed in a neutron beam at the TUNL facility. The energy-dependence of the nuclear quenching factor is modelled using a simple power law: $α$E$_{nr}^β$; we determine its parameters by simultaneously fitting the data collected with the detector over a range of energies. We measured the following parameters in Ne:CH$_{4}$ at \SI{2}{bar}: $α$ = 0.2801 $\pm$ 0.0050 (fit) $\pm$ 0.0045 (sys) and $β$ = 0.0867 $\pm$ 0.020 (fit) $\pm$ 0.006(sys). Our measurements do not agree with expected values from SRIM or Lindhard theory. We demonstrated the feasibility of performing quenching factor measurements at sub-keV energies in gases using SPCs and a neutron beam.
△ Less
Submitted 3 December, 2021; v1 submitted 2 September, 2021;
originally announced September 2021.
-
Pulseshape discrimination against low-energy Ar-39 beta decays in liquid argon with 4.5 tonne-years of DEAP-3600 data
Authors:
The DEAP Collaboration,
P. Adhikari,
R. Ajaj,
M. Alpízar-Venegas,
P. -A. Amaudruz,
D. J. Auty,
M. Batygov,
B. Beltran,
H. Benmansour,
C. E. Bina,
J. Bonatt,
W. Bonivento,
M. G. Boulay,
B. Broerman,
J. F. Bueno,
P. M. Burghardt,
A. Butcher,
M. Cadeddu,
B. Cai,
M. Cárdenas-Montes,
S. Cavuoti,
M. Chen,
Y. Chen,
B. T. Cleveland,
J. M. Corning
, et al. (104 additional authors not shown)
Abstract:
The DEAP-3600 detector searches for the scintillation signal from dark matter particles scattering on a 3.3 tonne liquid argon target. The largest background comes from $^{39}$Ar beta decays and is suppressed using pulseshape discrimination (PSD).
We use two types of PSD algorithm: the prompt-fraction, which considers the fraction of the scintillation signal in a narrow and a wide time window ar…
▽ More
The DEAP-3600 detector searches for the scintillation signal from dark matter particles scattering on a 3.3 tonne liquid argon target. The largest background comes from $^{39}$Ar beta decays and is suppressed using pulseshape discrimination (PSD).
We use two types of PSD algorithm: the prompt-fraction, which considers the fraction of the scintillation signal in a narrow and a wide time window around the event peak, and the log-likelihood-ratio, which compares the observed photon arrival times to a signal and a background model. We furthermore use two algorithms to determine the number of photons detected at a given time: (1) simply dividing the charge of each PMT pulse by the charge of a single photoelectron, and (2) a likelihood analysis that considers the probability to detect a certain number of photons at a given time, based on a model for the scintillation pulseshape and for afterpulsing in the light detectors.
The prompt-fraction performs approximately as well as the log-likelihood-ratio PSD algorithm if the photon detection times are not biased by detector effects. We explain this result using a model for the information carried by scintillation photons as a function of the time when they are detected.
△ Less
Submitted 6 April, 2021; v1 submitted 22 March, 2021;
originally announced March 2021.
-
A novel experimental system for the KDK measurement of the $^{40}$K decay scheme relevant for rare event searches
Authors:
M. Stukel,
B. C. Rasco,
N. T. Brewer,
P. C. F. Di Stefano,
K. P. Rykaczewski,
H. Davis,
E. D. Lukosi,
L. Hariasz,
M. Constable,
P. Davis,
K. Dering,
A. Fijałkowska,
Z. Gai,
K. C. Goetz,
R. K. Grzywacz,
J. Kostensalo,
J. Ninkovic,
P. Lechner,
Y. Liu,
M. Mancuso,
C. L. Melcher,
F. Petricca,
C. Rouleau,
P. Squillari,
L. Stand
, et al. (4 additional authors not shown)
Abstract:
Potassium-40 ($^{40}$K) is a long-lived, naturally occurring radioactive isotope. The decay products are prominent backgrounds for many rare event searches, including those involving NaI-based scintillators. $^{40}$K also plays a role in geochronological dating techniques. The branching ratio of the electron capture directly to the ground state of argon-40 has never been measured, which can cause…
▽ More
Potassium-40 ($^{40}$K) is a long-lived, naturally occurring radioactive isotope. The decay products are prominent backgrounds for many rare event searches, including those involving NaI-based scintillators. $^{40}$K also plays a role in geochronological dating techniques. The branching ratio of the electron capture directly to the ground state of argon-40 has never been measured, which can cause difficulty in interpreting certain results or can lead to lack of precision depending on the field and analysis technique. The KDK (Potassium (K) Decay (DK)) collaboration is measuring this decay. A composite method has a silicon drift detector with an enriched, thermally deposited $^{40}$K source inside the Modular Total Absorption Spectrometer. This setup has been characterized in terms of energy calibration, gamma tagging efficiency, live time and false negatives and positives. A complementary, homogeneous, method is also discussed; it employs a KSr$_2$I$_5$:Eu scintillator as source and detector.
△ Less
Submitted 27 July, 2021; v1 submitted 30 December, 2020;
originally announced December 2020.
-
Copper electroplating for background suppression in the NEWS-G experiment
Authors:
NEWS-G Collaboration,
:,
L. Balogh,
C. Beaufort,
A. Brossard,
R. Bunker,
J. -F. Caron,
M. Chapellier,
J. -M. Coquillat,
E. C. Corcoran,
S. Crawford,
A. Dastgheibi Fard,
Y. Deng,
K. Dering,
D. Durnford,
G. Gerbier,
I. Giomataris,
G. Giroux,
P. Gorel,
M. Gros,
P. Gros,
O. Guillaudin,
E. W. Hoppe,
I. Katsioulas,
F. Kelly
, et al. (26 additional authors not shown)
Abstract:
New Experiments with Spheres-Gas (NEWS-G) is a dark matter direct detection experiment that will operate at SNOLAB (Canada). Similar to other rare-event searches, the materials used in the detector construction are subject to stringent radiopurity requirements. The detector features a 140-cm diameter proportional counter comprising two hemispheres made from commercially sourced 99.99% pure copper.…
▽ More
New Experiments with Spheres-Gas (NEWS-G) is a dark matter direct detection experiment that will operate at SNOLAB (Canada). Similar to other rare-event searches, the materials used in the detector construction are subject to stringent radiopurity requirements. The detector features a 140-cm diameter proportional counter comprising two hemispheres made from commercially sourced 99.99% pure copper. Such copper is widely used in rare-event searches because it is readily available, there are no long-lived Cu radioisotopes, and levels of non-Cu radiocontaminants are generally low. However, measurements performed with a dedicated 210Po alpha counting method using an XIA detector confirmed a problematic concentration of 210Pb in bulk of the copper. To shield the proportional counter's active volume, a low-background electroforming method was adapted to the hemispherical shape to grow a 500-$μ$m thick layer of ultra-radiopure copper to the detector's inner surface. In this paper the process is described, which was prototyped at Pacific Northwest National Laboratory (PNNL), USA, and then conducted at full scale in the Laboratoire Souterrain de Modane in France. The radiopurity of the electroplated copper was assessed through Inductively Coupled Plasma Mass Spectrometry (ICP-MS). Measurements of samples from the first (second) hemisphere give 68% confidence upper limits of <0.58 $μ$Bq/kg (<0.24 $μ$Bq/kg) and <0.26 $μ$Bq/kg (<0.11 $μ$Bq/kg) on the 232Th and 238U contamination levels, respectively. These results are comparable to previously reported measurements of electroformed copper produced for other rare-event searches, which were also found to have low concentration of 210Pb consistent with the background goals of the NEWS-G experiment.
△ Less
Submitted 13 December, 2020; v1 submitted 7 August, 2020;
originally announced August 2020.
-
The liquid-argon scintillation pulseshape in DEAP-3600
Authors:
The DEAP collaboration,
P. Adhikari,
R. Ajaj,
G. R. Araujoand M. Batygov,
B. Beltran,
C. E. Bina,
M. G. Boulay,
B. Broerman,
J. F. Bueno,
A. Butcher,
B. Cai,
M. Cárdenas-Montes,
S. Cavuoti,
Y. Chen,
B. T. Cleveland,
J. M. Corning,
S. J. Daughertyand K. Dering,
L. Doria,
F. A. Duncan andM. Dunford,
A. Erlandson,
N. Fatemighomi,
G. Fiorillo,
A. Flower,
R. J. Ford,
R. Gagnon
, et al. (76 additional authors not shown)
Abstract:
DEAP-3600 is a liquid-argon scintillation detector looking for dark matter. Scintillation events in the liquid argon (LAr) are registered by 255 photomultiplier tubes (PMTs), and pulseshape discrimination (PSD) is used to suppress electromagnetic background events. The excellent PSD performance of LAr makes it a viable target for dark matter searches, and the LAr scintillation pulseshape discussed…
▽ More
DEAP-3600 is a liquid-argon scintillation detector looking for dark matter. Scintillation events in the liquid argon (LAr) are registered by 255 photomultiplier tubes (PMTs), and pulseshape discrimination (PSD) is used to suppress electromagnetic background events. The excellent PSD performance of LAr makes it a viable target for dark matter searches, and the LAr scintillation pulseshape discussed here is the basis of PSD.
The observed pulseshape is a combination of LAr scintillation physics with detector effects. We present a model for the pulseshape of electromagnetic background events in the energy region of interest for dark matter searches. The model is composed of a) LAr scintillation physics, including the so-called intermediate component, b) the time response of the TPB wavelength shifter, including delayed TPB emission at $\mathcal O$(ms) time-scales, and c) PMT response.
TPB is the wavelength shifter of choice in most LAr detectors. We find that approximately 10\% of the intensity of the wavelength-shifted light is in a long-lived state of TPB. This causes light from an event to spill into subsequent events to an extent not usually accounted for in the design and data analysis of LAr-based detectors.
△ Less
Submitted 8 June, 2020; v1 submitted 27 January, 2020;
originally announced January 2020.
-
Precision laser-based measurements of the single electron response of SPCs for the NEWS-G light dark matter search experiment
Authors:
NEWS-G Collaboration,
:,
Q. Arnaud,
J. -P. Bard,
A. Brossard,
M. Chapellier,
M. Clark,
S. Crawford,
E. C. Corcoran,
A. Dastgheibi-Fard,
K. Dering,
P. Di Stefano,
D. Durnford,
G. Gerbier,
I. Giomataris,
G. Giroux,
P. Gorel,
M. Gros,
P. Gros,
O. Guillaudin,
E. W. Hoppe,
A. Kamaha,
I. Katsioulas,
D. G. Kelly,
P. Knights
, et al. (15 additional authors not shown)
Abstract:
Spherical Proportional Counters (SPCs) are a novel gaseous detector technology employed by the NEWS-G low-mass dark matter search experiment for their high sensitivity to single electrons from ionization. In this paper, we report on the first characterization of the single electron response of SPCs with unprecedented precision, using a UV-laser calibration system. The experimental approach and ana…
▽ More
Spherical Proportional Counters (SPCs) are a novel gaseous detector technology employed by the NEWS-G low-mass dark matter search experiment for their high sensitivity to single electrons from ionization. In this paper, we report on the first characterization of the single electron response of SPCs with unprecedented precision, using a UV-laser calibration system. The experimental approach and analysis methodology are presented along with various direct applications for the upcoming next phase of the experiment at SNOLAB. These include the continuous monitoring of the detector response and electron drift properties during dark matter search runs, as well as the experimental measurement of the trigger threshold efficiency. We measure a mean ionization energy of $\mathrm{W}=27.6\pm0.2~\mathrm{eV}$ in $\mathrm{Ne + CH_4}$ $(2\%)$ for 2.8 keV X-rays, and demonstrate the feasibility of performing similar precision measurements at sub-keV energies for future gas mixtures to be used for dark matter searches at SNOLAB.
△ Less
Submitted 24 February, 2019;
originally announced February 2019.
-
First results from the NEWS-G direct dark matter search experiment at the LSM
Authors:
NEWS-G Collaboration,
:,
Q. Arnaud,
D. Asner,
J. -P. Bard,
A. Brossard,
B. Cai,
M. Chapellier,
M. Clark,
E. C. Corcoran,
T. Dandl,
A. Dastgheibi-Fard,
K. Dering,
P. Di Stefano,
D. Durnford,
G. Gerbier,
I. Giomataris,
P. Gorel,
M. Gros,
O. Guillaudin,
E. W. Hoppe,
A. Kamaha,
I. Katsioulas,
D. G. Kelly,
R. D. Martin
, et al. (12 additional authors not shown)
Abstract:
New Experiments With Spheres-Gas (NEWS-G) is a direct dark matter detection experiment using Spherical Proportional Counters (SPCs) with light noble gases to search for low-mass Weakly Interacting Massive Particles (WIMPs). We report the results from the first physics run taken at the Laboratoire Souterrain de Modane (LSM) with SEDINE, a 60 cm diameter prototype SPC operated with a mixture of…
▽ More
New Experiments With Spheres-Gas (NEWS-G) is a direct dark matter detection experiment using Spherical Proportional Counters (SPCs) with light noble gases to search for low-mass Weakly Interacting Massive Particles (WIMPs). We report the results from the first physics run taken at the Laboratoire Souterrain de Modane (LSM) with SEDINE, a 60 cm diameter prototype SPC operated with a mixture of $\mathrm{Ne}+\mathrm{CH}_{4}$ (0.7 %) at 3.1 bars for a total exposure of $9.7\;\mathrm{kg\cdot days}$. New constraints are set on the spin-independent WIMP-nucleon scattering cross-section in the sub-$\mathrm{GeV/c^2}$ mass region. We exclude cross-sections above $4.4 \times \mathrm{10^{-37}\;cm^2}$ at 90 % confidence level (C.L.) for a 0.5 $\mathrm{GeV/c^2}$ WIMP. The competitive results obtained with SEDINE are promising for the next phase of the NEWS-G experiment: a 140 cm diameter SPC to be installed at SNOLAB by summer 2018.
△ Less
Submitted 7 September, 2017; v1 submitted 15 June, 2017;
originally announced June 2017.
-
In-situ characterization of the Hamamatsu R5912-HQE photomultiplier tubes used in the DEAP-3600 experiment
Authors:
DEAP Collaboration,
P. -A. Amaudruz,
M. Batygov,
B. Beltran,
C. E. Bina,
D. Bishop,
J. Bonatt,
G. Boorman,
M. G. Boulay,
B. Broerman,
T. Bromwich,
J. F. Bueno,
A. Butcher,
B. Cai,
S. Chan,
M. Chen,
R. Chouinard,
S. Churchwell,
B. T. Cleveland,
D. Cranshaw,
K. Dering,
S. Dittmeier,
F. A. Duncan,
M. Dunford,
A. Erlandson
, et al. (77 additional authors not shown)
Abstract:
The Hamamatsu R5912-HQE photomultiplier-tube (PMT) is a novel high-quantum efficiency PMT. It is currently used in the DEAP-3600 dark matter detector and is of significant interest for future dark matter and neutrino experiments where high signal yields are needed.
We report on the methods developed for in-situ characterization and monitoring of DEAP's 255 R5912-HQE PMTs. This includes a detaile…
▽ More
The Hamamatsu R5912-HQE photomultiplier-tube (PMT) is a novel high-quantum efficiency PMT. It is currently used in the DEAP-3600 dark matter detector and is of significant interest for future dark matter and neutrino experiments where high signal yields are needed.
We report on the methods developed for in-situ characterization and monitoring of DEAP's 255 R5912-HQE PMTs. This includes a detailed discussion of typical measured single-photoelectron charge distributions, correlated noise (afterpulsing), dark noise, double, and late pulsing characteristics. The characterization is performed during the detector commissioning phase using laser light injected through a light diffusing sphere and during normal detector operation using LED light injected through optical fibres.
△ Less
Submitted 29 January, 2019; v1 submitted 29 May, 2017;
originally announced May 2017.
-
Application of the TPB Wavelength Shifter to the DEAP-3600 Spherical Acrylic Vessel Inner Surface
Authors:
B. Broerman,
M. G. Boulay,
B. Cai,
D. Cranshaw,
K. Dering,
S. Florian,
R. Gagnon,
P. Giampa,
C. Gilmour,
C. Hearns,
J. Kezwer,
M. Kuźniak,
T. Pollmann,
M. Ward
Abstract:
DEAP-3600 uses liquid argon contained in a spherical acrylic vessel as a target medium to perform a sensitive spin-independent dark matter search. Argon scintillates in the vacuum ultraviolet spectrum, which requires wavelength shifting to convert the VUV photons to visible so they can be transmitted through the acrylic light guides and detected by the surrounding photomultiplier tubes. The wavele…
▽ More
DEAP-3600 uses liquid argon contained in a spherical acrylic vessel as a target medium to perform a sensitive spin-independent dark matter search. Argon scintillates in the vacuum ultraviolet spectrum, which requires wavelength shifting to convert the VUV photons to visible so they can be transmitted through the acrylic light guides and detected by the surrounding photomultiplier tubes. The wavelength shifter 1,1,4,4-tetraphenyl-1,3-butadiene was evaporatively deposited to the inner surface of the acrylic vessel under vacuum. Two evaporations were performed on the DEAP-3600 acrylic vessel with an estimated coating thickness of 3.00 $\pm$ 0.02 $μ$m which is successfully wavelength shifting with liquid argon in the detector. Details on the wavelength shifter coating requirements, deposition source, testing, and final performance are presented.
△ Less
Submitted 6 April, 2017;
originally announced April 2017.
-
DEAP-3600 Dark Matter Search
Authors:
DEAP Collaboration,
P. -A. Amaudruz,
M. Batygov,
B. Beltran,
J. Bonatt,
M. G. Boulay,
B. Broerman,
J. F. Bueno,
A. Butcher,
B. Cai,
M. Chen,
R. Chouinard,
B. T. Cleveland,
K. Dering,
J. DiGioseffo,
F. Duncan,
T. Flower,
R. Ford,
P. Giampa,
P. Gorel,
K. Graham,
D. R. Grant,
E. Guliyev,
A. L. Hallin,
M. Hamstra
, et al. (32 additional authors not shown)
Abstract:
The DEAP-3600 experiment is located 2 km underground at SNOLAB, in Sudbury, Ontario. It is a single-phase detector that searches for dark matter particle interactions within a 1000-kg fiducial mass target of liquid argon. A first generation prototype detector (DEAP-1) with a 7-kg liquid argon target mass demonstrated a high level of pulse-shape discrimination (PSD) for reducing $β$/$γ$ backgrounds…
▽ More
The DEAP-3600 experiment is located 2 km underground at SNOLAB, in Sudbury, Ontario. It is a single-phase detector that searches for dark matter particle interactions within a 1000-kg fiducial mass target of liquid argon. A first generation prototype detector (DEAP-1) with a 7-kg liquid argon target mass demonstrated a high level of pulse-shape discrimination (PSD) for reducing $β$/$γ$ backgrounds and helped to develop low radioactivity techniques to mitigate surface-related $α$ backgrounds. Construction of the DEAP-3600 detector is nearly complete and commissioning is starting in 2014. The target sensitivity to spin-independent scattering of Weakly Interacting Massive Particles (WIMPs) on nucleons of 10$^{-46}$ cm$^2$ will allow one order of magnitude improvement in sensitivity over current searches at 100 GeV WIMP mass. This paper presents an overview and status of the DEAP-3600 project and discusses plans for a future multi-tonne experiment, DEAP-50T.
△ Less
Submitted 25 August, 2015; v1 submitted 27 October, 2014;
originally announced October 2014.
-
Improving Photoelectron Counting and Particle Identification in Scintillation Detectors with Bayesian Techniques
Authors:
M. Akashi-Ronquest,
P. -A. Amaudruz,
M. Batygov,
B. Beltran,
M. Bodmer,
M. G. Boulay,
B. Broerman,
B. Buck,
A. Butcher,
B. Cai,
T. Caldwell,
M. Chen,
Y. Chen,
B. Cleveland,
K. Coakley,
K. Dering,
F. A. Duncan,
J. A. Formaggio,
R. Gagnon,
D. Gastler,
F. Giuliani,
M. Gold,
V. V. Golovko,
P. Gorel,
K. Graham
, et al. (57 additional authors not shown)
Abstract:
Many current and future dark matter and neutrino detectors are designed to measure scintillation light with a large array of photomultiplier tubes (PMTs). The energy resolution and particle identification capabilities of these detectors depend in part on the ability to accurately identify individual photoelectrons in PMT waveforms despite large variability in pulse amplitudes and pulse pileup. We…
▽ More
Many current and future dark matter and neutrino detectors are designed to measure scintillation light with a large array of photomultiplier tubes (PMTs). The energy resolution and particle identification capabilities of these detectors depend in part on the ability to accurately identify individual photoelectrons in PMT waveforms despite large variability in pulse amplitudes and pulse pileup. We describe a Bayesian technique that can identify the times of individual photoelectrons in a sampled PMT waveform without deconvolution, even when pileup is present. To demonstrate the technique, we apply it to the general problem of particle identification in single-phase liquid argon dark matter detectors. Using the output of the Bayesian photoelectron counting algorithm described in this paper, we construct several test statistics for rejection of backgrounds for dark matter searches in argon. Compared to simpler methods based on either observed charge or peak finding, the photoelectron counting technique improves both energy resolution and particle identification of low energy events in calibration data from the DEAP-1 detector and simulation of the larger MiniCLEAN dark matter detector.
△ Less
Submitted 12 December, 2014; v1 submitted 8 August, 2014;
originally announced August 2014.
-
Radon backgrounds in the DEAP-1 liquid-argon-based Dark Matter detector
Authors:
P. -A. Amaudruz,
M. Batygov,
B. Beltran,
K. Boudjemline,
M. G. Boulay B. Cai T. Caldwell,
M. Chen,
R. Chouinard,
B. T. Cleveland,
D. Contreras,
K. Dering,
F. Duncan,
R. Ford,
R. Gagnon F. Giuliani,
M. Gold V. V. Golovko,
P. Gorel,
K. Graham,
D. R. Grant,
R. Hakobyan,
A. L. Hallin,
P. Harvey,
C. Hearns,
C. J. Jillings,
M. Kuźniak,
I. Lawson,
O. Li
, et al. (27 additional authors not shown)
Abstract:
The DEAP-1 \SI{7}{kg} single phase liquid argon scintillation detector was operated underground at SNOLAB in order to test the techniques and measure the backgrounds inherent to single phase detection, in support of the \mbox{DEAP-3600} Dark Matter detector. Backgrounds in DEAP are controlled through material selection, construction techniques, pulse shape discrimination and event reconstruction.…
▽ More
The DEAP-1 \SI{7}{kg} single phase liquid argon scintillation detector was operated underground at SNOLAB in order to test the techniques and measure the backgrounds inherent to single phase detection, in support of the \mbox{DEAP-3600} Dark Matter detector. Backgrounds in DEAP are controlled through material selection, construction techniques, pulse shape discrimination and event reconstruction. This report details the analysis of background events observed in three iterations of the DEAP-1 detector, and the measures taken to reduce them.
The $^{222}$Rn decay rate in the liquid argon was measured to be between 16 and \SI{26}{\micro\becquerel\per\kilogram}. We found that the background spectrum near the region of interest for Dark Matter detection in the DEAP-1 detector can be described considering events from three sources: radon daughters decaying on the surface of the active volume, the expected rate of electromagnetic events misidentified as nuclear recoils due to inefficiencies in the pulse shape discrimination, and leakage of events from outside the fiducial volume due to imperfect position reconstruction. These backgrounds statistically account for all observed events, and they will be strongly reduced in the DEAP-3600 detector due to its higher light yield and simpler geometry.
△ Less
Submitted 23 April, 2014; v1 submitted 5 November, 2012;
originally announced November 2012.