Manual TGA701
Manual TGA701
Manual TGA701
TGA701
Thermogravimetric
Analyzer
IN STRUCTION MAN UAL
Software Version 1.2x
Part Number 200-287
August 2007
1 Introduction ...............................................................................1–3
Illustrations .................................................................................1–6
Software License ..........................................................................1–7
Warranty .................................................................................. 1–11
Warning and Caution Symbols...................................................... 1–13
Interference to Other Devices ...................................................... 1–14
Declaration of Conformity ............................................................ 1–15
Accessing the Manual.................................................................. 1–16
About this Manual ...................................................................... 1–17
Equipment Packages................................................................... 1–18
Options..................................................................................... 1–19
Components and Accessories List ................................................. 1–20
Thermocouple Certificate of Compliance ........................................ 1–21
Specifications ............................................................................ 1–22
2 Installation ................................................................................2–1
Illustrations .................................................................................2–2
Analyzer Installation .....................................................................2–3
Gas Connections ..........................................................................2–4
Ventilation Connection ..................................................................2–5
Computer Installation ...................................................................2–6
Software Installation.....................................................................2–8
Copy Protection Key ................................................................... 2–11
Power On Instrument.................................................................. 2–13
Power Off Instrument ................................................................. 2–14
Balance, Pedestal, and Carousel Installation ................................... 2–15
Dual Instrument Installation ........................................................ 2–18
Facility Requirements.................................................................. 2–20
3 Option Installation .....................................................................3–1
External Blower............................................................................3–3
External Balance Installation ..........................................................3–6
8 Diagnostics ................................................................................8–1
Ambients ....................................................................................8–3
Solenoids and Switches .................................................................8–8
Network......................................................................................8–9
Communications ........................................................................ 8–11
Systems Check .......................................................................... 8–13
9 Service .......................................................................................9–1
Software Upgrade ........................................................................9–3
Register the Software ...................................................................9–4
Hardware Calibration ....................................................................9–5
Balance Adjustment.................................................................... 9–10
Balance Test.............................................................................. 9–13
Carousel ................................................................................... 9–15
Furnace .................................................................................... 9–19
Over Temperature Switch Reset ................................................... 9–24
Thermocouple............................................................................ 9–25
Smartline® ................................................................................ 9–28
Troubleshooting ......................................................................... 9–34
10 Illustrations .............................................................................10–1
11 Schematics ...............................................................................11–1
Illustrations ............................................................................... 11–2
TGA701 Cabinet Schematics Part List ............................................ 11–3
12 Index .......................................................................................13–1
I. Definitions
"Licensed Software" shall mean the software supplied with this
Software License, which is software developed and distributed by
LECO.
"Instruction Manual" shall mean a guide describing the use and
operation of the Licensed Software, together with any related
supporting documentation.
"Licensee" shall mean and include any of the Licensee's divisions and
departments within Licensee's organization, but shall not include
wholly or partially owned subsidiaries, affiliates, successors or assigns,
or independent third parties.
VII. Confidentiality
All information and materials regarding the software, provided to
Licensee by LECO which are not generally known in the trade are
confidential, are proprietary, and such information and materials are
furnished solely for the use of Licensee under this Software License.
Licensee agrees not to use or disclose any such information or material
otherwise without prior written consent from LECO and Licensee
further agrees to take all reasonable steps requested by LECO, or
otherwise necessary to protect the confidentiality of all such
information and materials. The foregoing obligations will survive
termination of this Software License, and Licensee agrees to return all
such information and materials to LECO upon request or any non-
renewal or termination of this Software License.
VIII. Miscellaneous
a) This Software License shall be binding upon and inure to the
benefit of the parties hereto and their respective successors and
assigns, provided that Licensee shall not have any right to assign its
rights or obligations under this Software License (whether by operation
of law or otherwise) without the prior written consent of LECO.
b) All notices hereunder shall be in writing and shall be sent by
registered or certified mail postage prepaid to LECO at the addresses
first written above. The date of postmark shall be deemed the date on
which such notice is given.
c) No delay on the part of LECO in exercising any right, power or
remedy hereunder shall operate as a waiver thereof, nor shall any
single or partial exercise of any such right, power or remedy preclude
any other or further exercise thereof or the exercise of any other right,
power or remedy.
Notice
The warranty on LECO equipment remains valid only when
genuine LECO replacement parts are employed. Since LECO has
no control over the quality or purity of consumable products not
manufactured by LECO, the specifications for accuracy of results using
LECO instruments are not guaranteed unless genuine LECO
consumables are employed in conjunction with LECO instruments. If
purchaser defaults in making payment for any parts or equipment, this
warranty shall be void and shall not apply to such parts and
equipment. No late payment or cure of default in payment shall extend
the warranty period provided herein.
LECO Corporation is not responsible for damage to any associated
instruments, equipment, or apparatus nor will LECO be held liable for
loss of profit or other special damages resulting from abuse, neglect,
or use in disregard of instructions. The Buyer, their employees,
agents, and successors in interest assume all risks and liabilities for
the operation, use, and/or misuse of the product(s) described herein
and agree to indemnify, hold harmless, and defend the seller from any
and all claims and actions arising from any cause whatsoever,
including seller's negligence for personal injury incurred in connection
Caution
The instrument should be operated only by technically qualified
individuals who have fully read and understand these instructions. The
instrument should be operated only in accordance with these
instructions.
The operator should follow all of the warnings and cautions set forth in
the manual and the operator should follow and employ all applicable
standard laboratory safety procedures.
This equipment has been tested and found to comply with the limits
for a Class B digital device, pursuant to Part 15 of the FCC Rules and
Regulations. These limits are designed to provide reasonable
protection against harmful interference in a residential installation.
This equipment generates, uses and can radiate radio frequency
energy and, if not installed and used in accordance with the
instructions, may cause interference to radio communications.
However, there is no guarantee that interference will not occur in a
particular installation. If this equipment does cause harmful
interference to radio or television reception, which can be determined
by turning the equipment off and on, the user is encouraged to try to
correct the interference by one or more of the following methods:
• Reorient or relocate the radio or television receiving antenna.
• Increase the separation between the equipment and the radio
or television receiver.
• Connect the equipment to an outlet on a circuit different from
that which the radio or television receiver is currently
connected.
• Consult the radio or television dealer or an experienced
radio/TV technician for help.
Use Bookmarks
After opening the pdf document, click the Bookmarks tab located on
the left side of the screen to access a tree that displays main section
headings. Click a heading to go immediately to the corresponding
section. If a + appears next to a chapter or section, click on it to
display more section headings.
Use Find
Click Edit and then Find on the menu bar to display a search box
where you can enter a word or phrase. Each time you click Next, the
software will highlight the next occurrence of the word or phrase in the
manual. Find is most useful when the word or phrase is used
infrequently in the manual.
Use Search
Click Edit and then click Search on the Menu bar to display a side bar
with a search box. Enter the desired word or phrase and each
occurrence of the word or phrase in context will display in the Results
area. Search makes it easier to find words or phrases that occur
several times in the manual. Search also provides the ability to search
words or phrases in several pdf documents, if applicable. To search the
instruction manual only, select the Search this Pdf Document Only
check box.
Use Links
In the manual, text or a page number highlighted in blue provides a
direct link to the associated topic. In the Index, click on the page
number to go directly to the associated topic.
Items listed below are repeated throughout this manual and are
subject to revision. Please consult the packing slip received with the
instrument.
NOTE For a complete listing of kits and packages, refer to Components and
Accessories List, page 1–20.
604-100-7op
259-043-CPK Software Kit with Key
259-043UPGR Software Kit Upgrade
501-291 Oxygen Regulator
502-335 Foil Crucible Liner (500/pk)
502-452 Standard 1% Green Sulfur Petroleum Coke Reference Material
529-048 Crucibles Cover Large, Ceramic, Glazed (5/pk)
614-442 Large Crucible Turntable
615-763 Smartline® Modem Kit
619-611 Grinder Assembly 110V
619-612 Grinder Assembly 230V
619-995 Bar Code Reader
621-001-110 Replacement Balance Kit
621-136 4" Exhaust Kit
621-145 Magnetic Switch
621-192 Add on 15-Inch Monitor
621-225 115V External Blower Kit
621-263 230V External Blower Kit
621-285 Exhaust Kit Upgrade
621-331 Large Glazed Crucibles
621-353 Exhaust Furnace Adapter Upgrade Kit (SN3564/down)
621-354 4-inch Furnace Exhaust Kit (SN3564/down)
621-362 Installation Qualification Kit
621-363 Operation Qualification Kit
621-433 56K Modem
621-434-110 Deskjet Printer HP6940
621-435 HP6940 Color Printer Cartridge
621-436 HP6940 Black Printer Cartridge
764-216 Helium Regulator
766-036 Pneumatic Regulator
776-978 Glass Sample Scoop
777-046 Heat-Resistant Glove
This section provides the Certificate of Compliance for the following part numbers:
621-183
621-184
621-185
Illustrations .................................................................................2–2
Analyzer Installation .....................................................................2–3
International Installation ..........................................................2–3
Gas Connections ..........................................................................2–4
Ventilation Connection ..................................................................2–5
External Exhaust .....................................................................2–5
Moisture Only Applications ...................................................2–5
Computer Installation ...................................................................2–6
Software Installation.....................................................................2–8
Copy Protection Key ................................................................... 2–11
Copy Protection Key Not Installed ............................................ 2–11
Copy Protection Key Installed While Starting Software ................ 2–11
Copy Protection Key Not Registered ......................................... 2–12
Power On Instrument.................................................................. 2–13
Power Off Instrument ................................................................. 2–14
Balance, Pedestal, and Carousel Installation ................................... 2–15
Dual Instrument Installation ........................................................ 2–18
Instruments Dialog Box .......................................................... 2–19
Transfer an Existing Database ............................................ 2–19
Facility Requirements.................................................................. 2–20
International Installation
For international installation, a HAR (Harmonized) approved line cord
for use in the end country must be obtained. This product requires a
disconnect device that may either be the plug of the power cord or a
disconnect device installed at the installation site if the equipment is
permanently wired. Refer to Figure 2-2, page 2–20.
When the plug on the power supply cord is intended to serve as the
disconnect device, the socket/outlet must be installed near the
equipment and shall be easily accessible. When the wiring is
permanently connected, a disconnect device (such as a circuit breaker
or switch) must be lockable, installed near the equipment, and be
easily accessible.
NOTE All gasses should be dry and oil-free. Refer to Specifications, page
1–22, for additional information.
1. Remove the cap plugs from the nitrogen, oxygen, and pneumatic
connections on the rear of the analyzer.
2. Connect the nitrogen supply.
A. Attach the nitrogen connector to the nitrogen regulator on
the nitrogen supply.
B. Attach the other end of the connector to the Nitrogen In
connection on the rear panel of the analyzer. Refer to
Figure 2-1, page 2–20.
C. Set the nitrogen gas supply to 35 psi (2.41 bars).
3. Connect the oxygen supply.
A. Attach the oxygen connector to the oxygen regulator on the
oxygen supply.
B. Attach the other end of the connector to the Oxygen In
connection on the rear panel of the analyzer. Refer to
Figure 2-1, page 2–20.
C. Set the oxygen gas supply to 35 psi (2.41 bars).
4. Connect the pneumatic supply.
A. Attach the pneumatic connector to the pneumatic regulator
on the pneumatic supply.
B. Attach the other end of the connector to the Pneu In
connection on the rear panel of the analyzer. Refer to
Figure 2-1, page 2–20.
C. Set the pneumatic gas supply to 45 psi (3.10 bars).
External Exhaust
NOTE If the computer system was purchased from LECO as part of the
TGA701 package, then the application software has already been
installed.
8. Click Yes. The computer will be restarted. Upon restart, the LECO
USB driver will be installed into the system.
9. Insert the Copy Protection Key into the USB port on the back of
the computer when prompted. Refer to Figure 2-3, page 2–21.
Refer to Copy Protection Key, page 2–11.
10. Remove, but do not dispose of, the software installation CD-ROM
from the CD-ROM disk drive (typically D:). Place in a secure
location for future use.
1. Ensure that all gas supplies are connected and enabled. Refer to
Gas Connections, page 2–4, for additional information.
2. Ensure that the USB cable is connected securely to the analyzer
and the computer. Refer to Figure 2-1, page 2–20, and
Figure 2-3, page 2–21.
3. Connect the computer, monitor, and printer to facility power. It is
strongly recommended that all the components of the computer
system are connected to the same outlet and that a surge
protector is used for the computer system.
4. Turn the computer, monitor, and printer on and wait for the
operating system to load.
5. Log in to the operating system.
6. Confirm that the LECO TGA701 software has been installed.
If the software is not installed, refer to Software Installation,
page 2–8, for additional information.
7. Turn the analyzer On using the On/Off switch. Refer to Figure
10-1, page 10–3.
8. Verify that the Copy Protection Key has been installed, page
2–11.
9. Activate the TGA701 software by double clicking the TGA701
desktop icon or by using the TGA701 menu. Refer to Getting
Started, page 4–4.
NOTE
If the software is not communicating properly, an error message will
display. Refer to Computer not Communicating with Instrument, page
9–36, for additional information.
10. The Main Window displays. Refer to Software Main Window, page
4–4.
NOTE If there are problems opening the software, refer to Troubleshooting,
page 9–34, for additional information.
11. After approximately 10 seconds, the word IDLE will display on the
status bar.
NOTE Do not adjust the height of the internal adjustments for the balance
assembly. This has been adjusted for proper alignment with the
furnace and the sample carousel.
12. Tighten the upper two pneumatic actuator coupling screws using
a 7/64 Allen wrench. Make certain that the lower two pneumatic
actuator coupling screws are tight and that the coupler is seated
against the home position dial.
13. Reinstall the analyzer front panel. Refer to Figure 10-25, page
10–22.
A. Reconnect the J401, J402, and J403 connectors to the
actuator button mounted on the front of the casting.
Gently push the analyzer front panel back and snap into the
side panels.
B. Reinstall the two Allen screws on the top of the front panel.
C. Tighten the two Allen screws on the bottom of the front
panel.
The dual instrument procedure should be used when the Dual TGA701
Thermogravimetric Analyzer package (TGA701D4C Package), page
1–18, was purchased. For installation outside the U.S.A., refer to
International Installation, page 2–3, for additional information.
1. Obtain the serial number from the back of each unit and enter
this information into the Serial Number column.
2. Click Add All to display the Network Address and Database
associated with each unit.
3. To Add or Remove units, click Add to add additional units (no
more than two total). Click Remove to delete a unit from the
Instruments dialog box.
4. Click in the Start column for each unit that you would like to run.
The Status column will display whether a unit is available or
running.
5. Close the Instruments dialog box by clicking Cancel or Continue.
Click Cancel to close the Instrument dialog box without making
any changes. Click Continue to accept changes made in
Instrument dialog box.
COUNTER TOP *
OPTIONAL PC
PRINTER TOWER
22-inches
TGA-701 (56 cm)
MONITOR
KEYBOARD
63-inches
(160 cm)
FLOOR
* 22-inches *
(56 cm)
62-inches to 80-inches
(157 cm to 203 cm)
Figure 2-1
Facility Requirements
PLUG BROWN
BROWN T
POWER Y Y
115 V~ GREEN/YELLOW GREEN/
GROUND G G YELLOW G W
230 V~ 115 V~
POWER X X BLUE
W W X
BLUE
CHASSIS
(CABINET)
RECEPTACLE GROUND
Figure 2-2
Wiring Diagram
KEYBOARD
(SERIAL PORT)
PRINTER
(USB PORT)
ANALYZER #1
(USB PORT)
COPY (SOFTWARE)
ANALYZER #2 PROTECTION KEY
(USB PORT) (USB PORT)
COM2 PORT
COM1 PORT
Figure 2-3
Computer Installation
External Blower............................................................................3–3
Installation .............................................................................3–3
Maintenance ...........................................................................3–4
Periodic Maintenance Schedule .............................................3–4
Performing External Blower Filter Maintenance........................3–4
External Balance Installation ..........................................................3–6
Prepare the External Balance ....................................................3–6
Set the External Balance ..........................................................3–6
Test the External Balance .........................................................3–7
Installation
The optional external blower is intended for customers who have an
external passive vent that does not satisfy the installation
requirements for the TGA701.
The instructions in the following section apply to both the 115V option
and the 230V option. Refer to Options, page 1–19, and Components
and Accessories List, page 1–20.
NOTE The external blower should be powered on whenever the analyzer is
also powered on.
1. Locate the external blower in a convenient location between the
right side of the TGA701 Analyzer and the external exhaust.
2. Position air filter in the top of external blower unit.
Maintenance
Figure 3-1
Balance Control Panel
TGA701 Menu
From the Start menu, click Programs/LECO/TGA701. The TGA701
menu is explained below.
The main window for the TGA701 software displays the Menu bar,
spreadsheet, results display, and plot. The Toolbar and Status bar will
display if enabled on the View menu, page 4–9.
Sample filters can be created to determine which sample results are
displayed in the spreadsheet. For example, sample results can be
grouped by method or by analysis date. For further information about
filters, refer to Determine Sample Filters, page 5–34.
The Title bar identifies the software and provides the buttons to
minimize, maximize, and close the software main window. The title bar
also identifies the filter that is currently applied. In the illustration
above, this filter is All Samples. Refer to Determine Sample Filters,
page 5–34.
The Menu bar, page 4–8, provides menus that display commands to
perform varous functions in the software.
The Toolbar provides quick access to frequently used functions. Refer
to Toolbar, page 4–12.
The Spreadsheet displays information about samples. The columns
displayed in the spreadsheet will vary, depending on the columns
designated as visible. Refer to Modify Columns, page 4–17, for further
information. The Analysis Date column has three display types:
aborted, date/time, and analyzing. When the analysis completes, the
date and time of analysis will display. When the analysis is aborted,
Aborted displays.
The Plot displays a graph of the weight loss versus time and furnace
temperature versus time. The reporting units that display in the plot
can be changed using the Elements dialog box. Refer to Configure Plot,
page 4–21.
A point sensitive ToolTip will display in the plot when the cursor is
held at a point on a plot line.
The first two lines in the ToolTip display the analysis time and furnace
temperature for that point in the analysis and are point sensitive to the
X axis. Any remaining lines in the ToolTip are displayed in the same
order as they display on the Plot Properties tab. Refer to Properties,
page 4–22.
Menu Bar
Click on any title on the Menu Bar to access a menu. The following
section explains the commands for each menu available from the Menu
bar.
Toolbar
Toolbar
Action
Button
F3 Login Creates a new entry in the sample pane.
Starts the analysis for the first batch of unanalyzed
F5 Analyze
sample.
Enters the weight from the external balance into the
F4 Balance
spreadsheet. Used with preweighed crucibles.
F6 Abort Stops the current analysis.
F7 Cover Raises and lowers the furnace cover.
F8 Carousel Rotates the carousel by one position.
F9 Home Locates the carousel home position.
Click to access a pdf (portable document file) of this
F1 Info
instruction manual.
Status Bar
Function
Segment
Delete Samples
Delete removes a selected sample or samples from the spreadsheet. If
logging was enabled when the database was created, refer to Archive
Log, page 4–59. When deleting a number of older samples from the
database, it is better to create an Archive Log File. Refer to Add a
Connection to a Log Archive Database, page 4–64.
1. On the main spreadsheet, select the sample to be deleted by
clicking on the row number to the left of the sample.
2. From the Samples menu, click Delete. The Delete Selected
Sample dialog box is displayed asking "Are you sure you want to
delete the selected samples?"
3. Click Yes to delete; click No to cancel.
Modify Font
The following section explains how to configure the font that displays
in a table or in the plot.
Recalculate
Click Recalculate to adjust the results of a previously analyzed sample
based on a new calibration.
1. Select the samples to be recalculated.
2. From the Samples menu, click Recalculate. A dialog box is
displayed verifying the results should be recalculated.
3. Click Yes to recalculate. The samples are automatically
recalculated. Click No to cancel recalculation.
Sample Signatures
Sign
Used to apply a signature to samples. Sign is only available if logging
was enabled when the database was created. Refer to Database
Management, page 4–60.
1. Select the desired sample in the spreadsheet.
In the Sign dialog box, System displays the name assigned to the
computer that identifies the computer on a network.
User displays the name of the user logged onto the Windows
Operating System.
Date and Time display the date and time the signature was
applied to the sample.
Local or UT will be checked depending on the selection made for
the presentation time. Refer to Presentation Time, page 4–56.
3. For Name, enter the name of the person applying the signature to
the sample.
4. For Description, enter information to help identify the sample, if
desired.
Label Points
When selected, points on the plot are labeled to show the order in
which they were selected from the Spreadsheet.
To label points on the plot,
1. Select multiple samples in the Spreadsheet.
2. Click in the desired plot.
Display Statistics
When Statistics is checked, the plotted result curves change to plotted
result points. A point is displayed for each sample and represents the
peak concentration. To enable or disable statistics,
1. Select multiple samples from the spreadsheet.
2. Click in the desired plot and select Statistics from the menu.
Plot Properties
The following section explains how to configure the plot display by
using the plot properties. To access the Properties tab, right click on
the plot and then select Properties. Refer to Configure Spreadsheet,
page 4–14 for further information about the Fields and Font Tabs.
Show Plot
1. If the Elements tab is not already displayed, right click in the plot
and then select Properties.
2. Select the desired plot from the Hidden Plots selection box by
clicking on it once.
3. Click Show. The plot will move to the Visible Plots Selection box.
4. Click OK.
Hide Plot
1. If the Elements tab is not already displayed, right click in the plot
and then select Properties.
2. Select the desired plot from the Visible Plots selection box by
clicking on it once.
3. Click Hide. The element will move to the Hidden Plots selection
box.
4. Click OK.
Significant Digits
Enter the number of significant digits in the entry box.
Spreadsheet Procedures
Navigate Spreadsheet
The spreadsheet can contain numerous rows and columns, so the
following commands can be used to easily navigate the data in the
spreadsheet. To access these commands, right click on either the
column scroll bar to page through the columns or the row scroll bar to
page through the rows.
Resize Columns
Columns can be sized to properly display sample data. The grid cells
can be resized by selecting one side of the column title frame and
dragging the frame to the desired size. Double clicking on the divider
line between the columns sizes the column to the largest entry.
Auto Widths
Columns can be sized automatically to the width of each cell in order
to accommodate the text inside the cell. From the Configuration menu,
click Auto Widths. The columns are automatically adjusted.
Enter Text
Text can be entered and edited right in the spreadsheet by either
entering data or selecting data from a list. Analysis results, however,
cannot be edited.
View Balance
The Balance display allows the user to monitor the amount of sample
that is being loaded into a crucible. To access the Balance display, click
Balance on the View menu.
The Balance display will show a target mass for a sample that is based
on the specified minimum weight and maximum weight values. Refer
to the System tab, page 4–26, for further information.
System Tab
1. From the Configuration menu, click System. The System
Configuration dialog box will display.
Equation Defined as
Value
cvi,j The current comparator value for
sample j
minitial,j The initial sample weight for sample j
mstep,j The weight loss for sample j for the
current method step
mi-1,j The preceding sample weight for sample
j
mi,j The current sample weight for sample j
D. For Minimum Weight, enter the minimum sample weight
allowed for an analysis.
E. For Maximum Weight, enter the maximum sample weight
allowed for an analysis.
Balance Instability
In most cases, balance instability is the result of trying to initiate the
next analysis before the furnace has completely cooled or reached
equilibrium. When the next analysis is initiated, the cover closes and
residual heat from the furnace will cause the furnace temperature to
increase and cause a balance instability. Balance instability can also be
the result of warm-up times associated with a balance and ambient
temperatures inside the instrument. Most modern analytical balances
have an auto zero feature built into the software that will maintain a
zero reading in the absence of user intervention. In the TGA701, this
feature is disabled since it is necessary to know the balance drift so
that it can be corrected during the course of the analysis. The Internal
Balance tab can be used to set the maximum zero deviation. Refer to
Internal Balance Tab, page 4–28.
1. For Port, click the arrow and select the port location from the list.
Default is COM2.
2. For Baud Rate, click the arrow and select the baud rate from the
list. Default is 1200 bps.
3. For Data Bits, click the arrow and select the number of data bits
from the list. Default is 7.
4. For Parity, click the arrow and select either odd or even. Default
is Even.
5. For Stop Bits, click the arrow and select either 1 or 2 from the
list. Default is 1.
6. For Readability, click the arrow and select 4, 5, or 6 from the list.
7. For Balance Mode, click the arrow and select either Normal or
Disabled from the list. Select Normal when a balance is connected
to system. Select Disabled when a balance is not connected to
system.
Balance Mode Settings
Very Stable: Operation in very stable conditions
Stable: Operation in normal conditions
Unstable: Operation in unstable conditions
Very unstable: Operation in very unstable conditions
Audible Alarm
To select an audible alarm to sound when an error occurs, from the
Configuration menu, select System, and then select the Sound Alarm
on Errors check box. When the check box is cleared, an alarm will not
sound for an error.
Audible Sounds
The TGA701 has an audible sound indicator when the lid opens or
closes, when it is time to load a sample, and when it is time to add or
remove crucible covers.
Filename Tab
Use the Filename tab to configure the file for the export.
1. For Filename, enter a name for the file or select an existing file by
clicking Browse.
2. Determine how the software will handle the file by selecting one
of the following check boxes:
Select Append to File, and the highlighted sample analysis data
will be attached to the previously saved file.
Select Automatically Increment Filename, and if the last
characters in the filename contain numbers, they will
automatically be incremented by one when the file is saved.
Select Prompt for Filename, and the export filename dialog box
will display and prompt the operator to enter a filename.
Fields Tab
Use the commands on the Fields tab to determine the information that
is exported.
Delimiters Tab
A delimiter is a character that separates fields of data. Use the
parameters on this tab to configure the display of the exported data.
1. Select the XML Format check box, and the data will display in XML
format.
2. Select the Column Headers check box, and the data will be
organized under column headers.
3. Select the Field Delimiter by clicking in the desired check boxes to
determine how the exported data will be separated.
Automation Tab
Select the Automatically Export Each New Analysis check box, and
each new analysis will export after analysis has completed.
Select the Advanced tab to enter and modify the fields that are
exported. The Advanced tab provides almost the same function as the
Fields and Preview tabs except that in the Advanced tab, export
information can be changed directly in the Export Format box. When
export information is changed using the Advanced tab, the Fields tab
and Preview tab automatically update.
To add a field to the Export Format, select the desired field from the
Fields selection box and then select Add Field to Transmit Format. The
field appears in the Export Format selection box.
To change the export format, select in the Export Format dialog box
and enter new information. To add a field to the Export Format, select
the desired field from the Fields selection box and then select Add Field
to Transmit Format. The field appears in the Export Format selection
box.
Numeric Field Formatting allows you to quickly enter number
formatting for all of the fields selected for export. Number Width
specifies the number of characters that will appear in the number
format and Precision specifies the number of places that display after
the decimal point.
To change these values, just enter a new value in the box for Precision
and/or for Width and then select Set to display the format in the
Export Format area. For example if 8 is entered for Number Width and
3 is entered for Number Precision, 8.3 will display for each field with
numeric value results in the Export Format area as shown below.
Preview Tab
Use the Preview tab to display the appearance of the data when it
exports. Any changes can then be made before the data is actually
exported. The following example displays the data with the Column
Headers check box checked, Comma checked for Delimiter, None
checked for start of Transmission, and the Carriage Return/Line Feed
(CR/LF) check box checked for End of Transmission. To change the
data display, refer to the Delimiters tab, page 4–36.
Fields Tab
Use the Fields tab to determine the information to transmit.
To add a field to transmit, select the field in Fields and then click Add.
The Field will automatically move to the Transmit column.
To remove a Field so that it will not transmit, select the field in
Transmit and then click Remove. The Field will automatically move to
the Fields column.
For Average, click the arrow and Select Off to transmit the individual
results for each sample. Select On to calculate and then transmit an
average value for each field selected. Select Group by Sample Name
Prefix to calculate an average value for each field and then transmit
the fields grouped by sample name.
Click Promote to move a field to be transmitted before another field.
Select the Field in the Transmit column and click Promote until the
field reaches desired position.
Click Demote to move a field to be transmitted after another field.
Select the Field in the Transmit column and click Demote until the field
reaches desired position.
1. Select the XML Format check box, and the data will display in XML
format.
2. Select the Column Headers check box, and the data will be
organized under column headers.
3. Select the Field Delimiter by clicking in the desired check box to
determine how the transmitted data will be separated.
Select comma, and the transmitted data is separated by
commas.
Select Tab, and the transmitted data is separated by a blank
space equivalent to one tab.
Select Other, and the transmitted data is separated by a
punctuation mark or symbol determined by the user.
4. Select the delimiter for the Start of Transmission.
Select None, and nothing will display before transmitted data.
Select Other and then type in the desired delimiter or click the
arrow on the menu to select a letter string that will display before
the transmitted data.
5. Select the delimiter for the end of transmission.
Select Carriage Return to have data followed by a carriage
return.
Select Carriage Return Followed by a line feed (CR/LF) to have
data followed by a carriage return and line feed.
Select Other and indicate the text that should display or click the
arrow on the menu to the letter string that will display after the
transmitted data.
Select the Automatically Transmit Each New Analysis check box, and
each new analysis will transmit after analysis has completed.
Advanced Tab
Select the Advanced tab to enter and modify the fields that are
transmitted. The Advanced tab provides almost the same function as
the Fields and Preview tabs except that in the Advanced tab, transmit
information can be changed directly in the Transmission Format box.
When transmit information is changed using the Advanced tab, the
Fields tab and Preview tab automatically update.
To add a field to the Transmission Format, select the desired field from
the Fields selection box and then select Add Field to Transmit Format.
The field appears in the Export Format selection box.
To change the transmit format, enter new information in the
Transmission Format box.
Preview Tab
When a sample or group of samples is selected in the spreadsheet, use
the Preview tab to display the appearance of the data before it is
transmitted. Any changes can then be made before the data is actually
transmitted. To change the display of the data, refer to the Delimiters
tab, page 4–42.
Export Buffer Format provides the means to select the columns and
specify the data format that is to be used when exporting sample data
buffers. The export buffer format that displays is based on the current
spreadsheet selection.
If it is necessary to return the export buffer format to the default
format, click the Set Defaults button.
2. Click Add to insert a new row into the Sample Templates table.
3. Click in the Method dialog box and then click the arrow to select
the method for the sample template.
4. Enter comments, if desired.
5. Enter a description, if desired.
6. Use the buttons on Sample Templates dialog box to determine
the sample template parameters. Refer to Dialog Box Buttons,
page 4–15.
Comparison Functions
The following table explains the comparisons that can be used in a
formula.
Function Definition
+ Plus
- Minus
* Multiply
/ Divide
^ Raise to the power
= Equal
> Greater than
>= Greater than or equal to
< Less than
<= Less than or equal to
Format Examples
*Format
Value Format for Result Result
Type
A 4.50360 A 4.50360000
0.123456 F4.5 0.1235
0.000789 F3.5 0.00079
Fn.m
0.000789 F3.2 0.00
321.8765 F3.5 322
0.123456 S4 1.235x10-1
Sn 0.000789 S3 7.89x10-4
0.000789 S2 7.9x10-4
321.8765 I3 322
In
321.8765 I2 320
* A = Default format
n = Total number of significant digits to be represented
m = Maximum number of digits after the decimal
I = Integer notation
S = Scientific notation: An exponential format (y.yyy x 10e) in which any number is
expressed as a number between 1 and 10 multiplied by a power of 10 that indicates
the correct position of the decimal in the original number.
F = Fixed point notation: Traditional decimal notation
@ABS(exp)
Absolute value
Example: @ABS([Moisture %]-[Ash %])
@Log10(exp)
Example: Log base 10
@Log10([Moisture %])
@Ln(exp)
Example: Natural log
@Ln([Moisture %])
@SQRT(exp)
Square root
Example: @SQRT([Moisture %])
@Maximum(exp1,exp2)
Finds Maximum
Example: @Maximum([Moisture %],[Ash %])
@Minimum(exp1,exp2)
Example: Finds Minimum
@Minimum([Moisture %],[Ash %])
Presentation Time
This feature is only available if logging was enabled when the database
was created. It is used to view the database as it existed at some
point of time in the past. For example, it could be used to view results
before a change in calibration was performed. When the presentation
time is set, changes are not allowed to the database, so samples
cannot be added and changes cannot be made to methods, etc.
NOTES After the Presentation Time and date are selected, the results from
that date will appear in the spreadsheet. Logging is disabled by
default.
A. For Date, click the arrow and then select the date from the
list that displays.
B. For Time, use the up and down arrows to determine the
presentation time.
C. Select either Local or UT (Universal Time). Select Local to
display the local time zone. Select UT to display the
universal time that corresponds to the UTC (Universal
Coordinated Time) as defined by the National Bureau of
Standards.
D. Select the Live check box and the other presentation fields
are disabled. When Live is not checked, the other
Presentation Time fields are activated. In this case, the time
and date are used to specify the moment in time to display
data using the Local or UT selections. The normal mode of
operation is to have the Live check box checked.
2. Uncheck the Live check box in order to activate the Date and
Time dialog boxes.
View History
This feature is only available if logging was enabled when the database
was created. View History displays a log of database changes that
could be used to identify the time/date when some particular change
took place. Since there is no description of what each change
represents, it can only be used to identify who made a change.
1. Click the Database menu and select View History.
Archive Log
This feature is only available if logging was enabled when the database
was created. Archive Log is used for copying log entries to another file
or reducing the amount of data stored in the database by offloading
old log entries.
1. Click the Database menu and select Archive Log. The Archive Log
screen will display.
2. Click the arrow and select the Stop Date. The Stop Date specifies
the cutoff date for log entries that are placed into the file. Any log
entries generated on or before this date will be placed into the
output file.
3. Select or clear the Delete Entries on Completion check box. When
this check box is selected, entries are deleted from the database
after they are copied to the output file. This checkbox should be
selected in order to reduce the amount of data stored in database
files. Note that the file size on disk may not change as a result of
this, but space will be made available inside the file for more data
to be stored.
View Size
View Size displays the file size of the database.
Database Management
Use the instructions in Database Management to access the Start
dialog box in order to add a new database connection, to select an
existing connection, to compact/repair a database, or to delete a
database connection.
2. Click Start, click Programs, click LECO, and then click TGA701.
3. From the TGA701 menu, select Database Management.
Database Validation
The software will perform database validation when the Validate check
box is checked on the Start dialog box. Refer to step 6, page
4–62. During database validation, the database files will be reviewed
for corruption. If orphan objects are found, refer to Orphan Objects,
page 4–63.
The Database Validation dialog box will display the status of the
database validation.
3. For Read Only, select the check box, and the database
information can be viewed but not changed.
4. For Log Activity, select the check box, and database activity will
be tracked. Logging can be disabled to save disk space, but then
a record of database activity will not be available. When logging is
disabled, log functions are not available. Refer to Database Menu
Functions, page 4–56.
Figure 5-1
Actuator Button Location
Control/Indicator Action
Enables and disables the power to the instrument. The
power switch is also a circuit breaker that provides
Power On/Off Switch
protection for the operator and instrument. Refer to
Figure 10-1, page 10–3.
The following section provides a brief overview of the steps that must
be followed in order to analyze samples using the TGA701.
Predefined Methods
The following is a list of predefined methods provided in the software.
For Testing the For Training the
For Running Samples:
Balance: Furnace:
ASTM D5142 Moisture Volatile Ash Quartz Test Furnace Training
ASTM D5142 Moisture Ash Balance Test Furnace Tuning
ASTM D5142 Moisture
ASTM D5142 Volatile
Flour Ash
Loss On Ignition
12. Enter the Result name, or click the arrow to select a Result name
from the list. Click OK, and the method result calculation will be
added to the table. In the table, the Result Name will display in
the Equation Name column. The equation name should match the
step name. The equation name references a particular method
result calculation in other software operations and printouts. Each
equation name must be unique to avoid ambiguity in calculations
or other software references.
For example: For the Equation Name Moisture, the software
will generate the following columns in the spreadsheet:
Moisture: The calculated percent weight loss for the step.
Moisture (Certified %): The user can specify a known
certified value for the standards
Moisture (+/-)%: The user can specify a known uncertainty
value for the standard.
Moisture (%Error): The software will calculate a relative
percent error based on the certified standard value.
13. Complete the Equation Text Table.
A. Click in the Equation Text dialog box and then click the
arrow to select an equation or enter an equation following
the defined format. The software will list the apparent choice
first. Equations vary depending on the step names defined in
the Method Step Parameters dialog box. The equation text
field allows the operator to specify the equation that is to be
used to calculate the result associated with the specified
equation name. The software provides a number of
predefined equations based on the method steps defined in
the method. The following mathematical operators are
available are plus (+), minus (-), multiply (*), divide (/),
power (^) and others. All method step names, equation
names and associated data fields can be used as parameters
in the result calculation.
B. Click in Result Format and then select a result format from
the menu or enter the result format manually. Refer to
Number Formatting, page 4–50, for additional information.
TGA 701 Equations associated with Step Names, Moisture, Volatile, Ash
Moisture (([Initial Mass]-[Moisture Mass])/[Initial Mass])*100
Volatile (([Moisture Mass]-[Volatile Mass])/[Initial Mass])*100
Ash ([Ash Mass]/[Initial Mass])*100
Fixed Carbon 100-([Moisture]+[Volatile]+[Ash])
Volatile Dry [Volatile]*(100/(100-[Moisture]))
Ash Dry [Ash]*(100/(100-[Moisture]))
Fixed Carbon Dry 100-([Volatile Dry]+[Ash Dry])
Create a Standard
1. From the Configuration menu, click Standards. The Standards
dialog box will display. The Standards columns are explained in
the Glossary, page 12–1.
2. Click Add to insert a new row into the Standard Table. The name
will display as Standard, Standard-1, Standard-2, and so forth,
depending on the number of times the Add button is clicked.
3. Double-Click the standard name in the Name column to change if
desired.
4. Enter the certified and the uncertainty values for the standard in
the appropriate columns.
5. From the Check Std. menu, select Yes if the standard is a check
standard; select No if the standard is not a check standard.
6. Use the commands on the Standards dialog box to determine the
display of the Standards table as explained below.
Click Fields to modify the Fields that display in the Standards
spreadsheet. Refer to Modify Columns, page 4–17.
Click Delete to remove a standard from the spreadsheet. Refer to
Delete a Standard, page 5–16.
Click Ascending, and Standards will automatically display in
alphabetical and/or numerical order from lowest to highest.
Click Descending, and Standards will automatically display in
alphabetical and/or numerical order from highest to lowest.
Log In a Standard
1. From the Samples menu, click Login or click the F3 Login button
on the toolbar. The Sample dialog box will display.
12. If there are numerous standards to log in, click the Keep Logging
in Samples until Cancel is pressed check box. This allows the
operator to continue to log in standards until the Cancel button is
clicked.
13. Click OK when finished.
Analyze a Standard
1. From the Samples menu, click Analyze or click the Analyze button
on the toolbar. The Login Standard Sample dialog box will be
displayed if no standards have been logged in. Refer to Log In a
Standard, page 5–15, for additional information.
2. If not open already, the furnace lid will open. Place all the empty
crucibles (including reference crucible) in the carousel.
3. Close the furnace lid by pressing the Actuator button on the front
panel. Refer to Figure 5-1 page 5–3.
4. The system will weigh the empty crucibles.
5. The furnace lid will open and the carousel will rotate to the first
crucible then lower to the weigh position.
6. Place the standard in the crucible.
Delete a Standard
1. From the Configuration menu, click Standards.
2. Select the standard to delete.
Print a Standard
1. From the Configuration menu, click Standards.
2. Select the standard to be printed.
3. Click Print. The Print dialog box will display.
4. Change the print settings as needed then click OK. The standard
is printed on the system printer.
5. Click OK to close the Standards dialog box.
Toolbar
Method
Click the arrow and then select the method to use for calibration. Refer
to Methods, page 5–5, for additional information.
Plot
Click the arrow to select the type of plot. Select Calibration to plot the
certified concentration versus the measured concentration of the
sample. Select Relative Error to plot the percent relative error versus
the certified concentration of the sample. Relative Error=100*[(xi-
xt)/xt] where xi is the Calculated Value and xt is the Certified Value.
Select Residual Error to plot the residual error versus the certified
concentration of the sample. Residual Error=Calculated-Certified
Standards
Click Standards to access the Standards Table. Refer to Standards,
page 5–14, for additional information.
Print
1. Open the desired calibration by clicking Configuration and then
selecting calibration procedure.
2. To print the calibration, click the Print button. The Select
Calibrations to Print dialog box will display.
Format Pane
The format pane will be active in New Calibration, Add Standard, and
Edit Calibration.
1. For Step, click the arrow to select the step that will be associated
with the calibration curve. The calculated calibration will be
applied to this step. A different calibration can be calculated and
applied to each step.
2. For Curve Type, select the type of calibration curve used in the
plot, page 5–22. Select a curve that intersects each calibration
result.
3. For Weighting, click the arrow to determine which calibration
results, on the calibration curve, get priority. Select Normal, and
each point gets equal priority. Select Manual, and a manual
weighting can be entered. Select 1/Certified for a weighting factor
that can be applied to the calibration data points. This will
counteract a bias for higher concentrations that develops
naturally from the curve fitting process.
4. Use Existing Calibration to view the current calibration equation.
5. Use Date and time to determine when the current calibration was
established.
6. Click in the Show check box to view the last calibration. The last
calibration will display on the graph in blue.
7. Use New Calibration to display the new calibration equation.
8. Use RMS Error to determine which calibration curve has a better
fit. The lower the RMS error the better the fit.
9. Click in the Manual check box to enter a new calibration equation.
This can be done if a calibration was lost and a previous
calibration recorded.
Calibration Spreadsheet
The spreadsheet displays the selected samples used for calibration.
Refer to the Glossary, page 12–1, for further information about the
Calibration Spreadsheet columns.
Calibration Results
The results display shows data when multiple samples are selected
from the spreadsheet. The statistics are calculated from the Error%
and Certified columns for the step selected in the Step box in the
Format Pane. These results are most useful for determining whether or
not a particular repetition for a standard should be included or
excluded. The results display can also be used to quickly determine the
minimum and maximum certified value used to create the calibration.
Refer to Calibration Results Columns, page 5–23, for further
information.
Calibration Procedures
View Calibration
View calibration permits the operator to view the current calibration.
The calibration cannot be changed from this screen.
1. From the Configuration menu, click Calibration, and then click
View Calibration.
NOTE Refer to Calibration Main Window, page 5–18, for additional
information. The calibration spreadsheet pane, page 5–22, will be in
red.
Edit Calibration
Edit calibration permits the operator to revise the current calibration.
1. From the Configuration menu, click Calibration, and then click edit
calibration.
NOTE Refer to Calibration Main Window, page 5–18, for additional
information. The calibration spreadsheet pane, page 5–22, will be in
red.
2. Select the analysis step, the curve type, and weighting to change.
The calibration curve should go through, or come close to every
calibration point. If not, select another curve type.
3. To change the known concentrations and uncertainty values of
the reference materials, click the Standards button. The
Standards dialog box will display. Refer to Standards, page 5–14,
for additional information.
4. When finished, click OK to save the calibration changes. Refer to
Save Calibration, page 5–24, for additional information.
Save Calibration
1. When the calibration updates have been completed, click the OK
button. The Save Calibration dialog box will display. The Save
Calibration dialog box will only be displayed if this is a new
calibration or if any changes to the calibration procedure were
made.
Add Standards
Add standards permits the operator to add reference materials to the
list of standards used for a previously defined calibration. Refer to
Standards, page 5–14, for additional information.
1. Select the analyzed standards from the spreadsheet. These will
be added to the list of standards used for calibration.
2. Click the Standards button. The Standards dialog box will display.
All samples with the same sample batch number will be analyzed
together. A maximum of 19 samples can be analyzed per batch.
1. Clean and dry sample scoop. The sample scoop available must be
cleaned and dried between each analysis. Refer to Options, page
1–19, for further information.
2. From the Samples menu, click Analyze or click the F5 Analyze
button on the toolbar. The Login Sample dialog box will be
displayed if no samples have been logged in. Refer to Log In a
Sample, page 5–26, for additional information.
3. If not already open, the furnace lid will open.
4. Obtain crucibles from dessicator.
5. Place the reference crucible in the zero position, which is marked
by a small 1/8-inch diameter hole.
7. Close the furnace lid by pressing the Actuator button on the front
panel. Refer to Figure 5-1, page 5–3.
8. The system will weigh the empty crucibles.
11. Press the Actuator button on the front panel to rotate the
carousel to the next crucible.
12. Repeat steps 6, page 5–28, and 7, page 5–28. Continue this
procedure until sample has been added to all of the crucibles.
13. After the sample has been added to the last crucible, press the
Actuator button on the front panel. The analysis will start
automatically.
14. After the analysis completes, clean the crucibles. Refer to Clean
Crucibles, page 6–8.
1. Verify the Crucible Mass column and Initial Mass column display
in the spreadsheet. Refer to Modify Columns, page 4–17. The
Crucible Mass column displays the weight of the crucible. The
Initial Mass column displays the weight of the sample.
2. Log in samples. Refer to Log In a Sample, page 5–26.
3. From the Configuration menu, click System. Use the System tab
to determine whether the crucible mass weight and initial mass
weight will be entered manually or by using the external balance.
Refer to Enter Weights Manually, below, Enter Weights Using
External Balance, page 5–31, or Enter Weights Using External
Balance (Weigh into Crucibles), page 5–31.
7. Close the furnace lid by pressing the actuator button on the front
panel. Refer to Figure 5-1, page 5-3. After the actuator button is
pressed, analysis will begin.
Add Filters
Use the Add command to create a new filter.
1. To add a new filter, click Add. "New Filter" will display in the
Filters dialog box. Each time Add is clicked "New Filter" will
increment by one and display as "New Filter 1," "New Filter 2,"
etc.
2. Click Properties to configure the filter. Refer to Properties, page
5–35.
3. In the Filter Name entry box, type in the name for the filter.
4. For Export Format, click the arrow to select the export format for
this filter.
5. For Transmit Format, click the arrow to select the transmit format
for this filter.
6. Add a parameter, delete a parameter, or modify columns for the
sample filter.
A. To add a parameter,
1) In the Field entry box, click the arrow and select the
desired field.
2) In the Compare entry box, click the arrow and select the
desired comparator. The comparator selections change
depending on the selected field.
3) In the With entry box, enter the desired parameter
depending on the selected field. For example, if Analysis
Date is selected for Field and On is selected for
Compare, in the With column, enter the desired date for
the analysis results.
4) Click Add to insert a new row.
5) In the And/Or entry box, click the arrow and make
desired selection.
6) Repeat steps 1) - step 5), above, until all parameters for
the filter have been determined.
Delete Filters
Use the Delete command to remove sample filters from the Filters
dialog box.
1. Click on the desired sample to remove.
2. Click Delete.
Predefined Filters
Filters that may be frequently used have been predefined in the
software. Default filters are All Samples and Today's Samples. The
remaining filters that display are based on created methods. To select
a predefined filter, click on the name of the filter, and then click OK.
In the Sample Filter dialog box, the Filter Name displays the name of
the predefined filter. The Description explains the filter and whether or
not the fields or automatic updates can be modified. For example, for
All Samples, only the Fields can be modified. To modify the fields, click
the Fields button. Refer to Modify Columns, page 4–17. To determine
the automatic update, refer to Automatic Update, page 5–36.
To display a Field that does not appear in the report, select the
field from Invisible Fields, and then click Show. The Field will
automatically move to the Visible Fields column.
To remove a Field so that it does not display in the report, select
the field from Visible Fields and then click Hide. The Field will
automatically move to the Invisible Fields column.
1. From the spreadsheet, select the desired samples that have been
analyzed to print. When several samples are selected, the
software will group the samples by sample name.
2. Click Samples and select Print. Click Print to access the Report
Settings dialog box.
5. Use the Page Setup dialog box to change the size, orientation,
and margins of the paper before printing.
6. Click OK. The Print screen will display.
7. Click OK to print.
Use Text Import Data to bring sample data from another software
application into the database. Sample login information, such as data
from an Excel spreadsheet, can be imported using the Import function.
1. From the Sample menu, click Text Import Data. The Open dialog
box will display.
Export Data permits the operator to select sample data and data
transmission parameters and transfer them to another storage location
such as a folder on the hard drive or a network folder.
1. Select desired sample or samples that have been analyzed from
the spreadsheet.
2. If an export format has not been created, refer to Create an
Export Format, page 4–34. If an export format has been created,
refer to step 3, below.
3. Click Samples and then select Text Export Data. The Select
Export dialog box will display.
The Text Export Buffers feature on the Samples menu allows the user
to export raw data buffers to another software application. The sample
data buffers contain the sample weight loss and furnace temperature
data for the analysis. The time of analysis is also indicated for each
buffer. Select samples from the spreadsheet and then select Text
Export Buffers to export the data buffers associated with the selected
samples to a file. The exported data buffer file can then be plotted
using a software program such as Windows Excel®.
1. Click and drag the mouse to select the desired sample rows from
the spreadsheet.
2. From the Samples menu, select Text Export Buffers. The Save As
dialog box will display.
Export
1. Select sample.
2. From the Samples menu, click Transfer Export.
3. In the Export dialog box, enter a File Name for the sample
transfer and then click Save to save the sample transfer in the
current location or select a new location by clicking the arrow for
Save in, select a new location, and click Save.
Backup Data
1. In the spreadsheet, select the data to back up.
2. Click the Configuration menu, select Backup/Restore and then
select Backup.
3. When the Save As dialog box displays. Enter the file name for the
data next to File name.
4. Select the location for the file from the arrow next to Save in.
5. Click Save.
6. Click OK when the confirmation dialog box displays.
3. Click Open.
4. When the Restore Hardware Parameters dialog box displays, click
Yes to overwrite the current hardware calibration parameters;
click No, and hardware calibration parameters will not be
restored.
Schedule
Below is a list of maintenance procedures that should be performed on
a regular basis. Periodic maintenance of the instrument will result in
improved instrument performance and will extend the life of the
instrument.
Refer to Configure Periodic Maintenance Counters, below, to set up a
counter for each maintenance item.
Frequency Area Cleaning or Replacement
Clean after
Consumable. Refer to Clean Crucibles, page 6–8.
every Crucible
Replace if damaged.
analysis.
Electronic
On the back of the analyzer. Refer to Electronic Chassis
Chassis
Fan Filter, page 6–6.
Fan Filter
As Needed Furnace Inside of the analyzer. Refer to Clean Furnace, page 6–7
The fan filters should be cleaned or replaced when they are visibly
dirty. Depending on the environment that the analyzer is in, the time
period for cleaning may vary considerably. Inspect the fan filters
weekly until a cleaning period can be determined.
2. Raise the furnace lid. From the toolbar, click the F7 Cover button.
The crucibles must be clean, dry, and free from static. Washing and
rinsing crucibles in water removes any static charge. It is important to
clean every crucible after every analysis because they will contain
residual sample material.
Illustrations .................................................................................7–2
Theory of Operation......................................................................7–3
+ - +-
99.5% Air SW9
(dry, oil-free)
25 no
45 psi psi 621-034 SV# 2
Poly
nc c
nc c - Lid Position
Rear panel 10-32 UNF, 10 Required
SV# 4
nc c - Weigh Position 1
773-661 Fitting, 10 Required
no c - Weigh Position 2
SV# 5
SV# 6 Other Inputs and Outputs
nc c - Rotate Carousel 1 SV# 14 Green LED
SV# Cooling Fan
no c - Rotate Carousel 2 SV# Alarm buzzer
Symbol LECO P/N Description SV# 7
SW# 0 Lid Position
SV# 8 620-625-101, 5 Required
nc c - Furnace Cover 1 SW# 1 Turn Position
618-758 Proportional Valve
SW# 2 Weigh Position
3-way valve with speed no
620-625-101 c - Furnace Cover 2 SW# 3 Carousel Advanced
adjustment SV# 9
616-567-106 2-way valve SV# 10 SW# 4 Carousel Return
nc c - Furnace Cover Height
SW# 5 Furnace Cover
?psig 621-036 Pressure Transducer
SV# 11 SW# 6 Home Position
25 25 PSI Pressure Switch, nc c - Crucible Actuator 1
618-122 SW# 7 Air Pressure
psi Normally Open
SW# 8 Nitrogen Pressure
616-567-108 3-way valve no c - Crucible Actuator 2
SW# 9 Oxygen Pressure
SV# 12
621-034 Pressure regulator, 35 PSI SW# 10 Front Panel Control
Figure 7-1
Flow Diagram
Ambients ....................................................................................8–3
Ambient Acceptable Range Table ...............................................8–3
Customize Ambient Monitor ......................................................8–4
Ambient Plot Menu Commands .............................................8–4
Customize Graph Lines........................................................8–5
Customize Ambient Column Display ......................................8–6
Solenoids and Switches .................................................................8–8
Network......................................................................................8–9
Communications ........................................................................ 8–11
Systems Check .......................................................................... 8–13
NOTE An electronic assembly that does not have the application firmware
programmed will still sign on to the network as long as it contains the
network software program. In this case, the Serial, Version, and
Checksum fields will display zeros.
2. Verify the Network Power is on. The power is on when the check
is displayed and off when the check is not displayed. It is only
necessary to turn the network power off if the network hardware
is being serviced.
Indicator
Status
Checkbox
Yellow light This item is currently being checked.
Green light This item has been checked and has passed.
This item has been checked and has failed the
Red light
system check.
3. Click Print to get a hard copy of the results from the systems
check procedure.
A. Click the Print button. The Print dialog box will display.
B. Click OK.
4. Click Close when finished.
5. Enter 3.5 as the value for the first set point. Click Next. The Wait
dialog box is displayed while the software sets the system flow to
this value.
7. Enter the actual flow rate (from the flow meter) and click Next.
The Enter The Set Point dialog box will display.
8. Enter 10.0 as the value for the second set point. Click Next. The
Wait dialog box is displayed while the software sets the system
flow to this value.
9. Click Next. The Enter the Actual Value dialog box will display.
10. Enter the actual flow rate (from the flow meter) and click Next.
11. After the flow has been calibrated, the Calibration Finished dialog
box will display. Click Finish.
12. Remove the external flow meter from the pneumatic flow
manifold.
A. Disconnect the external flow meter from the furnace lance
port of the pneumatic manifold. Refer to Figure 10-17, page
10–13.
B. Reconnect the stainless steel tubing to the furnace lance port
of the pneumatic manifold. Refer to Figure 10-17, page
10–13.
C. Reinstall the left instrument panel. Refer to Figure 10-25, page
10–22.
D. Reconnect J401, J402, and J403 to the push button mounted on
the front of the casting. Refer to Figure 10-25, page 10–22.
E. Reinstall the TGA701 front panel. Refer to Figure 10-25, page
10–22.
Balance Calibration
1. Remove the carousel. Refer to Figure 10-3, page 10–5.
2. From the Configuration menu, click Hardware Calibration. The
Hardware Calibration dialog box will display.
4. Place the draft shield over the balance pedestal. When prompted,
click Next.
5. Remove the draft shield from the balance pedestal and place the
100-gram weight on the balance.
6. Cover the weight with the draft shield. Click the Next button when
available.
7. Wait for the balance calibration to finish then Click Next.
8. After the message is displayed stating that the balance has been
calibrated, click Next.
9. Click Done when finished.
10. Remove the draft shield and 100 gram weight.
11. Reinstall the carousel. Refer to Figure 10-3, page 10–5.
5. Using the (2) adjusting screws, level the balance so that the level
bubble is indicating that the balance is centered.
6. Position the pedestal.
A. Loosen the screws securing the balance guide block. Refer to
Figure 10-16, page 10–12.
B. Move the balance guide block and the balance until the
pedestal is positioned in the center of the hole in the ceramic
shield in the lower furnace.
C. Tighten the balance guide block screws. Refer to
Figure 10-16, page 10–12.
7. Lock down the adjusting screws located in front of the balance.
Refer to Figure 10-16, page 10–12.
A. Tighten the jam nuts against the balance base.
B. Tighten the lockdown screws against the balance adjusting
screws.
The Balance Test checks the analytical performance and stability of the
balance within the instrument.
1. Verify system settings as follows: Weight range must be between
the minimum and the maximum settings. Batch mode has been
deactivated.
2. Create the Balance Test method. Refer to Create or Edit a
Method, page 5–5, for additional information.
A. From the Configuration menu, click Methods.
B. Click the Add button.
C. Enter Balance Test for the method name.
D. Click OK. The new method will be added.
E. Click the Properties button.
F. Enter the following method general parameters:
Crucible Type Ceramic
Crucible Density 3
Cover Density 3
Sample Type -
Sample Density 1.5
Installation
Carousel Installation
1. Press the actuator button on the front of the instrument to open
the furnace. Refer to Figure 5-1, page 5–3.
2. Place the spindle drive through the center hold in the bottom of
the lower oven and into the pneumatic actuator coupling. Do not
tighten the coupling at this time.
3. Place the carousel on the spindle drive. Line up the slot on the
spindle drive with the slot on the carousel.
Adjustments
Home Position
1. Remove the front panel. Refer to Figure 10-25, page 10–22.
A. Remove the two Allen screws on the top of the front panel.
B. Loosen the two Allen screws on the bottom of the front
panel.
C. Carefully insert a flat tip screwdriver in between the front
panel and the right side panel and twist. This will allow a gap
to form.
Speed Adjustments
Transition Speed
The carousel transition speed adjustments vary the up and down
speed. They do not vary the rotational speed.
1. Adjust the carousel transition speed from the weigh position to
the turn position using the adjuster located at the top of the
indexer cylinder. The carousel should rise relatively slowly and
smoothly.
A. From the Diagnostics menu, click Solenoids.
B. Move the carousel from the weigh position to the turn
position.
1) Click to enable the Weigh Position solenoid.
2) Click to disable the Weigh Position solenoid.
2. Adjust the carousel transition speed from the turn position to the
weigh position using the adjuster located at the middle of the
indexer cylinder. Refer to Figure 10-10, page 10–9. The carousel
should rise relatively slowly and smoothly.
A. From the Diagnostics menu, click Solenoids. The
Solenoids/Switches dialog box will display.
B. Move the carousel from the turn position to the weigh
position.
1) Click to disable the Weigh Position solenoid.
2) Click to enable the Weigh Position solenoid.
3. Click Close when finished.
Rotate Speed
1. Set the carousel to rotate continuously.
A. From the Diagnostics menu, click Solenoids.
B. From the Life Test menu, select Rotate Carousel.
C. Click the Start Life Test button.
2. Adjust the carousel rotate speed using the right and left
pneumatic adjusters. Refer to Figure 10-10, page 10–9.
A. The left pneumatic adjuster cushions the cylinder return.
B. The right pneumatic adjuster adjusts the speed of the
carousel advance.
3. Lock the adjuster screws by tightening the jam nut against the
pneumatic adjuster housing.
4. Stop the carousel rotation by clicking the Stop Life Test button.
5. Click Close when finished.
Seal Adjustment
Adjustment
Element Temperature
1. Loosen the clamp to the rear thermocouple.
2. Position the thermocouple so that it is as straight as possible.
A. The distance from the inside bottom of the furnace to the
top of the thermocouple is 1.5 inches.
B. The distance between the wall of the furnace and the
thermocouple ceramic is 0.125 inches.
3. Tighten the clamp to the rear thermocouple.
Replacement
The thermocouple replacement procedure should be used to replace
the furnace thermocouples in the event of a failure.
3. Remove the analyzer right and left side panels by sliding them
forward.
4. Remove the four screws securing the electronics module to the
rear panel.
5. Move the electronics module out far enough to expose connectors
J394D and J394F.
6. Disconnect the thermocouple lead wires from the J394 connector.
(The J394 connector depends on which thermocouple is
replaced.)
J394F is the furnace temperature thermocouple.
J394D is the element temperature thermocouple.
7. Remove the screw securing the clamp holding the thermocouple.
8. Install a new thermocouple into the thermocouple clamp and
tighten the clamp. Make sure the thermocouple is positioned
correctly. Refer to Figure 10-5, page 10–6.
9. Connect the thermocouple lead wires to the appropriate J394
connector. The J394 connector depends on which thermocouple is
replaced.
J394F is the furnace temperature thermocouple.
J394D is the element temperature thermocouple.
Modem Installation
1. Turn the analyzer and computer off.
2. Unpack the Smartline package. Refer to Component and
Accessories List, page 1–20, for package contents.
3. Using the modem cable, connect the modem to COM2 Port on the
rear of the computer. Refer to Figure 2-3, page 2–21.
4. Connect the power cube, supplied with the modem, to the power
input jack on the modem.
5. Connect the modem to the same outlet as the computer.
6. Turn the modem, computer, and analyzer on.
7. Install the modem driver. Manual installation from the Modem
Installation CD-ROM.
A. From the desktop, click the My Computer icon. The My
Computer window will display.
Create a Shortcut
A shortcut should be created on the computer desktop for easy access
to LECO.
1. Close or minimize all open programs.
2. Right-click on the desktop. A menu will display.
3. Click New. A second menu will display.
3. In the left corner of the window, there is the Wait For Call play
arrow. Click on the Wait For Call play arrow. The status bar will
display Waiting For Guest To Reply.
4. Contact LECO Service department (269-982-5497) to confirm
Smartline is working properly or wait for the LECO Service
department to initiate further testing.
Troubleshooting Table
Area or Assembly Symptom and Checks
Instability
Balance
Refer to Balance Instability, page 4–30.
Balance pedestal is touching the shield or the bottom of
the furnace
Crucibles vibration
Poor gas supplies and tubing may not provide sufficient flow to
maintain the input pressures required. Always make sure that
Flow
the customer uses high quality gas regulators and that the
tubing length and size does not restrict the gas flow, thereby
causing a pressure drop at the manifold inlet when high lance
flows are used.
Inspect the alignment and the seal between the upper and
lower furnaces. Realign or adjust as necessary.
Leaks around the pedestal shield
Inspect the o-ring seal with the balance assembly and the
lower furnace. Adjust as necessary.
Leaks inside the balance assembly
Troubleshooting Procedures
Autoexec.bat File
1. Exit the software to the DOS c:\prompt.
2. The >type autoexec.bat and press the Enter key.
3. The following should be displayed:
C:\DOS\SMARTDRV\EXE/X
@ECHO OFF
PROMP $p$g
PATH C:\DOS;C:\WINDOWS;C:\TGA701
SET TEMP=C:\DOS
Config.sys
1. Exit the software to the DOS c:\prompt.
2. The >type Config.sys and press the Enter key.
3. The following should be displayed:
DEVICE=C:\DOS\SEVER
DEVICE=C:\DOS\HIMEM.SYS
DEVICE=C:\DOS\EMM386.EXE X=D000-DFFF
DOS=HIGH
FILES=30
STACKS=9256
Figure 10-4
Pedestal Location in Furnace
Figure 10-6
Top Furnace Cover—Internal View
Figure 10-8
Furnace Cover—Rear View
Figure 10-12
Indexer—Rear View
Figure 10-14
Over Temperature Switch—Location
Figure 10-16
Balance Assembly
NC NC NC NC NC NO NC NO NC NO
NITROGEN INLET
FITTING SV0 SV1 SV2 SV3 SV4 SV5 SV6 SV7 SV8 SV9
(618-657)
PROPORTIONAL VALVE
(618-758-010)
Figure 10-17
Flow Manifold Assembly
Pressure Transducer
P1 ............................................ MPX4250D..................................................... 621-036
Proportional Valve
PV1........................................... 100 PSID....................................................... 618-758
Relay
K1 ............................................ SSR, Zero Cross ............................................. 769-335
K3 ............................................ Safety Relay, 25A ........................................... 607-157
Solenoid Valves
SV0 .......................................... Oxygen Supply, 2W ........................................ 616-567-106
SV1 .......................................... Nitrogen Supply, 2W ....................................... 616-567-106
SV2 .......................................... Air Supply, 2W ............................................... 616-567-106
SV3 .......................................... Lid Position, 3W.............................................. 616-567-108
SV4 .......................................... Weigh Position, 3W ......................................... 616-567-108
A
Accessories, and Components ................................................. 1–20
Actuator Button.......................................................................5–3
Ambient Monitor......................................................................8–3
Analysis, Overview ..................................................................5–4
Analyze a Sample.................................................................. 5–28
Analyze a Standard ............................................................... 5–16
Archive Log .......................................................................... 4–59
Atmosphere............................................................................5–9
Audible Sounds ..................................................................... 4–33
Alarm .................................................................. 4–27, 4–33
Auto Assign Crucible Location.................................................. 4–27
Auto Find Crucible Location..................................................... 4–27
Automatically Transmit Analysis .............................................. 4–43
B
Backup Data ......................................................................... 5–47
Balance
Adjustment ..................................................................... 9–10
Configure ........................................................................ 4–28
External, Optional - Refer to External Balance
Instability........................................................................ 4–30
Installation ...................................................................... 2–15
Poor Repeatability ............................................................ 9–34
Resolution – See Specifications........................................... 1–22
Test................................................................................ 9–13
Unstable Readings ............................................................ 9–34
View Balance – See View Menu.............................................4–9
Barometric Pressure .............................................................. 4–26
Batch Number, Change .......................................................... 4–15
Blower, External......................................................................3–3
Blower Maintenance............................................................3–4
Blower Filter Maintenance ....................................................3–4
C
Calibration ........................................................................... 5–18
Add Standard................................................................... 5–25
Edit Calibration ................................................................ 5–24
Format Pane .................................................................... 5–21
Manual Calibration – See Format Pane................................. 5–21
Manual Weighting – See Format Pane.................................. 5–21
New Calibration................................................................ 5–23
Plot ................................................................................ 5–22
Print ............................................................................... 5–20
Results Display ................................................................ 5–22
Save Calibration ............................................................... 5–24
Spreadsheet Pane............................................................. 5–22
View Calibration ............................................................... 5–24
Calibration Menu Bar ............................................................. 5–19
Carousel
Adjust............................................................................. 9–16
Advance Carousel Switch ................................................... 9–16
Alignment with Pedestal .................................................... 9–35
Carousel Return Switch ..................................................... 9–16
Home Position.................................................................. 9–17
Install ............................................................................. 2–15
Lid Position Switch............................................................ 9–16
Poor Flow Calibration ........................................................ 9–36
Position........................................................................... 4–27
Positioning Errors ............................................................. 9–35
Speed Too Fast ................................................................ 9–37
Turn Position Switch ......................................................... 9–15
View Carousel .................................................................. 4–25
Weigh Position Switch ....................................................... 9–15
Caution Symbols ................................................................... 1–13
Chassis, Electronic Fan Filter .....................................................6–6
Columns in Spreadsheet ......................................................... 4–17
Communications ................................................................... 8–11
Comparator .......................................................................... 5–10
Components ......................................................................... 1–20
Computer
Autoexec.bat File.............................................................. 9–40
D
Data, Backup and Restore ...................................................... 5–47
Data Transmit
Create and Modify Transmission Formats ............................. 4–40
Transmit Sample Data....................................................... 5–49
Database
Compact/Repair ............................................................... 4–68
Delete............................................................................. 4–69
Log Archive ..................................................................... 4–59
Management.................................................................... 4–60
Menu Functions ................................................................ 4–56
Select ............................................................................. 4–67
Declaration of Conformity ....................................................... 1–15
Diagnostics.............................................................................8–1
Diagnostics Menu .................................................................. 4–12
Dimensions – See Specifications .............................................. 1–22
Dual Instrument
Installation ...................................................................... 2–18
Instruments Dialog Box ..................................................... 2–19
E
Edit Menu...............................................................................4–9
F
Facility Requirements............................................................. 2–20
Filters
Furnace Fan Filter...............................................................6–6
Electronic Chassis Fan Filter .................................................6–6
Filters for Sample Results - Refer to Sample Filters .................... 5–34
Final Weight ......................................................................... 5–10
Flow, Poor Flow Calibration ..................................................... 9–36
Flow Rate ...............................................................................5–9
Furnace
Adjust Cover Speed .......................................................... 9–19
Calibration Problems ......................................................... 9–37
Clean................................................................................6–7
Cover Position.................................................................. 4–27
Cover Speed Too Fast ....................................................... 9–37
Excessive Smoke.............................................................. 9–37
Fan Filter ..........................................................................6–6
Overtemp........................................................................ 4–27
Power On/Off ................................................................... 9–37
G
Gas
Connections.......................................................................2–4
Flow Rate – See Specifications ........................................... 1–22
Purity – See Specifications ................................................. 1–22
H
Hardware Calibration ...............................................................9–5
Help Menu Commands ........................................................... 4–12
History, View ........................................................................ 4–58
Hold Time ..............................................................................5–8
Humidity – See Specifications ................................................. 1–22
I
Import Data ......................................................................... 5–42
Installation
Balance, Pedestal, Carousel ............................................... 2–15
Dual Instrument............................................................... 2–18
External Blower..................................................................3–3
TGA701 Analyzer................................................................2–3
Interference to Other Devices ................................................. 1–14
International Installation ..........................................................2–3
K
Key – See Copy Protection Key................................................ 2–11
L
Lid Position Switch................................................................. 9–16
Log Archive .......................................................................... 4–59
Log File, View .........................................................................6–9
M
Main Window, Software ............................................................4–4
Maintenance ...........................................................................6–1
External Blower..................................................................3–4
Periodic Schedule ...............................................................6–3
Reset Counters ..................................................................6–4
Maintenance Menu................................................................. 4–12
Max Time ...............................................................................5–9
N
Network .................................................................................8–9
Notice, Warranty ................................................................... 1–11
Number Formatting ............................................................... 4–50
O
Options................................................................................ 1–19
Overtemp............................................................................. 4–27
Oven Temperature – See Specifications .................................... 1–22
Over Temperature Switch Reset .............................................. 9–24
P
Pedestal............................................................................... 2–15
Pneumatic Actuator, Adjust..................................................... 9–16
Power On Instrument............................................................. 2–13
Presentation Time ................................................................. 4–56
Preweighed Crucibles ............................................................. 5–30
R
Ramp Rate .............................................................................5–8
Ramp Time.............................................................................5–8
Restore Data ........................................................................ 5–48
Results, Recalculate Results .................................................... 4–10
Result Format ....................................................................... 5–11
Result Units .......................................................................... 5–12
S
Sample
Analyze........................................................................... 5–28
Delete............................................................................. 4–14
Exclude........................................................................... 5–19
Include ........................................................................... 5–19
U
USB Driver, Reinstall.............................................................. 9–36
User
Add or Change User Permission for Instrument ..................... 4–53
Assign User to Windows .................................................... 4–53
User Defined Fields ................................................................ 4–48
V
Ventilation..............................................................................2–5
View History ......................................................................... 4–58
View Menu .............................................................................4–9
W
Warning Symbols .................................................................. 1–13
Warranty.............................................................................. 1–11
Weigh Position Switch ............................................................ 9–15
Window .................................................................................5–9